K Number
K231420

Validate with FDA (Live)

Device Name
Aura 10 PET/CT
Manufacturer
Date Cleared
2023-08-10

(86 days)

Product Code
Regulation Number
892.1200
Age Range
0 - 21
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AURA 10 PET/CT system is a cabinet diagnostic imaging device that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The AURA 10 PET/CT system images harvested specimens from various anatomical regions in order to provide rapid verfication that the correct tissue has been excised during the surgical procedure. Images can be obtained as CT only, or a combination of both by surgeon's discretion.

The AURA 10 cabinet PET/CT system provides an image of extent and degree of intensity radiopharmaceutical uptake in the specimen by PET and the anatomical information by CT, which will help the surgeon with further patient management.

The AURA 10 PET/CT system is not validated for margin detection.

Device Description

The AURA 10 PET/CT is a mobile, vertical-bore PET/CT system with a Field of View (FOV) suitable for small pathology specimens. It is intended to be used in both the operating room (OR) as well as the pathology department to image pathology specimens from various anatomical regions in order to provide rapid pathology imaging. The AURA 10 PET/CT is intended to image pathology specimens for a wide range of patient types, sizes, and extent of diseases. It is designed as a mobile cart so that it is easily portable by one person and can be moved to different surgical suites or between departments as needed.

AI/ML Overview

The provided text describes the regulatory clearance for the AURA 10 PET/CT system, which is a cabinet diagnostic imaging device for imaging harvested specimens. However, it explicitly states, "Not applicable. Clinical studies are not necessary to establish the substantial equivalence of this device." This means that the submission does not include information about clinical performance data or a study proving the device meets acceptance criteria derived from clinical studies.

The document focuses on demonstrating substantial equivalence to a predicate device (Vereos PET/CT) and a reference device (Faxitron VisionCT) through non-clinical performance data, primarily bench testing, electrical safety, EMC, and software verification/validation.

Therefore, it is not possible to provide the requested information regarding acceptance criteria and a study that proves the device meets those criteria from a clinical performance perspective. The clearance is based on technological characteristics and non-clinical testing demonstrating equivalency to existing cleared devices, not on a new clinical performance study.

Here's an explanation based on the provided text's limitations:

1. A table of acceptance criteria and the reported device performance:

  • Acceptance Criteria: No specific clinical acceptance criteria (e.g., sensitivity, specificity, accuracy against a clinical ground truth) are provided in the document. The acceptance for this device is based on meeting safety and performance standards through non-clinical testing and demonstrating technological equivalence to predicate devices.
  • Reported Device Performance (Non-Clinical):
    • Electrical Safety & EMC: "The device passed all tests" for IEC 61010-1, IEC 61010-2-091, IEC 61010-2-101, and IEC 61326-2-6.
    • Software Verification & Validation: "Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance... The software for this device was considered as a moderate level of concern." "The AURA 10 PET/CT was tested in accordance with IEC 62304:2006 + A1:2015...".
    • Bench Testing:
      • Analytical performance testing: "The AURA 10 PET/CT was tested in accordance with NEMA NU 4-2008 - Performance measurements of small animal positron emission tomographs (PETs)." (Specific performance numbers from this test, like spatial resolution or peak NEC rate, are listed in the "Technological characteristics" table, but it's not explicitly stated that these are the "acceptance criteria" for clearance, rather they are characteristics measured against the predicate.)
      • Usability testing: "The AURA 10 PET/CT was tested in accordance with IEC 62366-1:2015... Usability testing was conducted and documentation provided as recommended by FDA's Guidance...".

2. Sample size used for the test set and the data provenance:

  • Since no clinical studies were performed, there is no "test set" in the context of clinical data. The non-clinical tests involved hardware and software evaluations.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable, as no clinical ground truth was established by experts for a clinical test set.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable, as no clinical test set requiring adjudication was used.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No MRMC study was done, as explicitly stated, "Clinical studies are not necessary to establish the substantial equivalence of this device." This device is an imaging system, not an AI-assisted diagnostic tool for human readers in the traditional sense of improving reader performance. Its purpose is to provide images for verification.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • The device is a medical imaging system (PET/CT), not an algorithm that produces a diagnostic output on its own. Its "performance" is in image generation and quality, which was evaluated through bench testing.

7. The type of ground truth used:

  • For the non-clinical bench testing, the "ground truth" would be the known physical properties and performance characteristics of phantoms and test setups used in accordance with standards like NEMA NU 4-2008.

8. The sample size for the training set:

  • Not applicable, as this documentation does not describe an AI/ML model that would require a "training set" of data for its primary function. It's a hardware imaging system.

9. How the ground truth for the training set was established:

  • Not applicable.

In summary: The provided FDA 510(k) clearance letter and summary for the AURA 10 PET/CT system indicate that its clearance was based on demonstrating "substantial equivalence" to existing, legally marketed predicate devices through non-clinical testing (electrical safety, EMC, software V&V, and bench testing against industry standards like NEMA NU 4-2008) and a comparison of technological characteristics. No clinical performance studies, human reader studies, or AI algorithm performance studies were deemed necessary for this clearance.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

August 10, 2023

XEOS Medical % Bjorn Delbeecke Head of OARA and Legal Ottergemsesteenweg-Zuid 808 Bus 358 Gent, Oost-Vlaanderen 9000 BELGIUM

Re: K231420

Trade/Device Name: AURA 10 PET-CT Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: Class II Product Code: QXL, MWP Dated: May 5, 2023 Received: May 16, 2023

Dear Bjorn Delbeecke:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

D

Daniel M. Krainak, Ph.D. Assistant Director Magnetic Resonance and Nuclear Medicine Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K231420

Device Name AURA 10 PET/CT

Indications for Use (Describe)

The AURA 10 PET/CT system is a cabinet diagnostic imaging device that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The AURA 10 PET/CT system images harvested specimens from various anatomical regions in order to provide rapid verfication that the correct tissue has been excised during the surgical procedure. Images can be obtained as CT only, or a combination of both by surgeon's discretion.

The AURA 10 cabinet PET/CT system provides an image of extent and degree of intensity radiopharmaceutical uptake in the specimen by PET and the anatomical information by CT, which will help the surgeon with further patient management.

The AURA 10 PET/CT system is not validated for margin detection.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

K231420

This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR 807.92.

5.1. General information

Date Prepared:5 May 2023
Company:XEOS Medical NVOttergemsesteenweg Zuid 808 bus 3589000 GentBELGIUM
Primary ContactPerson:Bjorn DelbeeckeHead of QARA and Legal+32 (0)9 277 77 93Bjorn.delbeecke@xeos.care
Device:Trade Name: AURA 10 PET/CTCommon Name: Cabinet, Emission Tomography SystemCabinet, CT systemClassification Name (Regulation): Cabinet, Emission Tomography SystemCabinet, CT systemClassification Panel: RadiologyDevice Class Class IIProduct Code QXLSecondary product code MWP
Predicate Device:Trade Name: Vereos Pet/CtCompany: Philips Medical Systems Nederland B.V.510(k) Clearance: K210880Classification Name (Regulation): Emission computed tomography system (21 CFR 892.1200)Computed tomography x-ray system (21 CFR 892.1750)Classification Panel: RadiologyDevice Class: Class IIProduct Code: KPSSecondary Product Code: JAK
Reference Device:Trade Name: Faxitron VisionctCompany: Faxitron Bioptics Llc510(k) Clearance: K173309Classification Name (Regulation): Stationary x-ray system (21 CFR 892.1680)Classification Panel: RadiologyDevice Class: Class IIProduct Code: MWP
Device description:The AURA 10 PET/CT is a mobile, vertical-bore PET/CT system with a Field of View(FOV) suitable for small pathology specimens. It is intended to be used in both theoperating room (OR) as well as the pathology department to image pathologyspecimens from various anatomical regions in order to provide rapid pathologyimaging. The AURA 10 PET/CT is intended to image pathology specimens for a widerange of patient types, sizes, and extent of diseases. It is designed as a mobile cart sothat it is easily portable by one person and can be moved to different surgical suites orbetween departments as needed.
Intended Use /Indications for UseThe AURA 10 PET/CT system is a cabinet diagnostic imaging device that combinesPositron Emission Tomography (PET) and X-ray Computed Tomography (CT)

{4}------------------------------------------------

systems. The AURA 10 PET/CT system images harvested specimens from various anatomical regions in order to provide rapid verification that the correct tissue has been excised during the surgical procedure. Images can be obtained as CT only, PET only, or a combination of both by surgeon's discretion.

The AURA 10 cabinet PET/CT system provides an image of extent and degree of intensity radiopharmaceutical uptake in the specimen by PET and the anatomical information by CT, which will help the surgeon with further patient management.

The AURA 10 PET/CT system is not validated for margin detection.

Technological characteristics

The following table provides a detailed overview of the technological similarities and differences between the subject AURA 10 PET/CT, the predicate Vereos PET/CT and the reference device Faxitron Vision CT.

FeaturesSubject deviceAURA 10 PET/CTPredicate deviceVereos PET/CT(K210880)Reference deviceFaxitron Vision CT(K173309)
GENERAL
Size(s)162 x 52 x 65 cm³485 x 220 x 207 cm³ (L xW x H)180 x 70 x 65 cm³
Mass150 kg4220 kg140 kg
Battery or Mains poweredMainsMainsMains
Power requirementsSingle-phase110-250 VAC50/60 Hz1.2 kW3-phase200/208/240/380/400/415/480/500 VAC50/60 Hz150,000 VASingle-phase100 - 250 VAC50/60 Hz200 VA
MobilityMobileStationaryMobile
PET SUBSYSTEM
Scintillator Crystal MaterialLYSOLSO/LYSON/A
Photomultiplier technologySiliconPhotomultipliersSiliconPhotomultipliersN/A
FOV, Transaxial100 mm576 mmN/A
FOV, Axial60 mm164 mmN/A
Energy Window358 keV - 664 keV450 - 613 keVN/A
Spatial Resolution< 1.0 mm> 4.2 mmN/A
Peak NEC Rate38,395 kpcs at 26.6kBq/mL687 kcps @ 50kBq/mL222 kcps @ 5.3kBq/mLN/A
ToF ResolutionN/A310 psN/A
Scatter fraction16%34%N/A
Max Count rate ErrorN/A6.8%N/A
Energy resolution18%11%N/A
Sensitivity60 cps/kBq5.2 cps/kBqN/A
CT SUBSYSTEM
X-ray tube anode materialTungstenTungstenTungsten
X-ray tube kV settings50 kV80, 100, 120, 140 kV15 - 50 kV
X-ray tube current1.0 mA max (0.6 mAactual)20 – 665 mA1.0 mA max
X-ray tube power50 W max80,000 W max11.5 W max
Focal spot size0.050 mm typical0.5 x 1.0 mm² or 1.0 x1.0 mm²< 0.015 mm
Anode angle20 degrees7 degrees45 degrees
Focus-isocenter distance260 mm570 mm230 mm
Focus-detector distance350 mm1040 mm370 mm
Magnification1.35X1.82X1.61X
Transaxial Modulation TransferFunction (MTF)5 lp/mm (at cutoff)4 lp/mm (at 20%MTF)1.3 - 2.4 lp/mm (atcutoff)Unpublished
Axial MTF5 lp/mm (at cutoff)4 lp/mm (at 20%MTF)1.3 - 2.4 lp/mm (atcutoff) withappropriately thinslice sensitivity profileselectionUnpublished
Image matrix1024 x 1024Up to 1024 x 10241024 x 1024
Slice thickness0.1 mm0.5 - 12.5 mm (axialmode)0.1 mm
Number of slices62064, 128Up to 1024
Metal artifact reductionYesAvailable as optionYes
HOST AND NETWORKING
Network connectionYesYesYes
PRESCRIPTION / OVER THE COUNTER
Prescription / Over the counterPrescriptionPrescriptionPrescription

{5}------------------------------------------------

Summary of non-clinical performance data

Electrical Safety and Electromagnetic Compatibility (EMC)

The AURA 10 PET/CT was tested in accordance with IEC 61010-1 (2010 + A1 2016) Standard for Safety for Electrical Equipment for Measurement, Control and Laboratory Use – Part 1: General requirements, including US deviations. The device passed all tests.

The AURA 10 PET/CT was tested in accordance with IEC 61010-2-091:2012 Safety requirements for electrical equipment for measurement, control and laboratory use – Part 2-091: Particular requirements for cabinet X-ray systems. The device passed all tests. The device was also designed and developed in accordance to 21 CFR 1020.40 (Cabinet x-ray systems) and as recommended by FDA's Guidance for Industry and FDA Staff, "Compliance guide for cabinet X-ray systems."

The AURA 10 PET/CT was tested in accordance with IEC 61010-2-101:2018 2012 Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment.

The AURA 10 PET/CT was tested in accordance with IEC 61326-2-6 Electrical equipment for measurement, control and laboratory use – EMC requirements – Part 2-6: Particular requirements – In vitro diagnostic (IVD) medical equipment. The device passed all tests. The testing was conducted and documentation provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for electromagnetic compatibility (EMC) of medical devices."

Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a moderate level of concern.

The AURA 10 PET/CT was tested in accordance with IEC 62304:2006 + A1:2015: Medical device software -Software life-cycle processes (FDA Recognition Number: 13-79).

Bench Testing

The following bench testing was performed to demonstrate substantial equivalence:

  • Analytical performance testing. The AURA 10 PET/CT was tested in accordance with NEMA NU 4-2008 -Performance measurements of small animal positron emission tomographs (PETs).
  • -Usability testing. The AURA 10 PET/CT was tested in accordance with IEC 62366-1:2015: Medical devices - Part 1: Application of usability engineering to medical devices (FDA Recognition Number: 5-114). Usability testing was conducted and documentation provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance on Applying Human Factors and Usability Engineering to Medical Devices".

Information pursuant to the following FDA Guidance Documents was also included:

  • -Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
  • -Content of Premarket Submissions for Management of Cybersecurity in Medical Devices

{6}------------------------------------------------

  • -Recommended Content and Format of Non-Clinical Bench Performance Testing Information in Premarket Submissions
  • -Technical Performance Assessment of Quantitative Imaging in Radiological Device Premarket Submissions
  • -Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices
  • Guidance for Radio Frequency Wireless Technology in Medical Devices -
  • Guidance for Off-The-Shelf Software Use in Medical Devices -
  • Guidance on General Principles of Software Validation -
  • -Guidance for the Submission of Premarket Notifications for Emission Computed Tomography Devices and Accessories (SPECT and PET) and Nuclear Tomography Systems.

Summary of clinical performance data

Not applicable. Clinical studies are not necessary to establish the substantial equivalence of this device.

Final Conclusion

This summary includes information on the technological characteristics, performance data as well as verification and validation activities demonstrating that the AURA 10 PET/CT is as safe and effective as the predicate and does not raise different questions of safety and effectiveness, and therefore is substantially equivalent to the predicate device.

§ 892.1200 Emission computed tomography system.

(a)
Identification. An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.(b)
Classification. Class II.