K Number
K231173

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-07-21

(87 days)

Product Code
Regulation Number
870.2790
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The IRNF is a software-only mobile medical application that is intended to be used with the Apple Watch. The feature analyzes pulse rate data to identify episodes of irregular heart rhythms suggestive of atrial fibrillation (AFib) and provides a notification to the user. The feature is intended for over-the-counter (OTC) use. It is not intended to provide a notification on every episode of irregular rhythm suggestive of AFib and the absence of a notification is not intended to indicate no disease process is present; rather the feature is intended to opportunistically surface a notification of possible AFib when sufficient data are available for analysis. These data are only captured when the user is still. Along with the user's risk factors the feature can be used to supplement the decision for AFib screening. The feature is not intended to replace traditional methods of diagnosis or treatment.

The feature has not been tested for and is not intended for use in people under 22 years of age. It is also not intended for use in individuals previously diagnosed with AFib.

Device Description

IRNF 2.0 is comprised of a pair of mobile medical apps - One on Apple Watch and the other on the iPhone.

IRNE 2.0 is intended to analyze pulse rate data collected by the Apple Watch PPG sensor on Apple Watch Series 3-8, Series SE, and Apple Watch Ultra to identify episodes of irregular heart rhythms consistent with AFib and provides a notification to the user. It is a background screening tool and there is no way for a user to initiate analysis of pulse rate data. IRNF 2.0 iPhone App is part of the Health App, which allows users to store, manage, and share health and fitness data, and comes pre-installed on every iPhone.

IRNF 2.0 Watch App refers to the tachogram classification algorithm, confirmation cycle algorithm, and the AF notification generation. If an irreqular heart rhythm consistent with AFib is identified, IRNF 2.0 Watch App will transfer the AFib notification to IRNF 2.0 iPhone App through HealthKit sync. In addition to indicating the finding of signs of AFib, the notification will encourage the user to seek medical care.

IRNF 2.0 iPhone App contains the on-boarding and educational materials that a user must review prior to enabling AFib notifications. IRNF 2.0 iPhone App is designed to work in combination with IRNF 2.0 Watch App and will display a history of all prior AFib notifications. The user is also able to view a list of times when each of the irregular tachograms contributing to the notification was generated.

AI/ML Overview

The provided text describes the Irregular Rhythm Notification Feature (IRNF) 2.0. However, the document provided is a 510(k) summary and clearance letter for a Predetermined Change Control Plan (PCCP) for IRNF 2.0, rather than a standalone study proving the device meets acceptance criteria for initial clearance.

The document indicates that the subject device (IRNF 2.0) is identical to its predicate device (also IRNF 2.0, K212516), with the only difference being the implementation of a PCCP. This PCCP outlines anticipated modifications to the software and the methods for implementing those changes. Therefore, the acceptance criteria and study data for the initial clearance of IRNF 2.0 (K212516) would be the most relevant information, which is not entirely detailed in this document.

However, the PCCP does specify test methods and acceptance criteria that will be used to demonstrate substantial equivalence for future modifications made under the plan. I will extract information primarily related to these future modification criteria and the study that would be performed to meet them.

Here's a breakdown based on the provided text, focusing on the PCCP and what it implies for future studies:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria described here are for future modifications to the algorithm under the PCCP, showing substantial equivalence to the performance of the existing IRNF 2.0. The document does not provide the absolute performance of IRNF 2.0 itself in this section, but rather the performance target for modified algorithms relative to IRNF 2.0.

Category of ChangeAcceptance CriteriaReported Device Performance (as described for future modifications)
Modifications to Tachogram Classification AlgorithmSubstantial equivalence in sensitivity and specificity when compared to the performance of IRNF 2.0To be demonstrated in future validation activities under the PCCP, by meeting the specified substantial equivalence in sensitivity and specificity criteria.
Modifications to Confirmation Cycle AlgorithmSubstantial equivalence in positive predictive value relative to IRNF 2.0To be demonstrated in future validation activities under the PCCP, by meeting the specified substantial equivalence in positive predictive value criteria.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: The document states that for future modifications under the PCCP, "each will meet minimum demographic requirements for age, sex, race, and skin tone derived from the demographics of the United States." It does not specify an exact numerical sample size for the test set.
  • Data Provenance: The document implies that validation test datasets will be "representative of the intended use population" and mentions "demographics of the United States." This suggests the data will primarily be from the United States. It does not explicitly state whether the data will be retrospective or prospective for these future validation activities.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not specify the number of experts or their qualifications for establishing ground truth, either for the initial clearance of IRNF 2.0 or for the future modifications under the PCCP.

4. Adjudication Method for the Test Set

The document does not specify an adjudication method (e.g., 2+1, 3+1, none) for the test set, either for the initial clearance of IRNF 2.0 or for the future modifications under the PCCP.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

The document does not mention a Multi-Reader Multi-Case (MRMC) comparative effectiveness study. The IRNF is described as a "software-only mobile medical application" providing notifications to the user, not a tool for human readers to interpret.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the document implies that standalone performance studies were done (or will be done for future modifications). The device is described as "software-only" and "analyzes pulse rate data... and provides a notification to the user." The acceptance criteria for future modifications explicitly refer to the algorithm's sensitivity, specificity, and positive predictive value, which are metrics of standalone algorithm performance.

7. The Type of Ground Truth Used

The document does not explicitly state the type of ground truth used (e.g., expert consensus, pathology, outcomes data). In the context of "irregular heart rhythms suggestive of atrial fibrillation (AFib)," the ground truth would typically be established by a gold standard method such as a 12-lead ECG interpreted by a cardiologist, or a continuous ECG monitor.

8. The Sample Size for the Training Set

The document states that for future modifications to the tachogram classification algorithm, the plan is to "retrain algorithm with additional datasets." It does not specify the sample size for the training set, either for the original IRNF 2.0 or for the "additional datasets" mentioned for future retraining.

9. How the Ground Truth for the Training Set Was Established

The document does not specify how the ground truth for the training set was established, either for the original IRNF 2.0 or for future retraining datasets.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 21, 2023

Apple Inc. Bonnie Wu Regulatory Affairs Lead One Apple Park Way Cupertino, California 95014

Re: K231173

Trade/Device Name: Irregular Rhythm Notification Feature (IRNF) Regulation Number: 21 CFR 870.2790 Regulation Name: Photoplethysmograph Analysis Software For Over-The-Counter Use Regulatory Class: Class II Product Code: QDB Dated: April 25, 2023 Received: April 25, 2023

Dear Bonnie Wu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an

{1}------------------------------------------------

established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the quality systems (OS) regulation (21 CFR 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jennifer W. Shih -S

Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K231173

Device Name

Irregular Rhythm Notification Feature (IRNF)

Indications for Use (Describe)

The IRNF is a software-only mobile medical application that is intended to be used with the Apple Watch. The feature analyzes pulse rate data to identify episodes of irregular heart rhythms suggestive of atrial fibrillation (AFib) and provides a notification to the user. The feature is intended for over-the-counter (OTC) use. It is not intended to provide a notification on every episode of irregular rhythm suggestive of AFib and the absence of a notification is not intended to indicate no disease process is present; rather the feature is intended to opportunistically surface a notification of possible AFib when sufficient data are available for analysis. These data are only captured when the user is still. Along with the user's risk factors the feature can be used to supplement the decision for AFib screening. The feature is not intended to replace traditional methods of diagnosis or treatment.

The feature has not been tested for and is not in people under 22 years of age. It is also not intended for use in individuals previously diagnosed with AFib.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over The Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92:

5.1 Submitter

ApplicantApple Inc.One Apple Park WayCupertino, CA 95014
Primary CorrespondentBonnie WuRegulatory AffairsPhone: (408) 974-0617Email: bonnie_h_wu@apple.com
Secondary CorrespondentLuke OlsonRegulatory AffairsPhone: (408) 609-2001Email: luke_olson@apple.com
Date PreparedJune 23, 2023

5.2 Device Names and Classifications

Subject Device:

Name of DeviceIRNF 2.0
Classification NamePhotoplethysmograph Analysis Software For Over-The-Counter Use, 21 CFR870.2790
Regulatory ClassClass II
Product CodeQDB
510(k) Review PannelCardiovascular

Predicate Device:

PredicateManufacturerApple Inc.
Predicate Trade NameIRNF 2.0
Predicate 510(k)K212516

{5}------------------------------------------------

5.3 Device Description

IRNF 2.0 is comprised of a pair of mobile medical apps - One on Apple Watch and the other on the iPhone.

IRNE 2.0 is intended to analyze pulse rate data collected by the Apple Watch PPG sensor on Apple Watch Series 3-8, Series SE, and Apple Watch Ultra to identify episodes of irregular heart rhythms consistent with AFib and provides a notification to the user. It is a background screening tool and there is no way for a user to initiate analysis of pulse rate data. IRNF 2.0 iPhone App is part of the Health App, which allows users to store, manage, and share health and fitness data, and comes pre-installed on every iPhone.

IRNF 2.0 Watch App refers to the tachogram classification algorithm, confirmation cycle algorithm, and the AF notification generation. If an irreqular heart rhythm consistent with AFib is identified, IRNF 2.0 Watch App will transfer the AFib notification to IRNF 2.0 iPhone App through HealthKit sync. In addition to indicating the finding of signs of AFib, the notification will encourage the user to seek medical care.

IRNF 2.0 iPhone App contains the on-boarding and educational materials that a user must review prior to enabling AFib notifications. IRNF 2.0 iPhone App is designed to work in combination with IRNF 2.0 Watch App and will display a history of all prior AFib notifications. The user is also able to view a list of times when each of the irregular tachograms contributing to the notification was generated.

5.4 Indications for Use

The IRNF is a software-only mobile medical application that is intended to be used with the Apple Watch. The feature analyzes pulse rate data to identify episodes of irreqular heart rhythms suggestive of atrial fibrillation (AFib) and provides a notification to the user. The feature is intended for over-the-counter (OTC) use. It is not intended to provide a notification on every episode of irreqular rhythm suggestive of AFib and the absence of a notification is not intended to indicate no disease process is present; rather the feature is intended to opportunistically surface a notification of possible AFib when sufficient data are available for analysis. These data are only captured when the user is still. Along with the user's risk factors the feature can be used to supplement the decision for AFib screening. The feature is not intended to replace traditional methods of diagnosis or treatment.

The feature has not been tested for and is not intended for use in people under 22 vears of age. It is also not intended for use in individuals previously diagnosed with AFib.

5.5 Comparison of Technological Characteristics with the Predicate Device

The subject device is identical to the predicate device, having the same intended use, indications for use, and principles of operation as its predicate device. The technological characteristics of the subject device are identical to the predicate device, other than the implementation of a Predetermined Change Control Plan (PCCP) that specifies anticipated modifications to the IRNF

{6}------------------------------------------------

2.0 software and the methods that will be utilized to implement those modifications in a controlled manner such that the device is as safe and as effective as the predicate.

The PCCP includes a specific list of software modifications defining the region of potential changes that can be made to the algorithms in the device. The detailed description of the changes, associated requirements and test methods are summarized in the table below.

The PCCP includes an algorithm modification protocol describing the verification and validation activities that will support the proposed changes. The modification protocol incorporates impact assessment considerations and specifies requirements for data management, including data sources, collection, storage, and sequestration, as well as documentation and data re-use practices.

Specific test methods are specified in the PCCP to establish substantial equivalence relative to IRNF 2.0, and include sample size determination, analysis methods, and acceptance criteria. To help ensure validation test datasets are representative of the intended use population, each will meet minimum demographic requirements for age, sex, race, and skin tone derived from the demographics of the United States.

Detailed List ofChangesRequirementsTest Method
Modifications toTachogramClassificationAlgorithmAdjust the numericalthreshold at which atachogram is classifiedas AF (operating point).While maintaining thesame algorithmarchitecture andnumber of parameters,retrain algorithm withadditional datasetsNo change to sensor inputsignal typeNo change to output typeNo change to confirmationcycle algorithmNo change to intended use ofdeviceCan be fully verified/validatedby requirements of theModification Protocol andapplicable Test MethodSubstantialequivalence insensitivity andspecificity whencompared to theperformance of IRNF2.0
Modifications toConfirmationCycle AlgorithmModification to thenumber of sequentialtachograms that mustbe classified asirregular in a given timeperiod to surface anotificationModification to the timeperiod of confirmationcycleNo change to input ofconfirmation cycle algorithmNo change to output typeNo change to tachogramclassification algorithmNo change to intended use ofdeviceCan be fully verified/validatedby requirements of theModification Protocol andapplicable Test MethodSubstantialequivalence inpositive predictivevalue relative to IRNF2.0

In accordance with the PCCP, all algorithm modifications will be trained, tuned, and locked prior to release of the software to the field. The PCCP does not include provisions for implementation of adaptive algorithms that will continuously learn in the field. A procedure for Instructions for Use

{7}------------------------------------------------

updates has been established in order to inform users about algorithm changes implemented under this FDA authorized PCCP, including a summary of the changes and a characterization of algorithm performance, as well as the availability and compatibility of the feature. Apple will publish updated Instructions for Use on its website, and make them accessible within the Health App.

5.6 Conclusion

The subject device, IRNF 2.0, has the same intended use, indications for use, and principles of operation as its predicate device. The technological characteristics are identical with the exception of the implementation of a PCCP, and this difference does not raise new questions of safety and effectiveness. Thus, the subject device is substantially equivalent.

§ 870.2790 Photoplethysmograph analysis software for over-the-counter use.

(a)
Identification. A photoplethysmograph analysis software device for over-the-counter use analyzes photoplethysmograph data and provides information for identifying irregular heart rhythms. This device is not intended to provide a diagnosis.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Clinical performance testing must demonstrate the performance characteristics of the detection algorithm under anticipated conditions of use.
(2) Software verification, validation, and hazard analysis must be performed. Documentation must include a characterization of the technical specifications of the software, including the detection algorithm and its inputs and outputs.
(3) Non-clinical performance testing must demonstrate the ability of the device to detect adequate photoplethysmograph signal quality.
(4) Human factors and usability testing must demonstrate the following:
(i) The user can correctly use the device based solely on reading the device labeling; and
(ii) The user can correctly interpret the device output and understand when to seek medical care.
(5) Labeling must include:
(i) Hardware platform and operating system requirements;
(ii) Situations in which the device may not operate at an expected performance level;
(iii) A summary of the clinical performance testing conducted with the device;
(iv) A description of what the device measures and outputs to the user; and
(v) Guidance on interpretation of any results.