K Number
K212516

Validate with FDA (Live)

Device Name
IRNF App
Manufacturer
Date Cleared
2021-10-22

(73 days)

Product Code
Regulation Number
870.2790
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Irregular Rhythm Notification Feature is a software-only mobile medical application that is intended to be used with the Apple Watch. The feature analyzes pulse rate data to identify episodes of irregular heart rhythms suggestive of atrial fibrillation (AFib) and provides a notification to the user. The feature is intended for over-the-counter (OTC) use. It is not intended to provide a notification on every episode of irregular rhythm suggestive of AFib and the absence of a notification is not intended to indicate no disease process is present; rather is intended to opportunistically surface a notification of possible AFib when sufficient data are available for analysis. These data are only captured when the user is still. Along with the user's risk factors the feature can be used to supplement the decision for AFib screening. The feature is not intended to replace traditional methods of diagnosis or treatment.

The feature has not been tested for and is not in people under 22 years of age. It is also not intended for use in individuals previously diagnosed with AFib.

Device Description

Irregular Rhythm Notification Feature 2.0 (IRNF 2.0) is comprised of a pair of mobile medical apps - One on Apple Watch and the other on the iPhone.

IRNF 2.0 is intended to analyze pulse rate data collected by the Apple Watch PPG sensor on Apple Watch Series 3. Series 4. Series 5. and SE to identify exisodes of irreqular heart rhythms consistent with AFib and provide a notification to the user. It is a background screening tool and there is no way for a user to initiate analysis of pulse rate data. IRNF 2.0 iPhone App is part of the Health App. which allows users to store, manage, and share health and fitness data, and comes pre-installed on every iPhone.

IRNF 2.0 Watch App refers to the rhythm classification algorithm, confirmation cvcle algorithm, and the AFib notification generation. If an irreqular heart rhythm consistent with Afib is identified and confirmed through the confirmation cycle, IRNF 2.0 Watch app will notify the user and transfer the AFib notification to the iPhone App through HealthKit sync. In addition to indicating the finding of signs of AFib, the notification will encourage the user to seek medical care.

IRNF 2.0 iPhone App contains the onboarding and educational materials that a user must review prior to use. IRNF 2.0 iPhone App is designed to work in combination with IRNF 2.0 Watch App and will display a history of all prior AFib notifications. The user is also able to view a list of times of the irreqular rhythms contributing to the notification.

AI/ML Overview

The provided text describes the Irregular Rhythm Notification Feature (IRNF) 2.0 app, a software-only mobile medical application for detecting irregular heart rhythms suggestive of Atrial Fibrillation (AFib) using Apple Watch pulse rate data. Below is a detailed breakdown of the acceptance criteria and study proving the device meets them, based on the provided document:

Acceptance Criteria and Reported Device Performance

The document states that the clinical performance of the IRNF 2.0 app was assessed, with specific metrics reported:

MetricAcceptance Criteria (Implied)Reported Device Performance
Person-level Sensitivity (AFib)Not explicitly stated but inferred as non-inferior to predicate88.6%
Person-level Specificity (AFib)Not explicitly stated but inferred as non-inferior to predicate99.3%

Note: The document states "IRNF 2.0 person-level sensitivity (88.6%) and specificity (99.3%) were both demonstrated to be non-inferior to those of the predicate device." While a specific numerical acceptance criterion for sensitivity and specificity isn't explicitly listed, the demonstration of non-inferiority to the predicate device (which had a reported 78.9% sensitivity for concordant AFib and 98.2% for AFib and other clinically relevant arrhythmias) serves as the implicit acceptance benchmark.


Study Details

Here's the information about the study that proves the device meets the acceptance criteria:

2. Sample Size Used for the Test Set and Data Provenance:

  • Test Set Sample Size: The clinical validation study (referred to as the "clinical study" under 5.7 Clinical Performance) involved 573 participants.
    • For the primary endpoint analysis, 432 participants contributed data to determine sensitivity, with 140 of these presenting with AFib.
    • 292 participants contributed data to the analysis of device specificity.
  • Data Provenance: The document does not explicitly state the country of origin of the data. It is implied to be a prospective study, as it involved enrolled subjects wearing an Apple Watch and a reference ECG patch concurrently.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • The document does not specify the number or qualifications of experts used to establish the ground truth for the test set. It states that the reference was an "electrocardiogram (ECG) patch concurrently." This implies a medical standard for AFib diagnosis, but the human interpretation component by experts is not detailed.

4. Adjudication Method for the Test Set:

  • The document does not describe a specific adjudication method (e.g., 2+1, 3+1) for the test set. The ground truth appears to be established directly from the reference ECG patch data.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

  • No, an MRMC comparative effectiveness study involving human readers assisting with or without AI was not described for this device. The study primarily focuses on the standalone performance of the IRNF 2.0 app against a reference standard.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Yes, a standalone performance study was done. The reported sensitivity (88.6%) and specificity (99.3%) refer to the device's performance in identifying irregular heart rhythms suggestive of AFib based on pulse rate data, without human interpretation in the loop for the primary performance metrics. The feature provides a notification to the user, encouraging them to seek medical care, but its performance metrics are of the algorithm only.

7. The Type of Ground Truth Used:

  • The ground truth for the clinical study was established using a reference electrocardiogram (ECG) patch, worn concurrently by participants. AFib was "identified on the reference ECG patch."

8. The Sample Size for the Training Set:

  • The document states that the new rhythm classification algorithm "was trained extensively using data collected in a number of development studies." These studies "included over 2500 subjects." While the exact size of the training set itself is not broken out numerically, it's part of this larger "over 2500 subjects" pool. The data was split into Training, Validation, Testing, and Sequestration sets.

9. How the Ground Truth for the Training Set Was Established:

  • The ground truth for the training set was established from "data collected in a number of development studies" that included "over 3 million pulse rate recordings on a variety of rhythms including: atrial fibrillation, normal sinus rhythm, sinus arrhythmia, and other ectopic beats (PVCs, PACs)."
  • It's inferred that these "rhythms" were definitively classified to serve as ground truth for training the convolutional neural network. While the exact method of establishing ground truth for the training data (e.g., direct ECG correlation, expert annotation of ECGs) is not explicitly detailed, the mention of specific rhythm types suggests a medically validated classification.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 26, 2021

Apple Inc. Luke Olson Regulatory Affairs Associate 1 Apple Park Way Cupertino, California 95014

Re: K212516

Trade/Device Name: IRNF App Regulation Number: 21 CFR 870.2790 Regulation Name: Photoplethysmograph analysis software for over-the-counter use Regulatory Class: Class II Product Code: QDB

Dear Luke Olson:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated October 22, 2021. Specifically, FDA is updating this SE Letter due to a typo in the trade name as an administrative correction.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Jennifer Shih Kozen, Office of Cardiovascular Devices, 301-796-5813, Jennifer.Shih(@fda.hhs.gov.

Sincerely,

Jennifer W. Shih -S

Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue. Underneath the FDA logo is the word "ADMINISTRATION".

October 22, 2021

Apple Inc. Luke Olson Regulatory Affairs Associate 1 Apple Park Way Cupertino, California 95014

Re: K212516

Trade/Device Name: Irregular Ryhthm Notification Feature (IRNF) 2.0 App Regulation Number: 21 CFR 870.2790 Regulation Name: Photoplethysmograph analysis software for over-the-counter use Regulatory Class: Class II Product Code: QDB Dated: August 9, 2021 Received: August 10, 2021

Dear Luke Olson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{2}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-

542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@tda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jennifer W. Shih -S

Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K212516

Device Name Irregular Rhythm Notification Feature 2.0

Indications for Use (Describe)

The Irregular Rhythm Notification Feature is a software-only mobile medical application that is intended to be used with the Apple Watch. The feature analyzes pulse rate data to identify episodes of irregular heart rhythms suggestive of atrial fibrillation (AFib) and provides a notification to the user. The feature is intended for over-the-counter (OTC) use. It is not intended to provide a notification on every episode of irregular rhythm suggestive of AFib and the absence of a notification is not intended to indicate no disease process is present; rather is intended to opportunistically surface a notification of possible AFib when sufficient data are available for analysis. These data are only captured when the user is still. Along with the user's risk factors the feature can be used to supplement the decision for AFib screening. The feature is not intended to replace traditional methods of diagnosis or treatment.

The feature has not been tested for and is not in people under 22 years of age. It is also not intended for use in individuals previously diagnosed with AFib.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)X

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92:

5.1 Submitter

ApplicantApple Inc.One Apple Park WayCupertino, CA 95014
PrimaryCorrespondentLuke OlsonRegulatory AffairsPhone: (408) 609-2001Email: luke_olson@apple.com
SecondaryCorrespondentDachan KwonRegulatory AffairsPhone: (669) 268-5659Email: dachan_kwon@apple.com
Date PreparedAugust 09, 2021

5.2 Device Names and Classifications

Subject Device:

Name of DeviceIrregular Rhythm Notification Feature 2.0
Classification NamePhotoplethysmograph Analysis Software For Over-The-Counter Use,21 CFR 870.2790
Regulatory ClassClass II
Product CodeQDB
510(k) ReviewPanelCardiovascular

Predicate Device:

PredicateManufacturerApple Inc.
Predicate TradeNameIrregular Rhythm Notification Feature
Predicate 510(k)DEN180042

{5}------------------------------------------------

5.3 Device Description

Irregular Rhythm Notification Feature 2.0 (IRNF 2.0) is comprised of a pair of mobile medical apps - One on Apple Watch and the other on the iPhone.

IRNF 2.0 is intended to analyze pulse rate data collected by the Apple Watch PPG sensor on Apple Watch Series 3. Series 4. Series 5. and SE to identify exisodes of irreqular heart rhythms consistent with AFib and provide a notification to the user. It is a background screening tool and there is no way for a user to initiate analysis of pulse rate data. IRNF 2.0 iPhone App is part of the Health App. which allows users to store, manage, and share health and fitness data, and comes pre-installed on every iPhone.

IRNF 2.0 Watch App refers to the rhythm classification algorithm, confirmation cvcle algorithm, and the AFib notification generation. If an irreqular heart rhythm consistent with Afib is identified and confirmed through the confirmation cycle, IRNF 2.0 Watch app will notify the user and transfer the AFib notification to the iPhone App through HealthKit sync. In addition to indicating the finding of signs of AFib, the notification will encourage the user to seek medical care.

IRNF 2.0 iPhone App contains the onboarding and educational materials that a user must review prior to use. IRNF 2.0 iPhone App is designed to work in combination with IRNF 2.0 Watch App and will display a history of all prior AFib notifications. The user is also able to view a list of times of the irreqular rhythms contributing to the notification.

5.4 Indications for Use

The Irreqular Rhythm Notification Feature is a software-only mobile medical application that is intended to be used with the Apple Watch. The feature analyzes pulse rate data to identify episodes of irregular heart rhythms suggestive of atrial fibrillation (AFib) and provides a notification to the user. The feature is intended for over-the-counter (OTC) use. It is not intended to provide a notification on every episode of irregular rhythm suggestive of AFib and the absence of a notification is not intended to indicate no disease process is present; rather the feature is intended to opportunistically surface a notification of possible AFib when sufficient data are available for analysis. These data are only captured when the user is still. Along with the user's risk factors the feature can be used to supplement the decision for AFib screening. The feature is not intended to replace traditional methods of diagnosis or treatment.

The feature has not been tested for and is not intended for use in people under 22 years of age. It is also not intended for use in individuals previously diagnosed with AFib.

{6}------------------------------------------------

5.5 Comparison with the Predicate Device

ItemSubject DevicePredicate Device
IRNF 2.0 App (K212516)IRNF App (DEN180042)
ManufacturerApple Inc.Apple Inc.
SubmissionReferenceK212516DEN180042
Intended UsePhotoplethysmograph analysissoftware for over-the-counter use. Aphotoplethysmograph analysissoftware device for over-the-counteruse analyzes photoplethysmographdata and provides information foridentifying irregular heart rhythms.This device is not intended to providea diagnosis.Photoplethysmograph analysissoftware for over-the-counter use. Aphotoplethysmograph analysissoftware device for over-the-counteruse analyzes photoplethysmographdata and provides information foridentifying irregular heart rhythms.This device is not intended to providea diagnosis.
Subject DevicePredicate Device
ltemIRNF 2.0 App (K212516)IRNF App (DEN180042)
Indications forUseThe Irregular Rhythm NotificationFeature is a software-only mobilemedical application that is intended tobe used with the Apple Watch. Thefeature analyzes pulse rate data toidentify episodes of irregular heartrhythms suggestive of atrial fibrillation(AFib) and provides a notification tothe user. The feature is intended forover-the-counter (OTC) use. It is notintended to provide a notification onevery episode of irregular rhythmsuggestive of AFib and the absence ofa notification is not intended toindicate no disease process ispresent; rather the feature is intendedto opportunistically surface anotification of possible AFib whensufficient data are available foranalysis. These data are only capturedwhen the user is still. Along with theuser's risk factors the feature can beused to supplement the decision forAFib screening. The feature is notintended to replace traditionalmethods of diagnosis or treatment.The feature has not been tested forand is not intended for use in peopleunder 22 years of age. It is also notintended for use in individualspreviously diagnosed with AFibThe Irregular Rhythm NotificationFeature is a software-only mobilemedical application that is intended tobe used with the Apple Watch. Thefeature analyzes pulse rate data toidentify episodes of irregular heartrhythms suggestive of atrial fibrillation(AFib) and provides a notification tothe user. The feature is intended forover-the-counter (OTC) use. It is notintended to provide a notification onevery episode of irregular rhythmsuggestive of AFib and the absence ofa notification is not intended toindicate no disease process ispresent; rather the feature is intendedto opportunistically surface anotification of possible AFib whensufficient data are available foranalysis. These data are only capturedwhen the user is still. Along with theuser's risk factors the feature can beused to supplement the decision forAFib screening. The feature is notintended to replace traditionalmethods of diagnosis or treatment.The feature has not been tested forand is not intended for use in peopleunder 22 years of age. It is also notintended for use in individualspreviously diagnosed with AFib
Principle ofOperationThe IRN 2.0 acquires platform sensordata from Apple Watch. Afteracquisition, the IRN 2.0 algorithmsanalyze pulse rate data to identifyepisodes of irregular heart rhythmssuggestive of atrial fibrillation (AFib)and provides notification to the user.The IRN app acquires platform sensordata from Apple Watch. Afteracquisition, the IRN app algorithmsanalyze pulse rate data to identifyepisodes of irregular heart rhythmssuggestive of atrial fibrillation (AFib)and provides notification to the user.
ItemSubject DevicePredicate Device
IRNF 2.0 App (K212516)IRNF App (DEN180042)
ClinicalPerformanceApple conducted a clinical validationstudy to assess the performance ofthe subject IRNF 2.0 app relative tothat of the predicate device on acommon sensor dataset.IRNF 2.0 person-level sensitivity(88.6%) and specificity (99.3%) wereboth demonstrated to be non-inferiorto those of the predicate device.In a study of 226 participants aged 22years or older wearing Apple Watchand an electrocardiogram (ECG)patch concurrently, 57 participantsreceived AFib notifications.Of those, 78.9% (45/57) showedconcordant AFib on the ECG patch,while 98.2 % (56/57) showed AFiband other clinically relevantarrhythmias.
Compatibilitywith IntendedPlatformsiOS version 15.5 or laterwatchOS version 8.5 or laterApple Watch Series 3, 4, 5, SEiPhone 6s and lateriOS 12.1.1 and laterwatchOS 5.1.2 and laterApple Watch Series 1 and lateriPhone 5s and later

{7}------------------------------------------------

{8}------------------------------------------------

5.6 Performance Testing

IRNF 2.0 was verified and validated according to Apple's internal design control processes and in accordance with the special controls for Photoplethysmograph Analysis software for over-the-counter use (21 CFR 870.2790). The testing demonstrated that the device performed according to its specifications and that the technological and performance criteria are comparable to the predicate device.

IRNF 2.0 includes a new rhythm classification algorithm that leverages machine learning techniques to differentiate between AFib and non-AFib rhythms. The new rhythm classification algorithm uses a convolutional neural network based architecture and was trained extensively using data collected in a number of development studies. In total, the studies included over 2500 subjects and collected over 3 million pulse rate recordings on a variety of rhythms including: atrial fibrillation, normal sinus rhythm, sinus arrhythmia, and other ectopic beats (PVCs, PACs).

{9}------------------------------------------------

The studies used to train the convolutional network recruited demographically diverse populations with broad representation of age, sex, BMI, race, and skin tones. Table 2 below summarizes approximate development study demographic characteristics:

Age Group (years)
<5539.5%
>=55 to <6525.4%
>=6535.1%
Sex
Male49.6%
Female50.4%
BMI (kg/m2)
<18.52.2%
>=18.5 to <25.032.7%
>=25.0 to <30.032.2%
>=30.032.9%
Race
White71.5%
Black or African American18.0%
Other10.5%

Table 2. Development Study Subject Demographics

For the purpose of developing the algorithm, the data was split into four sets with matching distributions of rhythms and demographics: Traininq, Validation, Testing, and Sequestration sets. The model was trained on the Training set, with the Validation set used for early stopping and threshold selection. The model was then evaluated on the Testing set at reqular intervals during model development. When development was complete the model was locked, and then evaluated on the Sequestration set as a last test to ensure it had not been over-fit to the development data.

{10}------------------------------------------------

5.7 Clinical Performance

The performance of the Irreqular Rhythm Notification Feature (IRNF) was extensively tested in a clinical study of 573 participants ages 22 and older with a mix of diagnosed AFib and no known history of AFib. Study demographic characteristics are summarized in Table 3 below:

N=573
Age Group (years)
<55123 (21.5%)
>=55 to <65140 (24.4%)
>=65310 (54.1%)
Sex
Male286 (49.9%)
Female287 (50.1%)
Ethnicity
Hispanic or Latino38 (6.6%)
Non-Hispanic or Latino535 (93.4%)
Race
White502 (87.6%)
Black or African American57 (9.9%)
Other14 (2.4%)

Table 3. IRNF 2.0 Clinical Study Subject Demographics

Enrolled subjects wore an Apple Watch and a reference electrocardiogram (ECG) patch concurrently for up to 13 days. For those subjects contributing data to the primary endpoint analysis, 32.4% (n=140/432) presented with AFib as identified on the reference ECG patch and were included in determining the device sensitivity. Of those, 124 received an IRNF irregular rhythm notification with concordant AFib on the ECG patch, and the sensitivity was 88.6%. Of the 292 subjects who did not present with AFib on the ECG patch and contributed data to the analysis of device specificity, 290 did not receive a notification. The AF detection specificity was 99.3%. The remaining subjects (n=141/573) either contributed data to only secondary endpoint analyses and/or did not complete the study. These results support the device's effectiveness in detecting AFib.

{11}------------------------------------------------

5.8 Human Factors Testing

Compared to the predicate device, there is no change to the indications for use, intended user populations, intended part of body applied to, use environment, operating principle, user interactions, use related hazards, use scenarios and critical tasks for IRNF 2.0. As such, Apple leveraged the Usability Engineering Report generated during development of the predicate device.

5.9 Conclusion

IRNF 2.0 is substantially equivalent to IRNF as they are identical with respect to intended use and there are no differences in technological or performance characteristics that raise new questions of safety and effectiveness.

§ 870.2790 Photoplethysmograph analysis software for over-the-counter use.

(a)
Identification. A photoplethysmograph analysis software device for over-the-counter use analyzes photoplethysmograph data and provides information for identifying irregular heart rhythms. This device is not intended to provide a diagnosis.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Clinical performance testing must demonstrate the performance characteristics of the detection algorithm under anticipated conditions of use.
(2) Software verification, validation, and hazard analysis must be performed. Documentation must include a characterization of the technical specifications of the software, including the detection algorithm and its inputs and outputs.
(3) Non-clinical performance testing must demonstrate the ability of the device to detect adequate photoplethysmograph signal quality.
(4) Human factors and usability testing must demonstrate the following:
(i) The user can correctly use the device based solely on reading the device labeling; and
(ii) The user can correctly interpret the device output and understand when to seek medical care.
(5) Labeling must include:
(i) Hardware platform and operating system requirements;
(ii) Situations in which the device may not operate at an expected performance level;
(iii) A summary of the clinical performance testing conducted with the device;
(iv) A description of what the device measures and outputs to the user; and
(v) Guidance on interpretation of any results.