K Number
K230944

Validate with FDA (Live)

Device Name
MeMed BV
Date Cleared
2023-06-30

(87 days)

Product Code
Regulation Number
866.3215
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MeMed BV test is an automated semi-quantitative immunoassay that measures three non-microbial (host) proteins (TRAIL, IP-10, and CRP) in adult and pediatic serum and venous whole blood samples and is intended for use in conjunction with clinical assessments and other laboratory findings as an aid to differentiate bacterial from viral infection. MeMed BV is indicated for use in patients presenting to the emergency department or urgent care center and with samples collected at hospital admission from patients with suspected acute bacterial or viral infection, who have had symptoms for less than seven days. The MeMed BV test generates a numeric score that falls within discrete interpretation bins based on the increasing likelihood of bacterial infection.

Device Description

The MeMed BV® ("BV test" or the "test") is an In-Vitro-Diagnostic device that measures in parallel the blood concentrations of TRAIL, IP-10 and CRP. The test consists of an automated analyzer with built-in hardware and software that conduct chemiluminescence based analyte measurements of patient serum and venous whole blood samples and their computational integration (MeMed Key®), and a disposable cartridge that contains the reagents and controls needed to detect the analytes of interest (MeMed BV® cartridge). The test generates an answer to each sample, with a test run time of approximately 15 minutes.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided FDA 510(k) summary for MeMed BV:

The MeMed BV test is intended for use in conjunction with clinical assessments and other laboratory findings as an aid to differentiate bacterial from viral infection in patients presenting to the emergency department or urgent care center, or with samples collected at hospital admission, who have had symptoms for less than seven days. The device generates a numeric score that falls within discrete interpretation bins based on the increasing likelihood of bacterial infection.

1. Table of Acceptance Criteria and Reported Device Performance

The 510(k) summary details various analytical performance studies and a clinical study to support the expanded indications for use. Key acceptance criteria and reported performance include:

Test CategoryAcceptance CriteriaReported Device Performance
Analytical Performance
Limit of Quantitation (LoQ)Total Error (TE): TRAIL < 30%, IP-10 < 40%, CRP < 30%Serum Test Script: - TRAIL: Max TE at LLOQ (15 pg/mL) was 21%. (One sample at X0.8 showed 51% TE for TRAIL, but the defined LLOQ concentration level (X1.0) met criteria).- IP-10: Max TE at LLOQ (100 pg/mL) was 15%.- CRP: Max TE at LLOQ (1 mg/L) was 9%.Whole Blood (WB) Test Script: - TRAIL: Max TE at LLOQ (15 pg/mL) was 10%.- IP-10: Max TE at LLOQ (100 pg/mL) was 14%.- CRP: Max TE at LLOQ (1 mg/L) was 10%. All defined LLOQ concentrations for both serum and WB met the acceptance criteria.
Reproducibility/PrecisionMeasurands (TRAIL, IP-10, CRP): CV ≤ 15% (for concentrations above LoQ).MeMed BV® Test Score: SD < 12.5 score units.Serum Samples: All reported repeatability, intermediate precision, and reproducibility CVs for TRAIL, IP-10, and CRP were ≤ 11.3%. All reported SDs for the MeMed BV Score were ≤ 6.6 score units. WB Samples (Precision): All reported CVs for TRAIL, IP-10, and CRP were ≤ 12.0%. All reported SDs for the MeMed BV Score were ≤ 3 score units. All reported values met the pre-established acceptance criteria.
Lot-to-Lot ReproducibilityMeasurands (TRAIL, IP-10, CRP): CV ≤ 15% (for concentrations above LoQ).MeMed BV® Test Score: SD < 12.5 score units.All reported between lot CVs for TRAIL, IP-10, and CRP were ≤ 10.7%. All reported between lot SDs for the MeMed BV Score were ≤ 2.3 score units. All reported values met the pre-established acceptance criteria.
LinearityAllowable deviation from linearity (ADL) < 15% or 10 mg/L for CRP; 15% or 10 pg/mL for TRAIL; 20% or 50 pg/mL for IP-10.Serum Samples: Max observed % deviation from linearity was 6.8% (TRAIL).Whole Blood Samples: Max observed % deviation from linearity was 8.6% (TRAIL). All results were within acceptance criteria.
Hook EffectNo hook effect observed up to tested concentrations (TRAIL – 1,000 pg/mL, IP-10 – 10,000 pg/mL, CRP – 500 mg/L).All concentrations up to TRAIL – 1,000 pg/mL, IP-10 – 10,000 pg/mL, and CRP – 500 mg/L showed higher signal than the ULOQ sample. No hook effect observed.
Carry OverWB Samples: Difference between average score of high score sample run after low score sample and high score sample baseline average score of ≤ 12.5 score units. Difference between average score of low score sample run after high score sample and low score sample baseline average score of ≤ 12.5 score units.Maximal difference in score obtained for high score sample was 1.4 score unit difference. No carry-over occurred with the MeMed BV test.
Interference/Cross Reactivity95% Confidence Interval for bias within +/- 12.5 score units for all interferents and cross-reactants.Previously submitted data (K222332) demonstrated this. The recovery of TRAIL, IP-10 and CRP were within the predetermined +/- 10% of the sample nominal concentration. Assays are tolerant to high HAMA concentrations and no interference/cross-reactivity from tested compounds.
Correlation to Reference Standard (New Calibration Scheme vs. Legacy)1. <5% of samples have MCC scores deviating from legacy calibration scores by an amount placing them in non-adjacent bins.2. Pearson correlation > 0.95.3. Absolute bias < 12.5 units at bin cutoff points (10, 35, 65, 90).1. The study successfully met the clinically relevant criterion (no paired samples assigned to nonadjacent bins).2. Pearson correlation was 1 (Deming Regression slope=1.00, 95% CI 0.99-1.00; intercept 0.00-0.06).3. Estimated bias at cutoff points ranged from -0.57 to -0.55, with 95% CIs well within +/- 12.5 units. New MCC is equivalent to legacy calibration.
Sample In-Use Stability (WB)Allowable handling conditions demonstrated from blood draw to sample input.The minimal acceptable period of time was approximately 140 minutes for TRAIL viral sample 1. Formal in-use stability of WB sample type established at 120 minutes prior to testing on analyzer.
Clinical Studies
Matrix Equivalency (WB vs. Serum)Passing & Bablok Regression: Slope in range of 0.9-1.1; Intercept in range of (-5) to 5.Slope: 1.00 (95% CI 0.99-1.00). Intercept: 0.00-0.06. Both predefined acceptance criteria for analytical equivalency were fulfilled.
Bin Impact Analysis (WB vs. Serum)<5% of paired samples demonstrating a score deviation that causes a patient to be assigned to a nonadjacent bin.No paired samples demonstrated a score deviation that caused the patient to be assigned to a nonadjacent bin. This strengthens the conclusion of analytical equivalency.
Diagnostic Accuracy (Simulated WB against Adjudication)Cochran-Armitage (CA) Test for trend: Reject null hypothesis (no trend of increasing probability of bacterial infection with higher test score) for ≥ 95% of simulations.Likelihood Ratio (LR): 95% CI should exclude 1 for some bins (preferably Bins 1,2,4,5) for ≥ 95% of simulations.All-inclusive cohort: CA p<0.001 for 100% of 100K simulations. CI of LR for exactly 4 bins (1,2,4,5) excluded 1 in 100% of simulations.Suspected cohort: CA p<0.001 for 100% of 100K simulations. CI of LR for exactly 4 bins (1,2,4,5) excluded 1 in 99.98% of simulations (0.02% had all 5 bins exclude 1). Both acceptance criteria passed, validating diagnostic accuracy of simulated WB samples.

2. Sample Sizes Used for the Test Set and Data Provenance

  • Analytical Performance Test Sets:

    • LoQ: Samples used per test script (serum and whole blood) and two cartridge lots. Specific number of unique samples not detailed, but each was tested three times on three non-consecutive days with results across 4 concentration levels.
    • Reproducibility/Precision & Lot-to-Lot Reproducibility: Panel of 4 scores (representing various infection statuses) for serum, and 3 scores for WB. Serum study involved 90 replicates per panel member across 3 labs. WB study involved runs on 5 different analyzers. Lot-to-lot used 18 replicates per panel member.
    • Linearity: Five replicates of eleven dilutions for each measurand.
    • Hook Effect: 4 samples with varying high concentrations.
    • Carry Over: Two whole blood samples (one high score, one low score) run in sequences.
    • Correlation to Reference Standard (Calibration Scheme Comparison): 100 serum specimens.
  • Clinical Study Test Set (Perseverance Study):

    • Sample Size: 216 prospectively recruited subjects.
    • Data Provenance: Multi-center study from 5 medical centers (2 in the US, 3 in Israel).
    • Retrospective/Prospective: Prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document states that the Apollo study (NCT04690569), which provided the basis for the original serum MeMed BV clearance (K210254), used a rigorous reference standard based on etiological adjudication by experts provided with comprehensive patient data.

  • Number of Experts: Not explicitly stated how many experts for the adjudication, but plural "experts" is used.
  • Qualifications of Experts: Not explicitly stated but they were responsible for "etiological adjudication" which implies medical professionals with expertise in differential diagnosis of infections (e.g., infectious disease specialists, clinical microbiologists, relevant clinical physicians). The term "comprehensive patient data" suggests they had access to clinical, laboratory, and other relevant information.

4. Adjudication Method for the Test Set

The ground truth for the clinical utility (diagnostic accuracy) of the MeMed BV was based on an adjudication-based reference standard from the Apollo study. The document doesn't specify the exact adjudication method (e.g., 2+1, 3+1), but implies a consensus or majority decision by the experts based on comprehensive patient data.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not mentioned as part of this 510(k) submission. This submission primarily focuses on analytical equivalency of whole blood samples to serum samples and simulated diagnostic accuracy, not human reader improvement with AI assistance. The MeMed BV is an in vitro diagnostic device that provides a numeric score, not an imaging AI algorithm designed to assist human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The device's performance, represented by its numeric score and bin assignment, is inherently standalone in terms of its output. The analytical performance evaluations (LoQ, precision, linearity, etc.) and the clinical study's simulation of diagnostic accuracy (comparison of MeMed BV score to adjudicated ground truth) assess the algorithm's performance directly. The device is intended to be used "in conjunction with clinical assessments and other laboratory findings," meaning its score is an aid, but its performance itself is measured as a standalone diagnostic aid.

7. The Type of Ground Truth Used

  • For Analytical Performance: Ground truth is established by the known concentrations of analytes in controls/calibrators, established reference methods, or through robust statistical measurements.
  • For Clinical Performance (Diagnostic Accuracy Simulation): The ground truth was an etiological adjudication by experts provided with comprehensive patient data from the Apollo study. This is a form of expert consensus based on extensive clinical information.

8. The Sample Size for the Training Set

The document does not specify a separate "training set" sample size for the MeMed BV algorithm because it is an in vitro diagnostic device that measures specific biomarkers and computationally integrates them. The algorithm's "training" or development would have occurred prior to the studies presented for this 510(k) (which are validation studies). The previous 510(k) (K222332) for the serum-only version would have covered the initial development and validation, and this submission focuses on extending the indication to whole blood samples.

The phrase "new Master Calibration Curve (MCC)" suggests a change to the underlying measurement calculation, which might involve a new calibration dataset for the algorithm, but this is an analytical validation, not a "training set" in the sense of machine learning model development. For the calibration scheme comparison, 100 serum specimens were used as a test set.

9. How the Ground Truth for the Training Set Was Established

As noted above, a distinct "training set" for the MeMed BV algorithm itself (as a machine learning model might have) is not described in this regulatory submission. For the analytical validations:

  • Calibration: Established against reference standards and through controlled experiments to define the relationship between measured signals (RLU) and analyte concentrations. The "Master Calibration Curve" is related to how the instrument translates raw signals into quantitative measurements and ultimately into the BV score.
  • Clinical Utility (Reference Standard): If earlier development involved classification algorithms, their "ground truth" would have similarly been derived from expert adjudication or robust clinical diagnoses in previous studies (like the Apollo study mentioned, which provided the original basis for serum clearance). The current "clinical studies" section primarily describes validation of the expanded indication.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 30, 2023

MeMed Diagnostics Ltd. Efrat Hartog-David VP of Regulatory Affairs and Quality Assurance Nahum Het 7 Tirat Carmel, 3508506 Israel

Re: K230944

Trade/Device Name: MeMed BV Regulation Number: 21 CFR 866.3215 Regulation Name: Device To Detect And Measure Non-Microbial Analyte(s) In Human Clinical Specimens To Aid In Assessment Of Patients With Suspected Sepsis Regulatory Class: Class II Product Code: QPS Dated: April 4, 2023 Received: April 4, 2023

Dear Efrat Hartog-David:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Noel J. Gerald -S

Noel J. Gerald, Ph.D. Branch Chief Bacterial Respiratory and Medical Countermeasures Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known)

K230944

Device Name

MeMed BV

Indications for Use (Describe)

The MeMed BV test is an automated semi-quantitative immunoassay that measures three non-microbial (host) proteins (TRAIL, IP-10, and CRP) in adult and pediatic serum and venous whole blood samples and is intended for use in conjunction with clinical assessments and other laboratory findings as an aid to differentiate bacterial from viral infection. MeMed BV is indicated for use in patients presenting to the emergency department or urgent care center and with samples collected at hospital admission from patients with suspected acute bacterial or viral infection, who have had symptoms for less than seven days. The MeMed BV test generates a numeric score that falls within discrete interpretation bins based on the increasing likelihood of bacterial infection.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Z Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

MeMed Diagnostics Ltd.'s MeMed BV®

Submitter

MeMed Diagnostics Ltd. Nahum Het 7 Tirat Carmel, 3508506, Israel Phone: +972-4-8500302 Facsimile: +972-4-8500298 Contact Person: Efrat Hartog-David, Ph.D, Olga Boico, Ph.D Date Prepared: April 4, 2023

Name of Device: MeMed BV®

Common or Usual Name: MeMed BV®

Classification Name: Device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis

Regulatory Class: Class II, 21 CFR 866.3215

Product Code: QPS

Predicate Devices

MeMed Diagnostics LTD., MeMed BV® (K222332)

Device Description

The MeMed BV® ("BV test" or the "test") is an In-Vitro-Diagnostic device that measures in parallel the blood concentrations of TRAIL, IP-10 and CRP. The test consists of an automated analyzer with built-in hardware and software that conduct chemiluminescence based analyte measurements of patient serum and venous whole blood samples and their computational integration (MeMed Key®), and a disposable cartridge that contains the reagents and controls needed to detect the analytes of interest (MeMed BV® cartridge). The test generates an answer to each sample, with a test run time of approximately 15 minutes.

Intended Use / Indications for Use

The MeMed BV® test is an automated semi-quantitative immunoassay that measures three nonmicrobial (host) proteins (TRAIL, IP-10, and CRP) in adult and pediatric serum and venous whole blood samples and is intended for use in conjunction with clinical assessments and other laboratory findings as an aid to differentiate bacterial from viral infection. MeMed BV® is indicated for use in patients presenting to the emergency department or urgent care center and with samples collected at hospital admission from patients with suspected acute bacterial

{4}------------------------------------------------

or viral infection, who have had symptoms for less than seven days. The MeMed BV® test generates a numeric score that falls within discrete interpretation bins based on the increasing likelihood of bacterial infection.

Comparison with Predicate Device

The MeMed BV® is substantially equivalent to the predicate device, MeMed Diagnostics LTD. MeMed BV® (K222332). The FDA cleared MeMed BV® device has similar intended use and indications for use, as well as the same basic technological principles to the predicate device. The described changes in the indications for use (venous whole blood sample type) and technological characteristics (calibration scheme) are supported by the performance testing and do not raise any new questions of safety and efficacy. A substantial equivalence table summarizing the similarities and differences between the MeMed BV® and its predicate device is provided in the table below (MeMed Diagnostics, Ltd.'s MeMed BV® Test Substantial Equivalence Chart).

Intended Use /Indications forUseProposed modified MeMed BV® TestMeMed BV® Test (K222332)
The MeMed BV® test is an automatedsemi-quantitative immunoassay thatmeasures three non-microbial (host)proteins (TRAIL, IP-10, and CRP) in adultand pediatric serum and venous wholeblood samples and is intended for use inconjunction with clinical assessments andother laboratory findings as an aid todifferentiate bacterial from viral infection.MeMed BV® is indicated for use inpatients presenting to the emergencydepartment or urgent care center andwith samples collected at hospitaladmission from patients with suspectedacute bacterial or viral infection, whohave had symptoms for less than sevendays. The MeMed BV® test generates anumeric score that falls within discreteinterpretation bins based on theincreasing likelihood of bacterial infection.The MeMed BV® test is an automated semi-quantitative immunoassay that measuresthree non-microbial (host) proteins(TRAIL, IP-10, and CRP) in adult andpediatric serum samples and is intended foruse in conjunction with clinical assessmentsand other laboratory findings as an aid todifferentiate bacterial from viral infection.MeMed BV® is indicated for use inpatients presenting to the emergencydepartment or urgent care center and withsamples collected at hospital admissionfrom patients with suspected acute bacterialor viral infection, who have had symptoms forless than seven days. The MeMedBV® test generates a numeric score that fallswithin discrete interpretation bins based onthe increasing likelihood of bacterial infection.
User PopulationSameHealth Care Providers requesting samples tobe tested by clinical laboratory technicians
SpecimenHuman serum or Venus whole bloodHuman serum
Assay PrincipleSameSandwich immunoassay technology
Analytes ofInterestSameTRAIL, IP-10, and CRP
AssayTechniqueSameChemiluminescent immunoassay (CLIA)

Table 1. MeMed Diagnostics, Ltd.'s MeMed BV Test Substantial Equivalence Chart

{5}------------------------------------------------

Proposed modified MeMed BV® TestMeMed BV® Test (K222332)
DetectionMethodSameAutomated chemiluminescence-basedanalyte measurement using MeMed Key®Instrument
AssessmentProcessSameSoftware algorithm-based
Test ResultReportingSameNumerical values with risk bins
Time to ResultSameApproximately 15 minutes
CalibrationFrequencySameEvery Four weeks
CalibrationSchemeBackwards compatibility with the legacycalibration scheme (for serum samples)as well as a modified master calibrationcurve scheme (MCC; upon calibration,the factory-derived master curve isadjusted)Legacy calibration scheme
Volume forSample150 µL for venous whole blood100 µL for serum100 µL for serum

{6}------------------------------------------------

Performance Data

1. Analytical performance:

The analytical performance testing supports the two newly introduced elements, the performance of the MeMed BV® test (serum) with the introduction of a new Master Calibration Curve (MCC) and the performance of the MeMed BV® test using venous WB specimen (also with the use of the new calibration scheme). For the first element (serum) only, the raw data that have been collected for previous analytical validation studies (K222332) were utilized for the re-calculation and re-analysis by applying the MCC.

a. Limit of Quantitation

The Total Error and precision for the lowest concentration of each measurand that could be reliably measured (i.e., Limit of Quantification or LoQ) by the MeMed BV® Test was evaluated in accordance with CLSI EP17-A2、Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures. The study used two cartridge lots (per each test script; serum and whole blood) with one MeMed Key® analyzer and the samples described below. Each sample was tested three times on three non-consecutive days.

AnalyteTotal Error Accuracy Goal
TRAILTE < 30%
IP-10TE < 40%
CRPTE < 30%

Table 2. Predefined acceptance criteria for LOQ

The TE (total error) was calculated for each of the four concentration levels for three analytes as 2 x SD (Standard Deviation) observed.

The results obtained for lower limit of quantification ("LLOQ") testing for both the serum sample test script and the Whole Blood sample test script (each tested on two cartridge lots) are summarized in the two tables below.

Table 3. Total Error for LLOQ Measurements For Two Cartridge Lots using the Serum test script

Cartridge LotM23716M23926
SampleParameterTRAIL(pg/ml)IP-10(pg/ml)CRP(mg/L)TRAIL(pg/ml)IP-10(pg/ml)CRP(mg/L)
1(X0.8)CV26%4%7%15%7%3%
TE51%7%15%29%13%6%
2(X0.9)CV5%4%5%6%5%6%
TE10%8%10%13%9%11%
3CV11%6%4%10%7%4%

{7}------------------------------------------------

Cartridge LotM23716M23926
SampleParameterTRAIL(pg/ml)IP-10(pg/ml)CRP(mg/L)TRAIL(pg/ml)IP-10(pg/ml)CRP(mg/L)
(X1.0)TE21%11%9%20%15%8%
4CV7%3%5%7%4%6%
(X1.1)TE14%6%10%14%9%12%

The results for serum script test show that for all the tested samples, MeMed BV® test passes the acceptance criteria of TE (except for TRAIL LLOQ sample 1 (X0.8)), importantly, for the defined LLOQ concentration level of TRAIL, CRP, and IP10 (X1.0., TRAIL 15 pg/mL, CRP 1 mg/L, IP10 100 pg/mL) the results achieved the following maximal TE values: TRAIL 21%, CRP 9%, and IP10 15%.

Cartridge Lot U24140 U24827 TRAIL IP-10 CRP IP-10 CRP TRAIL Sample Parameter (pg/ml) (pg/ml) (mg/L) (pg/ml) (pg/ml) (mg/L) CV 6% 6% 7% 7% 4% 4% 1 (X0.8) 11% TE 15% 13% 7% 8% 12% CV 4% 6% 8% 6% 5% 4% 2 (X0.9) TE 8% 13% 17% 11% 10% 9% 3 CV 3% 7% 5% 5% 5% 4% (X1.0) TE 6% 14% 10% 10% 10% 8% CV 4% 6% 6% 4% 5% 5% 4 TE 7% 12% 13% 8% 10% 10% (X1.1)

Table 4. Total Error for LLOQ Measurements For Two Cartridge Lots using the WB test script

The results for WB script test show that for all the tested samples, MeMed BV® test passed the acceptance criteria of TE. For the defined LLOQ concentration level of TRAIL, CRP, and IP10 (X1.0., TRAIL 15 pg/mL, CRP 1 mg/L, IP10 100 pg/mL) the results achieved the following maximal TE values: TRAIL 10%, CRP 10%, and IP10 14%.

Reproducibility/Precision: b.

The repeatability, intermediate precision and reproducibility studies for each measurand (TRAI/IP-10/CRP) of the MeMed BV® test were conducted using the MeMed Key® Analyzer. The MeMed BV® test score used a panel of 4 scores representing infectious virus, equivocal and noninfectious scores during the studies. Studies were performed in accordance with CLSI EP05-A3 Evaluation of Precision of Quantitative Measurement Procedures.

The serum panel members representing the MeMed BV® test scores used for these studies are described below:

{8}------------------------------------------------

Panel memberSample typeScore
AInfectious serum specimenHigh (Score = 97)
BInfectious serum specimenMedium (Score = 51)
CInfectious serum specimenLow (Score = 1)
DHealthy serum specimenHealthy (Score = 4)

The study was performed in three laboratories. At each site, a single operator preformed tests on two different analyzers using one cartridge lot. Each panel member was run in triplicates on each analyzer, on each day, over 5 non-consecutive days. Calibration was performed on the first day on each analyzer; one calibrator lot was used. External controls were run daily using one lot of ECs.

For each measurand, TRAIL, IP-10, and CRP, the acceptance criteria for measurements was CV ≤ 15 %. This acceptance criteria were not applicable to IP-10 and CRP concentration of healthy specimens since the concentrations were expected to be below the LoQ of IP-10 and CRP assays. The acceptance criterion for the MeMed BV® test score was set at SD < 12.5 score units.

The results of the repeatability, intermediate precision and reproducibility studies for serum samples are summarized below.

RepeatabilityIntermediatePrecisionReproducibility
PanelmemberMeasurandor scoreMeanNSDCV%SDCV%SDCV%
ATRAIL49903.06.13.26.43.47.0
BTRAIL60904.57.44.57.44.67.7
CTRAIL165907.74.77.84.78.45.1
DTRAIL55903.86.93.87.14.37.9
AIP-104759022.54.726.35.527.35.7
BIP-104149022.35.423.65.724.15.8
CIP-101,5749074.74.790.45.7102.96.5
DIP-10100900.00.00.00.00.00.0
ACRP190.19018.99.920.110.621.611.3
BCRP59.8904.88.15.38.85.49.0
CCRP29.5902.48.02.48.12.48.2
DCRP1.0900.00.00.00.00.00.0
AScore98901.2NA1.2NA1.4NA
BScore61906.4NA6.4NA6.6NA
CScore1900.5NA0.5NA0.5NA
DScore9902.0NA2.0NA2.2NA

Table 6. Repeatability, Intermediate precision, and reproducibility results for four serum panel members

{9}------------------------------------------------

The reproducibility results complied with the pre-established acceptance criteria for score and individual analytes.

Unlike serum specimens that can be collected and then thawed for Intermediate Precision (between days variance) or Reproducibility assessment (between sites variance), Whole Blood (WB) specimens are unstable and prone to hemolysis when undergoing a freeze-thaw cycle. Therefore, the WB specimens in this study were tested only as fresh sample and the testing protocol is set accordingly.

Each specimen was analyzed in four runs on five different analyzers using one cartridge lot. The study was performed in one laboratory with a single operator. Calibration was performed on the first day on each analyzer; one calibrator lot was used.

Panel memberSample typeScore
BacterialVenous WBHigh (Score = 100)
EquivocalVenous WBMedium (Score = 38)
ViralVenous WBLow (Score = 1)

Table 7. Patient specimen panel members - Whole Blood samples

SampleParameterAverageSTDCV
ViralTRAIL (pg/mL)200.012.66.3%
IP-10 (pg/mL)557.017.53.1%
CRP (mg/L)19.41.05.3%
Score1.00.0NA
EquivocalTRAIL (pg/mL)38.41.84.6%
IP-10 (pg/mL)273.013.95.1%
CRP (mg/L)9.10.44.0%
Score64.13.0NA
BacterialTRAIL (pg/mL)24.61.76.9%
IP-10 (pg/mL)402.823.35.8%
CRP (mg/L)201.624.112.0%
Score99.90.4NA

Table 8. precision results for four WB panel members

The WB precision results comply with the pre-established acceptance criteria for score and individual analytes. Maximal CV% obtained for CRP Bacterial specimen (12.0%). Maximal score units difference obtained for Equivocal specimen (3).

C. Lot-to-Lot Reproducibility

A lot-to-lot reproducibility study was conducted to estimate lot-to-lot variance, for each MeMed BV® test measurand (TRAIL/IP-10/CRP) and the MeMed BV® test score for the four serum panel members as described above.

The lot-to-lot study was performed on 3 days with one operator at one site using three runs per day

{10}------------------------------------------------

for each of the four serum panel members using two lots of cartridges on one MeMed Key® Analyzer. Two calibration lots were used, one for each cartridge lot. External controls were run daily using one lot of External Control reagents. Since no change in cartridge reagents was introduced, the validation using serum specimens applies also to venous whole blood sample type.

For each of TRAIL, IP-10, and CRP, the acceptance criterion for measurement was set at CV ≤ 15 %. This acceptance criteria are not applicable to IP-10 and CRP concentration of healthy individual since it is expected to be below the LoQ of IP-10 and CRP assays. The acceptance criterion for the score was set at SD < 12.5 score units.

The results of the lot-to-lot reproducibility study are summarized below.

PanelmemberMeasurandor scoreMeanNBetween LotsUpper 95% Confidence Limit
SDCV%SDCV%
ATRAIL45180.00.0NANA
BTRAIL55180.00.0NANA
CTRAIL152187.04.64.2E+022.7E+02
DTRAIL49182.85.79.2E+021.9E+03
AIP-104601827.86.07.6E+021.7E+02
BIP-104001817.84.59.0E+022.3E+02
CIP-101,50218135.29.02.5E+031.6E+02
DIP-10100180.00.0NANA
ACRP184.01819.810.73.5E+021.9E+02
BCRP57.2180.00.0NANA
CCRP27.7180.00.0NANA
DCRP1.0180.00.0NANA
AScore98180.6NA1.6E+011.6E+01
BScore67180.0NANANA
CScore2180.4NA7.5E+013.8E+03
DScore13182.3NA2.6E+022.0E+03

Table 9. Between lots analysis of components of variance

The lot-to-lot reproducibility results comply with the pre-established acceptance criteria for score and individual analytes.

Linearity d.

Linearity of the MeMed BV test for each of the three measurands (TRAIL/IP-10/CRP) was evaluated in accordance with CLSI EP6-Ed2 Evaluation of the Linearity of Quantitative Measurement Procedures. The study was performed in one laboratory with one MeMed Key® Analyzer per cartridge lot, two lots of MeMed BV® cartridges, one lot of calibration reagents and one lot of External Control reagents. Calibration was performed before initiating the study for each cartridge lot. External Controls were run daily.

Five replicates of eleven dilutions of each MeMed BV® test measurand were measured in the linearity study. The order of measurement of the dilution series was random. Eleven dilutions were created by repeated pipetting of a single volume (Y) using a single pipette according to the table below.

{11}------------------------------------------------

DilutionVolume of low positivematerialVolume of high positivematerialTRAIL(pg/mL)IP-10(pg/mL)CRP(mg/L)
110 x Y µL-151001
29 x Y µL1 x Y µL43.529025.9
38 x Y µL2 x Y µL7248050.8
47 x Y µL3 x Y µL100.567075.7
56 x Y µL4 x Y µL129860100.6
65 x Y µL5 x Y µL157.51050125.5
74 x Y µL6 x Y µL1861240150.4
83 x Y µL7 x Y µL214.51430175.3
92 x Y µL8 x Y µL2431620200.2
101 x Y µL9 x Y µL271.51810225.1
11-10 x Y µL3002000250

Table 10. Preparation of dilutions for linearity testing

The measured value was calculated as the average measured concentration of CRP/IP10/TRAIL at each level. Predicted value was established as the calculated value using the fit model at each level (in accordance with chapter 3.6, CLSI EP06 Ed2). The allowable deviation from linearity criterion (ADL) was set to be less than 15% or 10mg/L for CRP, 15% or 10 pg/mL for TRAIL and 20% or 50 pg/mL for IP-10 of the value corresponding to the linear fit (predicted).

For the serum sample testing the measurement procedure shows linearity for all analytes for the interval tested, with deviation from linearity within acceptance criteria.

Lot 1: The maximum observed % deviation from linearity is 6.8% in TRAIL assay.

Lot 2: The maximum observed % deviation from linearity is -6.6% in CRP assay.

For the Whole Blood sample testing, the measurement procedure shows linearity for all analytes for the interval tested, with deviation from linearity within acceptance criteria.

Lot 1: The maximum observed % deviation from linearity is 7.1% in TRAIL assay.

Lot 2: The maximum observed % deviation from linearity is 8.6% in TRAIL assay.

  • Hook Effect e.
    This study was executed using Whole Blood test script. Since the workflow for the serum MeMed BV test was not modified, retest for serum sample type was not required.

A recombinant sample where each analyte was present at the upper limit of quantitation (ULOQ, sample #1) was used as well as three additional samples where each analyte was present at higher concentrations (samples 2-4).

The samples were prepared by spiking protein rich buffer with each of the three measurands (recombinant proteins). This approach was used to generate Sample 1 and 4 as indicated the table below. Sample 2 was prepared by mixing Sample 1 and 4 samples in a ratio of 2/3 and 1/3, respectively. Sample 3 sample was prepared by mixing Level 1 and 4 samples in a ratio of 1/3 and

{12}------------------------------------------------

2/3, respectively. For each concentration level, 3 runs were measured on one MeMed Key® Analyzer, on the same day.

SamplesTRAIL (pg/ml)IP-10 (pg/ml)CRP (mg/L)
Sample 1 (ULOQ)2835,582303
Sample 24786,500372
Sample 36677,307410
Sample 48218,046462

Table 11. Analyte concentrations levels to be tested for hook effect assessment

Hook effect was determined to be excluded if the responses obtained for concentrations up to level 4 were no less than the response obtained for upper limit of quantification (ULoQ). If one or more of the assessed concentration levels deviated from this criterion, hook effect concentration was established as the lowest concentration for which the obtained response was lower than the response corresponding to ULoQ.

For each concentration level the average relative light unit (RLU) signal was calculated and compared aqainst the average response obtained for sample 1 (ULOQ). The results are summarized in the table below.

Table 12. Measurements of high analyte concentration samples on MeMed Key® Analyzer
-----------------------------------------------------------------------------------------------
Measurement by analyzer (RLU)
SampleTRAILIP-10CRP
Sample 1 (ULOQ)3,671,32510,293,6607,109,370
Sample 26,189,60511,987,2508,755,138
Sample 38,621,90013,475,8109,676,356
Sample 410,611,70314,839,40410,900,659

All the concentrations levels show a higher signal than the signal obtained for sample 1 ULOQ. This means that no hook effect is observed for concentrations up to TRAIL – 1,000 pg/mL, IP-10 – 10,000 pg/mL, and CRP – 500 mg/L.

  • f. Carry over
    Carry over was previously evaluated as part of the prior 510(k) submission (K222332). Since the workflow for the serum MeMed BV test was not modified, retest for serum sample type was not required.

Because each specimen tested with the MeMed BV test is processed in a separate disposable cartridge and within the cartridge, each one of the three immunoassays is processed using a separate disposable filtered tip, with a unique tip dedicated to each measurand, the likelihood of carry over between specimens is negligible. For venous whole blood sample type a carry-over study was, nonetheless, conducted to address the low risk of potential carry over.

Sequential runs of ("L") and high score ("H") clinical samples were used in the study. No carry-over was assessed based on 1) the difference between average score of high score sample ran after low

{13}------------------------------------------------

score sample and high score sample baseline average score of no more than 12.5 score units; and 2) the difference between average score of low score sample ran after high score sample and low score sample baseline average score of no more than 12.5 score units.

Two whole blood (WB) samples were run in two sequences 1 and 2 of high to low scores and low to high scores, each sequence on a different MeMed Kev® analyzer: The high to low score and lo to high score are represented in the tables below.

SampleScore
High score99
High score98
High score99
High score97
High score99
Low score6
High score96
Low score7
High score99
Low score8
High score99
Low score5
High score99
Low score6
High score99
Baseline high score98
Test high score98
(Baseline mean) -(Test mean)difference0.0

Table 13: High to low score series

SampleScore
Low score5
Low score5
Low score4
Low score6
Low score4
High score97
Low score7
High score99
Low score6
High score99
Low score6
High score99
Low score6
High score99
Low score6
Baseline Low score5
Test low score6
(Baseline mean) –(Test mean)difference1.4

Table 14: Low to high score series

The maximal difference in score obtained for high score sample (1.4 score unit difference) demonstrates that no carry-over occurred with the MeMed BV test.

Interference/Cross Reactivity g.

Interfering substances and cross-reactants were evaluated as part of the prior 510(k) submission (K222332). Since no change in cartridge reagents (antibody or assay formulation) was introduced, no repeated testing was conducted. The previously submitted data (K222332) shows that the 95% confidence interval for the bias lies within +/-12.5 score units for all the interferants and cross-reactants in the indicated concentrations for both bacterial and viral clinical samples. Thus, it can be concluded that there is no interference or cross-reactivity caused by the tested compounds at the indicated concentrations.

h. Human Anti-Mouse Antibody (HAMA) Interference

{14}------------------------------------------------

Interference of human anti-mouse antibody (HAMA) was evaluated as part of the prior 510(k) submission (K222332). Since no change in cartridge reagents (antibody or assay formulation) was introduced, no repeated testing was conducted. The previously submitted data (K222332) shows that the recovery of TRAIL, IP-10 and CRP are within the predetermined +/- 10% of the sample nominal concentration. The results show that thee assays are tolerant to high HAMA concentrations.

i. Correlation to reference standard

Correlation between serum and WB has been assessed according to the clinical validation plan and covered in the Clinical Validation Report.

For the serum sample type, the purpose of this study was to verify that a strong correlation exists between the new calibration scheme (MCC) and the legacy calibration method. To that end, one hundred serum specimens with known TRAIL, CRP and IP-10 concentration were measured on both configurations on three analyzers, 1 repeat per analyzer. The study was designed in alignment with CLSI standard EP09-C 3rd edition (Measurement Procedure Comparison and Bias Estimation Using Patients Samples, 3rd edition) and the following criteria were used:

  1. The clinically relevant criterion is that in less than 5% of samples, the MCC scores deviate from the legacy calibration scores by an amount that would place the pair of scores in two non-adjacent bins.

  2. The Pearson correlation between the MCC and the legacy calibration scores should be greater than 0.95.

  3. The accuracy requirements are that in the range of score values (0 through 100), the absolute bias incurred by using the MCC method and the legacy calibration method is less than 12.5 units at the "bin" cutoff points (10, 35, 65, 90).

A plot of the regression line with 95% confidence bands, and the identity line (y=x) for reference are provided in the figure below.

Image /page/14/Figure/8 description: The image is a scatter plot that compares a test and reference. The x-axis is labeled 'Reference' and ranges from 0 to 100. The y-axis is labeled 'Test' and ranges from 0 to 80. The plot shows a strong positive correlation between the test and reference, with a Pearson's r value of 1. A Deming Regression Fit line is plotted with the equation -0.58 + 1 * Reference.

Deming Regression Fit ( ^ = 1 )

Figure 1. Deming regression analysis (λ=1)

Deming regression analysis was conducted using 100 observation pairs. The ratio of error variances was set at λ=1. Confidence bands were computed using bootstrap samples and the "BCa"

{15}------------------------------------------------

(accelerated Bias Correction) method for obtaining confidence bounds. The estimated fitted regression line consisting of the estimates of intercept and slope with their 95% confidence intervals are provided in the table below.

ValueLCIUCI
Intercept-0.58-0.93-0.35
Slope1.001.001.00

Table 15.The coefficients and the 95% confidence bounds

LCI- lower confidence interval, UCI- upper confidence interval.

The estimated bias at each cutoff point and the corresponding 95% confidence intervals for the bias at each point are provided in the table below. A plot of the estimated bias over the entire range with the 95% confidence band about the estimated bias is also provided in the below figure.

Table 16. The estimate of bias at the four cut-off points between the bins

LevelBiasLCIUCI
10-0.57-0.90-0.37
35-0.57-0.83-0.39
65-0.56-0.79-0.37
90-0.55-0.82-0.32

Image /page/15/Figure/7 description: The image is a plot that shows the relationship between "Reference" and "Bias". The x-axis represents "Reference", ranging from 0 to 100, while the y-axis represents "Bias", ranging from -0.9 to -0.3. A blue line is plotted on the graph, showing a slight positive correlation between "Reference" and "Bias". The shaded area around the blue line indicates the confidence interval.

Bias plot Deming Regression Fit ( 2 = 1 )

Figure 2. The bias plot over the entire range

The legacy calibration and the MCC methods can be considered equivalent methods for producing BV Scores for populations meeting the current indications for use of the BV Score for indicating bacterial versus viral infection.

This claim is with respect to both the clinically relevant criterion of pairs of scores not falling in nonadjacent bins, as well as the requirements based on the methodology for comparing measurement procedures.

{16}------------------------------------------------

j. Sample In-Use Stability

For the serum-based test, as In-Use stability is an inherent property of the sample type, the originally established stability (submitted under K210254) remains. For the venous whole blood sample type, In in-use stability study was conducted to demonstrate the allowable handling conditions from blood draw to sample input into the cartridge. Stability was assessed for each MeMed BV test measurand (TRAIL/IP-10/CRP) and the MeMed BV test resulting score for four patient whole-blood panel members representing two samples with 'low' scores and two samples with 'high' scores as described in the table below.

The study was performed in one laboratory on four days, one day per panel member. Two MeMed Key® analyzers and one lot of cartridges were used. Calibration was performed at the beginning of the study using one lot of calibration reagents.

Panel memberSample typeScoreNumber of patients
A1, A2Infectious whole-blood specimenHigh (score approximately 95)2
B1, B2Infectious whole-blood specimenLow (score approximately 5)2

Table 17. Patient specimens (panel members)

For each panel member, the incubations listed the table below were performed with the package insert indicated K2-EDTA tube before centrifugation and testing with the MeMed BV® Test. There was one run for each time point performed in parallel on two MeMed Key® analyzers.

Tube #Time at room temp (mins)
10-10
230
360
490
5120
6150

Table 18. Incubation Time at Room Temperature

The mean values, regression lines, confidence intervals and significance level of the difference of the slope from 0 were examined for each of the incubation times. The results show that the minimal acceptable period of time was obtained for TRAIL viral sample 1 of approximately 140 minutes. The formal in-use stability of WB sample type is established to be 120 minutes prior testing on analyzer.

k. Freeze-thaw stability

A study was conducted to validate stability between fresh and frozen serum specimens and was

{17}------------------------------------------------

submitted as part of the original 510(k) submission (K210254).

Importantly, venous whole blood samples are not intended to be frozen, hence such study was not conducted for this sample type.

し Calibrator Traceability

The company conducted metrological traceability testing of the MeMed BV® multi-standard calibrator material to ensure that analytical results used for patient care are accurate as well as consistent over time and when using different devices and systems. There was no change in the process that was provided in the original 510(k) submission for the MeMed BV ® (K210254) except for the introduction of Master calibration curve performed by the manufacturer. Both the master calibrator and user calibrator product are traceable to same secondary and primary reference standards discussed in K210254.

m. Calibrator, External Controls, and Cartridges Stability Testing

MeMed BV calibrators (i.e., CAL1, CAL2, and CAL3), external controls (ECs), and cartridges were previously subjected to real time stability, in-use stability, transportation stability, shelf life validation, stability monitoring, and calibration interval testing (for ECs and calibrators only) in order to show that they maintain their respective performance characteristics over a defined time interval under indicated storage conditions. MeMed BV cartridges were also subjected to calibration interval testing.

The results of this testing were submitted as part of the prior 510(k) submission (K210254 and K222332) and demonstrated that the calibrators, ECs have a shelf life of 3.5 months and the cartridges have a shelf life of 12 months.

Since no change in reagent composition of cartridge, calibrators and external controls was introduced, no repeated testing was conducted.

Clinical Studies

1. Perseverance Clinical Study

The analytical equivalency of the expanded indication MeMed BV® test was established by a prospective, multi-center study ("Perseverance). The Perseverance study was designed following guidance provided in h CLSI standard EP35Ed1E (Assessment of Equivalence or Suitability of Specimen Types) and EP09Ed3cE (Measurement Procedure Comparison and Bias Estimation) and included 216 prospectively recruited subjects from 5 medical centers (2 in the US and 3 in Israel). The study population comprised hospital admitted, emergency department, and urgent care center patients over the age of 90 days, with clinical suspicion of acute bacterial or viral infection. The bins used in the study correspond to the below table. Most patients (66.2%) vielded BV scores falling in the outer bins (0 ≤ s ≤ 10 and 90 ≤ s ≤100, bins 1 and 5). The bin with the lowest representation was bin 2 (10 < score <35) with 22 patients, 10.2% of study population.

Table 19.Sample Score Ranges and Specimen Numbers per Score

{18}------------------------------------------------

Score binInterpretation# of specimen tested
$90 \le s \le 100$High likelihood of bacterial infection72
$65 < s < 90$Moderate likelihood of bacterial infection25
$35 \le s \le 65$Equivocal26
$10 < s < 35$Moderate likelihood of viral infection22
$0 \le s \le 10$High likelihood of viral infection71

The clinical validation report includes three studies: a matrix equivalency study comparing scores measured from paired serum and whole blood samples; a study analyzing the impact of whole blood sample type on score bin assignment and associated test interpretation; and a study simulating the impact of whole blood sample type on the diagnostic accuracy of the test in the Apollo study (NCT04690569), where performance was assessed in comparison to a rigorous reference standard based on etiological adjudication by experts provided with comprehensive patient data. The data from the Apollo study were the basis for the clearance of MeMed BV for serum samples (K210254).

Demonstrating the equivalence of MeMed BV® score in paired venous whole blood and serum samples

Matrix equivalency study

The primary endpoint of analytical equivalency required attainment of the following pre-defined acceptance criteria based on Passing & Bablok regression analysis:

  • . Slope in the range of 0.9-1.1
  • . Intercept in the range of (-5) to 5

Passing& Bablok regression analysis of the BV scores measured in serum versus venous whole blood vielded a slope of 1.00, 95% Cl 0.99-1.00 and intercept of 0.00-0.06, fulfilling both predefined acceptance criteria for analytical equivalency. Therefore, BV score measurements conducted using paired serum and venous whole blood samples run on MeMed Key® were established as analytically equivalent.

{19}------------------------------------------------

Image /page/19/Figure/0 description: This image is a scatter plot that compares WB Score and Serum Score. The x-axis represents Serum Score, and the y-axis represents WB Score, both ranging from 0 to 100. The plot shows a positive correlation between the two scores, with data points clustered along a diagonal line, indicating that as Serum Score increases, WB Score also tends to increase.

Figure 3. Analytical equivalency of score measurements from serum and whole blood samples

Passing-Bablok regression: 1 [0.99-1.00]*X+0 [0.00-0.06]; R2=0.970; n=216

Bin impact analysis

in addition to examining the analytical equivalency of BV scores measured using paired serum versus whole blood samples, the potential impact of sample type on score assignment to bins was assessed for the study population of the Perseverance study. The acceptance criteria were pre-defined as less than 5% of the paired serum and whole blood samples from the Perseverance study demonstrating a score deviation that causes a patient to be assigned to a nonadjacent bin.

Results demonstrated that were no paired samples demonstrating a score deviation that caused the patient to be assigned to a nonadjacent bin.

This bin impact analysis strengthens the conclusion that measurements conducted using paired serum and venous whole blood samples run on MeMed Key® are analytically equivalent.

Comparison to adjudication-based reference standard

A third analysis was performed to assess the impact of matrix difference on performance against an adjudication-based reference standard. For this purpose, BV score and reference standard data from the Apollo study (NCT04690569, clinical study supporting the original MeMed BV submission K210254) were analyzed via simulation. Specifically, serum measurements from the Apollo study were

{20}------------------------------------------------

converted to whole blood measurements using multiple simulations that allow for variability in the conversion based on the observed bias and coefficient of variance. The simulated whole blood measurements were then compared to the adjudication-based reference standard (conducted both for the primary all-inclusive and secondary suspected cohorts).

Diagnostic accuracy was assessed by establishing that the probability of bacterial infection is an increasing function of the BV score (as output by whole blood simulations).

The acceptance criteria for this analysis are similar to the acceptance criteria used for the Apollo study, to support the original MeMed BV (Serum) clearance (K210254):

  • . The Cochran-Armitage (CA) Test for trend with a 2-sided 5% level of significance will be used to reject the null hypothesis that there is no trend of increasing probability of bacterial infection with higher test score for at least 95% of simulations
  • . Likelihood Ratio (LR) with 95% Confidence Interval (CI) should exclude 1 for some bins (preferably Bins 1,2,4,5) for at least 95% of simulations

The diagnostic accuracy of the simulated whole blood BV scores passed both acceptance criteria for the all-inclusive and suspected cohorts.

  • . For the all-inclusive cohort, the Cochran-Armitage p-value was significant (p<0.001) for a 100% of the 100K simulations and the Cl of the LR of exactly 4 bins (specifically bins 1.2,4 and 5) excluded 1 in 100% of the simulations.
  • . For the suspected cohort, the Cochran-Armitage p-value was significant (p<0.001) for a 100% of the 100K simulations and the Cl of the LR of exactly 4 bins (specifically bins 1,2,4 and 5) excluded 1 in 99.98% of the simulations. In 0.02% of the simulations, the Cl of the LR of all 5 bins excluded 1.

The study successfully passed the pre-defined acceptance criteria, validating the diagnostic accuracy of simulated whole blood sample measurements.

In conclusion, the MeMed BV® Whole Blood Clinical Validation documents the results of three studies that together were designed to demonstrate the clinical performance of the proposed modified MeMed BV device in support of expanding the indications for use of the previously cleared MeMed BV (K222332) to include whole blood samples.

Conclusions

Analytical and clinical performance data demonstrate that the proposed modified MeMed BV® (with the extended indication for the use of venous whole blood and modified calibration scheme) is as safe and effective as the FDA Cleared MeMed BV® (K222332). In addition, the technological differences between the proposed modified MeMed BV® test and its predicate devices (K222332) raise no new issues of safety or effectiveness. Thus, the MeMed BV® with the extended indication for the use of venous whole blood is substantially equivalent.

§ 866.3215 Device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis.

Link to an amendment published at 90 FR 55982, Dec. 4, 2025. (a)
Identification. A device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis is identified as an in vitro device intended for the detection and qualitative and/or quantitative measurement of one or more non-microbial analytes in human clinical specimens to aid in the assessment of patients with suspected sepsis when used in conjunction with clinical signs and symptoms and other clinical and laboratory findings.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the device's detailed Indications for Use statement describing what the device detects and measures, the results provided to the user, whether the measure is qualitative and/or quantitative, the clinical indications for which the test is to be used, and the specific population(s) for which the device use is intended.
(2) Premarket notification submissions must include detailed documentation of the device description, including (as applicable), all device components, software, ancillary reagents required but not provided, explanation of the device principle and methodology, and for molecular devices include detailed documentation of the primer/probe sequence, design, and rationale for sequence selection.
(3) Premarket notification submissions must include detailed documentation of applicable analytical studies, such as, analytical sensitivity (Limit of Detection, Limit of Blank, and Limit of Quantitation), precision, reproducibility, analytical measuring range, interference, cross-reactivity, and specimen stability.
(4) Premarket notification submissions must include detailed documentation of a prospective clinical study or, if appropriate, results from an equivalent sample set. This detailed documentation must include the following information:
(i) Results must demonstrate adequate device performance relative to a well-accepted comparator.
(ii) Clinical sample results must demonstrate consistency of device output throughout the device measuring range likely to be encountered in the Intended Use population.
(iii) Clinical study documentation must include the original study protocol (including predefined statistical analysis plan), study report documenting support for the Indications for Use(s), and results of all statistical analyses.
(5) Premarket notification submissions must include evaluation of the level of the non-microbial analyte in asymptomatic patients with demographic characteristics (
e.g., age, racial, ethnic, and gender distribution) similar to the Intended Use population.(6) As part of the risk management activities performed under 21 CFR 820.30 design controls, you must document an appropriate end user device training program that will be offered as part of your efforts to mitigate the risk of failure to correctly operate the instrument.
(7) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling, and a detailed explanation of the interpretation of the limitations of the samples (
e.g., collected on day of diagnosis) must be included in the device's 21 CFR 809.10(b)(10) compliant labeling.