K Number
K210254

Validate with FDA (Live)

Device Name
MeMed BV
Date Cleared
2021-09-01

(215 days)

Product Code
Regulation Number
866.3215
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MeMed BV™ test is an automated semi-quantitative immunoassay that measures three non-microbial (host) proteins (TRAIL, IP-10, and CRP) in adult and pediatric serum samples and is intended for use in conjunction with clinical assessments and other laboratory findings as an aid to differentiate bacterial from viral infection. MeMed BV™ is indicated for use in patients presenting to the emergency department or urgent care center and with samples collected at hospital admission from patients with suspected acute bacterial or viral infection, who have had symptoms for less than seven days. The MeMed BV™ test generates a numeric score that falls within discrete interpretation bins based on the increasing likelihood of bacterial infection.

Device Description

The MeMed BV™ ("BV test" or the "test") is an In-Vitro-Diagnostic device that measures in parallel the blood concentrations of TRAIL, IP-10 and CRP. The test consists of an automated analyzer with built-in hardware and software that conduct chemiluminescence-based analyte measurements of patient serum samples and their computational integration (MeMed Key™), and a disposable cartridge that contains the reagents and controls needed to detect the analytes of interest (MeMed BV™ cartridge). The test generates an answer to each sample, with a test run time of approximately 15 minutes.

AI/ML Overview

The MeMed BV™ test is an automated semi-quantitative immunoassay that measures three non-microbial (host) proteins (TRAIL, IP-10, and CRP) in adult and pediatric serum samples. It's intended for use in conjunction with clinical assessments and other laboratory findings to differentiate bacterial from viral infections in patients with suspected acute bacterial or viral infection, who have had symptoms for less than seven days, and are presenting to the emergency department or urgent care center or at hospital admission. The test generates a numeric score that falls within discrete interpretation bins based on the increasing likelihood of bacterial infection.

Here's a breakdown of the acceptance criteria and the study that proves the device meets them:

1. A table of acceptance criteria and the reported device performance:

The provided document details various analytical performance criteria and associated results. The primary clinical study acceptance criteria are related to the statistical significance of the trend and likelihood ratios.

Acceptance Criteria CategorySpecific Acceptance CriterionReported Device Performance
Analytical Performance
Limit of Quantitation (LoQ)Total Error (TE) for TRAIL < 30%, IP-10 < 40%, CRP < 30%.All tested samples passed the acceptance criteria for TE. Formal LLoQ established at TRAIL - 15 pg/mL, CRP - 1 mg/mL, IP-10 - 100 pg/mL.
Reproducibility/Precision (Measurands)CV ≤ 15% for TRAIL, IP-10, and CRP (excluding healthy specimens for IP-10 and CRP where concentrations are below LoQ).All measurands' CVs were within the acceptance criteria across repeatability, intermediate precision, and reproducibility studies (e.g., TRAIL CVs ranged from 6.6% to 12.7%, IP-10 CVs from 4.0% to 6.2%, CRP CVs from 4.9% to 11.6%).
Reproducibility/Precision (MeMed BV Score)SD < 12.5 score units.All scores' SDs were within the acceptance criteria (Repeatability SDs: 0.3-7.5; Intermediate Precision SDs: 0.3-7.7; Reproducibility SDs: 0.4-9.4).
Lot-to-Lot Reproducibility (Measurands)CV ≤ 15% for TRAIL, IP-10, and CRP (excluding healthy specimens for IP-10 and CRP).All measurands' CVs for between-lots analysis were within the acceptance criteria (e.g., TRAIL CVs 0.0-1.6%, IP-10 CVs 0.0-7.3%, CRP CVs 0.0-1.2%).
Lot-to-Lot Reproducibility (MeMed BV Score)SD < 12.5 score units.All scores' SDs for between-lots analysis were within the acceptance criteria (all 0.0).
LinearityMeasurement bias due to non-linearity < 10% or 10mg/L for CRP, 10% or 10 pg/mL for TRAIL, and 10% or 50 pg/mL for IP-10.The degree of non-linearity for CRP and IP-10 (Lot 1 and Lot 2) was within the acceptance criteria.
Hook EffectResponses obtained for concentrations up to Level 4 were no less than the response obtained for ULoQ.No hook effect was observed for TRAIL up to 1,000 pg/mL, IP-10 up to 10,000 pg/mL, and CRP up to 500 mg/L.
Carry OverDifference between average scores of high/low sequence no more than 12.5 score units.Maximal difference in score was 1 score unit, demonstrating no carry-over.
Interference/Cross ReactivityBias between spiked and non-spiked score results was ± 12.5 score units.The 95% confidence interval for the bias was within +/-12.5 score units for all tested interferents and cross-reactants.
HAMA InterferenceTRAIL, CRP, and IP10 concentrations, when run on clinical serum sample mixed with HAMA positive sample, shall measure concentrations within +/- 10% compared to their nominal concentration.For both HAMA samples, the recovery of TRAIL, IP-10, and CRP was within +/- 10%.
In-Use StabilityMean values, regression lines, confidence intervals, and significance level of the difference of the slope from 0 were examined.Allowable incubation time at room temperature before centrifugation was established at 120 minutes.
Freeze-Thaw StabilityAll scores within the 95% confidence interval are within the same or adjacent score categories and do not result in a move between non-adjacent scores.Frozen and fresh samples demonstrated score results corresponding to the same or adjacent scores within the 95% confidence interval, thus demonstrating equivalency.
Clinical Performance
Primary Objective Cohort: Cochran-Armitage (CA) TestSignificant trend in increasing probability of bacterial infection with higher MeMed BV™ score (p < 0.05).p < 0.0001; trend in increasing probability of bacterial infection was demonstrated.
Primary Objective Cohort: Likelihood Ratio (LR) for bins95% CI of the LR for some bins (Bins 1,2,4,5) excluded 1.95% CI for Bins 1 (0.1-0.2), 2 (0.3-0.6), 4 (2.1-3.1), and 5 (6.3-10.5) excluded 1.
Secondary Objective Cohort: Cochran-Armitage (CA) TestSignificant trend in increasing probability of bacterial infection with higher MeMed BV™ score (p < 0.05).p < 0.0001; trend in increasing probability of bacterial infection was demonstrated.
Secondary Objective Cohort: Likelihood Ratio (LR) for bins95% CI of the LR for some bins (Bins 1,2,4,5) excluded 1.95% CI for Bins 1 (0.0-0.1), 2 (0.0-0.3), 4 (1.5-4.0), and 5 (16.6-37.8) excluded 1.

2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

  • Clinical Study Test Set Sample Size:
    • Primary Objective Cohort: 1016 infectious subjects. This included 476 prospectively recruited adult and pediatric patients and 540 archived cases.
    • Secondary Objective Cohort: 872 subjects (after removing 144 indeterminate cases from the primary cohort).
    • Prospectively recruited patients (subgroup analysis): 476 patients.
    • Archived cases (subgroup analysis): 540 cases.
  • Data Provenance: The study was a prospective, multicenter, observational, and blinded study. Patients were recruited from 11 medical centers (9 in the US and 2 in Israel). Therefore, the data originates from both the United States and Israel, and it includes both prospective and retrospective (archived cases) data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

The document states that the ground truth for the clinical study was established using an "expert adjudication comparator method." It also mentions "experts blinded to C-reactive protein (CRP) and procalcitonin (PCT) values" for the primary objective, and "experts given CRP and PCT values" for the secondary objective.
The number of experts and their specific qualifications (e.g., "radiologist with 10 years of experience") are not explicitly detailed in the provided text.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

The term "expert adjudication comparator method (forced diagnosis for indeterminate cases)" is used for the primary objective, and "expert adjudication comparator method (indeterminate cases removed from analysis)" for the secondary objective.
The specific method of adjudication (e.g., if multiple experts made independent decisions and how discrepancies were resolved like "2+1" or "3+1") is not explicitly stated in the provided text.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

A multi-reader multi-case (MRMC) comparative effectiveness study focusing on human readers' improvement with vs. without AI assistance was not described in the provided text. The clinical study assessed the standalone diagnostic performance of the MeMed BV™ test itself (algorithm only, as an aid to differentiate infection), not its impact on human reader performance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Yes, a standalone performance study was done. The "Apollo Clinical Study" evaluated the diagnostic performance of the MeMed BV™ test (an automated immunoassay with computational integration) in differentiating bacterial from viral infection. The results tabulated in Tables 17 and 18, and Figures 1 and 2, demonstrate the performance of the device's algorithmic output (numeric score and interpretation bins) against the expert-adjudicated reference standard, without human-in-the-loop interaction being part of the primary outcome measurement itself.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

The primary type of ground truth used for the clinical study was expert adjudication. For the primary objective cohort, it involved a "forced diagnosis for indeterminate cases" where experts were blinded to CRP and PCT values. For the secondary objective cohort, expert adjudication was used, but indeterminate cases were removed from the analysis, and experts were given CRP and PCT values.

8. The sample size for the training set:

The provided document does not specify the sample size used for the training set for the MeMed BV™ test's algorithm. The document focuses on the test set used for analytical and clinical validation.

9. How the ground truth for the training set was established:

The document does not provide information on how the ground truth for any potential training set was established. It only details the ground truth establishment method for the clinical study's test set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 1, 2021

MeMed Diagnostics Ltd. Efrat Hartog-David Director of Regulatory Affairs and Quality Assurance Nahum Heth 5 Tirat Carmel, 3508504 Israel

Re: K210254

Trade/Device Name: MeMed BV Regulation Number: 21 CFR 866.3215 Regulation Name: Device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis Regulatory Class: Class II Product Code: QPS Dated: January 29, 2021 Received: January 29, 2021

Dear Efrat Hartog-David:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Kristian Roth, Ph.D. Branch Chief Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210254

Device Name MeMed BV™M

Indications for Use (Describe)

The MeMed BV test is an automated semi-quantitative immunoassay that measures three non-microbial (host) proteins (TRAIL, IP-10, and CRP) in adult and pediatric serum samples and is intended for use in conjunction with clinical assessments and other laboratory findings as an aid to differentiate bacterial from viral infection. MeMed BV is indicated for use in patients presenting to the emergency department or urgent care center and with samples collected at hospital admission from patients with suspected acute bacterial or viral infection, who have had symptoms for less than seven days. The MeMed BV test generates a numeric score that falls within discrete interpretation bins based on the increasing likelihood of bacterial infection.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

MeMed Diagnostics Ltd.'s MeMed BV™

Submitter

MeMed Diagnostics Ltd. Nahum Heth 5 Tirat Carmel, 3508504, Israel

Phone: +972-4-8500302 Contact Person: Efrat Hartog-David, Ph.D.

Date Prepared: January 29, 2021

Name of Device: MeMed BV™

Common or Usual Name: MeMed BV™

Classification Name: Immunoassay for host biomarkers of infection

Regulatory Class: Class II

Product Code: QPS

Predicate Devices

BioMerieux, Inc, VIDAS B.R.A.H.M.S. PCT (PCT) (K162827)

Device Description

The MeMed BV™ ("BV test" or the "test") is an In-Vitro-Diagnostic device that measures in parallel the blood concentrations of TRAIL, IP-10 and CRP. The test consists of an automated analyzer with built-in hardware and software that conduct chemiluminescencebased analyte measurements of patient serum samples and their computational integration (MeMed Key™), and a disposable cartridge that contains the reagents and controls needed to detect the analytes of interest (MeMed BV™ cartridge). The test generates an answer to each sample, with a test run time of approximately 15 minutes.

Intended Use / Indications for Use

The MeMed BV™ test is an automated semi-quantitative immunoassay that measures three nonmicrobial (host) proteins (TRAIL, IP-10, and CRP) in adult and pediatric serum samples and is intended for use in conjunction with clinical assessments and other laboratory findings as an aid to differentiate bacterial from viral infection. MeMed BV™ is indicated for use in patients presenting to the emergency department or urgent care center and with samples collected at hospital admission from patients with suspected acute bacterial or viral infection, who have had symptoms for less than seven days. The MeMed BV™ test generates a numeric score that falls within discrete interpretation bins based on the increasing likelihood of bacterial infection.

Comparison with Predicate Device

The MeMed BV is substantially equivalent to the predicate device the VIDAS B.R.A.H.M.S. PCT (PCT) (K162827). The MeMed BV has similar intended and indications for use, as

{4}------------------------------------------------

well as similar basic technological principles to the predicate device. The minor differences in technological characteristics are supported by the performance testing and do not raise any new questions of safety and efficacy. A substantial equivalence table summarizing the similarities and differences between the MeMed BV and its predicate device is provided below.

Performance Data

    1. Analytical performance:
  • a. Limit of Quantitation

The Total Error and precision for the lowest concentration of each measurand that could be reliably measured (i.e., Limit of Quantification or LoQ) by the MeMed BV test was evaluated in accordance with CLSI EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures. The study used two cartridge lots with one MeMed Key analyzer and the samples described in Table 1. Each sample was tested three times on three non-consecutive days.

AnalyteTotal Error Accuracy Goal
TRAILTE < 30%
IP-10TE < 40%
CRPTE < 30%

Table 1. Predefined acceptance criteria for LOQ

The TE (total error) was calculated for each of the four concentration levels for three analytes as 2 x SD observed.

The results obtained for LLOQ testing on two cartridge lots are summarized in Table 2.

Cartridge LotM21244M21532
SampleParameterTRAIL(pg/ml)IP-10(pg/ml)CRP(mg/L)TRAIL(pg/ml)IP-10(pg/ml)CRP(mg/L)
1MEAN13.490.30.913.980.40.9
1STD0.87.00.11.111.00.1
1CV6%8%7%8%14%8%
1TE11%15%13%16%27%16%
2MEAN15.098.71.015.885.81.0
2STD1.214.00.11.011.60.1
2CV8%14%7%6%13%11%
2TE16%28%14%12%27%22%
3MEAN17.3106.21.116.993.61.1
3STD2.55.40.01.62.80.1
3CV14%5%3%10%3%13%
3TE29%10%7%19%6%26%
4MEAN18.1111.11.218.1103.31.2
4STD1.14.40.11.615.70.1

Table 2. Total error for LLOQ measurements for two cartridge lots

{5}------------------------------------------------

Cartridge LotM21244M21532
SampleParameterTRAIL (pg/ml)IP-10 (pg/ml)CRP (mg/L)TRAIL (pg/ml)IP-10 (pg/ml)CRP (mg/L)
CV6%4%9%9%15%9%
TE12%8%18%18%30%18%

The results show that for all the tested samples, MeMed BV test passes the acceptance criteria of TE. The formal LLOQ is established to values corresponding to TRAIL -15pg/mL, CRP-1 mg/mL, IP-10 - 100 pg/mL as is set in MeMed Key analyzer.

b. Reproducibility/Precision:

The repeatability, intermediate precision and reproducibility studies for each measurand (TRAIL/IP-10/CRP) of the MeMed BV™ test were conducted using the MeMed Key™ Analyzer. The MeMed BV test score used a panel of 4 scores representing infectious bacteria, infectious virus, equivocal and noninfectious scores during the studies. Studies were performed in accordance with CLSI EP05-A3 Evaluation of Precision of Quantitative Measurement Procedures.

The serum panel members representing the MeMed BV test scores used for these studies included the following:

Panel memberSample typeScore
AInfectious serum specimenHigh (Score = 96)
BInfectious serum specimenMedium (Score = 53)
CInfectious serum specimenLow (Score = 1)
DHealthy serum specimenHealthy (Score = 4)

Table 3. Patient Specimen Panel Members

The study was performed in three laboratories. The measurements were performed over 5 nonconsecutive days. At each site, a single operator conducted the tests on two different MeMed Key analyzers using one cartridge lot, with three runs performed each day per panel member. Calibration was performed on the first day for each MeMed Key analyzer using one calibrator lot. External Controls were run daily using one lot of external controls.

For each measurand, TRAIL, IP-10, and CRP, the acceptance criteria for measurements was CV ≤ 15 %. This acceptance criteria were not applicable to IP-10 and CRP concentration of healthy specimens since the concentrations were expected to be below the LoQ of IP-10 and CRP assays. The acceptance criterion for the MeMed BV test score was set at SD < 12.5 score units.

The results of the repeatability, intermediate precision and reproducibility studies are summarized below.

{6}------------------------------------------------

RepeatabilityIntermediatePrecisionReproducibility
PanelmemberMeasurandor scoreMeanNSDCV%SDCV%SDCV%
ATRAIL34.0902.98.52.98.54.112.0
BTRAIL68.0906.08.96.39.37.611.1
CTRAIL266.59017.66.618.16.825.99.7
DTRAIL77.2907.09.17.79.99.812.7
AIP-10930.79040.14.343.24.748.65.2
BIP-10372.39020.05.420.55.521.35.7
CIP-10558.49022.34.024.44.425.84.6
DIP-10101.4906.26.16.26.16.36.2
ACRP126.09010.58.310.58.314.611.6
BCRP63.2905.28.35.79.06.510.2
CCRP60.9904.98.15.38.66.310.4
DCRP1.0900.14.90.15.00.15.0
AScore96.0901.31.31.8
BScore53.4907.57.79.4
CScore0.9900.30.30.4
DScore3.6901.01.21.4

Table 4. Repeatability, Intermediate precision and reproducibility results for four panel members

The reproducibility results complied with the pre-established acceptance criteria for score and individual analytes.

c. Lot-to-Lot Reproducibility

A lot-to-lot reproducibility study was conducted to estimate lot-to-lot variance, for each MeMed BV test measurand (TRAIL/IP-10/CRP) and the MeMed BV test score for the four panel members as described in Table 4 above.

The lot-to-lot study was performed on 3 days with one operator at one site using three runs per day for each of the four panel members using two lots of cartridges on one MeMed Key Analyzer. Two calibration lots were used, one for each cartridge lot. External controls were run daily using one lot of External Control reagents.

For each of TRAIL, IP-10, and CRP, the acceptance criterion for measurement was set at CV ≤ 15 %. This acceptance criteria are not applicable to IP-10 and CRP concentration of healthy individual since it is expected to be below the LoQ of IP-10 and CRP assays. The acceptance criterion for the score was set at SD < 12.5 score units.

PanelmemberMeasurandor scoreMeanNBetween Lots
SDCV%
ATRAIL33.1180.51.6
BTRAIL66.4180.00.0
CTRAIL258.8180.00.0
DTRAIL74.5180.00.0
AIP-10950.11869.77.3

Table 5. Between lots analysis of components of variance

{7}------------------------------------------------

PanelmemberMeasurandor scoreMeanNBetween Lots
SDCV%
BIP-10385.81816.94.4
CIP-10575.21844.47.7
DIP-10100.0180.00.0
ACRP117.6180.80.7
BCRP60.8180.00.0
CCRP58.5180.71.2
DCRP1.0180.00.6
AScore95.8180.00.0
BScore54.2180.00.0
CScore1.0180.00.0
DScore3.8180.00.0

The lot-to-lot reproducibility results comply with the pre-established acceptance criteria for score and individual analytes.

  • d. Linearity
    Linearity of the MeMed BV test for each of the three measurands (TRAIL/IP-10/CRP) was evaluated in accordance with CLSI EP06-A Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach. The study was performed in one laboratory with one MeMed Key Analyzer, two lots of cartridges, one lot of calibration reagents and one lot of External Control reagents. Calibration was performed before initiating the study for each cartridge lot. External Controls were run daily.

Four replicates of eleven dilutions of each MeMed BV test measurand were measured in the linearity study. The order of measurement of the dilution series was random. The eleven dilutions used in the study are represented in Table 6.

{8}------------------------------------------------

Dilution#Volumeof lowpositivematerial[mL]Volumeof highpositivematerial[mL]TRAIL(pg/mL)IP-10(pg/mL)CRP(ug/mL)
12015961
21.80.24930127
31.60.48050052
41.40.611169978
51.20.8142898103
6111731097129
70.81.22041295154
80.61.42351494180
90.41.62661693205
100.21.82971892231
11022901930289

Table 6. Preparation of dilutions for linearity testing

The linearity criterion is that measurement bias due to non-linearity is less than 10% or 10mg/L for CRP, 10% or 10 pg/mL for TRAIL and 10% or 50 pg/mL for IP-10 of the value corresponding to the linear fit. The absolute value thresholds are intended to reflect differences at low concentrations for which defined relative errors result in extremely low absolute value.

Polynomial regression analysis was performed for first-, second- and third-order polynomials.

For CRP and IP-10 Lot 1 and IP-10 Lot 2 significance was identified for second order coefficients. However, the degree of non-linearity is within the acceptance criteria.

e. Hook Effect

A recombinant sample where each analyte was present at the upper limit of quantitation (ULOQ, sample #1) was used as well as three additional samples where each analyte was present at higher concentrations (samples 2-4).

The samples were prepared by spiking protein rich buffer with each of the three measurands (recombinant proteins). This approach was used to generate Sample 1 and 4 as indicated in Table 7. Sample 2 was prepared by mixing Sample 1 and 4 samples in a ratio of 2/3 and 1/3, respectively. Sample 3 sample was prepared by mixing Level 1 and 4 samples in a ratio of 1/3 and 2/3, respectively. For each concentration level, 3 runs were measured on one MeMed Key Analyzer, on the same day.

SamplesTRAIL (pg/ml)IP-10 (pg/ml)CRP (mg/L)
Sample 1 (ULOQ)3006000250
Sample 25337333333
Sample 37678666417
Sample 4100010000500
Table 7. Analyte concentrations levels to be tested for hook effect assessment
------------------------------------------------------------------------------------

{9}------------------------------------------------

Hook effect was determined to be excluded if the responses obtained for concentrations up to level 4 were no less than the response obtained for upper limit of quantification (ULoQ). If one or more of the assessed concentration levels deviated from this criterion, hook effect concentration was established as the lowest concentration for which the obtained response was lower than the response corresponding to ULoQ.

For each concentration level average signal was calculated and compared against the average response obtained for sample 1 (ULOQ). The results are summarized in Table 8.

SampleMeasurement by analyzer (RLUs)
TRAILIP-10CRP
Sample 1(ULOQ)201897966456763656138
Sample 2320976180291444693431
Sample 3444854995087945348636
Sample 46111236109045086125845
SampleAnalyteExpectedValueMeMed KeyAnalyzerDifference% Difference
1CRP100 mg/L98.7 mg/L-1.3 mg/L-1.3%
2CRP100 mg/L98.8 mg/L-1.2 mg/L-1.2%
3PCT100 ng/mL98.7 ng/mL-1.3 ng/mL-1.3%
4PCT100 ng/mL98.8 ng/mL-1.2 ng/mL-1.2%

No hook effect was seen for concentrations up to TRAIL of 1,000 pg/mL, IP-10 of 10,000 pg/mL, and CRP of 500 mg/L.

f. Carry over

Because each specimen tested with the MeMed BV test is processed in a separate disposable cartridge and within the cartridge, each one of the three immunoassays is processed using a separate disposable filtered tip, with a unique tip dedicated to each measurand, the likelihood of carry over between specimens is negligible. A carry over study was, nonetheless, conducted to address the low risk of potential carry over. Sequential runs of ("L") and high score ("H") clinical samples were used in the study. No carry-over was assessed based on 1) the difference between average score of high score sample ran after low score sample and high score sample baseline average score of no more than 12.5 score units; and 2) the difference between average score of low score sample ran after high score sample and low score sample baseline average score of no more than 12.5 score units.

Two clinical samples were run in two sequences 1 and 2 of high to low scores and low to high scores, each sequence on a different MeMed Key analyzer: The high to low score (Table 9) and low to high score (Table 10) are represented below.

{10}------------------------------------------------

Table 9. High to low score seriesTable 10. Low to high score series
1High951Low3
2High882Low3
3High953Low5
4High964Low5
5High955Low4
6Low46High96
7High907Low2
8Low28High94
9High939Low3
10Low310High94
11High9711Low4
12Low712High98
13High9713Low5
14Low314High97
15High9715Low6
Baseline high score mean93.8Baseline low score mean4
Test high score mean94.8Test low score mean4
Difference1Difference0

The maximal difference in score obtained for high score sample (1 score unit difference) demonstrates that no carry-over occurred with the MeMed BV test.

g. Interference/Cross Reactivity

Interfering substances and cross-reactants were evaluated for the MeMed BV test score. Each interferent and cross-reactant was tested using two serum panel members that represented score a 'low' score of approximately 5 and 'high' score of approximately 95. The study was performed in one laboratory using one lot of cartridges, one lot of calibration reagents and one lot of External Control reagents. Each interferant and cross-reactant was tested by 8 repeat runs of spiked and non-spiked sample for each clinical sample using 2 MeMed Key Analyzers.

Master stock interferant/cross reactant solutions were spiked into serum specimens at concentrations of less than 10% v/v of the total sample volume. Detailed descriptions of each interferant/cross-reactant concentration is provided in Table 11 and Table 12. The Interference/cross reactivity was assessed by comparison of the spiked sample Test result to a non-spiked sample Test result (appropriate diluent/buffer was added to the non-spiked sample). Interference was determined as bias between spiked and non-spiked score results was ± 12.5 score units.

{11}------------------------------------------------

InterferentStock concentrationReferencematrixVolume spikedFinal Concentration tested
Acetaminophen (Tylenol)312 mg/dLAqueous5% v/v0.156 mg/mL
Acetyl Salicylic Acid(Aspirin)60 mg/dLEthanol5% v/v0.03 mg/mL
Alcohol100%Aqueous5% v/v0.5% V/V
Amoxicillin108 mg/dLAqueous5% v/v54 µg/mL
Ampicillin150 mg/dLAqueous5% v/v75 µg/mL
Azithromycin22.2 mg/dLEthanol5% v/v11.1 µg/mL
Bilirubin (conjugated)>400 mg/dLAqueous10% v/v0.4 mg/mL
Bilirubin (unconjugated)>400 mg/dL0.1M NaOH10% v/v0.4 mg/mL, in 0.1M NaOH
Caffeine216 mg/dLAqueous5% v/v108 µg/mL
Cetirizine HCL8.7 mg/dLAqueous5% v/v4.35 µg/mL
Dextramethorphan0.0312 mg/dLAqueous5% v/v15.6 ng/mL
Doxycycline36 mg/dLAqueous5% v/v18 µg/mL
HAMA (human α-mouseAb)168 ng/mL>640 (titer)N/AN/AN/A
Hemoglobin>10,000 mg/dLAqueous10% v/v10 mg/ mL, zero spike: 0.9%NaCl
Heparin66,000 U/LAqueous5% v/v3300 U/L
Human serum Albumin(HSA, total protein)powderAqueous10% v/v60 mg/mL, in serum
Ibuprofen (Motrin)438 mg/dLEthanol5% v/v219 µg/mL
Levofloxacin72 mg/dLHCl5% v/v36 µg/mL
Loratidine0.0174 mg/dLEthanol5% v/v87 ng/mL
Nicotine0.1983 mg/dLEthanol5% v/v969 ng/mL
Oxymetazoline HCl0.0012 µg/mLAqueous5% v/v0.0006 µg/mL
Phenylephrine0.06 mg/dLAqueous5% v/v30 ng/mL
Prednisolone0.198 mg/dLEthanol5% v/v1200 ng/mL
Rheumatoid Factor (RF)25 kU/LAqueous2% v/v500 IU/mL
Triglycerides/Triolein>15,000 mg/dLSucrose andNaCl10% v/v15 mg/mL, in Ethanol
Table 11. Panel of interferents for screening1
------------------------------------------------------
  1. Concentrations based on recommended testing concentrations in CLSI EP7-A2 Interference testing in clinical chemistry and CLSI EP37 supplemental tables for interference testing in clinical chemistry)
Cross-reactantStockReconstitutionvolumeReferencematrixVolumespikedFinalConcentrationtested
Recombinant Human 4-1BBLigand/TNFSF9 Protein25 µg250 µLPBS5% v/v50 ng/mL
Recombinant Human CD40Ligand/TNFSF5 (HEK293-expressed)25 µg250 µLPBS5% v/v50 ng/mL
Recombinant HumanLymphotoxin alpha1/beta2Protein25 µg100 µLPBS5% v/v50 ng/mL
Recombinant HumanLymphotoxin alpha2/beta1Protein10 µg100 µLPBS5% v/v50 ng/mL
Recombinant Human TNF-alpha Protein20 µg200 µLPBS5% v/v50 ng/mL
Recombinant Human10 µg100 µLPBS5% v/v50 ng/mL

{12}------------------------------------------------

Cross-reactantStockReconstitutionvolumeReferencematrixVolumespikedFinalConcentrationtested
Lymphotoxin-alpha/TNF-betaProtein
Recombinant HumanAdiponectin/Acrp30 Protein0.266mg/mLNAPBS5% v/v50 ng/mL
Recombinant HumanCXCL13/BLC/BCA-1 Protein25 µg250 µLPBS5% v/v50 ng/mL
Recombinant HumanCXCL5/ENA-78 Protein25 µg250 µLPBS5% v/v50 ng/mL
Recombinant HumanCXCL6/GCP-2 Protein25 µg250 µLPBS5% v/v50 ng/mL
Recombinant HumanCXCL1/GRO alpha Protein10 µg100 µLPBS5% v/v50 ng/mL
Recombinant HumanCXCL3/GRO gamma Protein10 µg100 µLPBS5% v/v50 ng/mL
Recombinant Human IFN-gamma Protein100 µg500 µLDDW5% v/v50 ng/mL
Recombinant Human IL-8/CXCL8 Protein50 µg500 µLPBS5% v/v50 ng/mL
Recombinant HumanCXCL11/I-TAC Protein25 µg250 µLPBS5% v/v50 ng/mL
Recombinant HumanCXCL7/NAP-2 Protein10 µg100 µLPBS5% v/v50 ng/mL
Recombinant HumanCXCL9/MIG Protein10 µg100 µLPBS5% v/v50 ng/mL
Recombinant Human/RhesusMacaque/Feline CXCL12/SDF-1a10 µg100 µLPBS5% v/v50 ng/mL
Recombinant Human/FelineCXCL12/SDF-1 beta aa 22-9310 µg100 µLPBS5% v/v50 ng/mL
Recombinant HumanPentraxin 2/SAP Protein50 µg200 µLDDW5% v/v50 ng/mL
Recombinant HumanPentraxin 3/TSG-14 Protein50 µg100 µLPBS5% v/v50 ng/mL

The data shows that the 95% confidence interval for the bias lies within +/-12.5 score units for all the interferants and cross-reactants in the indicated concentrations for both bacterial and viral clinical samples. Thus, it can be concluded that there is no interference or cross-reactivity caused by the above mentioned compounds at the indicated concentrations.

h. Human Anti-Mouse Antibody (HAMA) Interference

Interference of human anti-mouse antibody (HAMA) was conducted by a dose-response experimental design consisting of 5 serum levels with different amounts of HAMA: Low Pool as described in Table 13 below. The Interference of HAMA was assessed by a comparison of TRAIL, CRP, and IP10 concentrations in the measured serum samples to their nominal concentration values.

Each sample was run on 2 MeMed Key Analyzers with a total of 8 repeats for each sample. The acceptance criterion was that TRAIL, CRP, and IP10 concentrations, when run on clinical serum sample mixed with HAMA positive sample, shall measure concentrations within +/- 10% compared to their nominal concentration. The interference of HAMA on analytes concentration was examined by two independent experiments using two different HAMA-positive samples.

{13}------------------------------------------------

TRAILCRPIP10
SampleMeanmeasuredMeannominalRecovery %Pass/FailMeanmeasuredMeannominalRecovery %Pass/FailMeanmeasuredMeannominalRecovery %Pass/Fail
Sample1Level 1132.3109.51338.6
Level 2103.8107.097%Pass78.184.792%Pass991.81010.298%Pass
Level 381.784.097%Pass59.855.3108%Pass681.7756.890%Pass
Level 457.358.898%Pass32.130.4106%Pass409.7428.396%Pass
Level 535.81.0175.0
Sample2Level 1138.1109.71293.6
Level 2115.5115.7100%Pass83.785.098%Pass967.0977.399%Pass
Level 393.4100.293%Pass60.455.6109%Pass661.1697.395%Pass
Level 479.977.8103%Pass31.930.9103%Pass355.5381.193%Pass
Level 562.31.5101.1

Table 13. HAMA interference testing results

For both HAMA samples the recovery of TRAIL, IP-10 and CRP are within the +/- 10%. The results show that the three assays are tolerant to high HAMA concentrations.

i. In-Use Stability

An in-use stability study was conducted to demonstrate the allowable handling conditions from blood draw to serum sample input into the cartridge. Stability was assessed for each MeMed BV test measurand (TRAIL/IP-10/CRP) and the MeMed BV test resulting score for four patient serum panel members representing two samples with 'low' scores of approximately 5 and two samples with 'high' scores of approximately 95 as described in Table 14.

The study was performed in one laboratory on four days, one day per panel member. Three MeMed Key analyzers and one lot of cartridges were used. Calibration was performed at the beginning of the study using one lot of calibration reagents.

Panel memberSample typeScoreNumber of patients
A1, A2Infectious serum specimenHigh (scoreapproximately 95)2
B1, B2Infectious serum specimenLow (scoreapproximately 5)2

Table 14. Patient specimens (panel members)

For each panel member, the incubations listed Table 15 were performed with the package insert indicated SST tube before centrifugation and testing with the MeMed BV Test. There were three replicate runs for each time point performed in parallel on three MeMed Key analyzers.

SSTTube #Time at room temp before centrifugation (mins)
130 (shortest time for coagulation according to the IFU of BD SST Vacutainer)
260
390

Table 15. Incubation Time at Room Temperature

{14}------------------------------------------------

SSTTube #Time at room temp before centrifugation (mins)
4120
5150

The mean values, regression lines, confidence intervals and significance level of the difference of the slope from 0 were examined for each of the incubation times. The minimum acceptable incubation time was approximately 130 minutes and the incubation time before an observed failure was 120 minutes. Thus, the 120 minute time period at room temperature before centrifygation preparation for use in the MeMed BV test was established as 120 minutes.

j. Freeze-thaw stability

A study was conducted to validate stability between fresh and frozen specimens. This study examined the stability of the MeMed BV test result (score) using 40 paired fresh and frozen specimens as indicated in Table 16. The study was performed in one laboratory. Each sample was tested three times on the same MeMed Key analyzer using one lot of cartridges and one lot of calibrator reagents. Calibration was performed on the first day and repeated after two weeks.

Score binInterpretation# of specimen tested
$90 \le s \le 100$High likelihood ofbacterial infection12
$65 < s < 90$Moderate likelihood ofbacterial infection6
$35 \le s \le 65$Equivocal4
$10 < s <35$Moderate likelihood ofviral infection6
$0 \le s \le 10$High likelihood of viralinfection12

Table 16. Sample Score Ranges and Specimen Numbers per S

The acceptance criteria for equivalency between fresh and frozen specimens for the score was that all scores within the 95% confidence interval are within the same or adjacent score categories and do not result in a move between non-adjacent scores.

Passing Bablok regression was used to compare the means of results from the fresh and the frozen specimens. A 95% confidence interval for the regression was calculated using bootstraping.

The confidence intervals do not intersect the lines representing the adjacent test scores and demonstrates the frozen and fresh samples demonstrated score results corresponding to the same or adjacent scores within the 95% confidence interval, thus demonstrating equivalency between fresh and frozen samples for the MeMed BV test.

k. Calibrator Traceability

The company conducted metrological traceability testing of the MeMed BV™ multi-standard calibrator material to ensure that analytical results used for patient care are accurate as well as consistent over time and when using different devices and systems.

{15}------------------------------------------------

All three analytes are traceable to a standard. The material was produced following ISO Guide 34:2009. The test was performed to assign the TRAIL, CRP and IP10 portion of the Master Lot-Calibrator a concentration that is based on their respective reference standard, in order to assign a value to the MeMed BV Calibrators. The test was performed with a calibration curve.

Lastly, testing was conducted to verify that the released calibrators are able to accurately quantify known samples on the MeMed Key analyzer using MeMed BV cartridges. The Lot was released only if all the parameters comply with the acceptance criteria.

Calibrator, External Controls, and Cartridges Stability Testing 1.

MeMed BV calibrators (i.e., CAL1, CAL2, and CAL3), external controls (ECs), and cartridges were subjected to real time stability, iransportation stability, shelf life validation, stability monitoring, and calibration interval testing (for ECs and calibrators only) in order to show that they maintain their respective performance characteristics over a defined time interval under indicated storage conditions. MeMed BV cartridges were also subjected to calibration interval testing.

The results of this testing demonstrated that the calibrators, ECs have a shelf life of 3.5 months and the cartridges have a shelf life of 12 months.

m. Reference Interval

Reference Interval study, conducted based on CLSI EP28-A3c Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory, demonstrated that:

The lower limit of the score (corresponding to the 2.5th percentile) is 0 [90% Cl: 0-1].

The upper limit of the score (corresponding to the 97.5th percentile) is 46 [90% Cl: 36-75].

Clinical Studies

a. Apollo Clinical Study

The diagnostic performance of the MeMed BV™ test was established by a prospective, multicenter, observational, blinded study. The primary objective of the Apollo clinical study was to establish the diagnostic performance of the MeMed BV™ test for differentiating bacterial from viral infection in patients with suspected acute bacterial or viral infection using expert adjudication comparator method (forced diagnosis for indeterminate cases) with the experts blinded to Creactive protein (CRP) and procalcitonin (PCT) values. A secondary objective was to establish the diagnostic performance of MeMed BV™ for differentiating bacterial from viral infection using expert adjudication comparator method (indeterminate cases removed from analysis) with the experts given CRP and PCT values. This study included 1.016 infectious subjects (476 prospectively recruited adult and pediatric patients and 540 archived cases) from 11 medical centers (9 in the US and 2 in Israel). The study population comprised hospital admitted, Emergency Department

{16}------------------------------------------------

(ED) and urgent care center patients over the age of 90 days, with suspected acute bacterial or viral infection, and healthy subjects.

The bins used in the study correspond to Table 16.

Image /page/16/Figure/2 description: The image shows the text "Primary Objective Cohort (Forced Bacterial Viral Cohort Based on Adjudication Blinded to CRP and PCT)". The text is underlined. The text describes the primary objective cohort.

Image /page/16/Figure/3 description: This image is a decision tree showing the distribution of a study population (n=1016) into different bins based on a score. The population is first divided into "Forced Bacterial Viral Cohort" (n=1016) and "Indeterminate" (n=0). The "Forced Bacterial Viral Cohort" is further divided into five bins based on score ranges: Bin 1 (0 ≤ score ≤ 10, n=478), Bin 2 (10 < score < 35, n=173), Bin 3 (35 ≤ score ≤ 65, n=102), Bin 4 (65 < score < 90, n=94), and Bin 5 (90 ≤ score ≤ 100, n=169). Each bin is then divided into "Bacterial" and "Viral" categories with corresponding counts.

Figure 1: Primary Objective Cohort Disposition

  • Viral includes viral and non-infectious comparator method outcomes

The performance of MeMed BV™ test in differentiating between bacterial and viral infection was assessed in the primary objective cohort using two pre-specified statistical tests:

  • . The Cochran-Armitage (CA) Test demonstrated a significant trend in increasing probability of bacterial infection with higher MeMed BV™ score (p < 0.0001).
  • . The 95% Confidence Interval (CI) of the Likelihood Ratio (LR) for some of the bins (Bins 1,2,4,5) excluded 1 (Table 17).

These statistical tests validate that the higher the score, the higher the likelihood of a bacterial infection and therefore the study successfully achieves its the primary objective.

Score BinNN(ReferenceBacterial)N(ReferenceViral andNon-Infectious)%Patients%ReferenceBacterialPatients%ReferenceViral andNon-InfectiousPatientsLikelihoodRatio(ConfidenceInterval)
90 ≤ score ≤1001691026716.660.439.68.1 (6.3-10.5)
65 < score <909426689.327.772.32.1 (1.3-3.1)
35 ≤ score ≤ 65102168610.015.784.31.0 (0.6-1.7)
10 < score <35173916417.05.294.80.3 (0.2-0.6)
0 ≤ score ≤10478747147.11.598.50.1 (0.0-0.2)
Total1016160856100

Table 17: MeMed BV™ Performance in Primary Objective Cohort

{17}------------------------------------------------

Image /page/17/Figure/0 description: This image shows a flow chart of a study population. The study population starts with n=1016, which then splits into two groups: Suspected Bacterial Viral Cohort (n=872) and Indeterminate (n=144). The Suspected Bacterial Viral Cohort is then divided into five bins based on score ranges: Bin 1 (0 ≤ score ≤ 10, n=458), Bin 2 (10 < score < 35, n=155), Bin 3 (35 ≤ score ≤ 65, n=72), Bin 4 (65 < score < 90, n=65), and Bin 5 (90 ≤ score ≤ 100, n=122). Each bin is further divided into Bacterial and Viral categories with corresponding counts.

Secondary Objective Cohort (Suspected Bacterial Viral Cohort Based on Adjudication NOT

Figure 2: Secondary Objective Cohort Disposition

  • Viral includes viral and non-infectious comparator method outcomes

The performance of MeMed BV™ test in differentiating between bacterial and viral infection was assessed in the secondary objective cohort using two statistical tests:

  • The Cochran-Armitage (CA) Test demonstrated a significant trend in increasing probability of bacterial infection with higher MeMed BV™ score (p < 0.0001).
  • . The 95% Confidence Interval (CI) of the Likelihood Ratio (LR) for some of the bins (Bins 1,2,4,5) excluded 1 (Table 18).

These statistical tests validate that the higher the higher the likelihood of a bacterial infection and therefore the study successfully achieves its secondary objective.

Score BinNN(Reference Bacterial)N(Reference Viral andNon-Infectious)% Patients% ReferenceBacterialPatients% ReferenceViral andNon-InfectiousPatientsLikelihoodRatio(ConfidenceInterval)
90 ≤ score ≤100122992314.081.218.925.0 (16.6-37.8)
65 < score <906519467.529.270.82.4 (1.5-4.0)
35 ≤ score ≤ 65726668.38.391.70.5 (0.2-1.2)
10 < score <35155115417.80.799.40.0 (0.0-0.3)
0 ≤ score ≤10458345552.50.799.30.0 (0.0-0.1)
Total872128744100

Table 18: MeMed BV™ Performance in Secondary Objective Cohort

{18}------------------------------------------------

Table 19 summarizes the results of a bin analysis for the prospectively recruited patients (primary objective), and Table 20 summarizes the results of a bin analysis for the archived cases (primary objective).

Table 19: Diagnostic performance of MeMed BV™ for prospectively recruited patients
(primary objective)
Score BinnnBacterialnViral%Patients%BacterialPatients%ViralPatientsLikelihoodRatio
90 ≤ score ≤1004227158.864.335.712.25 (6.9-21.7)
65 < score <903913268.233.366.73.40 (1.8-6.3)
35 ≤ score ≤ 654610369.721.778.31.89 (1.0-3.6)
10 < score <357777016.29.190.90.68 (0.3-1.4)
0 ≤ score ≤10272426857.11.598.50.10 (0.0-0.3)
Total47661415100
Table 20: Diagnostic performance of MeMed BV™ for archived cases (primary objective)
----------------------------------------------------------------------------------------
Score BinnnBacterialnViral%Patients%BacterialPatients%ViralPatientsLikelihoodRatio
90 ≤ score ≤100127755223.559.140.96.42 (4.9-8.5)
65 < score <9055134210.223.676.41.38 (0.8-2.5)
35 ≤ score ≤ 655665010.410.789.30.53 (0.2-1.2)
10 < score <359629417.82.197.90.09 (0.0-0.4)
0 ≤ score ≤10206320338.21.598.50.07 (0.0-0.2)
Total54099441100

Subgroup Analyses

For both the primary and secondary objective cohort, the MeMed BV™ score increased significantly with the likelihood of bacterial infection irrespective of sex, age, ethnicity, race, preenrollment antibiotics, time from symptom onset, and comorbidities (hypertension, ischemic heart disease, hyperlipidemia, diabetes, and COPD adults). As expected, there was a difference in bacterial prevalence in children versus adults (e.g., in the primary objective cohort 10.6% versus 19.5%, respectively). Nevertheless, both statistical tests of performance were passed for all ages.

Based on the clinical performance as documented in the clinical study, the MeMed BV™ test has a safety and effectiveness profile that is similar to the predicate device.

{19}------------------------------------------------

Conclusions

The MeMed BV™ is as safe and effective as the VIDAS B.R.A.H.M.S. PCT (PCT) (K162827). The MeMed BV™ and the predicate test have the same general intended use and similar indications for use, technological characteristics and principles of operation.

In addition, the minor technological differences between the MeMed BV™ test and its predicate devices raise no new issues of safety or effectiveness. Performance data demonstrate that the MeMed BV™ is as safe and effective as the VIDAS B.R.A.H.M.S. PCT (PCT). Thus, the MeMed BV™ is substantially equivalent.

{20}------------------------------------------------

MeMed BV testVIDAS® B·R·A·H·M·S PCT™ Test(K162827)
Intended Use /Indications forUseThe MeMed BV™ test is an automatedsemi-quantitative immunoassay thatmeasures three non-microbial (host)proteins (TRAIL, IP-10, and CRP) in adultand pediatric serum samples and isintended for use in conjunction withclinical assessments and other laboratoryfindings as an aid to differentiate bacterialfrom viral infection. MeMed BV™ isindicated for use in patients presenting tothe emergency department or urgent carecenter and with samples collected athospital admission from patients withsuspected acute bacterial or viralinfection, who have had symptoms forless than seven days. The MeMed BV™test generates a numeric score that fallswithin discrete interpretation bins basedon the increasing likelihood of bacterialinfection.VIDAS® B·R·A·H·M·S PCT™ (PCT) is anautomated test for use on the instrumentsof the VIDAS® family for the determinationof human procalcitonin in human serum orplasma (lithium heparinate) using the ELFA(Enzyme-Linked Fluorescent Assay)technique.Used in conjunction with other laboratoryfindings and clinical assessments, VIDASB·R·A·H·M·S PCT is intended for use asfollows:• to aid in the risk assessment of criticallyill patients on their first day of ICUadmission for progression to severesepsis and septic shock,• to aid in assessing the cumulative 28-day risk of all-cause mortality forpatients diagnosed with severe sepsisor septic shock in the ICU or whenobtained in the emergency departmentor other medical wards prior to ICUadmission, using a change in PCT levelover time,• to aid in decision making on antibiotictherapy for patients with suspected orconfirmed lower respiratory tractinfections (LRTI) defined as community-acquired pneumonia (CAP), acutebronchitis, and acute exacerbation ofchronic obstructive pulmonary disease(AECOPD) - in an inpatient setting oran emergency department,• to aid in decision making on antibioticdiscontinuation for patients withsuspected or confirmed sepsis.
User PopulationHealth Care Providers requesting samplesto be tested by clinical laboratorytechniciansHealth Care Providers requesting samplesto be tested by clinical laboratorytechnicians
SpecimenHuman serumHuman serum or plasma (lithium heparinate)
Assay PrincipleSandwich immunoassay technologySandwich immunoassay technology
Analytes ofInterestTRAIL, IP-10, and CRPProcalcitonin (PCT)
Assay TechniqueChemiluminescent immunoassay (CLIA)ELFA (Enzyme-Linked Fluorescent Assay)
Detection MethodAutomated chemiluminescence-basedanalyte measurement using MeMed KeyInstrumentAutomated fluorescence-based analytemeasurement using VIDAS instrument
AssessmentProcessSoftware algorithm-basedSoftware algorithm-based
MeMed BV testVIDAS® B·R·A·H·M·S PCT TM Test(K162827)
Test ResultReportingNumerical values with risk binsNumerical values with risk bins
Time to ResultApproximately 15 minutesApproximately 20 minutes
CalibrationFrequencyEvery two weeksEvery 28 days
Volume forSample100 µL200 µL

Table 21. MeMed Diagnostics, Ltd.'s MeMed BV test Substantial Equivalence Chart

{21}------------------------------------------------

§ 866.3215 Device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis.

(a)
Identification. A device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis is identified as an in vitro device intended for the detection and qualitative and/or quantitative measurement of one or more non-microbial analytes in human clinical specimens to aid in the assessment of patients with suspected sepsis when used in conjunction with clinical signs and symptoms and other clinical and laboratory findings.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the device's detailed Indications for Use statement describing what the device detects and measures, the results provided to the user, whether the measure is qualitative and/or quantitative, the clinical indications for which the test is to be used, and the specific population(s) for which the device use is intended.
(2) Premarket notification submissions must include detailed documentation of the device description, including (as applicable), all device components, software, ancillary reagents required but not provided, explanation of the device principle and methodology, and for molecular devices include detailed documentation of the primer/probe sequence, design, and rationale for sequence selection.
(3) Premarket notification submissions must include detailed documentation of applicable analytical studies, such as, analytical sensitivity (Limit of Detection, Limit of Blank, and Limit of Quantitation), precision, reproducibility, analytical measuring range, interference, cross-reactivity, and specimen stability.
(4) Premarket notification submissions must include detailed documentation of a prospective clinical study or, if appropriate, results from an equivalent sample set. This detailed documentation must include the following information:
(i) Results must demonstrate adequate device performance relative to a well-accepted comparator.
(ii) Clinical sample results must demonstrate consistency of device output throughout the device measuring range likely to be encountered in the Intended Use population.
(iii) Clinical study documentation must include the original study protocol (including predefined statistical analysis plan), study report documenting support for the Indications for Use(s), and results of all statistical analyses.
(5) Premarket notification submissions must include evaluation of the level of the non-microbial analyte in asymptomatic patients with demographic characteristics (
e.g., age, racial, ethnic, and gender distribution) similar to the Intended Use population.(6) As part of the risk management activities performed under 21 CFR 820.10(c) design and development, you must document an appropriate end user device training program that will be offered as part of your efforts to mitigate the risk of failure to correctly operate the instrument.
(7) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling, and a detailed explanation of the interpretation of the limitations of the samples (
e.g., collected on day of diagnosis) must be included in the device's 21 CFR 809.10(b)(10) compliant labeling.