K Number
K230847

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-09-29

(185 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Chemotherapy Gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs. The proposed device is non-sterile, with fern green, midnight black and sky blue three colors.

Device Description

The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, nitrile, and tested for use with chemotherapy drugs. The gloves are offered in four sizes: small, medium, large, and extra-large and with fern green, midnight black and sky blue three colors for options to each sizes.

AI/ML Overview

The provided text describes the acceptance criteria and the study that proves Chemotherapy Gloves meets these criteria. The study specifically focuses on the physical properties of the gloves and their resistance to permeation by chemotherapy drugs.

Here’s a breakdown of the requested information:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria / TestReported Device PerformanceStandard Met?
Physical Properties (per ASTM D6319)
Before Aging: Tensile Strength≥ 14 MPaYes
Before Aging: Ultimate Elongation≥ 500%Yes
After Aging: Tensile Strength≥ 14 MPaYes
After Aging: Ultimate Elongation≥ 400%Yes
Freedom from Holes (per ASTM D5151)AQL=2.5Yes
Powder Content (per ASTM D6124)Fern Green: 0.6 mg/glove; Midnight black: 1.1 mg/glove; Sky blue: 1.0 mg/glove (All are ≤ 2 mg/glove, which is the predicate device's performance, implying this is the criterion.)Yes
Chemotherapy Drug Permeation (per ASTM D6978)
Carboplatin (10 mg/ml)> 240 minYes
Carmustine (3.3 mg/ml)Midnight black: 32.6 min; Sky blue: 24.5 min; Fern Green: 24.3 minYes (Reported)
Cisplatin (1.0 mg/ml)> 240 minYes
Cyclophosphamide (20.0 mg/ml)> 240 minYes
Dacarbazine (10 mg/ml)> 240 minYes
Docetaxel (10 mg/ml)> 240 minYes
Doxorubicin HCL (2.0 mg/ml)> 240 minYes
Epirubicin HCI (2 mg/ml)> 240 minYes
Etoposide (20 mg/ml)> 240 minYes
Fluorouracil (50.0 mg/ml)> 240 minYes
Gemcitabine (38 mg/ml)> 240 minYes
Ifosfamice (50 mg/ml)> 240 minYes
Irinotecan (20 mg/ml)> 240 minYes
Methotrexate (25 mg/ml)> 240 minYes
Mitomycin (0.5 mg/ml)> 240 minYes
Mitoxantrone HCl (2 mg/ml)> 240 minYes
Oxaliplatin (5 mg/ml)> 240 minYes
Paclitaxel (6.0 mg/ml)> 240 minYes
Thio Tepa (10.0 mg/ml)Midnight black: 58.7 min; Sky blue: 77.3 min; Fern Green: 67.9 minYes (Reported)
Vincristine Sulfate (1 mg/ml)> 240 minYes
Biocompatibility (per ISO 10993)
Irritation (ISO 10993-10)Not an irritantYes
Skin Sensitization (ISO 10993-10)Not a sensitizerYes
Cytotoxicity (ISO 10993-5)Potentially cytotoxic (However, the submission states that acute systemic toxicity test results show the device is safe, implying the overall biological safety is acceptable.)Yes (Mitigated/Accepted)
Acute Systemic Toxicity (ISO 10993-11)Does not elicit acute systemic toxicity response in the model animalYes

Note: For Carmustine and Thio Tepa, the reported breakthrough times are below 240 minutes, which is explicitly noted with a warning: "Please do not use with Carmustine and Thiotepa." This indicates that these specific drugs do not meet a >240 min criterion, but the performance is clearly reported, and the device is indicated to be used with other drugs, or with a warning for these specific drugs. Therefore, even with lower permeation times for these specific drugs, the device still "meets" acceptance criteria by explicitly disclosing the limitation and providing warnings.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not explicitly state the specific sample sizes used for each individual test (e.g., number of gloves tested for tensile strength or permeation). It mentions that "the proposed device was tested" for permeation, implying a test set was used.

The data provenance is not explicitly stated in terms of country of origin for the testing data or whether it was retrospective or prospective. However, the submitter, 3A Glove Sdn. Bhd., is located in Malaysia, and the submission correspondent, Shanghai Truthful Information Technology Co., Ltd., is in China. This suggests the testing may have been conducted in these regions or by contract labs associated with them. The testing is non-clinical performance testing (bench testing).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This section is Not Applicable as the device is a Chemotherapy Glove and the studies described are non-clinical (bench) performance and biocompatibility tests, not studies requiring expert interpretation of medical images or patient data. Ground truth in this context refers to established scientific or engineering methods as defined by the ASTM and ISO standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This section is Not Applicable for the same reason as point 3. The testing involves objective measurements based on specified standard test methods, not human interpretation that would require adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This section is Not Applicable. The device is a medical glove, not an AI-powered diagnostic or assistive tool for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This section is Not Applicable. The device is a medical glove, not a standalone algorithm.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

The "ground truth" for this device's performance is established by recognized, standardized test methods from organizations like ASTM (American Society for Testing and Materials) and ISO (International Organization for Standardization). These standards define specific procedures and criteria for evaluating physical properties (e.g., tensile strength, elongation), freedom from holes, powder content, and chemotherapy drug permeation. Biocompatibility is also assessed against ISO standards. Therefore, the ground truth is based on:

  • Standardized Test Methods: Adherence to ASTM D6319, ASTM D6124, ASTM D5151, ASTM D6978, ISO 10993-10, ISO 10993-5, and ISO 10993-11.
  • Objective Measurements: Quantitative data obtained through these standardized tests (e.g., MPa for tensile strength, percentage for elongation, mg/glove for powder content, minutes for breakthrough time, qualitative observations for irritation/sensitization/toxicity).

8. The sample size for the training set

This section is Not Applicable. The device is a physical product (a glove), not an artificial intelligence model that requires a training set.

9. How the ground truth for the training set was established

This section is Not Applicable for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 29, 2023

3A Glove Sdn. Bhd. % Boyle Wang General Manager Shanghai Truthful Information Technology Co., Ltd. Room 608, No. 738, Shangcheng Rd., Pudong Shanghai, Shanghai 200120 China

Re: K230847

Trade/Device Name: Chemotherapy Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC, OPJ Dated: September 7, 2023 Received: September 7, 2023

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Allan Guan -S

For Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K230847

Device Name Chemotherapy Gloves

Indications for Use (Describe)

Chemotherapy Gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs. The proposed device is non-sterile, with fern green, midnight black and sky blue three colors.

Chemotherapy DrugConcentrationBreakthrough Detection Time in Minutes (minutes)
Carboplatin10 mg/ml(10,000 ppm)> 240 min.
Carmustine3.3 mg/ml (3,300 ppm)Midnight black:32.6(33.2,32.6,33.0) min.Sky blue:24.5(25.2,24.5,26.1) minFern Green:24.3(28.1,26.3,24.3) min.
Cisplatin1.0 mg/ml(1,000 ppm)> 240 min.
Cyclophosphamide20.0 mg/ml(20,000 ppm)> 240 min.
Dacarbazine10 mg/ml (10,000 ppm)> 240 min.
Docetaxel10 mg/ml (10,000 ppm)> 240 min.
Doxorubicin HCL2.0 mg/ml(2,000 ppm)> 240 min.
Epirubicin HCI2 mg/ml (2,000 ppm)> 240 min.
Etoposide20 mg/ml (20,000 ppm)> 240 min.
Fluorouracil50.0 mg/ml(50,000 ppm)> 240 min.
Gemcitabine38 mg/ml (38,000 ppm)> 240 min.
Ifosfamice50 mg/ml (50,000 ppm)> 240 min.
Ifinotecan20 mg/ml (20,000 ppm)> 240 min.
Methotrexate25 mg/ml (25,000 ppm)> 240 min.
Mitomycin0.5 mg/ml (500 ppm)> 240 min.
Mitoxantrone HCl2 mg/ml (2,000 ppm)> 240 min.
Oxaliplatin5 mg/ml (5,000 ppm)> 240 min.
Paclitaxel6.0 mg/ml(6,000 ppm)> 240 min.
Thio Tepa10.0 mg/ml(10,000 ppm)Midnight black: 58.7(87.2,67.8,58.7)Sky blue: 77.3(77.3,85.8,79.8)Fern Green: 67.9(68.4,67.9,79.2)
Vincristine Sulfate1 mg/ml (1,000 ppm)> 240 min.

Please note that the following drugs have low permeation times:

Midnight black: Carmustine 3.3 mg/ml 32.6 Minutes (min.); Thio Tepa 10.0 mg/ml 58.7 Minutes (min.). Sky blue: Carmustine 3.3 mg/ml 24.5 Minutes (min.); Thio Tepa 10.0 mg/ml 77.3 Minutes (min.). Fern green: Carmustine 3.3 mg/ml 24.3 Minutes (min.); Thio Tepa 10.0 mg/ml 67.9 Minutes (min.). Warning: Please do not use with Carmustine and Thiotepa.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary K230847

This 510(k) Summary is being submitted in accordance with 21 CFR 807.92.

1.0 Submitter's Information

Name: 3A Glove Sdn. Bhd. Address: PTD 2058 & 2059, Jalan Cyber 4, Kawasan Perindustrian Senai (III), 81400 Senai, Johor, Malaysia Phone Number: +60127708756 Contact: Poh Seng Ping Date of Preparation: 2023.09.26

Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 Lujiazui East Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn

2.0 Device Information

Trade name: Chemotherapy Gloves Patient Examination Gloves Common name: Classification name: Non-powdered patient examination glove Model(s): S, M, L, XL

3.0 Classification

LZA.LZC. OPJ Product code: Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital

4.0 Predicate Device Information

Manufacturer: Ever Growth (Vietnam) Co., Ltd. Device: Disposable Powder Free Nitrile Examination Glove, Tested For Use With Chemotherapy Drugs, Disposable Powder Free Nitrile Examination Glove, Tested For Use With Chemotherapy

{5}------------------------------------------------

Drugs, Orange Color 510(k) number: K190860

5.0 Device Description

The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, nitrile, and tested for use with chemotherapy drugs. The gloves are offered in four sizes: small, medium, large, and extra-large and with fern green, midnight black and sky blue three colors for options to each sizes.

6.0 Indication for Use

Chemotherapy Gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs. The proposed device is non-sterile, with fern green, midnight black and sky blue three colors.

Chemotherapy DrugConcentrationBreakthrough DetectionTime in Minutes(minutes)
Midnight black carboplatin10 mg/ml(10,000 ppm)> 240 min.
Carmustine3.3 mg/ml (3,300 ppm)32.6(33.2,32.6,33.0) min.
Cisplatin1.0 mg/ml(1,000 ppm)> 240 min.
Cyclophosphamide20.0 mg/ml(20,000 ppm)> 240 min.
Dacarbazine10 mg/ml (10,000 ppm)> 240 min.
Docetaxel10 mg/ml (10,000 ppm)> 240 min.
Doxorubicin HCl2.0 mg/ml(2,000 ppm)> 240 min.
Epirubicin HCl2 mg/ml (2,000 ppm)> 240 min.
Etoposide20 mg/ml (20,000 ppm)> 240 min.
Fluorouracil50.0 mg/ml(50,000 ppm)> 240 min.
Gemcitabine38 mg/ml (38,000 ppm)> 240 min.
Ifosfamide50 mg/ml (50,000 ppm)> 240 min.
Irinotecan20 mg/ml (20,000 ppm)> 240 min.
Methotrexate25 mg/ml (25,000 ppm)> 240 min.
Mitomycin0.5 mg/ml (500 ppm)> 240 min.
Mitoxantrone HCl2 mg/ml (2,000 ppm)> 240 min.
Oxaliplatin5 mg/ml (5,000 ppm)> 240 min.

{6}------------------------------------------------

Paclitaxel6.0 mg/ml(6,000 ppm)> 240 min.
Thio Tepa10.0 mg/ml(10,000 ppm)58.7(87.2,67.8,58.7)
Vincristine Sulfate1 mg/ml (1,000 ppm)> 240 min.

Please note that the following drugs have low permeation times:

Carmustine 3.3 mg/ml 32.6 Minutes (min.);

Thio Tepa 10.0 mg/ml 58.7 Minutes (min.).

Warning: Please do not use with Carmustine and Thiotepa.

Chemotherapy DrugConcentrationBreakthrough DetectionTime in Minutes(minutes)
Sky blue10 mg/ml(10,000 ppm)> 240 min.
carboplatin10 mg/ml(10,000 ppm)> 240 min.
Carmustine3.3 mg/ml (3,300 ppm)24.5(25.2,24.5,26.1) min.
Cisplatin1.0 mg/ml(1,000 ppm)> 240 min.
Cyclophosphamide20.0 mg/ml(20,000 ppm)> 240 min.
Dacarbazine10 mg/ml (10,000 ppm)> 240 min.
Docetaxel10 mg/ml (10,000 ppm)> 240 min.
Doxorubicin HCL2.0 mg/ml(2,000 ppm)> 240 min.
Epirubicin HCI2 mg/ml (2,000 ppm)> 240 min.
Etoposide20 mg/ml (20,000 ppm)> 240 min.
Fluorouracil50.0 mg/ml(50,000 ppm)> 240 min.
Gemcitabine38 mg/ml (38,000 ppm)> 240 min.
Ifosfamice50 mg/ml (50,000 ppm)> 240 min.
Ifinotecan20 mg/ml (20,000 ppm)> 240 min.
Methotrexate25 mg/ml (25,000 ppm)> 240 min.
Mitomycin0.5 mg/ml (500 ppm)> 240 min.
Mitoxantrone HCI2 mg/ml (2,000 ppm)> 240 min.
Oxaliplatin5 mg/ml (5,000 ppm)> 240 min.
Paclitaxel6.0 mg/ml(6,000 ppm)> 240 min.
Thio Tepa10.0 mg/ml(10,000 ppm)77.3(77.3,85.8,79.8)
Vincristine Sulfate1 mg/ml (1.000 ppm)> 240 min.

Please note that the following drugs have low permeation times:

Carmustine 3.3 mg/ml 24.5 Minutes (min.); Thio Tepa 10.0 mg/ml 77.3 Minutes (min.).

Warning: Please do not use with Carmustine and Thiotepa.

Chemotherapy DrugConcentrationBreakthrough DetectionTime in Minutes(minutes)
Fern Green

{7}------------------------------------------------

carboplatin10 mg/ml(10,000 ppm)> 240 min.
Carmustine3.3 mg/ml (3,300 ppm)24.3(28.1,26.3,24.3) min.
Cisplatin1.0 mg/ml(1,000 ppm)> 240 min.
Cyclophosphamide20.0 mg/ml(20,000 ppm)> 240 min.
Dacarbazine10 mg/ml (10,000 ppm)> 240 min.
Docetaxel10 mg/ml (10,000 ppm)> 240 min.
Doxorubicin HCL2.0 mg/ml(2,000 ppm)> 240 min.
Epirubicin HCI2 mg/ml (2,000 ppm)> 240 min.
Etoposide20 mg/ml (20,000 ppm)> 240 min.
Fluorouracil50.0 mg/ml(50,000 ppm)> 240 min.
Gemcitabine38 mg/ml (38,000 ppm)> 240 min.
Ifosfamice50 mg/ml (50,000 ppm)> 240 min.
Ifinotecan20 mg/ml (20,000 ppm)> 240 min.
Methotrexate25 mg/ml (25,000 ppm)> 240 min.
Mitomycin0.5 mg/ml (500 ppm)> 240 min.
Mitoxantrone HCI2 mg/ml (2,000 ppm)> 240 min.
Oxaliplatin5 mg/ml (5,000 ppm)> 240 min.
Paclitaxel6.0 mg/ml(6,000 ppm)> 240 min.
Thio Tepa10.0 mg/ml(10,000 ppm)67.9(68.4,67.9,79.2)
Vincristine Sulfate1 mg/ml (1,000 ppm)> 240 min.

Please note that the following drugs have low permeation times:

Carmustine 3.3 mg/ml 24.3 Minutes (min.);

Thio Tepa 10.0 mg/ml 67.9 Minutes (min.).

Warning: Please do not use with Carmustine and Thiotepa.

7.0 Technological Characteristic Comparison Table

Table1-General Comparison

ItemSubject DeviceK230847Predicate Device(K190860)Remark
Product CodeLZA,LZC, OPJLZA,LZCSame
Regulation No.21CFR880.625021CFR880.6250Same
ClassIISame
Intended UseChemotherapy Gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves wereA patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves wereSame

{8}------------------------------------------------

tested for use withchemotherapy drugs inaccordance with ASTMD6978-05 StandardPractice for Assessment ofMedical gloves toPermeationbyChemotherapy Drugs.tested for use withchemotherapy drugs inaccordance with ASTMD6978-05 Standard Practicefor Assessment of Medicalgloves toPermeationbyChemotherapy Drugs.
Powdered orpowder freePowder freePowder freeSame
Design FeatureAmbidextrousAmbidextrousSame
SterilityNon-SterileNon-SterileSame
Labeling InformationSingle-use indication,powder free, device color,device name, glove size andquantity, Non-Sterile,a statement of standardASTM D6978-05compliance and asummary of the testingresults.Single-use indication,powder free, device color,device name, glove size andquantity, Non-Sterile, astatement of standardASTM D6978-05compliance and a summaryof the testing results.Similar

Table2 Device Dimensions Comparison

Size
DesignationXSSMLXLTolerance
PredicateDevice(K190860)Length, mm230230230230230min
Width, mm708095110120±10
Thickness, mm:
Finger0.05min
Palm0.05min
DesignationSizeTolerance
SMLXL
Subject DeviceLength, mm220230230230min
Width, mm8095110120±10
Thickness, mm:
Finger0.05min
Palm0.05min
RemarkAnalysis 1

Analysis1: The sizes and tolerances of proposed device are different with those of the predicate, but they all meet the requirements of ASTM D6319-19, so the differences do not raise any new safety or performance questions.

{9}------------------------------------------------

ItemSubject devicePredicate device (K190860)Remark
ColorFERN GREEN/MIDNIGHT BLACK/SKY BLUEWhite, OrangeAnalysis 2
PhysicalPropertiesBefore Aging:Tensile Strength14MPa, min14MPa, minSame
Before Aging:Ultimate Elongation500% min500% minSame
PhysicalPropertiesAfter Aging:Tensile Strength14MPa, min14MPa, minSame
After Aging:Ultimate Elongation400%min400%minSame
Comply with ASTM D6319Comply with ASTM D6319Same
Freedom from HolesBe free from holes when tested in accordance with ASTMD5151 AQL=2.5Be free from holes when tested in accordance with ASTMD5151 AQL=2.5Same
Powder ContentFern Green 0.6 mg/gloveMidnight black 1.1 mg/gloveSky blue 1.0 mg/glove<= 2 mg/gloveSimilar
Chemotherapy Drugs Tested with Minimum Breakthrough Detection Time as Tested per ASTM D 6978carboplatin10 mg/ml(10,000 ppm)> 240 min.NAAnalysis 3
Carmustine3.3 mg/ml (3,300 ppm)Fern Green:24.3(28.1,26.3,24.3) min.Sky blue: 24.5(25.2,24.5,26.1) min.Midnight black:32.6(33.2,32.6,33.0) min.Carmustine (BCNU)3.3 mg/ml:White:11.8 Minutes;Orange:31.6MinutesSimilar
Cisplatin1.0 mg/ml(1,000 ppm)> 240 min.Cisplatin 1.0 mg/ml:>240 MinutesSame
Cyclophosphamide20.0 mg/ml(20,000 ppm)> 240 min.Cyclophosphamide (Cytoxan)20.0 mg/ml: > 240 MinutesSame
DacarbazineDacarbazine (DTIC)Same

{10}------------------------------------------------

10 mg/ml (10,000 ppm)> 240 min.10.0 mg/ml:>240 Minutes
Docetaxel10 mg/ml (10,000 ppm)> 240 min.NAAnalysis 5
Doxorubicin HCL2.0 mg/ml(2,000 ppm)> 240 min.DoxorubicinHydrochloride 2.0 mg/ml:> 240 MinutesSimilar
Epirubicin HCI2 mg/ml (2,000 ppm)> 240 min.NAAnalysis 5
Etoposide20 mg/ml (20,000 ppm)> 240 min.Etoposide (Toposar)20.0 mg/ml:> 240 MinutesSimilar
Fluorouracil50.0 mg/ml(50,000 ppm)> 240 min.Fluorouracil 50.0 mg/ml:>240 MinutesSame
Gemcitabine38 mg/ml (38,000 ppm)> 240 min.NAAnalysis 5
Ifosfamice50 mg/ml (50,000 ppm)> 240 minNAAnalysis 5
Ifinotecan20 mg/ml (20,000 ppm)> 240 min.NAAnalysis 5
Methotrexate25 mg/ml (25,000 ppm)> 240 minNAAnalysis 5
Mitomycin0.5 mg/ml (500 ppm)> 240 min.NAAnalysis 5
Mitoxantrone HCI2 mg/ml (2,000 ppm)> 240 min.NAAnalysis 5
Oxaliplatin5 mg/ml (5,000 ppm)> 240 min.NAAnalysis 5
Paclitaxel6.0 mg/ml(6,000 ppm)> 240 min.Paclitaxel (Taxol) 6.0mg/ml:>240 MinutesSame
Thio Tepa10.0 mg/ml (10,000 ppm)> 240 min.Thio-Tepa10.0Similar

{11}------------------------------------------------

10.0 mg/ml(10,000 ppm)mg/ml:White:16.9
FernGreen:Minutes;
67.9(68.4,67.9,79.2)min.Orange: 72.5 Minutes
Skyblue:
77.3(77.3,85.8,79.8)min.
Midnightblack:
58.7(87.2,67.8,58.7) min.
Vincristine SulfateNAAnalysis 5
1 mg/ml (1,000 ppm)
> 240 min.

Analysis 2: The proposed device has different color to the predicate device, but all proposed devices are conducted the performance and biocompatibility tests, the test results shown that the color difference do not effect the safety and performance of proposed device.

Analysis 3: More Chemotherapy Drugs Tested with proposed devices than those with Predicate device. But all proposed devices are conducted the chemo drugs test, the test results shown that they comply with the requirements of standard ASTM D6978, so the differences do not raise any new safety or performance questions.

ItemProposed devicePredicate deviceRemark
MaterialNitrileNitrileSAME
BiocompatibilityIrritationUnder the conditions of the study, not an irritantComply with ISO10993-10SAME
SensitizationUnder conditions of the study, not a sensitizer.
CytotoxicityUnder the conditions of the study, the device is potentially cytotoxicComply with ISO10993-5Analysis 6
Acute Systemic toxicityUnder the conditions of the study, the device does not elicit acute systemic toxicity response in the model animal.Complies with ISO 10993-11 Third edition 2017-09
Label and LabelingMeet FDA's RequirementMeet FDA's RequirementSAME

Table4 Safety Comparison

Analysis 6: The proposed device is potentially cytotoxic, but all proposed devices are conducted the acute systemic toxicity test, the test results show that the proposed device is safe.

8.0 Summary of Non-Clinical Testing

Biocompatibility Testing

{12}------------------------------------------------

The biocompatibility evaluation for Chemotherapy Gloves was conducted in accordance with the following standards:

ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests for Irritation And Skin Sensitization.

ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity

ISO 10993-11 Third edition 2017-09 Biological evaluation of medical devices -Part 11: Tests for systemic toxicity

Performance Testing (Bench)

Physical performance qualities of the proposed device were evaluated per ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.

Permeation testing was conducted to support the addition of the labeling claim: Tested for use with chemotherapy drugs. The proposed device was tested according to ASTM D6978-05 (Reapproved 2019), Standard Practice for Assessment of Resistance of Medical Gloves to Permeation bv Chemotherapy Drugs, in which minimum breakthrough times were determined for a wide range of chemotherapy drugs.

In summary, the performance testing of the subject device was conducted to adequately demonstrate the effectiveness of the device in accordance with the relevant test methods cited below:

  • ASTM D6124-06 (Reapproved 2017), Standard Test Method for -Residual Powder on Medical Gloves
  • ASTM D5151-19, Standard Test Method for Detection of Holes in ー Medical Gloves.
  • ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.
  • D6978-05 (Reapproved 2019) Standard Practice । ASTM for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

9.0 Summary of Clinical Testing

Clinical testing is not needed for this device.

10.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject

{13}------------------------------------------------

device, Chemotherapy Gloves is as safe, as effective, and performs as well as or better than the legally marketed Predicate device under K190860.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.