K Number
K230461

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-03-24

(31 days)

Product Code
Regulation Number
864.5430
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Quantra® System is composed of the Quantra Hemostasis Analyzer, OPlus Cartridge, and Quantra Quality Controls Level 1 and 2. The Quantra System is intended for in vitro diagnostic use.

The Quantra Hemostasis Analyzer uses Sonic Estimation of Elasticity via Resonance (SEER) Sonorheometry, an ultrasound-based technology, to measure the shear modulus of whole blood during coagulation. The system is intended to be used by trained professionals at the point-of-care and in clinical laboratories to evaluate the viscoelastic properties of whole blood.

The QPlus Cartridge is a multi-channel cartridge that provides sem-quantitative indications of the coagulation state of a 3.2% citrated venous or arterial whole blood sample. The QPlus Cartridge includes tests to assess coagulation characteristics via the intrinsic pathway, via the extrinsic pathway, and includes tests with a heparin neutralizer. The QPlus Cartridge is indicated for use in cardiovascular or major orthopedic surgeries before, and following the procedure.

The QStat Cartridge is a multi-channel cartridge that provides semi-quantitative indications of the coagulation and clot lysis state of a 3.2% citrated venous whole blood sample. The QStat Cartridge includes tests to assess coagulation characteristics via the intrinsic pathway, via the extrinsic pathway, and includes a test with tranexamic acid to evaluate clot lysis characteristics. The QStat Cartridge is indicated for use in trauma and liver transplantation procedures.

The Quantra System is indicated for the evaluation of blood coagulation in perioperative patients age 18 years and older to assess possible hypocoagulable and hypercoagulable conditions. Results obtained with the Quantra System should not be the sole basis for patient diagnosis.

Device Description

The Quantra System is an in vitro diagnostic device designed to assess a patient's coagulation system by measuring the viscoelastic properties of a blood sample during clot formation and lysis in trauma, surgical and intensive care settings. The system consists of the Quantra Hemostasis Analyzer (instrument), single-use disposable cartridges, (OPlus and OStat cartridges) and Quantra Quality Controls (external Quality Control materials).

AI/ML Overview

This document describes a Special 510(k) submission for the HemoSonics Quantra Hemostasis Analyzer. The purpose of the submission is to implement a new operating system (Microsoft Windows 10 IoT Enterprise LTSC 2019) in the device. The FDA determined that this change does not affect the device's intended use or alter its fundamental scientific technology. Therefore, the information provided focuses on the substantial equivalence to the predicate devices (Quantra System cleared under K213917 and K223433), rather than a detailed clinical study demonstrating device performance against new acceptance criteria.

1. Table of Acceptance Criteria and Reported Device Performance

Since this 510(k) is for a software update (operating system change) that does not alter the fundamental scientific technology or intended use, traditional performance acceptance criteria in terms of clinical accuracy or diagnostic efficacy are not presented in this document. The "acceptance criteria" here refer to demonstrating that the new software version maintains the performance of the predicate device.

Aspect of PerformanceAcceptance Criteria (Implicit)Reported Device Performance (Implicit)
FunctionalityThe device with the new operating system (W10IoT) performs the same functions as the predicate device (Windows Embedded Standard 8).The submission asserts that the change does not affect the device's intended use or alter the device's fundamental scientific technology. All functionalities related to measuring shear modulus of whole blood during coagulation, multi-channel cartridge processing (QPlus, QStat), and quality controls are presumed to be maintained.
SafetyThe new operating system does not introduce new safety concerns or compromise existing safety measures.Implicitly deemed safe by the FDA's substantial equivalence determination for this software change. The FDA stated: "We have reviewed your Section 510(k) premarket notification... and have determined the device is substantially equivalent...".
EffectivenessThe device's effectiveness in evaluating blood coagulation, as per its Indications for Use, is maintained.No new clinical effectiveness data presented for this specific 510(k). The equivalence is based on the underlying technology and intended use remaining unchanged from the previously cleared predicate devices.
Software VersionUpdate to Quantra Hemostasis Analyzer Software to v2.2.16 and Embedded Windows Operating System to Microsoft Windows 10 IoT Enterprise LTSC 2019Achieved, as this is the subject of the submission.

2. Sample Size Used for the Test Set and Data Provenance

This submission is for a software update (operating system change) and does not involve a new clinical study with a test set of patient data as would be typical for a de novo device or a significant change in intended use/technology. The document does not specify a "test set" in the context of clinical data. It relies on the performance data from its predicate devices (K213917 and K223433) for which clinical data would have been submitted.

Therefore:

  • Sample size for test set: Not applicable (no new clinical test set for this specific software update).
  • Data provenance: Not applicable for a new clinical test set. The predicate devices' data would have included (presumably) prospective or retrospective clinical data, but its details are not in this document.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Not applicable for this software update submission. No new clinical ground truth establishment described here.

4. Adjudication Method for the Test Set

Not applicable for this software update submission. No new clinical adjudication described here.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not done as this submission pertains to a software (operating system) update, not a change requiring a new clinical evaluation of human reader performance with or without AI assistance. The Quantra System is a diagnostic instrument that measures viscoelastic properties of whole blood; it is not an AI-assisted diagnostic tool for imaging interpretation.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

This is not applicable in the context of a software update for an in vitro diagnostic device like the Quantra System. The device performs automated measurements. Its performance is inherent to the algorithm and instrument design, which are not changing beyond the operating system.

7. The type of ground truth used

Not applicable for this software update submission. The "ground truth" for the original device clearance (K213917, K223433) would have been established through correlation with established clinical methods for assessing coagulation status (e.g., standard coagulation assays, clinical outcomes, expert clinical judgment), but details are not provided in this document.

8. The Sample Size for the Training Set

Not applicable. This submission is about updating the operating system of an already cleared device, not about training a new algorithm with a specific training set.

9. How the ground truth for the training set was established

Not applicable, as no new algorithm training or associated ground truth establishment is described in this submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 24, 2023

HemoSonics, LLC Debbie Winegar Vice-President, Clinical Affairs 4020 Stirrup Creek Drive. Suite 105 Durham, North Carolina 27703

Re: K230461

Trade/Device Name: Quantra Hemostasis Analyzer Regulation Number: 21 CFR 864.5430 Regulation Name: Coagulation system for the measurement of whole blood viscoelastic properties in perioperative patients Regulatory Class: Class II Product Code: OFR Dated: February 21, 2023 Received: February 21, 2023

Dear Debbie Winegar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Min Wu)-

Min Wu, Ph.D. Branch Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K230461

Device Name Quantra Hemostasis Analyzer

Indications for Use (Describe)

The Quantra® System is composed of the Quantra Hemostasis Analyzer, OPlus Cartridge, and Quantra Quality Controls Level 1 and 2. The Quantra System is intended for in vitro diagnostic use.

The Quantra Hemostasis Analyzer uses Sonic Estimation of Elasticity via Resonance (SEER) Sonorheometry, an ultrasound-based technology, to measure the shear modulus of whole blood during coagulation. The system is intended to be used by trained professionals at the point-of-care and in clinical laboratories to evaluate the viscoelastic properties of whole blood.

The QPlus Cartridge is a multi-channel cartridge that provides sem-quantitative indications of the coagulation state of a 3.2% citrated venous or arterial whole blood sample. The QPlus Cartridge includes tests to assess coagulation characteristics via the intrinsic pathway, via the extrinsic pathway, and includes tests with a heparin neutralizer. The QPlus Cartridge is indicated for use in cardiovascular or major orthopedic surgeries before, and following the procedure.

The QStat Cartridge is a multi-channel cartridge that provides semi-quantitative indications of the coagulation and clot lysis state of a 3.2% citrated venous whole blood sample. The QStat Cartridge includes tests to assess coagulation characteristics via the intrinsic pathway, via the extrinsic pathway, and includes a test with tranexamic acid to evaluate clot lysis characteristics. The QStat Cartridge is indicated for use in trauma and liver transplantation procedures.

The Quantra System is indicated for the evaluation of blood coagulation in perioperative patients age 18 years and older to assess possible hypocoagulable and hypercoagulable conditions. Results obtained with the Quantra System should not be the sole basis for patient diagnosis.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Hemosonics. The logo consists of a red blood drop with sound waves emanating from the bottom left. To the right of the blood drop is the word "HEMOSONICS" in black, sans-serif font.

SECTION 5: 510(K) SUMMARY

A. APPLICANT INFORMATION

Submission Date:February 21, 2023
Submitter Information:HemoSonics, LLC4020 Stirrup Creek Drive, Suite 105Durham, NC 27703Phone: 919-244-6990
Contact Person:Deborah Winegar, PhDEmail: dwinegar@hemosonics.comPhone 919-244-6990

B. PROPRIETARY AND ESTABLISHED NAMES

Quantra® System

C. REGULATORY INFORMATION

Trade/Device Name:Quantra System
Regulation Number:21 CFR 864.5430
Regulation Name:Coagulation system for the measurement of whole bloodviscoelastic properties in perioperative patients
Regulatory Classification:Class II
Product Code:QFR

D. PURPOSE OF SUBMISSION

To implement a new operating system (Microsoft Windows 10 IoT Enterprise LTSC 2019) in the Quantra Hemostasis Analyzer. This change does not affect the device's intended use nor alter the device's fundamental scientific technology.

E. MEASURAND

The combination of clot time and clot stiffness parameters measured from the four channels of the cartridge provides information about the functional role of coagulation factors, fibrinogen, and platelets in the sample.

F. TYPE OF TEST

The Quantra System is an in vitro diagnostic device designed to assess a patient's coagulation system by measuring the viscoelastic properties of a blood sample during clot formation and

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for Hemosonics. The logo consists of a red blood drop with sound waves emanating from the bottom left. To the right of the blood drop is the word "HEMOSONICS" in a sans-serif font.

lysis in trauma, surgical and intensive care settings. The system consists of the Quantra Hemostasis Analyzer (instrument), single-use disposable cartridges, (OPlus and OStat cartridges) and Quantra Quality Controls (external Quality Control materials).

G. INTENDED USE/INDICATIONS FOR USE

The intended use/indications for use have not been modified from the intended use/indication for use cleared in K213917 and K223433.

The Quantra® System is composed of the Quantra Hemostasis Analyzer, QPlus Cartridge, QStat Cartridge, and Quantra Quality Controls Level 1 and 2. The Quantra System is intended for in vitro diagnostic use.

The Quantra Hemostasis Analyzer uses Sonic Estimation of Elasticity via Resonance (SEER) Sonorheometry, an ultrasound-based technology, to measure the shear modulus of whole blood during coagulation. The QPlus Cartridge is a multi-channel cartridge that provides semiquantitative indications of the coagulation state of a 3.2% citrated venous or arterial whole blood sample. The QPlus Cartridge includes tests to assess coagulation characteristics via the intrinsic pathway, via the extrinsic pathway, and includes tests with a heparin neutralizer. The OStat Cartridge is a multi-channel cartridge that provides semi-quantitative indications of the coagulation and clot lysis state of a 3.2% citrated venous whole blood sample. The QStat Cartridge includes tests to assess coagulation characteristics via the intrinsic pathway, via the extrinsic pathway, and includes a test with tranexamic acid to evaluate clot lysis characteristics.

The system is intended to be used by trained professionals at the point-of-care and in clinical laboratories to evaluate the viscoelastic properties of whole blood.

The Quantra System is indicated for the evaluation of blood coagulation in perioperative patients age 18 years and older to assess possible hypocoagulable and hypercoagulable conditions. The QPlus Cartridge is indicated for use in cardiovascular or major orthopedic surgeries before, during, and following the procedure. The QStat Cartridge is indicated for use in trauma and liver transplantation procedures.

Results obtained with the Quantra System should not be the sole basis for patient diagnosis.

H. DEVICE MODIFICATION DESCRIPTION

The Quantra System was previously cleared under K213917 and K223433. HemoSonics is submitting this Special 510(k) to implement a new operating system (W10IoT) in the Quantra Hemostasis Analyzer.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Hemosonics. The logo consists of a red blood drop with sound waves emanating from it on the left, and the word "HEMOSONICS" in black text on the right. The text is in a sans-serif font.

SUBSTANTIAL EQUIVALENCE INFORMATION

Predicate Device Name: Quantra System

K223433, K213917 Predicate 510(k) Number:

Comparison with the Predicate:

Table 5-1 provides an overall comparison of the modified Quantra System with the previously cleared Quantra System.

Table 5-1: Comparison between Quantra System (K223433, K213917) and Modified Quantra System

Modified DevicePredicate Device
Quantra System(Subject of Special 510(k))Quantra System(K223433, K213917)
Similarities
ManufacturerSame as predicate deviceHemoSonics, LLC
Trade NameSame as predicate deviceQuantra Hemostasis Analyzer
Common NameSame as predicate deviceWhole Blood Hemostasis System
Classification NameSame as predicate deviceCoagulation system for the measurement of whole bloodviscoelastic properties in perioperative patients.
Regulation NumberSame as predicate device21 CFR 864.5430
Product CodeSame as predicate deviceQFR (K213917, K223433)
Device ClassSame as predicate deviceII
Indications for UseSame as predicate deviceThe Quantra® System is composed of the Quantra HemostasisAnalyzer, QPlus Cartridge, QStat Cartridge, and Quantra QualityControls Level 1 and 2. The Quantra System is intended for in vitrodiagnostic use.The Quantra Hemostasis Analyzer uses Sonic Estimation ofElasticity via Resonance (SEER) Sonorheometry, an ultrasound-based technology, to measure the shear modulus of whole bloodduring coagulation. The system is intended to be used by trainedprofessionals at the point-of-care and in clinical laboratories toevaluate the viscoelastic properties of whole blood.The QPlus Cartridge is a multi-channel cartridge that providessemi-quantitative indications of the coagulation state of a 3.2%citrated venous or arterial whole blood sample. The QPlusCartridge includes tests to assess coagulation characteristics via theintrinsic pathway, via the extrinsic pathway, and includes tests witha heparin neutralizer. The QPlus Cartridge is indicated for use incardiovascular or major orthopedic surgeries before, during, andfollowing the procedure.The QStat Cartridge is a multi-channel cartridge that provides semi-quantitative indications of the coagulation and clot lysis state of a3.2% citrated venous whole blood sample. The QStat Cartridgeincludes tests to assess coagulation characteristics via the intrinsicpathway, via the extrinsic pathway, and includes a test with

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Hemosonics. The logo consists of a red blood drop with sound waves emanating from it on the left, and the word "HEMOSONICS" in black capital letters on the right. Below the logo is the word "mille" in a smaller font.

Quantra® QPlus® System Special 510(k)
Device Modification to K213917, K223433

Modified DeviceQuantra System(Subject of Special 510(k))Predicate DeviceQuantra System(K223433, K213917)
tranexamic acid to evaluate clot lysis characteristics. The QStatCartridge is indicated for use in trauma and liver transplantationprocedures.
The QStat Cartridge is a multi-channel cartridge that provides semi-quantitative indications of the coagulation and clot lysis state of a3.2% citrated venous whole blood sample. The QStat Cartridgeincludes tests to assess coagulation characteristics via the intrinsicpathway, via the extrinsic pathway, and includes a test withtranexamic acid to evaluate clot lysis characteristics. The QStatCartridge is indicated for use in trauma and liver transplantationprocedures.
The Quantra System is indicated for the evaluation of bloodcoagulation in perioperative patients age 18 years and older toassess possible hypocoagulable and hypercoagulable conditions.Results obtained with the Quantra System should not be the solebasis for patient diagnosis.
Intended UseSame as predicate deviceSame as indications for use
DisposablesSame as predicate deviceQPlus Cartridge (multichannel cartridge)QStat Cartridge (multichannel cartridge)Quantra Quality Controls (Level 1 and Level 2)
Quantra Hemostasis AnalyzerHS-002
Analyzer HardwareSame as predicate device
Differences
Quantra HemostasisAnalyzer Softwarev2.2.16v2.0.36.1 (K213917)v2.1.37 (K223433)
EmbeddedWindows OperatingSystemMicrosoft Windows 10 IoTEnterprise LTSC 2019Windows Embedded Standard 8

§ 864.5430 Coagulation system for the measurement of whole blood viscoelastic properties in perioperative patients.

(a)
Identification. A coagulation system for the measurement of whole blood viscoelastic properties in perioperative patients is an in vitro diagnostic device used to evaluate blood coagulation, fibrinolysis, or both, in perioperative patients, as an aid in the assessment of coagulopathies when used in conjunction with clinical signs and symptoms and other clinical and laboratory findings.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include detailed documentation of, and results from, the following:
(i) A study assessing precision using protocols determined to be acceptable by FDA, to cover the measurement range for each reported parameter (test output). Testing must include native specimens with coagulation profiles representative of the intended use population. In order to cover the measuring range, testing may include a limited number of contrived specimens, not to exceed 10 to 20 percent, or as otherwise deemed appropriate by FDA. The contrived specimens must be prepared to resemble clinical specimens. This testing must evaluate repeatability and reproducibility and provide assessments of within-run, within-day, between-run, between-day, between-reagent lot, between-instrument, between-site, and between-operator precision, as applicable to the system;
(ii) Studies that demonstrate the performance of each parameter (test output) throughout the claimed measurement range, to include linearity studies or dose-response studies, as applicable to the parameter (test output);
(iii) Potential interferent study that includes evaluation of hemolyzed and lipemic samples as potential interferents; exogenous and endogenous interferents associated with each patient population intended for use with the device, and which might be expected to affect assay performance, must be evaluated; and potential interferents that are specific for, or related to, the technology or methodology of the device. Evaluation of all potential interferents must be performed using a protocol determined to be acceptable to the FDA (
e.g., an FDA-recognized standard) and include both normal and abnormal specimens covering coagulation profiles representative of the intended use population;(iv) A study that evaluates specimen stability under the intended conditions for specimen collection, handling, and storage, using samples that cover the coagulation profiles representative of the intended use population, and using protocols determined to be acceptable by FDA;
(v) A multisite clinical study, determined to be acceptable by FDA, demonstrating performance, relative to clinically relevant and clinically validated laboratory test(s) for each parameter (test output). Further, the study must meet all of the following criteria:
(A) The study must be performed in the intended use population and include representation from all patient populations for whom the device is intended to be used. Potential endogenous and exogenous interferents for each target patient population must be evaluated or known prior to the study;
(B) The study must be conducted at a minimum of three external sites representative of the intended use setting by the intended operators;
(C) Test samples must be collected at time intervals relevant to the device's use in the intended use population;
(D) Clinical specimens, which cover coagulation profiles representative of the intended use population, must be evaluated at each of the three clinical sites in the study;
(E) Analysis of the concordance of clinical interpretation of patient coagulation status made from individual test parameter (test output) results as compared to clinical interpretation of coagulation status from a clinically relevant laboratory test or tests (
e.g., a comparative viscoelastic device or standard laboratory tests) must be conducted; and(F) Expected (reference) values for each parameter (test output) must be demonstrated by testing a statistically appropriate number of samples from apparently healthy normal individuals;
(vi) For a device with a user interface that has information that needs to be interpreted by the user in correctly using the device to achieve the intended test results or a device that does not provide a final output that is a comprehensive interpretation of all parameter (test output) results, a study evaluating the ability of device users to correctly interpret results;
(vii) For any device indicated to guide blood product use, a clinical outcome study determined to be acceptable by FDA that specifically validates the device's indicated use in guiding blood product use; and
(viii) For any device indicated to guide use of medication, a clinical outcome study determined to be acceptable by FDA that specifically validates the device's indicated use in guiding use of medication.
(2) The labeling required under § 809.10(b) of this chapter must include the following:
(i) A summary of results from the study required by paragraph (b)(1)(i) of this section, including repeatability, reproducibility, and assessments of within-run, within-day, between-run, between-day, between-reagent lot, between-instrument, between-site, and between-operator precision, as applicable to the system.
(ii) The claimed measurement range of each parameter (test output), as supported by demonstrated performance of the parameter (test output) throughout the claimed measurement range, including studies required by paragraphs (b)(1)(i) through (iii) and (v) of this section, and, if applicable, paragraphs (b)(1)(vii) and (viii) of this section.
(iii) Identification of known interferents, including all endogenous, exogenous, technology-specific, and patient population-specific interferents, specific to each parameter (test output). The information must include the concentration(s) or level(s) at which interference was found to occur and the concentration range or levels at which interference was not found to occur.
(iv) Information regarding the multisite clinical study required by paragraph (b)(1)(v) of this section, including:
(A) Each patient population evaluated;
(B) Each intended use setting and the operators;
(C) A summary of the results, including the concordance analysis to clinically relevant laboratory test(s); and
(D) Demonstrated expected (reference) values for each parameter (test output).
(3) The labeling required under § 809.10 of this chapter must include the following:
(i) A limiting statement that the result(s) from the device is(are) not intended to be used as the sole basis for a patient diagnosis.
(ii) Unless appropriate clinical outcome studies are done in accordance with paragraph (b)(1)(vii) of this section that specifically validate an indication for the device's use in guiding blood product use, a limiting statement that the device has not been evaluated to guide blood product use.
(iii) Unless appropriate clinical outcome studies are done in accordance with paragraph (b)(1)(viii) of this section that specifically validate an indication for the device's use in guiding use of medication, a limiting statement that the device has not been evaluated to guide use of medication.