K Number
K230423

Validate with FDA (Live)

Date Cleared
2024-02-23

(371 days)

Product Code
Regulation Number
868.1860
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This device is intended for monitoring PEF (Peak Expiratory Flow) for patient home use. The device is designed for children 5 years of age or older, adolescent and adult subjects.

Device Description

The device is made up of two elements – the Electronic Peak Flow Meter and a Mobile Medical Application for smartphones that communicate via Bluetooth. The device is a portable device that can measure the Peak Expiratory Flow (PEF). The Electronic Peak Flow Meter consists of the main unit, turbine, and blowing mouthpiece.

Electronic Peak Flow Meter software is embedded in the main control unit, which is used to control the whole system of operation. After compiling, the program is solidified into the hardware. The combination of software and hardware realizes the functions for electronic peak flow meter.

When Electronic Peak Flow Meter connected to APP of communication devices (such as a mobile) through Bluetooth. The changes of the patient's PEF data within one week can be monitored by APP, so that it is more convenient to monitor the respiratory health status.

AI/ML Overview

This document is a 510(k) premarket notification for an Electronic Peak Flow Meter (model DL-DF01). It states that performance testing was conducted according to the ATS/ERS Task Force: "Standardisation of lung function testing - Standardisation of spirometry 2005" to evaluate accuracy and repeatability of PEF measurements.

Here's a breakdown of the requested information based on the provided text, recognizing that some details directly related to the study's execution and ground truth might not be explicitly stated in this high-level summary:

  • 1. A table of acceptance criteria and the reported device performance
CriteriaAcceptance Criteria (from predicate)Reported Device Performance (Proposed Device DL-DF01)
Accuracy±10 L/min or ±10%, whichever is greaterPEF ± 20 L/min or PEF ± 10% of the reading
Repeatability<5% or 10 L/min, whichever is greater<5% or 10 L/min, whichever is greater

Note: The document lists the predicate's accuracy as "±10 L/min or ±10%, whichever is greater" for both PEF and FEV1. For the proposed device, it only lists PEF accuracy as "PEF ± 20 L/min or PEF ± 10% of the reading," which is a different (and less stringent) accuracy specification compared to the predicate, but it does state that "Measurements of PEF meet the requirements in the standard." The exact performance values beyond meeting the standard are not explicitly detailed for the proposed device, but it explicitly states the device meets the requirements of the standard for PEF measurements for accuracy and repeatability.

  • 2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
    The document does not explicitly state the sample size for the test set or the data provenance (country of origin, retrospective/prospective). It only mentions "Performance testing was conducted on the Electronic Peak Flow Meter for the measurement of peak expiratory flow (PEF)."

  • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
    Not applicable. For a peak flow meter, the ground truth is typically established by objective measurements from a calibrated spirometer/flow source, not expert interpretation.

  • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    Not applicable. Adjudication is not typically used for objective device performance testing of a peak flow meter. Performance is compared against a reference standard.

  • 5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This device is an Electronic Peak Flow Meter, which is a measurement device, not an AI-assisted diagnostic tool requiring MRMC studies.

  • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    The performance testing described is a standalone evaluation of the device's measurement capabilities. The text states: "Performance testing was conducted on the Electronic Peak Flow Meter for the measurement of peak expiratory flow (PEF)."

  • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
    The ground truth for performance testing (accuracy and repeatability) of a peak flow meter is established by a reference standard, typically a highly accurate and calibrated flow generator or spirometer that simulates breaths with known PEF values. The study followed the "ATS/ERS Task Force: Standardisation of lung function testing - Standardisation of spirometry 2005" which defines the methodologies for these ground truth measurements.

  • 8. The sample size for the training set
    Not applicable. This device is a measurement instrument, not an AI/ML device that requires a training set. The embedded software "is used to control the whole system of operation" and "is solidified into the hardware" after compiling.

  • 9. How the ground truth for the training set was established
    Not applicable. This device is a measurement instrument, not an AI/ML device that requires a training set and corresponding ground truth.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

February 23, 2024

Taian Dalu Medical Instrument Co., Ltd. % Liz Li Counselor Shenzhen Joyantech Consulting Co., Ltd. 1713A, 17th Floor, Block A, Zhongguan Times Square, Liuxian Avenue, Xili Town, Nanshan District Shenzhen, Guangdong 518000 China

Re: K230423

Trade/Device Name: Electronic Peak Flow Meter Regulation Number: 21 CFR 868.1860 Regulation Name: Peak-Flow Meter For Spirometry Regulatory Class: Class II Product Code: BZH Dated: January 11, 2024 Received: January 19, 2024

Dear Liz Li:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Rachana Visaria -S

Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and

{2}------------------------------------------------

Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K230423

Device Name Electronic Peak Flow Meter

Indications for Use (Describe)

This device is intended for monitoring PEF (Peak Expiratory Flow) for patient home use. The device is designed for children 5 years of age or older, adolescent and adult subjects.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

1. Contact Details

1.1 Applicant information

Applicant NameTaian Dalu Medical Instrument Co., Ltd.
AddressWest Part of Yitianmen Street, Hi-tech Zone, Taian271000, Shandong, P.R. China
Phone No.+86-538-8931961-8006
Contact personWang Yuqin
Contact person's e-mailmobilepef@163.com
Date PreparedFeb.15. 2023
1.2 Submission Correspondent
Correspondent NameShenzhen Joyantech Consulting Co., Ltd.
Address1713A, 17th Floor, Block A, Zhongguan Times Square,Liuxian Avenue, Xili Town, Nanshan District, Shenzhen,Guangdong Province, China
Phone No.+86 755-86069197
Contact personLiz Li
Contact person's e-mailliz@cefda.com; grace@cefda.com
Websitehttp://www.cefda.com

2. Device information


Trade nameElectronic Peak Flow Meter
Device ClassII
Classification namePeak-flow meter for spirometry
Product codeBZH
Regulation No.21 CFR 868.1860

3. Legally Marketed Predicate Device

Trade NameMSA100BT Peak Flow Meter
510(k) NumberK170281
Product CodeBZH
ManufacturerBeijing M&B Electronic Instruments Co., Ltd.

{5}------------------------------------------------

4. Device Description

The device is made up of two elements – the Electronic Peak Flow Meter and a Mobile Medical Application for smartphones that communicate via Bluetooth. The device is a portable device that can measure the Peak Expiratory Flow (PEF). The Electronic Peak Flow Meter consists of the main unit, turbine, and blowing mouthpiece.

Electronic Peak Flow Meter software is embedded in the main control unit, which is used to control the whole system of operation. After compiling, the program is solidified into the hardware. The combination of software and hardware realizes the functions for electronic peak flow meter.

When Electronic Peak Flow Meter connected to APP of communication devices (such as a mobile) through Bluetooth. The changes of the patient's PEF data within one week can be monitored by APP, so that it is more convenient to monitor the respiratory health status.

5. Intended use

This device is intended for monitoring PEF (Peak Expiratory Flow) for patient home use. The device is designed for children 5 years of age or older, adolescent and adult subjects.

6. Substantial Equivalence Comparison

{6}------------------------------------------------

ItemProposed DevicePredicateDevice:(K170281)Comments
Device name and modelElectronic Peak FlowMeter (DL-DF01)MSA100BT PeakFlow Meter
Regulation number21 CFR 868.186021 CFR868.1860Same
ClassificationIIIISame
Product codeBZHBZHSame
Intended use/Indicationsfor useThis device is intendedfor monitoring PEF(Peak Expiratory Flow)for patient home use.The device is designedfor children 5 years ofage or older,adolescent and adultsubjects.This device isintended formonitoring PEF(Peak ExpiredFlow Rate) andFEV1 (ForcedExpiratoryVolume in onesecond) forpatient homeuse.The deviceis designed forpediatric to adultpatients.The deviceis intended formonitoringrespiratoryconditions suchas asthma.Similar
Patient PopulationThe device is designedfor children greaterthan five years of age,adolescent and adultsubjects.The device isdesigned forpediatric to adultpatientsSimilar
OTC or RxOTCOTCSame
MaterialsPolypropylene(mouthpiece)Polypropylene(mouthpiece)Same
Measuring rangePEF from 50 to 900 L/minPEF from 50 to 900 L/minFEV1 from 0.01 to 9.99 LSame
Measuring methodFlow: Turbine sensorFlow: Turbine sensorVolume: Flow IntegrationSame
Accuracy±10 L/min or ±10%,whichever is greaterPEF ± 20 L/minor PEF ± 10% of the reading;FEV1 ± 0.05 L orFEV1 ± 3% of the readingSimilar
Repeatability<5% or 10 L/min,whichever is greaternot knownSimilar
Measuring resolutionPEF 1 L/minPEF 1 L/min;FEV1 0.01 LSame
Data transmissionBLE wirelesstransmissionBLE wirelesstransmissionSame
Memory volume500 recordings250 recordingsDifferent
Power sourceAAA 1.5×2 alkalinebatteriesAAA 1.5×2alkaline batteriesSimilar
Performance testingStandardisation ofspirometry (2005Revision)Standardisationof spirometry(2005 Revision)Same
Electrical Safety and EMCIEC 60601-1 and IEC60601-1-2IEC 60601-1 andIEC 60601-1-2Same
Shelf-Life/Use-Life3 yearsnot knownSimilar

{7}------------------------------------------------

7. Summary of Non-clinical Performance Testing

1) Electrical Safety, Electromagnetic Compatibility

The subject device has passed safety testing in according to following standards.

  • IEC 60601-1:2005, AMDI:2012 Medical electrical equipment Part 1: General ● requirements for basic safety and essential performance
  • IEC 60601-1-11:2015 Medical electrical equipment Part 1-11: General . requirements for basic safety and essential performance - Collateral Standard:

{8}------------------------------------------------

Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

  • . IEC 60601-1-2:2014 Medical electrical equipment -- Part 1-2: General requirements for basic safety and essential performance -- Collateral Standard: Electromagnetic disturbances -- Requirements and tests

2) Performance testing

Performance testing was conducted on the Electronic Peak Flow Meter for the measurement of peak expiratory flow (PEF). Technical parameters of accuracy and repeatability were evaluated in the performance testing, according to ATS/ERS Task Force: Standardisation of lung function testing - Standardisation of spirometry 2005. Measurements of PEF meet the requirements in the standard.

Operational/Environmental Testing IEC TR 60721-4-7:2001 Standardisation of spirometry (2005 Revision)

Transport testing: ISTA-3A:2018 Standardisation of spirometry (2005 Revision)

3) Biocompatibility test

Biocompatibility testing was conducted in accordance with the 2020 FDA guidance "Use of International Standard ISO 10993, Biological Evaluation of Medical Device Part 1: Evaluation and Testing."

The subject device has passed biocompatibility testing in according to following standards.

  • ISO 10993-5:2019 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
  • ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for . irritation and skin sensitization

4) Shelf/Use Life and Reprocessing

Shelf/Use life and reprocessing methods were validated in accordance with the following Standards:

  • ASTM F1980-21 Standardisation of spirometry (2005 Revision) .
  • AAMI TIR 12:2020 Designing, testing, and labeling medical devices intended for . processing by health care facilities: A guide for device manufacturers
  • . ISO 17664-1:2021 Processing of Health Care products- Information to be provided by the medical device manufacturer for the processing of medical devices

5) Software Verification and Validation

The subject device contains an embedded software system and an APP software. Software verification and validation testing were conducted as recommended by FDA's Guidance, "Guidance for the Content of Premarket Submissions for Software

{9}------------------------------------------------

Contained in Medical Devices", and the software for this device was considered as a "moderate" level of concern.

6) Cybersecurity Management

The device was evaluated for cybersecurity according to the following guidelines and standards

  • Content of Premarket Submissions for Management of Cybersecurity in Medical . Devices
  • IEEE/ANSI C63.27:2017

7) Human Factors

The device was evaluated for human factors and usability engineering.

8. Conclusion

The subject device, Electronic Peak Flow Meter (model DL-DF01), manufactured by Taian Dalu Medical Instrument Co., Ltd. is substantially equivalent to the predicate device (MSA100BT Peak Flow Meter) manufactured by Beijing M&B Electronic Instruments Co., Ltd. (K170281). This conclusion is based upon comparison on intended use, technological characteristics, and performance testing.

§ 868.1860 Peak-flow meter for spirometry.

(a)
Identification. A peak-flow meter for spirometry is a device used to measure a patient's maximum ventilatory flow rate.(b)
Classification. Class II (performance standards).