K Number
K170281

Validate with FDA (Live)

Date Cleared
2017-10-06

(249 days)

Product Code
Regulation Number
868.1860
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This device is intended for monitoring PEF (Peak Expired Flow Rate) and FEV 1 (Forced Expiratory Volume in one second) for patient home use. The device is designed for pediatric to adult patients. The device is intended for monitoring respiratory conditions such as asthma.

Device Description

MSA100BT Peak Flow Meter is a hand-held pulmonary function measuring medical device that measures patient's maximum possible exhalation which is called peak expiratory flow (PEF) and forced expiratory volume in 1 second (FEV1). The accuracy meets American Thoracic Society (ATS) Standard 2005 Revision. The MSA100BT is designed for pediatric to adult patients for home use and for single user, and is intended for monitoring respiratory conditions such as asthma. The MSA100BT has an automatic memory of 250 recordings and can connect to cell phone by Bluetooth. Patients can transfer the records to phone via BLE4.0, then using app to help save and track long term. The MSA100BT is capable of measuring PEF from 50 to 900 L/min and FEV1 from 0.01 to 9.99 L. Its measuring accuracy is ± 20 L/min or ± 10% of the reading for PEF, and ± 0.050 L or ± 3% of the reading for FEV1. The MSA100BT is composed of main unit, turbine sensor, and mouth piece. After each use, the mouthpiece and turbine of the device should be cleaned by soap solution and water within 30 minutes.

AI/ML Overview

The provided document is a 510(k) premarket notification for the MSA100BT Peak Flow Meter. It focuses on demonstrating substantial equivalence to a predicate device, the MSA100 Peak Flow Meter, rather than detailing a study that establishes acceptance criteria for a novel AI device and proves its performance.

Therefore, the specific information requested about acceptance criteria, a study proving device meeting criteria, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth types, and training set details for an AI-powered device cannot be extracted directly from this document. This document pertains to a physical medical device (peak flow meter) and its performance validation using non-clinical testing and comparison to an existing device, not an AI algorithm.

However, I can extract the performance criteria and the testing performed to demonstrate that the device complies with certain standards, which can be interpreted as fulfilling "acceptance criteria" in the context of this device's submission.

Here's the closest information I can provide based on the document:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Standard Requirement)Reported Device Performance
Accuracy:
PEF: ± 20 L/min or ± 10% of the readingPEF: ± 20 L/min or ± 10% of the reading (Stated as identical to predicate and meets ATS Standard 2005 Revision)
FEV1: ± 0.05 L or ± 3% of the readingFEV1: ± 0.05 L or ± 3% of the reading (Stated as identical to predicate and meets ATS Standard 2005 Revision)
Biocompatibility: (According to ISO 10993-1:2009/(R)2013)
Non-cytotoxicTest article confirmed non-cytotoxic (Complied with ISO 10993-5:2009)
Non-irritating (oral mucosa)Test article confirmed non-oral-mucosal-irritating (Complied with ISO 10993-10:2010)
Non-sensitizingTest article confirmed non-sensitizing (Complied with ISO 10993-10:2010)
Electrical Safety: (According to IEC 60601-1)Device complies with IEC 60601-1:2012
Electromagnetic Compatibility (EMC): (According to IEC 60601-1-2)Device complies with IEC 60601-1-2:2014
Wireless Compliance: (According to FDA Guidance on RF Wireless Tech)Device complies with FDA's guidance: "Guidance Radio Frequency Wireless Technology in Medical Devices Guidance for Industry and Food and Drug Administration Staff."
Performance Parameters: (According to Standards of Spirometry 2005)All of the tested parameters (accuracy, intra instrument repeatability, PEF inter instrument repeatability, performance of flow resistance) meet the requirements in the standards. The performance of the proposed device is demonstrated to be comparable with the predicate device. The accuracy meets American Thoracic Society (ATS) Standard 2005 Revision. (Range: PEF 50 to 900 L/min, FEV1 0.01 to 9.99 L; Measuring solution: PEF 1 L/min; FEV1 0.01 L).
Software Verification & Validation: (According to FDA Guidance)Software validation demonstrated that the software functions as specified in the software requirement specifications. (For "moderate" level of concern software)

Regarding the specific questions about AI/ML device studies, the document does not contain this information as it is not about such a device.

  1. Sample size used for the test set and the data provenance: Not applicable. The testing described is for a physical device, not an AI algorithm evaluated on a data test set from specific provenance.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. The ground truth for performance is established by reference standards for spirometry or laboratory tests for biocompatibility/electrical safety.
  3. Adjudication method: Not applicable.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No, this is not an AI-assisted diagnostic device.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: No, this is not an AI algorithm.
  6. The type of ground truth used:
    • Performance: Reference standards specified by Standards of Spirometry (2005 Revision) and American Thoracic Society (ATS) Standard 2005 Revision.
    • Biocompatibility: Laboratory testing results against ISO 10993 standards.
    • Electrical Safety & EMC: Compliance with IEC 60601 standards.
    • Wireless: Compliance with FDA guidance.
    • Software: Software requirement specifications.
  7. The sample size for the training set: Not applicable, as this is not an AI/ML device.
  8. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The caduceus is a symbol of medicine and healing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 6, 2017

Beijing M&B Electronic Instruments Co., Ltd. % Mike Gu Regulatory Affairs Manager Guangzhou Osmunda Medical Device Technical Services Co.,Ltd. 8-9Th Floor, R&D Building, No.26 Qinglan St, Panyu District Guangzhou. 510006 Cn

Re: K170281

Trade/Device Name: MSA100BT Peak Flow Meter Regulation Number: 21 CFR 868.1860 Regulation Name: Peak-Flow Meter for Spirometry Regulatory Class: Class II Product Code: BZH Dated: September 1, 2017 Received: September 5, 2017

Dear Mike Gu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{1}------------------------------------------------

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Tara A. Ryan -S

for

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K170281

Device Name MSA100BT Peak Flow Meter

Indications for Use (Describe)

This device is intended for monitoring PEF (Peak Expired Flow Rate) and FEV 1 (Forced Expiratory Volume in one second) for patient home use. The device is designed for pediatric to adult patients. The device is intended for monitoring respiratory conditions such as asthma.

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows a blue logo. The logo appears to be a stylized flower or plant with two petals or leaves extending upwards and outwards. There is a registered trademark symbol (®) located to the upper right of the logo.

510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

1. SUBMITTER

Beijing M&B Electronic Instruments Co., Ltd.

Room 6319, Building 1, No. 27, Yongwang Road, Beijing Bioengineering and Medicine Industry Base, Huangcun Town, Daxing district, Beijing, P.R. China

Phone: +86-10-61253716 Fax: +86-10-61253794

Contact Person:Yang Shisheng
Project Manager
Beijing M&B Electronic Instruments Co., Ltd.
Tel: (+86)-10-61253716-8037
Fax: (+86)-10-61253794

Date prepared Oct. 03, 2017

    1. DEVICE
Device Name:MSA100BT Peak Flow Meter
Common/Usual Name:Peak Flow Meter
Regulation number868.1860 Peak-flow meter for spirometry
Regulation Class:II
Product Code:BZH Meter, Peak Flow, Spirometry
    1. PREDICATE DEVICE
      K133975, MSA100 Peak Flow Meter

This predicate legally marketed device has not been subject to a design-related recall.

No reference device was used in this submission.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a logo with a stylized blue symbol that resembles a flame or a stylized flower. The symbol has a flowing, upward motion. To the upper right of the symbol is a circled "R", indicating that the logo is a registered trademark.

4. DEVICE DESCRIPTION

MSA100BT Peak Flow Meter is a hand-held pulmonary function measuring medical device that measures patient's maximum possible exhalation which is called peak expiratory flow (PEF) and forced expiratory volume in 1 second (FEV1). The accuracy meets American Thoracic Society (ATS) Standard 2005 Revision.

The MSA100BT is designed for pediatric to adult patients for home use and for single user, and is intended for monitoring respiratory conditions such as asthma.

The MSA100BT has an automatic memory of 250 recordings and can connect to cell phone by Bluetooth. Patients can transfer the records to phone via BLE4.0, then using app to help save and track long term.

The MSA100BT is capable of measuring PEF from 50 to 900 L/min and FEV1 from 0.01 to 9.99 L. Its measuring accuracy is ± 20 L/min or ± 10% of the reading for PEF, and ± 0.050 L or ± 3% of the reading for FEV1.

The MSA100BT is composed of main unit, turbine sensor, and mouth piece. After each use, the mouthpiece and turbine of the device should be cleaned by soap solution and water within 30 minutes.

5. INDICATIONS FOR USE

This device is intended for monitoring PEF (Peak Expired Flow Rate) and FEV1 (Forced Expiratory Volume in one second) for patient home use. The device is designed for pediatric to adult patients. The device is intended for monitoring respiratory conditions such as asthma.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a blue logo. The logo appears to be a stylized representation of a plant or flower with three petals or leaves. There is a registered trademark symbol to the right of the logo.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE 6. PREDICATE DEVICE

SpecificationPredicate MSA100 PeakFlow MeterProposedMSA100BT Peak Flow MeterDiscussion ofDifferences
K numberK133975
Device nameMSA100 Peak Flow MeterMSA100BT Peak Flow Meter
ManufacturerBeijing M&B Electronic Instruments Co., LtdIdentical
Indications for useThe device is intended for monitoring Peak Expired Flow(PEF) and Forced Expiratory Volume in one second (FEV1)for patient home use. The device is intended for monitoringrespiratory conditions such as asthma.Identical
Patient PopulationThe device is designed for pediatric to adult patientsIdentical
Environment of UseOTCIdentical
MaterialsPP (mouthpiece)Identical
Measuring rangePEF from 50 to 900 L/minFEV1 from 0.01 to 9.99 LIdentical
Measuring methodFlow: Turbine sensorVolume: Flow IntegrationIdentical
AccuracyPEF ± 20 L/min or PEF ± 10% of the reading;FEV1 ± 0.05 L or FEV1 ± 3% of the readingIdentical
Measuring solutionPEF 1 L/min; FEV1 0.01 LIdentical
Data safetyData rememorized by flash memoryIdentical
Data transmissionUSB wire transmissionBLE wireless transmissionDifferentThis difference does notraise different questionsof safety andeffectiveness.
Memory volume300 recordings250 recordingsDifferentThis difference does notraise different questionsof safety andeffectiveness.
Power sourceAAA 1.5×2 alkaline batteriesIdentical
Electrical Safety andEMCIEC 60601-1 and IEC 60601-1-2Identical

Table 6.1 Comparison with the Predicate Device

According to the above comparison, the proposed device is identical to the predicate device in indications for use, materials, measuring method, technical specification, power source and electrical safety.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows a stylized, abstract blue flower. The flower has three petals that are curved and appear to be blowing in the wind. The flower is set against a white background. In the upper right corner of the image, there is a small, circular icon with a symbol inside.

The differences are the way of data transmission and memory, which are USB wire transmission and 300 recordings for the predicate device and BLE (Blue Smart) wireless transmission and 250 recordings for the proposed device. The differences will not raise different questions of safety and effectiveness, as both devices were demonstrated to be electromagnetic compatibility and electrical safety by testing.

In conclusion, Beijing M&B Electronic Instruments Co., Ltd. believes that the MSA100BT Peak Flow Meter is as safe and effective, and performs in a substantially equivalent manner to the predicate device.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows a blue logo on a white background. The logo appears to be a stylized flower or plant with three distinct petals or leaves at the top and a stem extending downwards. There is a small circle with the registered trademark symbol (®) in the upper right corner of the logo.

7. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility testing

The mouthpiece of the MSA100BT Peak Flow Meter is considered to contact directly with the human body for a duration of less than 24 hours.

The mouthpiece of the MSA100BT Peak Flow Meter in its final finished form is identical to the mouthpiece of the MSA100 Peak Flow Meter in formulation and processing, and no other chemicals have been added.

The biocompatibility evaluation for the mouthpiece was conducted in accordance with the International Standard ISO 10993-1:2009/(R)2013, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" as recognized by FDA. The biocompatible testing included the following tests:

  • Cytotoxicity
  • Irritation
  • Skin sensitization

Testing for cytotoxicity, skin sensitization and oral mucosa irritation complied with ISO 10993-5:2009, Biological evaluation of medical devices - Part 5: Tests for In Vitro cytotoxicity and ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization. Testing demonstrates that the test article is non-cytotoxic, non-sensitizing, and non-oral-mucosal-irritating.

Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the MSA100BT Peak Flow Meter. The device complies with the IEC 60601-1:2012, standard for electrical safety and the IEC 60601-1-2:2014 standard for EMC. It demonstrates substantial equivalences to the predicate device.

Wireless testing was conducted on the MSA100BT Peak Flow Meter. The device complies with FDA's guidance: Guidance Radio Frequency Wireless Technology in Medical Devices Guidance for Industry and Food and Drug Administration Staff.

Performance testing

Performance testing was conducted on the MSA100BT Peak Flow Meter. Technical parameters of accuracy, intra instrument repeatability, PEF inter instrument repeatability, performance of flow resistance were evaluated in the performance

{8}------------------------------------------------

testing, according to Standards of Spirometry (2005 Revision). All of the tested parameters meet the requirements in the standards. The performance of the proposed device is demonstrated to be comparable with the predicate device in the comparison testing, so it is concluded that the proposed device is substantially equivalent to the predicate device.

Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern. Software validation demonstrated that the software functions as specified in the software requirement specifications. It demonstrates substantial equivalences to the predicate device.

Animal Study

The subject of this premarket submission, MSA100BT Peak Flow Meter, did not require animal studies to support substantial equivalence.

Clinical Study

The subject of this premarket submission, MSA100BT Peak Flow Meter, did not require clinical studies to support substantial equivalence.

8. CONCLUSION

The differences between the MSA100BT Peak Flow Meter and its predicate device do not raise different questions of safety and effectiveness. The non-clinical data including the performance testing supports that the MSA100BT Peak Flow Meter is substantially equivalent to the predicate device.

§ 868.1860 Peak-flow meter for spirometry.

(a)
Identification. A peak-flow meter for spirometry is a device used to measure a patient's maximum ventilatory flow rate.(b)
Classification. Class II (performance standards).