K Number
K223788

Validate with FDA (Live)

Date Cleared
2023-09-15

(270 days)

Product Code
Regulation Number
880.5200
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MedSource TrueSafe Blood Control I.V. Safety Catheter, and the MedSource ClearSafe Blood Control I.V. Safety Catheter are indicated to sample blood or administer fluids intravenously. The MedSource TrueSafe Blood Control I.V. Safety Catheter and the MedSource ClearSafe Blood Control I.V. Safety Catheter may be used for any patient population with consideration given to adequacy for the vascular anatomy, appropriateness for the solution being administered and duration of the therapy.

Device Description

The MedSource TrueSafe Blood Control I.V. Safety Catheter, and the MedSource ClearSafe Blood Control I.V. Safety Catheters are a safety medical device used for inserting a catheter into a patient's body for the delivery of fluids or drainage of fluids from a patient's body. This device is engineered to protect healthcare workers using it against accidental needle stick injury. In addition, this device also secures the needle (used to insert the catheter) within the catheter body with either a push button or slide safety mechanism, thus protecting healthcare and other personnel from accidental needle sticks.

The MedSource TrueSafe Blood Control I.V. Safety Catheter is comprised of a sharp needle attached to a needle hub with a catheter attached to a catheter hub. In the catheter hub there is a valve that, when activated, allows for the movement of fluids. In the 'inactivated' state prior to insertion of a male Luer, the movement of fluids is restricted. When the needle is retracted with the push button from the catheter the needle retracts into the body of the catheter and the needle is fully encapsulated.

The MedSource ClearSafe Blood Control I.V. Safety Catheter blood control mechanism is identical to the MedSource TrueSafe Blood Control I.V. Safety Catheter. The only difference in design features is the ClearSafe Blood Control has a slide safety mechanism that retracts the needle into the catheter body. The needle is also fully encapsulated to prevent accidental needlesticks. There is no difference in clinical use between the two models and the use of one over the other depends on user preference.

The MedSource TrueSafe and ClearSafe Blood Control Safety I.V. Catheter gauges are color coded for positive identification and range from 14G to 24G.

AI/ML Overview

The provided text does not contain information about acceptance criteria and a study that proves a device meets those criteria. Instead, it describes a medical device, its intended use, comparison with a predicate device, and the various performance and biocompatibility tests it underwent to achieve substantial equivalence for FDA clearance.

Therefore, I cannot fulfill your request for:

  1. A table of acceptance criteria and reported device performance.
  2. Sample size and data provenance for a test set.
  3. Number and qualifications of experts for ground truth.
  4. Adjudication method for the test set.
  5. MRMC study information.
  6. Standalone performance information.
  7. Type of ground truth used.
  8. Sample size for the training set.
  9. How ground truth for the training set was established.

This document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a legally marketed predicate device through engineering and performance testing, rather than a clinical study evaluating diagnostic or therapeutic efficacy against specific acceptance criteria in a human population.

The performance testing listed (ISO standards, simulated clinical use study) are applied to the device itself to ensure it functions as intended and safely, not typically to evaluate its effectiveness in a diagnostic or treatment context against a clinical ground truth. For example, ISO 10555-5 relates to "Intravascular catheters Sterile and single-use catheters Part 5: One-needle peripheral catheters" to ensure the device meets the physical and functional requirements for such catheters.

The document states: "In summary, based on the indication for use, technological characteristics and performance testing, the differences between The MedSource TrueSafe Blood Control I.V. Safety Catheter, and the MedSource ClearSafe Blood Control I.V. Safety Catheter and the predicate did not raise new or different questions of safety and effectiveness." This is the conclusion of the 510(k) process, indicating that the device is deemed safe and effective based on its similarity to existing devices and its adherence to relevant performance standards.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

September 15, 2023

MedSource International LLC Emilie Andrews Quality and Regulatory Compliance Engineer 8600 Shelby Court, Suite 100 Chanhassen, Minnesota 55317

Re: K223788

Trade/Device Name: The MedSource TrueSafe Blood Control I.V. Safety Catheter, and the MedSource ClearSafe Blood Control I.V. Safety Catheter

Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: Class II Product Code: FOZ Dated: August 15, 2023 Received: August 16, 2023

Dear Emilie Andrews:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

David Walloschek

David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223788

Device Name

The MedSource TrueSafe Blood Control I.V. Safety Catheter, and the MedSource ClearSafe Blood Control I.V. Safety Catheter

Indications for Use (Describe)

The MedSource TrueSafe Blood Control I.V. Safety Catheter, and the MedSource ClearSafe Blood Control I.V. Safety Catheter are indicated to sample blood or administer fluids intravenously. The MedSource TrueSafe Blood Control I.V. Safety Catheter and the MedSource ClearSafe Blood Control I.V. Safety Catheter may be used for any patient population with consideration given to adequacy for the vascular anatomy, appropriateness for the solution being administered and duration of the therapy.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

MEDSOURCELABS

K223788 - 510(k) Summary

Date of Prepared: September 15, 2023

A. Submitter:

MedSource International, LLC 8600 Shelby Court Chanhassen, MN 55317

B. Contact Person:

Emilie Andrews, Regulatory and Quality Compliance Engineer 8600 Shelby Court Chanhassen, MN 55317 Phone: 952-220-8875

C. Proposed Device:

Product Code: FOZ

Trade Name: The MedSource TrueSafe Blood Control I.V. Safety Catheter, and the MedSource ClearSafe Blood Control I.V. Safety Catheter

Common Name: Short-Term Less Than 30 Days, Therapeutic, Intravascular Catheter

Classification Name: Intravascular catheter

Regulatory Reference: 21 CFR §880.5200

Review Panel: General Hospital

Classification: Class II

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image shows the logo for MEDSOURCELABS. The text is in a bold, sans-serif font and is black. To the right of the text is a cluster of colorful hexagons.

D. Predicate:

510(K): K161779

Product Code: FOZ

Trade Name: MedSource TrueSafe Safety IV Catheter and MedSource TrueSafe Comfort Safety IV Catheter

Common Name: Short-Term Less Than 30 Days, Therapeutic, Intravascular Catheter

Classification Name: Intravascular catheter

Regulatory Reference: 21 CFR §880.5200

Review Panel: General Hospital

Classification: Class II

E. Indications for Use:

The MedSource TrueSafe Blood Control I.V. Safety Catheter, and the MedSource ClearSafe Blood Control I.V. Safety Catheter are indicated to sample blood or administer fluids intravenously. The MedSource TrueSafe Blood Control I.V. Safety Catheter and the MedSource ClearSafe Blood Control I.V. Safety Catheter may be used for any patient population with consideration given to adequacy for the vascular anatomy, appropriateness for the solution being administered and duration of the therapy.

F. Device Description:

The MedSource TrueSafe Blood Control I.V. Safety Catheter, and the MedSource ClearSafe Blood Control I.V. Safety Catheters are a safety medical device used for inserting a catheter into a patient's body for the delivery of fluids or drainage of fluids from a patient's body. This device is engineered to protect healthcare workers using it against accidental needle stick injury. In addition, this device also secures the needle (used to insert the catheter) within the catheter body with either a push button or slide safety mechanism, thus protecting healthcare and other personnel from accidental needle sticks.

{5}------------------------------------------------

Image /page/5/Picture/2 description: The image shows the logo for MEDSOURCELABS. The text is in a bold, sans-serif font and is black. To the right of the text is a cluster of colorful hexagons.

The MedSource TrueSafe Blood Control I.V. Safety Catheter is comprised of a sharp needle attached to a needle hub with a catheter attached to a catheter hub. In the catheter hub there is a valve that, when activated, allows for the movement of fluids. In the 'inactivated' state prior to insertion of a male Luer, the movement of fluids is restricted. When the needle is retracted with the push button from the catheter the needle retracts into the body of the catheter and the needle is fully encapsulated.

The MedSource ClearSafe Blood Control I.V. Safety Catheter blood control mechanism is identical to the MedSource TrueSafe Blood Control I.V. Safety Catheter. The only difference in design features is the ClearSafe Blood Control has a slide safety mechanism that retracts the needle into the catheter body. The needle is also fully encapsulated to prevent accidental needlesticks. There is no difference in clinical use between the two models and the use of one over the other depends on user preference.

The MedSource TrueSafe and ClearSafe Blood Control Safety I.V. Catheter gauges are color coded for positive identification and range from 14G to 24G

Orange = 14GGrey = 16GGreen = 18GPink = 20G
Blue = 22GYellow = 24G

G. SUMMARY OF COMPARISON TABLE OF PROPOSED DEVICE and PREDICATE

The proposed device is substantially equivalent to the predicate device the MedSource TrueSafe Safety I.V. Catheter (K161779). The MedSource IV Safety Catheter is also used as a reference device.

ComparisonPointSubmission DevicePredicate DeviceDiscussionnComments
The MedSourceTrueSafe BloodControl I.V. SafetyCatheter, and theMedSource ClearSafeThe MedSourceTruesafe Safety I.V.Catheter and theMedSource TrueSafe
Blood Control I.V.Safety CatheterComfort Safety I.V.Catheter(K161779)
Indicationsfor UseThe MedSourceTrueSafe Safety IVCatheter and theMedSource TrueSafeComfort Safety IVCatheter is indicatedto sample blood oradminister fluidsintravenously. TheMedSource TrueSafeSafety IV Catheterand the MedSourceTrueSafe ComfortSafety IV Cathetermay be used for anypatient populationwith considerationgiven to adequacy forthe vascular anatomyappropriateness forthe solution beingadministered andduration of thetherapy.The MedSourceTrueSafe BloodControl I.V. SafetyCatheter, and theMedSource ClearSafeBlood Control I.V.Safety Catheter areindicated to sampleblood or administerfluids intravenously.The MedSourceTrueSafe BloodControl I.V. SafetyCatheter and theMedSource ClearSafeBlood Control I.V.Safety Catheter maybe used for anypatient populationwith considerationgiven to adequacy forthe vascular anatomy,appropriateness forthe solution beingadministered andduration of thetherapySame
Principles ofOperationsThe needle is insertedto gain vascularaccess. Then bloodwill fill the flashbackchamber. After theneedle is placed itwill be removed fromThe needle is insertedto gain vascularaccess. Then bloodwill fill the flashbackchamber. After theneedle is placed itwill be removed fromSame
the catheter leavingthe catheter behind toallow for vascularaccess. As the needleis removed the needleis encapsulated insidethe catheter body.the catheter leavingthe catheter behind toallow for vascularaccess. As the needleis removed the needleis encapsulated insidethe catheter body.
CatheterOuterDimensions(O.D.) inmm24G - 0.722G - 0.920G - 1.118G - 1.316G – 1.814G - 2.124G - 0.722G - 0.920G - 1.118G - 1.316G - 1.714G – 2.1Same
CatheterEffectiveLength inmmGauge Length24G 1922G 2520G - TrueSafe 3020G - ClearSafe 3218G - TrueSafe 3018G - ClearSafe 3216G - TrueSafe 3016G - ClearSafe 3214G - TrueSafe 45Gauge Length24G 1922G 2520G 2518G 2516G 3014G 45DifferentClinical simulated usewas conducted tovalidate the additionof the blood controlvalve did not changethe effectiveness ofthe device.Performance testingwas done to ISO10555-5 and ISO10555-1 to verifythere was no changeto the effectiveness ofthe deviceBased on the testingdone the catheter
Number ofLumen(s)Single LumenSingle LumenSamebetween the two IVCatheters does notaffect the safety anddevice.
Shape ofLumen(s)Catheter hub –cylindrical with ataper down to thecatheterCatheter hub –cylindrical with ataper down to thecatheterSame
Flow Ratefor EachLumen(s) inmL/minGauge Flow Rate24G – TrueSafe 1824G – ClearSafe 2022G – TrueSafe 3322G – ClearSafe 3620G – TrueSafe 5520G – ClearSafe 6118G – TrueSafe 8518G – ClearSafe 9416G 20014G 270Gauge Flow Rate24G 2022G 3620G 6118G 9216G 22014G 297DifferentFlow rate variancebetween the two IVCatheters is withinthe allowance in ISO10555-1 ANNEX -EThis does not raisenew or differentquestions of safety oreffectiveness.
ProximalEndconfigurationCupperCupperSame
Distal EndconfigurationBeveledBeveledSame
Location ofoutlets orside portsNo portsNo portsSame
Diameter ofoutlets orside portsNot applicable noports presentNot applicable noports presentSame
Needle StickPreventionFeatureActive - Manualsafety mechanismSlide-ClearSafe andPush button-TrueSafeActive - Manualsafety mechanismPush buttonSameDifferentClinical simulated usewas conducted tovalidate the additionof the blood controlvalve did not changethe effectiveness ofthe device.Performance testingwas done to ISO23908:2011Based on the resultsof the testing thisdifference does notraise new or differentquestions of safety oreffectiveness.
SterilizationTerminal EthyleneOxide (EtO gas)Terminal EthyleneOxide (EtO gas)Same
CatheterPolyurethane (PUR)/ (PTFE)PolytetrafluoroethylenePolyurethane (PUR) /(PTFE)PolytetrafluoroethyleneSame
NeedleStainless StealStainless SteelSame
CatheterBodyPolypropylene (PP)K-ResinDifferentClinical simulated usewas conducted tovalidate the additionof the blood controlvalve did not changethe effectiveness ofthe device.Performance testingwas done to ISO10555-5ISO 10555-1, ISO594-1 and ISO 594-2to verify there wasnot change to theeffectiveness of thedeviceSee thebiocompatibility inSection H below forthe list ofbiocompatibilitytesting done to verifythere was no changein safetyBased on the resultsof the testing thisdifference does notraise new or differentquestions of safety oreffectiveness.
CatheterHolderPolyacetal (POM)Polyacetal (POM)Same
Needle HubK-Resin / MABS-TK-ResinDifferentClinical simulated usewas conducted tovalidate the additionof the blood controlvalve did not changethe effectiveness ofthe device.
10555-5,ISO 10555-1,ISO 594-1, ISO 594-2to verify there wasnot change to theeffectiveness of thedevice
See thebiocompatibility inSection H below forthe list ofbiocompatibilitytesting done to verifythere was no changein safety
Based on the resultsof the testing thisdifference does notraise new or differentquestions of safety oreffectiveness.
BloodControlValvePolypropylene (PP) /Silicon / POMAbsentDifferentThis valve is locatedinside the catheterhub. This is notpresent in theTrueSafe Safety IVmodels.This was added as anadditional feature tohelp control possibleblood flow out of thecatheter hub
Performance testingwas done followingISO 594-1 andISO 594-2 to verify
to the effectiveness ofthe device
Clinical simulated usewas conducted tovalidate the additionof the blood controlvalve did not changethe effectiveness ofthe device.
See thebiocompatibility inSection H below forthe list ofbiocompatibilitytesting done to verifythere was no changein safety
Based on the resultsof the testing thisdifference does notraise new or differentquestions of safety oreffectiveness.
FlashbackChamberPolypropylene (PP) /Polycarbonate (PC)ABSDifferentPerformance andbiocompatibilityevaluationdemonstrates that theaddition ofPolytetrafluoroethylene does not raiseadditional questionsof safety andeffectiveness.
Clinical simulated usewas conducted tovalidate the addition
of the blood controlvalve did not changethe effectiveness ofthe device.Performance testingwas done to ISO10555-5: and ISO10555-1 to verifythere was no changeto the effectiveness ofthe device
See thebiocompatibility inSection H below forthe list ofbiocompatibilitytesting done to verifythere was no changein safety
Based on the resultsof the testing thisdifference does notraise new or differentquestions of safety oreffectiveness.
Hydrophobic FilterPolyethylene (PE)Poly-ethyleneSame

{6}------------------------------------------------

MEDSOURCELABS

{7}------------------------------------------------

MEDSOURCELABS

{8}------------------------------------------------

Image /page/8/Picture/2 description: The image shows the logo for MEDSOURCELABS. The text "MEDSOURCELABS" is in a bold, sans-serif font. To the right of the text is a cluster of colorful hexagons arranged in a triangular shape. The hexagons are in various colors, including yellow, green, red, purple, black, teal, and orange.

{9}------------------------------------------------

MEDSOURCELABS

{10}------------------------------------------------

Image /page/10/Picture/2 description: The image shows the logo for MEDSOURCELABS. The text "MEDSOURCELABS" is in a bold, sans-serif font. To the right of the text is a cluster of colorful hexagons arranged in a triangular shape. The hexagons are in various colors, including yellow, green, red, purple, black, teal, and orange.

{11}------------------------------------------------

Image /page/11/Picture/2 description: The image shows the logo for MEDSOURCELABS in bold, black font. To the right of the logo is a cluster of colorful hexagons. Below the logo, there is text that states "Performance testing was done to ISO 10555-2".

{12}------------------------------------------------

10 of 13

Image /page/12/Picture/2 description: The image shows the logo for MEDSOURCELABS. The text "MEDSOURCELABS" is written in a bold, sans-serif font. To the right of the text is a cluster of colorful hexagons arranged in a triangular shape. The hexagons are various colors, including green, yellow, red, purple, black, teal, and orange.

{13}------------------------------------------------

11 of 13

Image /page/13/Picture/2 description: The image shows the logo for MEDSOURCELABS. The text is in a bold, sans-serif font and is black. To the right of the text is a cluster of colorful hexagons.

{14}------------------------------------------------

MEDSOURCELABS

H. Performance Data:

i. Performance Testing:

  • ISO 10555-5: Intravascular catheters Sterile and single-use catheters Part 5: Over-. needle peripheral catheters
  • ISO 10555-1 Intravascular catheters Sterile and single-use catheters Part 1: General . requirements
  • ISO 594-1Conical fittings with a 6% (Luer) taper for syringes, needles and certain other . medical equipment. - Part 1: General requirements
  • . ISO 594-2 Conical fittings with 6 % (Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock fittings
  • ISO 23908:2011 Sharps injury protection Requirements and test methods Sharps . protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling
  • Simulated Clinical Use Study ●

ii. Biocompatibility

This device has prolonged contact with circulating blood.

  • ISO 10993-5: 2009(E) Biological evaluation of medical devices Part 5: Tests for . in vitro cytotoxicity
  • ISO 10993-10: 2010 (E) Biological evaluation of medical devices Part 10: Tests ● for irritation and skin sensitization
  • . USP General Chapter <151> Pyrogen Test
  • ISO 10993-11: 2017(E) Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
  • . ASTM F756-17 Standard Practice for Assessment of Hemolytic Properties of Materials
  • ISO 10993-4: 2017 (E) Biological evaluation of medical devices Part 4: Selection . of tests for interactions with blood
  • ISO 10993-33:2015 Biological evaluation of medical devices Part 33: Guidance ● on tests to evaluate genotoxicity - Supplement to ISO 10993-3
  • ISO 10993-3:2014 Biological evaluation of medical devices Part 3: Tests for ● genotoxicity, carcinogenicity and reproductive toxicity

{15}------------------------------------------------

Image /page/15/Picture/2 description: The image shows the logo for MEDSOURCELABS. The text "MEDSOURCELABS" is in a bold, sans-serif font and is black. To the right of the text is a cluster of colorful hexagons arranged in a grid pattern.

  • ISO 10993-6:2016(E) Biological evaluation of medical devices - Part 6: Tests for local effects after implantation
  • . ASTM F2888-19 Standard Practice for Platelet Leukocyte Count-An In-Vitro Measure for Hemocompatibility Assessment of Cardiovascular Materials
  • ASTM F2382-18 Standard Test Method for Assessment of Circulating Blood-● Contacting Medical Device Materials on Partial Thromboplastin Time (PTT)

iii. Sterilitv

  • ISO 11135:2014 Sterilization of health-care products Ethylene oxide Requirements ● for the development, validation and routine control of a sterilization process for medical devices
  • ASTM D4332-22 Standard Practice for Conditioning Containers, Packages, or . Packaging Components for Testing
  • . ASTM D4169-22 Standard Practice for Performance Testing of Shipping Containers and Systems
  • ASTM F1980-21 Standard Guide for Accelerated Aging of Sterile Barrier Systems and . Medical Devices
  • ASTM F88/F88M-21 Standard Test Method for Seal Strength of Flexible Barrier ● Materials
  • ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical ● Packaging by Dye Penetration

I. Conclusion:

In summary, based on the indication for use, technological characteristics and performance testing, the differences between The MedSource TrueSafe Blood Control I.V. Safety Catheter, and the MedSource ClearSafe Blood Control I.V. Safety Catheter and the predicate did not raise new or different questions of safety and effectiveness.

§ 880.5200 Intravascular catheter.

(a)
Identification. An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.(b)
Classification. Class II (performance standards).