K Number
K223613

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-03-17

(102 days)

Product Code
Regulation Number
870.5150
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The InThrill Thrombectomy System is indicated for:

  • The non-surgical removal of thrombi and emboli from blood vessels, including arteriovenous grafts for dialysis access, and synthetic grafts.
  • Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel/graff.
    The InThrill Thrombectomy System is intended for use in the peripheral vasculature.
    The InThrill Thrombectomy System is not intended for use in deep vein thrombosis (DVT) treatment.
Device Description

The InThrill™ Thrombectomy System is a single-use, sterile medical device designed to remove thrombi and emboli from the peripheral vasculature. The InThrill™ Thrombectomy System consists of the InThrill Sheath ("Sheath") and the InThrill Thrombectomy Catheter ("Catheter"), each packaged separately.
The Sheath is an introducer sheath with a distal self-expanding funnel, proximal aspiration port, and proximal hub. A Dilator is provided to aid insertion and positioning of the Sheath. Radiopaque markers aid Sheath positioning under fluoroscopic visualization. The Sheath and Dilator tips are radiopaque, and there is a radiopaque marker band at the proximal end of the Sheath funnel.

AI/ML Overview

The provided text describes the 510(k) summary for the InThrill Thrombectomy System (K223613). While it details the device, its intended use, and a comparison to predicate devices, it does not contain information on acceptance criteria or the specifics of a study proving the device meets those criteria in the context of AI/ML performance.

The "Performance Tests" section on page 8 lists several non-clinical tests conducted, and states: "Test results demonstrated that all acceptance criteria were met; therefore, the device conforms to established product specifications." However, it does not provide the specific acceptance criteria values, nor the reported device performance values against those criteria. Furthermore, these appear to be engineering/mechanical performance tests, not AI/ML model performance evaluations.

The document explicitly states: "Clinical testing was not required for the determination of substantial equivalence." This means there was no human clinical study conducted to assess the device's performance in patients, and therefore no data on metrics like sensitivity, specificity, or AUC as would be typical for AI/ML device approval.

Therefore, based solely on the provided text, I cannot complete the requested information regarding acceptance criteria and a study proving the device meets them in the context of AI/ML performance.

The provided document is a 510(k) premarket notification for a medical device (an embolectomy catheter), which focuses on demonstrating substantial equivalence to predicate devices primarily through engineering and non-clinical testing. It does not describe an AI/ML component or a study with requirements such as ground truth establishment for a test set, expert adjudication, or MRMC studies.

If this was a request to describe the acceptance criteria and study for a different type of device or an AI/ML component that was not detailed in this specific 510(k) summary, then the information would need to come from a different source.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.

March 17, 2023

Inari Medical, Inc Ellen Nguyen Regulatory Affairs Specialist 6001 Oak Canyon, Suite 100 Irvine, California 92618

Re: K223613

Trade/Device Name: InThrill Thrombectomy System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: QEW, KRA Dated: February 14, 2023 Received: February 14, 2023

Dear Ellen Nguyen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Gregory W. Digitally signed by
Gregory W. Gregory W. O'connell -S O'connell -S Date: 2023.03.17

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223613

Device Name InThrill Thrombectomy System

Indications for Use (Describe) The InThrill Thrombectomy System is indicated for:

  • The non-surgical removal of thrombi and emboli from blood vessels, including arteriovenous grafts for dialysis access, and synthetic grafts.

  • Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel/graff.

The InThrill Thrombectomy System is intended for use in the peripheral vasculature.

The InThrill Thrombectomy System is not intended for use in deep vein thrombosis (DVT) treatment.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

PAGE 1 OF 6

510(K) SUMMARY

Date preparedMarch 16, 2023
NameInari Medical, Inc.6001 Oak Canyon, Suite 100Irvine, CA 92618877.923.4747
Contact personEllen NguyenRegulatory Affairs Specialist
Trade nameInThrill Thrombectomy System
Common nameEmbolectomy catheter
Regulation nameEmbolectomy catheter
Classification number21 CFR 870.5150
Product codeQEW
Secondary product codeKRA
Regulatory classII
Predicate device(s)Capture Vascular, MegaVac Mechanical Thrombectomy System (K171493)Hotspur Technologies, Inc., PTA-Plus PTA Balloon Catheter (K100842)Inari Medical, Mini-ClotTriever Thrombectomy System (K220887)
Reference device(s)Intramed Laboratories, Inc., Graft Thrombectomy Instruments (K942457)Rex Medical, Cleaner Rotational Thrombectomy System (K141617)
DescriptionThe InThrill™ Thrombectomy System is a single-use, sterile medical devicedesigned to remove thrombi and emboli from the peripheral vasculature. TheInThrill™ Thrombectomy System consists of the InThrill Sheath ("Sheath")and the InThrill Thrombectomy Catheter ("Catheter"), each packagedseparately.The Sheath is an introducer sheath with a distal self-expanding funnel,proximal aspiration port, and proximal hub. A Dilator is provided to aidinsertion and positioning of the Sheath.Radiopaque markers aid Sheath positioning under fluoroscopic visualization.The Sheath and Dilator tips are radiopaque, and there is a radiopaque markerband at the proximal end of the Sheath funnel.
Indications for UseThe InThrill Thrombectomy System is indicated for:The non-surgical removal of thrombi and emboli from blood vessels, including arteriovenous fistulae and arteriovenous grafts for dialysis access, and synthetic grafts. Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel/graft.

{4}------------------------------------------------

PAGE 2 OF 6

The InThrill Thrombectomy System is intended for use in the peripheral vasculature.

The InThrill Thrombectomy System is not intended for use in deep vein thrombosis (DVT) treatment.

Summary of substantial

A tabular comparison of the predicate and subject devices is provided below:

equivalence

Subject DeviceInThrillThrombectomySystemPrimary PredicateMegaVac MechanicalThrombectomySystemPredicatePTA-Plus PTABalloon Cather 5mmX 4cm, PTA-PlusPTA BalloonCatheter 6mm X 4cmPredicateMini-ClotTrieverThrombectomySystem
K NumberK223613K171493K100842K220887
ManufacturerInari Medical, Inc.Capture Vascular, Inc.HotspurTechnologies, Inc.Inari Medical, Inc.
Regulations21 CFR 870.5150Embolectomycatheter21 CFR 870.5150Embolectomycatheter21 CFR 870.5150Embolectomycatheter21 CFR 870.1250Percutaneous catheter21 CFR 870.5150Embolectomycatheter
Product CodeQEWKRAQEWQEXDXEDXELITQEW
Indications forUseThe InThrillThrombectomySystem is indicatedfor:• The non-surgicalremoval of thrombiand emboli fromblood vessels,includingarteriovenousfistulae andarteriovenous graftsfor dialysis access,and synthetic grafts.• Injection, infusion,and/or aspiration ofcontrast media andother fluids into orfrom a bloodvessel/graft.The InThrillThrombectomySystem is intendedfor use in theperipheralvasculature.The MegaVacMechanicalThrombectomySystem is indicatedfor:• The non-surgicalremoval of emboliand thrombi fromblood vessels.• The non-surgicalremoval of thrombifrom syntheticgrafts.• Use in temporaryblood vessel/graftocclusion.• Injection, infusion,and/or aspiration ofcontrast media andother fluids into orfrom a vessel/graft• Catheter placementover a guidewireThe PTA-Plus PTABalloon Catheter isindicated for usewithin syntheticarteriovenous dialysisfistulae to removeembolic material(thrombus/debris)and dilate stenosis fortreatment ofobstructive lesions.The Mini-ClotTrieverThrombectomySystem is indicatedfor:• The non-surgicalremoval of emboliand thrombi fromblood vessels.Injection, infusion,and/or aspiration ofcontrast media andother fluids into orfrom a blood vessel.The Mini-ClotTrieverThrombectomySystem is intendedfor use in theperipheralvasculature.The Mini-ClotTrieverThrombectomySystem is notintended for use indeep vein thrombosis(DVT) treatment.
The InThrillThe MegaVacThe PTA-Plus PTAThe Mini-ClotTriever
ThrombectomyMechanicalBalloon Catheter isThrombectomy
System is notThrombectomydesigned for de-System is a single-
intended for use inSystem is a single-clotting and treatinguse, over-the-wire,
deep vein thrombosisuse, over-the-wire,stenosis in syntheticcatheter-based system
(DVT) treatment.catheter-based systemdialysis fistulae. It isfor the minimally
DeviceThe InThrillfor intravascularan .035" guide-wireinvasive treatment of
DescriptionThrombectomymechanicalcompatible, PTAthromboemboli in the
System is a single-thrombectomy,balloon catheter withperipheral
use, over-the-wire,occlusion, aspirationa proprietary valvevasculature. The
catheter-based systemand embolectomy insystem which allowssystem comprises
for the minimallythe peripheralinjection of contrasttwo main
invasive treatment ofvasculature,and an embolectomycomponents
thromboemboliincluding grafts. Thecoil for mechanicalpackaged separately:
(including organizedsystem comprisesremoval of thrombus.• Mini-ClotTriever
thrombus andtwo mainSheath (8 Fr)
adherent thromboticcomponents:• Mini-ClotTriever
material) in the• MegaVac CatheterCatheter (8 Fr)
peripheral• ThromboWire clotThe Mini-ClotTriever
vasculature,retractorSheath is comprised
includingThe MegaVacof reinforced
arteriovenous fistulaeCatheter withpolymeric coaxial
and arteriovenousSafeSeal technologysheath shafts
grafts for dialysisutilizes a siliconeequipped with a self-
access, and syntheticcoated nitinol braidedexpanding distal
grafts. The systemfunnel that expandsmesh funnel, a
comprises two mainto occlude antegradeflush/aspiration port,
componentsblood flow proximaland a proximal
packaged separately:to the target workhemostatic valve. The
• InThrill Sheath (8zone creating a staticMini-ClotTriever
Fr)environment in whichCatheter is comprised
• InThrillto perform theof reinforced
Thrombectomyintervention, whilepolymeric coaxial
Catheter (8 Fr)also centering andshafts terminating in
The InThrill Sheath issecuring the catheteran expandable nitinol
comprised oftip position within thecoring element
reinforced polymericvessel. The MegaVac(basket). Other
coaxial sheath shaftscatheter's large-accessories provided
equipped with a self-mouth funnel andinclude a pre-dilator
expanding distalinner diameter allowsand dilator.
mesh funnel, afor strong aspiration
flush/aspiration port,while easily being
and a proximalable to pass the
hemostatic valve. TheThromboWire and
InThrillother lesion
Thrombectomydisruptive or
Catheter is comprisedtherapeutic devices
of reinforcedthrough it. The
polymeric coaxial
shafts terminating in
an expandable nitinol
coring element
(basket). Other
accessories provided
include a pre-dilatorand dilator.ThromboWireconsists of a nitinolembolectomyelement that whenexpanded by theproximal actuationhandle can serve togather and pull mattertowards and throughthe MegaVac catheterduring aspiration.
Target VesselPeripheral vessels (4-10 mm) that include:• native vessels• arteriovenousfistulae• arteriovenous grafts• synthetic graftsPeripheral vesselsthat include:• native vessels• synthetic graftsPeripheral vesselsthat include:• arteriovenousfistulae• arteriovenous grafts• synthetic graftsPeripheral vessels (4-10 mm) that include:• native vessels
SterilitySAL 10-6, EOSAL 10-6, EOSAL 10-6, EOSAL 10-6, EO
Shelf-life2 yearsUnknownUnknown2 years *
Guidewirecompatibility0.035"Up to 0.044" OD0.035"0.035"
SheathDimensionsOuter shaft: 0.154"OD/0.137" IDInner shaft: 0.137"OD/0.110" IDLength: 6 cmUp to 9 Fr ODLength: up to 300 cmUnknownOuter shaft: 0.154"OD/0.137" IDInner shaft: 0.137"OD/0.110" IDLength: 6 cm
Shaft MaterialsInner: Pebax 55D,ProPell, PTFE Liner,Stainless steel coil,Radiopaque markerbandOuter: Pebax 63D,ProPell, PTFE LinerUnknownUnknownInner: Pebax 55D,ProPell, PTFE Liner,Stainless steel coil,Radiopaque markerbandOuter: Pebax 40Dand 25D, ProPell,PTFE Liner
HemostasisValve8 Fr Garrote valveRotating swivel hubwith side portUnknownUnknown8 Fr Garrote valveRotating swivel hubwith side port
HandleYes; Slide actuatorenclosed withinhandle housingsYes; Operatorcontrolled – linearactuationUnknownYes; Slide actuatorenclosed withinhandle housings
Braided FunnelOD: 10 mmLength: 0.70"Braided nitinolfunnelOD: 2-12 mmSilicone coatedbraided nitinol funnelUnknownOD: 10 mmLength: 0.70"Braided nitinolfunnel
Side PortTygon tubing withLoctite adhesive1-way stopcock withfemale LuerconnectorUnknownUnknownTygon tubing withLoctite adhesive1-way stopcock withfemale Luerconnector
DilatorUnknownUnknownDilator
OD: 0.110"OD: 0.110"
Total length: 8.8"Total length: 8.3"
Tipped LDPE/HDPETipped LDPE/HDPE
extrusionextrusion
Dilator capDilator cap
Proximal flush portProximal flush port
Pre-DilatorUnknownUnknownOD: 0.13" (10 Fr)
OD: 0.13" (10 Fr)Polypropylene,
Polypropylene,HDPE
HDPELength: 10.2 cm
Length: 10.2 cmOD: 0.111"
CatheterNitinolUnknownOD: 0.111"
Proximal hub withProximal hub with
Tuohy BorstTuohy Borst
hemostasis Y-valvehemostasis Y-valve
and 1-way stopcockand 1-way stopcock
Materials:Materials:
Outer: PTFEOuter: PTFE
Liner, SS304VLiner, SS304V
Braid, RadiopaqueBraid,
marker band, 63DRadiopaque
Pebax Jacket, 63Dmarker band, 63D
Pebax Fluoro-safePebax Jacket, 63D
marker bandPebax Fluoro-safe
Middle: Braidedmarker band
polyimide,Middle: Braided
Radiopaquepolyimide,
marker band,Radiopaque
Pebax 72Dmarker band,
Inner: BraidedLoctite 3942
polyimide,Inner: Braided
Radiopaque 55Dpolyimide,
Pebax tip withRadiopaque 55D
ProPellPebax tip with
ProPell
Length65 cmUp to 300 cm55 cm65 cm
Coring ElementLaser-cut nitinolWoven nitinolUnknownLaser-cut nitinol
OD: 18 mmOD: 2-9 mmOD: 18 mm
Length: 88 mmLength: 88 mm

{5}------------------------------------------------

K223613

PAGE 3 OF 6

{6}------------------------------------------------

K223613

PAGE 4 OF 6

{7}------------------------------------------------

K223613

PAGE 5 OF 6

Summary of substantial equivalence

Biocompatibility

The material changes proposed in this submission have no impact on the established biocompatibility of the device. Therefore, the previous passing results demonstrating that the InThrill Thrombectomy System (K220887) and accessories meet biological safety requirements per ISO 10993-1 are still applicable.

Sterilization

The subject device, including its accessories, is sterilized using EtO to achieve a sterility assurance level (SAL) of 106 using a validated sterilization process in accordance with the principles of ISO 11135:2014/Amd 1:2018 and AAMI TIR 28:2016. There have been no changes proposed that would affect device

{8}------------------------------------------------

sterilization; therefore, the previous sterilization process per K220887 remains applicable.

Non-Clinical Testing

In accordance with the Design Failure Modes and Effects Analysis, verification and validation testing were identified to support the substantial equivalence of the InThrill Thrombectomy System to the predicate devices. These tests included:

Performance Tests

  • Graft Leak Testing (Pre-Simulated Use) ●
  • In-Graft Simulated Use, InThrill Thrombectomy System ●
  • . Graft Leak Testing (Post-Simulated Use)
  • Graft Visual Inspection (Post-Simulated Use) ●
  • Radial Force Testing
  • Prescale Contact Paper Pressure Testing ●
  • Graft Abrasion Testing
  • Comparative Adherent Clot Removal Testing ●
  • Visual Inspection of Vein-to-Graft Anastomosis (Pre-Simulated Use)
  • Simulated Use, InThrill Catheter through Vein-to-Graft Anastomosis ●
  • Visual Inspection of Vein-to-Graft Anastomosis (Post-Simulated ● Use)

Test results demonstrated that all acceptance criteria were met; therefore, the device conforms to established product specifications.

Pre-Clinical Study

A GLP Animal study was performed, which did not identify any new questions of safety or effectiveness for the InThrill Thrombectomy System in peripheral vessels within the indicated size range.

Clinical testing was not required for the determination of substantial equivalence.

Conclusion

The non-clinical tests demonstrate that the subject device does not raise new questions of safety or effectiveness and, therefore, is substantially equivalent to the predicate device.

§ 870.5150 Embolectomy catheter.

(a)
Identification. An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.(b)
Classification. Class II (performance standards).