K Number
K223584

Validate with FDA (Live)

Date Cleared
2023-08-12

(254 days)

Product Code
Regulation Number
880.5200
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Pre-Filled Normal Saline Flush Syringe is intended for use in flushing compatible intravenous administration sets and indwelling intravenous access devices. Use according to the recommendations of the manufacturer for the appropriate device.

Device Description

Pre-Filled Normal Saline Flush Syringe is a polypropylene syringe filled with 0.9% sodium chloride for injection. It contains 3ml, 5ml and 10ml and consists of tip cap, barrel, piston, and plunger. This is a single use, disposable device(s), provided sterile.

AI/ML Overview

The provided document is a 510(k) summary for a medical device called "Pre-Filled Normal Saline Flush Syringe." This document focuses on demonstrating substantial equivalence to a legally marketed predicate device through non-clinical testing. It does not describe a study involving an AI/CADe device or human readers. Therefore, I cannot provide information regarding acceptance criteria and studies typical for AI/CADe devices, such as those related to multi-reader multi-case studies, effect size of AI assistance, expert qualifications, or ground truth establishment for training and test sets in an AI context.

However, I can extract the acceptance criteria and performance data for the non-clinical (bench) tests conducted on the Pre-Filled Normal Saline Flush Syringe to demonstrate its safety and effectiveness.

Here's the information based on the provided text, specifically from Section VII "Summary of Non-clinical Testing (Bench)":


1. Table of Acceptance Criteria and the Reported Device Performance

ID#TestMethodAcceptance CriteriaConclusion (Reported Device Performance)
1.1LubricantISO7886-1ISO7886-1Pass
1.2Dead SpaceISO7886-1ISO7886-1Pass
1.3Limits for acidity or alkalinityISO7886-1ISO7886-1Pass
1.4Syringe Luer PerformanceISO 80369-7ISO 80369-7Pass
1.5Sealing performanceISO7886-1ISO7886-1Pass
2.1pH value (Sodium Chloride Injection, USP)USP<791>PH: 4.5-7.0Pass
2.2Oxidizable substance testUSP6-471USP6-471Pass
2.3Carbonate (Sodium Chloride Injection, USP)USP<191>USP<191>Pass
2.4Sulfate (Sodium Chloride Injection, USP)USP<191>USP<191>Pass
2.5Calcium (Sodium Chloride Injection, USP)USP<191>USP<191>Pass
2.6Ammonium (Sodium Chloride Injection, USP)USP<191>USP<191>Pass
2.7Iron test (Sodium Chloride Injection, USP)USP<241>< 2ppmPass
2.8Limits of extractable metalsUSP<233>, USP<232>USP<233>, USP<232>Pass
3Particulate ContaminationAAMI TIR42:2021≥10μm, ≤6000; ≥25μm, ≤600≥10μm, ≤361.5; ≥25μm, ≤0.0
4.1Bacterial Endotoxins TestUSP 43<85>Bacterial endotoxins ≤ 0.5EU/mLPass
4.2In Vitro CytotoxicityISO 10995-5:2009 (should be ISO 10993-5)Non-cytotoxicPass
4.3Intracutaneous Reactivity TestISO 10995-23:2021 (should be ISO 10993-23)Non-irritantPass
4.4Skin Sensitization TestISO 10993-10:2021Non-sensitizerPass
4.5Acute Systemic Toxicity TestISO 10993-11:2017No systemic toxicityPass
4.6Pyrogen TestISO 10993-11:2017Non-pyrogenPass
4.7In Vitro Hemolysis TestISO 10993-4:2017Non-hemolyticPass

Note on ISO standards in the table: There appears to be a typo for test IDs 4.2 and 4.3 in the provided document (ISO 10995 instead of ISO 10993 for biocompatibility tests). I have kept the text as written but noted the likely correct standard.

2. Sample size used for the test set and the data provenance

  • Sample Size: "Real time aged samples from three non-consecutive lots were tested for all performance criteria."
  • Data Provenance: The document does not specify the country of origin for the test samples or explicitly state if the study was retrospective or prospective. It implies prospective testing of manufactured lots.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This is a non-clinical, bench testing study for a physical medical device, not an AI/CADe device requiring expert-established ground truth from medical images or patient data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as this is bench testing. The results are based on objective measurements against established standard criteria.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study was not done. This device is not an AI/CADe device, and the study did not involve human readers or AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, in a sense. The "device" in this context is the physical syringe. The performance tests conducted are standalone tests of the device itself (its materials, chemical contents, physical properties, and biocompatibility), without specific human-in-the-loop performance measurement beyond standard laboratory procedures.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for this device's performance is established by internationally recognized consensus standards (e.g., ISO, USP, AAMI TIR) which define the acceptable limits and test methodologies for the physical, chemical, and biological properties of the device.

8. The sample size for the training set

Not applicable. This is not an AI/ML device that requires a training set.

9. How the ground truth for the training set was established

Not applicable. This is not an AI/ML device that requires a training set or ground truth establishment for it.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 12, 2023

Anhui Tianyang Pharmaceutical Co., Ltd. Zhang Shunlin Quality Manager 46 Tiantong Road, Tianchang City, Anhui Province Tianchang. Anhui China

Re: K223584

Trade/Device Name: Pre-Filled Normal Saline Flush Syringe Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: Class II Product Code: NGT Dated: July 13, 2023 Received: July 14, 2023

Dear Zhang Shunlin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Bifeng Qian -S

Bifeng Oian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223584

Device Name Pre-Filled Normal Saline Flush Syringe

Indications for Use (Describe)

Pre-Filled Normal Saline Flush Syringe is intended for use in flushing compatible intravenous administration sets and indwelling intravenous access devices. Use according to the recommendations of the manufacturer for the appropriate device.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary (K223584)

l 510(k) Submitter

Device Submitter:Anhui Tianyang Pharmaceutical Co., LTD.
No.46, Tiantong Road, Tianchang City, Anhui Province
Contact Person:Zhang Sunlin
Quality Manager
Phone:+86-15805509075
E-mail:752322508@qq.com

II Device

Trade Name of Device:Pre-Filled Normal Saline Flush Syringe
Regulation Number:21 CFR 880.5200
Classification Name:Saline, Vascular Access Flush
Product Code:NGT
Regulation Number:21 CFR 880.5200
Regulatory ClassII
Review PanelGeneral Hospital

III Predicate Devices

510k NumberK213522
Trade Name of Device:AMSafe® Pre-Filled Normal Saline Flush Syringe
Regulation Number:21 CFR 880.5200
Regulation Name:Intravascular Catheter
Regulatory ClassII
Product Code:NGT

IV Device Description

Pre-Filled Normal Saline Flush Syringe is a polypropylene syringe filled with 0.9% sodium chloride for injection. It contains 3ml, 5ml and 10ml and consists of tip cap, barrel, piston, and plunger. This is a single use, disposable device(s), provided sterile.

V Indications for use

Pre-Filled Normal Saline Flush Syringe is intended for use in flushing compatible intravenous administration sets and indwelling intravenous access devices. Use according to the recommendations of the manufacturer for the appropriate device.

VI Technological Characteristics Comparison

VI-1: Comparison of Pre-Filled 0.9% Normal Saline Flush Syringe

{4}------------------------------------------------

DeviceCharacteristicSubject Device(K223584)Predicate Device(K213522)Discussion
Indications forUsePre-Filled Normal Saline FlushSyringe is intended for use influshing compatibleintravenous administrationsets and indwellingintravenous access devices.Use according to therecommendations of themanufacturer for theappropriate device.The AMSafe® PreFilledNormal Saline FlushSyringe, is intended for usein flushing compatibleintravenousadministration sets andindwelling intravenousaccess devices. Useaccording to therecommendations of themanufacturer for theappropriate device.SimilarComment 1
Prescription/over-the counteruseFor Rx onlyFor Rx onlyIdentical
OperationPrincipleThe Pre-Filled Normal SalineFlush Syringe is flushed afterthe locking connector isconnected to the medicalcatheter connector for clinicaluse the liquid medicine (0.9%sodium chloride injection) waspushed into the medicalcatheter and used to closeand flush the end of thecatheter in the gap betweendifferent drugs.The AMSafe® PreFilledNormal Saline Flush Syringeis a three-piece, sterile,single use syringe with a 6%(Luer) connector pre-filledwith 0.9% Sodium ChlorideInjection, USP, and sealedwith a tip cap.Identical
DesignThe subject device with Luerlock connection fitting andnonvented, female Luer locktip cap.The device has modified toadd an extra thread to theplunger rod and inside ofplunger stopper, the femaleLuer cap has Changed toscrew type.DifferentComment 2
Chemicalcomposition0.9% Sodium chlorideinjection, USP0.9% Sodium chlorideinjection, USPIdentical
Syringe materialTip Cap, Barrel and Plunger:polypropylene;Piston: Bromobutyl rubberBarrel and plunger:polypropyleneStopper: Butyl rubber (notmade with natural rubberIdentical
DeviceCharacteristicSubject DevicePredicate Device(K213522)Discussion
latex)
Tip cap: polypropylene withwhite colorant
Syringe Sizeand Fill VolumesFill 3ml, 5ml, 10ml in 10ccsyringeFill 3ml, 5ml, 10ml in 10ccsyringeIdentical
Fill 3ml, 5ml in 5cc syringeFill 3ml, 5ml in 5cc syringe
SyringePackagingBOPP heat sealing filmPP wrapDifferentComment 3
Sterilizationmethod andSAL LevelTerminally sterilized by steam,10-6 SALTerminally sterilized bysteam, 10-6 SALIdentical
LabelednonpyrogenicYesYesIdentical
Singe Use OnlyYesYesIdentical
Shelf Life2 Years3 YearsSimilar

{5}------------------------------------------------

Comment 1: The Indications for Use of the predicate and subject device are same and the intended use is same, which does not affect the safety and effectiveness of the product.

Comment 2: The Luer cap of the predicate device has been modified, and the subject device is a standard Luer connector that meets the requirements of ISO 80369-7, so it does not affect the safety and effectiveness of the product.

Comment 3: The packaging materials of the subject and predicate are different, but the packaging of the subject device has been verified, and the sterility of the product can be guaranteed within the claimed shelf life of 2 years. Therefore, the safety and effectiveness of the product will not be affected.

VII Summary of Non-clinical Testing (Bench)

The non-clinical testing for Pre-Filled Normal Saline Flush Syringe was performed to demonstrate verification testing in conformance with the acceptance criteria of test methods and recognized consensus standards shown below. Real time aged samples from three non-consequtive lots were tested for all performance criteria.

Table VII-1. Performance testing was conducted on the subject device
ID#TestMethodAcceptance CriteriaConclusion
1Physical Testing of Syringe
1.1LubricantISO7886-1ISO7886-1Pass
1.2Dead SpaceISO7886-1ISO7886-1Pass
1.3Limits for acidity oralkalinityISO7886-1ISO7886-1Pass
1.4Syringe LuerPerformanceISO 80369-7ISO 80369-7Pass
1.5SealingperformanceISO7886-1ISO7886-1Pass
2Sodium Chloride Injection, USP Testing
2.1pH valueUSP<791>PH: 4.5-7.0Pass
2.2Oxidizablesubstance testUSP6-471USP6-471Pass
2.3CarbonateUSP<191>USP<191>Pass
2.4SulfateUSP<191>USP<191>Pass
2.5CalciumUSP<191>USP<191>Pass
2.6AmmoniumUSP<191>USP<191>Pass
2.7Iron testUSP<241>< 2ppmPass
2.8Limits ofextractable metalsUSP<233>USP<232>USP<233>USP<232>Pass
3ParticulateContaminationAAMI TIR42:2021≥10μm, ≤6000≥25μm, ≤600≥10μm, ≤361.5≥25μm, ≤0.0
4Biocompatibility Testing
4.1BacterialEndotoxins TestUSP 43<85>Bacterial endotoxins≤0.5EU/mLPass
4.2In Vitro CytotoxicityISO 10995-5:2009Non-cytotoxicPass
4.3IntracutaneousReactivity TestISO 10995-23:2021Non-irritantPass
4.4Skin SensitizationTestISO 10993-10:2021Non-sensitizerPass
4.5Acute SystemicToxicity TestISO 10993-11:2017No systemic toxicityPass
4.6Pyrogen TestISO 10993-11:2017Non-pyrogenPass
4.7In Vitro HemolysisTestISO 10993-4:2017Non-hemolyticPass

Table VII-1: Performance testing was conducted on the subject device

{6}------------------------------------------------

VIII Clinical Test Conclusion

No clinical study is included in this submission.

IX Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject Pre-Filled Normal

{7}------------------------------------------------

Saline Flush Syringe is as safe as effective, and performs as well as or better than the legally marketed device.

Date of Summary: August 8, 2023

§ 880.5200 Intravascular catheter.

(a)
Identification. An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.(b)
Classification. Class II (performance standards).