K Number
K213522

Validate with FDA (Live)

Date Cleared
2022-03-22

(139 days)

Product Code
Regulation Number
880.5200
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AMSafe® Pre-Filled Normal Saline Flush Syringe, is intended for use in flushing compatible intravenous administration sets and indwelling intravenous access devices. Use according to the recommendations of the manufacturer for the appropriate device.

Device Description

AMSafe® Pre-Filled Normal Saline Flush Syringe is a polypropylene plastic syringe filled with 0.9% sodium chloride for injection, USP, and capped with a polypropylene syringe tip cap.

AI/ML Overview

This document describes the premarket notification (510(k)) for the AMSafe® Pre-Filled Normal Saline Flush Syringe. The information provided outlines the device's technical characteristics, intended use, and comparison to a legally marketed predicate device (K183473). The study performed is a non-clinical bench testing study, not an AI/ML-based clinical study, therefore, many of the requested elements for an AI/ML study (e.g., sample size for test set, number of experts, MRMC study, training set details) are not applicable.

Here's a breakdown of the requested information based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria and reported device performance are detailed in "Table 5-2" on page 6 of the document.

ID#TestMethodAcceptance CriteriaConclusion
1Physical testing of syringeISO7886-1 ISO80369-7ISO7886-1 ISO80369-7Pass
Integrity test of packageASTM F2338-09No leakagePass
Dimension testISO80369-7ISO80369-7Pass
Lubricant of syringe testISO7886-1ISO7886-1Pass
Dead space testISO7886-1ISO7886-1Pass
Limits of acidity or alkalinity of syringeISO7886-1ISO7886-1Pass
2Sodium Chloride Injection, USP Testing
pH valueUSP<791>PH: 4.5-7.0Pass
Chemical Identification TestsUSP<191>USP<191>Pass
0.9% normal saline content testUSP6-4660.86% -- 0.94%Pass
Oxidizable substance testUSP6-471USP6-471Pass
Iron testUSP<241>< 2ppmPass
AmmoniumUSP<191>USP<191>Pass
CalciumUSP<191>USP<191>Pass
CarbonateUSP<191>USP<191>Pass
SulfateUSP<191>USP<191>Pass
Total organic carbonUSP<643>USP<643>Pass
Limits of extractable metalsUSP<233> USP<232>USP<233> USP<232>Pass
3Particulate matterUSP<788>≥10um, ≤6000 ≥25um, ≤600Pass
4Biocompatibility testing
Bacterial endotoxins testUSP<85>Bacterial endotoxins≤0.5EU/mLPass
Acute systemic toxicityISO10993-11No systemic toxicityPass
Intracutaneous reactivityISO10993-10Non-irritantPass
Pyrogen testISO10993-11Non-pyrogenPass
Skin sensitizationISO10993-10Non-sensitizerPass
In vitro cytotoxicityISO10993-5Non-cytotoxicPass
In vitro hemolysis propertiesASTM F756-17Non-hemolyticPass

Additionally, shelf life was evaluated:

  • Test: Shelf life
  • Method: ASTM F1980-16
  • Acceptance Criteria/Conclusion: Verify that the device will remain within specification during the prescribed shelf life when stored under labeled conditions. (Implied Pass, as per overall conclusion).

2. Sample Size Used for the Test Set and Data Provenance

The document describes non-clinical bench testing of manufacturing characteristics and chemical composition, not an AI/ML-based study with a "test set" composed of patient data. Therefore, the concept of sample size for a test set (e.g., number of images, patient records) and data provenance (country of origin, retrospective/prospective) as typically understood for AI/ML studies are not applicable. The tests were performed on samples of the device itself. The specific number of units tested for each criterion is not provided, but it would typically be a predetermined number based on the statistical requirements of the specific test methods (e.g., ISO, ASTM, USP standards).

3. Number of Experts Used to Establish Ground Truth and Qualifications

Not Applicable. This was a bench study, not an AI/ML study requiring expert labeling of data. The "ground truth" for these tests are the established physical, chemical, and biological standards and methodologies outlined in the referenced ISO, ASTM, and USP documents.

4. Adjudication Method for the Test Set

Not Applicable. This was a bench study; there was no human-in-the-loop expert adjudication process as would be typical for AI/ML algorithm evaluation.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC study was not done. This is a review of a physical medical device (pre-filled syringe), not an AI/ML diagnostic or assistive tool where human reader performance would be a relevant metric.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

Not Applicable. This document describes the performance of a physical medical device, not an algorithm.

7. The Type of Ground Truth Used

The "ground truth" for the non-clinical tests is based on established industry standards and pharmacopeial specifications. This includes:

  • Physical Properties: Standards like ISO7886-1, ISO80369-7, ASTM F2338-09.
  • Chemical Composition: United States Pharmacopeia (USP) monographs such as USP<791>, USP<191>, USP6-466, USP6-471, USP<241>, USP<643>, USP<233>, USP<232>.
  • Biocompatibility: ISO10993 series (ISO10993-11, ISO10993-10, ISO10993-5) and ASTM F756-17.
  • Sterility Assurance: The mention of 10-6 SAL (Sterility Assurance Level) indicates adherence to established sterilization standards.

8. The Sample Size for the Training Set

Not Applicable. This is a physical device, not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not Applicable. As there is no training set for a physical device, there is no ground truth establishment process in that context.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 22, 2022

Amsino International, Inc. Jane Gao VP of R&D 708 Corporate Center Drive Pomona, California 91768

Re: K213522

Trade/Device Name: AMSafe® Pre-Filled Normal Saline Flush Syringe Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: Class II Product Code: NGT Dated: March 8, 2022 Received: March 14, 2022

Dear Jane Gao:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213522

Device Name

AMSafe® Pre-Filled Normal Saline Flush Syringe

Indications for Use (Describe)

The AMSafe® Pre-Filled Normal Saline Flush Syringe, is intended for use in flushing compatible intravenous administration sets and indwelling intravenous access devices. Use according to the recommendations of the manufacturer for the appropriate device.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K213522 Summary

a) Submitter Information:

Submitter:Amsino International Inc.708 Corporate Center DrivePomona, CA 91768Phone: +1 (909)626-5888Fax: +1 (909)626-3888
Contact Person:Jane GaoVP of R&D of Amsino InternationalCell phone: +86 139 1614 7664Jane_gao@amsino.com
Date of Summary:March 22, 2022

b) Device Information:

Trade or Proprietary Name:AMSafe® Pre-Filled Normal Saline Flush Syringe
Common or Usual Name:Pre-Filled Normal Saline Flush Syringe
Classification Name:Saline, vascular access flush
Product Code:NGT
Regulation Number:880.5200
Device Classification:II
Review Panel:General Hospital

c) Identification of Legally Marketed Device(s):

AMSafe® Pre-Filled Normal Saline Flush Syringe, 510(k) number K183473.

d) Device Description:

AMSafe® Pre-Filled Normal Saline Flush Syringe is a polypropylene plastic syringe filled with 0.9% sodium chloride for injection, USP, and capped with a polypropylene syringe tip cap.

e) Indications for Use:

The AMSafe® Pre-Filled Normal Saline Flush Syringe, is intended for use in flushing compatible intravenous administration sets and indwelling intravenous access devices. Use according to the recommendations of the manufacturer for the appropriate device.

f) Technological Characteristics Comparison:

The main difference between subject device and the predicate device are different material suppliers and adding an extra thread to the plunger rod and inside of plunger stopper, and subject device is terminally sterilized by steam sterilization. The device will be marketed as 10mL syringe with a 3mL, 5mL or 10mL fill volume, and a 5mL syringe with 3mL or 5mL fill volume according to the market needs. The products are in PP wrapper as a dust cover for non

{4}------------------------------------------------

sterile field.

  • Technological characteristics: The subject device has the same technological characteristics and provide the same principle of operation as the predicate device.
  • Intended use: The subject device has the same intended use as the predicate device.

Comparison between subject device and predicate device.

able15
DeviceCharacteristicSubject devicePredicate device(K183473)Discussion
Indicationsfor UseThe AMSafe® Pre-Filled Normal SalineFlush Syringe, isintended for use influshing compatibleintravenousadministration setsand indwellingintravenous accessdevices. Useaccording to therecommendations ofthe manufacturer forthe appropriatedevice.The AMSafe 0.9%sodium chloride pre-filled normal salineflush syringe, isintended for use influshing compatibleintravenousadministration setsand indwellingintravenous accessdevices. Useaccording to therecommend at ionsof the manufacturerfor the appropriatedevice.Identical
Prescription/over-thecounter useFor Rx onlyFor Rx onlyIdentical
OperationPrincipleThe AMSafe® Pre-Filled Normal SalineFlush Syringe is athree-piece, sterile,single use syringewith a 6% (Luer)connector pre-filledwith 0.9% SodiumChloride Injection,USP, and sealedwith a tip cap.The AMSafe® Pre-Filled Normal SalineFlush Syringe is athree-piece, sterile,single use syringewith a 6% (Luer)connector pre-filledwith 0.9% SodiumChloride Injection,USP, and sealedwith a tip cap.Identical
DesignThe subject devicehas modified to addan extra thread tothe plunger rod andinside of plungerstopper, the femaleLuer cap hasPrefilled NormalSaline plastic pistonsyringe with Luerlock connectionfitting and non-vented, female Luerlock tip capIdentical
changed to screwtype.
Chemicalcomposition0.9% Sodiumchloride injection,USP0.9% Sodiumchloride injection,USPIdentical
SyringematerialBarrel and plunger:polypropyleneStopper:Chlorobutyl rubber(not made withnatural rubber latex)Tip cap:polypropylene withwhite colorantBarrel and plunger:polypropyleneStopper: Butylrubber (not madewith natural rubberlatex)Tip cap:polypropylene withwhite colorantDifferent
Syringe Sizeand FillVolumesFill 3ml, 5ml, 10mlin 10cc syringeFill 3ml, 5ml in 5ccsyringeFill 3ml, 5ml, 10mlvolume in 12ccsyringeFill 20ml volume in20cc syringeDifferent
SyringepackagingPP wrapPP wrap or Sterilebarrier Plastic peelpouch
Sterilizationmethod andSAL LevelTerminally sterilizedby steam, 10-6 SALTerminally sterilizedby Gammaradiation, 10-6 SALDifferent
Labeled non-pyrogenicYesYesIdentical
Single useonlyYesYesIdentical
Shelf Life3 years2 yearsDifferent

{5}------------------------------------------------

g) Summary of Non-clinical Testing (Bench):

The non-clinical testing for AMSafe® Pre-Filled Normal Saline Flush Syringe was performed to demonstrate verification testing in conformance with the acceptance criteria of test methods and recognized consensus standards shown below.

The following performance testing was conducted on the proposed device:

{6}------------------------------------------------

Table 5-2

ID#TestMethodAcceptance criteriaConclusion
1Physical testing ofsyringeISO7886-1ISO80369-7ISO7886-1ISO80369-7Pass
Integrity test ofpackageASTM F2338-09No leakagePass
Dimension testISO80369-7ISO80369-7Pass
Lubricant of syringetestISO7886-1ISO7886-1Pass
Dead space testISO7886-1ISO7886-1Pass
Limits of acidity oralkalinity of syringeISO7886-1ISO7886-1Pass
2Sodium Chloride Injection, USP Testing
pH valueUSP<791>PH: 4.5-7.0Pass
ChemicalIdentification TestsUSP<191>USP<191>Pass
0.9% normal salinecontent testUSP6-4660.86% -- 0.94%Pass
Oxidizablesubstance testUSP6-471USP6-471Pass
Iron testUSP<241>< 2ppmPass
AmmoniumUSP<191>USP<191>Pass
CalciumUSP<191>USP<191>Pass
CarbonateUSP<191>USP<191>Pass
SulfateUSP<191>USP<191>Pass
Total organic carbonUSP<643>USP<643>Pass
Limits of extractablemetalsUSP<233>USP<232>USP<233>USP<232>Pass
3Particulate matterUSP<788>≥10um, ≤6000≥25um, ≤600Pass
4Biocompatibility testing
Bacterial endotoxinstestUSP<85>Bacterialendotoxins≤0.5EU/mLPass
Acute systemictoxicityISO10993-11No systemic toxicityPass
IntracutaneousreactivityISO10993-10Non-irritantPass
Pyrogen testISO10993-11Non-pyrogenPass
Skin sensitizationISO10993-10Non-sensitizerPass
In vitro cytotoxicityISO10993-5Non-cytotoxicPass
In vitro hemolysis propertiesASTM F756-17Non-hemolyticPass

{7}------------------------------------------------

The shelf life of the final finished sterilized device was evaluated according to the recognized consensus standard ASTM F1980-16 to verify that the subject device will remain within specification during the prescribed shelf life when stored under the labeled storage conditions.

h) Conclusions:

The conclusions drawn from the nonclinical tests demonstrate that the AMSafe® Pre-Filled Normal Saline Flush syringe is as safe as effective, and performs as well as or better than the legally marketed device

§ 880.5200 Intravascular catheter.

(a)
Identification. An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.(b)
Classification. Class II (performance standards).