K Number
K223559

Validate with FDA (Live)

Date Cleared
2023-02-09

(73 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Nephron Nitrile Powder-Free Nitrile Examination Gloves (Tested For Use With Chemotherapy Drugs) is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

The following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes:

Bleomycin Sulfate (15.0 mg/ml)
Busulfan (6.0 mg/ml)
Carboplatin (10.0 mg/ml)
Cisplatin (1.0 mg/ml)
Cyclophosphamide (20.0 mg/ml)
Cytarabine (100.0 mg/ml)
Dacarbazine (10.0 mg/ml)
Daunorubicin HCl (5.0 mg/ml)
Docetaxel (10.0 mg/ml)
Doxorubicin HCl (2.0 mg/ml)
Epirubicin HCl (2.0 mg/ml)
Etoposide (20.0 mg/ml)
Fludarabine (25.0 mg/ml)
Fluorouracil (50.0 mg/ml)
Gemcitabine (38.0 mg/ml)
Idarubicin HCl (1.0 mg/ml)
Ifosfamide (50.0 mg/ml)
Irinotecan (20.0 mg/ml)
Mechlorethamine HCl (1.0 mg/ml)
Melphalan (5.0 mg/ml)
Methotrexate (25.0 mg/ml)
Mitomycin C (0.5 mg/ml)
Mitoxantrone HCl (2.0 mg/ml)
Paclitaxel (6.0 mg/ml)
Rituximab (10.0 mg/ml)
Trisenox (1.0 mg/ml)
Vincristine Sulfate (1.0 mg/ml)

The following chemotherapy drugs have low permeation times: Carmustine (3.3 mg/ml) : 33.8 minutes Thiotepa (10.0 mg/ml) : 128.1 minutes

Warning: Not for Use with: Carmustine, Thiotepa

Device Description

Nephron Nitrile Powder-Free Nitrile Examination Gloves (Tested For Use With Chemotherapy Drugs) is a Class I, patient examination gloves bearing the product codes LZA, LZC, OPJ (21CFR880.6250). They meet all the current specifications listed under the ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application and also complies with requirements for Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs as per ASTM D6978-05 (2019). They are made from Nitrile (NBR). These gloves are blue in color and are powder free. The product is non-sterile, fingertip textured, ambidextrous with beaded cuff and single use only.

Nephron Nitrile Powder-Free Nitrile Examination Gloves (Tested For Use With Chemotherapy Drugs) with sizes Medium, Large, X-Large and XX-Large are included in the submission.

AI/ML Overview

The provided text describes the acceptance criteria and performance of the Nephron Nitrile Powder-Free Nitrile Examination Gloves.

Here's the requested information:

1. A table of acceptance criteria and the reported device performance

Test Method (Standard)PurposeAcceptance CriteriaReported Device Performance
ASTM D6319-19 (Dimensions - Length)To determine the length of the glovesMedium: 230 mm minLarge: 230 mm minX-Large: 230 mm minXX-Large: 230 mm minMedium: 235 mmLarge: 237 mmX-Large: 250 mmXX-Large: 238 mm
ASTM D6319-19 (Dimensions - Width)To determine the width of the glovesMedium: 95+/-10 mmLarge: 110+/-10 mmX-Large: 120+/-10 mmXX-Large: 130+/-10 mmMedium: 95 mmLarge: 113 mmX-Large: 121 mmXX-Large: 129 mm
ASTM D6319-19 (Thickness)To determine the thickness of the glovesPalm: 0.05 mm min for all sizesFinger: 0.05 mm min for all sizesSize Palm (Avg value) Finger (Avg value)Medium 0.077 mm 0.111 mmLarge 0.106 mm 0.109 mmX-Large 0.089 mm 0.115 mmXX-Large 0.113 mm 0.107 mm
ASTM D6319-19 (Physical Properties - Tensile Strength)To determine the physical properties - Tensile strengthBefore Ageing Tensile Strength: 14MPa min for all sizesAfter Ageing Tensile Strength: 14MPa min for all sizesSize Before ageing After ageingMedium 34.0 MPa 37.3 MPa
ASTM D6319-19 (Physical Properties - Ultimate Elongation)To determine the physical properties - Ultimate ElongationBefore Ageing Ultimate Elongation: 500% min for all sizesAfter Ageing Ultimate Elongation: 400% min for all sizesSize Before ageing After ageingMedium 542% 503%
ASTM D5151-19 (Freedom from holes)To determine the holes in the glovesAQL 2.5Gloves Passes AQL 2.5
ASTM D6124-06 (Residual Powder)To determine the residual powder in the gloves≤ 2 mg/gloveMedium: 0.3516 mg/glove
ASTM D6978-05 (Chemotherapy Drugs Permeation)To determine the breakthrough detection time of chemotherapy drugsFor listed drugs (e.g., Bleomycin Sulfate, Busulfan, Carboplatin, Cisplatin, Cyclophosphamide, Cytarabine, Dacarbazine, Daunorubicin HCl, Docetaxel, Doxorubicin HCl, Epirubicin HCl, Etoposide, Fludarabine, Fluorouracil, Gemcitabine, Idarubicin HCl, Ifosfamide, Irinotecan, Mechlorethamine HCl, Melphalan, Methotrexate, Mitomycin C, Mitoxantrone HCl, Paclitaxel, Rituximab, Trisenox, Vincristine Sulfate): >240 Minutes. Carmustine (3.3 mg/ml): Not specified as a pass/fail criterion, but a low permeation time is noted. Thiotepa (10.0 mg/ml): Not specified as a pass/fail criterion, but a low permeation time is noted.For all 28 listed chemotherapy drugs, the breakthrough detection time was >240 Minutes. Carmustine (3.3 mg/ml) : 33.8 minutes Thiotepa (10.0 mg/ml) : 128.1 minutes
ISO 10993-23 (Primary Skin Irritation)To evaluate the local dermal irritation in rabbitsUnder the condition of study not an irritantUnder the conditions of the study, the test article met the requirements of the test.
ISO 10993-10 (Dermal Sensitization)To evaluate the skin sensitization in Guinea pigsUnder the conditions of the study, not a sensitizerUnder the conditions of the study, the test article was not considered a sensitizer.
ISO 10993-5 (In Vitro Cytotoxicity)To determine the potential to cause cytotoxicityUnder the conditions of the study, non-cytotoxicThe undiluted test article extract and 50% test article extract dilution did not meet the requirements of the test and the 25%, 12.5%, 6.25%, and 3.13% test article extract dilutions met the requirements of the test. The cytotoxic concern was addressed via acute systemic toxicity testing. (This result indicates some cytotoxicity at higher concentrations but was mitigated by systemic toxicity testing).
ISO 10993-11 (Acute Systemic Toxicity)To evaluate the acute systemic toxicity in miceUnder the conditions of study, the device extracts do not pose a systemic toxicity concernUnder the conditions of study, there was no mortality or evidence of systemic toxicity.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

The document does not explicitly state the sample sizes for each test in the level of detail (e.g., number of gloves tested for each specific property or chemotherapy drug). It only refers to compliance with the standards (e.g., AQL 2.5 for freedom from holes).

The data provenance (country of origin, retrospective/prospective) is not mentioned. These are bench tests, not human trials or medical imaging data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

This information is not applicable. The document describes bench testing of physical and chemical properties of gloves, not expert interpretation of medical images or clinical data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable. The tests are laboratory-based and follow standardized testing protocols, which do not typically involve adjudication methods like those used for expert review of clinical cases.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. The document describes the testing of medical examination gloves, which are physical products, not AI systems or medical imaging devices. Therefore, MRMC studies and AI assistance are irrelevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. The document describes the testing of medical examination gloves, which are physical products, not algorithms.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for these tests is defined by the technical specifications and performance requirements outlined in the referenced ASTM and ISO standards. For example:

  • Physical properties (length, width, thickness, tensile strength, elongation) are measured objectively against defined numerical limits.
  • Freedom from holes is assessed against an Acceptable Quality Limit (AQL), a statistical measure.
  • Resistance to chemotherapy drugs is measured by detecting breakthrough time using specific analytical methods defined in ASTM D6978.
  • Biocompatibility tests (irritation, sensitization, cytotoxicity, systemic toxicity) are assessed by observing biological responses in animal models or in vitro systems against established criteria for non-irritancy, non-sensitization, and non-toxicity according to ISO standards.

Essentially, the "ground truth" reflects compliance with regulatory and industry standards for glove performance and safety.

8. The sample size for the training set

This information is not applicable, as this is not a machine learning model.

9. How the ground truth for the training set was established

This information is not applicable, as this is not a machine learning model.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 9, 2023

Nephron Pharmaceuticals Corporation % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114

Re: K223559

Trade/Device Name: Nephron Nitrile Powder-Free Nitrile Examination Gloves (Tested For Use With Chemotherapy Drugs) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-powdered patient examination glove Regulatory Class: Class I. reserved Product Code: LZA, LZC, OPJ Dated: February 1, 2023 Received: February 1, 2023

Dear Prithul Bom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Allan Gu

For Bifeng Oian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223559

Device Name

Nephron Nitrile Powder-Free Nitrile Examination Gloves (Tested For Use With Chemotherapy Drugs)

Indications for Use (Describe)

Nephron Nitrile Powder-Free Nitrile Examination Gloves (Tested For Use With Chemotherapy Drugs) is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

The following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes:

Bleomycin Sulfate (15.0 mg/ml)Doxorubicin HCl (2.0 mg/ml)Mechlorethamine HCl (1.0 mg/ml)
Busulfan (6.0 mg/ml)Epirubicin HCl (2.0 mg/ml)Melphalan (5.0 mg/ml)
Carboplatin (10.0 mg/ml)Etoposide (20.0 mg/ml)Methotrexate (25.0 mg/ml)
Cisplatin (1.0 mg/ml)Fludarabine (25.0 mg/ml)Mitomycin C (0.5 mg/ml)
Cyclophosphamide (20.0 mg/ml)Fluorouracil (50.0 mg/ml)Mitoxantrone HCl (2.0 mg/ml)
Cytarabine (100.0 mg/ml)Gemcitabine (38.0 mg/ml)Paclitaxel (6.0 mg/ml)
Dacarbazine (10.0 mg/ml)Idarubicin HCl (1.0 mg/ml)Rituximab (10.0 mg/ml)
Daunorubicin HCl (5.0 mg/ml)Ifosfamide (50.0 mg/ml)Trisenox (1.0 mg/ml)
Docetaxel (10.0 mg/ml)Irinotecan (20.0 mg/ml)Vincristine Sulfate (1.0 mg/ml)

The following chemotherapy drugs have low permeation times: Carmustine (3.3 mg/ml) : 33.8 minutes Thiotepa (10.0 mg/ml) : 128.1 minutes

Warning: Not for Use with: Carmustine, Thiotepa

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY K223559 AS REQUIRED BY: 21CFR§807.92

A. APPLICANT INFORMATION

510(K) Owner's NameNephron Pharmaceuticals Corporation
Address4500 12th Street Extension, West Columbia, SC29172.
Phone1-803-569-3110
Fax1-803-926-9853
E-maillkennedy@nephronpharm.com
Contact PersonLou Kennedy
DesignationChief Executive Officer
Contact Number1-803-569-3110
Contact Emaillkennedy@nephronpharm.com
Date Submitted21 October 2022

B. DEVICE IDENTIFICATION

Name of the deviceNephron Nitrile Powder-Free Nitrile ExaminationGloves (Tested For Use With ChemotherapyDrugs)
Product proprietary or trade nameNephron Nitrile
Common or usual nameNitrile Examination Gloves (Tested for use withChemotherapy drugs)
Classification nameNon-Powdered Patient Examination Glove Specialty
Device ClassificationClass-1, Reserved
Product CodeLZA, LZC, OPJ
Regulation Number21 CFR 880.6250
Review PanelGeneral Hospital and Personal Use Devices

C. PREDICATE DEVICE

Predicate DevicePowder Free Nitrile Examination Gloves (Blue,Purple-Blue), Tested for Use with ChemotherapyDrugs
510(k) NumberK213440
Regulatory ClassClass-1
Product codeLZA, LZC
Reference DeviceSHOWA® Blue Nitrile Powder Free MedicalExamination Glove
510(k) NumberK211003
Regulatory ClassClass-1
Product codeLZA, LZC

{4}------------------------------------------------

D. DESCRIPTION OF THE DEVICE:

Nephron Nitrile Powder-Free Nitrile Examination Gloves (Tested For Use With Chemotherapy Drugs) is a Class I, patient examination gloves bearing the product codes LZA, LZC, OPJ (21CFR880.6250). They meet all the current specifications listed under the ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application and also complies with requirements for Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs as per ASTM D6978-05 (2019). They are made from Nitrile (NBR). These gloves are blue in color and are powder free. The product is non-sterile, fingertip textured, ambidextrous with beaded cuff and single use only.

Nephron Nitrile Powder-Free Nitrile Examination Gloves (Tested For Use With Chemotherapy Drugs) with sizes Medium, Large, X-Large and XX-Large are included in the submission.

E. INDICATION FOR USE OF THE DEVICE:

Nephron Nitrile Powder-Free Nitrile Examination Gloves (Tested For Use With Chemotherapy Drugs) is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

The following chemotherapy drugs and concentration had NO breakthrough detected up to
240 minutes:
Bleomycin Sulfate (15.0 mg/ml)Doxorubicin HCl (2.0 mg/ml)Mechlorethamine HCl (1.0 mg/ml)
Busulfan (6.0 mg/ml)Epirubicin HCl (2.0 mg/ml)Melphalan (5.0 mg/ml)
Carboplatin (10.0 mg/ml)Etoposide (20.0 mg/ml)Methotrexate (25.0 mg/ml)
Cisplatin (1.0 mg/ml)Fludarabine (25.0 mg/ml)Mitomycin C (0.5 mg/ml)
Cyclophosphamide (20.0 mg/ml)Fluorouracil (50.0 mg/ml)Mitoxantrone HCl (2.0 mg/ml)
Cytarabine (100.0 mg/ml)Gemcitabine (38.0 mg/ml)Paclitaxel (6.0 mg/ml)
Dacarbazine (10.0 mg/ml)Idarubicin HCl (1.0 mg/ml)Rituximab (10.0 mg/ml)
Daunorubicin HCl (5.0 mg/ml)Ifosfamide (50.0 mg/ml)Trisenox (1.0 mg/ml)
Docetaxel (10.0 mg/ml)Irinotecan (20.0 mg/ml)Vincristine Sulfate (1.0 mg/ml)

The following chemotherapy drugs have low permeation times: Carmustine (3.3 mg/ml) : 33.8 minutes Thiotepa (10.0 mg/ml) : 128.1 minutes

Warning: Not for Use with: Carmustine, Thiotepa

{5}------------------------------------------------

F. SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE

CHARACTERISTICSSTANDARDSDEVICE PERFORMANCEComparison
PREDICATEREFERENCESUBJECT
510(K) Number---K213440K211003K223559
Name of device---Powder FreeNitrileExaminationGloves (Blue,Purple-Blue),Tested for Use withChemotherapyDrugsSHOWA® BlueNitrile Powder FreeMedical ExaminationGloveNephron NitrilePowder-Free NitrileExamination Gloves(Tested For Use WithChemotherapy Drugs)Similartopredicatedevice
Product Code---LZA, LZCLZA, LZCLZA, LZC, OPJSimilar
Indication for use---Powder FreeNitrileExaminationGloves (Blue,Purple-Blue),Tested for Use withChemotherapyDrugs is adisposable deviceintended formedical purposesthat is worn on theexaminer's hand topreventcontaminationbetween patientand examiner.A patient examinationglove is a disposabledevice intended formedical purposes thatis worn on theexaminer's hands orfingers to preventcontaminationbetween patient andexaminer.These gloves weretested for use withchemotherapy drugs inaccordance withASTM D6978-05Standard Practice forAssessment ofMedical Glove toPermeation bychemotherapy drugs.Nephron NitrilePowder-Free NitrileExamination Gloves(Tested For Use WithChemotherapy Drugs)is a disposable deviceintended for medicalpurpose that is wornon the examiner'shand to preventcontaminationbetween patient andexaminer. In addition,these gloves weretested for use withchemotherapy drugsin accordance withASTM D6978-05(2019) StandardPractice forAssessment ofResistance of MedicalGloves to Permeationby ChemotherapyDrugs.Similar
Regulation Number---21 CFR 880.625021 CFR 880.625021 CFR 880.6250Same
Material---NitrileNitrileNitrileSame
Color---Blue, Purple-BlueBlueBlueSame
Size---XS, S, M, L, XLXS, S, M, L, XLM, L, XL, XXLSimilar

{6}------------------------------------------------

CHARACTERISTICSSTANDARDSDEVICE PERFORMANCEComparison
PREDICATEREFERENCESUBJECT
510(K) Number---K213440K211003K223559
Single Use---Single-useSingle-useSingle-useSame
Sterile/non sterile---Non SterileNon SterileNon SterileSame
Rx Only or OTC---OTCOTCOTCSame
Dimensions - LengthASTM D6319-19Minimum 230mmOverall Length (mm) = 220 mm (sizes XS – S) and 230 mm (sizes M – XL)Minimum 230 mmSimilar
Size Average value M 235 L 237 XL 250 XXL 238
Dimensions - WidthASTM D6319-19XS: 70±10S: 80±10M: 95±10L: 110±10XL: 120±10XXL: 130±10Width (± 10 mm)Size XS = 70 mmSize S = 80 mmSize M = 95 mmSize L = 110 mmSize XL = 120 mmM: 95±10 mmL: 110±10 mmXL: 120±10 mmXXL: 130±10 mmSame
Size Average value M 95 L 113 XL 121 XXL 129
Physical Properties- Tensile StrengthASTM D6319-2019Before aging 14MPa, minBefore aging (MPa) = 14 minBefore aging 14MPa, minAverage value = 34.0 MPa (Medium)Same
After aging 14MPa, minAfter aging (MPa) = 14 minAfter aging 14MPa, minAverage value = 37.3 MPa (Medium)Same
Physical Properties- Ultimate ElongationASTM D6319-2019Before aging 500%, minBefore aging (%) = 500 minBefore aging 500%, minAverage value = 542% (Medium)Same
After aging 400%, minAfter aging (%) = 400 minAfter aging 400%, minAverage value = 503% (Medium)Same
ThicknessASTM D6319-19Palm: Minimum 0.05 mmFinger: Minimum 0.05 mmPalm (mm) = 0.05 minFinger Tip (mm) = 0.05 minPalm: Minimum 0.05 mmFinger: Minimum 0.05 mmSame
Size Palm (Avg value) Finger (Avg value) M 0.077 0.111 L 0.106 0.109 XL 0.089 0.115 XXL 0.113 0.107

{7}------------------------------------------------

CHARACTERISTICSSTANDARDSDEVICE PERFORMANCEComparison
PREDICATEREFERENCESUBJECT
510(K) NumberK213440K211003
Powder Free ResidueASTMD6319-19≤ 2 mg per glove≤ 2.0 mg/pc≤2 mg per gloveAverage value = 0.3516mg/glove (Medium)Same
Freedom from holesASTM D5151-2019In accordance withASTM D 5151-19,following ASTMD6319-19,G-I, AQL 2.5AQL 2.5Inspection LevelG-1In accordance withASTM D 5151-19,following ASTMD6319-19,G-I, AQL 2.5Same
Chemotherapy DrugsTested with MinimumBreakthroughDetection TimeASTM D6978-05 (2019)Bleomycin Sulfate15 mg/ml (15,000 ppm)>240 MinutesBlenoxane(15.0 mg/ml)>240 MinutesBleomycin Sulfate15 mg/ml (15,000 ppm)>240 MinutesSame
Busulfan6 mg/ml (6,000 ppm)>240 MinutesBusulfan(6.0 mg/ml)>240 MinutesBusulfan6 mg/ml (6,000 ppm)>240 MinutesSame
Carboplatin10 mg/ml (10,000 ppm)>240 MinutesCarboplatin(10.0 mg/ml)>240 MinutesCarboplatin10 mg/ml (10,000 ppm)>240 MinutesSame
Carmustine3.3 mg/ml (3,300 ppm)25.5 MinutesCarmustine(3.3 mg/ml)73.7 MinutesCarmustine3.3 mg/ml (3,300 ppm)33.8 MinutesSimilar
Chloroquine50mg/ml (50,000ppm)>240 Minutes---Not testedOptional*
Cisplatin1 mg/ml (1,000 ppm)>240 MinutesCisplatin(1.0 mg/ml)>240 MinutesCisplatin1 mg/ml (1,000 ppm)>240 MinutesSame
Cyclophosphamide20 mg/ml (20,000 ppm)>240 MinutesCyclophosphamide(20.0 mg/ml)>240 MinutesCyclophosphamide20 mg/ml (20,000 ppm)>240 MinutesSame
Cyclosporin100 mg/ml (100,000 ppm)>240 Minutes---Not testedOptional*
Cytarabine HCL100 mg/ml (100,000 ppm)>240 Minutes---Not testedOptional*
Cytarabine(100.0 mg/ml)>240 MinutesCytarabine100 mg/ml (100,000 ppm)>240 MinutesSame asreferencedevice
Dacarbazine10 mg/ml (10,000 ppm)>240 MinutesDacarbazine(10.0 mg/ml)>240 MinutesDacarbazine10 mg/ml (10,000 ppm)>240 MinutesSame
CHARACTERISTICSSTANDARDSDEVICE PERFORMANCEComparison
510(K) NumberPREDICATEREFERENCESUBJECT
Chemotherapy DrugsTested with MinimumBreakthrough DetectionTimeASTMD6978-05(2019)Daunorubicin HCL5 mg/ml (5,000 ppm)>240 MinutesDaunorubicin HCl(5.0 mg/ml)>240 MinutesDaunorubicin HCl5 mg/ml (5,000 ppm)>240 MinutesSame
Docetaxel HCL10 mg/ml (10,000 ppm)>240 Minutes---Not testedOptional*
Docetaxel(10.0 mg/ml)>240 MinutesDocetaxel10 mg/ml (10,000 ppm)>240 MinutesSame asreferencedevice
Doxorubicin HCl2 mg/ml (2,000 ppm)>240 MinutesDoxorubicin HCl(2.0 mg/ml)>240 MinutesDoxorubicin HCl2 mg/ml (2,000 ppm)>240 MinutesSame
Epirubicin HCl2 mg/ml (2,000 ppm)>240 MinutesEpirubicin HCl (2.0mg/ml)>240 MinutesEpirubicin HCl2 mg/ml (2,000 ppm)>240 MinutesSame
Etoposide20 mg/ml (20,000 ppm)>240 MinutesEtoposide(20.0 mg/ml)>240 MinutesEtoposide20 mg/ml (20,000 ppm)>240 MinutesSame
Fludarabine25 mg/ml (25,000 ppm)>240 MinutesFludarabine(25.0 mg/ml)>240 MinutesFludarabine25 mg/ml (25,000 ppm)>240 MinutesSame
Fluorouracil50 mg/ml (50,000 ppm)>240 MinutesFluorouracil(50.0 mg/ml)>240 MinutesFluorouracil50 mg/ml (50,000 ppm)>240 MinutesSame
Gemcitabine38 mg/ml (38,000 ppm)>240 MinutesGemcitabine(38.0 mg/ml)>240 MinutesGemcitabine38 mg/ml (38,000 ppm)>240 MinutesSame
Idarubicin HCl1 mg/ml (1,000 ppm)>240 MinutesIdarubicin HCl (1.0mg/ml)>240 MinutesIdarubicin HCl1 mg/ml (1,000 ppm)>240 MinutesSame
Ifosfamide50 mg/ml (50,000 ppm)>240 MinutesIfosfamide(50.0 mg/ml)>240 MinutesIfosfamide50 mg/ml (50,000 ppm)>240 MinutesSame
Irinotecan20 mg/ml (20,000 ppm)>240 MinutesIrinotecan(20.0 mg/ml)>240 MinutesIrinotecan20 mg/ml (20,000 ppm)>240 MinutesSame
Mechlorethamine HCl1 mg/ml (1,000 ppm)>240 MinutesMechlorethamineHCl (1.0 mg/ml)>240 MinutesMechlorethamine HCl1 mg/ml (1,000 ppm)>240 MinutesSame
Melphalan5 mg/ml (5,000 ppm)>240 MinutesMelphalan(5.0 mg/ml)>240 MinutesMelphalan5 mg/ml (5,000 ppm)>240 MinutesSame
CHARACTERISTICSSTANDARDSDEVICE PERFORMANCEComparison
PREDICATEREFERENCESUBJECT
510(K) NumberK213440K211003
Chemotherapy DrugsTested with MinimumBreakthroughDetection TimeASTMD6978-05(2019)Methotrexate25 mg/ml (25,000 ppm)>240 MinutesMethotrexate(25.0 mg/ml)>240 MinutesMethotrexate25 mg/ml (25,000 ppm)>240 MinutesSame
Mitomycin0.5 mg/ml (500 ppm)>240 MinutesMitomycin C(0.5 mg/ml)>240 MinutesMitomycin C0.5 mg/ml (500 ppm)>240 MinutesSame
Mitoxantrone HCL2 mg/ml (2,000 ppm)>240 MinutesMitoxantrone HCl2 mg/ml (2,000 ppm)>240 MinutesSame aspredicatedevice
Mitoxantrone(2.0 mg/ml)>240 MinutesNot testedOptional*
Oxaliplatin5mg/ml (5,000ppm)>240 MinutesNot testedOptional*
Paclitaxel6 mg/ml (6,000 ppm)>240 MinutesPaclitaxel(6.0 mg/ml)>240 MinutesPaclitaxel6 mg/ml (6,000 ppm)>240 MinutesSame
Paraplatin/Carboplatin,10mg/ml (10,000ppm)>240 MinutesNot testedOptional*
Retrovir10mg/ml (10,000ppm)>240 MinutesNot testedOptional*
Rituximab10mg/ml (10,000 ppm)>240 MinutesRituximab(10.0 mg/ml)>240 MinutesRituximab10 mg/ml (10,000 ppm)>240 MinutesSame
Thiotepa10mg/ml (10,000 ppm)66.8 MinutesThioTepa(10.0 mg/ml)25.4 MinutesThiotepa10 mg/ml (10,000 ppm)128.1 MinutesSimilar
Topotecan1mg/ml (1,000ppm)>240 MinutesNot testedOptional*
Trisenox1 mg/ml (1,000 ppm)>240 MinutesTrisenox(1.0 mg/ml)>240 MinutesTrisenox1 mg/ml (1,000 ppm)>240 MinutesSame
Velcade (Bortezomib)1mg/ml (1,000ppm)>240 MinutesNot testedOptional*
Vincristine Sulfate1 mg/ml (1,000 ppm)>240 MinutesVincristine Sulfate(1.0 mg/ml)>240 MinutesVincristine Sulfate1 mg/ml (1,000 ppm)>240 MinutesSame
CHARACTERISTICS STANDARDSDEVICE PERFORMANCEComparison
PREDICATEREFERENCESUBJECT
510(K) Number---K213440K211003K223559
BiocompatibilityPrimary SkinIrritation- ISO10993-23: FirstEdition 2021-01Under the conditions ofthe study, not an irritantUnder theconditions of thestudy, not aprimary skinirritant.Under the conditions ofthe study, the test articlemet the requirements ofthe testSame
DermalSensitization-ISO 10993-10:Fourth Edition2021-11Under the conditions ofthe study, not a sensitizerUnder conditionsof the study, not acontact sensitizerUnder the conditions ofthe study, the test articlewas not considered asensitizerSame
In vitrocytotoxicity-ISO 10993-5:Third Edition2009-06-01Under the conditions ofthis study, the test articleextract showed potentialtoxicity---Under the conditions ofthe study, the undilutedtest article extract and50% test article extractdilution did not meet therequirements of the testand the 25%, 12.5%,6.25%, and 3.13% testarticle extract dilutionsmet the requirements ofthe testSimilar topredicatedevice
AcuteSystemicToxicity- ISO10993-11:Third Edition2017-09Under the conditions ofthis study, there was noevidence of systemictoxicity.Under conditionsof the ISO AcuteSystemic Injectiontest, not toxicUnder the conditions ofthe study, there was nomortality or evidence ofsystemic toxicitySame

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

  • Predicate device / reference device perform additional Chemotherapy drug test.

Reference device has been included in the 510k to prove the safety of subject device with respect to the use of chemotherapy drug Cytarabine and Docetaxel which was not tested in case of predicate device.

There are no significant differences between the products and are identical in terms of intended use, materials, design and manufacturing methods. The devices meet the ASTM standard D6319-19 and D6978-05 (2019).

{11}------------------------------------------------

G. NON-CLINICAL TESTING SUMMARY PERFORMANCE DATA

BENCH TEST DATA

TEST METHODPURPOSEACCEPTANCE CRITERIARESULT
ASTM D6319-19Standard Specification forNitrile Examination Glovesfor Medical Application.To determine thelength of the glovesMedium : 230 mm minLarge : 230 mm minX-Large : 230 mm minXX-Large : 230 mm minMedium : 235 mmLarge : 237 mmX-Large : 250 mmXX-Large : 238 mm
ASTM D6319-19Standard Specification forNitrile Examination Glovesfor Medical Application.To determine thewidth of the glovesMedium : 95+/-10 mmLarge : 110+/-10 mmX-Large : 120+/-10 mmX-Large : 130+/-10 mmMedium : 95 mmLarge : 113 mmX-Large : 121 mmXX-Large : 129 mm
ASTM D6319-19Standard Specification forNitrile Examination Glovesfor Medical Application.To determine thethickness of theglovesPalm: 0.05 mm minfor all sizesFinger: 0.05 mm minfor all sizesSize Palm Finger Medium 0.077 mm 0.111 mm Large 0.106 mm 0.109 mm X-Large 0.089 mm 0.115 mm XX-Large 0.113 mm 0.107 mm
ASTM D6319-19Standard Specification forNitrile Examination Glovesfor Medical Application.To determine thephysical properties-Tensile strengthBefore AgeingTensile Strength14MPa min for all sizesAfter AgeingTensile Strength14MPa min for all sizesSize Beforeageing Afterageing Medium 34.0 MPa 37.3 MPa
To determine thephysical properties-Ultimate ElongationBefore AgeingUltimate Elongation500% min for all sizesAfter AgeingUltimate Elongation400% min for all sizesSize Beforeageing Afterageing Medium 542% 503%
ASTM D5151-19 StandardTest Method for Detectionof Holes in Medical GlovesTo determine theholes in the glovesAQL 2.5Gloves Passes AQL 2.5
ASTM D6124-06(Reapproved 2017) StandardTest Method for ResidualPowder on Medical GlovesTo determine theresidual powder in thegloves≤ 2 mg/gloveMedium: 0.3516 mg/glove

{12}------------------------------------------------

TEST METHODPURPOSEACCEPTANCE CRITERIARESULT
ASTM D6978-05(Reapproved 2019)Standard Practice forAssessment ofResistance ofMedical Gloves toPermeation byChemotherapyDrugs.To determine thebreakthroughdetection time ofchemotherapydrugsBleomycin Sulfate 15 mg/ml(15,000 ppm) >240 MinutesBleomycin Sulfate 15 mg/ml(15,000 ppm) >240 Minutes
Busulfan 6 mg/ml(6,000 ppm) >240 MinutesBusulfan 6 mg/ml(6,000 ppm) >240 Minutes
Carboplatin 10 mg/ml(10,000 ppm) >240 MinutesCarboplatin 10 mg/ml(10,000 ppm) >240 Minutes
Cisplatin 1 mg/ml(1,000 ppm) >240 MinutesCisplatin 1 mg/ml(1,000 ppm) >240 Minutes
Cyclophosphamide 20 mg/ml(20,000 ppm) >240 MinutesCyclophosphamide 20 mg/ml(20,000 ppm) >240 Minutes
Cytarabine 100 mg/ml(100,000 ppm) >240 MinutesCytarabine 100 mg/ml(100,000 ppm) >240 Minutes
Dacarbazine 10 mg/ml(10,000 ppm) >240 MinutesDacarbazine 10 mg/ml(10,000 ppm) >240 Minutes
Daunorubicin HCl 5 mg/ml(5,000 ppm) >240 MinutesDaunorubicin HCl 5 mg/ml(5,000 ppm) >240 Minutes
Docetaxel 10 mg/ml(10,000 ppm) >240 MinutesDocetaxel 10 mg/ml(10,000 ppm) >240 Minutes
Doxorubicin HCl 2 mg/ml(2,000 ppm) >240 MinutesDoxorubicin HCl 2 mg/ml(2,000 ppm) >240 Minutes
Epirubicin HCl 2 mg/ml(2,000 ppm) >240 MinutesEpirubicin HCl 2 mg/ml(2,000 ppm) >240 Minutes
Etoposide 20 mg/ml(20,000 ppm) >240 MinutesEtoposide 20 mg/ml(20,000 ppm) >240 Minutes
Fludarabine 25 mg/ml(25,000 ppm) >240 MinutesFludarabine 25 mg/ml(25,000 ppm) >240 Minutes
Fluorouracil 50 mg/ml(50,000 ppm) >240 MinutesFluorouracil 50 mg/ml(50,000 ppm) >240 Minutes
Gemcitabine 38 mg/ml(38,000 ppm) >240 MinutesGemcitabine 38 mg/ml(38,000 ppm) >240 Minutes
Idarubicin HCl 1 mg/ml(1,000 ppm) >240 MinutesIdarubicin HCl 1 mg/ml(1,000 ppm) >240 Minutes
Ifosfamide 50 mg/ml(50,000 ppm) >240 MinutesIfosfamide 50 mg/ml(50,000 ppm) >240 Minutes
Irinotecan 20 mg/ml(20,000 ppm) >240 MinutesIrinotecan 20 mg/ml(20,000 ppm) >240 Minutes
Mechlorethamine HCl 1 mg/ml(1,000 ppm) >240 MinutesMechlorethamine HCl 1 mg/ml(1,000 ppm) >240 Minutes
Melphalan 5 mg/ml(5,000 ppm) >240 MinutesMelphalan 5 mg/ml(5,000 ppm) >240 Minutes
Methotrexate 25 mg/ml(25,000 ppm) >240 MinutesMethotrexate 25 mg/ml(25,000 ppm) >240 Minutes
Mitomycin C 0.5 mg/ml(500 ppm) >240 MinutesMitomycin C 0.5 mg/ml(500 ppm) >240 Minutes
Mitoxantrone HCl 2 mg/ml(2,000 ppm) >240 MinutesMitoxantrone HCl 2 mg/ml(2,000 ppm) >240 Minutes
Paclitaxel 6 mg/ml(6,000 ppm) >240 MinutesPaclitaxel 6 mg/ml(6,000 ppm) >240 Minutes
Rituximab 10 mg/ml(10,000 ppm) >240 MinutesRituximab 10 mg/ml(10,000 ppm) >240 Minutes
Trisenox 1 mg/ml(1,000 ppm) >240 MinutesTrisenox 1 mg/ml(1,000 ppm) >240 Minutes
Vincristine Sulfate 1 mg/ml(1,000 ppm) >240 MinutesVincristine Sulfate 1 mg/ml(1,000 ppm) >240 Minutes

{13}------------------------------------------------

BIOCOMPATIBILITY DATA

TEST METHODPURPOSEACCEPTANCECRITERIARESULT
ISO 10993-23 First edition2021-01 BiologicalEvaluation of MedicalDevices - Part 23, Tests forIrritation.To evaluate the localdermal irritation of a testarticle extract followingintracutaneous injection inrabbits.Under the condition ofstudy not an irritantUnder the conditions of thestudy, the test article met therequirements of the test
10993-10 Fourth edition2021-11 BiologicalEvaluation of MedicalDevices - Part 10, Tests forSkin Sensitization.To evaluate the test item,for the skin sensitizationin Guinea pigs bymaximization test.Under the conditions ofthe study, not asensitizerUnder the conditions of thestudy, the test article was notconsidered a sensitizer
ISO 10993-5 Third edition2009-06-01 BiologicalEvaluation of MedicalDevices - Part 5, Tests forIn Vitro Cytotoxicity.To determine the potentialof a test article to causecytotoxicityUnder the conditions ofthe study, non-cytotoxicThe undiluted test articleextract and 50% test articleextract dilution did not meetthe requirements of the testand the 25%, 12.5%, 6.25%,and 3.13% test article extractdilutions met therequirements of the test. Thecytotoxic concern wasaddressed via acute systemictoxicity testing
ISO 10993-11 Third edition2017-09 BiologicalEvaluation of MedicalDevices - Part 11, Tests forSystemic ToxicityTo evaluate the acutesystemic toxicity of a testarticle extract followinginjection in mice.Under the conditions ofstudy, the deviceextracts do not pose asystemic toxicityconcernUnder the conditions ofstudy, there was no mortalityor evidence of systemictoxicity

The performance test data of the non-clinical tests that support a determination of safety and equivalence is the same as mentioned above (ASTM Requirements).

The performance test data of the non-clinical tests meet following standards:

ASTM D6319-19 Standard Specification for Nitrile examination Gloves for Medical Application.

ASTM D5151-19 Standard Test Method for Detection of Holes in Medical Gloves.

ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual Powder on Medical Gloves.

ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

{14}------------------------------------------------

ISO 10993-23 First Edition 2021-01 Biological Evaluation of Medical Devices - Part 23, Tests for Irritation.

ISO 10993-10 Fourth Edition 2021-11 Biological Evaluation of Medical Devices - Part 10, Tests for Skin Sensitization.

ISO 10993-5 Third Edition 2009-06-01 Biological Evaluation of Medical Devices - Part 5, Tests for In Vitro Cytotoxicity.

ISO 10993-11 Third Edition 2017-09 Biological Evaluation of Medical Devices - Part 11, Tests for Systemic Toxicity.

H. CLINICAL TESTING SUMMARY

Not applicable - Clinical data is not needed for gloves.

I. CONCLUSION

The conclusions drawn from the non-clinical test demonstrate that the subject device in 510(K) submission, Nephron Nitrile Powder-Free Nitrile Examination Gloves (Tested For Use With Chemotherapy Drugs) is as safe, as effective, and performs as well as or better than the legally marketed predicate device K213440.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.