(209 days)
The NC Thermometer (Gen3) is used orally for the intermittent measurement and monitoring of human body temperature. The device can be used by adults and children over the age of 5 years old.
The purpose of this traditional 510(k) premarket notification is to introduce NC° Thermometer (Gen3). NC° Thermometer (Gen3) is a battery powered, thermistor- based Bluetooth Low Energy (BLE) and Near Field Communication (NFC) enabled adjusted mode digital thermometer that uses a predictive algorithm for the measurement and monitoring of human body temperature. Body temperature is measured orally with the NC° Thermometer. Temperature measurements are displayed on the thermometer and can also (optionally) be transmitted to the Natural Cycles application on the user's smartphone by women over the age of 18. The compatible application displays the thermometer temperature value. The thermometer is reusable for home use on adults and children ages 5 and above. The thermometer is made of biocompatible metals and resins. The thermometer consists of a Negative Temperature Coefficient (NTC) thermistor located in the probe tip to sense human body temperature, three input user facing buttons, an output buzzer for audio indications, a screen for user display, and back lights for illuminating the buttons.
The context provided is for a clinical electronic thermometer, not an AI-powered medical device. Therefore, the information typically requested in questions related to acceptance criteria and study designs for AI/ML-based medical devices (such as sample sizes for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, and training set details) is not directly applicable or available in this document.
The document primarily focuses on demonstrating substantial equivalence of the NC° Thermometer (Gen3) to a predicate device (Kinsa QuickCare Thermometer) based on technical characteristics, non-clinical testing, and a clinical validation study for its accuracy as a thermometer.
However, I can extract the relevant information regarding the device's performance and the study that proves it meets acceptance criteria as per the provided text for a clinical electronic thermometer.
Device: NC° Thermometer (Gen3)
Device Type: Clinical electronic thermometer
1. Table of Acceptance Criteria and Reported Device Performance
| Criteria Category | Specific Criterion / Requirement (Acceptance) | Reported Device Performance |
|---|---|---|
| Accuracy (Clinical) | Meets ISO 80601-2-56:2017 requirements for clinical accuracy of an adjusted thermometer. | The NC° Thermometer (Gen3) was found to meet the ISO 80601-2-56 requirements of clinical accuracy for an adjusted thermometer, as documented in the pivotal clinical study. |
| Accuracy (Technical) | ±0.2°C/0.3°F (Temperature < 37°C/98.0°F)±0.1°C/0.2°F (Temperature 37.0°C to 39.0°C/98.0°F to 102.0°F)±0.2°C/0.3°F (Temperature > 39.0°C/102.0°F) Meets ASTM E1112-00 (2018) and ISO 80601-2-56:2017. | Performance testing conducted on the subject device according to ASTM E1112 and ISO 80601-2-56 demonstrated that it meets these accuracy requirements. The device has higher accuracy than the predicate device from 37 to 39°C because ASTM E1112 has stricter accuracy requirements. |
| Response Time | Specified (nominal 40 seconds) and validated by testing per ISO 80601-2-56 and ASTM E1112. User expectations considered. | Nominally 40 seconds. Usability testing showed the measuring time is well within user expectations and far below the 3 minutes that would be required without a predictive algorithm. This specification has been validated by the testing per ISO 80601-2-56 and ASTM E1112. |
| Resolution of Display | Specified (0.01 °C / 0.01°F) and meets ISO 80601-2-56 and ASTM E1112 requirements. | Performance testing demonstrated that the subject device is in compliance with both ISO 80601-2-56 and ASTM E1112 requirements for display resolution. |
| Biocompatibility | Meets ISO 10993-1:2018, ISO 10993-5:2009, ISO 10993-10:2010. | All skin contacting materials successfully tested for biocompatibility (cytotoxicity, irritation, sensitization) in accordance with AAMI/ANSI/ISO 10993-5:2009/(R)2014 and AAMI/ANSI/ISO 10993-10:2010/(R)2014. |
| Electrical Safety | Meets AAMI/ANSI ES 60601-1:2005+A2:2020 and IEC 60601-1-11:2015. | Confirmed through compliance testing to AAMI/ANSI ES 60601-1:2005/(R)2012 + A1:2012, C1:2009/(R)2012 + A2:2010/(R)2012, and IEC 60601-1-11:2015 (Edition 2.1 2020-07). |
| ElectromagneticCompatibility (EMC) | Meets IEC 60601-1-2:2014 and FCC Part 15 Subpart C (§15.247 / RSS 247 and §15.225). | Confirmed through compliance testing to IEC 60601-1-2:2014, FCC Part 15 Subpart C § 15.247, and FCC Part 15 Subpart C § 15.225. |
| Software Performance | Firmware and software units meet software requirements specifications and system performs as intended. | Software Verification and Validation results confirmed the firmware and software units meet the software requirements specifications and the system performs as intended. Software documentation provided per FDA guidance. |
| Cybersecurity | Hazards and risks evaluated and managed. | Cybersecurity hazards and risks associated with the thermometer have been evaluated. Information provided per FDA Guidance "Content of Premarket Submission for Management of Cybersecurity in Medical Devices". |
| Wireless Interoperability | Able to communicate with the Natural Cycles application as intended. | Interoperability testing completed to ensure that the thermometer is able to communicate with the Natural Cycles application as intended. Performance data includes software verification, cybersecurity analysis, wireless coexistence testing, and usability testing. |
| Operating Environment | 15 to 40°C, < 95% relative humidity non-condensing. Meets IEC 60601-1-11 and ISO80601-2-56 standards. | Performance testing conducted according to IEC 60601-1-11 and ISO80601-2-56 demonstrates compliance within the specified operating environment. |
| Storage Environment | -25 to 50 °C, ≤ 95% relative humidity, 70kPa to 106kPa. Meets IEC 60601-1-11 and ISO80601-2-56 standards. | Performance testing conducted according to IEC 60601-1-11 and ISO80601-2-56 demonstrates compliance within the specified storage environment. |
Regarding the study that proves the device meets the acceptance criteria (Clinical Validation Study):
The document provides details of a clinical validation study specifically for the clinical accuracy of the thermometer.
-
Sample size used for the test set and the data provenance:
- Sample Size: 105 participants.
- Data Provenance: Not explicitly stated, but it's a clinical validation study performed to support the performance of the device, implying prospective data collection for this specific purpose. No mention of country of origin.
- Participant Demographics:
- 32 (30%) were febrile.
- All participants were over the age of 5 years old.
- 64 participants (61%) were female and 41 participants (39%) were male.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This question is not applicable in the context of a clinical electronic thermometer where the "ground truth" for temperature measurement is typically established by physical reference thermometers or established clinical methods, not by expert interpretation as in imaging AI.
- The study evaluated the clinical performance of the NC° Thermometer (Gen3) "to a reference clinical thermometer." This reference thermometer established the "ground truth" for temperature.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable. Adjudication methods are typically for subjective expert interpretations (e.g., in radiology image reading), not for objective measurements by a thermometer compared to a reference thermometer.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is a thermometer, not an AI-assisted diagnostic tool that would involve human readers or MRMC studies.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- The "standalone performance" of the device in this context refers to its intrinsic measurement accuracy and predictive algorithm performance. The non-clinical testing (Design Verification, Software Verification, etc., against standards like ISO 80601-2-56 and ASTM E1112) essentially assesses this standalone performance. The clinical study then validated this performance in a real-world setting.
- The device does use a "predictive algorithm for the measurement and monitoring of human body temperature," which is an algorithmic component. Its performance is evaluated through the described testing against the standards.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The ground truth for temperature measurement in the clinical study was established by a reference clinical thermometer compliant with relevant standards (implied, as the study itself was in accordance with ISO 80601-2-56).
-
The sample size for the training set:
- Not explicitly stated in this document. The document refers to a "predictive algorithm," but details on the dataset used to train this algorithm are not provided within this text. This document focuses on the validation of the final device.
-
How the ground truth for the training set was established:
- Not explicitly stated. As above, details regarding the training of the predictive algorithm are not within the scope of this FDA 510(k) summary, which focuses on demonstrating substantial equivalence and validation.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. To the left of the FDA logo is the Department of Health & Human Services logo.
June 20, 2023
Xiamen Intretech Inc. % Megan Callanan Regulatory and Compliance Manager Natural Cycles 510 5th Avenue 3rd Floor New York, New York 10036
Re: K223522
Trade/Device Name: NCº Thermometer (Gen 3), Model Number: NCTG3 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: May 19, 2023 Received: May 19, 2023
Dear Megan Callanan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
David Walloschek
David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K223522
Device Name NCº Thermometer (Gen3), Model Number: NCTG3
Indications for Use (Describe)
The NC Thermometer (Gen3) is used orally for the intermittent measurement and monitoring of human body temperature. The device can be used by adults and children over the age of 5 years old.
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K223522
510(k) Summary: NC° Thermometer (Gen3)
| Applicant: | Xiamen Intretech Inc.No.100 Dongfu West RoadHaicang District, XiamenFujian, China Zip:361027 |
|---|---|
| Applicant Contact: | Name: Sunny SunPhone: 86-15750706715Email: xmshj@intretech.com |
| Correspondent: | Name: Megan CallananPhone: (216)744-4524Email: Megan.callanan@naturalcycles.com |
| Date Prepared: | June 17, 2023 |
| Trade Name: | NC° Thermometer (Gen3),Model Number: NCTG3 |
| Common Name: | Clinical electronic thermometer |
| Proposed Class: | Class II |
| Classification Name: | Clinical electronic thermometer |
| Regulation Number: | 21 CFR 880.2910 |
| Product Code: | FLL |
| Predicate Device: | K173730 KSA-110 Clinical electronic thermometer from KINSA,INC ,marketed as the Kinsa QuickCare Thermometer. |
| Device Description: | The purpose of this traditional 510(k) premarket notification is to introduceNC° Thermometer (Gen3). NC° Thermometer (Gen3) is a battery powered,thermistor- based Bluetooth Low Energy (BLE) and Near FieldCommunication (NFC) enabled adjusted mode digital thermometer thatuses a predictive algorithm for the measurement and monitoring of humanbody temperature. Body temperature is measured orally with the NC°Thermometer. Temperature measurements are displayed on thethermometer and can also (optionally) be transmitted to the Natural Cyclesapplication on the user's smartphone by women over the age of 18. Thecompatible application displays the thermometer temperature value. The |
| thermometer is reusable for home use on adults and children ages 5 andabove. The thermometer is made of biocompatible metals and resins. Thethermometer consists of a Negative Temperature Coefficient (NTC)thermistor located in the probe tip to sense human body temperature,three input user facing buttons, an output buzzer for audio indications, ascreen for user display, and back lights for illuminating the buttons. | |
| Indications forUse: | The NC° Thermometer (Gen3) is used orally for the intermittentmeasurement and monitoring of human body temperature. The device canbe used by adults and children over the age of 5 years old. |
{4}------------------------------------------------
Summary of Technical Characteristics:
The technology of the NC° Thermometer (Gen3) is identical to the predicate device in terms of its principle of operation, thermistor, use of a prediction algorithm, and probe tip design. The industrial design is similar to the predicate except for the outer shell form.
| Element ofcomparison | Subject DeviceNC° Thermometer (Gen3)K223522 | Predicate DeviceKinsa QuickCareThermometerK173730 | Verdict |
|---|---|---|---|
| Thermometertype | Predictive digital | Predictive digital | Same |
| Indications forUse | The NC° Thermometer(Gen3) is used orally for theintermittent measurementand monitoring of humanbody temperature. Thedevice can be used byadults and children over theage of 5 years old. | The Kinsa QuickCareThermometer is used for theintermittent measurementand monitoring of humanbody temperature, orally,rectally and under the arm.The device is for the adultand pediatric population. | Similar1 |
| Fundamentaltechnology &Operatingprinciple | Utilizes an NTC thermistorlocated in the probe tip tosense human bodytemperature when placed inthe desired measuring site(mouth). The resultingchange in resistance issensed and monitored inorder to through use of | Utilizes an NTC thermistorlocated in the probe tip tosense human bodytemperature when placed inthe desired measuring site(mouth, underarm, andrectum). The resulting changein resistance is sensed andmonitored in order to through | Similar2 |
| predictive algorithmsestimate human bodytemperature. | use of predictive algorithmsestimate human bodytemperature. | ||
| Sensor | Thermistor based | Thermistor based | Same |
| Signalprocessingand display | Internal firmware and localscreen display- Also able to transfertransmit data to an auxiliarydevice for secondary display | Internal firmware and localLCD display- Also able to transfertransmit data to an auxiliarydevice for secondary display | Same |
| WirelessInterface | Bluetooth Low Energy (BLE)and Near FieldCommunication (NFC) | Bluetooth Low Energy (BLE) | Different1 |
| CompatibleApplicationName | Natural Cycles | Kinsa | Different2 |
| CompatibleApplicationDisplay ofThermometerRelated Data | Temperature, timestamp | Temperature, timestamp, tipsfrom our in-house clinician,share your illness timelinewith another caregiver oryour doctor | Different3 |
| Powerrequirements | Battery powered CR2032(3V) | Battery powered (CR2032(3V) | Same |
| PatientContactingMaterials | SUS316L, ABS, PMMA,ABS+SMMA | SU304 Stainless Steel, ABS,PMMA | Similar3 |
| Scale | °F / °C | °F / °C | Same |
| Measurementlocations | Oral | Oral, Axillary and Rectal | Different4 |
| MeasurementRange | 32 to 42 °C (89.6 to 107.6°F) | 32 to 42.8 °C (89.6 to 109.2°F) | Different5 |
| Number ofmeasurementsthat can besaved andviewed onthermometer | 10 | N/A- No history button onthermometer | Different6 |
| ThermometerButtons | Power button, settingbutton, history button | Power button | Different7 |
| OperatingEnvironment | 15 to 40°C< 95% relative humidity non- | 15 to 35 °C15 to 85% Relative humidity | Different8 |
| StorageEnvironment | -25 to 50 °C≤ 95% relative humidity70kPa to 106kPa | -25 to 70 °CUp to 90% relative humidity,non-condensing, up to 50 hPa | Different9 |
| Accuracy | ±0.2°C/0.3°F (Temperature lessthan 37°C/98.0°F)±0.1°C/0.2°F (Temperature37.0°C to 39.0°C/98.0°F to102.0°F)±0.2°C/0.3°F (Temperaturegreater than 39.0°C/102.0°F) | ± 0.2°C within measurementrange of 32 to 42.8°C (89.6to 109.2°F) | Different10 |
| ResponseTime | 40 seconds nominally | 8 seconds nominally | Different11 |
| Resolution ofDisplay | 0.01 °C / 0.01°F | 0.1 °C / 0.1°F | Different12 |
| Performance | Meets ISO 80601-2-56:2017 | Meets ISO 80601-2-56 | Same |
| Biocompatibility | Meets ISO 10993-1:2018,ISO 10993-5:2009, ISO10993-10:2010 | Meets ISO 10993-1:2009,ISO 10993-5:2009, ISO10993-10:2010 | Same |
| ElectricalSafety | Meets AAMI/ANSI ES60601-1:2005+A2:2020 | Meets AAMI/ANSI ES 60601-1:2005+A1:2012 | Same |
| ElectromagneticDisturbances | Meets IEC 60601-1-2:2014 | Meets IEC 60601-1-2:2014 | Same |
{5}------------------------------------------------
Traditional 510(k)
{6}------------------------------------------------
Traditional 510(k)
Discussion of Similar Elements of Comparison:
-
- Similar indications for use: The subject device can only be used for adults and children over the age of 5. The predicate device indications for use do not specify a pediatric age limit. As the subject device is intended for oral use only, the intended user was limited to children over the age of 5 years old and adults. Performance testing demonstrates that the subject device meets ISO 80601-2-56 and the different pediatric age limit does not raise any new safety and effectiveness questions.
-
- Similar fundamental technology and operating principle: Similar hardware is used in both devices. Performance testing demonstrates that the subject device meets ISO 80601-2-56 and ASTM E1112. The differences in fundamental technology and operating principle do not raise any new safety and effectiveness questions.
-
- Similar patient contacting materials: The thermometer materials are similar except for the probe tip and buttons. The subject device probe tip uses an equivalent biocompatible stainless steel and the subject device buttons include an additional thermoplastic, SMMA. All skin contacting materials in the subject device have been tested successfully for biocompatibility: cytotoxicity in accordance with AAMI/ANSI/ISO 10993-5:2009/(R)2014, Biological Evaluation of
{7}------------------------------------------------
Medical Devices Part 5: Tests for In Vitro Cytotoxicity, as well as irritation and sensitization in accordance with AAMI/ANSI/ISO 10993-10:2010/(R)2014. Biological Evaluation of Medical Devices Part 10: Tests for Irritation and Skin Sensitization. The materials used in the subject device do not raise any new questions of safety and effectiveness.
Discussion of Different Elements of Comparison:
-
- Different wireless interface: The subject device has NFC interface in addition to BLE. while the predicate device has BLE. NFC is a widely used technology and the subject device has passed electrical safety and electromagnetic compatibility testing in accordance with the FDA recognized consensus standards: AAMI/ANSI ES 60601-1:2005/(R)2012 + A1:2012, C1:2009/(R)2012 + A2:2010/(R)2012, IEC 60601-1-2:2014, IEC60601-1-11 Edition 2.1 2020-07, FCC Part 15 Subpart C & 15.247, and FCC Part 15 Subpart C & 15.225. Additionally, performance data consisting of software verification, cybersecurity analysis, wireless coexistence testing, usability testing demonstrate the safe and effective use of this feature. The essential performance of the subject device is not compromised as the primary display of the temperature readings on the thermometer's local screen is independent of the wireless transmission. The addition of the NFC wireless interface does not raise different questions of safety and effectiveness.
-
- Different compatible application name: Subject and predicate thermometers are compatible with different applications. Both applications are downloaded onto smartphones and allow users to see temperature and timestamp. The essential performance of the subject device is not compromised as the primary display of the temperature readings on the thermometer's local screen is independent of the wireless transmission. The subject device has passed electrical safety and electromaqnetic compatibility testing in accordance with the FDA recognized consensus standards: AAMI/ANSI ES 60601-1:2005/(R)2012 + A1:2012, C1:2009/(R)2012 + A2:2010/(R)2012, IEC 60601-1-2:2014, IEC60601-1-11 Edition 2.1 2020-07, FCC Part 15 Subpart C § 15.247, and FCC Part 15 Subpart C § 15.225. Additionally, performance data consisting of software verification, cybersecurity analysis, wireless coexistence testing, usability testing demonstrate the safe and effective use of the subject device and compatible application.
-
- Different compatible application display of temperature related data: The predicate thermometer's compatible app provides more information regarding illness but both apps function equivalently when receiving temperature and timestamp data and displaying it on the phone. The predicate thermometer's compatible application has additional features related to illness such as tips from clinician and illness timeline but these features do not impact the safe and effective use of the thermometer. The subject thermometer compatible app is a medical device while the predicate is not but this does not impact the app with respect to the thermometer. The subject device has passed electrical safety and electromagnetic compatibility testing in accordance with the FDA recognized
{8}------------------------------------------------
consensus standards: AAMI/ANSI ES 60601-1:2005/(R)2012 + A1:2012, C1:2009/(R)2012 + A2:2010/(R)2012, IEC 60601-1-2:2014, IEC60601-1-11 Edition 2.1 2020-07, FCC Part 15 Subpart C § 15.247, and FCC Part 15 Subpart C & 15.225. Additionally, performance data consisting of software verification, cybersecurity analysis, wireless coexistence testing, usability testing demonstrate the safe and effective use of the subject device and compatible application.
-
- Different measurement locations: The subject device is used for oral measurement only. The route of temperature measurement for the subject device is a subset of the routes available using the predicate device. Both the subject and predicate devices are used for oral measurements. The subject device meets the requirements of ISO 80601-2-56 and ASTM E1112 for oral use and the limitation on measuring routes does not raise any different questions of safety and effectiveness.
-
- Different measurement range: The subject device has a lower upper limit of measurement range than the predicate device. The ISO 80601-2-56 performance testing of the subject device demonstrated that the difference does not raise any new questions of safety and effectiveness.
-
- Different number of measurements that can be saved and viewed on thermometer: There is no history button on the predicate device and thus no measurements can be saved and viewed on the thermometer. The subject device has a history button, but this does not impact the essential performance of the subject device. Software verification, design verification, and usability testing demonstrates that the history button and display of previous measurements on the thermometer does not raise any new questions of safety and effectiveness.
-
- Different thermometer buttons: Compared to the predicate device, there are two more buttons (history button and settings button) on the subject device. The essential performance of the subject device is not compromised as temperature measurement and display of temperature readings is independent of History and Settings buttons. Performance data consisting of software verification, design verification, and usability testing demonstrates the safe and effective use of all buttons. It does not raise new questions of safety and effectiveness.
-
- Different operating environment: The subject device has larger operating range and higher humidity requirements than the predicate device. The operating environment requirements on the subject device and the predicate device both meet the IEC 60601-1-11 and ISO80601-2-56 standards and performance testing conducted according to these standards demonstrates that the difference in operating environment does not raise any new questions of safety and effectiveness.
-
- Different storage environment: The subject device has a smaller storage range and higher humidity limitation. The storage environment requirements on the subject device and the predicate device both meet the IEC 60601-1-11 and ISO80601-2-56 standards and performance testing conducted according to these standards demonstrates that the difference in storage environment does not raise any new questions of safety and effectiveness.
{9}------------------------------------------------
-
- Different accuracy: The subject device meets ISO80601-2-56 and ASTM E1112 requirements, while the predicate device meets ISO80601-2-56. The subject device has higher accuracy than the predicate device from 37 to 39°C because ASTM E1112 has stricter accuracy requirements. Performance testing was conducted on the subject device according to ASTM E1112 and ISO80601-2-56. The difference does not raise any new performance or safety concerns.
-
- Different response time: The subject device has longer measurement time than the predicate device, but usability testing showed the measuring time is well within user expectations and far below the 3 minutes that would be required without a predictive algorithm. This specification has been validated by the testing per ISO 80601-2-56 and ASTM E1112. The difference does not raise any new questions of safety and effectiveness.
-
- Different resolution of display: The subject device has a higher resolution of display than the predicate device. Performance testing demonstrated that the subject device is in compliance with both ISO 80601-2-56 and ASTM E1112 requirements. The difference in resolution of display does not raise any new questions of safety and effectiveness.
Testing to Demonstrate Substantial Equivalence:
Non-clinical Testing
Non-clinical performance reports were provided to document verification and validation activities intended to demonstrate substantial equivalence of the subject device to the predicate device:
-
- Design Verification results confirmed the device meets the product requirements set by and the performance standard requirements of ISO 80601-2-56:2017, Medical Electrical Equipment – Part 2-56: Particular Requirements For Basic Safety And Essential Performance Of Clinical Thermometers For Body Temperature Measurement and ASTM E 1112-00 (2018) Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature.
-
- Software Verification and Validation results confirmed the firmware and software units meet the software requirements specifications and the system performs as intended. Software documentation is provided in accordance with the FDA "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (May 11, 2005).
-
- Cybersecurity hazards and risks associated with the thermometer have been evaluated. Cybersecurity information is provided in accordance with the FDA Guidance "Content of Premarket Submission for Management of Cybersecurity in Medical Devices"
-
- All skin contacting materials have been tested successfully for biocompatibility: cytotoxicity in accordance with AAMI/ANSI/ISO 10993-5:2009 /(R)2014, Biological Evaluation of Medical Devices Part 5: Tests for In Vitro Cytotoxicity, as well as irritation and sensitization in accordance with AAMI/ANSI/ISO 10993-
{10}------------------------------------------------
10:2010/(R)2014, Biological Evaluation of Medical Devices Part 10: Tests for Irritation and Skin Sensitization.
-
- Electrical and Mechanical Safety as well as essential performance was confirmed through compliance testing to AAMI/ANSI ES 60601-1:2005/(R)2012, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance and IEC 60601-1-11:2015, Medical Electrical Equipment Part 1-11: Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment.
-
- Electromagnetic Compatibility was confirmed through compliance testing to IEC 60601-1-2:2014, Medical Electrical Equipment - Part 1-2: Electromagnetic Disturbances – Requirements and Tests.
-
- FCC compliance was confirmed through testing to FCC Part 15 Subpart C §15.247/ RSS 247.
-
- Interoperability testing was completed to ensure that the thermometer is able to communicate with the Natural Cycles application as intended.
Clinical Testing
To support the performance of the NC° Thermometer (Gen3), a clinical validation study was performed. The study evaluated the clinical performance of the NC° Thermometer (Gen3) to a reference clinical thermometer in accordance with ISO 80601-2-56. Clinical validation study reports were provided with this submission. The pivotal clinical study included 105 participants, 32 (30%) were febrile. All participants were over the age of 5 years old. 64 participants (61%) were female and 41 participants (39%) were male. Based on the clinical performance as documented in the pivotal clinical study, the NC° Thermometers (Gen3) was found to meet the ISO 80601-2-56 requirements of clinical accuracy for an adjusted thermometer.
Conclusion:
Based on the design features, the use of established well known materials, feature comparisons, and indications for use, the NC° Thermometer (Gen3) has demonstrated substantial equivalence to the identified predicate device.
§ 880.2910 Clinical electronic thermometer.
(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.