K Number
K223399

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-05-20

(558 days)

Product Code
Regulation Number
870.2700
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AVITA Pulse Oximeter is intended for measuring functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adults as non-invasive spot checking in professional caring environment. It is designed for fingers between 0.8cm and 2.5cm (0.3 inches to 1 inches) and for patients during no-motion condition. The device is prescription only.

Device Description

The AViTA Pulse Oximeter SPO2 is non-invasive spot checking, not provided sterile, multi-use device, which can measure and display user's oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) through finger during no-motion condition. The subject device is not for life-sustaining, not for implant and does not contain drug or biological products. The device is for prescription use only. The subject device consists of sensor, electronic circuits, display and plastic enclosures. It is a battery powered device and is adopted with two color OLED screen to display SpO2 and PR. It is software-driven and does not include alarms.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the AViTA Pulse Oximeter meets them, based on the provided text:

Acceptance Criteria and Device Performance

CriteriaReported Device Performance
SpO2 Accuracy (70% to 100% range)± 2%
Pulse Rate Accuracy±2 bpm or ±2%, whichever is greater
Clinical Performance (SpO2 under no motion)ARMS (Accuracy Root Mean Square) for SpO2 = 1.89%
Electrical SafetyComplied with IEC 60601-1
Electromagnetic Compatibility (EMC)Complied with IEC 60601-1-2
Biocompatibility - CytotoxicityComplied with ISO 10993-5 (in accordance with ISO 10993-1)
Biocompatibility - SensitizationComplied with ISO 10993-10 (in accordance with ISO 10993-1)
Biocompatibility - IrritationComplied with ISO 10993-23 (in accordance with ISO 10993-1)
Software Verification and ValidationComplied with FDA Guidance, "Moderate" level of concern
Cleaning ValidationDevice performance not affected after multiple cleanings
Usage Life> 18 hours typical operation
Water ResistanceIP22 Approved

Note: The SpO2 accuracy of ±2% for the 70-100% range and the Pulse Rate accuracy are specifications, while the ARMS of 1.89% is the result of the clinical performance study. The 1.89% ARMS falls within the generally accepted range for pulse oximeters, which is often considered to be within ±2-3% for this range of SpO2.


Study Details

The primary study proving the device meets its acceptance criteria is a clinical performance study and various performance tests conducted in accordance with international standards.

2. Sample size used for the test set and the data provenance:

  • Clinical Performance Study: The study was conducted on healthy, adult volunteers. The specific number of volunteers (sample size) is not explicitly stated in the provided text.
  • Data Provenance: The origin of the data is not specified beyond "healthy, adult volunteers." It is a prospective study as indicated by "clinical validation testing... under no motion."

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

The ground truth for pulse oximeters in clinical validation studies does not typically involve human expert consensus on the SpO2 values. Instead, it relies on a co-oximeter (or similar arterial blood gas analysis device), which provides a highly accurate, invasive measurement of arterial oxygen saturation (SaO2). The text states "oxygenated hemoglobin characterization testing" and "clinical validation testing of the SpO2 performance," which implies comparison against a reference method, likely a co-oximeter. Therefore, there is no mention of experts establishing ground truth in this context, nor their specific qualifications.

4. Adjudication method for the test set:

Not applicable for this type of medical device performance testing. Pulse oximeter accuracy is evaluated by comparing the device's readings against a precise, objective reference standard (like a co-oximeter), not by expert adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is typically relevant for interpretative diagnostic devices (e.g., imaging AI) where human readers interact with the output. The AViTA Pulse Oximeter is a direct measurement device; it does not involve human interpretation of complex data that would be enhanced by AI assistance.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

Yes, a standalone performance evaluation was done. The results presented ("ARMS for SpO2 under no motion was found to be 1.89%") directly reflect the performance of the device's algorithms and hardware in measuring SpO2, without any human interpretation component for the measured value itself.

7. The type of ground truth used:

The type of ground truth used for the clinical performance study, though not explicitly named, would be a highly accurate, invasive arterial oxygen saturation measurement (SaO2) obtained via a co-oximeter. This is the standard reference method for validating non-invasive pulse oximeters. The text refers to "oxygenated hemoglobin characterization testing" which supports this assumption.

8. The sample size for the training set:

Not applicable. The AViTA Pulse Oximeter is a hardware-based device that employs principles of spectrophotometry to mechanically and optically measure SpO2 and PR. While it is "software-driven," it is not described as using machine learning or AI algorithms that would require a distinct "training set" in the conventional sense (e.g., for image classification or prediction). Its algorithms are based on established physiological models of light absorption by blood.

9. How the ground truth for the training set was established:

Not applicable. As noted above, there is no indication of a "training set" for AI/ML models in this context. The core operating principles rely on physics and established physiological relationships.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and name on the right. The FDA acronym and name are in blue, with the acronym in a larger font size than the name. The name is written as "U.S. FOOD & DRUG ADMINISTRATION".

May 20, 2024

AViTA Corporation Maggie Chao 9F, No.78, Sec.1, Kwang-Fu Rd. New Taipei City, San-Chung District 24158 Taiwan

Re: K223399

Trade/Device Name: AViTA Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: February 1, 2024 Received: February 1, 2024

Dear Maggie Chao:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.

{2}------------------------------------------------

See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Bradley Q. Quinn -S

Bradley Quinn Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K223399

Device Name

AVITA Pulse Oximeter (SP61)

Indications for Use (Describe)

The AVITA Pulse Oximeter is intended for measuring functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adults as non-invasive spot checking in professional caring environment. It is designed for fingers between 0.8cm and 2.5cm (0.3 inches to 1 inches) and for patients during no-motion condition. The device is prescription only.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

This summary of 510(k) information is being submitted in accordance with the requirement of 21 CFR 807.92.

Submitter's Name: AViTA Corporation 1 Address: 9F, No.78, Sec.1, Kwang-Fu Rd., San-Chung District, New Taipei City 24158, Taiwan Telephone number: +886-2-8512-1568 ext.2010 Fax Number: +886-2-8512-1347 Applicant Contact: Meggie Chao The Date of the Summary: May 2, 2024

  • 2 Trade/Proprietary Name: AViTA Pulse Oximeter Common Name: Pulse Oximeter Model numbers: SP61 Review Panel: Anesthesiology Classification Product Code: DQA Classification name: Oximeter Regulation Number: 870.2700 Device Class: II

PREDICATE DEVICE 3

510(k) Number:K193350
Trade Name:Leadtek Fingertip Pulse Oximeter
Manufacturer:Leadtek Research Incorporation

4 DEVICE DESCRIPTION

The AViTA Pulse Oximeter SPO2 is non-invasive spot checking, not provided sterile, multi-use device, which can measure and display user's oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) through finger during no-motion condition. The subject device is not for life-sustaining, not for implant and does not contain drug or biological products. The device is for prescription use only.

The subject device consists of sensor, electronic circuits, display and plastic enclosures. It is a battery powered

{5}------------------------------------------------

device and is adopted with two color OLED screen to display SpO2 and PR. It is software-driven and does not include alarms. Table 1 shows the specifications for the subject device.

A ViT A pulse oximeters work by the principles of spectrophotometry, emitting two different wavelengths of light, typically red and infrared, through a pulsating capillary bed, such as a fingertip. The sensor on the other side of the tissue detects the light that emerges from the tissues. The device the intensity of red and infrared light that is transmitted through the capillary bed.

Based on the differences in absorption between oxygenated and deoxygenated blood at specific wavelengths, the device can calculate the ratio of oxygenated hemoglobin (HbO) to total hemoglobin in the blood, which is known as oxygen saturation (SpO2).

It's important to keep the finger or the measurement site stationary during the reading to avoid introducing motion artifacts that could affect the accuracy of the measurement. Additionally, it is recommended to use the pulse oximeter before or after engaging in sports activities rather than during physical exercise. Do not use for continuous monitoring. Table 1 shows the specifications for the subject device.

PropertySpecification
DimensionL68mm (2.68") x W37.8mm (1.49") x H28mm (1.1")
Weight(without battery)26g (0,92 ounces)
DisplayTwo color OLED
Auto on/offWhenever user inserts a finger, the device will turn onautomatically. Vice versa, the device will turn off automaticallywhen the finger is removed from it.
Input key© key for screen rotate
MeasurementMethodwavelength
SpO2 Range &ResolutionRange: 0% to 100%resolution: 1%
SpO2 AccuracyRange 70% to 100% range ± 2%,less than 70% are unspecified
Pulse RateRange &ResolutionRange: 30 to 250 bpm;resolution: 1 bpm
Pulse RateAccuracy±2 bpm or ±2%, whichever is greater
Water-resistanceAgainst water splash (IP22 Approved)
Battery Type1 AAA-size Alkaline battery
Usage Life> 18 hrs typical operation under default setting
Shelf Life3 years

Table 1 Technical Specification

{6}------------------------------------------------

AmbientTemperatureOperation: 5 °C - 40 °C (41 °F - 104 °F);Storage: -30°C ~ 70 °C (-22 °F ~ 158 °F)
AtmosphericPressureOperation & storage are both 700 hPa - 1060 hPa
HumidityOperation & storage are both 10% - 90%, non-condensing

INDICATIONS FOR USE 5

The A ViTA Pulse Oximeter is intended for measuring functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adults as non-invasive spot checking in professional caring environment. It is designed for fingers between 0.8cm and 2.5cm (0.3 inches to 1 inches) and for patients during no-motion condition. The device is prescription only.

The Indications for Use statement for the AVITA Pulse Oximeter is not identical to the predicate device; however, the differences do not alter the intended diagnostic use of the device nor do they affect the safety and effectiveness of the device relative to the predicate. Both the subject and predicate devices have intended use for measuring functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adults as noninvasive spot checking in professional caring environment.

ItemSubject DevicePredicate Device
Product NameAVITA Pulse OximeterLeadtek Fingertip Pulse Oximeter
Model No.SP618D01B and 8D01C
510(k) Information
RegulationNumber870.2700870.2700
ClassificationClass IIClass II
Product CodeDQADQA
Indication for Use
StatementThe AVITA Pulse Oximeter is intended for measuring functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adults as non-invasive spot checking in professional caring environment. It is designed for fingers between 0.8cm and 2.5cm (0.3 inches to 1 inches) and for patients during no-motion condition. The device is prescription only.The 8D01B and 8D01C are intended for measuring functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for both adults and adolescent as non-invasive spot checking in home and professional caring environment. It is designed for fingers between 0.8cm and 2.3cm (0.3 inches to 0.9 inches) and for patients during no-motion condition.
Populationadultsadults and adolescent
Application siteFingerFinger
Performancenormal conditionnormal condition
Stand-alone ormodulestand-alonestand-alone
Single use or notmultiple usemultiple use
Use environmentprofessional caring environmenthome and professional caring environment
Test Principle
PrincipleThe LED contains a red light and an infraredlight that are differentially absorbed byoxygenated and oxygenated hemoglobin. Basedon the relative absorption of the twowavelengths that is determined by the sensor,the POX determines the relative amount ofoxygenated and deoxygenated hemoglobin,which is calculated as SpO2. In order to makethe SpO2calculation independent of skin color, fingersize, etc., the pulse oximeter sensor uses onlythe time varying light absorption componentgenerated by the patient's pulse.Determine the functional oxygensaturation of arterial hemoglobin (SpO2)and pulse rate by measuring the absorptionof red and infrared light passing throughperfused tissue.Changes in absorption caused by thepulsation of blood in the vascular bed areused to determine SpO2 reading and pulserate.
WavelengthDual wavelength LED(660 nanometers @ 3.2mW and 905 nanometers@2.4mW; both as max averageDual wavelength LED(660 nanometers @ 0.8mW and 940nanometers @ 1.2mW; both as maxaverage
Energy
TypeBatteryBattery
BatteryAAA Alkaline battery x 1AAA Alkaline battery x 1
Operation Features
On/OffAutomatic turn on and offAutomatic turn on and off
DisplayTwo color OLEDFull color OLED
Input KeyA 4-directional keyA 5-directional key (8D01B) or a singlepush-down (8D01C) key
Warning/IndicatorVisual indicator8D01B: Audio and visual warning8D01C: Visual indicator
Warning /IndicatorFunctionReading starts to flash as an indicator to userwhen SpO2 and Pulse rate drop out of thesetting range.8D01B: Appear red color with beepsounds when SpO2 and pulse rate out ofthe setting range.Low SpO2 warning:default 87%; setting range: 50% to 95%High SpO2 warning:default off; setting range: 80% to 100%Low HR warning:default off; setting range: 30 to 110 bpmHigh HR warning:default off; settingrange: 75 to 250 bpm
Display RotationYesYes
General Specification
Usage Life> 24 hrs typical operation under default setting> 24 hrs typical operation under defaultsetting
OperatingTemperature5 °C to 40 °C (41 °F to 104 °F)5 °C to 40 °C (41 °F to 104 °F)
StorageTemperature-3070°C (-22158 °F)-30°C to 70 °C (-22 °F to 158 °F)
Humidity10% - 90% (non-condensing)10% to 90%, non-condensing for bothoperating and storage
AtmosphericPressure700 hPa – 1060 hPa for both operating andstorage700 hPa - 1013 hPa for both operating andstorage
WaterResistanceIP22IP22
Classification
Applied PartType BFType BF
SafetyIEC 60601-1IEC 60601-1
EMCIEC 60601-1-2IEC 60601-1-2
HarmonizedStandardISO 80601-2-61ISO 80601-2-61
Mode ofOperationSpot checkingSpot checking
Appearance
Weightweight without battery: 26g (0,92 ounces)weight without battery: 26g (0,92 ounces)
SizeL68mm (2.68") x W37.8mm (1.49") x H28mm(1.1")L67.5 mm (2.63”) x W38 mm (1.48”) xH25 mm (0.98")
Pulse Oximetry
Range70%-100%0% to 100%
ResolutionResolutionResolution
Accuracy70% to 100% range ± 2%, less than 70% areunspecified70% to 100% range ± 2%, less than 70%are unspecified
Biocompatibility Testing
CytotoxicityIn accordance with ISO 10993-1In accordance with ISO 10993-1
SkinsensitizationIn accordance with ISO 10993-1In accordance with ISO 10993-1
Skin irritationIn accordance with ISO 10993-1In accordance with ISO 10993-1
Heart Rate Specification
Range30 to 250 bpm30 to 250 bpm
Resolution1 bpm1 bpm

DEVICE COMPARISON TABLE 6

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

7 PERFORMANCE SUMMARY

The following tests were conducted to evaluate the safety and effectiveness of the subject device, and the test results indicated that the subject device is safe and effective.

Performance Data【807.92(b)】 7.1

All necessary performance testing was conducted oxygenated hemoglobin characterization testing on the subject device to support a determination of substantial equivalence to the predicate device in accordance with ISO 80601-2-61.

7.2 Electrical Safety and EMC Testing

The laboratory tests of electrical safety, electromagnetic compatibility, and reliability testing were conducted and showed that the subject device complied with IEC 60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance, IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests, IEC 60601-1-11 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in home healthcare environment, and ISO 80601-2-61 Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment.

7.3 Biocompatibility testing

The biocompatibility evaluation for the subject device was in accordance with the FDA Biocompatibility guidance (Use of International Standard ISO10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"). The standards below are tested and met the acceptance criteria.

  • Biological evaluation (ISO10993-1)
  • Cytotoxicity (ISO10993-5)
  • Sensitization (ISO10993-10)
  • Irritation (ISO10993-23)

7.4 Software Verification and Validation

Software verification and validation were provided in compliance with FDA Guidance "The Content of the Premarket Submission for Software Contained in Medical Devices". The verifications and validations demonstrate that the subject device work functionally. The subject device is considered as a "moderate" level of concern, which is identical to the predicate device. A failure or latent flaw in the software could not directly cause serious injury or death to the patient or operator, but a non-serious injury could occur. According to FDA Guidance document, the software validation documentation summarized the required for a Moderate level of concern device.

{10}------------------------------------------------

7.5 Cleaning Validation

Cleaning validation was executed in accordance with FDA Guidance "Reprocessing Medical Device in Health Care Setting: Validation Methods and Labeling" The performance of the subject device will not be affected after multiple cleaning procedures as illustrated in user manual.

7.6 Clinical Performance

Clinical performance was conducted per ISO 80601-2-61. The clinical validation testing of the SpO2 performance under no motion on healthy, adult volunteers in the range of 70% to 100%. The ARMS for SpO2 under no motion was found to be 1.89%. No adverse effects and complications happened during the clinical study.

SUBSTANTIAL EQUIVALENCE 8

Based upon equivalences in: intended use, patient population, conditions of use, operating principles, and the non-clinical performance data, the subject device have been shown to be safe and effective and to perform equivalently as compared to the legally marketed predicate device. Therefore, the subject devices are substantially equivalent to the legally marketed predicate device.

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).