K Number
K193350

Validate with FDA (Live)

Date Cleared
2020-04-27

(146 days)

Product Code
Regulation Number
870.2700
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Leadtek Fingertip Pulse Oximeter are intended for measuring functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for both adults and adolescent as non-invasive spot checking in home and professional caring environment. It is designed for fingers between 0.8cm and 2.3cm (0.3 inches to 0.9 inches) and for patients during no-motion condition.

Device Description

The main function of the Leadtek Fingertip Pulse Oximeter is to measure the functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate as non-invasive spot checking. The Fingertip Pulse Oximeter determines the functional oxygen saturation of arterial hemoglobin (SpO2) by measuring the absorption of red and infrared light passing through perfused tissue. Changes in absorption caused by the pulsation of blood in the vascular bed are used to determine SpO2 reading and pulse rate. The characteristic of the device is a stand-alone device with 1 AAA Alkaline battery. It includes an OLED display screen and a warning or indicator function for abnormal readings.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text for the Leadtek Fingertip Pulse Oximeter (K193350).

Key Takeaways from the Document:

  • This device is a Fingertip Pulse Oximeter intended for non-invasive spot checking of SpO2 and pulse rate in adults and adolescents.
  • It's seeking Substantial Equivalence (SE) to a predicate device (Onyx Vantage 9590, K112843).
  • The primary performance evaluation is for SpO2 accuracy and Heart Rate (HR) accuracy.
  • The device's performance is gauged against recognized a standard: ISO 80601-2-61:2011 (Medical electrical equipment – Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment).

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Set by Standard/Regulatory Expectation for Pulse Oximeters)Reported Device Performance (Leadtek Fingertip Pulse Oximeter)
SpO2 Accuracy Range: 70% to 100%70% to 100%
SpO2 Accuracy (Arms - Root-Mean-Square): ≤ 2% (within 70%-100% range)1.72% (within 70%-100% range)
SpO2 Resolution: 1%1%
SpO2 Range: 0% to 100%0% to 100%
Heart Rate Range: 30 to 250 bpm30 to 250 bpm
Heart Rate Resolution: 1 bpm1 bpm
Heart Rate Accuracy: ±1 bpm or ±1% (whichever is greater)±1 bpm or ±1% (whichever is greater)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: A total of 10 subjects were enrolled for the clinical evaluation.
  • Data Provenance:
    • Country of Origin: Not explicitly stated in the provided text.
    • Retrospective or Prospective: The clinical evaluation was "conducted," implying a prospective study where data was collected for this submission. The text doesn't mention the use of pre-existing data.

3. Number of Experts Used to Establish Ground Truth and Qualifications

  • The document describes a clinical evaluation where SpO2 accuracy was measured against a reference method (likely arterial blood gas analysis, though not explicitly stated, this is standard for pulse oximeter accuracy studies).
  • It does not mention the use of "experts" in the sense of human readers interpreting data, as would be common for an AI/imaging device. The ground truth for pulse oximetry is derived from instrumental measurements, not human interpretation. Therefore, the concept of "number of experts" and their "qualifications" for establishing ground truth is not applicable in this context.

4. Adjudication Method for the Test Set

  • Not applicable. As noted above, the ground truth for SpO2 accuracy is typically established by invasively measured arterial oxygen saturation (SaO2) via co-oximetry, not by human adjudication of readings from the device itself.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No, an MRMC study was not done. This type of study is relevant for diagnostic imaging devices where human readers interpret images, sometimes with AI assistance. For a pulse oximeter, the core performance is algorithmic (device output) and doesn't involve human interpretation to the same degree. The study here focuses on the device's accuracy against a clinical reference standard.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

  • Yes, a standalone performance evaluation was indeed done. The clinical evaluation specifically assessed the "Leadtek Fingertip Pulse Oximeter" and its "SpO2 accuracy" and "performance results" against a reference standard. This is the direct, unassisted performance of the device's algorithm.

7. The Type of Ground Truth Used

  • The ground truth for SpO2 accuracy was established through clinical evaluation conducted in accordance with ISO 80601-2-61. This standard typically requires comparing the device's SpO2 readings with values obtained from a reference co-oximeter measuring arterial oxygen saturation (SaO2) in desaturation studies. The document states "The functional oxygen saturation (SpO2) measurements were validated in accordance with ISO 80601-2-61." and that the accuracy was "calculated using the root-mean-square (Arms)." This strongly implies direct physiological measurements from subjects desaturated to various SpO2 levels.

8. The Sample Size for the Training Set

  • The document does not mention a "training set" or "training data." This device is a traditional medical device (hardware with embedded software for signal processing and calculation), not an AI/Machine Learning model in the modern sense that typically requires a large training dataset distinct from a test set. The performance validation relies on the device itself as a complete system, built according to specified engineering principles, and then tested for accuracy and safety.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable, as no separate "training set" is described or implied for this type of device and its validation.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Leadtek Research Inc. Sharon Peng Regulatory Affairs Specialist 18F, No.166, Jian-Yi Rd., Zhonghe Dist New Taipei City, 23511 Tw

Re: K193350

Trade/Device Name: Leadtek Fingertip Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: March 23, 2020 Received: April 8, 2020

Dear Sharon Peng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing. Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K193350

Device Name Leadtek Fingertip Pulse Oximeter

Indications for Use (Describe)

The Leadtek Fingertip Pulse Oximeter are intended for measuring functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for both adults and adolescent as non-invasive spot checking in home and professional caring environment. It is designed for fingers between 0.8cm and 2.3cm (0.3 inches to 0.9 inches) and for patients during nomotion condition.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K193350

This summary of 510(K) information is being submitted in accordance with the requirements of 21 CFR 807.92

1. Applicant Information

Applicant:Leadtek Research Incorporation
Address:18F, No.166, Jian-Yi Rd., Chung Ho Dist23511 New Taipei City, Taiwan
Phone:+886-2-8226-5800
Fax Number:+886-2-8226-5801
Contact Person:Sharon PengRegulatory Affairs Specialist
Date Prepared:March 26, 2020

2. Device Information

Proprietary Name:Leadtek Fingertip Pulse Oximeter
Common Name:Pulse Oximeter
Classification Name:Oximeter
Regulatory Class:Class II
Regulation Number:21CFR 870.2700
Product Code:DQA
Review Panel:Anesthesiology

3. Predicate Device

Proprietary Name:Onyx Vantage 9590
510(K) Number:K112843
Manufacturer:Nonin Medical Inc.

4. Intended Use

The Leadtek Fingertip Pulse Oximeter are intended for measuring functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate for both adults and adolescent as non-invasive spot checking in home and professional caring environment. It is designed for fingers between 0.8cm and 2.3cm (0.3 inches to 0.9 inches) and for patients during no-motion condition.

18F.,No.166, Jian-Yi Rd., Chung Ho Dist., New Taipei City, 23511, Taiwan

TEL: 886-2-8226-5800 FAX: 886-2-8226-5801

510(k) No. K193350

{4}------------------------------------------------

5. Device Description

The main function of the Leadtek Fingertip Pulse Oximeter is to measure the functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate as non-invasive spot checking. The Fingertip Pulse Oximeter determines the functional oxygen saturation of arterial hemoglobin (SpO2) by measuring the absorption of red and infrared light passing through perfused tissue. Changes in absorption caused by the pulsation of blood in the vascular bed are used to determine SpO2 reading and pulse rate. The characteristic of the device is a stand-alone device with 1 AAA Alkaline battery. It includes an OLED display screen and a warning or indicator function for abnormal readings.

6. Substantial Equivalence Information:

The intended use and test principle of Leadtek Fingertip Pulse Oximeter is similar to that of the predicates, Onyx Vantage 9590. The main function of Leadtek Fingertip Pulse Oximeter is to measure the functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate as non-invasive spot checking by measuring the absorption of red and infrared light passing through perfused tissue.

The main differences between the proposed device and predicated device is that Leadtek Fingertip Pulse Oximeter has an input key for the operation and a warning/ indicator function which will show RED and make beep sounds when the value of SpO2 or pulse rate is out of the setting range. The power supply source is same as the predicates with only one battery. A comparison table between the proposed device and the predicated device is in the following below.

ItemProposed DevicePredicate Device
ProductNameLeadtek Fingertip Pulse OximeterOnyx Vantage 9590
Model No.8D01B and 8D01COnyx 9590
510(k) Information
RegulationNumber870.2700870.2700
ClassificationClass IIClass II
Product CodeDQADQA
Indication for Use
StatementThe 8D01B and 8D01C are intendedThe Nonin® Onyx Vantage 9590 Finger
ItemProposed DevicePredicate Device
ProductNameLeadtek Fingertip Pulse OximeterOnyx Vantage 9590
Model No.8D01B and 8D01COnyx 9590
for measuring functional oxygensaturation of arterial hemoglobin(SpO2) and pulse rate for both adultsand adolescent as non-invasive spotchecking in home and professionalcaring environment. It is designed forfingers between 0.8cm and 2.3cm (0.3inches to 0.9 inches) and for patientsduring no-motion condition.Pulse Oximeter is a small, lightweight,portable device indicated for use inmeasuring and displaying functionaloxygen saturation of arterial hemoglobin(%SpO2) and pulse rate of patients whoare well or poorly perfused. It is intendedfor spot-checking of adult and pediatricpatients on digits, including the thumband toes, that are between 0.3 - 1.0 inch(0.8 -2.5 cm) thick. The device's intendeduse environments include hospitals,clinics, long-term care facilities, skillednursing facilities, emergency medicalservices, and home healthcare services.
Populationadults and adolescent patientsadult and pediatric patients
ApplicationsiteFingerFinger, thumb, and toes
Performancenormal conditionnormal and low perfusion conditions
Stand-aloneor modulestand-alonestand-alone
Single use ornotmultiple usemultiple use
Useenvironmenthome and professional caringenvironmenthome and professional caringenvironment
ComparisonThe proposed device and the predicated device have the same intended use andclassification. All changes in indications for use would not affect the safety andeffectiveness. The Leadtek Fingertip Pulse Oximeter is substantially equivalent tothe predicate device (K112843) concerning the same intended use.
Test Principle
PrincipleDetermine the functional oxygensaturation of arterial hemoglobinDisplays numerical values for functionaloxygen saturation of arterial hemoglobin
ItemProposed DevicePredicate Device
ProductNameLeadtek Fingertip Pulse OximeterOnyx Vantage 9590
Model No.8D01B and 8D01COnyx 9590
(SpO2) and pulse rate by measuringthe absorption of red and infraredlight passing through perfused tissue.Changes in absorption caused by thepulsation of blood in the vascular bedare used to determine SpO₂ readingand pulse rate.(SpO2) and pulse rate by measuring theabsorption of red and infrared (IR) lightpassing through perfused tissue.Changes in absorption caused by thepulsation of blood in the vascular bed areused to determine SpO₂ and pulse rate.
WavelengthDual wavelength LED(660 nanometers @ 0.8mW and 940nanometers @ 1.2mW; both as maxaverageDual wavelength LED(660 nanometers @ 0.8mW and 910nanometers @ 1.2mW; both as maxaverage
ComparisonThe difference between the proposed device and the predicate device is theinfrared wavelength. There is no additional question of safety and effectiveness ascompared to the predicate device. The Leadtek Fingertip Pulse Oximeter issubstantially equivalent to the predicated device (K112843) concerning the testprinciple.
Energy
TypeBatteryBattery
BatteryAAA Alkaline battery x 1AAA Alkaline battery x 2
ComparisonThere is no additional question of safety and effectiveness as compared to thepredicate device raised by the battery. The Leadtek Fingertip Pulse Oximeter issubstantially equivalent to the predicate device (K112843) concerning the energysource.
Operation Features
On/OffAutomatic turn on and offAutomatic turn on and off
DisplayFull color OLEDLED
Input KeyA 5-directional key (8D01B) or a singlepush-down (8D01C) keyNone
Warning /Indicator8D01B: Audio and visual warning8D01C: Visual indicatorVisual indicator
Warning /8D01B:3 color pulse quality indicator to provide
ItemProposed DevicePredicate Device
ProductNameLeadtek Fingertip Pulse OximeterOnyx Vantage 9590
Model No.8D01B and 8D01COnyx 9590
IndicatorFunctionAppear red color with beep soundswhen SpO2 and pulse rate out of thesetting range.assurance of reading quality■ Green Light: Good Reading■ Yellow Light: Fair Reading■ Red Light: Poor Reading.
■ Low SpO2 warning:default 87%; setting range: 50% to95%
■ High SpO2 warning:default off; setting range: 80% to100%
■ Low HR warning:default off; setting range: 30 to 110bpm
■ High HR warning:default off; setting range: 75 to 250bpm
DisplayRotationYesNone
ComparisonThe differences of operation features will not be considered as a NSE between theproposed device and the predicate device. The Leadtek Fingertip Pulse Oximeteris substantially equivalent to the predicate device (K112843) concerning thetechnological characteristics.
General Specification
Usage Life> 24 hrs typical operation underdefault setting6,000 spot checks or 36 hours ofoperation
OperatingTemp.5 ºC to 40 ºC ( 41 ºF to 104 ºF)-5 ºC to 40 ºC ( 23 ºF to 104 ºF)
StorageTemp.-30ºC to 70 ºC (-22 ºF to 158 ºF)-40ºC to 70 ºC (-40 ºF to 158 ºF)
Humidity10% to 90%, non-condensing for bothoperating and storage10% to 90%, non-condensing foroperating10% to 95%, non-condensing for storage
Atmospheric700 hPa - 1013 hPa for both operating
ItemProposed DevicePredicate Device
ProductNameLeadtek Fingertip Pulse OximeterOnyx Vantage 9590
Model No.8D01B and 8D01COnyx 9590
Pressureand storage
WaterResistanceIP22IP32
Classification
Applied PartType BFType BF
SafetyIEC 60601-1IEC 60601-1
EMCIEC 60601-1-2IEC 60601-1-2
HarmonizedStandardISO 80601-2-61ISO 80601-2-61
Mode ofOperationSpot checkingcontinuous monitoring
Appearance
Weightweight without battery: 26g (0,92 ounces)less than 60g (2 ounces)
SizeL67.5 mm (2.63") x W38 mm (1.48") xH25 mm (0.98")L55.9 mm (2.2") x W33 mm (1.3") xH32.3 mm (1.27")
ComparisonThe differences of general specification and appearance will not be considered as a NSE between the proposed device and the predicate device. The LeadtekFingertip Pulse Oximeter is substantially equivalent to the predicate device (K112843) concerning the general specification and appearance.
Pulse Oximetry and Heat Rate Specification
Range0% to 100%0% to 100%
Resolution1%1%
Accuracy70% to 100% range ± 2%, less than70% are unspecified70% to 100% range ± 2%, less than 70%are unspecified
Heat Rate Specification
Range30 to 250 bpm18 to 321 bpm
Resolution1 bpm1 bpm
Accuracy±1 bpm or ±1%, whichever is greater20 to 250 bpm ±3 digits
ItemProposed DevicePredicate Device
ProductNameLeadtek Fingertip Pulse OximeterOnyx Vantage 9590
Model No.8D01B and 8D01COnyx 9590
ComparisonThe difference of Heart Rate Specification will not be considered as a NSEbetween the proposed device and the predicate device. The Leadtek FingertipPulse Oximeter is substantially equivalent to the predicate device (K112843)concerning the performance specification.

18F.,No.166, Jian-Yi Rd., Chung Ho Dist., New Taipei City, 23511, Taiwan

8

TEL: 886-2-8226-5800 FAX: 886-2-8226-5801

{5}------------------------------------------------

TEL: 886-2-8226-5800

FAX: 886-2-8226-5801

{6}------------------------------------------------

{7}------------------------------------------------

8

TEL: 886-2-8226-5800 FAX: 886-2-8226-5801

{8}------------------------------------------------

TEL: 886-2-8226-5800

{9}------------------------------------------------

7. Functional and Safety Testing:

The Leadtek Fingertip Pulse Oximeter is designed and conducted in accordance with the related standards. The following tests were performed to verify and validate that Leadtek Fingertip Pulse Oximeter meets all requirements of related standards and demonstrates substantial equivalence to the predicates.

Electrical Safety and EMC Testing

The laboratory tests of electrical safety, electromagnetic compatibility, and reliability testing were conducted and show that the proposed device complies with IEC 60601-1:2012 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance, IEC 60601-1-2:2014 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests, IEC 60601-1-11:2015 Medical electrical equipment – Part 1-11: General requirements for basic safety and essential performance – Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in home healthcare environment, and ISO 80601-2-61:2011 Medical electrical equipment – Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment.

Biocompatibility Testing

The biocompatibility testing for Leadtek Fingertip Pulse Oximeter was in accordance with ISO 10993-1:2009 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process" as recognized by FDA. The studies of cytotoxicity, skin sensitization, and skin irritation were performed and show that the skin contacting materials do not cause the potential risks.

Software Verification and Validation

Software verification and validation were provided in compliance with FDA Guidance for the Content of the Premarket Submission for Software Contained in Medical Devices. These

18F.,No.166, Jian-Yi Rd., Chung Ho Dist., New Taipei City, 23511, Taiwan

TEL: 886-2-8226-5800 FAX: 886-2-8226-5801

{10}------------------------------------------------

verifications and validations demonstrate that Leadtek Fingertip Pulse Oximeter work functionally. The software for the device is considered as a "moderate" level of concern, which is identical to the predicate devices. A failure or latent flaw in the software could not directly cause serious injury or death to the patient or operator, but a non-serious injury could occur. According to FDA Guidance document, the software validation documentation summarized the required for a Moderate level of concern device.

Cleaning Validation

Cleaning and disinfection validations were executed in accordance with FDA Guidance "Reprocessing Medical Device in Health Care Setting: Validation Methods and Labeling" The performance of Leadtek Fingertip Pulse Oximeter will not be affected after multiple cleaning and disinfection procedures as illustrated in user manual.

Performance Testing

The laboratory and clinical evaluations for the performance of Leadtek Fingertip Pulse Oximeter demonstrate that the modifications do not raise different questions of safety and effectiveness. The root-mean-square (Arms) was used to analyze the accuracy of Leadtek Fingertip Pulse Oximeter, per ISO 80601-2-61:2011, the SpO2 accuracy results demonstrate that the root-mean-square (Arms) value of Leadtek Fingertip Pulse Oximeter is less than 2% within the range 70% - 100%.

Clinical Evaluation

The functional oxygen saturation (SpO2) measurements were validated in accordance with ISO 80601-2-61. The clinical evaluation for SpO2 accuracy was conducted on healthy male and female, light to dark skinned subjects, and total 10 subjects were enrolled. The SpO2 accuracy data was calculated using the root-mean-square (Arms); the performance results showed the Leadtek Fingertip Pulse Oximeter has an Arms of 1.72 during steady state conditions over the range of 70-100% and it is also in compliance with the specified performance claimed by the manufacturer.

8. Conclusion

The Leadtek Fingertip Pulse Oximeter has the same intended use and similar characteristics as that of the predicate device manufactured by Nonin Medical Inc. and cleared by FDA under K112843. Based on the information contained in this submission demonstrate that any different characteristics do not raise different questions of safety and effectiveness.

18F.,No.166, Jian-Yi Rd., Chung Ho Dist., New Taipei City, 23511, Taiwan

TEL: 886-2-8226-5800 FAX: 886-2-8226-5801

510(k) No. K193350

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).