K Number
K223326

Validate with FDA (Live)

Device Name
Axiom PSR System
Manufacturer
Date Cleared
2023-03-30

(150 days)

Product Code
Regulation Number
888.3110
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Axiom PSR System is intended to be used as patient specific surgical planning and instrumentation to assist in the positioning of total ankle replacement components intraoperatively, and in guiding bone cutting. The Axiom PSR System is intended for use with the Kinos Axiom Total Ankle System and its cleared indications for use.

Device Description

restor3d's Axiom PSR System is an ankle surgical guide system, designed for an individual patient from the patient's CT imaging scans. The subject Axiom PSR System is intended to aid in the placement of Kinos Axiom Total Ankle System (K192778) implant components when used in conjunction with the Kinos Axiom standard, reusable instruments. The Axiom PSR System includes preoperative plans developed with and approved by the operative surgeon, tibia and talus resection guides, manufactured from implant grade titanium, bone models manufactured from biocompatible polymer for use with the Kinos Axiom Total Ankle System.

AI/ML Overview

The restor3d Axiom PSR System is a patient-specific surgical planning and instrumentation system designed to assist in the positioning of total ankle replacement components intraoperatively and in guiding bone cutting. The information provided does not include specific acceptance criteria with numerical thresholds. However, the study conducted aims to demonstrate that the Axiom PSR System is at least as accurate as, or more accurate than, the previously cleared Kinos Axiom Total Ankle System standard instruments alone, with respect to implant placement relative to the pre-operative plan in six degrees of freedom.

Here's a breakdown of the requested information based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance:

The document does not provide a table with specific numerical acceptance criteria. Instead, it states a comparative performance objective.

Acceptance Criteria (Implied)Reported Device Performance
The Axiom PSR System should be at least as accurate or more accurate than the Kinos Axiom Total Ankle System standard instruments alone regarding implant placement relative to the pre-operative plan in six degrees of freedom.Cadaver validation testing demonstrated that the subject Axiom PSR System was at least as accurate or more accurate than using the previously cleared Kinos Axiom Total Ankle System standard instruments alone, with respect to implant placement relative to the preoperative plan in six degrees of freedom.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size for Test Set: The document mentions "cadaver testing" but does not specify the number of cadavers or individual surgical procedures performed.
  • Data Provenance: Retrospective or Prospective is not explicitly stated, but "cadaver testing" implies a controlled, experimental setting, likely
    prospective for the purpose of the study. The origin (country/region) of the cadavers is not mentioned.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

This information is not provided in the document. The ground truth ("pre-operative plans") is established as a digital design workflow, which is later compared to post-operative CT scans. The experts involved in creating these pre-operative plans and evaluating the post-operative scans are not specified.

4. Adjudication Method for the Test Set:

The document mentions that post-operative CT scans were obtained, and the final implant positions were evaluated in 3D space about six degrees of freedom and then compared to the pre-operative plans. It does not describe any specific multi-reviewer adjudication method (e.g., 2+1, 3+1). The evaluation method appears to be a direct comparison of physical outcomes (post-operative CT) to digital plans.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

A multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly stated or described. The study compared the device's performance to the standard method rather than comparing human readers with and without AI assistance. The "AI" aspect is in the patient-specific surgical planning and guiding, not in assisting human readers in diagnosis or interpretation.

6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance):

This aspect is not directly applicable in the typical sense of standalone AI performance for diagnostic algorithms. The Axiom PSR System is a surgical guidance system, and its "performance" is inherently tied to its use in a surgical setting. The digital design workflow is an algorithmic process that creates the patient-specific guides and plans, but the "standalone" performance isn't measured in isolation from the surgical act. The evaluation "compared to the pre-operative plans" could be considered an evaluation of the system's output (the guides) against its intended design.

7. Type of Ground Truth Used:

The ground truth for comparison was the "pre-operative plans." These plans are a digital design output, established through a digital design workflow. The "final implant positions" from post-operative CT scans were then compared against these pre-operative plans.

8. Sample Size for the Training Set:

The document does not provide any information regarding a training set sample size. The Axiom PSR System is a patient-specific device based on individual patient CT scans, not a machine learning model that requires a separate training set for algorithm development in the traditional sense. The "digital design workflow" is likely based on established biomechanical and engineering principles rather than supervised machine learning requiring a training dataset.

9. How the Ground Truth for the Training Set Was Established:

Since there is no mention of a training set or a machine learning model requiring such a set, this information is not applicable based on the provided document. The device's functionality is based on individual patient imaging and a digital design workflow to create custom surgical guides and plans, not on learning from a large labeled training dataset.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 30, 2023

restor3d, inc. Brianna Prindle Head of Regulatory 311 W. Corporation St. Durham, North Carolina 27701

Re: K223326

Trade/Device Name: Axiom PSR System Regulation Number: 21 CFR 888.3110 Regulation Name: Ankle Joint Metal/Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: OYK Dated: February 24, 2023 Received: February 24, 2023

Dear Brianna Prindle:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Limin Sun -S

Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223326

Device Name Axiom PSR System

Indications for Use (Describe)

The Axiom PSR System is intended to be used as patient specific surgical planning and instrumentation to assist in the positioning of total ankle replacement components intraoperatively, and in guiding bone cutting. The Axiom PSR System is intended for use with the Kinos Axiom Total Ankle System and its cleared indications for use.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K223326 page 1 of 3

Image /page/3/Picture/1 description: The image shows the word "restor3d" in a bold, sans-serif font. The first six letters, "restor", are solid black. The last two characters, "3d", are outlined in black, with a white fill.

311 W. Corporation St. Durham, NC 27701 984.888.0593

510(k) Summary

510(k) Number: K223326

This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.

A. 510(k) Owner:

restor3d, inc. 311 W. Corporation St. Durham, NC 27701

B. Primary Correspondent:

Brianna Prindle Head of Regulatory brianna@restor3d.com

C. Premarket Notification:

Submission Type:Traditional 510(k)
Trade Name:Axiom PSR System
Classification:Class II
Product Code:OYK
Classification Name:Ankle Arthroplasty Implantation System
Regulation Number:21 CFR 888.3110
Regulation Name:Ankle joint metal/polymer semi-constrained cementedprosthesis

D. Indications for Use:

The Axiom PSR System is intended to be used as patient specific surgical planning and instrumentation to assist in the positioning of total ankle replacement components intraoperatively, and in guiding bone cutting. The Axiom PSR System is intended for use with the Kinos Axiom Total Ankle System and its cleared indications for use.

E. Predicate Devices:

Kinos Axiom PSR is substantially equivalent to the following devices:

510(k)Trade NameManufacturer
Primary Predicate Device
K193432Vantage PSI SystemMedical Modeling – A 3DSystems Company
Secondary Predicate Devices

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the word "restor3d" in a bold, sans-serif font. The first six letters, "restor," are filled in with black, while the "3d" is outlined in black. The background is white.

311 W. Corporation St. Durham, NC 27701 984.888.0593

510(k)Trade NameManufacturer
K192778Kinos Axiom Total Ankle Systemrestor3d
K220523Tidal Lumbar Interbody FusionDevicerestor3d

F. Device Description:

restor3d's Axiom PSR System is an ankle surgical guide system, designed for an individual patient from the patient's CT imaging scans. The subject Axiom PSR System is intended to aid in the placement of Kinos Axiom Total Ankle System (K192778) implant components when used in conjunction with the Kinos Axiom standard, reusable instruments. The Axiom PSR System includes preoperative plans developed with and approved by the operative surgeon, tibia and talus resection guides, manufactured from implant grade titanium, bone models manufactured from biocompatible polymer for use with the Kinos Axiom Total Ankle System.

G. Comparison of Technological Characteristics

The intended use and technological characteristics of the subject Axiom PSR System are either identical or substantially equivalent to the predicate device (Vantage PSI System, K193432).

The principles of operation and technological characteristics are substantially equivalent between the Axiom PSR System and the identified predicate. Specifically, the predicate device includes pre-operative planning from patient CT imaging scans, 3D printing of patient specific bone models and resection guides. Both the subject and predicate system perform bone resection cuts directly through the patient specific guides and are used to aid in the placement of a previously cleared total ankle system when used in conjunction with standard instruments of their respective total ankle system.

The subject Axiom PSR System uses the same fundamental technologies employed in the digital workflow as the predicate device, including medical image transformation and manipulation in a CAD environment with digital outputs subsequently used for 3D printing of the patient specific devices.

The subject Axiom PSR System resection guides and bone models are made from implant grade titanium alloy (Ti-6Al-4V) and biocompatible polymer. Whereas the predicate device resection guides are made from polyamide. The biocompatibility of the subject Axiom PSR System resection guides and bone models has been assessed through testing in support of this and other 510(k) submissions and does not represent any new or increased risks with respect to technological characteristics.

H. Summary of Performance Testing

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the word "restor3d" in a bold, sans-serif font. The first six letters, "restor," are solid black, while the "3d" is outlined in black. The letters are closely spaced, creating a compact and modern look.

311 W. Corporation St. Durham, NC 27701 984.888.0593

Non-clinical, cadaver testing was conducted to validate the digital design workflow and assess the accuracy of the subject Axiom PSR System, compared to the previously cleared Kinos Axiom Total Ankle System. Post-operative CT scans were obtained, and the final implant positions were evaluated in 3D space about six degrees of freedom and then compared to the pre-operative plans. Cadaver validation testing demonstrated that the subject Axiom PSR System was at least as accurate or more accurate than using the previously cleared Kinos Axiom Total Ankle System standard instruments alone, with respect to implant placement relative to the preoperative plan in six degrees of freedom.

Process qualification was performed to demonstrate manufactured devices meet the output specifications of the digital design workflow.

Cleaning, biocompatibility, sterilization, and shelf life were evaluated against previously conducted validations. The evaluations demonstrate acceptable adoption into previously established manufacturing processes, materials and specifications employed in restor3d's previously cleared implant and instrument systems, including the restor3d Tidal Lumbar Interbody Fusion Device (K220523), therefore the subject device does not represent any new or increased concerns of safety or efficacy.

l. Conclusions:

Based on comparison of the intended use, indications for use and technological characteristics, the subject Axiom PSR System is substantially equivalent to the predicate Vantage PSI System. Minor differences in the material and technological considerations do not raise any new or different questions of safety or effectiveness. Additionally, the cadaveric testing demonstrates the system performs in accordance with its intended use and is as safe, effective and performs as well as the previously cleared Kinos Axiom Total Ankle System (K192778) predicate device.

§ 888.3110 Ankle joint metal/polymer semi-constrained cemented prosthesis.

(a)
Identification. An ankle joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace an ankle joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces and has no linkage across-the-joint. This generic type of device includes prostheses that have a talar resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a tibial resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).(b)
Classification. Class II.