K Number
K223298

Validate with FDA (Live)

Date Cleared
2023-02-22

(118 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sterile Latex Examination Gloves Powder Free is a disposable device intended for medical purposes that is worn on the examiners' hand or finger to prevent contamination between patient and examiner.

Device Description

The Sterile Latex Examination Gloves Powder Free are natural color, single use only, disposable examination gloves intended for medical purposes to be worn by examination between the patient and the examiner. The device is ambidextrous and can be worn on either the left or right hand. The gloves are offered in sizes small, medium, large and extra-large, and sterilized by Gamma radiation. The gloves are designed and manufactured in accordance with the ASTM D3578-19 standard.

AI/ML Overview

The provided text describes the 510(k) summary for "Sterile Latex Examination Gloves Powder Free". This document focuses on demonstrating substantial equivalence to a predicate device based on non-clinical performance and biocompatibility testing, rather than an AI-driven medical device requiring extensive clinical validation of diagnostic accuracy. As such, many of the requested elements for AI/ML device studies (like sample sizes for test sets, expert ground truth adjudication, MRMC studies, training set details) are not applicable to the information provided.

However, I can extract the acceptance criteria and performance results for the non-clinical tests conducted for these gloves:

1. Table of Acceptance Criteria and Reported Device Performance

Test Method / ItemPurposeAcceptance CriteriaReported Device PerformanceResult
Dimensions (width)Evaluate the physical dimension of the gloveWidth:Pass
S: 80 ± 10 mmS: 81mm min
M: 95 ± 10 mmM: 95mm min
L: 105 ± 10 mmL: 106mm min
XL: 115 ± 10 mmXL: 115mm min
Dimensions (length)Evaluate the physical dimension of the gloveLength: 230mm min238mm min lengthPass
Dimensions (thickness)Evaluate the physical dimension of the glovePalm – 0.08mm min.Finger – 0.08mm min.Palm – 0.11mm min.Finger – 0.11mm minPass
Physical properties (Before aging)Evaluate tensile strength and ultimate elongation before agingTensile Strength: 18 MPa, min.Elongation: 650%, min.Stress at 500% Elongation: 5.5MPa, max.Tensile Strength: 23.4MPa, min.Elongation: 650%, min.Stress at 500% Elongation: 5.3MPa, max.Pass
Physical properties (After aging)Evaluate tensile strength and ultimate elongation after agingTensile Strength: 14 MPa, min.Elongation: 500%, min.Tensile Strength: 22MPa, min.Elongation: 704%, min.Pass
Freedom from holesDetect holes in the glovesNo leakage at sampling level of G-1, AQL 1.5No leakage, 314 of 315 passedPass
Residual PowderDetect powder residue in the glove<2mg per gloveAverage 0.09 mg per glovePass
Residual ProteinDetect protein residue in the glove<200µg/dm²Average 70.880µg/dm²Pass
SterilityDetect the sterility of the gloveSterileSterilePass
Irritation (ISO 10993-10)Demonstrate the safety of the subject device (biocompatibility)Under the conditions of the study, not an irritant. Comply with ISO 10993-10.Under the conditions of the study, not an irritant. Comply with ISO 10993-10.Pass
Sensitization (ISO 10993-10)Demonstrate the safety of the subject device (biocompatibility)Under the conditions of the study, not a sensitizer. Comply with ISO 10993-10.Under the conditions of the study, not a sensitizer. Comply with ISO 10993-10.Pass
Acute systemic toxicity (ISO 10993-11)Demonstrate the safety of the subject device (biocompatibility)Under the conditions of the study, the device does not induce acute systemic toxicityUnder the conditions of the study, the device does not induce acute systemic toxicityPass

Since this is not an AI/ML-driven medical device, the following sections are either not applicable (N/A) or the information is not provided in the document:

2. Sample size used for the test set and the data provenance:

  • For the Freedom from holes test, the document states "314 of 315 passed," implying a test set size of 315 gloves for this specific attribute.
  • For other physical and material property tests (Dimensions, Physical Properties, Residual Powder, Residual Protein, Sterility), specific sample sizes are not explicitly stated, but standard testing procedures for medical devices of this type would typically involve a statistically appropriate random sample from production lots.
  • Data provenance (country of origin, retrospective/prospective) is not specified for the test data, but the manufacturer is based in China. This is a non-clinical/material performance study.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

  • N/A. This is a physical product, and ground truth is established through standardized laboratory testing methods (e.g., ASTM and ISO standards), not human expert interpretation of images or clinical data.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • N/A. Ground truth is determined by objective laboratory measurements and tests according to cited standards, not by human adjudication of ambiguous cases.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • N/A. This is not an AI/ML device; therefore, MRMC studies involving human readers are irrelevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • N/A. This is not an AI/ML device. The "standalone" performance refers to the device's inherent physical and material properties when tested according to standards.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Ground truth for this device is based on objective measurements and results from standardized laboratory testing protocols as defined by ASTM International and ISO standards (e.g., ASTM D3578-19, ASTM D6124-06, ASTM D5151-19, ASTM D5712-15, ISO 10993-10, ISO 10993-11, ISO 11137-1, ISO 11137-2).

8. The sample size for the training set:

  • N/A. This device does not involve a "training set" in the context of machine learning or AI.

9. How the ground truth for the training set was established:

  • N/A. This device does not involve a "training set" or "ground truth establishment" for training purposes as understood in AI/ML.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

February 22, 2023

Shandong Intco Medical Products Co. Ltd. % Ivv Wang Consultant Shanghai Sungo Management Consulting Co. Ltd. 14th Floor, 1500# Century Avenue Shanghai. 200122 China

Re: K223298

Trade/Device Name: Sterile Latex Examination Gloves Powder Free (S, M, L, Extra L) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LYY Dated: January 10, 2023 Received: January 12, 2023

Dear Ivy Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bifeng Qian -S

Bifeng Qian M.D., Ph. D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

Submission Number (if known)

K223298

Device Name

Sterile Latex Examination Gloves Powder Free (S, M, L, Extra L)

Indications for Use (Describe)

The Sterile Latex Examination Gloves Powder Free is a disposable device intended for medical purposes that is worn on the examiners' hand or finger to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for INTCO Medical. The logo features the word "INTCO" in a bold, sans-serif font, with the letters in a dark blue color. Above the word "INTCO" is a stylized blue flower-like design with several petals radiating outward from a central point. Below the word "INTCO" are the Chinese characters for "Yingke Medical" in a lighter shade of gray.

510(K) Summary - K223298

(As requirement by 21 CFR 807.92)

Date prepared: 2023-02-22

A. Applicant: Name: Shandong Intco Medical Products Co., Ltd. Address: 9888 Qiwang Road, Naoshan Industrial Park, Qingzhou, Shandong, China Contact: Mr. John Zhao Title: Manager Tel: +86-536-6136888 Email: johnzhao@basicmedical.com

Submission Correspondent: Primary contact: Ms. Ivy Wang Shanghai SUNGO Management Consulting Co., Ltd. Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: haiyu.wang@sungoglobal.com Secondary contact: Mr. Raymond Luo Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: fda.sungo@gmail.com

B. Device:

Trade Name: Sterile Latex Examination Gloves Powder Free (S, M, L, Extra L) Common Name: Latex Patient Examination Gloves (Powder Free)

Regulatory Information Classification Name: Latex Patient Examination Glove Classification: Class I Product code: LYY Regulation Number: 21 CFR 880.6250 Review Panel: General Hospital

C. Predicate device:

K171367 SANCARE STERILE LATEX EXAMINATION GLOVES Sanrea Healthcare Products Pvt Ltd

D. Indications for use of the device:

The Sterile Latex Examination Gloves Powder Free is a disposable device intended for medical purposes that is worn on the examiners' hand or finger to prevent contamination between patient and examiner.

{4}------------------------------------------------

E. Device Description:

The Sterile Latex Examination Gloves Powder Free are natural color, single use only, disposable examination gloves intended for medical purposes to be worn by examination between the patient and the examiner. The device is ambidextrous and can be worn on either the left or right hand. The gloves are offered in sizes small, medium, large and extra-large, and sterilized by Gamma radiation.

The gloves are designed and manufactured in accordance with the ASTM D3578-19 standard.

F. Summary of Technological Characteristics

DeviceProposed DevicePredicate DeviceResult
510K #K223298K171367-
Product NameSterile Latex Examination Gloves PowderFreeSANCARE STERILE LATEXEXAMINATION GLOVES-
Product CodeLYYLYYSame
RegulationNumber21 CFR 880.625021 CFR 880.6250Same
Indications foruseThe Sterile Latex Examination Gloves PowderFree is a disposable device intended formedical purposes that is worn on theexaminers' hand or finger to preventcontamination between patient andexaminer.The Sterile Latex Examination Gloves,Powder Free, is a disposable deviceintended for medical purposes that isworn on the examiners' hand or fingerto prevent contamination betweenpatient and examiner.Same
Powder freeYesYesSame
Design featureAmbidextrousAmbidextrousSame
MaterialNatural Rubber LatexNatural Rubber LatexSame
OTC useYesYesSame
SterilityGamma sterilizedEthylene Oxide sterilizedDifferent
UseSinge useSingle useSame

Table 1 General Comparison of Proposed and Predicate Devices

Analysis: The proposed device uses a different sterilization method than the predicate device. But both sterilization methods are current commonly used sterilization methods and the sterilization process validation has been conducted to demonstrate its effectiveness. Therefore, this difference does not raise any new safety or performance questions.

DesignationSize
XSSMLXLTolerance
Predicatedevice(K171367)Length, mm230230230230230min
Width, mm708085111120$\pm$ 10
Thickness, mm
Finger0.08min
Palm0.08min

Table 2 Device dimension comparison

{5}------------------------------------------------

DesignationSizeTolerance
Proposeddevice(K223298)Length, mmS220M230L230XL230min
Width, mm8095111120$\pm 10$
Thickness, mm
Finger0.08min
Palm0.08min
ResultSimilar

Analysis: The physical dimensions are slightly different from that of the predicate, but they all meet the requirements of ASTM D3578-19, so the differences do not raise any new safety or performance questions.

Table 3 Performance comparison

ItemProposed device(K223298)Predicate device(K171367)Result
ColorNatureNatureSame
PhysicalpropertiesBeforeagingTensile strength18MPa, min18MPa, minSame
Ultimate elongation650%, min650%, minSame
Stress at 500%Elongation5.5 MPa, Max5.5 MPa, MaxSame
AfteragingTensile strength14MPa, min14MPa, minSame
Ultimate elongation500%, min500%, minSame
Comply with ASTM D3578Comply with ASTMD3578Same
Freedom from holesBe free from holes whentested in accordance withASTM D3578 & ASTMD5151G-1, AQL 1.5Be free from holes whentested in accordancewith ASTM D3578 &ASTM D5151G-1, AQL 2.5Different
Residual PowderMeet the requirementsof ASTM D3578 & ASTMD6124Less than 2 mg per gloveMeet the requirementsof ASTM D3578 & ASTMD6124Less than 2 mg per gloveSame
Residual ProteinMeet the requirementsof ASTM D3578 & ASTMD5712Less than 200µg/dm²Meet the requirementsof ASTM D3578 & ASTMD5712Less than 200µg/dm²Same

Analysis: The subject device tested the leakage performance with a different AQL as compared to the predicate device. A more stringent acceptance criterion does not affect the safety and performance of the proposed device.

Table 4 safety comparison

ItemProposed devicePredicate deviceResult
-------------------------------------------------

{6}------------------------------------------------

(K223298)(K171367)
MaterialNatural rubber latexNatural rubber latexSame
BiocompatibilityIrritationISO 10993-10Under theconditions of thestudy, not anirritant. Complywith ISO 10993-10.Comply with ISO10993-10Same
SensitizationISO 10993-10Under theconditions of thestudy, not asensitizer. Complywith ISO 10993-10.
Acute systemictoxicityISO 10993-11Under the /conditions of thestudy, the devicedoes not induceacute systemictoxicity responseDifferent

Analysis: Toxicity information for the predicate device is not publicly available. This does not raise different safety or performance questions since the subject device has acceptable biocompatibility per the biocompatibility endpoint assessment.

G. Summary of Non-Clinical Testing

A Biocompatibility

The following tests for the subject device were conducted the biocompatibility of Sterile Latex Examination Gloves Powder Free:

  • ISO 10993-10: Primary Skin Irritation
  • ISO 10993-10: Dermal Sensitization
  • ISO 10993-11: Acute Systemic Toxicity

A Performance Testing

Physical performance testing of the proposed device was conducted as per ASTM D3578-19 Standard Specification for Rubber Examination Gloves.

To summarize, the performance testing of the subject device was conducted to adequately demonstrate the effectiveness of the device in accordance with the relevant test methods cited below:

  • ASTM D3578-19 Standard Specification for Rubber Examination Gloves
  • ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual Powder on Medical Gloves
  • ASTM D5151-19 Standard Test Method for Detection of Holes in Medical Gloves
  • ASTM D5712-15 Standard Test Method for Analysis of Aqueous Extractable Protein in Natural Rubber in Latex Natural Rubber and Elastomeric Products Using the Modified Lowry Method
Test MethodPurposeAcceptance CriteriaResults
Dimensions (width)The purpose of theWidth $80\pm10$ mm for Small sizePass
(thickness)test is to evaluatethe physicaldimension of theglove$95\pm10$ mm for Medium size$105\pm10$ mm for Large size$115\pm10$ mm for Extra largesize81mm min for S95mm min for M106mm min for L115mm min for Extra L
Length 230mm min238mm min length
Palm – 0.08mm min.Finger-0.08mm min.PassPalm – 0.11mm min.Finger—0.11mm min
Physical propertiesThe purpose of thetest is to evaluatethe tensile strengthand ultimateelongation beforeand after agingBefore Aging:Tensile Strength: 18 MPa, min.Elongation: 650%, min.Stress at 500% Elongation:5.5MPa, max.After Aging:Tensile Strength: 14 MPa, min.Elongation: 500%, min.PassBefore Aging:Tensile Strength: 23.4MPa, min.Elongation: 650%, min.Stress at 500% Elongation:5.3MPa, max.After Aging:Tensile Strength: 22MPa, min.Elongation: 704%, min.
Freedom from holesThe purpose of thetest is to detectholes in the glovesNo leakage at sampling level ofG-1, AQL 1.5PassNo leakage, 314 of 315 passed
Residual PowderThe purpose of thetest is to detect thepowder residue inthe glove<2mg per glovePassaverage 0.09 mg per glove
Residual ProteinThe purpose of thetest is to detect theprotein residue inthe glove<200µg/dm²PassAverage 70.880µg/dm²
SterilityThe purpose of thetest is to detect thesterility of the gloveSterilePassSterile
IrritationISO 10993-10The purpose of thetesting is todemonstrate thesafety of the subjectdevice.Under the conditions of thestudy, not an irritant. Complywith ISO 10993-10.Under the conditions of thestudy, not an irritant. Complywith ISO 10993-10.
SensitizationISO 10993-10Under the conditions of thestudy, not a sensitizer. Complywith ISO 10993-10.Under the conditions of thestudy, not a sensitizer. Complywith ISO 10993-10.
Acute systemictoxicityISO 10993-11Under the conditions of thestudy, the device does notinduce acute systemic toxicityUnder the conditions of thestudy, the device does notinduce acute systemic toxicity

{7}------------------------------------------------

{8}------------------------------------------------

A Sterilization

The proposed device is provided Gamma sterilized to achieve the sterility Assurance Level (SAL) of 10v. The sterilization process validation was conducted in accordance with ISO 11137-1 Sterilization of health care products Radiation Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices, and ISO11137-2 Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose.

H. Clinical Test Conclusion

No clinical study is included in this submission.

Conclusion l.

The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission, the Sterile Latex Examination Gloves Powder Free are as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K171367.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.