K Number
K171367

Validate with FDA (Live)

Date Cleared
2018-01-26

(262 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sterile Latex Patient Examination gloves, Powder Free, is a disposable device intended for medical purposes that is worn on the examiners' hand or finger to prevent contamination between patient and examiner.

Device Description

The subject device is a patient examination glove made of latex compound. It is sterile, Powder Free. The device is ambidextrous and can be worn on either the left or right hand. The device meets ASTM D3578-05: Standard specification for Rubber Examination Gloves. The subject device is substantially equivalent to legally marketed Latex Examination Powder Free gloves identified as Product code LYY. The device is for over-the counter single use.

AI/ML Overview

The provided document is a 510(k) premarket notification for a medical device: SANCARE STERILE LATEX EXAMINATION GLOVES.

It describes the device's characteristics and compares its performance against a legally marketed predicate device (K052728, Powder free Latex Sterile Patient Examination Gloves from M/S. KANAM LATEX INDUSTRIES PVT. LTD.) to establish substantial equivalence.

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the SANCARE STERILE LATEX EXAMINATION GLOVES are primarily based on established ASTM standards and FDA water leak tests, as compared to the performance of the predicate device. The document explicitly states that the device "Meets" all specified standards.

CharacteristicTest / StandardAcceptance Criteria (from Standard)Reported Device Performance (SANCARE STERILE LATEX EXAM GLOVES)Result
Freedom from PinholesFDA 1000 ml water leak test, ASTM D5151-06 (Re-approved 2011)ISO 2859-1 / G1 / AQL 2.5 (PASS/FAIL)PASSPASS
Dimensions
LengthASTM D3578-05ISO 2859-1 / S2 / AQL 4.0 (min 230 mm)> 230 mmPASS
WidthASTM D3578-05ISO 2859-1 / S2 / AQL 4.0 (70±10 mm to 120±10 mm (sizes XS to XL))70±10 mm to 120±10 mm (sizes XS to XL)PASS
ThicknessASTM D3578-05ISO 2859-1 / S2 / AQL 4.0 (min 0.08 mm (Cuff, palm & finger))> 0.08 mm (Cuff, palm& finger)PASS
Physical Properties
Before aging (Tensile Strength)ASTM D3578-05 / ASTM D412-06ISO 2859-1 / S2 / AQL 4.0 (> 18 Mpa)> 18 MpaPASS
Before aging (Ultimate Elongation)ASTM D3578-05 / ASTM D412-06ISO 2859-1 / S2 / AQL 4.0 (> 650%)>650%PASS
Before aging (Stress at 500% elongation)ASTM D3578-05 / ASTM D412-06ISO 2859-1 / S2 / AQL 4.0 (5.5 Mpa max)5.5 Mpa maxPASS
After Accelerated aging (Tensile Strength)ASTM D3578-05 / ASTM D573-04ISO 2859-1 / S2 / AQL 4.0 (> 14 Mpa)> 14 MpaPASS
After Accelerated aging (Ultimate Elongation)ASTM D3578-05 / ASTM D573-04ISO 2859-1 / S2 / AQL 4.0 (> 500%)> 500%PASS
Powder-free residueASTM D3578-05 / ASTM D6124-06N=5 (Less than 2 mg per glove)Less than 2 mg per glovePASS
Residual ProteinASTM D3578-05 / ASTM D5712-15N=3 (Less than 50µg/dm²)Less than 50µg/dm²PASS
Biocompatibility
Primary skin irritationISO 10993-10Under the conditions of the study, the device is not an irritantNon-irritantPASS
Skin SensitizationISO 10993-10Under the conditions of the study, the device is not a sensitizerNon-sensitizerPASS
SterilityISO 11737-2SterileSterilePASS
Expiration dating/Shelf lifeASTM D7160-05(Not explicitly stated as Pass/Fail but confirmed)Three yearsMeets
SterilizationISO 11135-2014(E)(Not explicitly stated as Pass/Fail but confirmed)Meets (Ethylene Oxide)Meets

2. Sample Sizes Used for Test Set and Data Provenance

Specific sample sizes for each test are provided within the table under "Sampling plan / Inspection level / AQL":

  • Freedom from Pinholes: ISO 2859-1 / G1 / AQL 2.5
  • Dimensions (Length, Width, Thickness): ISO 2859-1 / S2 / AQL 4.0 for each
  • Physical Properties (Before & After aging): ISO 2859-1 / S2 / AQL 4.0 for each
  • Powder-free residue: N=5
  • Residual Protein: N=3

The document does not explicitly state the country of origin of the data or whether the studies were retrospective or prospective. Given that it's a 510(k) submission for a finished medical device (gloves), the performance data would typically be generated by the manufacturer through in-house testing or contracted labs, following international standards (ASTM, ISO).

3. Number of Experts Used to Establish Ground Truth and Qualifications

This information is not applicable to this type of device and submission. This is a submission for examination gloves, which are evaluated based on physical, chemical, and biological performance characteristics against recognized standards, not clinical diagnostic accuracy requiring expert interpretation of images or patient data.

4. Adjudication Method for the Test Set

This information is not applicable. The device's performance is determined by objective laboratory tests against pre-defined thresholds from standards, not through adjudication of expert opinions.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices (e.g., AI algorithms for medical imaging) that involve human readers' interpretation of outputs. For examination gloves, the evaluation criteria are entirely different, focusing on physical properties, safety, and sterility.

6. If a Standalone Study Was Done (Algorithm Only Without Human-in-the-loop Performance)

No, a standalone "algorithm only" study was not done. This concept is specific to AI or software as a medical device (SaMD). The SANCARE STERILE LATEX EXAMINATION GLOVES are a physical, Class I medical device. The performance data presented (e.g., tensile strength, freedom from holes) are inherent characteristics of the gloves themselves, independent of any algorithm.

7. The Type of Ground Truth Used

The "ground truth" for this device's evaluation is based on established international performance standards (ASTM D3578-05, ASTM D5151-06, ASTM D412-06, ASTM D6124-06, ASTM D5712-15, ISO 10993-10, ASTM D7160-05, ISO 11135-2014(E), ISO 11737-2) and FDA-specific tests (FDA 1000ml water leak test). These standards define acceptable ranges or conditions for physical properties, biocompatibility, and sterility.

8. The Sample Size for the Training Set

This information is not applicable. There is no "training set" in the context of examination gloves. This term typically refers to data used to train machine learning models, which are not involved in the manufacturing or evaluation of this device.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo is a stylized representation of a human figure, while the FDA part includes the acronym "FDA" in blue, followed by "U.S. FOOD & DRUG" and "ADMINISTRATION" in a smaller font size.

January 26, 2018

Sanrea Healthcare Products Pvt Ltd Jose Paul M Manager QA & RA Plot#P-56, Pearl Road, Kinfra IIT Park, Kanjikode, Palakkad, 678 621, Kerala, IN

Re: K171367

Trade/Device Name: SANCARE STERILE LATEX EXAMINATION GLOVES Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LYY Dated: December 26, 2017 Received: December 28, 2017

Dear Jose Paul M:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K171367

Device Name SANCARE STERILE LATEX EXAMINATION GLOVES

Indications for Use (Describe)

The Sterile Latex Patient Examination gloves, Powder Free, is a disposable device intended for medical purposes that is worn on the examiners' hand or finger to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K171367

1.0 SUBMITTER

1.1 Company Name:SANREA HEALTHCARE PRODUCTS PVT LTD
1.2 Address:Plot#P-56, Pearl Road, Kinfra Integrated Industrial &Textile Park, Kanjikode, PalakkadKerala, India– 678 621
1.3 Telephone:+ 91-491-2970145
1.4 Email:josepaul@primusgloves.com, sanreaqa@gmail.com,qa@sanrea.com
1.5 Contact Person:Mr. JOSE PAUL MMANAGER QA & RA

2.0 OFFICIAL CORRESPONDENT

2.1 Company Name:SANREA HEALTHCARE PRODUCTS PVT LTD
2.2 Address:Plot#P-56, Pearl Road, Kinfra Integrated Industrial &Textile Park, Kanjikode, PalakkadKerala, India– 678 621
2.3 Telephone:+ 91-491 -2970145
2.4 Email:josepaul@primusgloves.com, sanreaqa@gmail.com,qa@sanrea.com
2.5 Contact Person:Mr. JOSE PAUL MMANAGER QA & RA

3.0 PREPARATION DATE :

{4}------------------------------------------------

4.0 NAME OF THE DEVICE

4.1 Device Name:STERILE LATEX PATIENT EXAMINATIONGLOVES, POWDER FREE
4.2 Trade Name:SANCARE STERILE LATEX EXAMINATION GLOVES (alsomarketed as GLOVTEK STERILE LATEX EXAMINATION GLOVES)
4.3 Common Name:PATIENT EXAMINATION GLOVES
4.4 Classification:PATIENT EXAMINATION GLOVES
4.5 Class:CLASS I
4.6 Product Code:LYY

IDENTIFICATION OF LEGALLY MARKETED PREDICATE DEVICE 5.0

5.1 Device Name:Powder free Latex Sterile Patient Examination Gloves
5.2 510(k) Number:K052728
5.3 Company:M/S. KANAM LATEX INDUSTRIES PVT. LTD.Ooppoottil Buildings, K.K. Road,Kottayam, Kerala, 686001
5.4 Device Description:Powder free Latex Sterile Patient Examination Gloves
5.5 Classification:PATIENT EXAMINATION GLOVES
5.6 Class:CLASS I
5.7 Product Code:LYY
5.8 Classification Panel:General Hospital

{5}------------------------------------------------

6.0 DESCRIPTION OF THE DEVICE

The subject device is a patient examination glove made of latex compound. It is sterile, Powder Free. The device is ambidextrous and can be worn on either the left or right hand. The device meets ASTM D3578-05: Standard specification for Rubber Examination Gloves. The subject device is substantially equivalent to legally marketed Latex Examination Powder Free gloves identified as Product code LYY. The device is for over-the counter single use.

7.0 INDICATIONS FOR USE

The Sterile Latex Patient Examination gloves, Powder Free, is a disposable device intended for medical purposes that is worn on the examiners' hand or finger to prevent contamination between patient and examiner.

8.0 SUMMARY OF PERFORMANCE DATA

There is no difference in technological characteristics compared to the predicate device. Gloves are made from Latex compound, which is Sterile and Powder Free. The gloves have the same technological characteristics compared to ASTM or equivalent standards as given below,

CharacteristicsStandardsPerformance of Latexpatient examinationgloves, Powder Free
Freedom from HolesASTM D3578-05/ ASTM D5151-06Meets
DimensionsASTM D3578-05Meets
Physical PropertiesASTM D3578-05/ ASTM D412-06Meets
Powder-free residueASTM D3578-05/ ASTM D6124-06Meets
Residual ProteinASTM D3578-05/ ASTM D5712-15Meets
Bio-compatibilityPrimary skin irritationISO 10993-10Non-irritant
Skin SensitizationNon-sensitizer

{6}------------------------------------------------

ISO 10993-10
Expiration dating/Shelf lifeASTM D7160-05Three years
SterilizationISO 11135-2014(E)Sterilization of healthcare products -Ethylene Oxide - Requirements forthe development, validation androutine control of a sterilizationprocess for medical devicesMeets
SterilityISO 11737 -2Sterilization of Medical devices-Microbiological methods Part 2:Test of sterility performed in thedefinition, validation andmaintenance of sterilization processSterile

Performance data of gloves based on ASTM D3578-05and FDA 1000ml water leak test

Г

ASTM D3578-05and FDA 1000 ml water leak test
CharacteristicsTestTest standardSamplingplan /Inspectionlevel / AQLSterile, Powder Free,Latex PatientExamination Gloves-Sancare / GlovtekResult
Freedom fromPin holesFDA 1000ml waterleak testASTM D5151 -06(Re-approved 2011)ISO 2859-1 /G1/AQL 2.5PASSPASS
DimensionsLengthASTM D3578-05ISO 2859-1 /S2/AQL 4.0> 230 mmPASS
WidthASTM D3578-05ISO 2859-1 /S2/AQL 4.070±10 mm to 120±10mm(sizes XS to XL)PASS
ASTM D3578-05ISO 2859-1 /> 0.08 mmPASS

{7}------------------------------------------------

ThicknessS2/AQL 4.0(Cuff, palm& finger)
PhysicalpropertiesBeforeagingASTM D3578-05andASTM D412-06ISO 2859-1 /S2/AQL 4.0Tensile strength :> 18 MpaPASS
Ultimate Elongation :>650%PASS
Stress at 500%elongation :5.5 Mpa maxPASS
AfterAcceleratedagingASTM D3578-05andASTM D573-04ISO 2859-1 /S2/AQL 4.0Tensile strength :> 14 MpaPASS
Ultimate Elongation :> 500%PASS
Powder-freeresiduePowder-freeresidueASTM D3578-05andASTM D6124-06N=5Less than 2 mg perglovePASS
Residual ProteinResidualProteinASTM D3578-05 andASTM D 5712 – 15N=3Residual Protein : Lessthan 50µg/dm²PASS
BiocompatibilityPrimary skinirritationISO 10993 -10Under the conditions of the study thedevice is not an irritantPASS
SkinSensitizationISO 10993 -10Under the conditions of the study thedevice is not a sensitizerPASS
SterilitySterilityISO-11737-2SterilePASS

9.0 Predicate Comparison Chart

CharacteristicsPREDICATE – 510(K) :SUBJECT DEVICE :Comparison
K052728
ManufacturerM/S. KANAM LATEXINDUSTRIES PVT. LTD.Ooppoottil Buildings,K.K. Road,Kottayam, Kerala, 686001SANREA HEALTHCAREPRODUCTS PVT LTD. Plot #P-56, Pearl Road, KinfraIntegrated Industrial & TextilePark, Kanjikode, Palakkad,Kerala, India – 678 621-
Product NamePowder free Latex SterilePatient Examination GlovesSterile Latex PatientExamination Gloves, PowderFreeSimilar
Intended UseIntended for medical purposeIntended for medical purposeSimilar
that is worn on the Examinersthat is worn on the
hand or finger to preventExaminers hand or finger to
contamination betweenprevent contamination
patient and examinerbetween patient and
examiner
Indication forA Powder Free PatientThe Sterile Latex PatientSimilar
useExamination Glove is aExamination gloves, Powder
Disposable Device made ofFree is a disposable device
Natural Rubber Latex and isintended for medical
intended to be worn on thepurposes that is worn on the
hands, for medical purposesexaminers hand or finger to
to provide a barrier againstprevent contamination
potentially infectiousbetween patient and
materials and otherexaminer
contaminants.
DescriptionSterile Powder Free, PatientSterile Powder Free, PatientSimilar
Examination gloves made ofExamination gloves made of
natural rubber latex. Thenatural rubber latex. The
gloves are provided in Sizesgloves are provided in Sizes
Extra Small, Small, Medium,Extra Small, Small, Medium,
Large and Extra LargeLarge and Extra Large
PresentationSterile gloves are provided inSterile gloves are provided inIdentical
pouchespouches
MaterialNatural Rubber LatexNatural Rubber LatexIdentical
Product codeLYYLYYIdentical
Non-sterile orSterileSterileIdentical
sterile
Single UseYesYesIdentical
AmbidextrousYesYesIdentical
Meets ASTM D3578Meets ASTMD3578 -Identical
Overall length min 230 mm,
Dimensionswidth varies from 70 mm for
XS size to 120 mm for XL
size, thickness in cuff, finger
and palm has a minimum
0.08 mm
Meets ASTM D 3578andMeets ASTM D 3578 andIdentical
ASTM D412, D573ASTM D412, D573 -
Tensile StrengthTensile strength 18 Mpa min
for before aging and 14 Mpa
min for after aging
Aging done at 70 ±2 deg C
for 166±2 hrs or 100±2deg C
for 22±0.3 hrs
Meets ASTM D 3578Meets ASTM D 3578Identical
Ultimate elongation 650 %
Ultimatemin for before aging and 500
Elongation% min for after aging. Aging
done at 70 ±2 deg C for
166±2 hrs or 100±2 deg C
for 22±0.3 hrs
Stress at 500%Meets ASTM D 3578Meets ASTM D3578 -Identical
ElongationStress at 500% Elongation
for before aging is 5.5Mpa(max)
Freedom fromMeets ASTM D 5151 andMeets ASTM D 5151 andIdentical
pinholesASTM D3578ASTM D3578
Meets ASTM D 6124Identical
Residual PowderMeets ASTM D 6124Less than 2 mg per glove
Residual ProteinMeets ASTM D 5712Meets ASTM D 5712 –Less than 50µg/dm²Identical
BiocompatibilityTestsNon-irritant –Primary Skin Irritation InRabbitsUnder the conditions of thestudy the device is not anirritantIdentical
ISO 10993-10Non-sensitizer –Skin Sensitization in GuineapigsUnder the conditions of thestudy the device is not asensitizerIdentical
SterilityEthylene Oxide sterilizedEthylene Oxide sterilizedIdentical
Labeling* Powder Free* Latex patient exam glove* Sterile* Single use only* Ambidextrous* Manufactured for* Lot No* Intended use* Quantity* Country of origin*Powder Free*Latex Patient exam glove*Sterile*Single use only*Ambidextrous*Manufactured for*Lot No*Intended use*Quantity*Country of originIdentical

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

10.0 CONCLUSION

Based on the intended use, physical properties and technological characteristics, the subject device is as safe, as effective and performs as well as the legally marketed predicate device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.