K Number
K223281

Validate with FDA (Live)

Date Cleared
2022-12-07

(43 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable Nitrile Examination Glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

Device Description

The subject device is powder free nitrile examination gloves. The subject device is blue. It can be available in five specifications: XS, S, M, L and XL. The subject device is non-sterile.

AI/ML Overview

The provided document is a 510(k) Premarket Notification for a Disposable Nitrile Examination Glove, not an AI/ML medical device. Therefore, the information requested regarding acceptance criteria and study designs that are specific to AI/ML device performance (e.g., sample size for test/training sets, expert ground truth establishment, MRMC studies, standalone performance, etc.) is not applicable.

The document focuses on demonstrating substantial equivalence of the new glove to a legally marketed predicate device based on material, performance, and biocompatibility standards.

Here's a breakdown of the relevant information from the document related to performance and acceptance criteria for this non-AI/ML device:


1. Table of Acceptance Criteria and Reported Device Performance

Test MethodPurposeAcceptance CriteriaReported Device Performance
ASTM D6319Physical Dimensions TestLength (mm): XS/S: ≥220; M/L/XL: ≥230; Width (mm): XS: 70±10; S: 80±10; M: 95±10; L: 110±10; XL: 120±10; Thickness (mm): Finger: ≥0.05; Palm: ≥0.05Length (mm): > 230 / Pass; Width (mm): XS: 74-77 / Pass; S: 85-86 / Pass; M: 94-97 / Pass; L: 103-108 / Pass; XL: 114-118 / Pass; Thickness (mm): Finger: 0.08-0.09 / Pass; Palm: 0.06-0.07 / Pass
ASTM D5151Watertightness Test for Detection of HolesMeet the requirements of ASTM D5151 AQL 2.50/125 / Pass (This likely means 0 failures out of 125 samples, meeting AQL 2.5)
ASTM D6124Powder ContentMeet the requirements of ASTM D6124 < 2.0mg0.29-0.33mg / Pass
ASTM D412Physical Properties (Before Aging)Tensile Strength: ≥14MPa; Ultimate Elongation: ≥500%Tensile Strength: 14-30 MPa / Pass; Ultimate Elongation: 509-586 % / Pass
ASTM D412Physical Properties (After Aging)Tensile Strength: ≥14MPa; Ultimate Elongation: ≥400%Tensile Strength: 14-25 MPa / Pass; Ultimate Elongation: 418-533 % / Pass
ISO 10993-5CytotoxicityNon-Toxicity"Under conditions of the study, device extract is cytotoxic." (Note: This might suggest a failure, but often with gloves, extract cytotoxicity is expected and evaluated in context of direct skin contact, which is covered by irritation/sensitization tests. Given the "Pass" results for irritation/sensitization, this might be a nuanced finding that doesn't preclude clearance.)
ISO 10993-11Acute Systemic ToxicityNon-acute systemic toxicity"Under conditions of the study, did not show acute systemic toxicity in vivo. / Pass"
ISO 10993-23IrritationNon-irritating"Under the conditions of the study, not an irritant. / Pass"
ISO 10993-10SensitizationNon-sensitizing"Under conditions of the study, not a sensitizer. / Pass"

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample sizes for each specific test, other than for the Watertightness Test (125 samples). The tests are non-clinical (laboratory-based) and conducted to verify compliance with specified ASTM and ISO standards. The data provenance is implied to be from the manufacturer (Anhui Kindguard Medical Supplies Technology Co., Ltd) in China, as it's part of their 510(k) submission. These are experimental results from a lab, not patient data, so "retrospective or prospective" is not applicable in the typical AI/ML sense.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable. The device is a medical glove, and its performance is assessed against technical specifications and standardized laboratory tests (e.g., dimensions, tensile strength, hole detection, biocompatibility assays), not through expert interpretation of clinical data like medical images.

4. Adjudication Method for the Test Set

Not applicable. Performance is determined by objective measurements and standardized test procedures, not human adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI/ML device.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an AI/ML device.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for this device is defined by the acceptance criteria specified in internationally recognized standards (ASTM D6319, ASTM D5151, ASTM D6124, ASTM D412, ISO 10993 series). These standards provide objective measurement methodologies and thresholds for physical properties and biocompatibility.

8. The Sample Size for the Training Set

Not applicable. This is not an AI/ML device subject to training data.

9. How the Ground Truth for the Training Set was Established

Not applicable. This is not an AI/ML device subject to training data.


In summary, the provided document is for a conventional Class I medical device (disposable examination gloves) and illustrates the data and tests required for a 510(k) clearance based on substantial equivalence. The questions posed are highly specific to AI/ML medical devices, which are not relevant to this submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

December 7, 2022

Anhui Kindguard Medical Supplies Technology Co., Ltd % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.1801,No.161 Lujiazui East Rd.,Pudong Shanghai, Shanghai 200120 China

Re: K223281

Trade/Device Name: Disposable Nitrile Examination Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: October 21, 2022 Received: October 25, 2022

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bifeng Qian -S

Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223281

Device Name Disposable Nitrile Examination Glove

Indications for Use (Describe)

The Disposable Nitrile Examination Glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92. The assigned 510(k) number is K223281.

1.0 Submitter's Information

Name: Anhui Kindguard Medical Supplies Technology Co., Ltd Address: Southwest of Intersection of Tonghe RD & HWY 343, The Economic Dvpt. Zone, Si County, 234300 Suzhou, Anhui, PEOPLE'S REPUBLIC OF CHINA Phone Number: +86-13485097856 Contact: Guo Hua Date of Preparation: Oct.21, 2022

Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 East Lujiazui Rd., Pudong,Shanghai 200120 ,China Tel: +86-21-50313932 Email: Info@truthful.com.cn

2.0 Device Information

Trade name:Disposable Nitrile Examination Glove
Common name:Patient Examination Gloves
Classification name:Non-powdered patient examination glove
Model(s):XS, S, M, L, XL

3.0 Classification

Production code:LZA
Regulation number:21CFR880.6250
Classification:Class I
Panel:General Hospital

4.0 Predicate Device Information

Manufacturer: Ever Global (Vietnam) Enterprise Corp Device: Disposable Powder Free Nitrile Examination Glove, White/ Blue/ Black/ Pink Color 510(k) number: K171422

{4}------------------------------------------------

5.0 Indication for Use

The Disposable Nitrile Examination Glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

6.0 Device Description

The subject device is powder free nitrile examination gloves. The subject device is blue.

It can be available in five specifications: XS, S, M, L and XL. The subject device is non-sterile.

7.0 Technological Characteristic Comparison Table

Table1-General Comparison
ItemSubject Device(K223281)Predicated Device(K171422)Remark
Product CodeLZALZASame
Regulation No.21CFR880.625021CFR880.6250Same
ClassIISame
Intended Use /Indications for UseThe Disposable NitrileExamination Glove is adisposable device intendedfor medical purposes that isworn on the examiner's handsto prevent contaminationbetween patient andexaminer.The Disposable Powder FreeNitrile Examination Glove,White/ Blue/ Black/ Pink Coloris a disposable deviceintended for medicalpurposes that is worn on theexaminer's hands to preventcontamination betweenpatient and examiner.Same
MaterialNitrileNitrileSame
Powdered orPowered freePowdered freePowdered freeSame
Design FeatureAmbidextrousAmbidextrousSame
ColorantBlueWhite/ Blue/ Black/ PinkDifferentAnalysis 1
Labeling InformationSingle-use indication, powderfree, device color, devicename, glove size andquantity, Non-SterileSingle-use indication, powderfree, device color, devicename, glove size andquantity, Non-SterileSame
Dimensions(mm)Length:XS/S:≥220; M/L/XL:≥230;Length:S/M/L/XL:≥230;DifferentAnalysis 2
Width:Width:
XS: 70±10;XS: 75±5;
S: 80±10;S: 85±5;
M: 95±10;M: 95±5;
L: 110±10;L: 105±5;
XL: 120±10XL: 115±5
Thickness(mm)Finger: ≥0.05;Palm: ≥0.05Finger: ≥0.05;Palm: ≥0.05Same
PhysicalPropertiesBeforeAgingTensileStrengthTensileStrengthSame
$14MPa$ , min$14MPa$ , min
UltimateElongationUltimateElongation
500% min500% minSame
AfterAgingTensileStrengthTensileStrengthSame
$14MPa$ , min$14MPa$ , min
UltimateElongationUltimateElongation
400%min400%minSame
Freedom from HolesBe free from holes whentested in accordance withASTMD5151 AQL=2.5Be free from holes whentested in accordance withASTMD5151 AQL=2.5Same
Powder ContentMeet the requirements ofASTM D6124Meet the requirements ofASTM D6124Same
BiocompatibilityISO 10993-10;Under the conditions of thestudy, not a sensitizerISO 10993-23Under the conditions of thestudy, not an irritantISO 10993-5Under conditions of the study,device extract is cytotoxicISO 10993-11;Under thecondition of acutesystemic toxicity test,the test article did notshow acute systemictoxicity in vivoISO 10993-10;Under the conditions of thestudy, not an irritant or asensitizerDifferentAnalysis 3
//
//

Table1-General Comparison

{5}------------------------------------------------

Analysis 1: The subject device (Blue) has different color to the predicate device (White/ Blue/ Black/ Pink), but all proposed devices are conducted the biocompatibility test.

Analysis 2: The physical dimensions are different with that of the predicate, but they all meet the requirements of ASTM D6319.

{6}------------------------------------------------

Analysis 3: The subject device conducted the irritation test under the ISO 10993-23 First edition 2021-01 Biological evaluation of medical devices - Part 23: Tests for irritation, while the predicate device conducted the irritation test under the old edition ISO 10993-10. But under the conditions of the study, both of the devices are not irritant.

8.0 Summary of Non-clinical Testing

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

ISO 10993-5 Third edition 2009-06-01, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity.

ISO 10993-10 Fourth edition 2021-11, Biological evaluation of medical devices - Part 10: Tests for skin sensitization.

ISO 10993-23 First edition 2021-01, Biological evaluation of medical devices - Part 23: Tests for irritation.

ISO 10993-11 Third edition 2017-09, Biological evaluation of medical devices - Part 11: Tests for systemic toxicity.

ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves

ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.

ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.

TestMethodPurposeAcceptance CriteriaResults
ASTMD6319PhysicalDimensionsTestLength (mm):XS/S: ≥220;M/L/XL: ≥230;Width (mm):XS: 70±10;S: 80±10;M: 95±10;L: 110±10;XL: 120±10.Thickness (mm):Finger: ≥0.05Length(mm):> 230/Pass;Width(mm):XS: 74-77 / PassS: 85-86 / PassM: 94-97 / PassL: 103-108 / PassXL:114-118 / PassThickness (mm):Finger:

Table 2 - Summary of non-clinical performance testing

{7}------------------------------------------------

Palm: ≥0.050.08-0.09/Pass
Palm:
0.06-0.07/Pass
ASTMD5151Watertightness Test forDetection ofHolesMeet the requirements of ASTM D5151 AQL 2.50/125/Pass
ASTMD6124PowderContentMeet the requirements of ASTM D6124 < 2.0mg0.29-0.33mg/Pass;
ASTMD412PhysicalpropertiesBefore AgingTensile Strength≥14MPa14-30 MPa/Pass;
Ultimate Elongation≥500%509-586 %/Pass;
After AgingTensile Strength≥14MPa14-25 MPa/Pass;
Ultimate Elongation≥400%418-533 %/Pass;
ISO10993-5CytotoxicityNon-ToxicityUnder conditions of the study, device extract is cytotoxic.
ISO10993-11AcutesystemictoxicityNon-acute systemictoxicityUnder conditions of` the study, did not show acute systemic toxicity in vivo. / Pass
ISO10993-23IrritationNon-irritatingUnder the conditions of the study, not an irritant. / Pass
ISO10993-10SensitizationNon-sensitizingUnder conditions of the study, not a sensitizer. / Pass

9.0 _Summary of Clinical Testing

Clinical testing is not needed for this device.

{8}------------------------------------------------

10.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device Disposable Nitrile Examination Glove is as safe, as effective, and performs as well as or better than the legally marketed predicated device K171422.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.