K Number
K223179

Validate with FDA (Live)

Date Cleared
2023-09-20

(344 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cholestech LDX™ System is a small, portable analyzer and test cassette system is for in vitro diagnostic use only and should not be used for testing in children under the age of 2 years. The Cholestech LDX™ System is comprised of the Cholestech LDX Analyzer and the following cassettes:

The Lipid Profile GLU cassette is for the quantitative determination of total cholesterol, HDL (high-density Ilpoprotein) cholesterol, triglycerides and glucose in whole blood. The TC/HDL (total cholesterol) ratio and estimated values for LDL (low-density lipoprotein) and non-HDL cholesterol are also reported.

The TC+HDL GLU cassette is for the quantitative determination of total cholesterol, HDL (high-density lipoprotein) cholesterol, and glucose in whole blood.

The TC GLU cassette is for the quantitative determination of total cholesterol and glucose in whole blood.

The Lipid Profile cassette is for the quantitative determination of total cholesterol. HDL (high-density lipoprotein) cholesterol, and triglycerides in whole blood. The TC/HDL (total cholesterol) ratio and estimated values for LDL (low-density lipoprotein) and non-HDL cholesterol are also reported.

The TC+HDL cassette is for the quantitative determination of total cholesterol and HDL (high-density lipoprotein) cholesterol in whole blood.

The TC cassette is for the quantitative determination of total cholesterol in whole blood.

· Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.

· HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.

· Triglyceride measurements are used in the diagnosis and treatment with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders.

· Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Device Description

The Cholestech LDX ™ system combines enzymatic methodology and solid-phase technology to measure total cholesterol, HDL cholesterol, triglycerides and glucose. Samples used for testing can be whole blood from a fingerstick (collected in a lithium heparin-coated capillary tube) or venipuncture. The sample is applied to the Cholestech LDX™ cassette®.

The cassette is then placed into the Cholestech LDX™ Analyzer where a unique system on the cassette separates the plasma from the blood cells. A portion of the plasma flows to the right side of the cassette and is transferred to both the total cholesterol and triglyceride reaction pads. Simultaneously, plasma flows to the left side of the cassette where the low- and very low-density lipoproteins (LDL and VLDL) are precipitated with dextran sulfate (50,000 MW) and magnesium acetate precipitating reagent.The filtrate, containing both glucose and HDL cholesterol, is transferred to both the glucose and HDL cholesterol reaction pads.

The Cholestech LDX ™ Analyzer measures total cholesterol and HDL cholesterol by an enzymatic method based on the method formulation of Allain et al, and Roeschlau. Cholesterol esterase hydrolyzes the cholesterol esters in the filtrate or plasma to free cholesterol and the corresponding fatty acid. Cholesterol oxidase, in the presence of oxygen, oxidizes free cholesterol to cholest-4-ene-3-one and hydrogen peroxide. In a reaction catalyzed by horseradish peroxidase, the peroxide reacts with 4-Aminoantipyrine and N-ethyl-N-sulfohydroxypropyl-m-toluidine, sodium sale (TOOS) to form a purple-colored quinoneimine dye proportional to the total cholesterol and HDL cholesterol concentrations of the sample.

The analyzer measures triglycerides by an enzymatic method based on the hydrolysis of triglycerides by lipase to glycerol and free fatty acids. Glycerol, in a reaction catalyzed by glycerol kinase, is converted to glycerol-3-phosphate. In a third reaction, glycerol-3phosphate is oxidized by glycerol phosphate oxidase to dihydroxyacetone phosphate and hydrogen peroxide. The color reaction utilizing horseradish peroxidase is the same as for the total cholesterol and HDL cholesterol. Estimated LDL cholesterol and non-HDL cholesterol and a TC/HDL ratio are calculated using the measured values for TC, HDL, and Triglycerides.

The analyzer measures glucose by an enzymatic method that uses glucose oxidase to catalyze the oxidation of glucose to gluconolactone and hydrogen peroxide. The color reaction utilizing horseradish peroxidase is the same as that for total cholesterol, HDL cholesterol and triglycerides. The resultant color in all the reactions is measured by reflectance photometry.

A brown (magnetic) stripe on each cassette contains the calibration information required for the Cholestech LDX ™ Analyzer to convert the reflectance reading (% R) to the total cholesterol, HDL cholesterol, triglycerides and glucose concentrations.

AI/ML Overview

The provided text is a 510(k) summary for the Cholestech LDX™ System and primarily discusses device modifications and comparison to a predicate device. It certifies that verification studies were performed as required by risk analysis and all acceptance criteria were met. However, it does not provide the specific details of the acceptance criteria or the reported device performance for these studies. It also does not contain information about the sample size, data provenance, number of experts, adjudication methods, MRMC studies, standalone algorithm performance, or how ground truth was established for test and training sets.

Therefore, based solely on the provided text, I cannot fulfill most of the requested information regarding the study that proves the device meets the acceptance criteria. The document states that such studies were done and met acceptance criteria, but omits the specifics.

Here's what can be inferred or stated from the provided text, and what is missing:


Table of Acceptance Criteria and Reported Device Performance

Information Not Available in the Text: The document explicitly states, "Verification studies were performed as required by risk analysis and all acceptance criteria were met." However, it does not list the specific acceptance criteria (e.g., specific accuracy thresholds, precision ranges, etc.) or the detailed reported device performance (e.g., actual measured accuracy, precision values, etc.) from these studies. The modification pertains to updating the performance claim related to conjugated and unconjugated Bilirubin interference. While it mentions that less than 10% interference was seen at specified levels for various substances, this is a general statement from the predicate device's limitations, not a specific acceptance criterion for the current modification or the exact performance data achieved.


2. Sample Size Used for the Test Set and the Data Provenance

Information Not Available in the Text: The document states that "verification studies" were performed, but it does not specify the sample size (e.g., number of patients, number of samples) used for any test set or the provenance of the data (e.g., country of origin, retrospective or prospective nature of the data collection).


3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

Information Not Available in the Text: The document details changes to an in vitro diagnostic (IVD) device for measuring cholesterol, triglycerides, and glucose. For IVD devices, ground truth is typically established by reference laboratory methods, not by human experts interpreting images or clinical cases. Therefore, the concept of "experts" as in radiologists or pathologists establishing ground truth is not applicable here. Even if it were (e.g., for method comparison studies requiring expert clinical correlation), the document does not mention any role for experts in establishing ground truth.


4. Adjudication Method for the Test Set

Information Not Available in the Text: Since the ground truth for an IVD device is generally established using reference methods (as opposed to human interpretation needing adjudication), an adjudication method as typically used in AI studies of imaging (e.g., 2+1, 3+1) is not applicable or described in this document.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

Information Not Applicable/Available in the Text: This is an in vitro diagnostic (IVD) device, not an AI-assisted diagnostic imaging device. Therefore, MRMC studies comparing human readers with and without AI assistance are not relevant to this type of device and are not mentioned in the documentation.


6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

Information Not Applicable/Available in the Text: The Cholestech LDX™ System is a chemical analyzer, not an AI algorithm. Its performance is inherent to the device's enzymatic and solid-phase technology. The concept of "standalone algorithm performance" without human-in-the-loop is not directly applicable in the same way it would be for a software-as-a-medical-device (SaMD) that processes and interprets data for human review. The document describes the device's direct measurement capabilities.


7. The Type of Ground Truth Used

Inferred from Text: For an in vitro diagnostic device measuring analytes (cholesterol, HDL, triglycerides, glucose), the ground truth is typically established by reference laboratory methods (e.g., highly accurate and precise methods run on core laboratory instruments). While the document does not explicitly state "reference laboratory comparison" for ground truth, the context of an IVD device submission, especially one measuring these specific analytes, strongly implies this method.


8. The Sample Size for the Training Set

Information Not Applicable/Available in the Text: This is a chemical analyzer, not a machine learning or AI-based device that requires a "training set" in the computational sense. The device's operation is based on established enzymatic and chemical reactions, not on data-driven learning. Therefore, there is no "training set" in the context of AI/ML.


9. How the Ground Truth for the Training Set Was Established

Information Not Applicable/Available in the Text: As noted above, there is no "training set" for this type of IVD device in the context of AI/ML. The device's calibration and performance are based on chemical principles and validation studies, not on learning from a training dataset.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

September 20, 2023

Alere San Diego, Inc. Nathifa Bradshaw Manager, Regulatory Affairs 9942 Mesa Rim Road San Diego, California 92121

Re: K223179

Trade/Device Name: Cholestech LDX TM System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: CGA, CHH, LBS, JGY, JJE Dated: June 6, 2023 Received: June 6, 2023

Dear Nathifa Bradshaw:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR

{1}------------------------------------------------

  1. for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Paula V. Caposino -S

Paula Caposino, Ph.D. Acting Deputy Division Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K223179

Device Name Cholestech LDX™ System

Indications for Use (Describe)

The Cholestech LDX™ System is a small, portable analyzer and test cassette system is for in vitro diagnostic use only and should not be used for testing in children under the age of 2 years. The Cholestech LDX™ System is comprised of the Cholestech LDX Analyzer and the following cassettes:

The Lipid Profile GLU cassette is for the quantitative determination of total cholesterol, HDL (high-density Ilpoprotein) cholesterol, triglycerides and glucose in whole blood. The TC/HDL (total cholesterol) ratio and estimated values for LDL (low-density lipoprotein) and non-HDL cholesterol are also reported.

The TC+HDL GLU cassette is for the quantitative determination of total cholesterol, HDL (high-density lipoprotein) cholesterol, and glucose in whole blood.

The TC GLU cassette is for the quantitative determination of total cholesterol and glucose in whole blood.

The Lipid Profile cassette is for the quantitative determination of total cholesterol. HDL (high-density lipoprotein) cholesterol, and triglycerides in whole blood. The TC/HDL (total cholesterol) ratio and estimated values for LDL (low-density lipoprotein) and non-HDL cholesterol are also reported.

The TC+HDL cassette is for the quantitative determination of total cholesterol and HDL (high-density lipoprotein) cholesterol in whole blood.

The TC cassette is for the quantitative determination of total cholesterol in whole blood.

· Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.

· HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.

· Triglyceride measurements are used in the diagnosis and treatment with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders.

· Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Abbott, a healthcare company. The logo consists of a stylized letter "A" in blue, positioned above the company name "Abbott" in black. The "A" is designed with rounded edges and a gap in the middle, giving it a modern and abstract appearance. The overall design is clean and professional, reflecting the company's focus on health and innovation.

510(K) SUMMARY 5

The following 510(k) Summary was prepared in accordance with 21 CFR 807.92

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for Abbott, a healthcare company. The logo features a stylized letter 'a' in blue, positioned above the company name 'Abbott' in black, sans-serif font. The 'a' symbol is a rounded, geometric shape with a gap in the upper right corner, giving it a modern and distinctive look. The overall design is clean and professional, reflecting the company's focus on health and innovation.

General Information 5.1

Document #:K223179 & CW220017
Type of 510(k):Dual Submission- 510(k) and CLIA Waiver
Applicant Name:Alere San Diego, Inc.9942 Mesa Rim RdSan Diego, California 92121United States
Establishment Registration:#2027969
Company Contact:Nathifa BradshawRegulatory Affairs ManagerPhone: +1-561-428-2203Email: nathifa.bradshaw@abbott.com
Date Prepared:September 19, 2023

5.2 Device Identification

Device Common NameGlucose and Cholesterol (total) test system
Device Trade NameCholestech LDX™ System

Classification:

Product CodeClassificationRegulation SectionClassification Panel
CGAClass II21 CFR 862.1345Clinical Chemistry
CHHClass I21 CFR 862.1175Clinical Chemistry
JGYClass I21 CFR 862.1705Clinical Chemistry
JJEClass I21 CFR 862.2160Clinical Chemistry
LBSClass I21 CFR 862.1475Clinical Chemistry

Predicate Device:

Cholestech LDX™ System (K120615)

The following are the legally marketed devices covered under the predicate device clearance, K120615, and their prior clearances:

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for Abbott, a healthcare company. The logo features a stylized, lowercase "a" in a light blue color. Below the symbol is the word "Abbott" in a bold, black font. The logo is simple and modern, conveying a sense of professionalism and innovation.

The original Cholestech LDX™ lipid monitoring system was cleared under premarket notification K901900 for use with Cholestech LDX ™ analyzer.

In addition to total cholesterol, the analytes glucose, HDL, triglycerides were cleared under premarket notification K932727.

The Cholestech multianalyte controls and calibration verification materials were originally cleared under premarket notification K102700.

Cholestech LDX™ software update to incorporate a humidity sensor was cleared under premarket notification K120615.

5.3 Device Description

The Cholestech LDX ™ system combines enzymatic methodology and solid-phase technology to measure total cholesterol, HDL cholesterol, triglycerides and glucose. Samples used for testing can be whole blood from a fingerstick (collected in a lithium heparin-coated capillary tube) or venipuncture. The sample is applied to the Cholestech LDX™ cassette®.

The cassette is then placed into the Cholestech LDX™ Analyzer where a unique system on the cassette separates the plasma from the blood cells. A portion of the plasma flows to the right side of the cassette and is transferred to both the total cholesterol and triglyceride reaction pads. Simultaneously, plasma flows to the left side of the cassette where the low- and very low-density lipoproteins (LDL and VLDL) are precipitated with dextran sulfate (50,000 MW) and magnesium acetate precipitating reagent.The filtrate, containing both glucose and HDL cholesterol, is transferred to both the glucose and HDL cholesterol reaction pads.

The Cholestech LDX ™ Analyzer measures total cholesterol and HDL cholesterol by an enzymatic method based on the method formulation of Allain et al, and Roeschlau. Cholesterol esterase hydrolyzes the cholesterol esters in the filtrate or plasma to free cholesterol and the corresponding fatty acid. Cholesterol oxidase, in the presence of oxygen, oxidizes free cholesterol to cholest-4-ene-3-one and hydrogen peroxide. In a reaction catalyzed by horseradish peroxidase, the peroxide reacts with 4-

a In this submission Cholestech LDX™ cassette refers to all six available configurations as indicated in the Package Insert. Section 5: 510(k) Summary

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the logo for Abbott, a healthcare company. The logo features a stylized, rounded letter 'a' in blue, positioned above the company name, "Abbott," which is written in a bold, sans-serif font. The overall design is clean and modern, reflecting a corporate identity.

Aminoantipyrine and N-ethyl-N-sulfohydroxypropyl-m-toluidine, sodium sale (TOOS) to form a purple-colored quinoneimine dye proportional to the total cholesterol and HDL cholesterol concentrations of the sample.

Image /page/7/Figure/3 description: This image shows three chemical reactions. The first reaction shows Cholesterol esters and H2O reacting with Cholesterol esterase to produce Free cholesterol and Fatty acids. The second reaction shows Cholesterol and O2 reacting with Cholesterol oxidase to produce Cholest-4-ene-3-one and H2O2. The third reaction shows 2 H2O2, 4-Aminoantipyrine, and TOOS reacting with Peroxidase to produce Quinoneimine dye and 4 H2O.

The analyzer measures triglycerides by an enzymatic method based on the hydrolysis of triglycerides by lipase to glycerol and free fatty acids. Glycerol, in a reaction catalyzed by glycerol kinase, is converted to glycerol-3-phosphate. In a third reaction, glycerol-3phosphate is oxidized by glycerol phosphate oxidase to dihydroxyacetone phosphate and hydrogen peroxide. The color reaction utilizing horseradish peroxidase is the same as for the total cholesterol and HDL cholesterol. Estimated LDL cholesterol and non-HDL cholesterol and a TC/HDL ratio are calculated using the measured values for TC, HDL, and Triglycerides.

Image /page/7/Figure/5 description: This image shows a series of chemical reactions. In the first reaction, triglycerides and H2O react in the presence of Lipase to produce Glycerol and Free fatty acids. In the second reaction, Glycerol and ATP react in the presence of Glycerol kinase and Mg2+ to produce Glycerol-3-phosphate and ADP. In the third reaction, Glycerol-3-phosphate and O2 react in the presence of Glycerol phosphate Oxidase to produce Dihydroxyacetone phosphate and H2O2. In the fourth reaction, 2 H2O2, 4-Aminoantipyrine, and TOOS react in the presence of Peroxidase to produce Quinoneimine dye and 4 H2O.

The analyzer measures glucose by an enzymatic method that uses glucose oxidase to catalyze the oxidation of glucose to gluconolactone and hydrogen peroxide. The color reaction utilizing horseradish peroxidase is the same as that for total cholesterol, HDL cholesterol and triglycerides. The resultant color in all the reactions is measured by reflectance photometry.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Abbott, a healthcare company. The logo consists of a stylized letter 'a' in a light blue color, positioned above the word "Abbott" in bold, black font. The 'a' symbol is a rounded, geometric shape, and the overall design is clean and modern.

Cholestech LDX™ System Dual Submission- 510(k) and CLIA Waiver K223179 & CW22017

Glucose + O₂Glucose oxidase o-D-gluconolactone + H₂O₂
2 H₂O₂ + 4-Aminoantipyrine + TOOSPeroxidase Quinoneimine dye + 4 H₂O

A brown (magnetic) stripe on each cassette contains the calibration information required for the Cholestech LDX ™ Analyzer to convert the reflectance reading (% R) to the total cholesterol, HDL cholesterol, triglycerides and glucose concentrations.

5.4 Intended Use/Indications for Use

The Cholestech LDX ™ System is a small, portable analyzer and test cassette system. The System is for in vitro diagnostic use only and should not be used for testing in children under the age of 2 years. The Cholestech LDX™ System is comprised of the Cholestech LDX Analyzer and the following cassettes:

The Lipid Profile•GLU cassette is for the quantitative determination of total cholesterol, HDL (high-density lipoprotein) cholesterol, triglycerides and glucose in whole blood. The TC/HDL (total cholesterol/HDL cholesterol) ratio and estimated values for LDL (low-density lipoprotein) and non-HDL cholesterol are also reported.

The TC • HDL • GLU cassette is for the quantitative determination of total cholesterol, HDL (high-density lipoprotein) cholesterol, and glucose in whole blood.

The TC •GLU cassette is for the quantitative determination of total cholesterol and glucose in whole blood.

The Lipid Profile cassette is for the quantitative determination of total cholesterol, HDL (high-density lipoprotein) cholesterol, and triglycerides in whole blood. The TC/HDL (total cholesterol/HDL cholesterol) ratio and estimated values for LDL (low-density lipoprotein) and non-HDL cholesterol are also reported.

The TC • HDL cassette is for the quantitative determination of total cholesterol and HDL (high-density lipoprotein) cholesterol in whole blood.

The TC cassette is for the quantitative determination of total cholesterol in whole blood.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Abbott. The logo consists of a stylized letter 'A' in blue, positioned above the word "Abbott" in black, bold font. The 'A' symbol is a rounded, geometric shape, and the overall design is clean and modern.

  • . Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.
  • . HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
  • Triglyceride measurements are used in the diagnosis and treatment of patients ● with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders.
  • Glucose measurements are used in the diagnosis and treatment of carbohydrate ● metabolism disorders including diabetes mellitus, idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Abbott, a healthcare company. The logo consists of a stylized letter "a" in a light blue color, positioned above the company name "Abbott" in bold, black font. The "a" symbol is a rounded, geometric shape, and the overall design is clean and modern.

AttributePredicate Device – K120615Candidate Device-K223179
Cholestech LDX™ analyzer andCholestech Lipid Profile•GLU CassetteModified Cholestech LDX™ System
Similarities
AssayprincipleThe Cholestech LDX ™ System combinesenzymatic methodology5 and solid-phasetechnology to measure total cholesterol, HDLcholesterol, triglycerides and glucose.Samples used for testing can be whole bloodfrom a fingerstick (collected in a lithiumheparin-coated capillary tube) orvenipuncture. The sample is applied to anCholestech LDX ™ cassette.Same
Sample TypeWhole bloodSame
AnalyzerCholestech LDX™ analyzerSame
User InterfaceUser display and operating instructions inlabelingSame
AttributePredicate Device - K120615Cholestech LDX™ analyzer andCholestech Lipid Profile•GLU CassetteCandidate Device-K223179Modified Cholestech LDX™ System
Warning andPrecautionsFor professional in vitro diagnostic use only.All blood samples, containers, capillary tubesand materials that have come in contact withblood should be handled as if capable oftransmitting infectious disease and discardedinto a biohazardous waste container after use.Same
Differences
Intended UseThe Alere Cholestech LDX™ System is asmall, portable analyzer and test cassettesystem. The System is for in vitrodiagnostic use only. The LipidProfile•GLU Cassette is for thequantitative determination of totalcholesterol, HDL (high-densitylipoprotein) cholesterol, triglycerides andglucose in whole blood. A TC/HDL (totalcholesterol/HDL cholesterol) ratio andestimated values for LDL (low-densitylipoprotein) and non-HDL cholesterol arecalculated by the Alere Cholestech LDX™Analyzer.The CholestechLDX™ System is a small,portable analyzer and test cassette system.The system is for In Vitro diagnostic useonly. The Lipid Profile•GLU Cassette is forthe quantitative determination of totalcholesterol, HDL (high-density lipoprotein)cholesterol, triglycerides and glucose inwhole blood. A TC/HDL (totalcholesterol/HDL cholesterol) ratio andestimated values for LDL (low-densitylipoprotein) and non-HDL cholesterol arecalculated by the Cholestech LDX™analyzer. Cholestech LDX™ system shouldnot be used for testing in children under theage of 2 years. The Cholestech LDX™
AttributePredicate Device - K120615Cholestech LDX™ analyzer andCholestech Lipid Profile•GLU CassetteCandidate Device-K223179Modified Cholestech LDX™ System
• Cholesterol measurements are used inthe diagnosis and treatment of disordersinvolving excess cholesterol in the bloodand lipid and lipoprotein metabolismdisorders.• HDL (lipoprotein) measurements areused in the diagnosis and treatment oflipid disorders (such as diabetesmellitus), atherosclerosis, and variousliver and renal diseases.• Triglyceride measurements are used inthe diagnosis and treatment of patientswith diabetes mellitus, nephrosis, liverobstruction, other diseases involvinglipid metabolism, or various endocrinedisorders.• Glucose measurements are used in thediagnosis and treatment of carbohydratemetabolism disorders including diabetesmellitus, neonatal hypoglycemia, andidiopathic hypoglycemia, and ofpancreatic islet cell carcinoma.System is comprised of the Cholestech LDXAnalyzer and the following cassettes:The Lipid Profile•GLU cassette is for thequantitative determination of totalcholesterol, HDL (high-density lipoprotein)cholesterol, triglycerides and glucose inwhole blood. The TC/HDL (totalcholesterol/HDL cholesterol) ratio andestimated values for LDL (low-densitylipoprotein) and non-HDL cholesterol arealso reported.The TC•HDL•GLU cassette is for thequantitative determination of totalcholesterol, HDL (high-density lipoprotein)cholesterol, and glucose in whole blood.The TC•GLU cassette is for the quantitativedetermination of total cholesterol andglucose in whole blood.The Lipid Profile cassette is for thequantitative determination of totalcholesterol, HDL (high-density lipoprotein)cholesterol, and triglycerides in wholeblood. The TC/HDL (total cholesterol/HDLcholesterol) ratio and estimated values forLDL (low-density lipoprotein) and non-HDLcholesterol are also reported.The TC•HDL cassette is for the quantitativedetermination of total cholesterol and HDL
AttributePredicate Device – K120615Cholestech LDX™ analyzer andCholestech Lipid Profile•GLU CassetteCandidate Device-K223179Modified Cholestech LDX™ System
(high-density lipoprotein) cholesterol inwhole blood.The TC cassette is for the quantitativedetermination of total cholesterol in wholeblood.Cholesterol measurements are usedin the diagnosis and treatment ofdisorders involving excesscholesterol in the blood and lipid anlipoprotein metabolism disorders. HDL (lipoprotein) measurements areused in the diagnosis and treatmentof lipid disorders (such as diabetesmellitus), atherosclerosis, andvarious liver and renal diseases. Triglyceride measurements are usedin the diagnosis and treatment ofpatients with diabetes mellitus,nephrosis, liver obstruction, otherdiseases involving lipid metabolism,or various endocrine disorders. Glucose measurements are used inthe diagnosis and treatment ofcarbohydrate metabolism disordersincluding diabetes mellitus,idiopathic hypoglycemia, and ofpancreatic islet cell carcinoma.
LimitationsSome substances may cause inaccurateresults with enzymatic tests. Thesubstances listed below were tested forinterference with all analytes. Less thanSome substances may cause inaccurateresults with enzymatic tests. The substanceslisted below were tested for interference
AttributePredicate Device – K120615Cholestech LDX™ analyzer andCholestech Lipid Profile•GLU CassetteCandidate Device-K223179Modified Cholestech LDX™ System
10% interference was seen at the levels shown.with all analytes. Less than 10% interference was seen at the levels shown.
Substance Concentration (mg/dL)Substance Concentration (mg/dL)
Ascorbic Acid 1Ascorbic Acid 1 Glutathione 1
Hemoglobin 125Conjugated Bilirubin 2.5 Hemoglobin 1
Bilirubin 5Unconjugated Bilirubin 2.5 Lactose 1
Lactose 100Creatinine 30 Lovastatin (Mevacor) 4
Creatinine 30Cysteine 10 Nicotinic Acid (Niacin) 10
Lovastatin (Mevacor) 4Fructose 30 Urea 500
Cysteine 10Gemfibrozil (Lopid) 15 Uric Acid 15
Nicotinic Acid (Niacin) 10
Fructose 30
Urea 500
Gemfibrozil (Lopid)15
Uric Acid 15
Glutathione 1
• Hematocrits between 30% and 49% do not affect results.• Hematocrits between 30% and 49% do not affect results.
• Blood collection tubes with glycerol should not be used for the triglyceride test.• Blood collection tubes with glycerol should not be used for the triglyceride test.
• Hand creams and soaps with glycerol may cause falsely high triglyceride results.• Hand creams and soaps with glycerol may cause falsely high triglyceride results.
• The triglyceride test measures triglycerides and free glycerol. Free glycerol usually is less than 20 mg/dL.
• There may be a 6–7% difference in the glucose levels of fingerstick and venous blood.
AttributePredicate Device – K120615Cholestech LDX™ analyzer andCholestech Lipid Profile•GLU CassetteCandidate Device-K223179Modified Cholestech LDX™ System
The triglyceride test measures triglycerides and free glycerol.Free glycerol usually isless than 20 mg/dL. There may be a 6–7% difference in the glucose levels of fingerstick and venous blood.

5.5 Comparison with Predicate

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image contains the logo for Abbott, a healthcare company. The logo features a stylized letter 'a' in blue, positioned above the company name "Abbott" in bold, black font. The 'a' symbol is a rounded, geometric shape, and the overall design is clean and modern.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image contains the logo for Abbott, a healthcare company. The logo features a stylized, rounded "a" in a light blue color. Below the symbol is the word "Abbott" in a bold, black font. The logo is simple and modern, reflecting the company's focus on innovation and health.

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for Abbott. The logo consists of a stylized letter 'a' in blue, positioned above the word "Abbott" in black, sans-serif font. The 'a' is designed with rounded corners and a continuous line, giving it a modern and clean appearance. The overall design is simple, recognizable, and corporate.

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image contains the logo for Abbott Laboratories. The logo features a stylized, sans-serif letter 'a' in blue, positioned above the company name, "Abbott," which is written in a bold, sans-serif font. The overall design is clean and modern, reflecting the company's focus on innovation and healthcare.

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image contains the logo for Abbott Laboratories. The logo features a stylized, sans-serif letter 'a' in a bright blue color. Below the symbol, the word "Abbott" is written in a bold, black, sans-serif font. The logo is simple and modern, conveying a sense of professionalism and trustworthiness.

5.6 Technological Characteristics

Cholestech LDX ™ system maintains the same intended use and indications for use. The principle of operations, and chemical composition remains unchanged.

5.7 Description of Device Modification

The modification to the labeling of the Cholestech LDX™ system is to update the performance claim of the device as it pertains to interference level of conjugated and unconjugated Bilirubin. The population of the intended use of the device has been further specified to indicate that the system should not be used for testing in children under the age of 2 years; and for Glucose measurements, to remove use in neonatal hypoglycemia.

5.8 Design Control Activities

The verification studies of the device modification have been performed under design control. The design control activities were based on risk analysis, and acceptance criteria were set to maintain the performance and safety of the Cholestech LDX ™ system. The risk analysis identified the appropriate in-house analytical performance verification studies to demonstrate that the analytical performance and risk of erroneous results for Cholestech LDX ™ system are not adversely affected by using the proposed modification device.

5.9 Conclusion

Verification studies were performed as required by risk analysis and all acceptance criteria were met. The technological characteristics of the predicate device cleared under K120615 are the same as the modified device, and the system maintains the same Section 5: 510(k) Summary Page 12 of 13

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image contains the logo for Abbott Laboratories. The logo features a stylized, rounded letter 'a' in blue, positioned above the company name, "Abbott," which is written in a bold, black font. The logo is simple and modern, conveying a sense of stability and innovation.

intended use and indications for use as the predicate device. In addition, the update to the Bilirubin Interference claim and further specification of the population in the indications for use/intended use did not raise new or different questions of safety and effectiveness; therefore, the Cholestech LDX™ system is found to be substantially equivalent to the predicate device.

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.