K Number
K223078

Validate with FDA (Live)

Date Cleared
2023-06-12

(255 days)

Product Code
Regulation Number
862.1110
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Atellica® CH Diazo Direct Bilirubin (D DBil) assay is for in vitro diagnostic use in the quantitative determination of direct bilirubin in human serum and plasma using the Atellica® CH Analyzer. Measurement of direct bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, is used in the diagnosis and treatment of liver, hemolytic-hematological, and metabolic disorders, including hepatitis and gall bladder block.

Device Description

Atellica® CH Diazo Direct Bilirubin is a Photometric test using 2,4-dichloroaniline (DCA). Direct bilirubin in presence of diazotized 2,4-dichloroaniline forms a red colored azocompound in acidic solution. Absorbance is measured at 545/658 nm.

AI/ML Overview

The provided text describes the performance characteristics and acceptance criteria for the Atellica® CH Diazo Direct Bilirubin (D DBil) assay. Here's a breakdown of the requested information:

Acceptance Criteria and Device Performance

1. Table of Acceptance Criteria and Reported Device Performance:

Performance CharacteristicAcceptance Criteria (Design Goal)Reported Device Performance
Detection CapabilityLoQ ≤ 0.10 mg/dLLoQ = 0.10 mg/dL
Assay ComparisonCorrelation coefficient (r) ≥ 0.950 Slope: 1.00 ± 0.10r = 0.993 Slope y = 0.95x - 0.03 mg/dL (0.95, within 1.00 ± 0.10)
Interferences (HIL)≤ 10% bias from hemoglobin, bilirubin (presumably total bilirubin as an icteric substance), and lipemia. Bias > 10% is considered interference.Hemoglobin: Interference observed above 12.5 mg/dL. Lipemia: No interference ≤ 1000 mg/dL
Non-Interfering SubstancesBias due to these substances ≤ 10%All tested substances showed ≤ 10% bias at specified concentrations.

Note: Specific acceptance criteria for precision and reproducibility are not explicitly listed as single values but are implied by the comprehensive presentation of the data, demonstrating acceptable variability for a diagnostic assay. The document states that the results "support that the Candidate Device... is substantially equivalent."


2. Sample size used for the test set and the data provenance:

  • Assay Comparison: N = 100 samples
  • Specimen Equivalency (Plasma vs. Serum): N = 53 samples for each plasma type (Lithium heparin, Sodium heparin, K2(EDTA)).
  • Precision: N = 80 for each serum level (4 serum levels tested, total 320 measurements).
  • Reproducibility: N = 225 for each serum level (4 serum levels tested, total 900 measurements).
  • Interferences (HIL and Non-interfering Substances): The number of samples for interference testing is not explicitly stated as a single 'N' for the test set. However, the tables indicate specific analyte concentrations tested (e.g., for Hemoglobin, Lipemia, Acetaminophen, etc.), implying multiple measurements were performed for each interference level.

Data Provenance: The document does not explicitly state the country of origin or if the data was retrospective or prospective.


3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not applicable as the device is an in vitro diagnostic assay for quantitative determination of direct bilirubin. The "ground truth" in this context refers to the measured concentration of direct bilirubin, which is established by established laboratory methods, standard reference materials, and comparison to a predicate device, rather than expert interpretation of images or clinical cases.


4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

This information is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in studies involving subjective interpretation (e.g., medical imaging interpretation) where multiple readers assess cases and discrepancies are resolved by a super-reader. For a quantitative diagnostic assay, the "ground truth" is determined by objective measurement rather than expert consensus on subjective findings.


5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This information is not applicable. The device is an in vitro diagnostic assay, not an AI-assisted diagnostic tool that would involve human readers interpreting cases. Therefore, an MRMC study or evaluation of human reader improvement with AI assistance is not relevant to this submission.


6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

The device is a standalone in vitro diagnostic assay. Its performance is measured independently of human interpretation in the clinical setting, although laboratory personnel operate the analyzer and interpret the numerical results in the context of a patient's overall clinical picture. The studies described (Precision, Reproducibility, Assay Comparison, Specimen Equivalency, Interferences) all reflect the standalone performance of the assay on the Atellica CH Analyzer.


7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The "ground truth" for this device's performance studies is established by:

  • Reference Methods/Predicate Device: The "Assay Comparison" section uses the Wako Direct Bilirubin V assay as the comparative method.
  • Internal Reference Standards: The assay's traceability is to internal reference standards manufactured by gravimetric methods.
  • Control Samples/Spiking: For precision, reproducibility, and interference studies, samples are prepared with known concentrations of analyte or interferents.

8. The sample size for the training set:

This information is not provided in the document. This type of detail is typically associated with machine learning or AI algorithm development, which is not the primary focus of this in vitro diagnostic device submission. The device involves a chemical reaction and photometric measurement, not a "training set" in the machine learning sense.


9. How the ground truth for the training set was established:

This information is not provided and is not applicable as the device does not involve a "training set" in the context of machine learning. The assay mechanism is based on a defined chemical reaction (diazo colorimetry) rather than a trained algorithm.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

June 12, 2023

Siemens Healthcare Diagnostics Inc. Anthony Calabro Regulatory Affairs Specialist 500 GBC Drive, M/S 514, PO Box 6101 Newark, Delaware 19714

Re: K223078

Atellica® CH Diazo Direct Bilirubin (D DBil) Trade/Device Name: Regulation Number: 21 CFR 862.1110 Regulation Name: Bilirubin (Total Or Direct) Test System Regulatory Class: Class II Product Code: CIG Dated: February 21, 2023 Received: February 21, 2023

Dear Anthony Calabro:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Paula V. Caposino -S

Paula Caposino, Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

K223078 Device Name

Atellica® CH Diazo Direct Bilirubin (D DBil)

Indications for Use (Describe)

The Atellica® CH Diazo Direct Bilirubin (D DBil) assay is for in vitro diagnostic use in the quantitative determination of direct bilirubin in human serum and plasma using the Atellica® CH Analyzer. Measurement of direct bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, is used in the diagnosis and treatment of liver, hemolytic-hematological, and metabolic disorders, including hepatitis and gall bladder block.

Type of Use (Select one or both, as applicable)

< | Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92 and the Safe Medical Device Act of 1990.

The assigned 510(k) Number is: K223078

1. Date Prepared

June 9, 2023

2. Applicant Information

Contact:Anthony CalabroRegulatory Affairs Specialist
Address:Siemens Healthcare Diagnostics Inc.P.O. Box 6101, M/S 514Newark, DE 19714-1601
Email:anthony.calabro@siemens-healthineers.com

3. Regulatory Information

Atellica® CH Diazo Direct Bilirubin (D_DBil) assay

Trade Name: Atellica® CH Diazo Direct Bilirubin (D DBil) Common Name: Bilirubin (total or direct) test system Classification Name: Diazo Colorimetry, Bilirubin FDA Classification: Class II Review Panel: Chemistry Product Code: CIG Regulation Number: 21 CFR 862.1110

4. Predicate Device Information

Predicate Device Name: Wako Direct Bilirubin V 510(k) Number: K053132

{4}------------------------------------------------

5. Intended Use / Indications For Use

The Atellica® CH Diazo Direct Bilirubin (D_DBil) assay is for in vitro diagnostic use in the quantitative determination of direct bilirubin in human serum and plasma using the Atellica® CH Analyzer. Measurement of direct bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, is used in the diagnosis and treatment of liver, hemolytichematologcial, and metabolic disorders, including hepatitis and gall bladder block.

Special Conditions for Use Statement: For Prescription Use Only

6. Device Description

Atellica® CH Diazo Direct Bilirubin is a Photometric test using 2,4-dichloroaniline (DCA). Direct bilirubin in presence of diazotized 2,4-dichloroaniline forms a red colored azocompound in acidic solution. Absorbance is measured at 545/658 nm.

7. Purpose of Submission

The purpose of this submission is a premarket notification for a new device: Atellica CH Diazo Direct Bilirubin (D DBil) assay

8. Comparison of Candidate Device and Predicate Device

The table below describes the similarities and differences between the Atellica CH Diazo Direct Bilirubin assay (Candidate Device) and the Wako Direct Bilirubin V (Predicate Device).

Substantial equivalence was demonstrated by testing several performance characteristics including intended use/indications for use, specimen types, units of measure, expected values, and measuring interval.

{5}------------------------------------------------

FeatureCandidate DevicePredicate Device
Atellica® CH Diazo Direct Bilirubin (D_DBil)Wako Direct Bilirubin V(K053132)
Intended UseThe Atellica CH Diazo Direct Bilirubin(D_DBil) assay is for in vitro diagnostic usein the quantitative determination of directbilirubin in human serum and plasma usingthe Atellica CH Analyzer. Measurement ofdirect bilirubin, an organic compoundformed during the normal and abnormaldestruction of red blood cells, is used in thediagnosis and treatment of liver,hemolytic-hematologcial, and metabolicdisorders, including hepatitis and gallbladder block.For in vitro diagnostic use in thequantitative determination ofdirect bilirubin in human serum orplasma. Such measurements areuseful in the screening of liverfunction disorders or in thediagnosis of jaundice.
Sample TypeHuman serum and plasma (lithium heparin,sodium heparin, dipotassium EDTA)Human Serum and plasma (Lithiumheparin)
Units of Measuremg/dLmg/dL
Assay Range / MeasuringInterval0.10 mg/dL -12.00 mg/dL0.1-20.0 mg/dL
Expected Values≤ 0.30 mg/dL (5.13 µmol/L).≤ 0.30 mg/dL (5.13 µmol/L).
FeatureCandidate DevicePredicate Device
Atellica® CH Diazo Direct Bilirubin (D_DBil)Wako Direct Bilirubin V (K053132)
Assay PrinciplePhotometric test using 2,4-dichloroaniline (DCA).Direct bilirubin in presence of diazotized 2,4-dichloroaniline forms a red coloredazocompound in acidic solution. Absorbance ismeasured at 545/658 nm.Bilirubin is oxidized by vanadate at aboutpH 3 to produce biliverdin. In the presenceof detergent and vanadate, conjugated(direct) bilirubin is oxidized. This oxidationreaction causes a decrease in the opticaldensity of the yellow color, which isspecific to bilirubin. The decrease inoptical density at 451/545 nm isproportional to the direct bilirubinconcentration in the sample. Theconcentration is measured as an endpointreaction.
TraceabilityTraceable to internal reference standardsmanufactured by gravimetric methodsTraceable to reference method, whichuses reference materials from the NationalInstitute of Standards and Technology(NIST).
FeatureCandidate DevicePredicate Device
Atellica® CH Diazo Direct Bilirubin (D_DBil)Wako Direct Bilirubin V (K053132)
CalibrationSingle level calibrationSingle level calibration
CalibratorsAtellica CH Bilirubin Calibrator (BILI CAL)Siemens Chemistry Calibrator
ReagentsTwo liquid reagents, ready to useReady-for-use liquid reagents
CompositionPack 1:Direct Bilirubin Reagent 1:
Well 1 Reagent 1:23.5mLEDTA-NA2(0.1 mmoll/L); NaCl (150mmol/L); sulfamicacid (100 mmol/L)68 mL in 70-mLcontainersTartrate buffer, pH 2.9(0.1 mol/L)Detergent
Well 2 Reagent 1:23.5mLEDTA-NA2(0.1 mmoll/L); NaCl (150mmol/L); sulfamicacid (100 mmol/L)Direct Bilirubin Reagent 2:25 mL in 70-mL containersPhosphate buffer, pH 7.0(10mmol/L)Sodium metavanadate (4 mmol/L)
Pack 2:
Well 1 Reagent 2:8.8mL2,4-Dichloroaniline (5 mmol/L); HCl (920mmol/L);EDTA-Na2 (0.13mmol/L); Na-Nitrite (0.5 mmol/L)
Well 2 Reagent 2:8.8mL2,4-Dichloroaniline (5 mmol/L); HCl (920mmol/L);EDTA-Na2 (0.13mmol/L); Na-Nitrite (0.5 mmol/L)
InterferencesHemoglobin:Hemoglobin:
No Interference ≤ 12.5 mg/dLNo Interference ≤ 750 mg/dL
Lipemia:Lipemia:
No Interference ≤ 1000 mg/dLNo Interference ≤ 1000mg/dL

{6}------------------------------------------------

{7}------------------------------------------------

9. Standard/Guidance Document References

The following recognized standards from Clinical Laboratory Standards Institute (CLSI) were used as a basis of the study procedures described in this submission:

• Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline—Third Edition. (CLSI EP05-A3).

· Interference Testing in Clinical Chemistry (CLSI EP07).

• Measurement Procedure Comparison and Bias Estimation Using Patient Samples (CLS) EP09-A3).

• Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline-Second Edition (EP17-A2).

  • · Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline (CLSI EP25-A).
  • Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory;
  • Approved Guideline Third Edition (CLSI EP28-A3c).
  • Establishing and Verifying an Extended Measuring Interval Through Specimen Dilution and Spiking (CLSI EP34-ED1)
  • · Metrological Traceability and Its Implementation; CLSI EP32-R)
  • Supplemental tables for Interference Testing in Clinical Chemistry (CLSI EP37-ED1)
  • Evaluation of the Linearity of Quantitative Measurement Procedures -20d Edition (CLSI EP06 ED2)

10. Performance Characteristics for Atellica ® CH Diazo Direct Bilirubin (D DBil)

10.1 Detection Capability

The Limit of Blank (LoB) corresponds to the highest measurement result that is likely to be observed for a blank sample. The assay is designed to have an LoB ≤ Limit of Detection (LoD).

The Limit of Detection (LoD) corresponds to the lowest concentration of direct bilirubin that can be detected with a probability of 95%. The assay is designed to have an LoD ≤ Limit of Quantitation(LoQ).

The Limit of Quantitation (LoQ) corresponds to the lowest concentration of direct bilirubin that met the required analyte level but did not reach 20%CV. The assay is designed to have a LoQ of ≤ 0.10mg/dL.

Detection capability was determined in accordance with CLSI Documents EP17-A2.

{8}------------------------------------------------

The following results were obtained:

Specimen TypeDetection CapabilityResult mg/dL
Serum/PlasmaLoB0.00
LoD0.02
LoQ0.10

10.2 Precision

Precision was determined in accordance with CLSI Document EP05-A3. Samples were assayed on the Atellica CH Analyzer in duplicate in 2 runs per day for 20 days. The following results were obtained.

MeanRepeatabilityWithin-Lab
SpecimenTypeNMean(mg/dL)Mean(µmol/L)SD(mg/dL)SD(µmol/L)CV(%)SD(mg/dL)SD(µmol/L)CV(%)
Serum 1800.376.330.0070.1201.90.0070.1201.9
Serum 2800.9916.930.0040.0680.40.0070.1200.7
Serum 3804.9584.650.0140.2390.30.0370.6330.7
Serum 4809.16156.640.0310.5300.30.0701.1970.8

10.3 Reproducibility

Reproducibility was determined in accordance with CLSI Document EP05-A3. Samples were assayed with 5 replicates per run for 5 days using 3 instruments/sites and 3 reagent lots. The data was analyzed to calculate the following components of precision: repeatability, between-day, between-lot, between-instrument, and reproducibility (total). The following results were obtained.

SpecimenTypeNMeanmg/dLMeanµmol/LRepeatabilityBetween-DayBetween-LOTBetween SYSTEMReproducibility
SDmg/dLSDµmol/L%CVSDmg/dLSDµmol/L%CVSDmg/dLSDµmol/L%CVSDmg/dLSDµmol/L%CVSDmg/dLSDµmol/L%CV
Serum2250.366.160.0060.1031.70.0010.0170.30.0090.1542.50.0020.0340.60.0110.1883.1
Serum2250.9716.590.0050.0860.50.0050.0860.50.0120.2051.20.0080.1370.80.0160.2741.6
Serum2255.0285.840.0170.2910.30.0160.2740.30.1182.0182.40.0510.8721.00.1312.2402.6
Serum2259.37160.230.0270.4620.30.0260.4450.30.2904.9593.10.0721.2310.80.3015.1473.2

{9}------------------------------------------------

10.4 Assay Comparison

The Atellica CH Diazo Direct Bilirubin (D DBil) assay was designed to have correlation coefficient of ≥ 0.950 and a slope of 1.00 ± 0.10 compared to the Wako Direct Billrubin V assay. The following results were obtained.

SpecimenTypeComparisonAssay (x)Regression Equationmg/dL ( $\u03bcmol$ /L)Sample Rangemg/dL ( $\u03bcmol$ /L)Nr
SerumWako DirectBilirubin V$y=0.95x-0.03mg/dL$ $(y=0.95x-.51)\u03bcmol/L$0.10 - 11.10(1.71 - 189.81)1000.993

N – Number of samples

r – Correlation coefficient

10.5 Specimen Equivalency

The specimen equivalency was determined using the Deming regression model in accordance with CLSI Document EP90c. The following results were obtained:

SpecimenTypeReferenceSpecimenRegression Equationmg/dL (µmol/L)Sample Rangemg/dL (µmol/L)Nr
Plasma(Lithiumheparin)Serum$y=1.00x - 0.02$( $y=1.00x - 0.34$ )0.10 - 10.27(1.71 - 175.62)530.999
Plasma(SodiumHeparin)Serum$y=0.98x + 0.00$( $y=0.98 + 0.00$ )0.10 - 10.27(1.71 - 175.62)530.999
PlasmaK2(EDTA)Serum$y=0.98x - 0.02$( $y=0.98x - 0.34$ )0.10 - 10.27(1.71 - 175.62)530.999

{10}------------------------------------------------

10.6 Interferences

10.6.1 Hemolysis, Icterus, and Lipemia (HIL)

The Atellica CH Diazo Direct Bilirubin (D DBil) assay is designed to have ≤ 10% interference from hemoglobin, bilirubin, and lipemia. Bias is the difference in the results between the control sample (does not contain the interferent) and the test sample (contains the interferent) expressed in a percentage. Bias > 10% is considered interference. Analyte results should not be corrected based on this bias.

Interference testing was performed in accordance with CLSI Document EP07. The following results were obtained:

Interferent concentrations in the table below are the highest concentration tested with no interference.

SubstanceSubstance ConcentrationConventional Units (SI Units)Analyte ConcentrationConventional Units (SI Units)Bias
Hemoglobin12.5 mg/dL (0.125 g/L)0.33 mg/dL (5.64 μmol/L)-0.10 mg/dL
Hemoglobin12.5 mg/dL (0.125 g/L)4.84 mg/dL (82.76 μmol/L)-6.2%
Hemoglobin25.0 mg/dL (0.250 g/L)0.31 mg/dL (5.30 μmol/L)-0.07 mg/dL
Hemoglobin25.0 mg/dL (0.250 g/L)4.50 mg/dL (76.95 μmol/L)-8.0%
Hemoglobin37.5 mg/dL (0.375 g/L)0.31 mg/dL (5.30 μmol/L)-0.11 mg/dL
Hemoglobin37.5 mg/dL (0.375 g/L)4.50 mg/dL (76.95 μmol/L)-12.2%
Hemoglobin50.0 mg/dL (0.500 g/L)0.31 mg/dL (5.30 μmol/L)-0.14 mg/dL
Hemoglobin50.0 mg/dL (0.500 g/L)4.50 mg/dL (76.95 μmol/L)-15.1%
Hemoglobin75.0 mg/dL (0.750 g/L)0.31 mg/dL (5.30 μmol/L)-0.17 mg/dL
Hemoglobin75.0 mg/dL (0.750 g/L)4.50 mg/dL (76.95 μmol/L)-22.3%
Hemoglobin112.5 mg/dL (1.125 g/L)0.31 mg/dL (5.30 μmol/L)-0.19 mg/dL
Hemoglobin112.5 mg/dL (1.125 g/L)4.50 mg/dL (76.95 μmol/L)-30.8%
Hemoglobin125.0 mg/dL (1.250 g/L)0.31 mg/dL (5.30 μmol/L)-0.20 mg/dL
Hemoglobin125.0 mg/dL (1.250 g/L)4.50 mg/dL (76.95 μmol/L)-32.8%
Hemoglobin150.0 mg/dL (1.500 g/L)0.31 mg/dL (5.30 μmol/L)-0.20 mg/dL
Hemoglobin150.0 mg/dL (1.500 g/L)4.50 mg/dL (76.95 μmol/L)-36.8%
Hemoglobin250.0 mg/dL (2.500 g/L)0.31 mg/dL (5.30 μmol/L)-0.25 mg/dL
Hemoglobin250.0 mg/dL (2.500 g/L)4.50 mg/dL (76.95 μmol/L)-46.2%
Hemoglobin375.0 mg/dL (3.750 g/L)0.31 mg/dL (5.30 μmol/L)-0.26 mg/dL
Hemoglobin375.0 mg/dL (3.750 g/L)4.50 mg/dL (76.95 μmol/L)-54.0%
Hemoglobin500.0 mg/dL (5.000 g/L)0.31 mg/dL (5.30 μmol/L)-0.26 mg/dL
Hemoglobin500.0 mg/dL (5.000 g/L)4.50 mg/dL (76.95 μmol/L)-59.0%
Lipemia1000 mg/dL (10.00g/L)0.31 mg/dL (3.50 μmol/L)-0.07 mg/dL
Lipemia1000 mg/dL (10.00g/L)4.55 mg/dL (77.81 μmol/L)-4.8%

{11}------------------------------------------------

Do not use hemolyzed samples, as they may cause significant interference with this assay. Hemolyzed samples will give falsely negative results.

An H index above 0 (≥1) indicates hemolysis has been detected and results should not be reported out of the laboratory. It is strongly recommended that the HIL index capability be turned on in order to detect hemolysis in samples.

Assay results obtained at individual laboratories may vary from the data presented.

{12}------------------------------------------------

10.6.2 Non-interfering Substances

The following substances do not interfere with Atellica CH Diazo Direct Bilirubin (D_DBil) assay when present in serum and plasma at the concentrations indicated in the table below. Bias due to these substances is ≤ 10%.

Interference testing was performed in accordance with CLSI Document EP07. The following results were obtained:

InterferentInterferentConcentration(SI)ObservedAnalyte (SI)ObservedBias fromControl
Acetaminophen20 mg/dL0.29 mg/dL
(1323.1 µmol/L)(4.96 µmol/L)-0.01 mg/dL
Acetaminophen20 mg/dL5.23 mg/dL
(1323.1 µmol/L)(89.43 µmol/L)-1.0%
Acetylsalicylic acid100 mg/dL0.29 mg/dL
(5555.6 µmol/L)(4.96 µmol/L)-0.01 mg/dL
Acetylsalicylic acid100 mg/dL5.21 mg/dL
(5555.6 µmol/L)(89.09 µmol/L)0.0%
Albumin6 g/dL0.29 mg/dL
(60 g/L)(4.96 µmol/L)-0.01 mg/dL
Albumin6 g/dL5.48 mg/dL
(60 g/L)(93.71 µmol/L)-4.0%
Ascorbic acid5 mg/dL0.30 mg/dL
(284.1 µmol/L)(5.13 µmol/L)-0.02 mg/dL
Ascorbic acid5 mg/dL5.18 mg/dL
(284.1 µmol/L)(88.58 µmol/L)-1.0%
Carbenicillin3 mg/dL0.29 mg/dL
(79.3 µmol/L)(4.96 µmol/L)0.00 mg/dL
Carbenicillin3 mg/dL5.23 mg/dL0.0%
InterferentInterferentConcentration(SI)ObservedAnalyte (SI)ObservedBias fromControl
(79.3 µmol/L)(89.43 µmol/L)
Cholesterol500 mg/dL(12.9 mmol/L)0.30 mg/dL(5.13 µmol/L)0.00 mg/dL
Cholesterol500 mg/dL(12.9 mmol/L)5.52 mg/dL(94.39 µmol/L)-1.0%
Diazepam20 µg/mL(70.2 µmol/L)0.27 mg/dL(4.62 µmol/L)0.00 mg/dL
Diazepam20 µg/mL(70.2 µmol/L)5.09 mg/dL(87.04 µmol/L)0.0%
Eltrombopag25 µg/mL(56.6 µmol/L)0.28 mg/dL(4.79 µmol/L)-0.01 mg/dL
Eltrombopag25 µg/mL(56.6 µmol/L)5.03 mg/dL(86.01 µmol/L)-1.0%
Ethanol800 mg/dL(173.5 mmol/L)0.28 mg/dL(4.79 µmol/L)0.00 mg/dL
Ethanol800 mg/dL(173.5 mmol/L)5.08 mg/dL(86.87 µmol/L)-1.0%
Ibuprofen50 mg/dL(2427.2 µmol/L)0.29 mg/dL(4.96 µmol/L)-0.01 mg/dL
Ibuprofen50 mg/dL(2427.2 µmol/L)5.18 mg/dL(88.58 µmol/L)-1.0%
IgG5 g/dL(333.3 µmol/L)0.28 mg/dL(4.79 µmol/L)-0.01 mg/dL
IgG5 g/dL(333.3 µmol/L)4.97 mg/dL(84.99 µmol/L)-1.0%
Levodopa300 µg/mL(1522.8 µmol/L)0.27 mg/dL(4.62 µmol/L)0.05mg/dL
InterferentInterferentConcentration(SI)ObservedAnalyte (SI)ObservedBias fromControl
Levodopa300 µg/mL4.85 mg/dL
(1522.8 µmol/L)(82.94 µmol/L)1.0%
Oxytetracycline50 mg/dL0.28 mg/dL
(1085.9 µmol/L)(4.79 µmol/L)0.01 mg/dL
Oxytetracycline50 mg/dL5.19 mg/dL
(1085.9 µmol/L)(88.75 µmol/L)0.0%
Phloroglucinol1500 ng/mL0.27 mg/dL
(11.9 µmol/L)(4.62 µmol/L)0.01 mg/dL
Phloroglucinol1500 ng/mL5.26 mg/dL
(11.9 µmol/L)(89.95 µmol/L)-1.0%
Rheumatoid Factor510 IU/mL0.28 mg/dL
(510 IU/mL)(4.79 µmol/L)0.00 mg/dL
Rheumatoid Factor510 IU/mL5.05 mg/dL
(510 IU/mL)(86.36 µmol/L)0.0%
Rifampicin6 mg/dL0.30 mg/dL
(72.9 µmol/L)(5.13 µmol/L)0.08 mg/dL
Rifampicin6 mg/dL5.24 mg/dL
(72.9 µmol/L)(89.60 µmol/L)0.0%
Total Protein12 g/dL0.30 mg/dL
(120 g/L)(5.13 µmol/L)-0.01 mg/dL
Total Protein12 g/dL5.46 mg/dL
(120 g/L)(93.37 µmol/L)-8.0%

{13}------------------------------------------------

{14}------------------------------------------------

Clinical Study 11.

Not applicable.

{15}------------------------------------------------

Expected Values 11.1

Siemens Healthineers has verified the reference interval for serum and plasma for the Atellica CH Diazo Direct Bilirubin assay, in accordance with CLSI Document EP28-A3c is ≤ 0.30 mg/dL (5.13 µmol/L).

12. Traceability

The Atellica CH D_DBil assay is traceable to internal reference standards manufactured by gravimetric methods

Clinical Cut-off 13.

Not applicable

Conclusion 14.

The results from the performance studies support that the Candidate Device, Atellica CH Diazo Direct Bilirubin (D_DBil) assay is substantially equivalent to the Predicate Device, Wako Direct Bilirubin V (K053132)

§ 862.1110 Bilirubin (total or direct) test system.

(a)
Identification. A bilirubin (total or direct) test system is a device intended to measure the levels of bilirubin (total or direct) in plasma or serum. Measurements of the levels of bilirubin, an organic compound formed during the normal and abnormal distruction of red blood cells, if used in the diagnosis and treatment of liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block.(b)
Classification. Class II.