K Number
K222712

Validate with FDA (Live)

Date Cleared
2023-11-22

(440 days)

Product Code
Regulation Number
870.2900
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Patient Monitoring Cables are intended to be used with SpO2, Temperature, BP, ECG, Cardiac Output, and related monitoring devices. The Patient Monitoring Cables are used to connect electrodes, catheters, and/or sensors placed at appropriate sites on the patient to a monitoring devices, or between monitoring devices, for general monitoring and/or diagnostic evaluation by health care professionals.

Device Description

The Patient Monitoring Cables are replacements for similar accessory cables manufactured or specified by Original Equipment Manufacturers (OEM) for their respective transducers, monitors, and peripheral equipment. These are non-patient-contact, insulated, electrical cables with a connector (plug or receptacle) at each end, designed to transmit electrical power and/or signal (data) between medical devices (e.g., to connect SpO2 sensor, IBP transducer to a patient monitor). They are not intended to connect to the mains (i.e., not mains power cables), do not generate any type of power and/or signal, and have no additional non-electrical conducting or processing functionality.

AI/ML Overview

The provided text describes a 510(k) premarket notification for "Fogg System Patient Monitoring Cables." This document focuses on the substantial equivalence of the cables to a predicate device based on bench testing of physical and electrical properties, not on a clinical "study" involving AI or human-in-the-loop performance. Therefore, many of the questions related to AI device performance evaluation (such as MRMC studies, training set details, and expert ground truth establishment for AI) are not applicable to this submission.

Here's a breakdown of the acceptance criteria and the "study" (nonclinical test data) presented in the document:

1. A table of acceptance criteria and the reported device performance

The document does not present a formal table of acceptance criteria with corresponding performance results. Instead, it states that the Fogg System Patient Monitoring Cables "met all acceptance criteria from the testing and analyses". The testing and analyses performed were:

Acceptance Criteria (Implied)Reported Device Performance
Cable assembly resistance and continuityConformed to standards
Cable material signal integrityConformed to standards
Electromagnetic interference and safety analysisConformed to standards
Conformance to applicable sections of standardsMet IEC 60601 and AAMI/ANSI BP22

2. Sample size used for the test set and the data provenance

  • Sample Size: The document mentions "worst case representative cable assemblies and cable material" were used for performance testing. It does not provide a specific number for the sample size.
  • Data Provenance: The testing was "nonclinical" and performed on the physical devices/materials. The document does not specify a country of origin for the data, but given it's a US FDA submission, it's implied that the testing was conducted in a manner acceptable for US regulatory purposes. The testing is not retrospective or prospective clinical data; it's laboratory bench testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This question is not applicable. The "ground truth" for this type of device (patient monitoring cables) is established by engineering specifications, international standards (IEC 60601, AAMI/ANSI BP22), and established testing methodologies for electrical and material properties, not by human expert review of clinical cases or images.

4. Adjudication method for the test set

This question is not applicable. Adjudication methods like 2+1 or 3+1 are used for establishing ground truth in clinical imaging studies, often involving expert readers. For physical device testing, the "ground truth" is determined by whether the device's electrical and material properties meet established engineering and safety standards. There is no human adjudication process described for this type of nonclinical testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This question is not applicable. This submission is for patient monitoring cables, which are passive accessories, not an AI-powered diagnostic device. No MRMC study was conducted.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable. This submission is for patient monitoring cables, not a standalone AI algorithm.

7. The type of ground truth used

The "ground truth" for this device's performance is based on engineering specifications and adherence to recognized international standards such as IEC 60601 and AAMI/ANSI BP22 for patient transducer and electrode cables. It is a technical ground truth related to electrical continuity, signal integrity, electromagnetic compatibility, and material properties, not clinical outcomes or expert consensus on clinical data.

8. The sample size for the training set

This question is not applicable. There is no AI component requiring a training set for this device.

9. How the ground truth for the training set was established

This question is not applicable. As there is no AI component, no training set or its ground truth was established for this submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 22, 2023

Fogg System Company, Inc. Jared Walkenhorst Novare Medical Consulting LLC. 1765 Dusty Boot Dr Lafayette, Colorado 80026

Re: K222712

Trade/Device Name: Fogg System Patient Monitoring Cables Regulation Number: 21 CFR 870.2900 Regulation Name: Patient Transducer And Electrode Cable (Including Connector) Regulatory Class: Class II Product Code: DSA Dated: September 8, 2022 Received: September 8, 2022

Dear Jared Walkenhorst:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jennifer W. Shih -S

Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K222712

Device Name

Fogg System Patient Monitoring Cables

Indications for Use (Describe)

The Patient Monitoring Cables are intended to be used with SpO2, Temperature, BP, ECG, Cardiac Output, and related monitoring devices. The Patient Monitoring Cables are used to connect electrodes, catheters, and/or sensors placed at appropriate sites on the patient to a monitoring devices, or between monitoring devices, for general monitoring and/or diagnostic evaluation by health care professionals.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY

Submitter's Name:Fogg System Company, Inc.
Submitter's Address:15592 East Batavia Drive
Aurora, CO 80011
United States
Contact Person:Jared Walkenhorst
Novare Medical Consulting
720.215.9244
novaremedllc@gmail.com
Date Summary was Prepared:16 Nov 2023
Trade or Proprietary Name:Fogg System Patient Monitoring Cables
Common or Usual Name:Patient transducer and electrode cable (including connector)
Classification:Class II per 21 CFR §870.2900
Product Code:DSA

DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:

The Patient Monitoring Cables are replacements for similar accessory cables manufactured or specified by Original Equipment Manufacturers (OEM) for their respective transducers, monitors, and peripheral equipment. These are non-patient-contact, insulated, electrical cables with a connector (plug or receptacle) at each end, designed to transmit electrical power and/or signal (data) between medical devices (e.g., to connect SpO2 sensor, IBP transducer to a patient monitor). They are not intended to connect to the mains (i.e., not mains power cables), do not generate any type of power and/or signal, and have no additional non-electrical conducting or processing functionality.

INDICATIONS FOR USE

The Patient Monitoring Cables are intended to be used with SpO2, Temperature, BP, ECG, Cardiac Output, and related monitoring devices. The Patient Monitoring Cables are used to connect electrodes, catheters, and/or sensors placed at appropriate sites on the patient to a monitoring device, or between monitoring devices, for general monitoring and/or diagnostic evaluation by health care professionals.

TECHNOLOGICAL CHARACTERISTICS

The Fogg System Patient Monitoring Cables are identical in intended use and similar in basic shape, material, and performance characteristics to the predicate device. There are no differences in the fundamental scientific technology shared by both the subject and predicate devices.

DescriptionSubject DevicePredicate Device (K203635)
Device NamePatient Monitoring CablesPatient Monitoring Cables
Indicationsfor UseThe Patient Monitoring Cables areintended to be used with SpO2,Temperature, BP, ECG, Cardiac Output,and related monitoring devices. TheThe Patient Monitoring Cables areintended to be used with ECG, EKG, Spo2and BP monitoring devices. The PatientMonitoring Cables are used to connect

Table 1 Substantial Equivalence Table

{4}------------------------------------------------

Patient Monitoring Cables are used toconnect electrodes, catheters, and/orsensors placed at appropriate sites on thepatient to a monitoring device, orbetween monitoring devices, for generalmonitoring and/or diagnostic evaluationby health care professionals.electrodes, catheters, and/or sensorsplaced at appropriate sites on the patientto a monitoring device for generalmonitoring and/or diagnostic evaluationby health care professional.
Design/AppearanceCables with various connectors(please see the table below)Cables with various connectors
LengthVarious specified standard lengths(please see the table below)Various specified standard lengths
MaterialTin copper, PVC, Santoprene,ThermoplasticsTin copper, PA66, PVC, ABS
UsageReusable, non-patient-contactingReusable and disposable, patient-contacting and non-patient-contacting
SterilizationNon-SterileNon-Sterile

NONCLINICAL TEST DATA

Performance testing and analyses were performed on worst case representative cable assemblies and cable material to support substantial equivalence to the predicate device. Testing and analysis included the following:

  • . Cable assembly resistance and continuity testing
  • Cable material signal integrity testing
  • Electromagnetic interference and safety analysis

CONCLUSION FROM NONCLINICAL TEST DATA

The Fogg System Patient Monitoring Cables met all acceptance criteria from the testing and analyses and concluded to be in conformance to applicable sections of standards IEC 60601 and AAMI/ANSI BP22.

SUBSTANTIAL EQUIVALENCE CONCLUSION

The Fogg System Patient Monitoring Cables have intended use and indications for use as the predicate device. The subject devices use the same operating principle, incorporate the same basic design, configurations, and labeling as the predicate devices. Performance testing and analysis indicates conformance to the applicable sections of standards, the same as the predicate devices.

As such, the Fogg System Patient Monitoring Cables have been determined to be as safe and effective as the predicate device and no new or different questions were raised regarding the safety and effectiveness when compared to the predicate device. Therefore, the devices have been found to be substantially equivalent.

Production CodeDescription
7162-0XXXInterface Cable - Connects specified Transducer Simulators to CompatibleMonitors

{5}------------------------------------------------

-តចេច
7204-0XXXInterface Cable - Connects specified Transducers to Compatible Monitors
7206-0XXXInterface Cable - Connects specified Transducer Simulators to Compatible Monitors
7213-0XXXInterface Cable - Connects specified Transducers to Compatible Monitors
7216-0XXXInterface Cable - Connects specified Transducers to Compatible Monitors
7228-0XXXInterface Cable - Connects specified Transducers to Compatible Monitors
7229-0XXXInterface Cable - Connects specified Transducers to Compatible Monitors
7233-0XXXInterface Cable - Connects specified Simulators to Compatible Monitors
7238-0XXXInterface Cable - Connects specified Transducers to Compatible Monitors
7245-0XXXInterface Cable - Connects specified Transducers to Compatible Monitors
7251-0XXXInterface Cable - Connects specified Transducers to Compatible Monitors
7254-0XXXInterface Cable - Connects specified Transducers to Compatible Monitors
7268-0XXXInterface Cable - Connects specified Transducer Control Unit to Compatible Monitors
7279-0XXXInterface Cable - Connects specified Transducers to Compatible Monitors
7286-0XXXInterface Cable - Connects specified Simulators to Compatible Monitors
7287-0XXXInterface Cable - Connects specified Transducers to Compatible Monitors
7295-0XXXInterface Cable - Connects specified Pressure Wire Interfaces to Compatible Monitors
7297-0XXXInterface Cable - Connects specified Simulators to Compatible Monitors
7303-0XXXInterface Cable - Connects specified Transducers to Compatible Monitors
7304-0XXXInterface Cable - Connects specified Transducers to Compatible Monitors
7305-0XXXInterface Cable - Connects specified Transducers to Compatible Monitors
7307-0XXXInterface Cable - Connects specified Transducers to Compatible Monitors
7310-0XXXInterface Cable - Connects specified Transducer / Monitor Systems to Compatible Monitors
7316-0XXXInterface Cable - Connects specified Transducers to Compatible Monitors
7330-0XXXInterface Cable - Connects specified Transducer / Monitor Systems to Compatible Monitors
7339-0XXXInterface Cable - Connects specified Transducers to Compatible Monitors
7340-0XXXInterface Cable - Connects specified Transducer / Monitor Systems to Compatible Monitors
7342-0XXXInterface Cable - Connects specified Transducers to Compatible Monitors
7343-0XXXInterface Cable - Connects specified Simulators to Compatible Monitors
7353-0XXXInterface Cable - Connects specified Transducers to Compatible Monitors
7354-0XXXInterface Cable - Connects specified Transducers to Compatible Monitors
7360-0XXXInterface Cable - Connects specified Transducer / Monitor Systems to Compatible Monitors
7364-0XXXInterface Cable - Connects specified Transducer / Monitor Systems to Compatible Monitors
7367-0XXXInterface Cable - Connects specified Transducer / Monitor Systems to Compatible Monitors
7368-0XXXInterface Cable - Connects specified Transducer / Monitor Systems to Compatible Monitors

{6}------------------------------------------------

ר7
-1-SS
7378-0XXXInterface Cable - Connects specified Transducers to Compatible Monitors
7379-0XXXInterface Cable - Connects specified Transducers to Compatible Monitors
7380-0XXXInterface Cable - Connects specified Transducers to Compatible Monitors
7381-0XXXInterface Cable - Connects specified Transducers to Compatible Monitors
7383-0XXXInterface Cable - Connects specified Transducer / Monitor Systems to Compatible
Monitors
7389-0XXXInterface Cable - Connects specified Transducers to Compatible Monitors
7161-0XXX-0XXXAdapter Cable - Adapts Compatible Transducer with Interface Cable to aCompatible Monitor
7185-0756Analog Cable - Cable for Pressure.
7185-0805Analog Cable - Cable for Pressure.
7185-0809Analog Cable - Cable for Pressure.
7242-0001Analog Cable - Cable for Temperature Probe Simulator.
7242-0002Analog Cable - Cable for Temperature Probe Simulator.
7242-0004Analog Cable - Cable for Temperature Probe Simulator.
7242-0006Analog Cable - Cable for Temperature Probe Simulator.
7242-0009Analog Cable - Cable for Temperature Probe Simulator.
7335-0007Analog Cable - Cable for ECG.
7335-0008Analog Cable - Cable for ECG.
7349-0001Analog Cable - Cable for ECG & Pressure.
7349-0002Analog Cable - Cable for ECG & Pressure.
7349-0003Analog Cable - Cable for ECG & Pressure.
7349-0005Analog Cable - Cable for ECG & Pressure.
7349-0006Analog Cable - Cable for ECG & Pressure.
7349-0008Analog Cable - Cable for ECG & Pressure.
7349-0009Analog Cable - Cable for ECG & Pressure.
7350-0001Analog Cable - Cable for Pressure.
7350-0002Analog Cable - Cable for Pressure.
7356-0001Analog Cable - Cable for ECG & Pressure.
7356-0002Analog Cable - Cable for ECG & Pressure.
7356-0003Analog Cable - Cable for ECG & Pressure.
7356-0004Analog Cable - Cable for Pressure.
7356-0005Analog Cable - Cable for Pressure.
7356-0006Analog Cable - Cable for ECG.
7356-0007Analog Cable - Cable for ECG.
7356-0008Analog Cable - Cable for Pressure.
7356-0014Analog Cable - Cable for Pressure.
7356-0016Analog Cable - Cable for ECG & Pressure.
7358-0012Analog Cable - Cable for Pressure.
7382-0001Analog Cable - Cable for Pressure.
7382-0002Analog Cable - Cable for Pressure.
7382-0003Analog Cable - Cable for Pressure.

{7}------------------------------------------------

K222712 - Traditional 510(k): Fogg System Patient Monitoring Cables
7382-0004Analog Cable - Cable for Pressure.
7382-0005Analog Cable - Cable for Pressure.
7382-0006Analog Cable - Cable for Pressure.
7382-0007Analog Cable - Cable for Pressure.
7382-0008Analog Cable - Cable for Pressure.
7382-0009Analog Cable - Cable for Pressure.
7382-0010Analog Cable - Cable for Pressure.
7390-0002Analog Cable - Cable for Pressure.
7390-0004Analog Cable - Cable for Pressure.

§ 870.2900 Patient transducer and electrode cable (including connector).

(a)
Identification. A patient transducer and electrode cable (including connector) is an electrical conductor used to transmit signals from, or power or excitation signals to, patient-connected electrodes or transducers.(b)
Classification. Class II (performance standards).