K Number
K222498

Validate with FDA (Live)

Date Cleared
2022-10-26

(69 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Medical Examination Glove (Nitrile) is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Device Description

The Medical Examination Gloves (Nitrile) are non-sterile, single use only, disposable examination gloves intended for medical purposes to be worn by examiners to prevent contamination between the patient and the examiner. The gloves are blue color, powder free, nitrile ambidextrous gloves are offered in sizes extra-small, small, medium, large, extra-large packed in a paper box. The gloves are designed and manufactured in accordance with the ASTM D6319-19 standard.

AI/ML Overview

This request asks to describe the acceptance criteria and the study that proves the device meets the acceptance criteria based on the provided text. However, the provided text is an FDA 510(k) clearance letter for Medical Examination Gloves (Nitrile). This type of device is a low-risk product and its clearance documentation does not involve AI-driven diagnostics, clinical trials, or the detailed study methodologies (like blinding, sample sizes for test/training sets, expert adjudication, MRMC studies) that would be present for a complex medical software device.

The information provided pertains to the physical and chemical properties of examination gloves, focusing on:

  • Biocompatibility: Tests for irritation, sensitization, and acute systemic toxicity.
  • Performance Testing (Physical): Dimensions, tensile strength, elongation, freedom from holes, and residual powder.

Therefore, many of the requested elements (like "multi reader multi case (MRMC) comparative effectiveness study," "standalone (i.e. algorithm only without human-in-the-loop performance)," "number of experts used to establish the ground truth," etc.) are not applicable to the documentation for a medical glove.

I will interpret the request based on the actual content related to the medical gloves, explaining the "acceptance criteria" and "study that proves the device meets the acceptance criteria" in that context.


Device: Medical Examination Gloves (Nitrile) (XS, S, M, L, XL)
Product Code: LZA
Regulation Number: 21 CFR 880.6250 - Non-Powdered Patient Examination Glove
Regulatory Class: Class I, reserved

1. Table of Acceptance Criteria and the Reported Device Performance

The "study" in this context refers to non-clinical laboratory testing of the glove's physical, chemical, and biological properties.

Biocompatibility Testing:

TestPurposeAcceptance CriteriaReported Device Performance (Results)
Irritation (ISO 10993-10)To verify that the proposed device extract is non-irritating.Under the conditions of the study, not an irritant.Pass
Sensitization (ISO 10993-10)To verify that the proposed device extract is non-sensitizing.Under the conditions of the study, not a sensitizer.Pass
Acute Systemic Toxicity (ISO 10993-11)To verify that the proposed device extract is non-toxic.Under the conditions of the study, the device is non-toxic.Pass

Physical Performance Testing (against ASTM D6319-19 and other relevant standards):

TestPurposeAcceptance CriteriaReported Device Performance (Results)
Dimensions (Width, Length, Thickness - Palm/Finger)To evaluate the physical dimension of the glove.Width: 70mm min; Length: 220mm min; Palm: 0.05mm min; Finger: 0.05mm min.Pass: 76mm min width, 234mm min length, Palm: 0.08mm min, Finger: 0.09mm min
Physical Properties (Tensile Strength & Elongation)To evaluate the tensile strength and ultimate elongation before and after aging.Tensile Strength: Before Aging $\geq$ 14 MPa, After Aging $\geq$ 14 MPa. Elongation: Before Aging 500% min, After Aging 400% min.Pass: Tensile Strength: Before Aging 24.44 MPa, After Aging 27.21 MPa. Elongation: Before Aging 513%, After Aging 541%.
Freedom from Holes (ASTM D5151-19)To detect holes in the gloves.No leakage at sampling level of G-1, AQL 2.5.Pass: No leakage, 80 of 80 passed of each size.
Residual Powder (ASTM D6124-06)To detect the powder residue in the glove.<2mg per glove.Pass: average 0.84 mg per glove.

2. Sample sizes used for the test set and the data provenance

  • Sample Sizes:
    • Freedom from holes: 80 pieces per size were tested (total of 400 gloves for 5 sizes if all were tested equally). All 80 pieces passed for each size.
    • Other tests (Biocompatibility, Dimensions, Physical Properties, Residual Powder): Specific sample sizes are not explicitly stated in the provided text beyond "80 of 80 passed" for the holes test. However, these are standard laboratory tests typically performed on a representative sample set according to the respective ASTM/ISO standards.
  • Data Provenance: The tests were non-clinical, laboratory-based. The location or "country of origin" of the testing itself is not explicitly stated, but the applicant (Humanwell Healthcare Group Medical Supplies Co., Ltd.) is based in Wuhan, Hubei Province, China. The testing would have been conducted to demonstrate compliance with international standards (ASTM, ISO) accepted by the FDA. The data is retrospective in the sense that it was collected prior to the 510(k) submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. For medical gloves, "ground truth" is established by adherence to internationally recognized consensus standards (ASTM, ISO) for physical, chemical, and biological properties, not by human expert interpretation of images or other subjective assessments. The results are objective measurements from laboratory tests.

4. Adjudication method for the test set

Not applicable. The tests performed are objective, quantitative measurements following defined protocols (e.g., ASTM D6319-19, ASTM D5151-19, ISO 10993 series). There is no "adjudication" in the sense of resolving inter-reader variability as there would be for an AI-powered diagnostic device.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a medical examination glove, not an AI-powered diagnostic tool. MRMC studies are used for evaluating diagnostic performance, typically of imaging devices or AI algorithms.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a medical examination glove, not an AI algorithm.

7. The type of ground truth used

The "ground truth" for this device is based on objective, verifiable measurements against established consensus standards (e.g., ASTM and ISO standards for physical properties, biocompatibility). It is not based on expert consensus, pathology, or outcomes data in the way an AI diagnostic would be.

8. The sample size for the training set

Not applicable. This is a physical medical device, not an AI algorithm requiring a training set.

9. How the ground truth for the training set was established

Not applicable. As this is not an AI algorithm, there is no training set or associated ground truth establishment process.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged in a circular pattern. To the right of the seal, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" below. The logo is simple and professional, reflecting the FDA's role in regulating food and drugs.

October 26, 2022

Humanwell Healthcare Group Medical Supplies Co., Ltd. % Ivy Wang Technical Manager Shanghai Sungo Management Consulting Company Limited 14th F, 1500# Century Avenue Shanghai. 200122 China

Re: K222498

Trade/Device Name: Medical Examination Gloves (Nitrile) (XS, S, M, L, XL) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: August 18, 2022 Received: August 18, 2022

Dear Ivy Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Bifeng Oian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

Submission Number (if known)

K222498

Device Name

Medical Examination Gloves (Nitrile) (XS, S, M, L, XL)

Indications for Use (Describe)

The Medical Examination Glove (Nitrile) is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K222498

510(K) Summary

(As requirement by 21 CFR 807.92)

Date prepared: 2022-10-25

A. Applicant:

Name: Humanwell Healthcare Group Medical Supplies Co., Ltd.

Address: 169 Shendun Yi Lu, East Lake Hi-tech Development Zone, Wuhan, Hubei Province, People's Republic of China Contact: Ms. Aixue Yao

Title: Sales Manager

Tel: +86- 18627726855

Email: 674599374@qq.com

Submission Correspondent: Primary contact: Ms. Ivy Wang Shanghai SUNGO Management Consulting Co., Ltd. Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: haiyu.wang@sungoglobal.com Secondary contact: Mr. Raymond Luo Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: fda.sungo@gmail.com

B. Device:

Trade Name: Medical Examination Gloves (Nitrile) (XS, S, M, L, XL) Common Name: Nitrile Patient Examination Gloves (Powder Free)

Regulatory Information Classification Name: Polymer Patient Examination Glove Classification: Class I Product code: LZA Regulation Number: 21 CFR 880.6250 Review Panel: General Hospital

C. Predicate device:

K171422 Disposable Powder Free Nitrile Examination Glove, White/Blue/ Black/ Pink Color Ever Global (Vietnam) Enterprise Corp

D. Indications for use of the device:

The Medical Examination Glove (Nitrile) is a disposable device intended for medical purposes that is worn on

{4}------------------------------------------------

K222498

the examiner's hand to prevent contamination between patient and examiner.

E. Device Description:

The Medical Examination Gloves (Nitrile) are non-sterile, single use only, disposable examination gloves intended for medical purposes to be worn by examiners to prevent contamination between the patient and the examiner. The gloves are blue color, powder free, nitrile ambidextrous gloves are offered in sizes extra-small, small, medium, large, extra-large packed in a paper box.

The gloves are designed and manufactured in accordance with the ASTM D6319-19 standard.

F. Summary of Technological Characteristics

Table 1 General Comparison of Proposed and Predicate Devices

DeviceProposed DevicePredicate DeviceResult
510K #K222498K171422-
Product NameMedical Examination Gloves (Nitrile)DisposablePowderFreeNitrileExaminationGlove,White/Blue/Black/ Pink Color-
Product CodeLZALZASame
RegulationNumber21 CFR 880.625021 CFR 880.6250Same
IndicationsforuseThe Medical Examination Glove (Nitrile) is adisposable device intended for medicalpurposes that is worn on the examiner's handto prevent contamination between patientand examiner.The Disposable Powder Free NitrileExamination Glove, White/ Blue/ Black/Pink Color is a disposable deviceintended for medical purposes that isworn on the examiner's hands toprevent contamination between patientand examiner.Same
Powder freeYesYesSame
Design featureAmbidextrousAmbidextrousSame
MaterialNitrileNitrileSame
OTC useYesYesSame
SterilityNon-sterileNon-sterileSame
UseSingle useSingle useSame
LabelSingle-use, indication, powder free, devicecolor, device name, glove size and quantity,Nitrile Glove Powder Free Blue, Non-SterileSingle-use, indication, powder free,device color, device name, glove size andquantity, Disposable Powder FreeNitrile Examination Glove, Non-SterileSame

Table 2 Device dimension comparison

Predicatedevice(K171422)DesignationSizeTolerance
XSSMLXL
Length, mm230230230230230min
Width, mm758590105115$\u00b15$

{5}------------------------------------------------

Thickness, mm
Finger0.05 min
Palm0.05 min
ProposeddeviceDesignationSizeTolerance
XSSMLXL
Length, mm220220230230230min
Width, mm708095110120$\pm$ 10
Thickness, mm
Finger0.05 min
Palm0.05 min
ResultSimilar

Analysis: The physical dimensions are different with that of the predicate, but they all meet the requirements of ASTM D6319-19, so the differences do not raise any new safety or performance questions.

ItemProposed devicePredicate device (K171422)Result
ColorantBlueWhite/ Blue/ Black/ PinkSimilar
Physical propertiesBefore aging Tensile strength14MPa, min14MPa, minSame
Ultimate elongation500%, min500%, minSame
After aging Tensile strength14MPa, min14MPa, minSame
Ultimate elongation400%, min400%, minSame
Comply with ASTM D6319Comply with ASTM D6319Same
Freedom from holesBe free from holes when tested in accordance with ASTM D5151 G-1, AQL 2.5Be free from holes when tested in accordance with ASTM D5151 G-1, AQL 2.5Same
Residual PowderMeet the requirements of ASTM D6124Meet the requirements of ASTM D6124Same

Table 3 Performance comparison

Analysis: The subject device is only available in a single color (blue,) however, the predicate is available in multiple colors (white, blue, black, pink). Biocompatibility testing was successfully completed for the subject device, demonstrating that any color differences do not affect the safety of the proposed device.

Table 4 Biocompatibility comparison

ItemProposed devicePredicate device(K171422)Result
MaterialNitrileNitrileSame
BiocompatibilityIrritationISO 10993-10Under theconditions of theComply with ISO10993-10Same

{6}------------------------------------------------

study, not anirritant. Complywith ISO 10993-10.
SensitizationISO 10993-10Under theconditions of thestudy, not asensitizer. Complywith ISO 10993-10.
Acute systemictoxicityISO 10993-11Under the /conditions of thestudy, the device isnon-toxic.Different

Analysis: Toxicity information for the predicate device is not publicly available. This does not raise different safety or performance questions since the subject device has acceptable biocompatibility per the biocompatibility endpoint assessment.

G. Summary of Non-Clinical Testing

A Biocompatibility

The following tests for the subject device were conducted the biocompatibility of Medical Nitrile Examination Gloves:

  • ISO 10993-10: Primary Skin Irritation
  • ISO 10993-10: Dermal Sensitization
  • · ISO 10993-11: Acute Systemic Toxicity
TestPurposeAcceptance CriteriaResults
IrritationISO 10993-10The purpose of the test is to verifythat the proposed device extract isnon-irritating.Under the conditions of thestudy, not an irritant.Pass
SensitizationISO 10993-10The purpose of the test is to verifythat the proposed device extract isnon-sensitizingUnder the conditions of thestudy, not a sensitizer.Pass
Acute systemictoxicityISO 10993-11The purpose of the test is to verifythat the proposed device extract isnon-cytotoxic.Under the conditions of thestudy, the device is non-toxic.Pass

A Performance Testing

Physical performance testing of the proposed device was conducted as per ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application.

To summarize, the performance testing of the subject device was conducted to adequately demonstrate the effectiveness of the device in accordance with the relevant test methods cited below:

  • ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application
  • ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual Powder on Medical Gloves

{7}------------------------------------------------

K222498

Test MethodPurposeAcceptance CriteriaResults
Dimensions (width)(thickness)The purpose of thetest is to evaluatethe physicaldimension of thegloveWidth 70mm minLength 220mm minPalm - 0.05mm min.Finger-0.05mm min.Pass76mm min width234mm min lengthPassPalm - 0.08mm min.Finger-0.09mm min
Physical propertiesThe purpose of thetest is to evaluatethe tensile strengthand ultimateelongation beforeand after agingTensile Strength:Before Aging $\geq$ 14 MPa, min.After Aging $\geq$ 14 MPa, min.Elongation:Before Aging 500%, min.After Aging 400%, min.PassTensile Strength:Before Aging 24.44 MPa, min.After Aging 27.21 MPa, min.Elongation:Before Aging 513%, min.After Aging 541%, min.
Freedom from holesThe purpose of thetest is to detectholes in the glovesNo leakage at sampling level ofG-1, AQL 2.5PassNo leakage, 80 of 80 passed ofeach size
Residual PowderThe purpose of thetest is to detect thepowder residue inthe glove<2mg per glovePassaverage 0.84 mg per glove

• ASTM D5151-19 Standard Test Method for Detection of Holes in Medical Gloves

H. Clinical Test Conclusion

No clinical study is included in this submission.

l. Conclusion

The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission, the Medical Examination Gloves (Nitrile) are as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K171422.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.