K Number
K222451

Validate with FDA (Live)

Device Name
SAVE GBR
Manufacturer
Date Cleared
2023-02-07

(176 days)

Product Code
Regulation Number
872.3630
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

SAVE GBR is a metal device intended for use with a dental implant to stabilize and support of bone graft in dentoalveolar bony defect sites.

Device Description

The SAVE GBR consists of Healing Cap, Cover Cap and Spacer. The SAVE GBR are manufactured by Ti-6A1-4V ELI and used with a dental implant to stabilize and support of bone graft in dento-alveolar bony defect sites.

The subject device is compatible with the OssBuilder membrane cleared in K172354.

The dimension of the device is as below:

No.ProductDiameterLength
1Healing CapØ4.5 and 5.56.0 and 7.0mm
2Cover CapØ4.54.0mm
3SpacerØ4.57.2 and 8.2mm

The healing cap and cover cap have same functions and uses but have different design and name. The SAVE GBR are removed from the patient after such time when sufficient bone regeneration is done.

AI/ML Overview

I am sorry, but the provided text (K222451_Decision_Letter[1].pdf) does NOT contain information about acceptance criteria, device performance results, sample sizes for test and training sets, expert qualifications, ground truth establishment, or any details related to an AI/ML-driven medical device study.

The document is a 510(k) premarket notification letter from the FDA regarding a dental device called "SAVE GBR." This device is a metal component used with dental implants to stabilize and support bone grafts. The core of this FDA submission is to demonstrate substantial equivalence to already legally marketed predicate devices, not to present performance data from an AI/ML model.

Therefore, I cannot fulfill your request for information regarding acceptance criteria and study details for an AI-driven device based on this input. The document focuses on material composition, dimensions, sterilization, shelf life, and intended use as compared to predicate devices, rather than AI performance metrics.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below.

Dentis Co., Ltd. % April Lee Consultant Withus Group Inc 106 Superior Irvine, California 92620

Re: K222451

Trade/Device Name: SAVE GBR Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: January 5, 2023 Received: January 6, 2023

Dear April Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS)

regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Andrew I. Steen -S

Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

Device Name SAVE GBR

Indications for Use (Describe)

SAVE GBR is a metal device intended for use with a dental implant to stabilize and support of bone graft in dentoalveolar bony defect sites.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Submitter

Dentis Co., Ltd. Gyu Ri Kim 99, Seongseoseo-ro, Dalseo-gu Daegu, 42718 Korea Email: kgr1026@dentis.co.kr Tel. +82-53-589-3541 Fax. +82-53-289-7922

Device Information

  • . Trade Name: SAVE GBR
  • Common Name: GBR System
  • Classification Name: Endosseous Dental Abutment ●
  • . Product Code: NHA
  • Panel: Dental
  • Regulation Number: 872.3630
  • Device Class: Class II
  • Date prepared: 02/07/2023 ●

Predicate Devices:

Primary Predicate K172354, Ossbuilder System by Osstem Implant Co., Ltd

Reference Device K171027, Dentis Dental Implant System by Dentis Co., Ltd. K181854, Ossbuilder System by Osstem Implant Co., Ltd K140600, SMARTbuilder by Osstem Implant Co., Ltd. K210080, Dentis s-Clean s-Line Mini by Dentis Co., Ltd.

Indication for Use:

SAVE GBR is a metal device intended for use with a dental implant to stabilize and support of bone graft in dento-alveolar bony defect sites.

Device Description:

The SAVE GBR consists of Healing Cap, Cover Cap and Spacer. The SAVE GBR are manufactured by Ti-6A1-4V ELI and used with a dental implant to stabilize and support of bone graft in dento-alveolar bony defect sites.

The subject device is compatible with the OssBuilder membrane cleared in K172354.

The dimension of the device is as below:

No.ProductDiameterLength
1Healing CapØ4.5 and 5.56.0 and 7.0mm
2Cover CapØ4.54.0mm
3SpacerØ4.57.2 and 8.2mm

Official Correspondent

Withus Group Inc. April Lee 106 Superior, Irvine, CA 92620 USA Email: withus6664@gmail.com Phone: 1-909-274-9971 Fax: 1-909-460-8122

{4}------------------------------------------------

The healing cap and cover cap have same functions and uses but have different design and name. The SAVE GBR are removed from the patient after such time when sufficient bone regeneration is done. The compatible fixture with the subject device is as below.

No.Trade NameProduct NameFixture SizeK number
1s-Clean SQ-SL ImplantSystem Regulars-Clean SQ-SL FixtureØ5.2 x 7.0, 7.5, 9.5,11.5 and 13.5K210132

Summaries of Technology Characteristics

The subject device is substantially equivalent to the current cleared devices. They are substantially equivalent in intended use, material and product spec as diameter and length. Comparison demonstrating Substantial Equivalence follows:

1. Healing Cap
Subject DevicePrimary PredicateReference Device
ApplicantDentis Co., Ltd.Osstem Implant Co., LtdOsstem Implant Co., Ltd
Trade NameSAVE GBROssbuilder SystemOssbuilder System
510(k) No.K222451K172354K181854
Classification NameEndosseous Dental AbutmentEndosseous Dental AbutmentEndosseous Dental Abutment
Product CodeNHAJEY, NHADZL, NHA
ClassClass IIClass IIClass II
DescriptionImage: Subject DeviceImage: Primary PredicateImage: Reference Device
MaterialTi-6Al-4V ELI (ASTM F136)Pure Titanium (ASTM F67)Ti-6Al-4V ELI (ASTM F136)
DiameterØ4.5 and Ø5.5Ø4.0 and Ø5.0Ø4.0 and Ø5.0
Length (mm)6.0 and 7.0mm7.0 and 8.0mm7.0 and 8.0mm
SterileGamma SterilizationGamma SterilizationGamma Sterilization
Shelf Life8 years8 years8 years
Indications For Use/Intended UseSAVE GBR is a metal deviceintended for use with a dentalimplant to stabilize andsupport of bone graft in dento-alveolar bony defect sites.OssBuilder System is a metaldevice intended for use with adental implant to stabilize andsupport of bone graft indento-alveolar bony defectsites.OssBuilder System is a metaldevice intended for use with adental implant to stabilize andsupport of bone graft indento-alveolar bony defectsites.
FeaturesHealing Cap is used withinternal type of anchor.Healing Cap is used withinternal type of anchor.Healing Cap is used withinternal type of anchor.
Principles of operationUsed with non-resorbablemembrane for boneregenerationUsed with non-resorbablemembrane for boneregenerationUsed with non-resorbablemembrane for boneregeneration
SubstantialEquivalenceComparisonThe subject healing cap is substantially equivalent in designs, material, indications,sterilization, shelf life, compatible membrane, and technological characteristics with theprimary device (K172354). The difference between the subject and primary predicate ismaterial. To support the different material, K181854 was added which is Ti-6Al-4V ELI(ASTM F136). Therefore, the subject device is substantial equivalent.
  1. Healing Cap

{5}------------------------------------------------

  1. Cover Cap
Subject DevicePrimary Device
ApplicantDentis Co., Ltd.Osstem Implant Co., Ltd
Trade NameSAVE GBROssbuilder System
510(k) No.K222451K172354
Classification NameEndosseous Dental AbutmentEndosseous Dental Abutment
Product CodeNHAJEY, NHA
ClassClass IIClass II
DescriptionImage: T shaped device
MaterialTi-6Al-4V ELI (ASTM F136)Ti-6Al-4V ELI (ASTM F136)
DiameterØ4.5Ø4.0
Length (mm)4.0mm4.3mm
SterileGamma SterilizationGamma Sterilization
Shelf Life8 years8 years
Indications For Use/Intended UseSAVE GBR is a metal device intended for usewith a dental implant to stabilize and supportof bone graft in dento-alveolar bony defectsites.OssBuilder System is a metal device intendedfor use with a dental implant to stabilize andsupport of bone graft in dento-alveolar bonydefect sites.
FeatureThe Cover Cap is used with internal type ofAnchor.The Cover Cap is used with internal type ofanchor
Principles of operationUsed with non-resorbable membrane for boneregenerationUsed with non-resorbable membrane for boneregeneration
SubstantialEquivalenceComparisonThe subject cover cap is substantially equivalent in designs, material, indications,sterilization, shelf life, compatible membrane, and technological characteristics with theprimary predicate (K172354). The difference between the subject and primary predicate isdimension. The difference of diameter and length is not important to affect the function ofsafety. Therefore, the subject device is substantial equivalent.

3. Spacer

Subject DevicePrimary DeviceReference Device
ApplicantDentis Co., Ltd.Osstem Implant Co., LtdOsstem Implant Co., Ltd
Trade NameSAVE GBROssbuilder SystemOssbuilder System
510(k) No.K222451K172354K140600
Classification NameEndosseous Dental AbutmentEndosseous Dental AbutmentEndosseous Dental Abutment
Product CodeNHAJEY, NHANHA
ClassClass IIClass IIClass II
DescriptionImage: Subject DeviceImage: Primary DeviceImage: Reference Device
MaterialTi-6Al-4V ELI (ASTM F136)Ti-6Al-4V ELI (ASTM F136)Ti-6Al-4V ELI (ASTM F136)
DiameterØ4.5Ø3.1~5.0Ø3.2~5.1

{6}------------------------------------------------

Length (mm)7.2 and 8.2mm4.45~9.57.0~15
SterileGamma SterilizationGamma SterilizationGamma Sterilization
Shelf Life8 years8 years8 years
Indications For Use/Intended UseSAVE GBR is a metal deviceintended for use with a dentalimplant to stabilize andsupport of bone graft indento-alveolar bony defectsites.OssBuilder System is a metaldevice intended for use with adental implant to stabilize andsupport of bone graft in dento-alveolar bony defect sites.SMART Builder System is ametal device intended for usewith a dental implant tostabilize and support of bonegraft in dento-alveolar bonydefect sites.
FeatureThe Spacer is external type ofanchorThe OB Anchor is internal typeof anchorThe SB Anchor is externaltype of anchor
Principles of operationUsed with non-resorbablemembrane for boneregenerationUsed with non-resorbablemembrane for boneregenerationUsed with non-resorbablemembrane (K120951) forbone regeneration
SubstantialEquivalenceComparisonThe subject spacer is substantially equivalent in general design, dimensions, material,indications, sterilization, shelf life, compatible membrane, and technological characteristicswith the primary predicate (K172354). The difference between the subject and primarypredicate is design, especially head screw shape. To support this discrepancy, K140600 wasadded, which has the similar shape of the device. Therefore, the subject device is substantialequivalent.

Non-Clinical Testing

Non-clinical testing data are submitted, referenced, or relied upon to demonstrate substantial equivalence.

Sterilization and Shelf Life:

The sterilization and shelf life test were performed on predicate healing abutment, K171027 and can be leveraged for subject device since both devices have the same material, sterilization method, packaging method, and manufacturing process as the subject device according to ISO11137-1, ISO11137-2 and ISO11737-3.

Bacterial Endotoxin Test according to ANSI/AAMI ST72:2011, USP <161>, and USP <85> for subject device was referenced in K210080.

Since endotoxin testing cannot be conducted on every batch in our company, alternative to batch testing was chosen with the sampling plan used for the in-process testing and/or finished product release, as recommended in the FDA guidance, Pyrogen and Endotoxins Testing: Questions and Answers".

Reverse Engineering

The pores of the Osstem non-resorbable membranes (K172354) which attach to the abutment were measured and compared to the diameters of the subject device in order to demonstrate compatibility.

Biocompatibility Evaluation:

Biocompatibility tests according to ISO 10993-1 performed on the abutments of K210080 can be leveraged for the subject device because both devices have same material and manufacturing process.

{7}------------------------------------------------

MR Environment Condition:

Non-clinical worst-case MRI review was performed to evaluate the GBR System in the MRI environment using scientific rationale and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan. " Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795), based on the entire system including all variations (all compatible implant bodies, dental abutments, and fixation screws) and material composition. Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque

Conclusion

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification, Dentis Co., Ltd. concludes that the SAVE GBR is substantially equivalent to the predicate devices as herein.

§ 872.3630 Endosseous dental implant abutment.

(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)