K Number
K222449

Validate with FDA (Live)

Date Cleared
2022-08-27

(12 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Gloves have been tested for use with chemotherapy drugs and Fentanyl Citrate using ASTM D6978-05(2019)

Device Description

Nitrile Powder Free Examination Gloves, Tested For Use With Chemotherapy Drugs And Fentanyl Citrate (Dark Blue) Are Class I Patient Examination Gloves and Specialty Chemotherapy Gloves. They are ambidextrous and come in different sizes - Extra Small, Medium, Large, Extra Large and XXL. Gloves meet the specification of ASTM D6319-19 and have been tested for resistance to permeation by chemotherapy drugs and Fentanyl Citrate as per ASTM D6978-05(2019). The gloves are single use, disposable, and provided non-sterile.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Nitrile Powder Free Examination Gloves, based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

MethodologyTest PerformedAcceptance CriteriaReported Device Performance
ASTM D6319-19Physical Dimensions LengthMinimum 230mm for all sizesPass
ASTM D6319-19Physical Dimensions Palm Width (XS, S, M, L, XL, XXL)XS: 70±10mm, S: 80±10mm, M: 95±10mm, L:110±10mm, XL: 120±10mm, XXL: 130±10mmPass
ASTM D6319-19Physical Dimensions Thickness (Finger, Palm)Finger: 0.05mm (min), Palm: 0.05mm (min)Pass
ASTM D6319-19, ASTM D412-16(2021)Physical Properties (Tensile Strength, Elongation)Tensile Strength (Min 14 MPa) and Elongation (Before Aging 500% and after aging 400%) Min AQL 2.5 (ISO 2859-1)Pass
ASTM D6319-19, ASTM D5151-19Water leak testAQL 2.5 (ISO 2859-1)Pass
ASTM D6319-19, ASTM D6124-06 (2017)Powder ResidueMax 2mg/glovePass
ASTM D6978-05 (2019)Permeation by Chemotherapy DrugsRefer to the table below for specific drug BDTsPass (Based on individual BDTs)
ISO 10993-10:2010Irritation and Skin SensitizationNo Skin sensitization and Skin irritationIs non-sensitization and Non-irritation
ISO 10993-5:2009CytotoxicityNo Cytotoxicity reactivityshowed potential toxicity to L929 cells. (Note: Cytotoxicity concern was addressed by acute systematic toxicity testing to still meet acceptance)
ISO 10993-11:2017Acute systemic toxicity studySubject showed no adverse biological reactionno evidence of systemic toxicity

Chemotherapy Drug and Fentanyl Citrate Permeation (Breakthrough Detection Time - BDT)

Chemotherapy DrugMinimum Breakthrough Detection Time (BDT) in Minutes (Acceptance Criteria)Reported Device Performance (BDT in Minutes)
Carmustine 3.3 mg/ml (3,300 ppm)Explicit criteria not listed, but compared to predicate and results are 11.2 min, with a caution against use.11.2
Cisplatin 1mg/ml (1,000 ppm)>240 (Implicitly, as predicate is >240 and proposed is same)>240
Cyclophosphamide 20mg/ml (20,000 ppm)>240 (Implicitly)>240
Dacarbazine 10 mg/ml (10,000 ppm)>240 (Implicitly)>240
Doxorubicin HCL, 2 mg/ml (2,000 ppm)>240 (Implicitly)>240
Etoposide, 20 mg/ml (20,000 ppm)>240 (Implicitly)>240
Fluorouracil, 50mg/ml (50,000ppm)>240 (Implicitly)>240
Methotrexate, 25mg/ml (25,000ppm)>240 (Implicitly)>240
Paclitaxel, 6mg/ml (6,000ppm)>240 (Implicitly)>240
Thiotepa, 10mg/ml (10,000ppm)Explicit criteria not listed, but compared to predicate and results are 29.4 min, with a caution against use.29.4
Fentanyl Citrate Injection (100 mcg/2ml)>240 (Implicitly)>240

Note on Permeation Acceptance: For chemotherapy drugs where the predicate device showed ">240 minutes", the proposed device also achieving ">240 minutes" indicates it meets or exceeds the previous performance. For Carmustine and Thiotepa, while the BDT is lower, the acceptance is ultimately related to the explicit warning not to use them with these drugs, indicating that the observed breakthrough times are treated as insufficient for safe handling.

2. Sample size used for the test set and the data provenance

The document does not specify the exact sample sizes used for each individual test against the acceptance criteria (e.g., number of gloves tested for each physical dimension, water leak, or permeation). However, it implies standardized testing according to relevant ASTM and ISO standards. The data provenance is industrial testing, most likely performed in a lab setting, to comply with the specified ASTM and ISO standards.

  • Data Provenance: The device manufacturer, Better Care Plastic Technology Co., Ltd. (located in Shenze County, Hebei, China), would have overseen these tests, either internally or through a qualified testing laboratory. The data is thus of industrial origin, specifically from the manufacturer's testing or a third-party lab they used.
  • Retrospective/Prospective: These are likely prospective tests conducted specifically for this 510(k) submission to demonstrate compliance with the standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable to this type of device and study. The "ground truth" for examination gloves is established by objective, standardized laboratory tests (e.g., measuring physical dimensions, tensile strength, breakthrough detection time for chemicals, biological reactivity as per ISO standards). It does not involve human expert consensus on interpretations of images or clinical outcomes.

4. Adjudication method for the test set

This is not applicable. Adjudication methods like "2+1" or "3+1" are typically used in clinical studies or image-based diagnostic device evaluations where expert opinions need to be reconciled. For the physical and chemical testing of gloves, the results are quantitative measurements or direct observations (e.g., pass/fail for a water leak test) and do not require expert adjudication in the same sense.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. An MRMC study assesses the performance of human readers (e.g., radiologists, pathologists) in interpreting medical data, often with and without AI assistance, across a range of cases. This device (examination gloves) is a physical product and does not involve human readers or AI assistance in its primary function, nor in the testing performed for its clearance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable. This question pertains to AI/algorithm performance. The device is a physical product, not an algorithm.

7. The type of ground truth used

The "ground truth" for the performance of these gloves is defined by:

  • Standardized Test Methods and Specifications: Specific values and criteria (e.g., minimum tensile strength, maximum powder residue, minimum breakthrough detection time) defined by national and international standards like ASTM D6319-19, ASTM D6978-05, ISO 10993 series.
  • Direct Measurement and Observation: The device's physical properties and chemical resistance are directly measured and observed in laboratory settings according to these established protocols.
  • Biological Reactivity Endpoints: For biocompatibility tests (irritation, sensitization, cytotoxicity, systemic toxicity), the "ground truth" is determined by established biological endpoints and criteria within the ISO 10993 standards.

8. The sample size for the training set

This is not applicable. This question refers to the training data for an AI algorithm. This device is a physical medical product, and no AI algorithm development is described or relevant to its clearance.

9. How the ground truth for the training set was established

This is not applicable for the reasons stated in point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

August 27, 2022

Better Care Plastic Technology Co., Ltd. % Kathy Liu Project Manager Hongray USA Medical Products Inc. 3973 Schaefer Avenue Chino, California 91710

Re: K222449

Trade/Device Name: Nitrile Powder Free Examination Gloves, Tested For Use With Chemotherapy Drugs And Fentanyl Citrate (Dark Blue) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC, ODO Dated: August 10, 2022 Received: August 15, 2022

Dear Kathy Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222449

Device Name

Nitrile Powder Free Examination Gloves, Tested For Use With Chemotherapy Drugs And Fentanyl Citrate (Dark Blue)

Indications for Use (Describe)

The glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Gloves have been tested for use with chemotherapy drugs and Fentanyl Citrate using ASTM D6978-05(2019)

Chemotherapy DrugMinimum Breakthrough Detection Time (BDT) in Minutes
Carmustine 3.3 mg/ml (3,300 ppm)11.2
Cisplatin 1mg/ml (1,000 ppm)>240
Cyclophosphamide 20mg/ml (20,000 ppm)>240
Dacarbazine 10 mg/ml (10,000 ppm)>240
Doxorubicin HCL, 2 mg/ml (2,000 ppm)>240
Etoposide, 20 mg/ml (20,000 ppm)>240
Fluorouracil, 50mg/ml (50,000ppm)>240
Methotrexate, 25mg/ml (25,000ppm)>240
Paclitaxel, 6mg/ml (6,000ppm)>240
Thiotepa, 10mg/ml (10,000ppm)29.4
Fentanyl Citrate Injection (100 mcg/2ml)>240
Fentanyl Citrate Injection (100 mcg/2ml)>240

Please note that the following drugs have extremely low permeation times: Carmustine: 11.2 minutes, Thio Tepa: 21.6 minutes * Warning: Do not use with Carmustine and Thio Tepa.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Fuqian Xi Road, West district of Shenze, Industrial Base, Shenze County, Hebei, 050000, China

510K Summary

The assigned 510(K) numbers: _K222449

Date Prepared: August 10, 2022

1. Owner's Identification:

Better Care Plastic Technology Co., Ltd. Fuqian Xi Road, West district of Shenze, Industrial Base, Shenze County, Hebei, 050000, China Tel:86-311-66179668 Contact: Ms. Kathy Liu, Project Manager Address: 3973 Schaefer Avenue, Chino, CA 91710, USA Tel:909-590-1611 Email: kathyliu@hongrayusa.com or fdareg@hongray.com.cn

2. Name of the Device:

Trade / Product Name: Nitrile Powder Free Examination Gloves, Tested For Use With Chemotherapy Drugs And Fentanyl Citrate (Dark Blue) Common Name: Exam Gloves Classification Name: Patient Examination Glove Specialty Classification Regulation: 21 CFR 880.6250 Product Code: LZA, LZC, QDO Classification Panel: General Hospital Device Class: Class I

3. Predicate Device Information:

Comfort Rubber Gloves Industries Sdn. Bhd. Blue Colored, Power Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (K192954)

4. Device Description:

Nitrile Powder Free Examination Gloves, Tested For Use With Chemotherapy Drugs And Fentanyl Citrate (Dark Blue) Are Class I Patient Examination Gloves and Specialty Chemotherapy Gloves. They are ambidextrous and come in different sizes - Extra Small, Medium, Large, Extra Large and XXL.

Gloves meet the specification of ASTM D6319-19 and have been tested for resistance to permeation by chemotherapy drugs and Fentanyl Citrate as per ASTM D6978-05(2019). The gloves are single use, disposable, and provided non-sterile.

5. Indications for Use:

The Glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Gloves have been tested for use with chemotherapy drugs and Fentanyl Citrate using ASTM D6978-05(2019)

{4}------------------------------------------------

Fuqian Xi Road, West district of Shenze, Industrial Base, Shenze County, Hebei, 050000, China

510K Summary

The following chemicals have been tested with these gloves:

Chemotherapy DrugMinimum Breakthrough Detection Time (BDT)in Minutes
Carmustine 3.3 mg/ml (3,300 ppm)11.2
Cisplatin 1mg/ml (1,000 ppm)>240
Cyclophosphamide 20mg/ml (20,000 ppm)>240
Dacarbazine 10 mg/ml (10,000 ppm)>240
Doxorubicin HCL, 2 mg/ml (2,000 ppm)>240
Etoposide, 20 mg/ml (20,000 ppm)>240
Fluorouracil, 50mg/ml (50,000ppm)>240
Methotrexate, 25mg/ml (25,000ppm)>240
Paclitaxel, 6mg/ml (6,000ppm)>240
Thiotepa, 10mg/ml (10,000ppm)29.4
Fentanyl Citrate Injection (100 mcg/2ml)>240
  • Please note that the following drugs have extremely low permeation times:

Carmustine: 11.2 minutes, Thiotepa: 29.4 minutes

  • Warning: Do not use with Carmustine and Thiotepa.

6. Comparison of Subject Device and Predicate Device:

General Comparison Table:

Proposed DeviceK222449Predicate DeviceK192954Comparison
Trade NameNitrile Powder FreeExamination Gloves, TestedFor Use With ChemotherapyDrugs And Fentanyl Citrate(Dark Blue)Blue Colored, Power Free NitrileExamination Gloves Tested forUse with Chemotherapy Drugsand Fentanyl CitrateSimilar
Product CodeLZA, LZC, QDOLZA, LZC,QDOSame
Regulation Number21 CFR 880.625021 CFR 880.6250Same
ClassIISame

{5}------------------------------------------------

Fuqian Xi Road, West district of Shenze, Industrial Base, Shenze County, Hebei, 050000, China

510K Summary

Indications for UseNitrile Powder FreeExamination Gloves, TestedFor Use With ChemotherapyDrugs And Fentanyl Citrate(Dark Blue) is a disposabledevice intended for medicalpurposes that is worn on theexaminer's hand to preventcontamination between patientand examiner. Gloves havebeen tested for use withchemotherapy drugs andFentanyl Citrate using ASTMD6978Blue Colored, Powder Free NitrileExamination Gloves, Non-sterile,and Tested for Use withChemotherapy Drugs and FentanylCitrate is a patient medical examglove which is a disposable deviceintended for medical purpose thatis worn on the examiner's hand orfinger to prevent contaminationbetween examiner and patient.Glove was tested for use withChemotherapy Drugs and FentanylCitrate as per ASTM D6978Similar
MaterialNitrileNitrileSame
Powder or Powder FreePowder FreePowder FreeSame
ColorDark BlueBlueSimilar
Single useSingle useSingle useSame
Chemotherapy Drugsand Fentanyl CitrateClaimSee below comparison tableSee below comparison tableSee belowcomparisontable

Technological Characteristic Comparison Table:

TechnologicalCharacteristicsProposed DevicePredicate DeviceK192954Comparison
LengthMinimum 230mmMinimum 240mmSimilar
Palm Width (size) (mm)
XS70±1070±10Same
S80±1080±10Same
M95±1095±10Same
L110±10110±10Same
XL120±10120±10Same
XXL130±10N/ADifferent
Thickness(mm)
FingerMinimum 0.05Minimum 0.05Same
PalmMinimum 0.05Minimum 0.05Same
Tensile Strength, BeforeAging14MPa, min14MPa, minSame

{6}------------------------------------------------

Fuqian Xi Road, West district of Shenze, Industrial Base, Shenze County, Hebei, 050000, China

510K Summary

Ultimate Elongation,Before Aging500%, min500%, minSame
Tensile Strength, AfterAccelerated Aging14MPa, min14MPa, minSame
Ultimate Elongation, AfterAccelerated Aging400%, min400%, minSame
Watertight (1000ml)21 CFR 800.20ASTM D5151AQL 2.521 CFR 800.20ASTM D5151AQL 2.5Same
Powder-Content≤ 2 mg per glove≤ 2 mg per gloveSame
10993-10:2010 SkinIrritation StudyUnder the conditions ofthe study, not an irritantUnder the conditions of thestudy, the subject device isnon-irritatingSame
10993-10:2010 MaximizationSensitization StudyUnder the conditions of thestudy, not a sensitizerUnder the conditions of thestudy, the subject device isnon sensitizationSame
10993-5:2009 CytotoxicityTestUnder the conditions of thisstudy, the test article extractshowed potential toxicity toL929 cells.Cytotoxicity concern wasaddressed by acute systematictoxicity testing.Exhibits severe cytotoxicityreactivity at 100%, and 66%extract concentrations andno cytotoxicity reactivity at44%, 30%, 20% and 15%extract concentrations underthe condition of this test.Cytotoxicity concern wasaddressed by acutesystematic toxicity testing.Same
ISO 10993-11:2017Acute Systemic toxicity studyUnder the conditions of thisstudy, there was no evidenceof systemic toxicity.Under the conditions of thestudy, the subject showed noadverse biological reaction.Same

Chemotherapy Permeation and Fentanyl Citrate Comparison Claim:

Tested Chemotherapy Drug andConcentrationMinimum BDT (Minutes)Comparison
Proposed DevicePredicate DeviceK192954
Carmustine 3.3 mg/ml (3,300 ppm)11.218.2Similar
Cisplatin 1mg/ml (1,000 ppm)>240>240Same
Cyclophosphamide 20mg/ml (20,000 ppm)>240>240Same
Dacarbazine 10 mg/ml (10,000 ppm)>240>240Same
Doxorubicin HCL, 2 mg/ml (2,000 ppm)>240>240Same
Etoposide, 20 mg/ml (20,000 ppm)>240>240Same
Fluorouracil, 50mg/ml (50,000ppm)>240>240Same

{7}------------------------------------------------

Fuqian Xi Road, West district of Shenze, Industrial Base,

Shenze County, Hebei, 050000, China

510K Summary

Methotrexate, 25mg/ml (25,000ppm)>240N/ADifferent
Paclitaxel, 6mg/ml (6,000ppm)>240>240Same
Thiotepa, 10mg/ml (10,000ppm)29.457.3Similar
Fentanyl Citrate Injection (100 mcg/2ml)>240>240Same

7. Summary of Non-Clinical Performance Data

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device met the performance criteria with the following standards:

MethodologyTest PerformedAcceptance CriteriaResults
ASTM D6319- 19Physical Dimensions LengthMinimum 230mm for all sizesPass
ASTM D6319- 19Physical Dimensions Palm WidthXS: 70±10mmS: 80±10mmM: 95±10mmL:110±10mmXL: 120±10mmXXL: 130±10mmPass
ASTM D6319- 19Physical Dimensions ThicknessFinger: 0.05mm (min)Palm: 0.05mm (min)Pass
ASTM D6319- 19ASTM D412-16(2021)Physical PropertiesTensile Strength (Min14 MPa) and Elongation(Before Aging 500% and after aging 400%) MinAQL 2.5 (ISO 2859-1)Pass
ASTM D6319- 19ASTM D5151-19Water leak testAQL 2.5 (ISO 2859-1)Pass
ASTM D6319- 19ASTM D6124-06(2017)Powder ResidueMax 2mg/glovePass
ASTM D6978-05(2019)Permeation by Chemotherapy DrugsRefer the above tablePass
ISO 10993-10:2010Irritation and Skin SensitizationNo Skin sensitization and Skin irritationIs non-sensitization and Non-irritation
ISO 10993-5:2009CytotoxicityNo Cytotoxicity reactivityshowed potential toxicity to L929 cells.
ISO 10993-11:2017Acute systemic toxicity studySubject showed no adverse biological reactionno evidence of systemic toxicity

ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application. .

  • ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves. ●
  • ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical ●

{8}------------------------------------------------

Fuqian Xi Road, West district of Shenze, Industrial Base, Shenze County, Hebei, 050000, China

510K Summary

Gloves

  • . ASTM D412-16 (2021) Standard Test Methods for Vulcanized Rubber and Thermoplastic Elastomers-Tension
  • ASTM D6978-05 (Reapproved 2019), Assessment of Reissuance of Medical Gloves to Permeation . by Chemotherapy Drugs.
  • ISO 10993-10:2010 Biological Evaluation of Medical Devices Part 10: Tests For Irritation And . Skin Sensitization.
  • . ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity
  • . ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity

8. Clinical Performance Data

N/A

9. Conclusion:

The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicated device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.