K Number
K222415

Validate with FDA (Live)

Date Cleared
2022-10-25

(76 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Rexious Spinal Fixation System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disease; spondylolisthesis; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis).

The Rexious Spinal Fixation System is intended for use as a posterior, non-pedicle system indicated for degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

Device Description

The Rexious Spinal Fixation System is a top-loading multiple component, posterior spinal fixation system which consists of fixation system which consists of pedicle screws, rods, set screws, connectors, hooks, and transverse (cross) linking mechanisms.

The Rexious Spinal Fixation System will allow surgeons to build a spinal implant construct to stabilize and promote spinal fusion. The Rexious implant components are supplied non-sterile, single use and fabricated from titanium or titanium alloy (Ti-6Al-4V ELI) as specified in ASTM F67, F136, and F1295 and from Cobalt-Chromium-Molybdenum (CoCr) as specified in ASTM F1537. Various sizes of these implants are available.

AI/ML Overview

This document is a 510(k) summary for the Rexious Spinal Fixation System. It explicitly states that no performance testing was required for this submission. The basis for clearance is that the device is identical to previously cleared predicate devices.

Therefore, the requested information regarding acceptance criteria, study details, sample sizes, ground truth establishment, expert involvement, and MRMC studies cannot be provided from the given text because such studies were not conducted for this specific submission.

The relevant section from the document is:

  • Performance Data: "The subject and predicate devices are identical and therefore, no performance testing is required. Submission is only transferring name of a system that has already been cleared under K100765, K113324, K111362, and K173131. No testing is required."

This means the device's acceptance was based on its substantial equivalence to a previously cleared device, not on new performance data demonstrating it meets specific acceptance criteria through a new study.

{0}------------------------------------------------

October 25, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA acronym and the words "U.S. FOOD & DRUG" are in a blue sans-serif font, with "ADMINISTRATION" in a smaller font below.

Dio Medical Corporation Milan George Vice President of R&D 2100 Campus Lane, Suite 100 East Norriton, Pennsylvania 19403

Re: K222415

Trade/Device Name: Rexious Spinal Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB, KWP Dated: August 5, 2022 Received: August 10, 2022

Dear Milan George:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222415

Device Name Rexious Spinal Fixation System

Indications for Use (Describe)

The Rexious Spinal Fixation System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disease; spondylolisthesis; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis).

The Rexious Spinal Fixation System is intended for use as a posterior, non-pedicle system indicated for degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K222415 Page 1 of 2

510(k) SUMMARY

Dio Medical Rexious Spinal Fixation System

Sponsor:Manufacturer:Dio Medical Corp.2100 Campus Lane, Suite 100East Norriton, PA 19403
Official Contact:Milan GeorgeEmail: mgeorge@dio-us.comPhone: 1-877-394-5407 ext.102
Date Prepared:August 5, 2022
Device Name:Rexious Spinal Fixation System
Common Name:Pedicle Screw Spinal Fixation System
Classification Name:Thoracolumbosacral Pedicle Screw System
Classification Number:21 CFR 888.3070
Product Code/Classification:NKB, KWP, Class II
Description:The Rexious Spinal Fixation System is a top-loading multiplecomponent, posterior spinal fixation system which consists offixation system which consists of pedicle screws, rods, setscrews, connectors, hooks, and transverse (cross) linkingmechanisms.
The Rexious Spinal Fixation System will allow surgeons to build aspinal implant construct to stabilize and promote spinal fusion. TheRexious implant components are supplied non-sterile, single useand fabricated from titanium or titanium alloy (Ti-6Al-4V ELI) asspecified in ASTM F67, F136, and F1295 and from Cobalt-Chromium-Molybdenum (CoCr) as specified in ASTM F1537.Various sizes of these implants are available.
Indicationsfor Use:The Rexious Spinal Fixation System is intended to provideimmobilization and stabilization of spinal segments in skeletallymature patients as an adjunct to fusion in the treatment of thefollowing acute and chronic instabilities or deformities of thethoracic, lumbar, and sacral spine: degenerative disc disease;
spondylolisthesis; fracture; dislocation; scoliosis; kyphosis; spinaltumor; and failed previous fusion (pseudarthrosis).
The Rexious Spinal Fixation System is intended for use as aposterior, noncervical, non-pedicle system indicated fordegenerative disc disease (defined as back pain of discogenicorigin with degeneration of the disc confirmed by history andradiographic studies); spondylolisthesis; trauma (i.e., fracture ordislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis,and/or lordosis); tumor; pseudoarthrosis; and failed previousfusion.
PredicateDevice:Primary predicate:Huvexel Co. Ltd. - Rexious Spinal Fixation System (K173131)
Additional predicates:Huvexel Co. Ltd. : Rexious Hook System (K113324)Huvexel Co. Ltd. - Rexious Spinal Fixation System (K100765,K111362)
SubstantialEquivalence:The Rexious Spinal Fixation System is identical to the predicatedevice, the Huvexel - Rexious Spinal Fixation System and RexiousHook System. The Subject device has the same intended uses andsimilar indications, technological characteristics, and principles ofoperation as its predicate device. There are no technologicaldifferences between the Subject device and its predicate devicesresulting in no new issues of safety or effectiveness. Thus, theDio Medical Rexious Spinal Fixation System is identical/substantially equivalent to the predicates.
PerformanceData:The subject and predicate devices are identical and therefore, noperformance testing is required. Submission is only transferringname of a system that has already been cleared under K100765,K113324, K111362, and K173131. No testing is required.
Conclusion:The Dio Medical Rexious Spinal Fixation System has the sameintended uses and similar indications, technologicalcharacteristics, and principles of operation as its predicate device.Thus, the subject device is identical/substantially equivalent to thepredicate device.

{4}------------------------------------------------

N/A