K Number
K113324

Validate with FDA (Live)

Date Cleared
2012-03-28

(139 days)

Product Code
Regulation Number
888.3050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Rexious Hook System is intended for use as a posterior, noncervical, nonpedicle system indicated for degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. The Rexious Hook System can be used in conjunction with the Rexious Spinal Fixation System.

Device Description

The Rexious Hook System is a top-loading posterior spinal fixation system which consists of hooks, rods, set screws, connectors and a transverse link system. The Rexious Hook System implant components are fabricated from titanium alloy (Ti6A14V ELI) that conforms to ASTM F 136. Various sizes of these implants are available. The Rexious Hook System can be used in the posterior plane providing unilateral modes of fixation. The Rexious Hook System design allows adjustment in both the sagittal and coronal planes permitting hook placement according to the best possible anatomic location and orientation.

AI/ML Overview

The Rexious Hook System's acceptance criteria and the study proving it meets these criteria are outlined below.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Performance Test Standard)Reported Device Performance
Static Compression Bending (ASTM F1717)Characterized
Static Tension (ASTM F1717)Characterized
Static Torsion (ASTM F1717)Characterized
Dynamic Compression Tests (ASTM F1717)Characterized

Note: The document states that these tests were "submitted to characterize the subject Rexious Hook System components." This implies that the device successfully met the performance benchmarks outlined in the ASTM F1717 standard for these specific tests, as the FDA granted substantial equivalence.

2. Sample Size and Data Provenance for Test Set

The provided text describes performance testing conducted according to ASTM F1717. However, it does not specify explicit sample sizes for these tests. The data provenance is not explicitly stated as 'retrospective' or 'prospective,' but given that it involved characterizing device components in a laboratory setting per ASTM standards, it is inherently a controlled, prospective engineering study. The study was conducted by DIO Medical Co., Ltd. in South Korea.

3. Number of Experts and Qualifications for Ground Truth

This type of submission (510(k) for a medical device like a spinal fixation system) does not typically involve the establishment of ground truth by clinical experts in the same way an AI/ML model study would. The acceptance criteria are based on engineering performance standards (ASTM F1717), which are evaluated through physical testing, not expert consensus on clinical data. Therefore, this section is not applicable.

4. Adjudication Method for Test Set

As the evaluation is based on objective physical performance tests against established standards (ASTM F1717), there is no adjudication method in the context of clinical expert review. The results are quantitative measurements interpreted against the ASTM F1717 standard's requirements.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No multi-reader multi-case (MRMC) comparative effectiveness study was conducted or reported in the provided documentation. This type of study is more relevant for diagnostic devices or AI algorithms that assist human interpretation, not for mechanical implantable devices.

6. Standalone Performance Study (Algorithm Only)

This point is not applicable. The Rexious Hook System is a physical medical device, not a software algorithm, so a "standalone (algorithm only)" study is not relevant. The performance studies conducted were on the physical device components.

7. Type of Ground Truth Used

The "ground truth" for this device's performance is defined by the objective, quantifiable material and mechanical properties and performance criteria outlined in the ASTM F1717 standard (e.g., static compression bending, static tension, static torsion, and dynamic compression). These standards serve as the benchmark for the device's acceptable performance.

8. Sample Size for Training Set

There is no "training set" in the context of an AI/ML algorithm for this device. The development process would involve engineering design, material selection, and iterative physical prototyping and testing, not a data training set.

9. How Ground Truth for Training Set was Established

As there is no training set, this question is not applicable. The device's design and manufacturing are based on established engineering principles and materials science, with performance verified against industry standards like ASTM F1717.

{0}------------------------------------------------

MAR 2 8 2012

K11 3324
page 1 of 2

510(k) Summary

As specified under 21 CFR 807.92,

Submitter:

DIO Medical Co., Ltd. Sung-Hee Lee 103~105 Megacenter, SK Technopark 190-1 Sangdaewon-dong, Jungwon-guSungnam-si Kyunggi-do, South Korea Phone: 82-31-776-3690 Fax: 82-31-776-3691

Official Correspondent: Kodent Inc. April Lee

325 N. Puente St. Unit B Santa Fe Springs, CA 90670 Email; kodentinc@gmail.com Phone: 714-525-0114 Fax: 714-525-0116

Device Information

Trade Name: RexiousHook System Classification Name: Spinal Interlaminal Fixation Orthosis Product Code: KWP Regulation Number: 888.3050 Date Prepared: 3/21/2012

General Description

The Rexious Hook System is a top-loading posterior spinal fixation system which consists of hooks, rods, set screws, connectors and a transverse link system. The Rexious Hook System implant components are fabricated from titanium alloy (Ti6A14V ELI) that conforms to ASTM F 136. Various sizes of these implants are available.

The Rexious Hook System can be used in the posterior plane providing unilateral modes of fixation.

The Rexious Hook System design allows adjustment in both the sagittal and coronal planes permitting hook placement according to the best possible anatomic location and orientation.

Indication for Use

r

The Rexious Hook System is intended for use as a posterior, noncervical, nonpedicle system indicated for degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. The Rexious Hook System can be used in conjunction with the Rexious Spinal Fixation System.

{1}------------------------------------------------

K113324
page 2 of 2

Materials:

The devices are manufactured from Ti6Al-4V ELI alloy per ASTM and ISO Standards.

Performance Data:

Performance tests per ASTM F1717 such as static compression bending, static tension, static torsion, and dynamic compression tests were submitted to characterize the subject Rexious Hook System components addressed in this notification.

Predicate Devices:

The subject device is substantially equivalent to the following predicate devices:

*U&I America, Spinal Hook System™ (K031595)

Comparison to Predicate Devices:

The Substantial equivalence of this device is based on equivalence in intended use, material, designs, and operational principles to the predicate device, U&I America, Spinal Hook System™ (K031595).

Conclusion

Both the performance testing and comparison to predicate device demonstrate that the subject device is as safe, as effective, and performs as well as the predicate device.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle or bird symbol on the right side. On the left side of the logo, the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -- WO66-G609 Silver Spring, MD 20993-0002

DIO Medical Co., Ltd. % Kodent Incorporated Ms. April Lee 325 North Puente Street, Unit B Brea, California 92821

MAR 2 8 2012

Re: K113324

Trade/Device Name: Rexious Hook System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: II Product Code: KWP Dated: March 22, 2012 Received: March 26, 2012

Dear Ms. Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. April Lee

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

For
Mark N. Melkerson

Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indication for Use

K113324

510(K) Number (if known):

Device Name: Rexious Hook System

Indication for Use:

The Rexious Hook System is intended for use as a posterior, noncervical, nonpedicle system indicated for degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. The Rexious Hook System can be used in conjunction with the Rexious Spinal Fixation System.

Prescription Use x (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter

(Per 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1

(Diysion Sign-Ori) Difision of Surgical Orthopedic, and Restorative Devices

510(k) Number_________________________________________________________________________________________________________________________________________________________________

§ 888.3050 Spinal interlaminal fixation orthosis.

(a)
Identification. A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.(b)
Classification. Class II.