K Number
K222358

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-12-20

(138 days)

Product Code
Regulation Number
870.5150
Age Range
0 - 21
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

INDIGO Aspiration Catheters and Separators:

As part of the INDIGO Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems, and for the treatment of pulmonary embolism.

INDIGO Aspiration Tubing:

As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Aspiration Pump.

Penumbra Aspiration Pump:

The Penumbra Aspiration Pump is indicated as a vacuum source for the Penumbra Aspiration Systems.

Device Description

The INDIGO® Aspiration System is comprised of the several devices:

  • INDIGO Aspiration Catheter
  • Penumbra Aspiration Pump
  • INDIGO Aspiration Pump Canister
  • INDIGO Aspiration Tubing
  • INDIGO Separator

The INDIGO Aspiration System is designed to remove thrombus from the vasculature using mechanical aspiration. The INDIGO Aspiration Catheter targets aspiration from the pump directly to the thrombus. The INDIGO Separator may be used to clear the lumen of the INDIGO Aspiration Catheter should it become blocked with thrombus. The INDIGO Aspiration Catheter is introduced through a guide catheter or vascular sheath into the peripheral vasculature and guided over a guidewire to the site of the primary occlusion. The INDIGO Aspiration Catheter is used with the Penumbra Aspiration Pump to aspirate thrombus from an occluded vessel. As needed, an INDIGO Separator may be deployed from the INDIGO Aspiration Catheter to assist with thrombus removal. The INDIGO Separator is advanced and retracted through the INDIGO Aspiration Catheter at the proximal margin of the primary occlusion to facilitate clearing of the thrombus from the INDIGO Aspiration Catheter tip. The devices are visible under fluoroscopy. For the aspiration source, the INDIGO Aspiration Catheter is used in conjunction with the Penumbra Aspiration Pump, which is connected using the INDIGO Aspiration Tubing and the INDIGO Aspiration Pump Canister. The INDIGO Aspiration Catheter may be provided with a steam shaping mandrel, rotating hemostasis valve, Select Catheter, and introducer. The INDIGO Separator may be provided with an introducer and torque device.

AI/ML Overview

The provided text is a compilation of FDA documents related to the 510(k) clearance for the Penumbra, Inc. Indigo® Aspiration System – Lightning® Flash. It describes the device, its indications for use, and a comparison to predicate devices, along with summaries of non-clinical data (biocompatibility and bench-top testing).

Based on the information provided, here's an analysis regarding acceptance criteria and the study that proves the device meets them:

1. A table of acceptance criteria and the reported device performance:

The document does not explicitly present a table of acceptance criteria with corresponding device performance metrics in a quantitative manner as one might expect for an AI/ML device. Instead, it describes general compliance and qualitative outcomes for various tests.

The reported performance is described in a narrative format, indicating that tests "met all acceptance criteria" and "performed similarly to the predicate devices."

Acceptance Criteria CategoryReported Device Performance
Biocompatibility- Cytotoxicity: Passed (Lightning Flash Aspiration Tubing, Flash Aspiration Catheter, RHV, Reducer, Connector Cap)
- Sensitization: Passed (Lightning Flash Aspiration Tubing, Flash Aspiration Catheter, RHV, Reducer, Connector Cap)
- Irritation: Passed (Lightning Flash Aspiration Tubing, Flash Aspiration Catheter, RHV, Reducer, Connector Cap)
- Systemic Toxicity (Acute Systemic Injection, Material Mediated Pyrogen): Passed (Flash Aspiration Catheter, RHV, Reducer, Connector Cap)
- Hemocompatibility (Thrombogenicity, Prothrombin Time (PT), Partial Thromboplastin Time (PTT), Complement Activation, Direct/Indirect hemolysis): Passed (Flash Aspiration Catheter, RHV, Reducer, Connector Cap)
Conclusion: "The results from the testing performed showed the subject devices to be biocompatible."
Design Verification (Bench-Top) Testing- Dimensional/Visual Inspection: Met acceptance criteria (Lightning Flash Aspiration Tubing, Flash Aspiration Catheter and Select Catheter)
- Performance/Simulated Use Testing: Met acceptance criteria (Lightning Flash Apiration Tubing, Flash Aspiration Catheter and Select Catheter)
- Tensile Testing: Met acceptance criteria (Lightning Flash Apiration Tubing)
- Friction Testing: Met acceptance criteria (Flash Aspiration Catheter and Select Catheter)
- Torsion Testing: Met acceptance criteria (Flash Aspiration Catheter and Select Catheter)
- Vacuum Test: Met acceptance criteria (Flash Aspiration Catheter and Select Catheter)
- Indigo Aspiration System Compatibility: Met acceptance criteria
- Valve Sense Testing: Met acceptance criteria
- Coating Integrity Testing: Met acceptance criteria
- Particulate Testing: Met acceptance criteria
- Hub Air Aspiration: Met acceptance criteria
- Catheter Pressure: Met acceptance criteria
- Hub / Shaft Tensile Strength: Met acceptance criteria
- Catheter Shaft Tensile: Met acceptance criteria
- Elongation to Failure: Met acceptance criteria
- Corrosion: Met acceptance criteria
Conclusion: "The in vitro bench-top tests demonstrated that the subject Lightning Flash met all acceptance criteria and performed similarly to the predicate devices. Performance data demonstrate that the Lightning Flash devices function as intended and does not raise any new questions of safety and effectiveness compared to the predicate devices."
Electrical Safety/EMC Testing- Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, and ISO 10079-1: Complied
Software Verification and Validation- Compliance with FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices": Provided and tested (Minor Level of Concern)

2. Sample size used for the test set and the data provenance:

The document does not provide details on the specific sample sizes used for each non-clinical (bench-top) test or biocompatibility test. It mentions "testing was conducted" and "tests were performed" without quantifying the number of units or replicates involved.

Data provenance (e.g., country of origin, retrospective/prospective) is not applicable as the studies described are non-clinical bench-top tests and biocompatibility assessments, not clinical studies involving patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not applicable. The device is an embolectomy catheter system, which is a physical medical device, not an AI/ML diagnostic or prognostic tool that requires expert interpretation to establish ground truth for a test set. The "ground truth" for its performance is established through measurable physical and biological parameters in laboratory settings (e.g., tensile strength, aspiration efficacy in simulated models, biocompatibility of materials).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

This information is not applicable for the reasons stated in point 3. The studies are technical verification and validation, not studies requiring adjudication of image interpretations or clinical outcomes by multiple experts.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

No. An MRMC comparative effectiveness study was not done. This type of study is relevant for AI/ML-driven diagnostic tools, not for a mechanical medical device like an aspiration system. The document explicitly states: "No clinical study was conducted as bench and previously performed animal testing was determined sufficient for verification and validation purposes."

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

No. This concept is not applicable as the Indigo® Aspiration System – Lightning® Flash is a mechanical medical device, not an algorithm or AI system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The "ground truth" for this device's performance is based on well-established engineering standards, material science principles, and biological safety assessments. This includes:

  • Quantitative measurements from bench-top tests (e.g., dimensions, tensile strength, vacuum pressure).
  • Adherence to recognized standards (e.g., ISO, ASTM for materials and performance, IEC for electrical safety).
  • Biocompatibility testing results (e.g., lack of cytotoxicity, sensitization).
  • Functional performance in simulated use models (e.g., removal of simulated thrombi).

8. The sample size for the training set:

Not applicable. This device is not an AI/ML product that undergoes a "training set" for model development. The development process involves engineering design, prototyping, and testing of physical units.

9. How the ground truth for the training set was established:

Not applicable, as there is no "training set" in the context of this mechanical medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol of the Department of Health & Human Services. To the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The word "FDA" is in a larger font and is placed inside a blue square.

January 30, 2023

Penumbra, Inc. Deanna Kimlinger Senior Regulatory Specialist One Penumbra Place Alameda, California 95132

Re: K222358

Trade/Device Name: Indigo® Aspiration System - Lightning® Flash Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: QEW

Dear Deanna Kimlinger:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated December 20, 2022. Specifically, FDA is updating this SE Letter to correct a typo in the device trade name as an administrative correction.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Gregory O'Connell, OHT2: Office of Cardiovascular Devices, by phone (301-796-6075), or email (gregory.oconnell(@fda.hhs.gov).

Sincerely,

Gregory W. Digitally signed by
Gregory W. Gregory W. O'connell -S O'connell -S Date: 2023.01.30
12:06:11 -05'00'

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 20, 2022

Penumbra, Inc. Deanna Kimlinger Senior Regulatory Specialist One Penumbra Place Alameda, California 95132

Re: K222358

Trade/Device Name: Indigo® Aspiration System - Lighting® Flash Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: QEW Dated: November 17, 2022 Received: November 18, 2022

Dear Deanna Kimlinger:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/ofdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{2}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/2/Picture/6 description: The image shows a digital signature. The signature is for Gregory W. O'connell. The date of the signature is 2022.12.20, and the time is 15:45:58 -05'00'.

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K222358

Device Name Indigo® Aspiration System - Lightning® Flash

Indications for Use (Describe)

INDIGO Aspiration Catheters and Separators:

As part of the INDIGO Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems, and for the treatment of pulmonary embolism.

INDIGO Aspiration Tubing:

As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Aspiration Pump.

Penumbra Aspiration Pump:

The Penumbra Aspiration Pump is indicated as a vacuum source for the Penumbra Aspiration Systems.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary 1

(as required by 21 CFR 807.92)

Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Penumbra, Inc. is providing the summary of Substantial Equivalence for the Indigo "Aspiration System - Lightning" Flash.

1.1 Sponsor/Applicant Name and Address

Penumbra, Inc. One Penumbra Place Alameda, CA 94502 USA

1.2 Sponsor Contact Information

Deanna Kimlinger Senior Regulatory Affairs Specialist Phone: (925) 212-9088 FAX: (510) 217-6414 Email: dkimlinger@penumbrainc.com

1.3 Date of Preparation of 510(k) Summary

December 20, 2022

1.4 Device Trade or Proprietary Name

Indigo

1.5 Device Classification

Regulatory Class:II
Classification Panel:Cardiovascular
Classification Name:Catheter, Embolectomy
Regulation Number:21 CFR §870.5150
Product Code:QEW

1.6 Predicate and Reference Devices - Lightning Flash

510(k) NumberName of Device
Lightning Flash Aspiration Tubing
PredicateK210323Indigo Aspiration System – Lightning Aspiration Tubing
ReferenceK202821Indigo Aspiration System - Aspiration Catheter 12 and Separator 12
Flash Aspiration Catheter
PredicateK192981Indigo Aspiration System - Aspiration Catheter 12 and Separator 12
ReferenceK202821Indigo Aspiration System - Aspiration Catheter 12 and Separator 12
ReferenceK180466FlowTriever Retrieval/Aspiration System
Select Catheter (Accessory)
PredicateK111380Neuron MAX System

{5}------------------------------------------------

1.7 Predicate Comparison

System NameIndigo® Aspiration System
Lightning Aspiration Tubing[Predicate]CAT12 & SEP12[Predicate]CAT12 & SEP12 withPE Treatment Indication[Reference]Lightning Flash[Subject]
ClassificationClass II, QEWClass II, QEWClass II, QEWSAME
510(k) no.K210323K192981K202821K222358
IndicationINDIGO Aspiration Cathetersand Separators:As part of the INDIGO™Aspiration System, the INDIGOAspiration Catheters andSeparators are indicated for theremoval of fresh, soft emboliand thrombi from vessels of theperipheral arterial and venoussystems, and for the treatmentof pulmonary embolism.INDIGO Aspiration Tubing:As part of the INDIGO™Aspiration System, the INDIGOSterile Aspiration Tubing isindicated to connect theINDIGO Aspiration Catheters tothe Penumbra Aspiration Pump.Penumbra Aspiration Pump:The Penumbra Aspiration Pumpis indicated as a vacuum sourcefor the Penumbra AspirationINDIGO Aspiration Catheters andSeparators:As part of the INDIGO AspirationSystem, the INDIGO AspirationCatheters and Separators areindicated for the removal offresh, soft emboli and thrombifrom vessels of the peripheralarterial and venous systems.INDIGO Aspiration Tubing:As part of the INDIGO AspirationSystem, the INDIGO SterileAspiration Tubing is indicated toconnect the INDIGO AspirationCatheters to the PenumbraAspiration Pump.Penumbra Aspiration Pump:The Penumbra Aspiration Pumpis indicated as a vacuum sourcefor Penumbra AspirationSystems.INDIGO Aspiration Cathetersand Separators:As part of the INDIGO™Aspiration System, the INDIGOAspiration Catheters andSeparators are indicated for theremoval of fresh, soft emboliand thrombi from vessels of theperipheral arterial and venoussystems, and for the treatmentof pulmonary embolism.INDIGO Aspiration Tubing:As part of the INDIGO™Aspiration System, the INDIGOSterile Aspiration Tubing isindicated to connect the INDIGOAspiration Catheters to thePenumbra Aspiration Pump.Penumbra Aspiration Pump:The Penumbra Aspiration Pumpis indicated as a vacuum sourcefor the Penumbra AspirationSystems.SAME AS PREDICATEK210323 andREFERENCE K202821
Aspiration TubingLightning Aspiration Tubing[Predicate]Lightning Flash Aspiration Tubing[Subject]
510(k) No.K210323K222358
Materials
MaterialsBiocompatible, commonly utilized forinterventional devicesSAME
Dimensions
Proximal ODAppropriately sized for the AspirationCatheterSAME plus larger diameters
Proximal ID
Distal OD
Distal ID
Effective Length
Aspiration CatheterIndigo System – CAT12[Predicate]Indigo System – Flash AspirationCatheter[Subject]
510(k) No.K192981K222358
Materials
MaterialsBiocompatible, commonly utilized forinterventional devicesSAME with subset of colorants
CoatingHydrophilicSAME
Dimensions
Proximal OD
Proximal IDSAME plus larger diameters
Distal ODAppropriately sized for the target vessel
Distal ID
Effective Length
Accessories
Sheath: GrilamidSAME
IntroducerID Band: PolyolefinSAME
RotatingHemostasisMaterialsPolycarbonate/Silicone/PTFEPolycarbonate/Silicone/PVC/PTFE/Polyolefin with foil/Loctite/Thermoplastic Polymer
Valve(RHV)FrenchSize10F13F
Select Catheter(Accessory)Neuron MAX - 6F Select Catheter[Predicate]Indigo System – Select Catheter[Subject]
510(k) No.K111380K222358
Materials
MaterialsBiocompatible, commonly utilized forinterventional devicesSAME

{6}------------------------------------------------

{7}------------------------------------------------

Dimensions
OD [Maximum]ID [Minimum]Working LengthAppropriately sized for the introductionof interventional devices into the targetvesselSAME plus expanded workinglength range
Accessories
Torque DevicePolycarbonate/Brass/PolypropyleneSAME
Packaging MaterialsCommonly used materials for medicaldevicesSAME
Aspiration SourcePenumbra Aspiration PumpSAME
SterilizationEOSAME
Shelf-Life36 Months12 Months
UseSingle use, disposableSAME

1.8 Device Description

The INDIGO® Aspiration System is comprised of the several devices:

  • INDIGO Aspiration Catheter
  • Penumbra Aspiration Pump
  • INDIGO Aspiration Pump Canister
  • INDIGO Aspiration Tubing
  • INDIGO Separator

The INDIGO Aspiration System is designed to remove thrombus from the vasculature using mechanical aspiration. The INDIGO Aspiration Catheter targets aspiration from the pump directly to the thrombus. The INDIGO Separator may be used to clear the lumen of the INDIGO Aspiration Catheter should it become blocked with thrombus. The INDIGO Aspiration Catheter is introduced through a guide catheter or vascular sheath into the peripheral vasculature and guided over a guidewire to the site of the primary occlusion. The INDIGO Aspiration Catheter is used with the Penumbra Aspiration Pump to aspirate thrombus from an occluded vessel. As needed, an INDIGO Separator may be deployed from the INDIGO Aspiration Catheter to assist with thrombus removal. The INDIGO Separator is advanced and retracted through the INDIGO Aspiration Catheter at the proximal margin of the primary occlusion to facilitate clearing of the thrombus from the INDIGO Aspiration Catheter tip. The devices are visible under fluoroscopy. For the aspiration source, the INDIGO Aspiration Catheter is used in conjunction with the Penumbra Aspiration Pump, which is connected using the INDIGO Aspiration Tubing and the INDIGO Aspiration Pump Canister. The INDIGO Aspiration Catheter may be provided with a steam shaping mandrel, rotating hemostasis valve, Select Catheter, and introducer. The INDIGO Separator may be provided with an introducer and torque device.

{8}------------------------------------------------

Lightning Aspiration Tubing

The Lightning Aspiration Tubing is a component to the currently available Indigo Aspiration System. The Lightning Aspiration Tubing facilitates the transfer of vacuum between the INDIGO Aspiration Catheter and the Penumbra Aspiration Pump while providing aspiration. Intended users for this device are physicians who have received appropriate training in surgical procedures and/or interventional techniques. The device is provided sterile, non-pyrogenic, and intended for single use only.

1.9 Indications for Use

INDIGO Aspiration Catheters and Separators

As part of the INDIGO Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems, and for the treatment of pulmonary embolism.

INDIGO Aspiration Tubing

As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Aspiration Pump.

Penumbra Aspiration Pump

The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

1.10 Summary of Non-Clinical Data/ Performance Data

As required under Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, a summary of any information regarding substantial equivalence of the devices follows. Included in this section are summary descriptions of the testing which substantiates the performance of the subject Lightning Flash.

1.10.1 Biocompatibility

Biocompatibility testing was conducted on the subject Lightning Flash. Biocompatibility testing was leveraged for the Introducer (K192981) and Select Catheter (K111380) accessories. The following tests were performed:

{9}------------------------------------------------

  • Lightning Flash Aspiration Tubing
    • O Cytotoxicity
    • O Sensitization
    • o Irritation
  • . Flash Aspiration Catheter
    • Cytotoxicity O
    • Sensitization O
    • O Irritation
    • Systemic Toxicity O
      • Acute Systemic Injection
      • I Material Mediated Pyrogen
    • Hemocompatibility O
      • . Thrombogenicity
      • Prothrombin Time (PT)
      • . Partial Thromboplastin Time (PTT)
      • . Complement Activation
      • . Direct/Indirect hemolysis
  • . RHV, Reducer, Connector Cap (Accessories)
    • Cytotoxicity O
    • Sensitization O
    • O Irritation
    • Systemic Toxicity O
      • . Acute Systemic Injection
      • Material Mediated Pyrogen
    • Hemocompatibility o
      • . Direct/Indirect hemolysis

The results from the testing performed showed the subject devices to be biocompatible.

1.10.2 Design Verification (Bench-Top) Testing

Non-clinical laboratory testing was performed on the subject Lightning Flash devices to determine substantial equivalence. The following tests were performed:

  • Lightning Flash Aspiration Tubing
  • . Flash Aspiration Catheter and Select Catheter
  • Dimensional/Visual Inspection O
  • Performance/Simulated Use O Testing
  • Tensile Testing O
  • Dimensional/Visual Inspection o
  • Friction Testing O
  • Performance/Simulated Use and Torsion o Testing
  • Vacuum Test o

{10}------------------------------------------------

  • o Indigo Aspiration System Compatibility
  • o Valve Sense Testing
  • Coating Integrity Testing O
  • Particulate Testing O
  • O Hub Air Aspiration
  • Catheter Pressure O
  • Hub / Shaft Tensile Strength O
  • Catheter Shaft Tensile O
  • Elongation to Failure O
  • o Corrosion

The in vitro bench-top tests demonstrated that the subject Lightning Flash met all acceptance criteria and performed similarly to the predicate devices. Performance data demonstrate that the Lightning Flash devices function as intended and does not raise any new questions of safety and effectiveness compared to the predicate devices.

1.10.3 Performance Data – Clinical

No clinical study was conducted as bench and previously performed animal testing was determined sufficient for verification and validation purposes. A review was conducted considering published clinical study articles that featured the predicate device and other devices with similar dimensions used for direct aspiration. The literature review was used to support the determination of substantial equivalence by leveraging clinical outcomes from devices that are considered technologically equivalent.

1.10.4 Electrical Safety/EMC Testing

Electrical Safety and EMC testing were conducted on the Lightning Flash Aspiration Tubing. The subject device complies with the requirements of IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, and ISO 10079-1.

1.10.5 Software

Software verification and validation testing and documentation for the Lightning Flash Aspiration Tubing was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (issued May 11, 2005). The software for this device was considered as a Minor Level of Concern.

1.11 Summary of Substantial Equivalence

The subject Indigo Aspiration System - Lightning Flash is substantially equivalent to the predicate devices, provided in Section 1.6 with regards to intended use, operating principle, design concept, materials, sterilization processes and packaging processes.

§ 870.5150 Embolectomy catheter.

(a)
Identification. An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.(b)
Classification. Class II (performance standards).