K Number
K202821

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-11-18

(55 days)

Product Code
Regulation Number
870.5150
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

INDIGO Aspiration Catheters and Separators:

As part of the INDIGO Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems, and for the treatment of pulmonary embolism.

INDIGO Aspiration Tubing:

As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Aspiration Pump.

Penumbra Aspiration Pump:

The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

Device Description

The INDIGO® Aspiration System is comprised of:

  • INDIGO Aspiration Catheter .
  • Penumbra Aspiration Pump .
  • INDIGO Aspiration Pump Canister ●
  • INDIGO Aspiration Tubing ●
  • INDIGO Separator™ .

The INDIGO Aspiration System is designed to remove thrombus from the vasculature using mechanical aspiration. The INDIGO Aspiration Catheter targets aspiration from the pump directly to the thrombus. The INDIGO Separator may be used to clear the lumen of the INDIGO Aspiration Catheter should it become blocked with thrombus. The INDIGO Aspiration Catheter is introduced through a guide catheter or vascular sheath and into the peripheral vasculature and guided over a guidewire to the site of the primary occlusion. The INDIGO Aspiration Catheter is used with the Penumbra Aspiration Pump to aspirate thrombus from an occluded vessel. As needed, an INDIGO Separator may be deployed from the INDIGO Aspiration Catheter to assist with thrombus removal. The INDIGO Separator is advanced and retracted through the INDIGO Aspiration Catheter at the proximal margin of the primary occlusion to facilitate clearing of the thrombus from the INDIGO Aspiration Catheter tip. The devices are visible under fluoroscopy. For the aspiration source, the INDIGO Aspiration Catheter is used in conjunction with the Penumbra Aspiration Pump, which is connected using the INDIGO Aspiration Tubing and the INDIGO Aspiration Pump Canister. The INDIGO Aspiration Catheter may be provided with a steam shaping mandrel, rotating hemostasis valve, and introducer. The INDIGO Separator may be provided with an introducer and torque device.

AI/ML Overview

This FDA 510(k) Premarket Notification is for the Indigo Aspiration System - Aspiration Catheter 12 and Separator 12. The document states that the subject device is identical to the predicate device (K192981) and leverages clinical data from a reference device (K192833). Therefore, the information below focuses on the clinical study (EXTRACT-PE) mentioned as supporting the substantial equivalence for the subject device.

1. Table of Acceptance Criteria and Reported Device Performance:

The document does not explicitly present a table of acceptance criteria with corresponding device performance for the subject device. Instead, it states that the EXTRACT-PE trial (for the reference device K192833) "demonstrated that the Indigo Aspiration System showed substantially equivalent safety and effectiveness outcomes for acute PE. The primary efficacy and safety endpoints were met."

  • Acceptance Criteria (Implied from the statement): Meeting primary efficacy and safety endpoints for acute Pulmonary Embolism (PE) treatment, demonstrating substantial equivalence to existing treatments. Specific numerical acceptance criteria are not provided in this document.
  • Reported Device Performance: Achieved substantially equivalent safety and effectiveness outcomes for acute PE, and met primary efficacy and safety endpoints.

2. Sample Size Used for the Test Set and Data Provenance:

The document refers to the EXTRACT-PE clinical study for the reference device (K192833). However, it does not provide the specific sample size used for this trial nor the data provenance (country of origin, retrospective/prospective) within this 510(k) summary. Further documentation would be needed to extract this detail.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

This information is not provided in the given document. The summary only mentions "primary efficacy and safety endpoints were met," implying clinical assessment, but not the specifics of ground truth establishment by experts.

4. Adjudication Method for the Test Set:

This information is not provided in the given document.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs without AI assistance:

This is not applicable. The device described is an aspiration system (a medical device used for removing emboli and thrombi), not an AI-based diagnostic or imaging device that would involve human readers interpreting output with or without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

This is not applicable. The device is a physical medical device, not an algorithm. Performance is assessed through its use in a clinical setting by medical professionals.

7. The Type of Ground Truth Used:

For the EXTRACT-PE study, the ground truth would have been established through clinical outcomes data related to the efficacy (e.g., successful removal of thrombi, resolution of PE symptoms, improvements in relevant physiological parameters) and safety (e.g., adverse events, complications) of the device in treating acute PE.

8. The Sample Size for the Training Set:

This information is not provided in the given document. As it is a non-AI medical device, the concept of a "training set" in the context of machine learning does not directly apply. The clinical study for the reference device would have involved a patient cohort, but the specific number is not stated.

9. How the Ground Truth for the Training Set was Established:

As above, the concept of a "training set" is not directly applicable. For the clinical study that demonstrated the device's performance, the "ground truth" (clinical outcomes, safety events, etc.) would have been established through standard clinical trial methodologies, including objective measurements, physician assessments, and follow-up data collection on patients enrolled in the trial.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 18, 2020

Penumbra, Inc. Michaela Mahl Senior Manager of Regulatory Affairs One Penumbra Place Alameda, California 94502

Re: K202821

Trade/Device Name: Indigo Aspiration System - Aspiration Catheter 12 and Separator 12 Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: QEW Dated: September 22, 2020 Received: September 24, 2020

Dear Michaela Mahl:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202821

Device Name

Indigo Aspiration System - Aspiration Catheter 12 and Separator 12

Indications for Use (Describe)

INDIGO Aspiration Catheters and Separators:

As part of the INDIGO Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems, and for the treatment of pulmonary embolism.

INDIGO Aspiration Tubing:

As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Aspiration Pump.

Penumbra Aspiration Pump:

The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the word "Penumbra" in a red, sans-serif font. To the right of the word is a circular logo, also in red, with a white stylized "P" inside. The logo has a white horizontal line extending from the "P" to the edge of the circle. The overall design is clean and modern.

K202821

1 510(k) Summary

(as required by 21 CFR 807.92)

Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Penumbra, Inc. is providing the summary of Substantial Equivalence for the Indigo® System - Aspiration Catheter 12 and Separator 12.

1.1 Sponsor/Applicant Name and Address

Penumbra, Inc. One Penumbra Place Alameda, CA 94502 USA

1.2 Sponsor Contact Information

Michaela Mahl Senior Manager of Regulatory Affairs Phone: (510) 748-3288 FAX: (510) 217-6414 Email: mmahl@penumbrainc.com

1.3 Date of Preparation of 510(k) Summary

November 18, 2020

Device Trade or Proprietary Name 1.4

Indigo® Aspiration System - Aspiration Catheter 12 and Separator 12

1.5 Device Classification

Regulatory Class:II
Classification Panel:Cardiovascular
Classification Name:Catheter, Embolectomy
Regulation Number:21 CFR §870.5150
Product Code:QEW

Predicate and Reference Devices 1.6

510(k) NumberName of DeviceName ofManufacture
Predicate Device
K192981Indigo Aspiration System – Aspiration Catheter 12and Separator 12Penumbra, Inc.
Reference Device
K192833Indigo Aspiration SystemPenumbra, Inc.
K180466FlowTriever Retrieval/Aspiration SystemInari Medical, Inc.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "Penumbra" in a red sans-serif font, followed by a red circle with a white "P" inside. The "P" is formed by a horizontal line that extends beyond the circle. The logo is simple and modern, and the red color gives it a sense of energy and excitement. The registered trademark symbol is located at the bottom right of the red circle.

Predicate Comparison 1.7

System NameIndigo® Aspiration System
Predicate DeviceReference DeviceSubject Device
ClassificationClass II, QEWSAME
510(k) no.K192981K192833K202821
IndicationINDIGO Aspiration Catheters andSeparators:As part of the INDIGO AspirationSystem, the INDIGO AspirationCatheters and Separators areindicated for the removal of fresh,soft emboli and thrombi from vesselsof the peripheral arterial and venoussystems.INDIGO Aspiration Tubing:As part of the INDIGO AspirationSystem, the INDIGO SterileAspiration Tubing is indicated toconnect the INDIGO AspirationCatheters to the Penumbra AspirationPump.Penumbra Aspiration Pump:The Penumbra Aspiration Pump isindicated as a vacuum source forPenumbra Aspiration Systems.INDIGO Aspiration Catheters andSeparators:As part of the INDIGO AspirationSystem, the INDIGO AspirationCatheters and Separators are indicatedfor the removal of fresh, soft emboliand thrombi from vessels of theperipheral arterial and venous systems,and for the treatment of pulmonaryembolism.INDIGO Aspiration Tubing:As part of the INDIGO AspirationSystem, the INDIGO Sterile AspirationTubing is indicated to connect theINDIGO Aspiration Catheters to thePenumbra Aspiration Pump.Penumbra Aspiration Pump:The Penumbra Aspiration Pump isindicated as a vacuum source forPenumbra Aspiration Systems.Same as ReferenceDeviceK192833
SystemComponentsReperfusion Catheter, Separator,Aspiration Tubing, Aspiration SourceSAMESAME
Materials,Packaging &ConfigurationsCommonly utilized for interventionaldevicesSAMESAME
AspirationSourcePenumbra Aspiration PumpSAMESAME
SterilizationEOSAMESAME
Shelf-Life36 Months36 MonthsSAME

Device Description 1.8

The INDIGO® Aspiration System is comprised of:

  • INDIGO Aspiration Catheter .
  • Penumbra Aspiration Pump .
  • INDIGO Aspiration Pump Canister ●
  • INDIGO Aspiration Tubing ●
  • INDIGO Separator™ .

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Penumbra company logo. The word "Penumbra" is written in a red sans-serif font. To the right of the word is a red circle with a white "P" inside. The "P" is stylized with a line extending from the top of the "P" to the edge of the circle.

K202821

The INDIGO Aspiration System is designed to remove thrombus from the vasculature using mechanical aspiration. The INDIGO Aspiration Catheter targets aspiration from the pump directly to the thrombus. The INDIGO Separator may be used to clear the lumen of the INDIGO Aspiration Catheter should it become blocked with thrombus. The INDIGO Aspiration Catheter is introduced through a guide catheter or vascular sheath and into the peripheral vasculature and guided over a guidewire to the site of the primary occlusion. The INDIGO Aspiration Catheter is used with the Penumbra Aspiration Pump to aspirate thrombus from an occluded vessel. As needed, an INDIGO Separator may be deployed from the INDIGO Aspiration Catheter to assist with thrombus removal. The INDIGO Separator is advanced and retracted through the INDIGO Aspiration Catheter at the proximal margin of the primary occlusion to facilitate clearing of the thrombus from the INDIGO Aspiration Catheter tip. The devices are visible under fluoroscopy. For the aspiration source, the INDIGO Aspiration Catheter is used in conjunction with the Penumbra Aspiration Pump, which is connected using the INDIGO Aspiration Tubing and the INDIGO Aspiration Pump Canister. The INDIGO Aspiration Catheter may be provided with a steam shaping mandrel, rotating hemostasis valve, and introducer. The INDIGO Separator may be provided with an introducer and torque device.

Indications for Use 1.9

INDIGO Aspiration Catheters and Separators:

As part of the INDIGO Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems, and for the treatment of pulmonary embolism.

INDIGO Aspiration Tubing:

As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Aspiration Pump.

Penumbra Aspiration Pump: The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

1.10 Summary of Non-Clinical Data

The subject and predicate Indigo System devices are identical. Therefore, previous device performance data regarding substantial equivalence described below remain unchanged.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "Penumbra" in a red, sans-serif font. To the right of the word is a circular logo, also in red, with a white "P" shape inside. The logo is a solid red circle with a thin white border, and the "P" shape is formed by a thick white line that runs vertically and then curves to the right at the top.

1.10.1 Summary of Biocompatibility

There are no changes to the previously provided biocompatibility data of the Indigo System - Aspiration Catheter 12 and Separator 12 sterile device materials cleared in K192981 (predicate device).

1.10.2 Summary of Performance Testing - Bench-Top

There are no changes to the previously provided bench-top data of the devices cleared in K192981 (predicate and reference device).

1.11 Summary of Performance Data - Clinical

The Indigo System, clinical study (EXTRACT-PE) was cleared in K192833 (reference device). The EXTRACT-PE trial demonstrated that the Indigo Aspiration System showed substantially equivalent safety and effectiveness outcomes for acute PE. The primary efficacy and safety endpoints were met. No additional clinical study was conducted as bench and previously performed animal testing was determined sufficient for verification and validation purposes. A review of the technological characteristics of the subject and reference device supported leveraging the clinical outcomes of the EXTRACT-PE Clinical study for the subject device.

1.12 Summary of Shelf-Life

There are no changes to the previously provided shelf-life data of the devices cleared in K192981 (predicate device).

1.13 Summary of Packaging

There are no changes to the packaging material listing or the packaging process for the devices cleared in K192981 (predicate device).

1.14 Summary of Substantial Equivalence

The subject device is substantially equivalent to the predicate and reference devices with regard to indications, intended use, design, performance, materials, sterilization, and packaging.

§ 870.5150 Embolectomy catheter.

(a)
Identification. An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.(b)
Classification. Class II (performance standards).