K Number
K222095

Validate with FDA (Live)

Manufacturer
Date Cleared
2023-04-20

(276 days)

Product Code
Regulation Number
870.5150
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BASHIR™ Endovascular Catheter and BASHIR™ S-B Endovascular Catheter are mechanical thrombolysis catheters indicated for the:

  • · Controlled and selective infusion of physician-specified fluids, including thrombolytics, into the pulmonary arteries for treatment of pulmonary embolism.
  • · Infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature, enabling the restoration of blood flow in patients with venous thrombus.
Device Description

The BASHIR™ Endovascular Catheter is a device intended for mechanical thrombolysis via expanding the distal infusion basket, allowing localized infusion of physician-specified fluids, including thrombolytics, into the peripheral or pulmonary vasculature. The distal infusion segment of the device is 12.50 cm (4.94 in.) long and consists of an expandable basket with a nitinol core surrounded by mini-infusion catheters, each with multiple infusion holes (hereafter referred to as the "infusion basket"). The infusion basket is expanded and closed using the red actuator located on the handle at the proximal end of the device. The BASHIR™ S-B Endovascular Catheter is a second model intended for smaller anatomies, with an infusion basket that is 10.0 cm (3.94 in.). The devices are otherwise the same.

The physician-specified fluid is then delivered via the basket infusion line located on the handle. Pulse sprays are followed by infusion through laser-drilled holes in the infusion limbs of the basket after placement in the peripheral vasculature or in the pulmonary artery. After the completion of fluid delivery, the infusion limbs are collapsed to a straight position and then the catheter is retracted into the sheath and removed from the patient.

The BASHIR™ Endovascular Catheter and BASHIR™ S-B Endovascular catheters are composed of common biocompatible materials used in vascular catheters are mechanical and contain no electrical components and require no capital equipment.

AI/ML Overview

I am sorry, but the provided text is a 510(k) premarket notification for a medical device (Bashir Endovascular Catheter) submitted to the FDA. It details information about the device, its intended use, comparison to predicate devices, and various performance tests (biocompatibility, sterility, bench testing, animal testing, and clinical testing).

However, this document does not contain information about an AI/ML powered device, specifically:

  • Acceptance criteria and reported device performance for an AI/ML model
  • Sample sizes for test sets and data provenance for an AI/ML model
  • Number and qualifications of experts for ground truth establishment for an AI/ML model
  • Adjudication methods for an AI/ML model
  • Multi-reader multi-case (MRMC) comparative effectiveness studies for an AI/ML model
  • Standalone performance of an AI/ML algorithm
  • Type of ground truth used for an AI/ML model
  • Sample size for the training set of an AI/ML model
  • How ground truth for the training set was established for an AI/ML model

Therefore, I cannot fulfill your request based on the provided text as it does not describe an AI-powered device or its associated study results.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.

April 20, 2023

Thrombolex, Inc. % Diane Horwitz Consultant Eminence Clinical Research Inc. 5 Lake Como Ct. Greenville, South Carolina 29609

Re: K222095

Trade/Device Name: BASHIRTM Endovascular Catheter, Ref. No. 7201 BASHIRTM S-B Endovascular Catheter, Ref. No. 7101 Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: QEY, KRA Dated: March 17, 2023 Received: March 17, 2023

Dear Diane Horwitz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Gregory W. Digitally signed by O'connell -S os:44:59 -04'00'

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222095

Device Name

BASHIR™ Endovascular Catheter, Ref. No. 7201 BASHIR™ S-B Endovascular Catheter, Ref. No. 7101

Indications for Use (Describe)

The BASHIR™ Endovascular Catheter and BASHIR™ S-B Endovascular Catheter are mechanical thrombolysis catheters indicated for the:

  • · Controlled and selective infusion of physician-specified fluids, including thrombolytics, into the pulmonary arteries for treatment of pulmonary embolism.
  • · Infusion of physician-specified fluids, including thrombolytics, into the peripheral vasculature, enabling the restoration of blood flow in patients with venous thrombus.
Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for THROMBOLEX. The logo is in red and consists of the word "THROMBOLEX" in all capital letters, with a stylized "X" at the end. Below the logo is the text "INNOVATIVE ENDOVASCULAR CATHETERS" in a smaller, sans-serif font. The logo appears to be for a medical device company specializing in catheters.

510(k) SUMMARY

GENERAL INFORMATION 1.

1.1 Submitter and 510(k) Owner

Thrombolex, Inc. 75 Britain Dr. New Britain PA 18901

1.2 Official Correspondent

Diane Horwitz, Ph.D. 5 Lake Como Ct. Greenville, SC 29609 Telephone: 703.307.2921 Email: dhorwitz(@ecr-inc.com

1.3 Date of Preparation

April 19, 2023

NAME OF THE DEVICE 2.

2.1.1 Trade/Proprietary Name

BASHIR™ Endovascular Catheter, Ref. No. 7201 BASHIR™ S-B Endovascular Catheter, Ref. No. 7101

Common/Usual Name 2.1.2

Continuous Flush Catheter

2.1.3 Classification Information

Classification Name: Mechanical Thrombolysis Catheter Classification Regulation: 21 CFR 870.5150 Class: 2 Product Code: QEY, KRA Panel: Cardiovascular

  1. PREDICATE DEVICE

Ekosonic Endovascular Device, Boston Scientific, K211080

Reference Devices: BASHIR™ Endovascular Catheter, K211061 BASHIR™ S-B Endovascular Catheter, K192598

{4}------------------------------------------------

4. DESCRIPTION OF THE DEVICE

The BASHIR™ Endovascular Catheter is a device intended for mechanical thrombolysis via expanding the distal infusion basket, allowing localized infusion of physician-specified fluids, including thrombolytics, into the peripheral or pulmonary vasculature. The distal infusion segment of the device is 12.50 cm (4.94 in.) long and consists of an expandable basket with a nitinol core surrounded by mini-infusion catheters, each with multiple infusion holes (hereafter referred to as the "infusion basket"). The infusion basket is expanded and closed using the red actuator located on the handle at the proximal end of the device. The BASHIR™ S-B Endovascular Catheter is a second model intended for smaller anatomies, with an infusion basket that is 10.0 cm (3.94 in.). The devices are otherwise the same.

The physician-specified fluid is then delivered via the basket infusion line located on the handle. Pulse sprays are followed by infusion through laser-drilled holes in the infusion limbs of the basket after placement in the peripheral vasculature or in the pulmonary artery. After the completion of fluid delivery, the infusion limbs are collapsed to a straight position and then the catheter is retracted into the sheath and removed from the patient.

The BASHIR™ Endovascular Catheter and BASHIR™ S-B Endovascular catheters are composed of common biocompatible materials used in vascular catheters are mechanical and contain no electrical components and require no capital equipment.

5. INTENDED USE / INDICATION FOR USE

The BASHIR™ Endovascular Catheter and BASHIR™ S-B Endovascular Catheter are mechanical thrombolysis catheters indicated for the:

  • . Controlled and selective infusion of physician-specified fluids, including thrombolytics, into the pulmonary arteries for treatment of pulmonary embolism.
  • Infusion of physician-specified fluids, including thrombolytics, into the peripheral . vasculature, enabling the restoration of blood flow in patients with venous thrombus.

INTENDED USE COMPARED TO THE PREDICATE 6.

The intended use of the BASHIR™ and the BASHIR™ S-B Endovascular Catheters is the same as that of the predicate device. Ekosonic Endovascular Device (K211080): They are mechanical thrombolysis catheters intended to infuse physician-specified fluids into a thrombus in either the peripheral or pulmonary circulation and for treatment of pulmonary embolism.

The indication for use for the subject devices has been expanded to include use in the pulmonary arteries to treat pulmonary embolism. The use of the BASHIR™ and BASHIR™ S-B Endovascular Catheters for treatment of pulmonary embolism were evaluated in the First-in-Human study and the pivotal RESCUE Clinical Trial, and the results support a substantially equivalent safety and effectiveness profile to the predicate device.

TECHNOLOGICAL CHARACTERISTICS COMPARED TO THE PREDICATE 7.

The BASHIR™ and the BASHIR™ S-B Endovascular Catheters have technological similarities to the predicate device (i.e., general catheter dimensions, catheter placement, compatibility with fluids, biocompatible nature of materials, radiopacity, and hydraulic mechanism for fluid dispersion). The subject and predicate devices differ in that the predicate device is a software-controlled device, whereas the subject devices are mechanical in nature and do not require capital equipment. The two reference devices (BASHIR™ and the BASHIR™ S-B Endovascular Catheters, K211061 and K 192598) are technologically identical to the subject devices.

{5}------------------------------------------------

8. PERFORMANCE TESTING

This 510(k) provides clinical performance data to support the expanded indication for use for treatment of pulmonary embolism and to establish the substantial equivalence of the BASHIR™ and the BASHIR™ S-B Endovascular Catheters to the predicate device.

Biocompatibility: Biocompatibility evaluation has been performed to show the finished, sterilized device is biocompatible and suitable for its intended use according to ISO 10993 and FDA Draft Guidance "Use of International Standard ISO 10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a Risk Management Process." Testing included cytotoxicity, sensitization, intracutaneous toxicity, acute systemic toxicity, hemolysis, complement, and pyrogenicity testing. All tests passed.

Sterility: Sterilization validation studies were conducted with ethylene oxide sterilization to a Sterilization Assurance Level (SAL) of 10° CFU according to international standards. Bacteriostasis and fungistasis and ethylene oxide residual testing passed. All sterilization process parameters met the established protocols and the requirements were satisfied.

Performance Testing - Bench: Thrombolex conducted design validation testing and human factors testing and showed that the subject devices meet product requirements and design specifications. Simulated use studies over 20 hours confirmed functionality of the subject device over 20 hours. Performance testing included an evaluation of: Kink radius, trackability, advancement force, slider actuator force, catheter retraction, radial force, delivery flow rate, infusion pressure through various lumens and at various flow rates, guidewire compatibility, dimensional verification, compliance of injection hubs, air leakage, fluid leakage, stress cracking, resistance to separation, torque strength, corrosion resistance, joint tensile strength and particulate generation.

Thrombolex performed an In-Use Study to evaluate the cumulative impact of in-use materials and conditions on the product quality attributes of a representative drug, before vs. after the fluid was infused through the subject device at a rate and concentration representative of those used in clinical trials. Testing demonstrated no impact on the device or the pharmacological fluid.

Performance Testing - Animal: Thrombolex performed a GLP animal study of deployment of the BEC in the peripheral vasculature and the pulmonary artery in a swine model. The BASHIR™ Endovascular Catheter had no adverse effects systemically, on gross or histopathology evaluation, and resulted in no animal mortality.

Performance Testing - Clinical: A First in Human (FIH) with 9 patients and pivotal study with 109 patients were conducted for the BASHIR™ Endovascular Catheter. Please refer to published literature for appropriate use of the BASHIR™ Endovascular Catheter in pulmonary endovascular cases.

CONCLUSIONS 9.

The information presented in this 510(k) submission demonstrates that the BASHIR™ and the BASHIR™ S-B Endovascular Catheters are substantially equivalent to the predicate device and supports the use of these devices in the pulmonary artery for treatment of pulmonary embolism and in the peripheral circulation for the restoration of blood flow in patients with venous thrombus.

§ 870.5150 Embolectomy catheter.

(a)
Identification. An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.(b)
Classification. Class II (performance standards).