K Number
K221800

Validate with FDA (Live)

Date Cleared
2022-12-14

(176 days)

Product Code
Regulation Number
880.2910
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Infrared Thermometer (Model: Y2001, Y20002) is intended for the intermittent measurement and monitoring of human body temperature from forehead. The device is indicated for use for people of two months and above at homecare and in hospital.

Device Description

Infrared thermometer (Model: Y20001 , Y20002) are a hand-held, battery powered, infrared Thermometer that coverts a user's forehead temperature, using the infrared energy emitted in the area around the user's forehead when measure from 1-3cm of the subject's forehead with no contact. The temperature will displayed on the screen, and the reference body site is axilla. Infrared thermometer (Model: Y20001, Y20002) use a thermopile sensor with integrated thermistor for the target reading and a thermistor mounted in the thermometer for ambient temperature readings. The device consists of measuring sensor, PCB, buttons, a LCD and an enclosure.

The difference between two models are size, weight and button settings. Y20001 has four buttons(Test button, °C/°F button, "B/S"button, "M" key) and Y20002 has two buttons(Function button, Measurement button). The functions of the two models are same. Infrared thermometer (Model: Y20001,Y20002) have the following features:

  1. The device is intended to be reusable for home use and clinical use.
  2. Switching of temperature unit between °C and °F.
  3. The latest 32 sets of memory for measuring human body and object; the user can view or delete the previous measurement results.
  4. Buzzer on or off to set the prompt tone on or off.
  5. Prompt tone function.
  6. Low battery indication, and auto power-off
AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the Infrared Thermometer (Models: Y20001, Y20002), based on the provided FDA 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

The core performance acceptance criteria for this medical device revolve around its accuracy in measuring body temperature. These are primarily derived from recognized international and national standards.

Acceptance CriteriaReported Device PerformanceReference Standard(s)
Measurement Range34.0°C to 42.9°C (93.2°F to 109.2°F)ISO 80601-2-56:2017, ASTM E1965-98 (2016)
Accuracy± 0.3°C / 0.5°FISO 80601-2-56:2017, ASTM E1965-98 (2016)
Display Resolution0.1°C / 0.1°FISO 80601-2-56:2017, ASTM E1965-98 (2016)
Clinical AccuracyComplied with ISO 80601-2-56:2017+AMD2018 requirementsISO 80601-2-56:2017+AMD2018
Electrical SafetyComplied with IEC 60601-1:2005+AMD 1:2012, IEC 60601-1-11:2015IEC 60601-1, IEC 60601-1-11
Electromagnetic Compatibility (EMC)Complied with IEC 60601-1-2:2014IEC 60601-1-2
BiocompatibilityNon-cytotoxic, non-sensitizing, non-irritatingISO 10993-5, ISO 10993-10, ISO 10993-23

Note on Accuracy: While the table states "± 0.3°C / 0.5°F", the comparison table (Note 4) mentions the subject device's accuracy being wider than the predicate within the range of 36.0°C - 39.0°C. However, the overall claim is that it still meets the performance standards based on the test reports. This implies that the ±0.3°C/0.5°F stated is the overarching acceptable tolerance.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Each clinical study evaluated 150 subjects.
  • Data Provenance: Not explicitly stated regarding the country of origin of the data. The studies were designed to evaluate subjects across specific age groups and a portion of fever patients, suggesting it was a prospective study conducted for regulatory submission.

The subjects were divided into three age groups with 50 subjects in each group:
* Infants: two months to one year
* Children: greater than one to five years
* Adults: greater than five years old

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not explicitly state the number of experts or their specific qualifications (e.g., "radiologist with 10 years of experience") used to establish the ground truth for the clinical accuracy test. However, for clinical thermometers, the "ground truth" (reference temperature) is typically obtained using a highly accurate, calibrated reference thermometer according to the specified standard (ISO 80601-2-56). The measurement itself, and hence the ground truth, relies on the precision of the reference device and adherence to standardized measurement protocols, rather than expert interpretation of an image or similar data.

4. Adjudication Method for the Test Set

Not applicable/Not mentioned. For a clinical thermometer's accuracy assessment, adjudication methods like 2+1 or 3+1 are typically used for subjective evaluations (e.g., image interpretation). Here, the comparison is quantitative: the device's reading versus a reference standard's reading.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not done. This type of study is more relevant for diagnostic imaging where human readers interpret results, and the AI's impact on their performance is being evaluated. For a thermometer, the performance is measured purely on its accuracy against a known standard.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, the primary clinical accuracy testing described is a standalone performance test of the device. The "clinical accuracy testing" evaluates the device's ability to measure human body temperature accurately as per ISO 80601-2-56 requirements, without human intervention affecting the measurement result itself. It compares the thermometer's reading directly to a reference method, acting as an algorithm-only performance assessment in the context of a medical device that outputs a direct measurement.

7. The Type of Ground Truth Used

The ground truth for the clinical accuracy test was established by comparing the infrared thermometer's readings to reference temperature measurements obtained according to the requirements of ISO 80601-2-56:2017. This typically involves a highly accurate, calibrated reference thermometer measuring core body temperature from a different site or a more invasive method, or using a specified (e.g., oral or rectal) reference thermometer, against which the non-contact device is compared. This is a type of objective measurement standard ground truth.

8. The Sample Size for the Training Set

The document does not provide any information about a training set. This device is a measurement tool, not an AI/ML algorithm that requires a separate training data set for model development. The performance is based on the physical design, sensors, and firmware logic, which are validated through bench and clinical testing.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as no training set is mentioned or implied for this type of device. The device's operational parameters and calibration would be established through engineering design and bench testing, not through supervised learning with a specific "training set" in the AI/ML sense.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and name on the right. The FDA acronym and name are in blue, with the acronym in a larger font size than the name. The name is written as "U.S. FOOD & DRUG ADMINISTRATION".

December 14, 2022

Zhangzhou Easepal Medical Science And Technology Co.,Ltd. % Jet Li Regulation Manager Guangdong Jianda Medical Technology Co., Ltd. 906 Room, Longxiang Garden, Tianhe District Guangzhou, Guangdong 510630 China

Re: K221800

Trade/Device Name: Infrared Thermometer (Model: Y20001, Y20002) Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: November 14, 2022 Received: November 14, 2022

Dear Jet Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

(24. SER. B. — 967) 1. hdl., (24. SER. B.

Page 2

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Danil Walloschek

David Wolloscheck, Ph.D. For Joyce M. Whang, Ph.D. Acting Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K221800

Device Name Infrared Thermometer (Model: Y20001, Y20002)

Indications for Use (Describe)

The Infrared Thermometer (Model: Y2001, Y20002) is intended for the intermittent measurement and monitoring of human body temperature from forehead. The device is indicated for use for people of two months and above at homecare and in hospital.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K221800

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 870.1130. Summary Prepared Date:14 December 2022

1. Submitter Information

  • Sponsor Company Name: Zhangzhou Easepal Medical Science And Technology Co.,Ltd. �
  • � Address: West of the 4th Floor, No. 7 Building, Easepal Factory, No. 228 Jiaosong Road, Jiaomei Town, Zhangzhou Taiwanese Investment Zone, Fujian, China
  • � Phone: +86-18905963518
  • Contact Person (including title): Flower Huang/ Engineer �
  • E-mail: flower@easepal.com.cn

Application Correspondent:

  • Guangdong Jianda Medical Technology Co.,Ltd. �
  • Address: 906 Room, Longxiang Garden, Tianhe district, Guangzhou, China �
  • Contact Person: Mr. Jet Li �
  • Tile: Regulation Manager �
  • � Tel: +86- 13512755282
  • Email: jianda-lee@foxmail.com

2. Subject Device Information

Common Name:Infrared Thermometer
Trade Name:Infrared Thermometer
Model names:Y20001,Y20002
Classification Name:Thermometer, electronic, clinical
Review Panel:General Hospital
Product Code:FLL
Regulation Number:21 CFR 880.2910
Regulation Class:II

{4}------------------------------------------------

3. Predicate Device Information

Sponsor:SHENZHEN EVERBEST MACHINERY INDUSTRY CO.,LTD.
Common Name:Infrared Thermometer
Trade Name:Infrared Thermometer
Classification Name:Clinical Electronic Thermometer
Model names:DT-8806S, DT-8807S
510(k) number:K203170
Review Panel:General Hospital
Product Code:FLL
Regulation Number:21 CFR 880.2910
Regulation Class:II

4. Device Description

Infrared thermometer (Model: Y20001 , Y20002) are a hand-held, battery powered, infrared Thermometer that coverts a user's forehead temperature, using the infrared energy emitted in the area around the user's forehead when measure from 1-3cm of the subject's forehead with no contact. The temperature will displayed on the screen, and the reference body site is axilla. Infrared thermometer (Model: Y20001, Y20002) use a thermopile sensor with integrated thermistor for the target reading and a thermistor mounted in the thermometer for ambient temperature readings. The device consists of measuring sensor, PCB, buttons, a LCD and an enclosure.

The difference between two models are size, weight and button settings. Y20001 has four buttons(Test button, °C/°F button, "B/S"button, "M" key) and Y20002 has two buttons(Function button, Measurement button). The functions of the two models are same. Infrared thermometer (Model: Y20001,Y20002) have the following features:

  1. The device is intended to be reusable for home use and clinical use.

  2. Switching of temperature unit between °C and °F.

  3. The latest 32 sets of memory for measuring human body and object; the user can view or delete the previous measurement results.

  4. Buzzer on or off to set the prompt tone on or off.

{5}------------------------------------------------

  1. Prompt tone function.
    1. Low battery indication, and auto power-off

5. Indication for use

The Infrared Thermometer (Model: Y20001, Y20002) is intended for the intermittent measurement and monitoring of human body temperature from forehead. The device is indicated for use for people of two months and above at homecare and in hospital.

6. Comparison to Predicate Device

Elements ofComparisonSubject DevicePredicate Device 1#Verdict
ManufacturerZhangzhou Easepal MedicalScience And Technology Co.,Ltd.SHENZHEN EVERBESTMACHINERY INDUSTRY CO.,LTD.--
510 (k) NumberK221800K203170--
Product NameInfrared ThermometerInfrared Thermometer--
ModelsY20001, Y20002DT-8806S, DT-8807S--
Intended UseThe Infrared Thermometer(Model: Y20001, Y20002) isintended for the intermittentmeasurement and monitoring ofhuman body temperature fromforehead. The device is indicatedfor use for people of two monthsand above at homecare and inhospital.The Infrared thermometer is intendedfor the intermittent measurement ofhuman body temperature fromforehead for people of all ages. Thedevice is reusable for home use andclinical use.Different,See Note 1
MeasurementMethodInfrared radiation detection,adjusted modeInfrared radiation detection, adjustedmodeSame
Principle ofoperationIt detects the radiant energy ofinfrared radiation emitted byobjects or human body andconverts it into electrical signals,and eventually into a temperaturevalueIt detects the radiant energy ofinfrared radiation emitted by objectsand converts it into electrical signals,and eventually into a temperaturevalueSame
MeasurementPlaceForeheadSurfaceForeheadSurfaceSame
Sensor typeThermopileThermopileSame
MeasurementDistance1-3cm8806S: 1 cm - 10 cm8807S: 1 cm-4 cmDifferent,See Note 2
MeasuringRange34.0°C to 42.9°C (93.2 to 109.2 °F)32.0°C -42.5°C (89.6 to 108.5 ° F)Different,See Note 3
Accuracy$\pm$ 0.3°C/0.5°F±0.2°C(0.4°F) within 36.0°C -39.0°C, (96.8°F - 102.2°F)Different,See Note 4
Elements ofComparisonSubject DevicePredicate Device 1#Verdict
±0.3°C(0.6°F) within 32°C -35.9°C(89.6°F -96.6°F) and 39.1 °C-42.5°C(102.3°F -108.5°F)
DisplayResolution0.1°C/0.1°F0.1°C/0.1°FSame
Display ScreenLCDLCDSame
BuzzerYesYesSame
°C/°FswitchableYesYesSame
Memory32 sets32 setsSame
Auto power-offSame while nooperationYesYesSame
Power Supply2 * 1.5V AAA2 * 1.5V AAASame
ContactmaterialsABSABSSame
OperatingConditionsTemperature:10°C-40°C;Relative humidity: ≤95%Temperature: 10.0°C~40°C (50.0°F~104.0°F);Relative humidity: ≤85%Different,See Note 5
StorageConditionsTemperature:-20°C-55°C;Relative humidity: ≤95%RH0 to 50°C (32 to 122°F) ;RH≤85%
DimensionY20001:145 mm * 93 mm * 38mmY20002:155 mm *40 mm *40mm128 * 74 * 36 mmDifferent,See Note 6
WeightY20001: 110 g(excludingbatteries )Y20002: 59g (excluding batteries)104.5 g
PerformanceCompliance with ASTME 1965;ISO80601-2-56Compliance with ASTME 1965;ISO80601-2-56Same
ClinicalaccuracyMeet the requirements of ISO80601-2-56:2017, ASTM 1965-98(2016)Meet the requirements of ISO 80601-2-56:2017, ASTM 1965-98(2016)Same
BiocompatibilityAll the patient contactingmaterials are compliance withISO 10993-1,ISO 10993-5,ISO 10993-10,ISO 10993-23.Under the condition of this studyAll the patient contacting materialsare compliance with ISO 10993.Under the condition of this study theSame
the device is non-cytotoxic , non-sensitizing and non-irritating.device is non-cytotoxic, non-sensitizing and non-irritating.
Elements ofComparisonSubject DevicePredicate Device 1#Verdict
Electrical SafetyCompliance with IEC 60601-1Compliance with IEC 60601-1Same
EMCCompliance with IEC 60601-1-2Compliance with IEC 60601-1-2Same

{6}------------------------------------------------

{7}------------------------------------------------

Note 1

The population of the predicate device is wider than the subject device. The user population of the subject device is subset of the predicate. Therefore, this difference does not raise new safety and effectiveness questions.

Note 2

The measurement distance is different. But we had conducted clinical accuracy testing on subject device according to ISO 80601-2-56:2017, and the clinical accuracy test result can demonstrate that the subject device can meet the requirement on this measurement distance. So, the different does not raise different questions of safety and effectiveness.

Note 3

The measurement range of subject device is different from the predicate device. The performance testing shows that the subject device complies with the performance standard ISO 80601-2-56 and ASTM E1965-98. Therefore, this difference does not affect the performance and accuracy.

Note 4

The subject device's accuracy are wider than the predicate device within 36.0℃ - 39.0℃, (96.8°F -102.2°F) ,these are based on device's performance standards. These differences do not affect the performance and accuracy according to our Test Report (ASTM E1965- 98), ISO 80601-2-56 Test Report, and our Clinical Test Report.

Note 5

The operation and storage environment of subject device is different from the predicate device, but the measurement accuracy of subject device has been demonstrated to comply with the requirements of standards IEC 60601-1 and ISO 80601-2-56 in operation and storage environment. Therefore, we considered that the difference does not raise any issues on the device safety and effectiveness.

Note 6

The "Dimensions", "Weight" of subject device is similar with predicate devices, both of them meet the requirement of safety and essential performance standard IEC 60601-1-2 and IEC 60601-1-2 and IEC 60601-1

{8}------------------------------------------------

  1. The differences between the predicate device and subject device will not affect the safety and effectiveness of the subject device.

7. Non-Clinical Test Summary

7.1. Electromagnetic Compatibility and Electrical Safety Test

The subject device has been tested in compliance with the following standards:

  1. AAMI/IEC 60601-1:2005+AMD 1: 2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

  2. IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests 3) IEC 60601-1-11:2015 Medical electrical equipment - Part 1-11: General requirementsfor basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

7.2. Biocompatibility Test

The subject device is considered for surface contacting, limited exposure (Less than 24 hours), it has been tested in compliance with the following standards:

  1. ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity

  2. ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization

  3. ISO 10993-23:2021 Biological evaluation of medical devices - Part 23:Tests for irritation

7.3. Performance Test-Bench

The subject device has been tested in compliance with the following standards:

  1. ISO 80601-2-56:2017+AMD2018 Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement 2) ASTM E1965-98 (Reapproved 2016) Standard Specification for InfraredThermometers for Intermittent Determination of Patient Temperature

7.4. Software Verification

The softw are documentation of the subject device was provided in accordance with FDA guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued May 11, 2005.

8. Clinical Accuracy Test Summary

{9}------------------------------------------------

Both models(Y20001, Y20002) have undergone clinical accuracy testing, both clinical studies are based on the requirement of ISO 80601-2-56:2017. The test reports showed the clinical performance of the subject device complied with the requirement of ISO 80601-2-56:2017+AMD2018 .

Each clinical study evaluated 150 subjects, and the infrared thermometer was evaluated for the patient populations for two months and above. All subjects were divided into three age groups with 50 subjects in each group, including (1)Infants –two months to one year, (2)Children - greater than one to five years; (3)Adults - greater than five years old. The proportion of fever patient is more than 30% in each age groups.

9. Conclusion

The subject device Infrared Thermometer (Model: Y20002) are substantially equivalent to the predicate device (K203170). This conclusion is based upon comparison of the intended use, technological characteristics and applicable safety standards. Any differences in the technological characteristics do not raise any new issues or different questions of safety or effectiveness.

§ 880.2910 Clinical electronic thermometer.

(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.