K Number
K221783

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-10-24

(125 days)

Product Code
Regulation Number
878.4850
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is designed for obtaining capillary blood sample.

Device Description

The SafetiCET Safety Lancet is single use only blood lancet used to puncture the skin to obtain a drop of blood for diagnostic purposes. It is a disposable blood lancet intended for a single use and the device is comprised of a needle encased in a housing made with plastic material commonly used for medical devices. The mechanism of the device includes an integral sharps injury prevention feature which allows the device to be used only once and then renders it inoperable and incapable of further use.

AI/ML Overview

The provided text describes the SafetiCET Safety Lancet and its substantial equivalence to a predicate device (VeriFine Safety Lancet) based on performance testing. However, it does not provide specific acceptance criteria values or detailed study results that prove the device meets these criteria in a quantitative sense for each test. Instead, it lists the types of performance tests conducted and generally states that the device's performance is comparable to the predicate or meets requirements.

Here's a breakdown of the requested information based on the provided text, with notes on what is not available:


Description of Acceptance Criteria and Study to Prove Device Meets Acceptance Criteria

The SafetiCET Safety Lancet is a single-use blood lancet designed to obtain capillary blood samples. Its substantial equivalence to the predicate device (VeriFine Safety Lancet, K221368) is asserted based on similarities in intended use, indications for use, sterilization methods, and performance characteristics.

The testing conducted aimed to confirm the safety and effectiveness of the device by evaluating various physical and mechanical properties relevant to its function and specified safety features. While specific numerical acceptance criteria for each test are not explicitly stated, the summary implies that the results of these tests were found to be acceptable and comparable to the predicate device's performance.

1. A table of acceptance criteria and the reported device performance

Performance Test CategoryAcceptance Criteria (Implicit from context)Reported Device Performance (from text)
SterilizationSterility Assurance Level (SAL) of 1$0^{-6}$ as per ISO 11137-1:2006Meets the Sterility Assurance Level of 1$0^{-6}$ as per the requirement of ISO 11137-1:2006.
Penetration DepthIn the same range as the needle length of the predicate device.Results are in the same range of the needle length of the predicate device.
Twist Break ForceAllows the protective cap to break the encapsulation and detach, preparing the device for activation.Measured as part of device performance assessment (implies it met functional requirements for cap detachment).
Drop Test(Not explicitly stated, but implies structural integrity and functionality retention after drops)Test conducted (implies satisfactory outcome for substantial equivalence).
Joint Strength of Needle-to-Needle Holder(Not explicitly stated, but implies sufficient strength to prevent needle detachment during use)Test conducted (implies satisfactory outcome for substantial equivalence).
Joint Strength of Top Cover to Housing(Not explicitly stated, but implies sufficient strength for proper assembly and function)Test conducted (implies satisfactory outcome for substantial equivalence).
Puncture Resistance(Not explicitly stated, but implies proper functioning of the needle)Test conducted (implies satisfactory outcome for substantial equivalence).
Needle RetractionAutomatic retraction into the housing after penetration to prevent re-use and needlestick injuries.Return Spring pushes the needle back into the housing preventing any needlestick issue. Needle cannot be pushed out for a second round.
Mechanical Properties Test of Material(Not explicitly stated, but implies materials meet specifications for intended use)Test conducted (implies satisfactory outcome for substantial equivalence).
BiocompatibilityCompliance with ISO 10993 series.ISO 10993 series testing performed.

Note: The document only lists the types of tests and generally states that the device's performance is comparable to or meets the requirements, rather than providing specific numerical acceptance thresholds or detailed performance metrics. For example, for "Penetration depth test," it says "results...are in the same range," but doesn't provide the range itself or the specific measurements.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size: Not specified for any of the performance tests.
  • Data Provenance: Not specified. It can be inferred that the testing was performed by NSP Tech Pte Ltd, based in Singapore, but details about the origin of test data (e.g., country, retrospective/prospective) are not provided.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. The performance tests described are mechanical and physical property tests of the medical device itself, not diagnostic performance tests requiring expert interpretation or ground truth establishment by clinical experts.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as this refers to clinical evaluation requiring adjudication (e.g., in cases of disagreement among expert readers). The tests described are objective, physical, and mechanical performance evaluations of the device.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a blood lancet, a mechanical instrument, not an AI-powered diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a blood lancet, a mechanical instrument, not an AI algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

Not applicable in the clinical sense. For the physical and mechanical tests, the "ground truth" would be established by objective measurements against predefined engineering specifications and relevant ISO standards (e.g., ISO 11137-1:2006 for sterility, ISO 10993 series for biocompatibility).

8. The sample size for the training set

Not applicable. This device is a physical medical instrument, not a machine learning model that requires a training set.

9. How the ground truth for the training set was established

Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.

10/24/2022

NSP Tech Pte Ltd Hc Tan Operations Manager 10 Admiralty Street, Northlink Building, #02-06 Singapore, 757695 Singapore

Re: K221783

Trade/Device Name: SafetiCET (SC-150);SafetiCET (SC-180);SafetiCET (SC-220);SafetiCET (SC-250);SafetiCET (SC-250B) Regulation Number: 21 CFR 878.4850 Regulation Name: Blood Lancets Regulatory Class: Class II Product Code: FMK Dated: August 29, 2022 Received: September 1, 2022

Dear Hc Tan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K221783

Device Name SafetiCET Safety Lancet

Indications for Use (Describe)The device is designed for obtaining capillary blood sample.
----------------------------------------------------------------------------------------------
Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K221783

510(k) Summary SafetiCET Blood Lancet

SUBMITTER: 1.

Applicant Name:

NSP Tech Pte Ltd 10 Admiralty Street, Northlink Building, #02-06 Singapore 757695

Contact Person:

Name: HC Tan
Phone: (65)98180450
Fax: (65)67476533
Email: hockchoon.tan@nsptech.com.sg
Establishment registration number

3008337059

Date prepared

: 21st October 2022

2. DEVICE

Trade Name:SafetiCET Safety Lancet
Common Name:Blood Lancet
Regulatory Class:Class II (special controls)
Classification Name:Single Use Only Blood Lancet with an IntegralSharps Injury Prevention Feature
Product Code:FMK
Regulation Number:21 CFR § 878.4850 (special controls)
Review Panel:General & Plastic Surgery
Device 510K number:K221783

PREDICATE DEVICE 3.

VeriFine Safety Lancet Regulatory Class II Product Code FMK General & Plastic Surgery Predicate device 510K number: K221368

DEVICE DESCRIPTION: 4.

The SafetiCET Safety Lancet is single use only blood lancet used to puncture the skin to obtain a drop of blood for diagnostic purposes. It is a disposable blood lancet intended for a single use and the device is comprised of a needle encased in a housing made with plastic material commonly used for medical devices. The mechanism of the device includes an integral sharps injury prevention feature which allows the device to be used only once and then renders it inoperable and incapable of further use.

{4}------------------------------------------------

NSP Tech Premarket Notification 510(k) Submission Under 21 CFR § 807.87 for SafetiCET

510(k) Summary SafetiCET Blood Lancet

The SafetiCET Safety Lancet is a sterile device and the sterilization is done using the Gamma Irradiation means.

5. INTENDED USE

The device is used for obtaining capillary blood sample.

6. INDICATION FOR USE STATEMENT

The device is designed for obtaining capillary blood sample.

7. SUBSTANTIAL EQUIVALENCE COMPARISON

TABLE OF COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE DEVICE

SafetiCET Safety LancetVeriFine Safety Lancet
510 (k) NumberK221783K 221368
ClassificationClass IIClass II
Product CodeFMKFMK
Intended UseThe device is designed forobtaining capillary blood sample.It is intended for capillary bloodsampling.
Indication ForUseThe device is designed forobtaining capillary blood samples.It is intended for capillary bloodsampling.
ManufacturerNSP Tech Pte LtdPromisemed Hangzhou MeditechCo., Ltd.
SterilizerGrandtenNot known
Needle Gauge21G, 23G, and 28G18G, 21G, 23G, 25G, 26G, 28G,30G

{5}------------------------------------------------

510(k) Summary SafetiCET Blood Lancet

SafetiCET Safety LancetVeriFine Safety Lancet
SterilizationGamma Radiation sterilizationGamma Radiation sterilization
SterilityMeet the Sterility Assurance Levelof 10-6 as per the requirement ofISO 11137-1:2006Not known
UseSingle-Use onlySingle-Use only
BiocompatibilityISO 10993 seriesISO 10993 series
Shelf Life5 yearsNot mentioned

8. TECHNOLOGICAL CHARACTERISTICS COMPARISON

The SafetiCET Safety Lancet has similarities with the predicate device in areas such as Intended Use, Indications For Use, Sterilization means.

The results of measuring penetration depth of the subject device are in the same range of the needle length of the predicate device. There is no significant difference in technological characteristics between the subject device and the predicate device.

Performance tests and comparison tests on the device quality confirmed the safety and effectiveness of the device.

In summary, the subject device and predicate device are said to have same technological characteristics.

{6}------------------------------------------------

510(k) Summary SafetiCET Blood Lancet

9. PRODUCT SAFETY AND EFFECTIVENESS DATA

Product SafetyDesign andEffectivenessThe housing of the device encases the Needle and device has aprotective cap attached to the housing. This combination prevents anyneedle stick issue with user or patient before the device is activated.The safety relating to needle stick issue is further enhanced with theneedle encapsulated by a plastic molded over it when residing in thehousing of the device. This encapsulation is only removed when theprotective cap is detached from the housing with a force known as theTwist Break Force preparing the device for activation and use. Theencapsulation of the needle also allows it to remain sterile after thedevice has undergone sterilization process. This prevents anyinadvertent pathogen contamination or contact to user or patient beforeuse.The device is activated by pushing the device on the blood collectionarea of the patient. Once activated, the needle is pushed out of thehousing by the Trigger Spring. After the device has performed thepiercing or penetration operation, the Return Spring pushes the needleback into the housing preventing any needlestick issue. The TriggerSpring helps in attaining the effectiveness of the device in providing thepenetration operation essential for the blood sample collection. TheReturn Spring helps in accomplishing the safety of the device byensuring that the needle retracts into the housing after the penetrationoperation is completed.
Test PrincipleBoth the subject device and the predicate device adopt the same testoperation principle of capillary blood sample collection. The device isfirst placed on the blood collection area of the patient and using contactactivation means, the automatic mechanism is activated with theneedle coming out of the device housing to make the piercing orpenetration operation. After the penetration operation is completed, theneedle automatically retracts into the housing. Because of theconstruction of the mechanism, the needle cannot be pushed out of thehousing for a second round of penetration preventing the device frombeing reused. After use, the device is to be discarded in a suitablebiohazard container.The blood collection operation is controlled for its depth of thepenetration. This is done by the design of the parts operating in thedevice. Hence, the device is checked for Penetration Depth as part ofits performance assessment.The protective cap is able to break the encapsulation of the needlewhen the twisting force is applied on the cap. This twisting force alsoallows the cap to be detached from the housing preparing the devicefor the penetration operation. Here, the Twist Break Force is measuredas part of device performance assessment.

{7}------------------------------------------------

510(k) Summary SafetiCET Blood Lancet

10. PERFORMANCE TESTING SUMMARY

The following performance testing done on the subject device is compared against that of the predicate device.

  • Sterilization testing and Validation ●
  • . Penetration depth test
  • Twist break force test ●
  • Drop test
  • Joint strength of needle-to-needle holder test ●
  • Joint strength of top cover to housing test ●
  • Puncture resistance test ●
  • Needle retraction test ●
  • Mechanical properties test of material

11. CONCLUSION

The comparison and evaluation of the device features, intended use, indication for use statement, and performance tests demonstrate substantial equivalence between the subject device and the predicate device. For the Safety and Effectiveness test, the subject device provides same safety and effective performance as the predicate device. From the evaluation above, we can deem that the Safety Lancet is substantially equivalent to the predicate device.

§ 878.4850 Blood lancets.

(a)
Single use only blood lancet with an integral sharps injury prevention feature —(1)Identification. A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base (including an integral sharps injury prevention feature) that is used to puncture the skin to obtain a drop of blood for diagnostic purposes. The integral sharps injury prevention feature allows the device to be used once and then renders it inoperable and incapable of further use.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and must include a sharps injury prevention feature.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use and that the integral sharps injury prevention feature will irreversibly disable the device after one use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device and its sharps injury prevention feature.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(b)
Single use only blood lancet without an integral sharps injury prevention feature —(1)Identification. A disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that the structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) Handwashing instructions for the user before and after use of the device.
(C) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(D) Instructions for the safe disposal of the device.
(E) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vi) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Discard the entire device after use.”
(B) “Warning: Not intended for more than one use. Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested.”
(c)
Multiple use blood lancet for single patient use only —(1)Identification. A multiple use capable blood lancet intended for use on a single patient that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class II (special controls). The special controls are:(i) The design characteristics of the device must ensure that:
(A) The lancet blade can be changed with every use, either manually or by triggering a blade storage unit to discard the used blade and reload an unused blade into the reusable base; and
(B) The structure and material composition are consistent with the intended use and address the risk of sharp object injuries and bloodborne pathogen transmissions and allow for validated cleaning and disinfection.
(ii) Mechanical performance testing must demonstrate that the device will withstand forces encountered during use.
(iii) The device must be demonstrated to be biocompatible.
(iv) Sterility testing must demonstrate the sterility of any device component that breaches the skin (
e.g., blade).(v) Validation testing must demonstrate that the cleaning and disinfection instructions are adequate to ensure that the reusable lancet base can be cleaned and low level disinfected.
(vi) Labeling must include:
(A) Detailed descriptions, with illustrations, of the proper use of the device.
(B) The Environmental Protection Agency (EPA) registered disinfectant's contact time for disinfectant use.
(C) Handwashing instructions for the user before and after use of the device.
(D) Instructions on preparation (
e.g., cleaning, disinfection) of the skin to be pierced.(E) Instructions on the cleaning and disinfection of the device.
(F) Instructions for the safe disposal of the device.
(G) Instructions for use must address the safe storage of the reusable blood lancet base between uses to minimize contamination or damage and the safe storage and disposal of the refill lancet blades.
(H) Labeling must be appropriate for the intended use environment.
(
1 ) For those devices intended for health care settings, labeling must address the health care facility use of these devices, including how these lancets are to be used with personal protective equipment, such as gloves.(
2 ) For those devices intended for use in the home, labeling must be written so that it is understandable to lay users.(vii) Labeling must also include the following statements, prominently placed:
(A) “For use only on a single patient. Disinfect reusable components according to manufacturer's instructions between each use.”
(B) “Used lancet blades must be safely discarded after a single use.”
(C) “Warning: Do not use on more than one patient. Improper use of blood lancets can increase the risk of inadvertent transmission of bloodborne pathogens, particularly in settings where multiple patients are tested. The cleaning and disinfection instructions for this device are intended only to reduce the risk of local use site infection; they cannot render this device safe for use for more than one patient.”
(d)
Multiple use blood lancet for multiple patient use —(1)Identification. A multiple use capable blood lancet intended for use on multiple patients that is comprised of a single use blade attached to a solid, reusable base that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.(2)
Classification. Class III (premarket approval).(3)
Date PMA or notice of completion of a PDP is required: A PMA or a notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before May 22, 2024, for any multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976, or that has, on or before May 22, 2024, been found to be substantially equivalent to a multiple use blood lancet for multiple patient use described in paragraph (d)(1) of this section that was in commercial distribution before May 28, 1976. Any other multiple use blood lancet for multiple patient use shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.