K Number
K221068

Validate with FDA (Live)

Date Cleared
2022-10-04

(175 days)

Product Code
Regulation Number
880.5860
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sterile Hypodermic Syringes for Single Use with Needles is intended for use in the aspiration of fluids for medical purpose.

Device Description

The Sterile Hypodermic Syringes for Single Use with Needles is intended for manual and single use only to aspirate and inject of fluids for medical purpose, which consists of piston, barrel, plunger and a hypodermic needle. The proposed device is available in a variety combination of syringe volume and needle size.

AI/ML Overview

The provided text describes a 510(k) premarket notification for "Sterile Hypodermic Syringes for Single Use with Needles." This document focuses on demonstrating substantial equivalence to a predicate device, rather than proving the device meets specific acceptance criteria through a clinical study or a standalone algorithm evaluation.

Therefore, many of the requested sections about acceptance criteria, clinical studies, reader studies, and ground truth establishment are not applicable to this type of regulatory submission and the provided document.

Here's an analysis based on the available information:


1. A table of acceptance criteria and the reported device performance

The document does not present a table of acceptance criteria and reported device performance in the way a clinical study or standalone algorithm study would. Instead, it demonstrates compliance with recognized international standards for medical devices. The "reported device performance" is essentially that the device complied with these standards.

Acceptance Criteria (Standards Met)Reported Device Performance
ISO 7886-1:2017 (Syringes for Manual Use)Complied with ISO 7886-1 (Syringe Performance)
ISO 80369-7:2016 (Small-bore connectors)Complied with ISO 80369-7 (Luer Connector Performance)
ISO 80369-20:2015 (Common test methods for connectors)Complied with ISO 80369-20
ISO 7864:2016 (Sterile Hypodermic Needles)Complied with ISO 7864 (Needle performance)
ISO 9626:2016 (Stainless Steel Needle Tubing)Complied with ISO 9626 (Needle performance)
ISO 10993-1 (Biocompatibility)Biocompatibility testing conducted, no adverse effects shown
USP <788> (Particulate Matter in Injections)Met USP acceptance criteria for particulate matter
ISO 11135:2014 (EO Sterilization)Validated to achieve SAL of 10^-6
ISO 10993-7:2008 (EO and ECH residuals)Residuals below specified limits
USP <85> (Bacterial Endotoxin Levels)Below level of 20 EU/device
ASTM F1886 / F1886M-16 (Visual Inspection)Performed
ASTM F88/F88-15 (Seal Strength)Performed
ASTM F1929-15 (Dye penetration)Performed
ASTM D4169-16 (Simulated Transportation)Performed
ASTM F1980-16 (Accelerated Aging for Shelf-life)Shelf-life of 5-years validated

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the document. The document describes non-clinical laboratory testing (e.g., physical performance, biocompatibility, sterilization validation) but does not detail the sample sizes for these tests or the data provenance. The tests are general in nature for medical device components and manufacturing processes, not related to patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is not applicable to this submission. Ground truth established by medical experts (like radiologists) is relevant for diagnostic devices or AI algorithms that interpret medical data. This submission is for a physical medical device (syringes and needles). The "ground truth" here is compliance with engineering and material standards, which is assessed through laboratory testing by qualified personnel, not by medical experts establishing diagnostic ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This is not applicable to this submission. Adjudication methods are typically used in clinical trials or reader studies where multiple human readers or algorithms provide interpretations that need to be resolved to establish a definitive ground truth. The testing described here is based on objective measurements against established international standards.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic devices or decision support systems. The device in question is a sterile hypodermic syringe with needles, which is a physical device used for aspiration and injection, not an AI or diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

A standalone performance study for an algorithm was not done. This device is not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for the device's performance is established by international standards and regulations for the physical, chemical, and biological properties of sterile hypodermic syringes and needles. This includes:

  • Physical performance requirements (e.g., according to ISO 7886-1, ISO 7864, ISO 9626)
  • Biocompatibility (e.g., according to ISO 10993-1)
  • Sterility Assurance Level (SAL) (e.g., according to ISO 11135)
  • Particulate matter levels (e.g., according to USP <788>)
  • Endotoxin levels (e.g., according to USP <85>)

These are objective, measurable criteria, not subjective interpretations requiring expert consensus or pathology reports.

8. The sample size for the training set

This is not applicable. This device is a physical medical product, not an AI algorithm that requires a training set.

9. How the ground truth for the training set was established

This is not applicable. As stated above, this device is not an AI algorithm and therefore does not have a "training set" or "ground truth for a training set."

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 4, 2022

Sichuan Prius Biotechnology Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O.box 120-119 Shanghai, 200120 China

Re: K221068

Trade/Device Name: Sterile Hypodermic Syringes for Single Use with Needles Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF, FMF, FMI Dated: August 26, 2022 Received: September 8, 2022

Dear Diana Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809 ); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

CAPT Alan Stevens Assistant Director Injection Team Devision of Drug Delivery and General Hospital and General Hospital Devices Office of GastroRenal , OBGYN, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K221068

Device Name

Sterile Hypodermic Syringes for Single Use with Needles

Indications for Use (Describe)

The Sterile Hypodermic Syringes for Single Use with Needles is intended for use in the aspiration of fluids for medical purpose.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K221068 -510(k) Summary

    1. Date of Preparation: 10/08/2022
    1. Sponsor Identification

Sichuan Prius Biotechnology Co., Ltd.

No.2 Prius Road, Luo Long Industrial Park Nanxi District, 644104 Yibin City, Sichuan Province, PEOPLE'S REPUBLIC OF CHINA

Establishment Registration Number: Not registered yet

Contact Person: Yan Liu Position: Management Representative Tel: +86-831-3839889 Fax: +86-831-3839887 Email: 48363603@qq.com

    1. Designated Submission Correspondent
      Ms. Diana Hong (Primary Contact Person) Ms. Tingting Su (Alternative Contact Person)

Mid-Link Consulting Co., Ltd

P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net

{4}------------------------------------------------

Identification of Proposed Device 4.

Trade Name: Sterile Hypodermic Syringes for Single Use with Needles Common Name: Syringe with Needle

Regulatory Information Classification Name: Syringe, Piston Classification: II; Product Code: FMF; Regulation Number: 21CFR 880.5860; Review Panel: General Hospital;

Classification Name: Hypodermic, Syringe Lumen Classification: II; Product Code: FMI; Regulation Number: 21CFR 880.5570; Review Panel: General Hospital;

    1. Identification of Predicate Device
      Predicate Device 510(k) Number: K211329 Product Name: Sterile Disposable Syringes with Needles
  • Device Description 6.
    The Sterile Hypodermic Syringes for Single Use with Needles is intended for manual and single use only to aspirate and inject of fluids for medical purpose, which consists of piston, barrel, plunger and a hypodermic needle. The proposed device is available in a variety combination of syringe volume and needle size.

Syringe VolumeNeedle Sizes
Needle Length (mm)Needle gaugeWall
1ml,3216GTW
2ml,3816GTW
3ml,3218GTW
5ml,3818GTW
10ml,3219GTW
20ml,3819GTW
30ml,3220GTW
50ml,3820GTW

{5}------------------------------------------------

60ml,100ml3221GTW
3821GTW
3222GTW
3822GTW
1923GTW
2523GTW
3223GTW
1924GRW
2524GRW
3224GRW
1625GRW
1925GRW
2525GRW
1326GRW
1626GRW
1926GRW
1327GRW
1627GRW
1927GRW
1328GRW
1628GRW
1329GRW
1629GRW
1330GRW
1630GRW

All gauges can be used with 1mL, 2mL, 5mL, 10mL, 20mL, 30mL, 50mL, 60mL, 100mL syringe.

The proposed devices are sterilized by Ethylene Oxide Gas to achieve a SAL of 10th and supplied sterility maintenance package which could maintain the sterility of the device during the shelf life of 5 years.

    1. Indication for Use:
CharacterizationProposed devicePredicate Device
Sterile Hypodermic Syringes forSingle Use with NeedlesK221068Sterile Disposable Syringe withNeedleK211329
Indication for UseThe Sterile Hypodermic Syringesfor Single Use with Needles isThe Sterile Disposable Syringe withNeedle is intended for use in the

{6}------------------------------------------------

intended for use in the aspirationaspiration and injection of fluids for
and injection of fluids for medicalmedical purpose.
purpose.
Prescription or OTC (overthe counter)Prescription usePrescription use

Technology Characteristics 8.

The table below includes a comparison of the technological characteristics between the new device and those of the predicate.

TechnologicalCharacteristicProposed DevicePredicate DeviceRemark
Sterile Hypodermic Syringesfor Single Use with NeedlesSterile Disposable Syringewith Needle
K221068K211329
ConfigurationBarrelBarrelSame
PlungerPlunger
PistonPiston
Needle hubNeedle hub
Protective capProtective cap
NeedleNeedle
OperationModeFor manual use onlyFor manual use onlySame
Single UseSingle UseSingle UseSame
Volume1ml, 2ml, 3ml, 5ml, 10ml,20ml, 30ml, 50ml, 60ml,100ml1ml, 2ml, 3ml, 5ml, 10ml,20ml, 30ml, 50ml, 60mlDifferentSeecomment#1
ConnectorTypeLuer Lock/ Luer SlipLuer Lock/ Luer SlipSame
SyringePerformanceComplied withISO 7886-1Complied withISO 7886-1Same
LuerConnectorPerformanceComplied withISO 80369-7Complied withISO 80369-7Same
NeedlePerformanceComplied withISO 7864,ISO 9626Complied withISO 7864,ISO 9626Same
Needle Gauge16G, 18G, 19G, 20G, 21G,23GDifferent

Table 1 Comparison of Technology Characteristics

{7}------------------------------------------------

22G, 23G, 24G, 25G, 26G, 27G, 28G, 29G, 30GSee comment #2
Needle Length13mm, 16mm, 19mm, 25mm, 32mm, 38mm13mm, 16mm, 20mm, 25mm, 32mm, 38mmDifferent See comment #3
Wall typeTW: 16G, 18G, 19G, 20G, 21G, 22G, 23GRW: 24G, 25G, 26G, 27G, 28G, 29G, 30GTWDifferent See comment #4
Bevel DesignLB/SBLB/SBSame
Syringe HubsCentral/EccentricCentral/EccentricSame
Material
BarrelPolypropylene (PP)Polypropylene (PP)Same
PlungerPolypropylene (PP)Polypropylene (PP)Same
PistonPolyisoprenePolyisopreneSame
NeedleStainless Steel SUS304Stainless Steel SUS304Same
Needle hubPolypropylene (PP)Polypropylene (PP)Same
Needle capPolypropylene (PP)Polypropylene (PP)Same
LubricantsPolydimethylsiloxaneSilicone oilDifferent See comment #5
AdhesiveEpoxy adhesiveUV adhesiveDifferent See comment #5

Discussions of differences in technological characteristics

Comment #1 – Syringe Volume

The proposed device has the additional 100ml syringe volume which is out of the volume range of predicate device. This does not change the intended use and does not raise new questions of safety and effectiveness. In addition, the syringe performance has been tested and test results demonstrate that the syringe meets the requirements of ISO 7886.

Comment #2 – Needle Length

The range of needle lengths is the same as the predicate. The predicate device does not have a 19mm needle length. This difference is not a new intended use and does not raise new questions of safety and

{8}------------------------------------------------

effectiveness. In addition, the needle performance has been tested and the results demonstrate that the needle meets the requirements of ISO 7864 and ISO 9626.

Comment #3 – Needle Gauge

The subject device has the more gauges compared to the predicate device. This difference is not a new intended use and does not raise new questions of safety and effectiveness. In addition, all the needle gauges of proposed device has been tested. The needle performance has been tested and results demonstrate that the needle meets the requirements of ISO 7864 and ISO 9626. Therefore, this difference will not raise new questions on safety and effectiveness of the proposed device.

Comment #4- Wall Type

The proposed device is available in thin wall and normal wall two types, the predicate device only has thin wall type. This difference is not a new intended use and does not raise new questions of safety and effectiveness. In addition, the all the wall type of proposed device has been tested and results demonstrate that the needle meets the requirements of ISO 7864 and ISO 9626.

Comment #5 - Material of Lubricants and Adhesive

The proposed device has different adhesive and lubricants material compared to the predicate device However, biocompatibility test has been conducted on the proposed device and the test result does not show any adverse effect. This difference is not a new intended use and does not raise new questions of safety and effectiveness.

9. Non-Clinical Test Conclusion

The device described in this summary Sterile Hypodermic Syringes for Single Use with Needles, were tested and demonstrated to be in conformance with the following FDA recognized standards:

  • ISO 7886-1: 2017 Sterile Hypodermic Syringes for Single Use - Part 1: Syringes for Manual Use

  • ISO 80369-7: 2016 Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications

  • ISO 80369-20:2015 Small-bore connectors for liquids and gases in healthcare applications-Part 20: Common test methods

  • ISO 7864: 2016, Sterile Hypodermic Needles for Single Use. >
  • ISO 9626:2016 Stainless Steel Needle Tubing for the Manufacture of Medical Devices A

Biocompatibility

In accordance with ISO10993-1 the device are classified as: Externally Communicating Device, Blood Path Indirect, Limited Contact (< 24hrs). The following testing was conducted:

  • Cytotoxicity

  • Irritation A
  • A Skin Sensitization

{9}------------------------------------------------

  • Acute Systemic Toxicity

  • Pyrogenicity

  • Hemolysis

  • Material Mediated Toxicity

Particulate matter testing was conducted in accordance with USP <788> Particulate Matter in Injections and met the USP acceptance criteria.

Sterility, Shipping, and Shelf -life

The proposed device sterilization process using Ethylene Oxide (EO) has been validated in accordance with ISO 11135:2014 to achieve a sterility assurance level (SAL) of 10th. EO and Ethylene Chlorohydrin (ECH) residuals were below the limits specified in ISO 10993-7:2008. Bacterial Endotoxin Levels were below the level of 20 EU/device in accordance with USP <85>. Both baseline and accelerated shelf life testing were conducted demonstrating the device will perform as intended to support the proposed 5 year shelf-life.

  • . Sterile Barrier Packaging performed on the proposed device:
    • Visual Inspection ASTM F1886 / F1886M-16
    • 0 Seal Strength ASTM F88/F88-15
    • o Dye penetration ASTM F1929-15
  • . Simulated transportation testing in accordance to ASTM D4169-16 on final, packaged, and sterile device.
  • Shelf-life of 5-years is validated using FDA recognized standard ASTM F1980-16 Standard . Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices.

10. Clinical Test Conclusion

No clinical study is included in this submission.

11. Conclusion

The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. The Sterile Hypodermic Syringes for Single Use with Needles is as safe, as effective, and performs as well as the legally marketed predicate device cleared under K211329.

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).