K Number
K220949

Validate with FDA (Live)

Date Cleared
2022-10-27

(209 days)

Product Code
Regulation Number
866.3175
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ARCHITECT CMV IgG assay is a chemiluminescent microparticle immunoassay (CMIA) used for the qualitative detection of IgG antibodies to cytomegalovirus in human serum, serum separator, and plasma tubes (lithium heparin, lithium heparin separator, and tripotassium EDTA) on the ARCHITECT i System.

The ARCHITECT CMV IgG assay is to be used as an aid in the diagnosis of infection with cytomegalovirus and as an aid in the determination of serological status to cytomegalovirus in individuals including women of child-bearing age.

The ARCHITECT CMV IgG assay has not been cleared for use in screening blood, plasma, or tissue donors.

Device Description

The ARCHITECT CMV IgG assay is a chemiluminescent microparticle immunoassay (CMIA) for the qualitative detection of IgG antibodies to cytomegalovirus. The reagent kit contains microparticles coated with CMV virus lysate, murine anti-human IgG acridinium-labeled conjugate, and assay diluent. The assay is performed on the ARCHITECT i System. The presence or absence of anti-CMV IgG is determined by comparing the chemiluminescent relative light unit (RLU) in the reaction to a cutoff RLU from an active calibration.

AI/ML Overview

Here's an analysis of the acceptance criteria and study information for the ARCHITECT CMV IgG device, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly stated as numerical targets in the document. However, based on the non-clinical and clinical performance summaries, we can infer performance goals for the device's accuracy and precision. The "Predicate Device" (ADVIA Centaur CMV IgG Assay, K181213) serves as a benchmark for substantial equivalence.

Performance MetricAcceptance Criteria (Inferred from Study Design & Predicate Equivalence)Reported Device Performance (ARCHITECT CMV IgG)
Within-Laboratory Precision (Repeatability)Low variability; %CV values typically < 5-10% for quantitative assays, ideally < 5%Controls: Negative Control (NA), Positive Control 1 (2.8%). Panels: Panel 1 (NA), Panel 2 (NA), Panel 3 (NA), Panel 4 (2.7%), Panel 5 (2.1%), Panel 6 (2.1%).
Within-Laboratory Precision (Total Variability)Low variability; %CV values typically < 5-10% for quantitative assays, ideally < 5%Controls: Negative Control (NA), Positive Control 1 (3.0%). Panels: Panel 1 (NA), Panel 2 (NA), Panel 3 (NA), Panel 4 (2.9%), Panel 5 (2.6-2.8%), Panel 6 (2.4-2.5%).
System ReproducibilityLow variability across sites, lots, and days; %CV values typically < 5-10%Controls: Negative Control (NA), Positive Control 1 (4.5%). Panels: Panel 1 (NA), Panel 2 (NA), Panel 3 (NA), Panel 4 (4.8%), Panel 5 (4.2%), Panel 6 (3.9%).
Analytical Specificity (Interference)No significant interference from common endogenous substances or drugs; generally defined as impact within a specific threshold (e.g., +/- 0.6 AU/mL for low concentrations, +/- 10% for high concentrations).Endogenous Substances: No significant interference observed for unconjugated/conjugated bilirubin, hemoglobin, total protein, and triglycerides (up to 1650 mg/dL). Interference (>+0.6 AU/mL) noted for Triglycerides at 2475 mg/dL on 4.0 AU/mL sample.
Drugs/Other Substances: No significant interference observed for Acetaminophen, Ascorbic Acid, Beta Carotene, Biotin, Cidofovir, Diphenhydramine, Folic Acid, Foscarnet, Gangciclovir, Ibuprofen, Valganciclovir.
Analytical Specificity (Other Conditions)Low rate of false positives/reactives with specimens from individuals with unrelated medical conditions.1 reactive result out of 187 specimens tested across various medical conditions (Toxoplasmosis (IgG) specimen was reactive). 1 equivocal parvovirus B19 (IgG) specimen.
CDC Panel Agreement (Positive)High positive percent agreement, ideally >90%100% (91.59%, 100%)
CDC Panel Agreement (Negative)High negative percent agreement, ideally >90%92.11% (78.62%, 98.34%)
CDC Panel Agreement (Overall)High overall percent agreement, ideally >90%96.25% (89.43%, 99.22%)
Clinical Percent Agreement (Positive) - Routine OrderHigh positive percent agreement with comparator, ideally >90%97.7% (96.1%, 98.7%)
Clinical Percent Agreement (Negative) - Routine OrderHigh negative percent agreement with comparator, ideally >90%99.2% (97.3%, 99.8%)
Clinical Percent Agreement (Positive) - Pregnant FemalesHigh positive percent agreement with comparator, ideally >90%99.0% (94.5%, 99.8%)
Clinical Percent Agreement (Negative) - Pregnant FemalesHigh negative percent agreement with comparator, ideally >90%100.0% (96.4%, 100.0%)

2. Sample Size Used for the Test Set and Data Provenance

  • Within-Laboratory Precision (20-Day): 117-120 replicates per panel/control across 3 reagent lots/calibrator lot combinations.
  • Analytical Specificity (Interference): Not explicitly stated, but each substance was tested at 3 analyte levels.
  • Analytical Specificity (Other Conditions): 187 specimens.
  • CDC Panel Agreement: 80 samples (either CMV IgG negative or CMV IgG positive). Data provenance is the Centers for Disease Control and Prevention (CDC). The document doesn't specify if this was retrospective or prospective, but the description "masked, characterized serum panel" suggests it was a curated, retrospective panel.
  • Clinical Percent Agreement:
    • Routine Order: 591 specimens collected in the US and 198 specimens collected outside of the US.
    • Pregnant Females: 200 specimens collected in the US.
    • Further Evaluation: 4 specimens (3 routine order, 1 pregnant female) with equivocal/grayzone results by the comparator assay.
    • Data provenance for the clinical study is multi-site (Indianapolis Indiana, Lewisville Texas, and Palo Alto California) for the US samples, and unspecified international locations for the "outside of the US" routine order specimens. This was a clinical study (method comparison), suggesting prospective collection relative to the comparison against the investigational assay.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

There is no information provided about experts being used to establish a ground truth for the test set in the conventional sense (e.g., for image interpretation).

  • For the CDC Panel Agreement, the CDC provided "their result interpretation for each sample," implying the CDC's own characterization (likely by reference methods/consensus) served as the ground truth. No specific number or qualifications of experts are given.
  • For the Clinical Percent Agreement, the ground truth was established by a "current, FDA-cleared, commercially available anti-CMV IgG assay" (the comparator assay), essentially establishing a reference standard from an already approved diagnostic device. For 4 equivocal/grayzone samples, "consensus testing" with "2 additional current, FDA-cleared, commercially available anti-CMV IgG assays" was used. The document does not mention human experts establishing ground truth for these studies.

4. Adjudication Method for the Test Set

  • CDC Panel Agreement: The results were compared directly to CDC's result interpretation. No explicit adjudication method is described beyond this direct comparison.
  • Clinical Percent Agreement: The primary comparison was against a single comparator assay. For specific equivocal/grayzone cases (4 specimens), a form of adjudication was used where the result was "negative by consensus testing" or "negative based on the consensus result from the comparator assay and 2 additional current, FDA-cleared, commercially available anti-CMV IgG assays." This implies a form of multi-comparator consensus for ambiguous cases, but not necessarily human expert adjudication in the traditional sense.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study was done. This device is an automated in vitro diagnostic (IVD) assay (chemiluminescent microparticle immunoassay), not an AI-powered diagnostic imaging device or tool that assists human readers. Therefore, the concepts of human readers, AI assistance, and effect size in improving human reader performance are not applicable.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

The device itself, the ARCHITECT CMV IgG assay on the ARCHITECT i System, is a standalone algorithm/device for qualitative detection. Its performance is evaluated entirely as an automated system without human interpretation in the loop influencing the direct output of IgG antibody detection. The "human-in-the-loop" would be the clinician interpreting the result provided by the device (e.g., Reactive, Nonreactive, Grayzone/Equivocal) in the context of a patient's overall clinical picture, but not in generating the result itself.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • CDC Panel Agreement: Ground truth was based on CDC's own "result interpretation" of their characterized serum panel. This likely represents a highly robust reference method or internal consensus.
  • Clinical Percent Agreement: Ground truth was primarily established by a "current, FDA-cleared, commercially available anti-CMV IgG assay" (comparator assay). For a few ambiguous cases, consensus results from multiple FDA-cleared comparator assays were used. This is a form of reference method comparison.

8. The Sample Size for the Training Set

The document does not explicitly mention a "training set" in the context of machine learning or AI. As this is a traditional immunoassay, it undergoes development and validation using various biological samples. However, no specific number is provided for samples used during the development or optimization phases that could be analogized to a training set. The clinical study samples described are specifically for performance evaluation (test set).

9. How the Ground Truth for the Training Set Was Established

As no explicit training set is defined (because it's not an AI/ML device), this question is not applicable. The assay's parameters (e.g., cutoff values) would have been established during product development using a large set of characterized samples and optimized for performance, but this is a different process than establishing ground truth for an AI training set.

{0}------------------------------------------------

October 27, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below that.

Abbott Laboratories Shannon Reibling Regulatory Affairs Project Manager Dept 09AA, Bldg. AP8A, 100 Abbott Park Rd. Abbott Park, Illinois 60064

Re: K220949

Trade/Device Name: Architect CMV IgG Regulation Number: 21 CFR 866.3175 Regulation Name: Cytomegalovirus Serological Reagents Regulatory Class: Class II Product Code: LFZ Dated: March 31, 2022 Received: April 1, 2022

Dear Shannon Reibling:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Maria Garcia, Ph.D. Assistant Director Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

Device Name ARCHITECT CMV IgG ARCHITECT CMV IgG Calibrators ARCHITECT CMV IgG Controls

Indications for Use (Describe) ARCHITECT CMV IgG

The ARCHITECT CMV IgG assay is a chemiluminescent microparticle immunoassay (CMA) used for the qualitative detection of IgG antibodies to cytomegalovirus in human serum, serum separator, and plasma tubes (lithium heparin, lithium heparin separator, and tripotassium EDTA) on the ARCHITECT i System.

The ARCHITECT CMV IgG assay is to be used as an aid in the diagnosis of infection with cytomegalovirus and as an aid in the determination of serological status to cytomegalovirus in individuals including women of child-bearing age.

The ARCHITECT CMV IgG assay has not been cleared for use in screening blood, plasma, or tissue donors.

ARCHITECT CMV IgG Calibrators

The ARCHITECT CMV IgG Calibrators are for the calibration of the ARCHITECT i System when used for the qualitative detection of IgG antibodies to cytomegalovirus in human serum and plasma.

ARCHITECT CMV IgG Controls

The ARCHITECT CMV IgG Controls are for the estimation of test precision and the detection of systematic analytical deviations of the ARCHITECT i System when used for the qualitative detection of IgG antibodies to cytomegalovirus in human serum and plasma.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5: 510(k) Summary (Summary of Safety and Effectiveness)

This summary of the 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.

I. Applicant Name

Abbott Diagnostics Department 09AA, Building AP8A, 100 Abbott Park Road Abbott Park, IL 60064

Primary contact person for all communications:

Shannon Reibling, Regulatory Affairs Project Manager Abbott Diagnostics Division Telephone Number: (224) 668-3735 Fax Number: (224) 667-4836

Jacob Richards, Associate Director, Regulatory Affairs Abbott Diagnostic Division Telephone Number: (224) 668-5877 Fax Number: (224) 667-4836

{4}------------------------------------------------

II. Device Name

ARCHITECT CMV IgG

Reagents

Trade Name: ARCHITECT CMV IgG Reagent Kit Device Classification: Class II Classification Name: Cytomegalovirus serological reagents Governing Regulation: 21 CFR § 866.3175 Code: LFZ

Calibrator

Trade Name: ARCHITECT CMV IgG Calibrator Device Classification: Class II, 510(k) Exempt Classification Name: Calibrator Governing Regulation: 21 CFR § 862.1150 Code: JIT

Controls

Trade Name: ARCHITECT CMV IgG Controls Device Classification: Class I, 510(k) Exempt Classification Name: Control Governing Regulation: 21 CFR § 862.1660 Code: JJX

III. Predicate Device

ADVIA Centaur CMV IgG Assay (K181213)

{5}------------------------------------------------

IV. Description of Device

Reagents

The ARCHITECT CMV IgG reagent kit contains:

  • . Microparticles: (1 bottle x 6.6 mL per 100-test / 1 bottle x 27.0 mL per 500-test) CMV virus lysate (strain AD169) coated microparticles in TRIS buffered saline with protein (bovine). Minimum concentration: 0.08% solids. Preservatives: ProClin 300 and antimicrobial agents.
  • . Conjugate: (1 bottle x 5.9 mL per 100-test / 1 bottle x 26.3 mL per 500-test). Murine anti-human IgG acridinium-labeled conjugate in MES buffer with protein (bovine). Minimum concentration: 44 ng/mL. Preservatives: sodium azide and antimicrobial agents.
  • Assay Diluent: (1 bottle x 10.0 mL per 100-test / 1 bottle x 50.9 mL per 500-test). . Calf serum and MES buffer with protein (bovine). Preservatives: ProClin 300 and ProClin 950.

Calibrators

The ARCHITECT CMV IgG Calibrators:

  • Calibrator A - 1 Bottle (4.0 mL): contains recalcified human plasma with protein (ovine) stabilizer. Preservative: sodium azide.
  • Calibrators B through F 5 bottles (4.0 mL each): contain recalcified human plasma . and are reactive for IgG antibodies to cytomegalovirus (anti-CMV IgG). Preservative: sodium azide.

Calibrators cover the calibration range of the assay (0.0 to 250.0 AU/mL). The calibrators are at the following anti-CMV IgG concentrations:

CalibratorTarget Anti-CMV IgG Concentration (AU/mL)
A0.0
B10.0
C50.0
D75.0
E125.0
F250.0

{6}------------------------------------------------

The Calibrators B through F are referenced to internal reference standards at each concentration level.

Controls

The ARCHITECT CMV IgG Controls:

  • Negative Control 1 Bottle (8.0 mL): contains recalcified human plasma with . protein (ovine) stabilizer. Preservative: sodium azide.
  • . Positive Control 1 - 1 Bottle (8.0 mL): contains recalcified human plasma and is reactive for IgG antibodies to cytomegalovirus (anti-CMV IgG). Preservative: sodium azide.

The controls are at the following proposed target anti-CMV IgG concentrations and ranges:

ControlTarget Anti-CMV IgGConcentration (AU/mL)Control Range (AU/mL)
Negative Control (Control -)N/A$\leq$ 3.1
Positive Control 1 (Control +1)30.015.0 to 45.0

The Positive Control 1 is referenced to an internal reference standard.

Principles of the Procedure

This assay is an automated, two-step immunoassay for the qualitative detection of IgG antibodies to CMV (anti-CMV IgG) in human serum and plasma using chemiluminescent microparticle immunoassay (CMIA) technology.

Pre-diluted sample, CMV virus lysate (strain AD169) coated paramagnetic microparticles, and assay diluent are combined and incubated. The anti-CMV IgG present in the sample binds to the CMV virus lysate (strain AD169) coated microparticles. The mixture is washed. Murine anti-human IgG acridinium-labeled conjugate is added to create a reaction mixture and incubated. Following a wash cycle, Pre-Trigger and Trigger Solutions are added.

The resulting chemiluminescent reaction is measured as a relative light unit (RLU). There is a direct relationship between the amount of anti-CMV IgG in the sample and the RLU detected by the system optics.

The presence or absence of anti-CMV IgG in the sample is determined by comparing the chemiluminescent RLU in the reaction to the cutoff RLU determined from an active calibration.

{7}------------------------------------------------

V. Intended Use of the Device

The ARCHITECT CMV IgG assay is a chemiluminescent microparticle immunoassay (CMIA) used for the qualitative detection of IgG antibodies to cytomegalovirus in human serum, serum separator, and plasma tubes (lithium heparin, lithium heparin separator, and tripotassium EDTA) on the ARCHITECT i System.

The ARCHITECT CMV IgG assay is to be used as an aid in the diagnosis of infection with cytomegalovirus and as an aid in the determination of serological status to cytomegalovirus in individuals including women of child-bearing age.

The ARCHITECT CMV IgG assay has not been cleared for use in screening blood, plasma, or tissue donors.

VI. Comparison of Technological Characteristics

The ARCHITECT CMV IgG assay (subject device) utilizes a CMIA methodology for the qualitative in vitro detection of IgG antibodies to cytomegalovirus and is intended for use on the ARCHITECT i System.

The similarities and differences between the subject device and the predicate assay are presented in the Table below.

{8}------------------------------------------------

Similarities and Differences Between
Device &PredicateDevice(s):Device: ARCHITECT CMV IgGPredicate Device: ADVIA Centaur CMV IgG Assay (K181213)
General DeviceCharacteristicSimilarities
Intended UseThe ARCHITECT CMV IgG assay is a chemiluminescentmicroparticle immunoassay (CMIA) used for the qualitativedetection of IgG antibodies to cytomegalovirus in humanserum, serum separator, and plasma tubes (lithium heparin,lithium heparin separator, and tripotassium EDTA) on theARCHITECT i System.The ARCHITECT CMV IgG assay is to be used as an aid inthe diagnosis of infection with cytomegalovirus and as an aidin the determination of serological status to cytomegalovirusin individuals including women of child-bearing age.The ARCHITECT CMV IgG assay has not been cleared foruse in screening blood, plasma, or tissue donors.The ADVIA Centaur CMV IgG (CMV IgG) assay is for in vitrodiagnostic use in the qualitative detection of IgG antibodies tocytomegalovirus (CMV) in human pediatric and adult serum and plasma(dipotassium EDTA, lithium heparin) using the ADVIA Centaur CPsystem. This assay is used to determine CMV IgG serological status andas an aid in the diagnosis of CMV infection in individuals for whom aCMV IgG test was ordered, including pregnant women.The ADVIA Centaur CMV IgG assay is not intended for blood andtissue donor screening.
Controls2 (Negative and Positive)2 (Negative and Positive)
MethodologyChemiluminescent microparticle immunoassayChemiluminometric Technology
Type of SpecimenSerum and PlasmaSerum and Plasma
General DeviceCharacteristicDifferences
Antigen UsedCMV Virus lysate (strain AD169)Heterogeneous mixture of CMV viral lysate antigens
Interpretation ofResultsNonreactive: < 6.0 AU/mLGrayzone/Equivocal: 6.0 to <15.0 AU/mLReactive: ≥ 15.0 AU/mLNegative: < 1.00 IndexReactive: ≥ 1.00 Index
ComponentsMicroparticles – CMV virus lysate (strain AD169) coatedmicroparticles in TRIS buffered saline with protein (bovine).Minimum concentration: 0.08% solids. Preservatives:ProClin 300 and antimicrobial agents.Conjugate - Murine anti-human IgG acridinium-labeledconjugate in MES buffer with protein (bovine). Minimumconcentration: 44 ng/mL. Preservatives: sodium azide andantimicrobial agents.Assay Diluent – Calf serum and MES buffer with protein(bovine). Preservatives: ProClin 300 and ProClin 950.Solid Phase Reagent – Streptavidin-coated paramagneticmicroparticles preformed with biotinylated CMV viral lysate antigens(~0.5 mg/mL) in buffer with surfactants, sodium caseinate, and sodiumazide (< 0.1%)Lite Reagent - Mouse monoclonal anti-human IgG antibody labeledwith acridinium ester (~0.06 µg/mL) in buffer with surfactant, bovineserum albumin (BSA), and sodium azide(< 0.1%)Diluent - Potassium thiocyanate (~0.55 M), surfactant, sodiumcaseinate, BSA, and preservatives
Calibrators6 Calibrators2 Calibrators
Calibration StorageMaximum of 30 days14 days

{9}------------------------------------------------

{10}------------------------------------------------

VII. Summary of Nonclinical Performance

A. Within- Laboratory Precision (20-Dav)

A study was performed based on guidance from CLSI EP05-A3. Testing was conducted using 3 lots of the ARCHITECT CMV IgG reagents, 3 lots of the ARCHITECT CMV IgG Calibrators, 3 lots of the ARCHITECT CMV IgG Controls, and 1 instrument. Two controls and 6 recalcified human plasma panels (representing serum matrix) were tested in a minimum of 2 replicates at 2 separate times per day on 20 days on 3 reagent lot/calibrator lot combinations, where a unique reagent lot and a unique calibrator lot are paired. The performance from a representative combination is shown in the following table.

SampleNMean(AU/mL)Within-Run(Repeatability)Within-Laboratorya
SD%CVSD(Rangeb)%CV(Rangeb)
Negative Control1180.10.04NAc0.04(0.00-0.04)NAc
Positive Control 111827.20.752.80.81(0.71-0.89)3.0(2.6-3.0)
Panel 11170.40.05NAc0.05(0.04-0.05)NAc
Panel 21183.30.13NAc0.14(0.13-0.15)NAc
Panel 31206.80.24NAc0.25(0.23-0.25)NAc
Panel 411917.60.472.70.52(0.50-0.54)2.9(2.8-2.9)
Panel 512037.30.792.10.96(0.96-1.06)2.6(2.6-2.8)
Panel 6118200.94.142.14.86(4.05-5.00)2.4(2.2-2.5)

a Includes within-run (repeatability), between-run, and between-day variability.

b Minimum and maximum SD or %CV across all reagent lot/calibrator lot combinations.

C Not applicable

* Clinical and Laboratory Standards Institute (CLSI). Evaluation of Quantitative Measurement Procedures: Approved Guideline-Third Edition. CLSI Document EP05-A3. Wayne, PA: CLSI; 2014.

{11}------------------------------------------------

B. Analytical Specificity

Potentially Interfering Endogenous Substances

A study was performed based on guidance from CLSI EP07, 3rd ed. * and CLSI EP37, 1st ed. Each substance was tested at 3 levels of the analyte (approximately 4.0 AU/mL, 12.0 AU/mL, and 20.0 AU/mL).

No significant interference (interference ≤ +0.6 AU/mL for samples < 6.0 AU/mL, ≤ +1.5 AU/mL for samples between 6.0 AU/mL and < 15.0 AU/mL, and ≥ -10% for samples ≥ 15.0 AU/mL) was observed at the following concentrations.

No Significant Interference
Potentially Interfering SubstanceInterferent Level
Unconjugated Bilirubin40 mg/dL
Conjugated Bilirubin40 mg/dL
Hemoglobin1000 mg/dL
Total Protein15 g/dL
Triglycerides1650 mg/dL

Interference greater than +0.6 AU/mL for samples < 6.0 AU/mL was observed at the concentration shown below for the following substance.

PotentiallyInterferingSubstanceInterferent LevelAnalyte LevelInterference(95% CI)
Triglycerides2475 mg/dL4.0 AU/mL0.7 AU/mL(0.6 AU/mL,0.8 AU/mL)

* Clinical and Laboratory Standards Institute (CLSI). Interference Testing in Clinical Chemistry. 3rd ed. CLSI Guideline EP07. Wayne, PA: CLSI; 2018.

* Clinical and Laboratory Standards Institute (CLSI). Supplemental Tables for Interference Testing in Clinical Chemistry. 1st ed. CLSI supplement EP37. Wayne, PA: CLSI; 2018.

{12}------------------------------------------------

Potentially Interfering Drugs and Other Substances

A study was performed based on guidance from CLSI EP07, 3rd ed. * and CLSI EP37, 1st ed. * Each substance was tested at 3 levels of the analyte (approximately 4.0 AU/mL, 12.0 AU/mL, and 20.0 AU/mL).

No significant interference (interference ≤ +0.6 AU/mL for samples < 6.0 AU/mL, ≤ +1.5 AU/mL for samples between 6.0 AU/mL and < 15.0 AU/mL, and ≥ -10% for samples ≥ 15.0 AU/mL) was observed at the following concentrations.

No Significant Interference
Potentially Interfering SubstanceInterferent Level
Acetaminophen250 mg/L
Ascorbic Acid300 mg/L
Beta Carotene6 mg/L
Biotin3510 ng/mL
Cidofovir240 mg/L
Diphenhydramine77.4 µg/dL
Folic Acid100 nmol/L
Foscarnet4320 mg/L
Gangciclovir800 mg/L
Ibuprofen500 mg/L
Valganciclovir900 mg/L

* Clinical and Laboratory Standards Institute (CLSI). Interference Testing in Clinical Chemistry. 3rd ed. CLSI Guideline EP07. Wayne, PA: CLSI; 2018.

* Clinical and Laboratory Standards Institute (CLSI). Supplemental Tables for Interference Testing in Clinical Chemistry. 1st ed. CLSI supplement EP37. Wayne, PA: CLSI; 2018.

{13}------------------------------------------------

Potentially Interfering Other Conditions

A total of 187 specimens from individuals with medical conditions unrelated to cytomegalovirus infection and specimens containing potentially interfering substances were evaluated. One specimen tested resulted in a reactive result.

CategoryNNumber of ARCHITECT CMVIgG Reactive Results
Anti-dsDNA Antibodies100
Anti-nuclear Antibody (ANA)80
Epstein-Barr Virus (EBV) IgG100
Hepatitis A90
Hepatitis B100
Hepatitis C100
Herpes Simplex Virus Types 1 (IgG)100
Herpes Simplex Virus Types 2 (IgG)60
High titer CMV IgM10
HAMA100
Human Herpesvirus 6 (HHV6)100
Human Immunodeficiency Virus (HIV)60
Hyper IgG70
Influenza vaccine recipient100
Measles (IgG)100
Parvovirus B19 (IgG)100ᵃ
Rheumatoid Factor100
Rubella (IgG)100
Syphilis100
Toxoplasmosis (IgG)101ᵇ
Varicella Zoster Virus100
Total1871

a One parvovirus B19 (IgG) specimen was equivocal with the ARCHITECT CMV IgG assay and positive with the comparator assay.

b One toxoplasmosis (IgG) specimen was reactive with the ARCHITECT CMV IgG assay and positive with the comparator assay.

{14}------------------------------------------------

C. CDC Panel Agreement

One CDC CMV IgG human serum panel, comprised of 80 samples that are either CMV IgG negative or CMV IgG positive, was obtained from the Centers for Disease Control and Prevention (CDC) and tested using the ARCHITECT CMV IgG assay. The results were submitted to the CDC. The CDC added their result interpretation for each sample.

The percent (%) agreement values of the ARCHITECT CMV IgG assay relative to the CDC results were calculated. The positive % agreement and corresponding two-sided 95% CI was 100% (91.59%, 100%). The negative % agreement and corresponding two-sided 95% CI was 92.11% (78.62%, 98.34%). The overall % agreement and corresponding two-sided 95% CI was 96.25% (89.43%, 99.22%).

The results are presented as a means to convey further information on the performance of this assay with a masked, characterized serum panel. This does not imply endorsement of the assay by the CDC.

VIII. Summary of Clinical Performance

A. Expected Values

Representative performance data are provided in this section. Results obtained in individual laboratories may vary.

It is recommended that each laboratory determine its own reference range based upon its particular locale and population characteristics.

Of the 989 specimens included in the ARCHITECT CMV IgG clinical study, 791 were from the intended use population in the US. Of the 791 specimens, 591 (74.7%) were routine order (325 female and 266 male, 0 to 91 years old) and 200 (25.3%) were pregnant females (19 to 45 years old). The mean age across the 791 subjects was 40 years.

The ARCHITECT CMV IgG assay was reactive in 508 (64.2%) of the collected specimens in the intended use population in the US (n = 791). Testing of the specimens was performed at 3 clinical testing sites located in Indianapolis Indiana, Lewisville Texas, and Palo Alto California.

The distribution of ARCHITECT CMV IgG reactive, grayzone/equivocal, and nonreactive results by age and sex is summarized in the following table.

{15}------------------------------------------------

AgeRange(Years)SexARCHITECT CMV IgG Result
Number ofReactive (%)Number ofGrayzone/Equivocal(%)Number ofNonreactive(%)Total
0 to 12Female4 (33.3%)1 (8.3%)7 (58.3%)12
Male7 (38.9%)0 (0.0%)11 (61.1%)18
13 to 21Female16 (47.1%)0 (0.0%)18 (52.9%)34
Male12 (60.0%)1 (5.0%)7 (35.0%)20
22 to 29Female90 (65.2%)0 (0.0%)48 (34.8%)138
Male17 (63.0%)0 (0.0%)10 (37.0%)27
30 to 39Female85 (50.0%)3 (1.8%)82 (48.2%)170
Male21 (56.8%)0 (0.0%)16 (43.2%)37
40 to 49Female38 (82.6%)0 (0.0%)8 (17.4%)46
Male25 (59.5%)2 (4.8%)15 (35.7%)42
50 to 59Female44 (81.5%)1 (1.9%)9 (16.7%)54
Male29 (72.5%)1 (2.5%)10 (25.0%)40
60 to 64Female16 (88.9%)0 (0.0%)2 (11.1%)18
Male24 (80.0%)0 (0.0%)6 (20.0%)30
65 to 100Female40 (75.5%)1 (1.9%)12 (22.6%)53
Male40 (76.9%)0 (0.0%)12 (23.1%)52
Total508 (64.2%)10 (1.3%)273 (34.5%)791

The ARCHITECT CMV IgG results for each category in the intended use population are summarized in the following table.

ARCHITECT CMV IgG Result
Specimen CategoryNumber of Reactive(%)Number ofGrayzone/Equivocal(%)Number ofNonreactive (%)Total
Routine Order412 (69.7%)8 (1.4%)171 (28.9%)591
Pregnant Females96 (48.0%)2 (1.0%)102 (51.0%)200
Total508 (64.2%)10 (1.3%)273 (34.5%)791

{16}------------------------------------------------

B. System Reproducibility

A study was performed based on guidance from CLSI EP05-A3. * Testing was conducted at each of 3 testing sites using 3 lots of the ARCHITECT CMV IgG reagents, 2 lots of the ARCHITECT CMV IgG Calibrators, 1 lot of the ARCHITECT CMV IgG Controls, and 1 instrument. Two controls and 6 recalcified human plasma panels (representing serum matrix) were tested in a minimum of 3 replicates at 2 separate times per day on 5 different days.

MeanRepeatabilityWithin-LaboratoryaReproducibilityb
SampleN(AU/mL)SD%CVSD%CVSD%CV
Negative Control3590.00.02NAc0.03NAc0.05NAc
Positive Control 136027.30.802.91.003.71.244.5
Panel 13600.30.04NAc0.05NAc0.12NAc
Panel 23603.10.13NAc0.16NAc0.30NAc
Panel 33606.50.23NAc0.27NAc0.49NAc
Panel 435917.20.583.40.694.00.844.8
Panel 536036.71.113.01.403.81.554.2
Panel 6360188.94.152.25.623.07.453.9

a Includes repeatability, between-run, and between-day variability.

b Includes repeatability, between-run, between-day, between-lot, and the site-lot interaction variability.

C Not applicable

* Clinical and Laboratory Standards Institute (CLSI). Evaluation of Quantitative Measurement Procedures: Approved Guideline-Third Edition. CLSI Document EP05-A3. Wayne, PA: CLSI; 2014.

{17}------------------------------------------------

C. Percent Agreement

A clinical study (method comparison) was performed in the US based on guidance from CLSI EP12-A2 * to evaluate the percent agreement between the ARCHITECT CMV IgG investigational assay and a current, FDA-cleared, commercially available anti-CMV IgG assay with routine order specimens collected in the US (n = 591) and outside of the US ( n= 198) and specimens collected from pregnant females in the US (n = 200).

Further evaluation of 4 specimens (3 from routine order and 1 from pregnant females) with an equivocal/grayzone result by comparator assay was performed with 2 additional current, FDA-cleared, commercially available anti-CMV IgG assays.

Comparator Result

The percent agreement between ARCHITECT CMV IgG and the comparator assay was evaluated.

ARCHITECTComparator ResultPositive %Negative %
SpecimenCategoryCMV IgGResultPositiveEquivocalNegativeAgreement(95% CI)aAgreement(95% CI)a
RoutineOrderReactive5140097.7(514/526)(96.1, 98.7)99.2(261/263)(97.3, 99.8)
Equivocal702
Nonreactive23c261
PregnantFemalesReactive98b0099.0(98/99)(94.5, 99.8)100.0(102/102)(96.4, 100.0)
Equivocal11d0
Nonreactive00102

a The 95% confidence interval (CI) for negative percent and positive percent agreement were estimated using the Wilson Score method.

b Two routine order specimens were from pregnant females and therefore, were also included in the pregnant female category.

C Of the 3 specimens that were nonreactive by ARCHITECT CMV IgG and equivocal/grayzone by comparator assay, 2 were negative and 1 was equivocal/grayzone by consensus testing. One specimen that was concordant equivocal/grayzone by ARCHITECT CMV IgG and comparator assay was negative by consensus testing.

d One specimen from pregnant females that was concordant grayzone/equivocal by ARCHITECT CMV IgG and comparator assay was negative based on the consensus result from the comparator assay and 2 additional current, FDA-cleared, commercially available anti-CMV IgG assays.

* Clinical and Laboratory Standards Institute (CLSI). User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline-Second Edition. CLSI Document EP12-A2. Wayne, PA: CLSI; 2008.

{18}------------------------------------------------

IX. Conclusion Drawn from Nonclinical and Clinical Laboratory Studies

The results presented in this 510(k) premarket notification demonstrate that the subject device (ARCHITECT CMV IgG) performance is substantially equivalent to the predicate assay (VIDAS CMV IgG assay, K920661).

§ 866.3175 Cytomegalovirus serological reagents.

(a)
Identification. Cytomegalovirus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to cytomegalovirus in serum. The identification aids in the diagnosis of diseases caused by cytomegaloviruses (principally cytomegalic inclusion disease) and provides epidemiological information on these diseases. Cytomegalic inclusion disease is a generalized infection of infants and is caused by intrauterine or early postnatal infection with the virus. The disease may cause severe congenital abnormalities, such as microcephaly (abnormal smallness of the head), motor disability, and mental retardation. Cytomegalovirus infection has also been associated with acquired hemolytic anemia, acute and chronic hepatitis, and an infectious mononucleosis-like syndrome.(b)
Classification. Class II (performance standards).