K Number
K181213

Validate with FDA (Live)

Date Cleared
2018-07-30

(84 days)

Product Code
Regulation Number
866.3175
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ADVIA Centaur® CMV IgG (CMV IgG) assay is for in vitro diagnostic use in the qualitative detection of IgG antibodies to cytomegalovirus (CMV) in human pediatric and adult serum and plasma (dipotassium EDTA, lithium heparin) using the ADVIA Centaur XP system. The assay is used to determine CMV IgG serological status and as an aid in the diagnosis of CMV infection in individuals for whom a CMV IgG test was ordered, including pregnant women. The ADVIA Centaur CMV IgG assay is not intended for blood and tissue donor screening.

The ADVIA Centaur® CMV IgG) Quality Control material is for in vitro diagnostic use for monitoring the performance of the ADVIA Centaur CMV IgG (CMV IgG) assay on ADVIA Centaur systems.

Device Description

CMV IgG Assay Kit (100-Tests) consists of 1 ReadyPack containing ADVIA Centaur CMV IgG Lite Reagent and Solid Phase reagent, and 1 ReadyPack ancillary reagent pack that contains ADVIA Centaur CMV IgG Diluent, and a set of Calibrators (1 vial each of Low and High, with fill volume of 2 mL each). The reagents and calibrators are packaged together in the ADVIA Centaur CMV IgG Assay kit, while the associated ADVIA Centaur CMV IgG Quality Control is packaged separately.

The ADVIA Centaur CMV IgG assay is a fully automated, 2-step sandwich immunoassay using indirect chemiluminometric technology.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the ADVIA Centaur CMV IgG assay, based on the provided FDA 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are generally inferred from the "Assay Cut-off" section, where the cutoff value was set to achieve a certain agreement with a comparator device. The performance is then demonstrated in the clinical studies.

Acceptance Criteria CategorySpecific Acceptance Criteria (Inferred)Reported Device Performance (Combined Population)
Agreement with Comparator≥ 95% Positive Agreemen99.6% (1037/1041)
≥ 95% Negative Agreemen96.8% (637/658)
Precision (Within-Lab)Not explicitly stated as acceptance criteria, but demonstrated by low %CV values.Ranges from NA (for very low mean values) up to 3.9% CV for positive control.
ReproducibilityNot explicitly stated as acceptance criteria, but demonstrated by low %CV values across sites.Ranges from NA (for very low mean values) up to 7.0% CV for a high positive serum pool.
No InterferenceNo change in clinical interpretation at specified interferent concentrations.No change in clinical interpretation throughout the assay range.
Cross-ReactivityAgreement with comparator assay (no significant false positives/negatives due to cross-reactants).Generally good agreement with comparator, minimal discrepancies identified for equivocal samples.
Matrix EquivalenceGood correlation (high 'r' value, slope near 1, intercept near 0) for different plasma types compared to serum.Correlation Coefficient (r) = 0.99 (Dipotassium EDTA plasma vs. Serum), 0.96 (Lithium heparin plasma vs. Serum). Slopes were 1.01.
CDC Panel AgreementNot explicitly stated as acceptance criteria, but 100% agreement indicates strong performance.100% agreement with CDC serological status (39 positive, 41 negative).

2. Sample Sizes Used for the Test Set and Data Provenance

  • Initial Cut-off Study: 389 remnant clinical samples (negative: 196, positive: 186, equivocal: remainder - exact number not specified but implied to be few).
  • Method Comparison (Primary Test Set): A total of 1842 samples
    • Prospective Study: 1699 specimens
      • 684 general population subjects
      • 348 pregnant subjects
      • 229 pediatric subjects (2-21 years old)
      • 44 HIV-positive subjects
      • 394 transplant-patient subjects
      • Provenance: 6 collection sites in the United States.
    • Retrospective Study: 143 HIV-positive specimens
      • Provenance: 1 site (implied to be in the US, given the overall context of the submission).
  • CDC Panel: 80 previously characterized serum samples.
    • Provenance: Centers for Disease Control (CDC).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not explicitly provided in the document. The ground truth for the comparison studies was established by a "comparator CMV IgG assay" (bioMerieux VIDAS CMV IgG (CMVG) Assay, K920661). For equivocal samples, additional testing was done on "2 other CMV IgG comparator assays" to achieve a "two out of three consensus result."

There is no mention of human experts (e.g., radiologists, pathologists) establishing ground truth for the primary clinical studies. The ground truth for the CDC panel samples was "previously characterized" by the CDC, implying expert determination at some point in the past, but details are absent.

4. Adjudication Method for the Test Set

  • For the primary method comparison, the comparator CMV IgG assay served as the reference.
  • For samples that tested equivocal on the comparator assay, an adjudication method of "two out of three consensus result" was used, comparing the initial comparator result with results from two other CMV IgG comparator assays.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

This is not applicable. The device is an in-vitro diagnostic (IVD) assay for detecting antibodies, not an AI-assisted diagnostic imaging or pathology device that involves human readers interpreting cases. Therefore, no MRMC study or effect size for AI assistance is provided or expected.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, this is a standalone device. The ADVIA Centaur CMV IgG assay is a fully automated immunoassay that generates a qualitative result (reactive or non-reactive) based on an Index value. Its performance is evaluated entirely without human interpretation of raw assay data. The results are reported directly by the system.

7. The Type of Ground Truth Used

The primary ground truth for the clinical studies was established by a comparator CMV IgG assay. This is a surrogate ground truth based on the performance of an already legally marketed and accepted diagnostic test. For equivocal samples, an expert consensus of other comparator assays (two out of three consensus) was used. For the CDC panel, the ground truth was previously characterized serological status provided by the CDC.

8. The Sample Size for the Training Set

The document does not explicitly describe a separate training set for the device in the context of machine learning or AI. The term "training set" is typically used in the development of AI algorithms. This device is an immunoassay, and its development would involve optimization and calibration rather than distinct "training" in the AI sense.

However, the "Assay Cut-off" section describes a study performed to determine the cutoff value:

  • A comparison study involving 389 remnant clinical samples was used to determine the placement of the cutoff value. While not a "training set" in the AI sense, these samples were used to establish a critical decision-making parameter for the assay.

9. How the Ground Truth for the Training Set Was Established

As discussed above, there isn't a "training set" in the AI context. For the 389 samples used to establish the cut-off value:

  • The ground truth for these samples was established by a comparator CMV IgG assay. The aim was to set the cutoff such that the new device's results agreed with the comparator.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

July 30, 2018

Siemens Healthcare Diagnostics, Inc. Kira Gordon Sr. Regulatory Affairs Specialist 511 Benedict Ave Tarrytown, New York 10591

Re: K181213

Trade/Device Name: ADVIA Centaur CMV IgG ADVIA Centaur CMV IgG Quality Control Regulation Number: 21 CFR 866.3175 Regulation Name: Cytomegalovirus serological reagents Regulatory Class: Class II Product Code: LFZ, OCH Dated: May 1, 2018 Received: May 7, 2018

Dear Kira Gordon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR

{1}------------------------------------------------

Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Steven R. Gitterman -S for

Uwe Scherf, Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K181213

Device Name ADVIA Centaur® CMV IgG ADVIA Centaur® CMV IgG Quality Control

Indications for Use (Describe)

ADVIA Centaur CMV lgG:

The ADVIA Centaur® CMV IgG (CMV IgG) assay is for in vitro diagnostic use in the qualitative detection of IgG antibodies to cytomegalovirus (CMV) in human pediatric and adult serum and plasma (dipotassium EDTA, lithium heparin) using the ADVIA Centaur XP system. The assay is used to determine CMV IgG serological status and as an aid in the diagnosis of CMV infection in individuals for whom a CMV IgG test was ordered, including pregnant women. The ADVIA Centaur CMV IgG assay is not intended for blood and tissue donor screening.

ADVIA Centaur® CMV IgG Quality Control:

The ADVIA Centaur® CMV IgG) Quality Control material is for in vitro diagnostic use for monitoring the performance of the ADVIA Centaur CMV IgG (CMV IgG) assay on ADVIA Centaur systems.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

1. Applicant:

Contact:Kira Gordon, PhDSr. Regulatory Affairs Specialist
Address:Siemens Healthcare Diagnostics Inc.511 Benedict AveTarrytown, NY 10591
Phone:(914) 524-2996
FAX:(914) 524-3579
Email:kira.gordon@siemens-healthineers.com
    1. Date: May 01, 2018

3. Proprietary and Established Names:

ADVIA® Centaur CMV IgG ADVIA® Centaur CMV IgG Quality Control

4. Regulatory Information:

Classification Names:Cytomegalovirus serological reagents
Regulation Number:21 CFR § 866.3175
Classification:Class II
Product Code:LFZ
Panel:Microbiology (83)

4. Predicate Device:

Device Name: bioMerieux VIDAS CMV IgG (CMVG) Assay 510(k) Number: K920661 Manufacturer: bioMerieux

5. Indications for Use:

ADVIA Centaur CMV lgG:

The ADVIA Centaur® CMV IgG Assay is for in vitro diagnostic use in the qualitative detection of IgG antibodies to cytomegalovirus (CMV) in human pediatric and adult serum and plasma (dipotassium EDTA, lithium heparin) using the ADVIA Centaur XP system. The assay is used to determine CMV IgG serological status and as an aid in the diagnosis of CMV infection in individuals for whom a CMV IgG test was ordered, including pregnant women.

The ADVIA Centaur CMV IgG assay is not intended for blood and tissue donor screening.

ADVIA Centaur® CMV IgG Quality Control: The ADVIA Centaur® CMV IgG (CMV IgG) Quality Control material is for in vitro

{4}------------------------------------------------

diagnostic use for monitoring the performance of the ADVIA Centaur CMV IgG (CMV IgG) assay on ADVIA Centaur systems.

9. Device Description:

CMV IgG Assay Kit (100-Tests) consists of 1 ReadyPack containing ADVIA Centaur CMV IgG Lite Reagent and Solid Phase reagent, and 1 ReadyPack ancillary reagent pack that contains ADVIA Centaur CMV IgG Diluent, and a set of Calibrators (1 vial each of Low and High, with fill volume of 2 mL each). The reagents and calibrators are packaged together in the ADVIA Centaur CMV IgG Assay kit, while the associated ADVIA Centaur CMV IgG Quality Control is packaged separately.

10. Test Principle

The ADVIA Centaur CMV IgG assay is a fully automated, 2-step sandwich immunoassay using indirect chemiluminometric technology. The patient specimen is diluted with ADVIA Centaur CMV IgG Diluent and incubated with the Solid Phase reagent. The Solid Phase reagent contains a heterogeneous mixture of biotinylated CMV viral lysate antigens, preformed to streptavidin-coated magnetic particles. The antigen-coated particles subsequently capture CMV-specific antibodies in the specimen. The antibody-antigen complex is washed and Lite reagent is added. The Lite reagent consists of an acridinium-ester (AE)-labeled anti-human IgG mouse monoclonal antibody. The entire complex is washed and the signal is generated in the presence of Lite Reagent bound to the Solid Phase via the CMV IgG-CMV antigen complex.

The system automatically performs the following steps:

  • Dispenses 20 uL of sample into a cuvette. -
  • Dispenses 195 µL of ADVIA Centaur CMV IgG Diluent into the cuvette with the sample. -
  • Removes 100 uL of the diluted sample from the cuvette and dispenses it into a second cuvette.
  • -Dispenses 200 uL of Solid Phase and incubates the mixture for 18.0 minutes at 37°C.
  • -Separates the Solid Phase from the mixture and aspirates the unbound reagent.
  • -Washes the cuvette with ADVIA Centaur Wash 1.
  • -Re-suspends the washed particles in 250 uL of ADVIA Centaur Wash 1.
  • Dispenses 100 uL Lite Reagent and incubates the mixture for 18.0 minutes at 37°C. -
  • -Separates the Solid Phase from the mixture and aspirates the unbound reagent.
  • Washes the cuvette with ADVIA Centaur Wash 1. -
  • Dispenses 300 uL of ADVIA Centaur Acid Reagent and 300 uL of ADVIA Centaur Base -Reagent to initiate the chemiluminescent reaction.
  • Reports results according to the selected option, as described in the system operating -Instructions.

A direct relationship exists between the amount of bound anti-CMV IgG present in the patient specimen and the amount of relative light units (RLUs) detected by the system. A result of reactive or nonreactive is determined according to the Index established with the calibrators:

  • . Samples with an Index Value of < 1.00 are considered nonreactive for CMV

{5}------------------------------------------------

IgG antibodies.

  • . Samples with an Index Value of ≥ 1.00 are considered reactive for CMV IgG antibodies.

11. Substantial Equivalence Information:

Device Name: bioMerieux VIDAS CMV IgG (CMVG) Assay 510(k) Number: K920661

Comparison with Predicate:

ItemNew Device:Predicate Device:
Intended UseThe ADVIA Centaur® CMV IgG Assay isfor in vitro diagnostic use in the qualitativedetection of IgG antibodies tocytomegalovirus (CMV) in human pediatricand adult serum and plasma (dipotassiumEDTA, lithium heparin) using the ADVIACentaur XP system. The assay is used todetermine CMV IgG serological status andas an aid in the diagnosis of CMV infectionin individuals for whom a CMV IgG testwas ordered, including pregnant women.The ADVIA Centaur CMV IgG assay is notintended for blood and tissue donorscreening.The VIDAS CMV IgG (CMVG)Assay is intended for use on theinstruments of the VIDAS family(Vitek ImmunoDiagnostic AssaySystem) as a semi-quantitativeautomated enzyme-linkedfluorescent immunoassay (ELFA).It is intended for use indetermination of CMVimmunological experience from asingle serum sample, or as an aid inthe diagnosis of current CMVinfection through evaluation ofpaired sera for a significant increasein CMV-specific IgG. It is notintended for use in testing(screening) blood or plasma donors.
InstrumentADVIA CentaurVIDAS/mini-VIDAS
MeasurementQualitativeSemi-Quantitative
TechnologyChemiluminescenseenzyme-linked fluorescentimmunoassay
Assay Protocolsandwich immunoassaysandwich immunoassay
Sample typeSerum, PlasmaSerum
Cut-Offs< 1.0 (Index) - nonreactive≥ 1.0 (Index) - reactive< 4 AU/mL - negative> 4 to < 6 AU/mL - equivocal≥ 6 AU/mL - positive
Sample Volume20 µL100 µL
Calibrators matrixHuman plasmaHuman serum
Calibrator fill volumeLiquid, 2 mLLiquid, 2 mL
ControlsNegative and PositiveNegative and Positive

Non-Clinical Studies

a. Assay Cut-off

A comparison study between ADVIA Centaur CMV IgG assay and a comparator CMV IgG assay was performed to determine the placement of the cutoff value at a 1.00 Index by testing 389 remnant clinical samples. Of these, 196 samples were negative, 186 samples were positive, and the rest were equivocal by the comparator device. The cutoff value was set at ≥ 95% positive agreement and ≥ 95% negative agreement to the comparator device. Assigned values for

{6}------------------------------------------------

calibrators and controls are based on the cutoff Index determination.

b. Precision

Precision was evaluated according to CLSI document EP05-A3. Repeatability and Within-Lab imprecision were evaluated by testing 5 serum-based samples (serum sample pools), and 2 plasma-based samples (controls. negative and positive). The samples were assayed in duplicate over the course of 20 days, 2 runs per day, for a total of 40 runs and 80 replicates. The results

{7}------------------------------------------------

are shown below
SampleTypeNMeanRepeatabilityBetween-RunBetween-DayWithin-Lab
SD% CVSD% CVSD% CVSD% CV
Control 1 (negative)800.080.00NAa0.000.00.000.00.01NA
Control 2 (positive)804.540.092.00.122.60.092.00.183.9
Serum Pool 1800.070.00NA0.000.60.00NA0.01NA
Serum Pool 2800.900.022.10.011.30.022.30.033.4
Serum Pool 3801.360.032.50.021.10.022.10.053.5
Serum Pool 4802.700.062.20.072.50.071.70.103.7
Serum Pool 5809.680.202.10.121.30.121.80.293.0

a NA = not applicable

c. Reproducibility

The reproducibility study was conducted at three external sites using 2 reagent lots. The protocol was run over 5 days, 2 runs per day, and 3 replicates per run for the sample pools, and 6 replicates per run for the negative and positive control materials. Reproducibility data was pooled for each reagent lot across three sites. Data is presented for 1 representative reagent lot.

SampleTypeNMeanRepeatabilityBetween-RunBetween-DayBetween-SiteReproducibility
SD% CVSD% CVSD% CVSD% CVSD% CV
Control 1(negative)1800.100.01NAa0.00NA0.00NA0.02NA0.02NA
Control 2(positive)1804.550.122.70.071.50.020.30.153.40.214.6
SerumPool 1900.100.01NA0.00NA0.00NA0.01NA0.01NA
SerumPool 289b0.800.011.80.021.90.011.10.056.10.056.7
SerumPool 3901.230.021.90.022.00.000.00.075.40.076.0
SerumPool 4902.980.061.90.061.90.051.70.134.20.165.3
SerumPool 59025.250.451.80.401.60.261.01.646.51.777.0

a NA = not applicable.

b One run had 2 replicates instead of 3.

{8}------------------------------------------------

d. Interference

Interference by endogenous substances in the ADVIA Centaur CMV IgG assay was evaluated at four CMV IgG levels (low negative, high negative, low positive, and high positive). Interfering substances at the levels indicated were tested as described in CLSI Document EP07-A2. There was no change in clinical interpretation throughout the assay range at the levels indicated.

InterferentConcentration
Hemolyzed500 mg/dL of hemoglobin
Icteric20 mg/dL of conjugated bilirubin
Icteric20 mg/dL of unconjugated bilirubin
Lipemic3000 mg/dL of intralipids (Triglycerides)
Total Protein*9g/dL of protein
Immunoglobulins*3g/dL of immunoglobulin
Biotin4500 ng/mL of biotin
Cholesterol400 mg/dL of cholesterol
  • Total protein and immunoglobulins were tested using clinical samples with high serum protein and samples from multiple myeloma patients respectively.

e. Cross-reactivity

The ADVIA Centaur CMV IgG assay was evaluated for potential cross-reactivity in specimens with other viral and microbial antibodies and other disease states. The CMV IgG status of each

{9}------------------------------------------------

sample was compared using the comparator assay.

Clinical CategoryNumber TestedADVIA Centaur CMV IgG ReactiveComparator CMV IgG Reactive
Chlamydia IgG1144
CMV IgM1066
Epstein-Barr virus (EBV) IgG1300
Epstein-Barr virus (EBV) IgM1044
Graves' disease3*00
Hepatitis A infection (HAV) IgG1033
Hepatitis B Core (HBc) IgG1034
Hepatitis C infection (HCV) IgG1011
Herpes simplex virus 1 (HSV1) IgG1333
Herpes simplex virus 2 (HSV2) IgG1200
Human anti-mouse antibody (HAMA)1476
Human chorionic gonadotropin (hCG)1100
Human herpes virus (HHV6) IgG1100
Human immunodeficiency virus (HIV) Antibodies1699
Influenza Antibodies1177
Measles IgG1100
Multiparity201818
Multiple myeloma231111
Parvovirus B19 IgG1133
Rheumatoid factor (RF)1011
Rubella IgG1300
Sjogren's Syndrome4*00
Systemic lupus erythematosus (SLE)*3*00
Syphilis IgG1144
Toxoplasma IgG1044
Varicella zoster virus (VZV) IgG1600
Total2978888

*Results may not be conclusive due to low number of samples tested.

f. Matrix Comparison

The ADVIA Centaur CMV IgG assay was evaluated using different specimen matrices.

{10}------------------------------------------------

ADVIA Centaur CMV IgG results that ranged from 0.74-28.55 Index were analyzed using orthogonal regression. The following results were obtained:

Serum (x) vs.NMean(Index)SlopeIntercept(Index)Correlation Coefficient (r)
Dipotassium EDTA plasma3810.761.01-0.080.99
Lithium heparin plasma (y)3810.761.01-0.070.96

12. Clinical Studies

  • a. Method Comparison with predicate device
    Percent agreement was determined by comparing the performance of the ADVIA Centaur CMV IgG assay to a comparator CMV IgG assay. A total of 1842 samples that were sent for CMV IgG testing were analyzed, including:

  • · 1699 prospectively collected specimens

  • 684 general population subjects sent for CMV IgG testing o

  • 348 pregnant subjects o

  • o 229 pediatric subjects (2-21 years old)

  • 44 HIV-positive subjects o

  • o 394 transplant-patient subjects

  • · 143 retrospective HIV-positive specimens

Prospective Study

A total of 1699 clinical routine specimens (from people aged 6 months-91 years, both male and female) were obtained from 6 collection sites in the United States. The specimens were from subjects sent for CMV IgG testing, pediatric subjects (aged 2-21 years), pregnant women. HIVpositive subjects, and transplant patients. The following results were obtained:

Prospective Study - Combined Population

Comparator-CMV IgG Assay
ADVIA Centaur CMV IgG AssayPositiveEquivocalNegativeTotal
Reactive103711101058
Nonreactive13637641
Total1038146471699
ADVIA Centaur CMV IgG AssayAgreement (%)95% Confidence Interval
Positive agreement99.6% (1037/1041)99.0%-99.9%
Negative agreement96.8% (637/658)95.2%-98.0%

Note: Fourteen samples that tested equivocal on the comparative assay were further tested on 2 other CMV IgG comparator assays. Of these 14 samples, 10 remained equivocal, 3 agreed and 1 did not agree with the ADVIA Centaur CMV IgG assay when compared to the two out of three consensus result.

The results obtained with the different subgroup prospective populations are presented in the

{11}------------------------------------------------

sections that follow.

Subjects Sent for CMV IgG Testing

A total of 684 general population prospective serum samples sent for CMV IgG testing were analyzed. The following results were obtained:

ADVIA Centaur CMV IgG AssayComparator CMV IgG Assay
PositiveEquivocalNegativeTotal
Reactive37567388
Nonreactive11294296
Total3767301684

Subjects Sent for CMV IgG Testing

ADVIA Centaur CMV IgG AssayAgreement (%)95% Confidence Interval
Positive agreement99.5% (375/377)98.1%-99.9%
Negative agreement95.8% (294/307)92.9%-97.7%

Note: Seven samples that tested equivocal on the comparator assay were further tested on 2 other CMV IgG assays. Of these 7 samples, 5 remained equivocal and 2 agreed with the ADVIA Centaur CMV IgG assay when compared to the two out of three consensus result.

Pregnant Women Subgroup

Serum samples sent for CMV IgG testing from 348 pregnant women were prospectively collected from 4 sites and tested. The following results were obtained:

Comparator CMV IgG Assay
ADVIA Centaur CMV IgG AssayPositiveEquivocalNegativeTotal
Reactive28701288
Nonreactive006060
Total287061348

Pregnant Women Subgroup

ADVIA Centaur CMV IgG AssayAgreement (%)95% Confidence Interval
Positive agreement100.0% (287/287)98.7%-100.0%
Negative agreement98.4% (60/61)91.2%-99.9%

Pediatric Subgroup

A total of 229 pediatric prospective serum samples were tested at 3 sites. Specimens were obtained from 4 sites, from males who were not pregnant. The subjects' ages

{12}------------------------------------------------

ranged from 2-21 years. The following results were obtained:

Pediatric Subgroup
Comparator CMV IgG Assay
ADVIA Centaur CMV IgG AssayPositiveEquivocalNegativeTotal
Reactive801182
Nonreactive01146147
Total802147229
ADVIA Centaur CMV IgG AssayAgreement (%)95% Confidence Interval
Positive agreement98.8% (80/81)93.3%-99.9%
Negative agreement98.6% (146/148)95.2%-99.8%

Note: Two samples that tested equivocal on the comparator assay were further tested on 2 other CMV IgG assays and the two out of three consensus result remained equivocal.

Human Immunodeficiency Virus (HIV) Patient Subgroup

Serum specimens sent for CMV IgG testing were prospectively collected at 5 sites, from 44 HIV-positive patients and tested. The following results were obtained:

HIV-Positive Patient Subgroup

Comparator CMV IgG Assay
ADVIA Centaur CMV IgG AssayPositiveEquivocalNegativeTotal
Reactive430043
Nonreactive0011
Total430144
ADVIA Centaur CMV IgG AssayAgreement (%)95% Confidence Interval
Positive agreement100.0% (43/43)91.8%–100.0%
Negative agreement100.0% (1/1)2.5%–100.0%

Transplant Patient Subgroup

Serum samples from 238 preoperative transplant patients from 3 sites were tested. The

{13}------------------------------------------------

following results were obtained:

Transplant Patient Subgroup - Preoperative
------------------------------------------------
ADVIA Centaur CMV IgG AssayComparator CMV IgG Assay
PositiveEquivocalNegativeTotal
Reactive15100151
Nonreactive008787
Total151087238
ADVIA Centaur CMV IgG AssayAgreement (%)95% Confidence Interval
Positive agreement100.0% (151/151)97.6%–100.0%
Negative agreement100.0% (87/87)95.8%–100.0%

Serum samples from 156 transplant patients were collected from 3 sites that perform kidney, pancreas, bone and marrow stem cell, liver, face and limb, heart, and lung transplants. After testing the following results were obtained

Transplant Patient Subgroup - Postoperative

Comparator CMV IgG Assay
ADVIA Centaur CMV IgG AssayPositiveEquivocalNegativeTotal
Reactive10141106
Nonreactive014950
Total101550156
ADVIA Centaur CMV IgG AssayAgreement (%)95% Confidence Interval
Positive agreement99.0% (101/102)94.7%-99.9%
Negative agreement90.7% (49/54)79.7%-96.9%

Note: Five samples that tested equivocal by the comparator assay were further tested on 2 other CMV IgG assays. Of these 5 samples, 3 remained equivocal, 1 agreed and 1 did not agree with the ADVIA Centaur CMV IgG assay when compared to the two of three consensus result.

Retrospective Study - HIV-Positive Patient Subgroup

A total of 143 retrospective remnant samples from the HIV-positive patient subgroup collected at 1

{14}------------------------------------------------

site were tested. The following results were obtained:

Comparator CMV IgG Assay
ADVIA Centaur CMV IgG AssayPositiveEquivocalNegativeTotal
Reactive13500135
Nonreactive0088
Total13508143

Retrospective HIV-Positive Patient Subgroup

ADVIA Centaur CMV IgG AssayAgreement (%)95% Confidence Interval
Positive agreement100.0% (135/135)97.3%-100.0%
Negative agreement100.0% (8/8)63.1%-100.0%

g. Centers for Disease Control (CDC) Panel:

A panel of 80 previously characterized serum samples was obtained from the CDC and evaluated with the ADVIA Centaur CMV IgG assay to determine the performance of the assay. There was 100% agreement with the serological status provided by the CDC.

Expected CDC Panel Results
ADVIA Centaur CMV IgG AssayPositiveNegativeTotal
Reactive39039
Nonreactive04141
Total394180

Note: The results are presented as a means to convey further information on the performance of this assay with a characterized serum panel from the CDC. This does not imply an endorsement of the assay by the CDC.

13. Conclusions

From the above comparative testing results of the ADVIA Centaur CMV IgG assay we conclude that it is substantially equivalent to the bioMerieux VIDAS CMV IgG Assay.

§ 866.3175 Cytomegalovirus serological reagents.

(a)
Identification. Cytomegalovirus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to cytomegalovirus in serum. The identification aids in the diagnosis of diseases caused by cytomegaloviruses (principally cytomegalic inclusion disease) and provides epidemiological information on these diseases. Cytomegalic inclusion disease is a generalized infection of infants and is caused by intrauterine or early postnatal infection with the virus. The disease may cause severe congenital abnormalities, such as microcephaly (abnormal smallness of the head), motor disability, and mental retardation. Cytomegalovirus infection has also been associated with acquired hemolytic anemia, acute and chronic hepatitis, and an infectious mononucleosis-like syndrome.(b)
Classification. Class II (performance standards).