K Number
K220772

Validate with FDA (Live)

Date Cleared
2022-12-07

(266 days)

Product Code
Regulation Number
880.5200
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The HydroMID is indicated for short term access (< 30 days) to the peripheral venous access system for intravenous therapy, including but not limited to, the administrations, and the sampling of blood and blood products.

Maximum Power Injection Flow Rate: -4Fr Single Lumen, 20cm: 6 mL/sec

Device Description

The HydroMID catheter is a 4 French, single lumegemidfifie catheter comprised of a radiopaque, hydrophilic catheter with a suture wing, Luer lock hub, and extension tube made from materials commonly used in the manufacture of catheters. Catheters are provided packaged in kit configurations with the appropriate accessories for placement in the appropriate clinical environments. The maximum power injection flow rate for the lumen is indicated on the extension tube clamp.

HydroMID has been shown to be effective in reducing thrombus accumulation and thrombotic occlusions. Reduction of thrombus accumulation and thrombotic occlusions were evaluated using in vitro and in vivo models. Pre-clinical in vivo evaluations do not necessarily predict clinical performance with respect to thrombus formation.

AI/ML Overview

The provided text describes the 510(k) summary for the HydroMID 4F Single Lumen Midline Catheter. It focuses on demonstrating substantial equivalence to a predicate device (HydroMID 4F Single Lumen Catheter, K203069) rather than presenting a performance study with specific acceptance criteria and reported device performance in the format requested.

Therefore, for aspects like "Table of acceptance criteria and the reported device performance," "Sample size used for the test set and data provenance," "Number of experts used to establish ground truth," "Adjudication method," "MRMC comparative effectiveness study," "Standalone performance," "Type of ground truth," "Sample size for the training set," and "How the ground truth for the training set was established," the document does not provide the requested quantitative information typically found in clinical or AI algorithm performance studies.

Instead, the document primarily addresses the following:

1. Acceptance Criteria and Device Performance (Qualitative Description):

The device is deemed to meet its intended purpose and be substantially equivalent to the predicate device based on various bench tests and biocompatibility testing. The "acceptance criteria" are implied to be adherence to various consensus standards and FDA guidance.

TestConsensus Standard/FDA Guidance/DescriptionReported Device Performance
Bench testing (Catheter length, marking spacing, surface continuity, in-life durability)"Confirm that the device meets intended product specifications"The device passed these tests, confirming it meets intended product specifications.
Biocompatibility testingFDA Final Guidance Document, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" (September 2020)The device passed biocompatibility testing according to the specified FDA guidance and ISO 10993 standards (1, 4, 5, 10, 12).

The document states, "All necessary performance testing has been conducted on the HydroMID to assure substantial equivalence to the predicate devices and to demonstrate the device performs as intended. All testing was performed on test units representative of finished devices."

Regarding the other requested information:

  • Sample size used for the test set and the data provenance: Not specified. The document mentions "test units representative of finished devices" for bench testing.
  • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable or not specified. This type of information is typically for AI/diagnostic device studies.
  • Adjudication method: Not applicable or not specified.
  • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI diagnostic device.
  • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a medical catheter, not an algorithm.
  • The type of ground truth used (expert consensus, pathology, outcomes data, etc): For the physical characteristics, the ground truth would be engineering specifications and measurements. For biocompatibility, it's the results of standardized biological assays.
  • The sample size for the training set: Not applicable. This is a physical medical device, not an AI model requiring a training set.
  • How the ground truth for the training set was established: Not applicable.

Summary of what the document focuses on instead:

The document focuses on demonstrating substantial equivalence to a predicate device (HydroMID 4F Single Lumen Catheter, K203069) by highlighting:

  • Identical intended use, principle of operation, and similar technological characteristics.
  • Minor differences (e.g., removal of certain warnings for midline therapies, addition of Copper Phthalocyanine Blue 15:1 material for catheter markings, addition of catheter markings) are addressed and supported by bench testing and biocompatibility data.
  • Adherence to various international and FDA consensus standards for medical devices (ISO 10555, ISO 11135, AAMI TIR28, ISO 15223, ISO 10993).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 7, 2022

Access Vascular, Inc. Brian Hanley VP. R&D and Operations 749 Middlesex Turnpike Billerica, Massachusetts 01821

Re: K220772

Trade/Device Name: HydroMID 4F Single Lumen Midline Catheter Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular catheter Regulatory Class: Class II Product Code: FOZ Dated: November 4, 2022 Received: November 7, 2022

Dear Brian Hanley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

David Walloschek

David Wolloscheck, Ph.D. For Joyce M. Whang, Ph.D. Acting Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220772

Device Name HydroMID 4F Single Lumen Midline Catheter

Indications for Use (Describe)

The HydroMID is indicated for short term access (< 30 days) to the peripheral venous access system for intravenous therapy, including but not limited to, the administrations, and the sampling of blood and blood products.

Maximum Power Injection Flow Rate: -4Fr Single Lumen, 20cm: 6 mL/sec

Type of Use (Select one or both, as applicable)
( Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Access Vascular. The word "ACCESS" is in a dark blue font, and the word "VASCULAR" is in a lighter blue font. There is a line that connects the two words, starting from the bottom left of the "A" in "ACCESS" and ending at the top left of the "V" in "VASCULAR".

K220772 510(k) Summary

Preparation Date: December 7, 2022

Submitter:

Access Vascular Inc. 749 Middlesex Turnpike Billerica, MA 01821

Contact:

Brian M. Hanley VP, R&D and Operations Phone: 781.538.6594 x101 bhanley@accessvascularinc.com

Subject Device

Trade Name:HydroMID 4F Single Lumen Midline Catheter
Regulation Name:Intravascular Catheter
Regulation Number:21 CFR 880.5200
Common Name:Catheter, Intravascular, therapeutic, Short-Term less than 30 days
Regulatory Class:Class II
Product Code:FOZ
Manufacturer:Access Vascular Inc
Manufacturer ERN:3015060232

Predicate Device

Trade Name:HydroMID 4F Single Lumen Catheter
Manufacturer:Access Vascular
510(k) Number:K203069
Regulation Name:Intravascular Catheter
Regulation Number:21 CFR 880.5200
Common Name:Catheter, intravascular, therapeutic, Short-Term less than 30 days
Regulatory Class:Class II
Product Code:FOZ

Reference Device

Trade Name:BioFlo Midline Catheter
510(k) Number:K161866

The reference devices are provided because they are a comparable midline catheter, with catheter markings.

Description of Use

HydroMID

The HydroMID catheter is a 4 French, single lumegemidfifie catheter comprised of a radiopaque, hydrophilic catheter with a suture wing, Luer lock hub, and extension tube made from materials

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Access Vascular. The logo is in blue and features the word "Access" in a bold, sans-serif font. Below "Access" is the word "Vascular" in a lighter shade of blue. The "A" in "Access" and the "V" in "Vascular" are connected to form a stylized "AV" symbol.

commonly used in the manufacture of catheters. Catheters are provided packaged in kit configurations with the appropriate accessories for placement in the appropriate clinical environments. The maximum power injection flow rate for the lumen is indicated on the extension tube clamp.

HydroMID has been shown to be effective in reducing thrombus accumulation and thrombotic occlusions. Reduction of thrombus accumulation and thrombotic occlusions were evaluated using in vitro and in vivo models. Pre-clinical in vivo evaluations do not necessarily predict clinical performance with respect to thrombus formation.

Indication for Use

The HydroMID is indicated for short term access (< 30 days) to the peripheral venous access system for intravenous therapy, including but not limited to, the administration of fluids, medications, and the sampling of blood and blood products.

Maximum Power Injection Flow Rate:

-4Fr Single Lumen, 20cm: 6 mL/sec

Comparison of Technological Characteristics with the Predicate Device

The Proposed Devices and Predicate Devices, respectively, are similar in indications for use, intended use, technological characteristics, and principles of operation.

The differences between the Proposed Device and the Predicate Device are minor and raise no different questions of safety and effectiveness, thus it was concluded that the Proposed Device is substantially equivalent to the Predicate Device. In accordance with 21CFR807.92(a)(6) a summary of how the technological characteristics of the Proposed Device compares to the Predicate Device is provided below.

SpecificationHydroMID 4F SingleLumenAccess Vascular Inc.Subject DeviceHydroMID-142 4FAccess Vascular Inc.Predicate DeviceK203069Comparison
Intended UseIntended for short-term access to theperipheral venoussystemIntended for short-termaccess to the peripheralvenous systemSame
SpecificationHydroMID 4F SingleHydroMID-142 4FComparison
LumenAccess Vascular Inc.
Access Vascular Inc.Predicate Device
Subject DeviceK203069
Indications for UseThe HydroMID isindicated for shortterm access (< 30 days)to the peripheralvenous access systemfor intravenoustherapy, including butnot limited to, theadministration of fluids,medications, and thesampling of blood andblood products.Maximum PowerInjection Flow Rate:-4Fr Single Lumen,20cm: 6 mL/secThe HydroMID isindicated for short termaccess (< 30 days) to theperipheral venousaccess system forintravenous therapy,including but not limitedto, the administration offluids, medications, andthe sampling of bloodand blood products.Therapies notappropriate for midlinecatheters includecontinuous vesicanttherapy, parentalnutrition, infusates withpH less than 5 or greaterthan 9, and infusateswith an osmolaritygreater than600mOsm/L.HydroMID is rated for aMaximum PowerInjection Flow Rate of 6mL/sec.Difference - seecomment #1: TheInfusion Nurses Society(INS) standards languageslightly changed, butintended use is still thesame.
WarningsDo Not use midlinecatheter for continuousvesicant therapy, PN, orinfusates withextremes of pH orosmolarityTherapies notappropriate for midlinecatheters includecontinuous vesicanttherapy, parentalnutrition, infusates withpH less than 5 or greaterthan 9, and infusateswith an osmolaritygreater than600mOsm/LDifference - seecomment #1: The INSstandards languageslightly changed, butintended use is still thesame.
Device ClassificationClass IIClass IISame
Product CodeFOZFOZSame
Regulation21 CFR 880.520021 CFR 880.5200Same
Prescription DeviceYesYesSame
Intended populationAdult and Pediatric useAdult and Pediatric useSame
SpecificationHydroMID 4F SingleLumenAccess Vascular Inc.Subject DeviceHydroMID-142 4FAccess Vascular Inc.Predicate DeviceK203069Comparison
Catheter OuterDiameter French Size4Fr (1.40mm)(Post Hydrated)4Fr (1.40mm)(Post Hydrated)Same
Catheter OuterDiameter as supplied1.30mm(Supplied Dehydrated)1.30mm(Supplied Dehydrated)Same
Catheter InnerDiameter1mm1mmSame
Useable Length20 cm20 cmSame and 20 cm
Priming Volume< 1.0mL< 1.0mLSame
GuidewireCompatibilityØ.018"Ø.018"Same
Catheter ShaftDesignTaperTaperSame
Number of LumensSingle Lumen (SL)Single Lumen (SL)Same
Key DeviceComponentsCatheter Shaft, SutureWing, Extension Tube,Luer Hub, ClampCatheter Shaft, SutureWing, Extension Tube,Luer Hub, ClampSame
Short or Long TermAccessShortShortSame
Use with PowerInjection andSpecified Flow RateYes6mL/secYes6mL/secSame
Catheter MaterialsRadiopaque hydrophilicpolyol catheter withCopper PhthalocyanineBlue 15:1 cathetermarkings, Luer lockhub, polyurethaneextension tubing, andsuture wingRadiopaque hydrophilicpolyol catheter withLuer lock hub,polyurethane extensiontubing, and suture wingDifference - seecomment #2:Copper PhthalocyanineBlue 15:1 material isadded to the outersurface of the cathetershaft to incorporatecatheter markings assupported by benchtesting andbiocompatibility testing.
X-Ray ConfirmationRequiredYesYesSame
Sterilization MethodEthylene OxideEthylene OxideSame
Single UseYesYesSame
MRI SafetyMRI ConditionalMRI ConditionalSame
How SuppliedConvenience Kit:Basic IR KitFull Nursing KitMaximal Barrier KitConvenience Kit:Basic IR KitFull Nursing KitMaximal Barrier KitSame
Catheter MarkingsMarkings (every cm)NoneDifference – seecomment #3:Catheter markings areadded to the subjectdevice to assist the user
SpecificationHydroMID 4F SingleLumenAccess Vascular Inc.Subject DeviceHydroMID-142 4FAccess Vascular Inc.Predicate DeviceK203069Comparison
catheter and monitoringcatheter migration

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image is a logo for Access Vascular. The word "ACCESS" is in a dark blue, sans-serif font, stacked on top of the word "VASCULAR" which is in a lighter blue, sans-serif font. The "A" in "ACCESS" has a line extending from the bottom left of the "A" to the top of the "V" in "VASCULAR", creating a visual connection between the two words.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image is a logo for Access Vascular. The word "Access" is in a dark blue font, and the word "Vascular" is in a lighter blue font. There is a line that connects the two words, and it looks like a heartbeat. The logo is simple and modern.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for Access Vascular. The word "ACCESS" is in a dark blue font, and the word "VASCULAR" is in a lighter blue font. A blue line connects the two words, starting from the bottom left of the "A" in "ACCESS" and ending at the bottom left of the "V" in "VASCULAR".

The purpose of listing the K161866 reference device in this submission is intended to provide information that PICC catheters with distance markings are effective in assisting users in both positioning the catheter and monitoring catheter migration. In prior Access Vascular submission K203069 we utilized the BioFlo Midline device a predicate device and demonstrated substantial equivalence despite minor differences in technological characteristics (i.e. catheter shaft material).

Comment #1 – Removal of statement about therapies not appropriate for midlines

The proposed HydroMID Catheter indication for use statement removes mention of restricted therapies for midline catheters to align with slight changes to INS standards without impacting the intended use. The proposed indication for use is identical to other marketed midline catheters. A warning was related to therapies not appropriate for midline catheters include those requiring central access. The difference between the subject device and the predicate device are minor and do not raise any different question of safety and effectiveness.

Comment #2 – Addition of Copper Phthalocyanine Blue 15:1 material

Materials of construction of the HydroMID Catheter with catheter markings are the same as those of the HydroMID catheter (K203069), with minor changes to accommodate the introduction of catheter markings. Bench testing and biocompatibility testing was done to support the ink used in the catheter distance markings. The difference between the subject device and predicate device are minor and do not raise any different question of safety and effectiveness as determined through testing.

Comment #3 - Addition of Catheter Markings

The addition of catheter markings to the subject device assists the user in positioning the catheter and monitor catheter migration. This change is proposed in response to user feedback and is consistent with currently marketed Midline catheters. Testing included bench testing and biocompatibility testing. The difference between the subject device and predicate device are minor and do not raise any different question of safety and effectiveness as determined through testing.

The HydroMID is used for the same intended use in the same anatomical location using the same principles of operation as the predicate device. Therefore, the HydroMID can be considered substantially equivalent to the predicate device.

Performance Data

All necessary performance testing has been conducted on the HydroMID to assure substantial equivalence to the predicate devices and to demonstrate the device performs as intended. All testing was performed on test units representative of finished devices.

The device passed the following tests, which were conducted in accordance with noted

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains the logo for Access Vascular. The logo is comprised of the word "ACCESS" in a dark blue font, stacked on top of the word "VASCULAR" in a lighter blue font. To the left of the words is a stylized "A" that is also in a dark blue font.

standards:

TestConsensus Standard/FDA Guidance/Description
Bench testing, including Catheter length testing Catheter marking spacing Surface continuity In-life durability testingConfirm that the device meets intended productspecifications
Biocompatibility testingFDA Final Guidance Document, "Use ofInternationalStandard ISO 10993-1, "Biological evaluation ofmedicaldevices - Part 1: Evaluation and testing within ariskmanagement process" (September 2020)

The following standards were used to support testing:

  • EN ISO 10555-1:2013 Intravascular Catheters Sterile and Single Use Catheters Part 1: General ● Requirements
  • . EN ISO 10555-3:2013 – Intravascular Catheters – Sterile and Single Use Catheters Part 3: Central Venous Catheters
  • . EN ISO 11135:2014 Sterilization of Healthcare Products - Ethylene Oxide - Requirements for the Development, Validation, and Routine Control of a Sterilization Process for Medical Devices
  • AAMI TIR28:2016 Product Adoption and Process Equivalence for Ethylene Oxide Sterilization
  • ISO 15223-1:2016 – Symbols to be used with medical device labels, labelling, and information to be supplied - Part 1: General Requirements
  • . ISO 10993-1: 2018 - Biological evaluation of medical devices – Evaluation and testing within a risk management process
  • ISO 10993-4: 2017 - Biological evaluation of medical devices – Selection of tests for interactions with blood
  • . ISO 10993-5: 2009 - Biological evaluation of medical devices - Tests for in vitro cytotoxicity
  • ISO 10993-10: 2010 - Biological evaluation of medical devices – Tests for irritation and skin sensitization
  • ISO 10993-12: 2021 - Biological evaluation of medical devices - sample preparation and reference materials

Conclusion:

Upon reviewing the information provided in this submission and comparing the intended use, principle of operation and overall technological characteristics, the HydroMID 4F Single Lumen Midline Catheter is substantially equivalent to existing legally marketed devices.

§ 880.5200 Intravascular catheter.

(a)
Identification. An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.(b)
Classification. Class II (performance standards).