K Number
K220609

Validate with FDA (Live)

Date Cleared
2022-05-27

(86 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, the glove was tested for chemotherapy drugs permeation.

Device Description

The proposed Powder-Free Nitrile Examination Glove (White, Blue, Black, Orange and Green Color) Tested for use with Chemotherapy Drugs and Fentanyl is a disposable device intended for over-the-counter use and is provided powder-free and non-sterile. These patient examination gloves are formulated using Nitrile. The patient examination gloves are not made with natural rubber latex.

AI/ML Overview

The provided document describes the acceptance criteria and study results for Powder-Free Nitrile Examination Gloves.

Here's the requested information:

1. Table of Acceptance Criteria and Reported Device Performance:

CharacteristicStandard / Acceptance CriteriaDevice Performance (Color: White, Blue, Black, Orange, Green)
Dimension (Length)ASTM D6319-19 (XS: 220mm, S: 220mm, M: 230mm, L: 230mm, XL: 230mm, XXL: 230mm minimum)All sizes and colors meet or exceed minimum length requirements (e.g., XS: 239-244mm, M: 230-245mm).
Dimension (Width)ASTM D6319-19 (XS: 70±10mm, S: 80±10mm, M: 95±10mm, L: 110±10mm, XL: 120±10mm, XXL: 130±10mm)All sizes and colors meet width requirements (e.g., XS: 75-77mm, M: 86-97mm).
Dimension (Finger Thickness)ASTM D6319-19 (0.05mm minimum)All sizes and colors meet or exceed minimum finger thickness (e.g., White: 0.07-0.10mm, Orange: 0.27-0.28mm).
Dimension (Palm Thickness)ASTM D6319-19 (0.05mm minimum)All sizes and colors meet or exceed minimum palm thickness (e.g., White: 0.05-0.06mm, Orange: 0.18-0.22mm).
Freedom from HolesASTM D6319-19, 21 CFR 800.20 (AQL 2.5)Passed for all sizes and colors.
Physical Property (Tensile Strength Before Aging)ASTM D6319-19 (14 MPa minimum)All sizes and colors meet or exceed 14 MPa (e.g., White: 20.5-23.2 MPa).
Physical Property (Elongation Before Aging)ASTM D6319-19 (500% minimum)All sizes and colors meet or exceed 500% (e.g., White: 520-536%).
Physical Property (Tensile Strength After Aging)ASTM D6319-19 (14 MPa minimum)All sizes and colors meet or exceed 14 MPa (e.g., White: 20.2-26.8 MPa).
Physical Property (Elongation After Aging)ASTM D6319-19 (400% minimum)All sizes and colors meet or exceed 400% (e.g., White: 427-488%).
Residual PowderASTM D6319-19 (≤ 2.0 mg of residual powder per glove)All sizes and colors meet this requirement (e.g., White: 0.74-0.98 mg).
Chemotherapy Drug Permeation (Breakthrough Detection Time)>240 minutes for most drugs (except Carmustine and ThioTepa)Performance is explicitly listed in Table 5.1 and Table 6.3. For example, Cisplatin: >240 minutes for all colors. Carmustine and ThioTepa have lower breakthrough times and are explicitly warned against.
Biocompatibility (Skin Irritation)ISO 10993-10 (Non-irritant)Non-irritant under the conditions of the study for all colors.
Biocompatibility (Skin Sensitization)ISO 10993-10 (Non-sensitizer)Non-sensitizer under the conditions of the study for all colors.
Biocompatibility (Acute Systemic Toxicity)ISO 10993-11 (Not acutely systemically toxic)Not an acute systemic toxic under the condition of the study for all colors.

2. Sample Size Used for the Test Set and Data Provenance:

The document does not explicitly state the specific sample sizes used for each test (e.g., how many gloves were tested for dimensions, holes, or chemical permeation). It lists "Device Performance (Minimum)" in tables, implying that measurements were taken across a sample.

The data provenance is not explicitly mentioned (e.g., country of origin for the testing or if it was retrospective/prospective). However, the submitter's address is Malaysia: "Lot 18256, Kawasan Perindustrian Lot Q, Kertih Bio-Polymer Park, 24300 Kemaman, Terengganu, Malaysia." The testing standards cited are international (ASTM, ISO).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

Not applicable. This device is a medical glove, and the "ground truth" for its performance is established through objective, standardized physical and chemical permeation tests, and biocompatibility studies, not through expert consensus or interpretation of medical images/conditions. The performance is measured against established scientific and engineering standards (ASTM and ISO).

4. Adjudication Method for the Test Set:

Not applicable. The tests are objective measurements against defined standards, not subjective interpretations requiring adjudication.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This is not an AI-powered diagnostic or interpretive device.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done:

Not applicable. This is a physical medical device (glove), not an algorithm.

7. The Type of Ground Truth Used:

The ground truth for the performance of the medical gloves is established through:

  • Standardized Physical and Mechanical Testing: Against ASTM D6319-19 requirements for dimensions, tensile strength, elongation, and residual powder.
  • Permeation Testing: Against ASTM D6978-05 for chemotherapy drugs.
  • Biocompatibility Testing: Against ISO 10993-5 (In Vitro Cytotoxicity - although the proposed device showed potential cytotoxicity, it was then deemed safe by the Acute Systemic Toxicity test), ISO 10993-10 for skin irritation and sensitization, and ISO 10993-11 for acute systemic toxicity.

These are all objective, laboratory-based measurement criteria rather than expert consensus, pathology, or outcomes data in a clinical setting.

8. The Sample Size for the Training Set:

Not applicable. This is a physical medical device, not a machine learning model requiring a training set.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as there is no training set for a physical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol is a stylized representation of a caduceus, a traditional symbol of medicine, with intertwined snakes and wings. To the right of the symbol is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue, with "FDA" in a larger font size than the rest of the text.

May 27, 2022

Encompass Industries Sdn. Bhd. Nur Muhyiddin OA Manager Lot 18256, Kawasan Perindustrian Lot Q Kertih BioPolymer Park Kemaman, Terengganu 24300 Malaysia

Re: K220609

Trade/Device Name: Powder-Free Nitrile Examination Glove (White, Blue, Black, Orange and Green Color) Tested for Use with Chemotherapy Drugs and Fentanyl Regulation Number: 21 CFR 880.6250 Regulation Name: Non-powdered patient examination glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: February 24, 2022 Received: March 2, 2022

Dear Nur Muhyiddin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bifeng Qian, M.D., Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220609

Device Name

Powder-Free Nitrile Examination Glove (White, Blue, Black, Orange and Green Color) Tested for Use with Chemotherapy Drugs and Fentanyl

Indications for Use (Describe)

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, the glove was tested for chemotherapy drugs permeation.

The following chemotherapy drugs have been tested with the glove:

Chemotherapy DrugsConcentrationBreakthrough Detection Time in Minutes
WhiteBlueBlackOrangeGreen
Carmustine3.3 mg/ml11.210.711.067.367.1
Cisplatin1.0 mg/ml>240>240>240>240>240
Cyclophosphamide20.0 mg/ml>240>240>240>240>240
Dacarbazine10.0 mg/ml>240>240>240>240>240
Doxorubicin HCL2.0 mg/ml>240>240>240>240>240
Etoposide20.0 mg/ml>240>240>240>240>240
Fluorouracil50.0 mg/ml>240>240>240>240>240
Methotrexate25.0 mg/ml>240>240>240>240>240
Paclitaxel6.0 mg/ml>240>240>240>240>240
ThioTepa10.0 mg/ml25.221.511.147.8169.1
Mitomycin C0.5 mg/ml>240>240>240>240>240
Vincristine Sulfate1.0 mg/ml>240>240>240>240>240
Fentanyl Citrate100 mcg/2mL>240>240>240>240>240

*Please note that the following 2 drugs have extremely low permeation times:

    1. Carmustine
    1. Thiotepa

Warning: Do not use with Carmustine and ThioTepa.

Type of Use (Select one or both, as applicable)

] Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

K220609

Preparation date: May 27, 2022

1. Submitter
Name:Encompass Industries Sdn. Bhd.
Address:Lot 18256, Kawasan Perindustrian Lot Q,Kertih Bio-Polymer Park, 24300 Kemaman,Terengganu, Malaysia.
Telephone No.:+609 831 8866
Contact Person:Nur Atikah binti Muhammad Muhyiddin
E-mail:regulatory@encompass-medical.com atikah@ems-inc.com

2. Identification of Device

Trade/Proprietary Name(s): 1. Powder-Free Nitrile Examination Glove (White, Blue, Black, Orange and Green Color) Tested for use with Chemotherapy Drugs and Fentanyl2. Other clients' trade name and private labeling
Common Name(s): Powder-Free Nitrile Examination Glove
Classification Name:: 1. Non-powdered Patient Examination Glove (21 CFR 880.6250)2. Patient Examination Glove, Specialty (Product Code: LZC)3. Polymer Patient Examination Glove (Product Code: LZA)
Device Classification: Class I

3. Identification of Legally Marketed Device as Predicate

Predicate name: Black Colored, Powder Free Nitrile Examination Gloves, Non-sterile,Tested for Use with Chemotherapy Drugs and Fentanyl Citrate
510(k) number: K200181
Company: Comfort Rubber Glove Industries Sdn Bhd

4. Description of Device

The proposed Powder-Free Nitrile Examination Glove (White, Blue, Black, Orange and Green Color) Tested for use with Chemotherapy Drugs and Fentanyl is a disposable device intended for over-the-counter use and is provided powder-free and non-sterile. These patient examination gloves are formulated using Nitrile. The patient examination gloves are not made with natural rubber latex. General specifications of the glove are as below:

1. Overall Length:230 mm minimum
2. Width:95 ± 5 mm minimum (for medium glove)
3. Palm Thickness:0.05 mm minimum
4. Finger Thickness:0.05 mm minimum
5. Tensile Strength
a. Before Aging:14 MPa minimum

{5}------------------------------------------------

b. After Aging:14 MPa minimum
6. Ultimate Elongation
a. Before Aging:500 % minimum
b. After Aging:400 % minimum
7. Pinhole AQL:2.5

5. Intended Use / Indication for Use:

A Powder-Free Nitrile Examination Glove (White, Blue, Black, Orange and Green Color) Tested for use with Chemotherapy Drugs and Fentanyl is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, these gloves were tested for chemotherapy drugs permeation test.

Table 5.1 Summary Test Results for Resistance of Powder-Free Nitrile Examination Gloves (White, Blue, Black, Orange and Green Color) to Permeation by Chemotherapy Drugs

Chemotherapy DrugConcentrationMinimum Breakthrough Detection Time (minutes)
WhiteBlueBlackOrangeGreen
Carmustine (BCNU)*3.3 mg/ml11.210.711.067.367.1
Cisplatin1.0 mg/mL>240>240>240>240>240
Cyclophosphamide20.0 mg/mL>240>240>240>240>240
Dacarbazine10.0 mg/mL>240>240>240>240>240
Doxorubicin HCL2.0 mg/mL>240>240>240>240>240
Etoposide20.0 mg/mL>240>240>240>240>240
Fluorouracil50.0 mg/mL>240>240>240>240>240
Methotrexate25.0 mg/mL>240>240>240>240>240
Paclitaxel6.0 mg/mL>240>240>240>240>240
ThioTepa*10.0 mg/mL25.221.511.147.8169.1
Mitomycin C0.5 mg/mL>240>240>240>240>240
Vincristine Sulfate1.0 mg/mL>240>240>240>240>240
Fentanyl Citrate100.0 mcg/2mL>240>240>240>240>240
*Please note that the following drugs have extremely low permeation time:
1. Carmustine (3.3 mg/mL)
  1. Thiotepa (10 mg/mL)

Warning: Do not use with Carmustine and ThioTepa.

6. Comparative Technological Characteristics & Performance Information Summary

Table 6.1 Summary of Technological Characteristics Comparison between Proposed Device and Predicate Device

Device Performance / CharacteristicComparison Analysis
Character-isticsRequirement /StandardPredicate Device(K200181)Subject DeviceComparisonAnySafety &Effective-ness Issue
Material-NitrileNitrileSimilarNo
Color-BlackWhiteBlueBlackOrangeGreenMinorDifferent(Supported byBiocompatibility Study)No
ProductCodeLZALZCLZALZCSimilarNo
Type ofUse-Single UseSingle UseSimilarNo
Material-NitrileNitrileSimilarNo
Powder-Free-YesYesSimilarNo
Sterility-Non-sterileNon-SterileSimilarNo
IntendedUse /Indicationsfor Use-The Black Colored,Powder Free NitrileExamination Gloves,Non-Sterile, and Testedfor Use withChemotherapy Drugs andFentanyl Citrate is aspecialty medical glovewhich is a disposabledevice intended formedical purposes that isworn on the examiner'shand or finger to preventcontamination betweenexaminer and patient. Inaddition, these gloves areworn to protect thewearer against exposureto chemotherapy drugsand Fentanyl Citrate.Powder-Free NitrileExamination Glove (White,Blue, Black, Orange andGreen Color) Tested forUse with ChemotherapyDrugs and Fentanyl is adisposable device intendedfor medical purposes that isworn on the examiner'shand, in order to preventcontamination betweenpatient and examiner, andalso prevent transmission ofa wide variety of diseases toboth patient and health carepersonnel. In addition,these gloves were tested forchemotherapy drugpermeation test.SimilarNo
DimensionASTM D6319-10Meets requirementMeets requirementSimilarNo
PhysicalPropertiesASTM D6319-10Meets requirementMeets requirementSimilarNo
Freedomfrom HolesASTM D6319-10(Tested accordingto ASTM D5151-06)Meets requirements ofASTM D6319-10(ASTM D5151)Meets ASTM D6319-10,21 CFR 800.20 (Tested inaccordance with ASTMD5151-06)SimilarNo
PowderResidualASTM D6124-06Meets requirementMeets requirementSimilarNo
Biocompatibility.In VitroCytotoxicity-ISO 10993-5Under the conditions ofthe study, the device isnon-cytotoxicUnder the condition ofstudy, the device extractshows potential toxicity.Different(Supported byAcuteSystemicToxicityStudy)No
Acute SystemicToxicity - ISO10993-11-Not an acute systemic toxicunder the condition of thestudy.Different(Device isproven safe toused)No
Skin Sensitization- ISO 10993-10Under the condition ofthe study, the device isnon-sensitizationNot a sensitizer under theconditions of the study.SimilarNo
Skin Irritation -ISO 10993-10Under the conditions ofthe study, the device isnon irritatingNot a primary skin irritantunder the conditions of thestudy.SimilarNo

{6}------------------------------------------------

{7}------------------------------------------------

CharacteristicPredicate Device(K200181)Proposed DeviceComparison AnalysisJustification
ColorBlackWhiteBlueBlackOrangeGreenMinor Different(Supported byBiocompatibilityStudy)Device in different color hasthe same non-sensitization,non-irritating and non-toxiccharacteristics.
Biocompatibility(In VitroCytotoxicity)Meets ISO 10993:5Under theconditions of thestudy, the device isnon-cytotoxicISO 10993:5Under the condition ofthe study, thedevice extract shows potentialcytotoxicity.Different(Supported byAcute SystemicToxicity Study)Device has potential cytotoxiceffect. Therefore, AcuteSystemic Toxicity test isconducted to prove thatdevice does not have potentialtoxicity.
Biocompatibility(Acute SystemicToxicity)-Meets ISO 10993-11Not an acute systemictoxic under thecondition of the study.Different(Device is provensafe to use)Device is does not havepotential toxicity effect.Therefore, similar with thepredicate.

Table 6.2 Summary of Difference in Technological Characteristics

Table 6.3: Test Results for Resistance of Medical Gloves to Permeation by Chemotherapy Drugs (ASTM D6978-05)

ChemotherapyDrugConcentrationMinimum Breakthrough Detection Time (minutes)
Predicate Device(K200181)Proposed Device
WhiteBlueBlackOrangeGreen
Carmustine (BCNU)*3.3 mg/ml54.111.210.711.067.367.1
Cisplatin1.0 mg/mL>240>240>240>240>240>240
Cyclophosphamide20.0 mg/mL>240>240>240>240>240>240
Dacarbazine10.0 mg/mL>240>240>240>240>240>240
Doxorubicin HCL2.0 mg/mL>240>240>240>240>240>240
Etoposide20.0 mg/mL>240>240>240>240>240>240
Fluorouracil50.0 mg/mL>240>240>240>240>240>240
Methotrexate25.0 mg/mL->240>240>240>240>240
Paclitaxel6.0 mg/mL>240>240>240>240>240>240
ThioTepa*10.0 mg/mL1.325.221.511.147.8169.1
Mitomycin C0.5 mg/mL->240>240>240>240>240
Vincristine Sulfate1.0 mg/mL->240>240>240>240>240
Fentanyl Citrate100.0 mcg/2mL>240>240>240>240>240>240
*Warning Statement:The maximum testing time is 240 minutes. Please note that the following drugs have extremely low permeation time:1. Carmustine (3.3 mg/mL)2. Thiotepa (10 mg/mL)

{8}------------------------------------------------

7. Summary of Non-Clinical Test:

Below are the non-clinical tests that was conducted and the purposes:

TestPurpose
Dimension TestTo evaluate whether the device meets the current ASTM D6319-19
Freedom From HoleTo evaluate whether the subject meets current 21 CFR 800.20, ASTMD6319-19 and test according to ASTM D5151-19
Physical Property TestTo evaluate whether the subjects meet current ASTM D6319-19 and testaccording to ASTM D412 and ASTM D573
Residual Powder TestTo evaluate whether the subjects meet current ASTM D6319-19 and testaccording to ASTM D6124-06
Biocompatibility Test(Skin Irritation)To determine whether skin irritation potential of device meets ISO 10993-10.
Biocompatibility Test(Skin Sensitization)To determine whether skin sensitization potential of device meets ISO10993-10.
Biocompatibility Test(Acute SystemicToxicity)To determine whether leachable extracted from the device would causeacute systemic toxicity and meet ISO 10993-11.

Table 7.1 Summary of Dimension Test

StandardMinimumDevice Performance (Minimum)
CharacteristicsSizeRequirementWhiteBlueBlackOrangeGreen
LengthASTMD6319-19XS220 mm239 mm235 mm242 mm243 mm244 mm
S220 mm236 mm237 mm242 mm244 mm241 mm
M230 mm230 mm230 mm240 mm245 mm245 mm
L230 mm234 mm237 mm255 mm243 mm242 mm
XL230 mm232 mm246 mm240 mm240 mm245 mm
XXL230 mm238 mm238 mm242 mm240 mm245 mm
WidthASTMD6319-19XS$70 \pm 10$ mm75 mm77 mm77 mm77 mm77 mm
S$80 \pm 10$ mm85 mm87 mm86 mm87 mm85 mm
M$95 \pm 10$ mm93 mm86 mm92 mm97 mm97 mm
L$110 \pm 10$ mm104 mm107 mm104 mm108 mm107 mm
XL$120 \pm 10$ mm112 mm111 mm114 mm118 mm117 mm
XXL$130 \pm 10$ mm128 mm128 mm127 mm124 mm126 mm
Thickness(Finger)ASTMD6319-19XS0.05 mm0.08 mm0.08 mm0.09 mm0.27 mm0.25 mm
S0.08 mm0.09 mm0.05 mm0.27 mm0.27 mm
M0.10 mm0.11 mm0.10 mm0.28 mm0.27 mm
L0.09 mm0.08 mm0.10 mm0.27 mm0.27 mm
XL0.07 mm0.06 mm0.10 mm0.28 mm0.27 mm
XXL0.07 mm0.08 mm0.07 mm0.28 mm0.26 mm
Thickness(Palm)ASTMD6319-19XS0.05 mm0.05 mm0.05 mm0.06 mm0.19 mm0.19 mm
S0.05 mm0.05 mm0.06 mm0.19 mm0.19 mm
M0.06 mm0.06 mm0.05 mm0.19 mm0.18 mm
L0.05 mm0.05 mm0.06 mm0.22 mm0.20 mm
XL0.05 mm0.05 mm0.06 mm0.22 mm0.18 mm
XXL0.05 mm0.05 mm0.06 mm0.20 mm0.19 mm

{9}------------------------------------------------

CharacteristicsStandardSizeMinimumRequirementDevice Performance (Minimum)
WhiteBlueBlackOrangeGreen
Freedom fromHolesA;STMD6319-1921 CFR800.20(Testedas perASTMD5151-19)XSAQL 2.5PassedPassedPassedPassedPassed
SPassedPassedPassedPassedPassed
MPassedPassedPassedPassedPassed
LPassedPassedPassedPassedPassed
XLPassedPassedPassedPassedPassed
XXLPassedPassedPassedPassedPassed

Table 7.2 Summary of Freedom of Hole Test

Table 7.3 Summary of Physical Property Test

CharacteristicsStandardSizeMinimumRequirementDevice Performance (Minimum)
WhiteBlueBlackOrangeGreen
Before Aging(Tensile [MPa])XS14 MPa23.221.021.320.721.1
S21.224.523.620.020.1
M22.821.622.321.028.8
L20.723.721.621.420.9
XL20.520.224.320.922.0
XXL22.823.622.223.022.1
Before Aging(Elongation[%])ASTMD6319-19XS500 %528520515527533
S527549518515556
M520522530549548
L528521527531523
XL536521528533524
XXL526524528525525
After Aging(Tensile [MPa])(Testedas perASTMD412-16)XS14 MPa23.020.223.024.521.4
S20.224.626.622.720.4
M26.819.822.722.226.4
L24.926.121.425.221.7
XL26.323.128.822.224.6
XXL23.321.020.027.420.1
After Aging(Elongation[%])XS400 %432461414430465
S427466421412446
M444457425466440
L488423436447455
XL458424434432431
XXL437435433425425

{10}------------------------------------------------

CharacteristicsStandardSizeMinimumRequirementDevice Performance (Minimum)
WhiteBlueBlackOrangeGreen
ResidualPowder(mg)ASTMD6319-19(Testedas perASTMD6124-06)XS≤ 2.0 mg ofresidualpowder perglove0.980.680.800.820.96
S0.860.720.860.980.82
M0.920.880.940.920.92
L0.780.760.920.780.90
XL0.840.980.840.680.86
XXL0.740.820.920.840.98

Table 7.4 Summary of Residual Powder Test

Table 7.5 Summary of Biocompatibility Test

CharacteristicsStandardMinimumRequirementDevice Performance
WhiteBlueBlackOrangeGreen
Skin IrritationISO 10993-10N/ANon-irritant under the conditions of the study.
SkinSensitizationISO 10993-10N/ANon-sensitizer under the conditions of the study.
Acute SystemicToxicityISO 10993-11N/ANot an acute systemic toxic under the condition ofthe study.

8. Summary of Clinical Test:

Clinical data is not required.

9. Conclusion:

The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.