K Number
K220578

Validate with FDA (Live)

Date Cleared
2022-05-25

(86 days)

Product Code
Regulation Number
882.5890
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Transcutaenous Electrical Nerve Stimulator (Model RJTENS-1) is an electrical nerve stimulator indicated for use for pain relief by applying an electrical current to electrodes on a patient's skin to treat pain. In particular, this device is indicated for use for symptomatic relief of chronic intractable pain and adjunctive treatment of post surgical and post traumatic pain.

Device Description

Transcutaneous Electrical Nerve Stimulator (Model RJTENS-1) sends gentle electrical current to underlying nerves and muscle group via electrodes applied on the skin to relieve pain. The device has 16 programs (12 standard programs and 4 editable programs). It is a battery-powered portable device, comprising electronic stimulatory module and accessories of lead wires, electrodes and 4x1.5V AAA alkaline batteries. Two outlet sockets are used to connect skin electrodes by lead wires. The accessories of electrodes is 510(k) cleared device (K213879), Size 50*50mm.

AI/ML Overview

The provided text describes the 510(k) summary for the Transcutaneous Electrical Nerve Stimulator (Model RJTENS-1). Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided information:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for this device are implicitly derived from its substantial equivalence to the predicate device (K202893). The "Comparison" column in the table indicates "Same" or "Similar" for most parameters, meaning the new device's performance aligns with the established performance of the predicate. Note 1 explicitly addresses the difference in deviation from amplitude, frequency, and pulse width.

ParameterAcceptance Criteria (Predicate Device K202893)Reported Device Performance (Model RJTENS-1)
Indications for UseSame as predicateSame as predicate
Type of UsePrescription usePrescription use
Power Source(s)1.5Vx4 AAA alkaline battery1.5Vx4 AAA alkaline battery
Method of Line Current IsolationN/AN/A
Patient Leakage Current - Normal Condition2μΑ2μΑ
Patient Leakage Current - Single Fault ConditionNANA
Average DC current through electrodes<0.01μA<0.01μA
Number of programs1616
Number of Output channels22
Synchronous or Alternating?SynchronousSynchronous
Method of Channel IsolationBy TransformerBy Transformer
Regulated Current or Regulated Voltage?Current controlCurrent control
Software/Firmware/Microprocessor Control?YesYes
Automatic Overload Trip?NoNo
Automatic No-Load Trip?NoNo
Automatic Shut Off?YesYes
User Override Control?YesYes
Indicator Display - On/Off Status?YesYes
Indicator Display - Low Battery?YesYes
Indicator Display - Voltage/Current Level?YesYes
Timer Range (minutes)10~90 min10~90 min
Compliance with Voluntary Standards?ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-10ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-10
Compliance with 21 CFR 898?YesYes
Weight (grams)Approx. 96g without batteryApprox. 96g without battery
Dimensions (mm)1406428 mm1406428 mm
Housing Materials & ConstructionABSABS
WaveformMonophasicMonophasic
ShapeRectangular pulseRectangular pulse
Maximum Output Voltage (volts)$30V±20% @500Ω$$30V±10% @500Ω$
Maximum Output Current (mA)$60mA±20% @500Ω$$60mA±10% @500Ω$
Pulse width (μsec)$75-300μs±20%$$75-300μs±10%$
Pulse Period (msec)8.33-1000ms8.33-1000ms
Max. pulse frequency (Hz)$1-120Hz±20%$$1-120Hz±10%$
Net Charge (µC per pulse)$0.65μC @500Ω$$0.65μC @500Ω$
Maximum Phase Charge (μC)$18μC @500Ω$$18μC @500Ω$
Maximum Average Current (mA)$2.16mA @500Ω$$2.16mA @500Ω$
Maximum Average Power Density (mW/cm²)2.59mW/cm² @500Ω2.59mW/cm² @500Ω
AccessoriesElectrodes, cables, batteryElectrodes, cables, battery

Study to Prove Acceptance Criteria:

The study proving the device meets the acceptance criteria is a non-clinical bench test study.

2. Sample Size for the Test Set and Data Provenance:

  • Sample Size: Not explicitly stated as "sample size" in the traditional sense for an AI/diagnostic study. For bench tests, it typically refers to the number of devices tested. This information is not provided in the document.
  • Data Provenance: Not applicable in the context of clinical data for this type of submission, as no clinical data was included. For bench tests, the data would have originated from laboratory testing of the manufactured devices. The manufacturer is Bozhou Rongjian Medical Appliance Co., Ltd. in China.

3. Number of Experts Used to Establish Ground Truth and Qualifications:

  • This information is not applicable/not provided. The assessment of these types of electrical stimulators primarily relies on objective physical and electrical measurements against established standards, not necessarily on expert interpretation of ambiguous data points.

4. Adjudication Method for the Test Set:

  • This information is not applicable/not provided. Adjudication methods like 2+1 or 3+1 are used for expert consensus in interpretive tasks (e.g., radiology reads), which is not the nature of the bench tests reported here.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

  • No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Clinical data was not including in this submission."

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done:

  • Not applicable. This device is a Transcutaneous Electrical Nerve Stimulator, not an AI algorithm. Its performance is evaluated through its physical and electrical characteristics according to safety and performance standards.

7. The Type of Ground Truth Used:

  • The ground truth for this device's performance evaluation is established through compliance with recognized voluntary standards and objective physical and electrical measurements as defined by those standards.
    • ANSI AAMI ES60601-1: 2005/(R) 2012 And A1: 2012, C1: 2009/(R) 2012 And A2: 2010/(R) 2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1: 2005, MOD)
    • IEC 60601-2-10 Edition 2.1 2016-04, Medical Electrical Equipment - Part 2-10: Particular Requirements For The Basic Safety And Essential Performance Of Nerve And Muscle Stimulators
    • IEC 60601-1-2 Edition 4.0 2014-02 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests

8. The Sample Size for the Training Set:

  • Not applicable. The device is an electrical stimulator, not an AI/machine learning algorithm requiring a training set.

9. How the Ground Truth for the Training Set Was Established:

  • Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

May 25, 2022

Bozhou Rongjian Medical Appliance Co., Ltd. Wu Zhifang General Manager Jianghuai Supply Base, Zhongkai Group, Mengcheng County Bozhou, Anhui 233500 China

Re: K220578

Trade/Device Name: Transcutaneous Electrical Nerve Stimulator (Model RJTENS-1) Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: GZJ Dated: February 18, 2022 Received: February 28, 2022

Dear Wu Zhifang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220578

Device Name

Transcutaneous Electrical Nerve Stimulator (Model RJTENS-1)

Indications for Use (Describe)

The Transcutaenous Electrical Nerve Stimulator (Model RJTENS-1) is an electrical nerve stimulator indicated for use for pain relief by applying an electrical current to electrodes on a patient's skin to treat pain. In particular, this device is indicated for use for symptomatic relief of chronic intractable pain and adjunctive treatment of post surgical and post traumatic pain.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

[As required by 21 CFR 807.92]

1. Submission Information
510(k) Number:K220578
Date:April 25, 2022
Type of 510(k) Submission:Traditional 510(k)
Submitter/Manufacturer:Bozhou Rongjian Medical Appliance Co.,Ltd.
Jianghuai Supply Base, Zhongkai Group, Mengcheng County, BozhouCity, 233500 Anhui, China
Contactor:Wu ZhifangE-mail: doris.d@ceve.org.cnTel: +86-558-7662968
2. Device Description
Proprietary Name:Transcutaneous Electrical Nerve Stimulator (Model RJTENS-1)
Common Name:TENS (Transcutaneous Electrical Nerve Stimulator)
Classification Name:Transcutaneous electrical nerve stimulator for pain relief
Product Code:GZJ
Device Class:2
Regulation Number:21 CFR 882.5890
Review Panel:Neurology
Indications for use:Transcutaneous Electrical Nerve Stimulator (Model RJTENS-1) isan electrical nerve stimulator indicated for use for pain relief byapplying an electrical current to electrodes on a patient's skin to treatpain. In particular, this device is indicated for use for symptomatic reliefof chronic intractable pain and adjunctive treatment in the management ofpost surgical and post traumatic pain.
Device Description:Transcutaneous Electrical Nerve Stimulator (Model RJTENS-1) sendsgentle electrical current to underlying nerves and muscle group viaelectrodes applied on the skin to relieve pain.The device has 16 programs (12 standard programs and 4 editableprograms). It is a battery-powered portable device, comprising electronicstimulatory module and accessories of lead wires, electrodes and 4×1.5VAAA alkaline batteries.Two outlet sockets are used to connect skin electrodes by lead wires. Theaccessories of electrodes is 510(k) cleared device (K213879), Size50*50mm.

3. Predicate Device Identification

Predicate 510(k) Number:K202893
Marketing clearance date:June 18, 2021
Product name:Transcutaneous Electrical Nerve Stimulator
Manufacturer:Wuxi Jiajian Medical Instrument Co., Ltd

4. Substantially Equivalent Comparison Conclusion

{4}------------------------------------------------

ParametersNew DevicePredicate DeviceComparison
510(k) NumberK220578K202893--
Device NameTranscutaneous ElectricalNerve Stimulator (ModelRJTENS-1)Transcutaneous ElectricalNerve Stimulator--
ManufacturerBozhou Rongjian MedicalAppliance Co.,Ltd.Wuxi Jiajian MedicalInstrument Co., Ltd--
Indication for UseTranscutaneous Electrical NerveStimulator ((Model RJTENS-1) isan electrical nerve stimulatorindicated for use for pain reliefby applying an electricalcurrent to electrodes on apatient's skin to treat pain. Inparticular, this device is indicatedfor use for symptomatic relief ofchronic intractable pain andadjunctive treatment in themanagement of post surgical andpost traumatic pain.Transcutaneous ElectricalNerve Stimulator is an electricalnerve stimulator indicated foruse for pain relief by applyingan electrical current toelectrodes on a patient's skin totreat pain. In particular, thisdevice is indicated for use forsymptomatic relief of chronicintractable pain and adjunctivetreatment in the management ofpost surgical and post traumaticpain.Same
Type of usePrescription usePrescription useSame
Power Source(s)1.5Vx4 AAA alkaline battery1.5Vx4 AAA alkaline batterySame
- Method of LineCurrent IsolationN/AN/ASame
- Patient LeakageCurrent----Same
- NormalCondition (μΑ)2μΑ2μΑSame
- Single FaultCondition (μΑ)NANASame
Average DC currentthrough electrodeswhen device is on butno pulses are beingapplied (μΑ)<0.01μA<0.01μASame
Number of program1616Same
Number of Outputchannels:22Same
- Synchronousor Alternating?SynchronousSynchronousSame
- Method ofChannel IsolationBy TransformerBy TransformerSame
Regulated Current orRegulated Voltage?Current controlCurrent controlSame
Software/Firmware/YesYesSame

{5}------------------------------------------------

MicroprocessorControl?
Automatic OverloadTrip?NoNoSame
Automatic No-LoadTrip?NoNoSame
Automatic Shut Off?YesYesSame
User OverrideControl?YesYesSame
On/OffStatus?YesYesSame
IndicatorDisplayLowBattery?YesYesSame
Voltage/CurrentLevel?YesYesSame
Timer Range(minutes)10~90 min10~90 minSame
Compliance withVoluntary Standards?ANSI AAMI ES60601-1, IEC60601-1-2, IEC 60601-2-10ANSI AAMI ES60601-1, IEC60601-1-2, IEC 60601-2-10Same
Compliance with 21CFR 898?YesYesSame
Weight (grams)Approx.96g without batteryApprox.96g without batterySame
Dimensions (mm) [Wx H x D]1406428 mm1406428 mmSame
Housing Materials &ConstructionABSABSSame
WaveformMonophasicMonophasicSame
ShapeRectangular pulseRectangular pulseSame
Maximum OutputVoltage (volts)$30V±10% @500Ω$$30V±20% @500Ω$
Maximum OutputCurrent (specifyunits)$60mA±10% @500Ω$$60mA±20% @500Ω$SimilarNote 1
Pulse width (μsec)$75-300μs±10%$$75-300μs±20%$
Pulse Period (msec)8.33-1000ms8.33-1000ms
Max. pulse frequency(Hz) [or Rate (pps)]$1-120Hz±10%$$1-120Hz±20%$
Net Charge (µC perpulse)$0.65μC @500Ω$$0.65μC @500Ω$
Maximum PhaseCharge, (μC)$18μC @500Ω$$18μC @500Ω$Same
Maximum AverageCurrent, (mA)$2.16mA @500Ω$$2.16mA @500Ω$

{6}------------------------------------------------

Density, (mA/cm², r.m.s.)
Maximum AveragePower Density,(mW/cm²)2.59mW/cm² @500Ω2.59mW/cm² @500Ω
AccessoriesElectrodes, cables, batteryElectrodes, cables, batterySame

Comparison in details:

Note 1:

The deviation from the amplitude, frequency and pulse width of the proposed device are different from those of the predicate device. The proposed device has less deviation than the predicate device, which means that its output is more stable. And both the proposed device and predicate device passed ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-10 standards tests. Therefore, this difference will not raise any safety or effectiveness issue.

5. Non-Clinical Test Conclusion

Bench tests were conducted on Transcutaneous Electrical Nerve Stimulator (Model RJTENS-1) to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The following tests were performed on the proposed device:

• ANSI AAMI ES60601-1: 2005/(R) 2012 And A1: 2012, C1: 2009/(R) 2012 And A2: 2010/(R) 2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1: 2005, MOD);

· IEC 60601-2-10 Edition 2.1 2016-04, Medical Electrical Equipment - Part 2-10: Particular Requirements For The Basic Safety And Essential Performance Of Nerve And Muscle Stimulators;

· IEC 60601-1-2 Edition 4.0 2014-02 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests;

6. Clinical Test

Clinical data was not including in this submission.

7. Conclusions

The conclusions drawn from the non-clinical tests demonstrate that the device is as safe, as effective, and performs as well as the legally marketed device identified in the subject device is substantially equivalent to the predicate device.

§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).