K Number
K213879

Validate with FDA (Live)

Date Cleared
2022-01-05

(23 days)

Product Code
Regulation Number
882.1320
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Self-adhesive Electrode is intended to be used to transmit electrical stimulation current to the patient's skin. Example electrical stimulations for current applications of the electrodes are: TENS and EMS.

Device Description

The Self-adhesive electrode, lead wire type and snap button type, are non-sterile flexible structures, composed of materials commonly used in this application:
First layer: Non-woven fabric tape
Second layer: Polyurethane (PU) electrically conductive carbon cloth (Hebei Kangshengda Electronic Technology Co., Ltd)
Third layer: Biocompatible conductive hydrogel coupling media (ValueTrode Carbon, K970426), which has passed the required skin sensitivity testing criteria as specified in ISO 10993-10 and cytotoxicity testing criteria as specified in ISO 10993-5.
The electrodes are designed for single patient / multiple application use. It can be used for low-frequency or medium-frequency nerve or muscle stimulators, as the conduction film adhered to body skin.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the "Self-adhesive Electrode" (K213879), structured according to your requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The FDA 510(k) summary does not explicitly list "acceptance criteria" for each parameter in the way a formal study report might. Instead, it demonstrates substantial equivalence to a predicate device (K090198) by comparing various characteristics and reporting that the new device met specific test standards or manufacturer requirements.

The table below presents the comparative data and implies the "acceptance criteria" are generally that the new device's performance is similar to or meets the standards met by the predicate device, or adheres to established industry standards for such devices.

ParameterAcceptance Criteria (Implied / Indirect)Reported Device Performance (New Device)Remark (from document)
Formulation/Design Characteristics
Trade Name(Not applicable for acceptance criteria)Self-adhesive Electrode--
Common NameSame as predicateCutaneous electrodeSame
Classification NameSame as predicateElectrode, cutaneousSame
Product CodeSame as predicateGXYSame
Device ClassSame as predicateIISame
Regulation NumberSame as predicate882.1320Same
Medical SpecialtySame as predicateNeurologySame
Indications for useSame as predicateThe Self-adhesive Electrode is intended to be used to transmit electrical stimulation current to the patient's skin. Example electrical stimulations for current applications of the electrodes are: TENS and EMS.Same
Target populationSame as predicate (Single patient use and multiple application)Single patient use and multiple applicationSame
Type of useSame as predicate (OTC and Prescription use)OTC and Prescription useSame
Design (shape & connection)Similar to predicate, customized specification allowedRound, Rectangle, Elliptical, Calabash, Butterfly, Palm shape according to customized specification. Lead wire with female socket, or snap button with male snap connector.Similar
MaterialsSimilar to predicate and biocompatible standards met- Non-woven fabric tape- Polyurethane (PU) electrically conductive carbon cloth- Biocompatible conductive hydrogel coupling media (ValueTrode Carbon, K970426)Similar
Electrode Pad SizeWithin ranges comparable to predicateRound Min.Ø20mm; Max.Ø85mmRectangle Min.26×26mm; Max.190×110mmElliptical Min.85×45mm;Calabash Min.90×47mm; Max.200×88mmButterfly Min.73×53mm; Max.195×95mmPalm shape 74×47mmSimilar
Performance Characteristics
Electrode Impedance of Electrode PadWithin ranges comparable to predicateRound: 302577ΩRectangle: 303646ΩElliptical: 407676ΩCalabash: 401602ΩButterfly: 402626ΩPalm shape: 402638ΩSimilar
Patient contact area of electrodeWithin ranges comparable to predicateRound Min.3.14cm²; Max.56.716cm²Rectangle Min.6.76cm²; Max.209cm²Elliptical Min.38.25cm²; Max.240cm²Calabash Min.42.3cm²; Max.176cm²Butterfly Min.38.69cm²; Max.185.25cm²Palm shape 34.78cm²Similar
Max. current Density (IRMS=10mA calculation)Within ranges comparable to predicateRound: 3.18mA/cm²Rectangle: 1.48mA/cm²Elliptical: 0.52mA/cm²Calabash: 0.24mA/cm²Butterfly: 0.52mA/cm²Palm shape: 0.29mA/cm²Similar
Hydrogel thickness35mils ± 5mils (0.89mm ± 0.13mm)35mils ± 5mils (0.89mm ± 0.13mm)Same
Hydrogel pH4.2 ± 1.04.2 ± 1.0Same
Hydrogel volume resistivity1500 ohm-cm max1500 ohm-cm maxSame
Compliance/Safety Standards
Standards meetCompliance with AAMI/ANSI ES 60601-1, ANSI AAMI IEC 60601-2-2, FDA's Impedance Test, ASTM F1980, and manufacturer's requirements (peeling force, simulation use)• Lead wires test per 8.5.2.3 of AAMI/ANSI ES 60601-1;• Impedance test, Conformability test and Fluid tolerance test per 201.15.101.6 and 201.15.101.7 of ANSI AAMI IEC 60601-2-2 Edition 6.0 2017-03;• Impedance Test (Dispersion Test) according to FDA's requirement;• Peeling force test and Simulation use test according to manufacturer's requirement;• Shelf life test per ASTM F1980:2016Similar
BiocompatibilityISO10993-5; ISO10993-10ISO10993-5; ISO10993-10Same
Sterility StatusNon-sterile (as per predicate)Non-sterileSame
Electrical safetyLead wire meets Clause 8.5.2.3 of AAMI/ANSI ES60601-1Lead wire meets Clause 8.5.2.3 of AAMI/ANSI ES60601-1Same
Other PerformanceGood electrical conductivity, good adhesive property (as per predicate)Good electrical conductivity, good adhesive propertySame

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a separate "test set" in terms of patient data. The evaluation relies on non-clinical bench tests performed on the "Self-adhesive Electrode" device itself.

  • Sample Size: The document does not explicitly state the number of electrodes tested for each bench test. For example, it lists "Impedance test," "Conformability test," "Fluid tolerance test," "Peeling force test," and "Simulation use test" but doesn't provide the quantity of devices subjected to these tests.
  • Data Provenance: The tests were "Bench tests" conducted on the proposed device. The country of origin for the device manufacturer is China (Bozhou Rongjian Medical Appliance Co.,Ltd., Anhui, China). The data is retrospective in the sense that these tests were completed before the 510(k) submission.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable as the study did not involve human interpretation or a "ground truth" derived from expert consensus on clinical data. The tests performed were objective bench tests against established standards and comparative measurements.

4. Adjudication Method for the Test Set

This is not applicable as there was no test set involving subjective human assessment or consensus building. The assessments were based on direct measurements and compliance with technical standards.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a MRMC comparative effectiveness study was not done. The submission explicitly states: "Clinical data was not including in this submission."

6. If a Standalone (Algorithm Only Without Human-in-the-loop Performance) Was Done

This is not applicable. The device is a self-adhesive electrode, a physical medical device, not an algorithm or software. Therefore, an "algorithm only" performance study is irrelevant. The compliance for the device involves physical and electrical characteristics.

7. The Type of Ground Truth Used

For this device, the "ground truth" or reference for performance and safety is established by:

  • Established industry standards: Such as ISO 10993 series for biocompatibility, AAMI/ANSI ES 60601-1, and IEC 60601-2-2 for electrical safety and performance.
  • Manufacturer's internal requirements: For tests like peeling force and simulation use.
  • Predicate device characteristics: Demonstrating that the new device's properties (materials, dimensions, electrical impedance ranges) are similar to those of a legally marketed predicate device (K090198).

There is no clinical ground truth (e.g., pathology, outcomes data) as no clinical study was performed.

8. The Sample Size for the Training Set

This information is not applicable. This is a physical medical device, not an AI/ML algorithm that requires a "training set."

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 5, 2022

Bozhou Rongjian Medical Appliance Co., Ltd. % Doris Dong Manager Shanghai CV Technology Co., Ltd. Room 903, No.19 Dongbao Road, Songjiang Area Shanghai, Shanghai 201613 China

Re: K213879

Trade/Device Name: Self-adhesive Electrode Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode Regulatory Class: Class II Product Code: GXY Dated: November 24, 2021 Received: December 13, 2021

Dear Doris Dong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213879

Device Name Self-adhesive Electrode

Indications for Use (Describe)

The Self-adhesive Electrode is intended to be used to transmit electrical stimulation current to the patient's skin. Example electrical stimulations for current applications of the electrodes are: TENS and EMS.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

[As required by 21 CFR 807.92]

1. Submission Information

510(k) Number:K213879
Date:November 24, 2021
Type of 510(k) Submission:Traditional 510(k)
Manufacturer/Applicant:Bozhou Rongjian Medical Appliance Co.,Ltd.Jianghuai Supply Base, Zhongkai Group, Mengcheng County, BozhouCity, 233500 Anhui, ChinaTel: +86-558-7963368E-mail: rongjianzl@rjmed.com.cn
Contactor:Doris Dong (Consultant)Shanghai CV Technology Co., Ltd.Rm. 903, No. 19 Dongbao Rd., Songjiang Area, Shanghai, 201613 ChinaE-mail: doris.d@ceve.org.cnTel: 86 21-31261348

2. Device Description

2. Device Description
Proprietary Name:Self-adhesive Electrode
Common Name:Cutaneous electrode
Classification Name:Electrode, cutaneous
Product Code:GXY
Device Class:II
Regulation Number:21 CFR 882.1320
Review Panel:Neurology
Indications for use:The Self-adhesive Electrode is intended to be used to transmit electrical stimulation current to the patient's skin. Example electrical stimulations for current applications of the electrodes are: TENS and EMS.
Device Description:The Self-adhesive electrode, lead wire type and snap button type, are non-sterile flexible structures, composed of materials commonly used in this application:First layer: Non-woven fabric tapeSecond layer: Polyurethane (PU) electrically conductive carbon cloth (Hebei Kangshengda Electronic Technology Co., Ltd)Third layer: Biocompatible conductive hydrogel coupling media (ValueTrode Carbon, K970426), which has passed the required skin sensitivity testing criteria as specified in ISO 10993-10 and cytotoxicity testing criteria as specified in ISO 10993-5.The electrodes are designed for single patient / multiple application use. It can be used for low-frequency or medium-frequency nerve or muscle stimulators, as the conduction film adhered to body skin.

{4}------------------------------------------------

There are six shapes of round, rectangle, elliptical, calabash, butterfly and palm shape. For the electrical connection, Rongjian provides lead wire type and snap button type:

Lead wire assembly - at least 40mm long wire with 2mm/2.5mm diameter female socket.

Snap button assembly - with 2.5~5mm diameter male socket.

The lead wire assembly is in compliance with the requirements of FDA performance standard 21 CFR part 898 by testing under ANSI/AAMI ES60601-1, subclause 8.5.2.3.

3. Predicate Device Identification

K090198 - JIAJIAN Self-adhesive Electrode

4. Non-Clinical Test Conclusion

Bench tests were conducted on Self-adhesive Electrode to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The following tests were performed on the proposed device:

· ASTM F1980-16, Standard guide for accelerated aging of sterile barrier systems for medical devices. (Sterility)

• ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity.

· ISO 10993-10 Third Edition 2010-08-01, Biological evaluation of medical devices - Part 10: Tests for Irritation and Skin Sensitization.

• AAMI / ANSI ES60601-1:2005/(R)2012 And A1:2012, C1:2009/(R)2012 And A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And

Essential Performance (IEC 60601-1:2005, MOD).

· Impedance test, Conformability test and Fluid tolerance test per 201.15.101.6 and 201.15.101.7 of IEC 60601-2-2 Edition 6.0 2017-03.

· Impedance Test (Dispersion Test) according to FDA's requirement.

· Peeling force test and Simulation use test according to manufacturer's requirement.

5. Substantial Equivalent Based on Assessment of Clinical PerformanceData

Clinical data was not including in this submission

{5}------------------------------------------------

ParametersNew DevicePredicate DeviceRemark
1510(k) NumberTo be assignedK090198--
2510(k) HolderBozhou Rongjian MedicalAppliance Co.,Ltd.Wuxi Jiajian MedicalInstrument Co., Ltd--
3Trade NameSelf-adhesive ElectrodeJiajian Self- adhesive Electrode--
4Common NameCutaneous electrodeCutaneous electrodeSame
5ClassificationNameElectrode, cutaneousElectrode, cutaneousSame
6Product CodeGXYGXYSame
7RegulationNumber882.1320882.1320Same
8MedicalSpecialtyNeurologyNeurologySame
9Device ClassIIIISame
10Indications foruseThe Self-adhesive Electrode isintended to be used to transmitelectrical stimulation current to thepatient's skin. Example electricalstimulations for currentapplications of the electrodes are:TENS and EMS.The self-adhesive electrode isintended to be used to applyelectrical stimulation current tothe patient's skin. Exampleelectrical stimulations forcurrent applications of theelectrodes are: TENS and EMS.Same
11TargetpopulationSingle patient use and multipleapplicationSingle patient use and multipleapplicationSame
12Type of useOTC and Prescription useOTC and Prescription useSame
13Design (shape &connection)Round, Rectangle, Elliptical,Calabash, Butterfly, Palm shapeaccording to customizedspecification.Lead wire with female socket, orsnap button with male snapconnector.Round, Rectangle, Oval, Gourd,Butterfly, Saddle according tocustomized specification.Lead wire with female socket.Similar
14Materials- Non-woven fabric tape- Polyurethane (PU) electricallyconductive carbon cloth (HebeiKangshengda ElectronicTechnology Co., Ltd)- Biocompatible conductivehydrogel coupling media(ValueTrode Carbon, K970426)- Non-woven fabric tape- Electrically conductive carboncloth (ValueTrode Carbon,K970426)- Biocompatible conductivehydrogel coupling media(ValueTrode Carbon, K970426)Similar
15Electrode PadSizeRound Min.Ø20mm;Max.Ø85mmRound Min.Ø20mm;Max.Ø80mmSimilar
Rectangle Min.26×26mm;Max.190×110mmRectangle Min.3×10mm;Max.80×130mm
Elliptical Min.85×45mm;Oval Min.50×120mm;

6. Substantially Equivalent Comparison Conclusion

{6}------------------------------------------------

Max.240×100mmMax.100×240mm
CalabashMin.90×47mm;Max.200×88mmGourdMin.85×50mm;Max.170×100mm
ButterflyMin.73×53mm;Max.195×95mmButterflyMin.55×75mm;Max.95×165mm
Palm shape74×47mmSaddleMin.75×110mm;Max.150×220mm
16ElectrodeImpedance ofElectrode PadRound302~577ΩRound290~1000ΩSimilar
Rectangle303~646ΩRectangle
Elliptical407~676ΩOval
Calabash401~602ΩGourd
Butterfly402~626ΩButterfly
Palm shape402~638ΩSaddle
17--Patient contactarea of electrodeRoundMin.3.14cm²;Max.56.716cm²RoundMin.3.14cm²;Max.50.24cm²Similar
RectangleMin.6.76cm²;Max.209cm²RectangleMin.0.3cm²;Max.104cm²
EllipticalMin.38.25cm²;Max.240cm²OvalMin.60cm²;Max.240cm²
CalabashMin.42.3cm²;Max.176cm²GourdMin.42.5cm²;Max.170cm²
ButterflyMin.38.69cm²;Max.185.25cm²ButterflyMin.41.25cm²;Max.156.75cm²
Palm shape34.78cm²SaddleMin.82.5cm²;Max.330cm²
18Max. currentDensity ofElectrode (UseIRMS=10mA forcalculation )Round3.18mA/cm²(Ø20mm: 3.14cm²)Round3.18mA/cm²(Ø20mm:3.14cm²)Similar
Rectangle1.48mA/cm²(26×26mm: 6.76cm²)Rectangle33.3mA/cm²(3×10mm:0.3cm²)
Elliptical0.52mA/cm²(85×45mm/2:19.13cm²)Oval0.17mA/cm²(50×120mm:60cm²)
Calabash0.24mA/cm²(90×47mm: 42.3cm²)Gourd0.24mA/cm²(85×50mm:42.5cm²)
Butterfly0.52mA/cm²(73×53mm/2:19.35cm²)Butterfly0.48mA/cm²(55×75mm/2:20.625cm²)
Palm shape0.29mA/cm²(74×47mm:34.78cm²)Saddle0.24mA/cm²(75×110mm/2:41.25cm²)
19Hydrogel35mils ± 5mils (0.89mm ±35mils ± 5mils (0.89mm ±Same

{7}------------------------------------------------

thickness0.13mm)0.13mm)
20Hydrogel pH4.2 ± 1.04.2 ± 1.0Same
21Hydrogelvolumeresistivity1500 ohm-cm max1500 ohm-cm maxSame
22Standards meet• Lead wires test per 8.5.2.3 ofAAMI/ANSI ES 60601-1;• Impedance test, Conformabilitytest and Fluid tolerance test per201.15.101.6 and 201.15.101.7 ofANSI AAMI IEC 60601-2-2Edition 6.0 2017-03;• Impedance Test (Dispersion Test)according to FDA's requirement;• Peeling force test and Simulationuse test according tomanufacturer's requirement;• Shelf life test per ASTMF1980:2016ISO14971;ISO 13485;EN 980;ANSI/AAMI EC12;IEC 60601-1.Similar
23BiocompatibilityISO10993-5; ISO10993-10ISO10993-5; ISO10993-10Same
24Sterility StatusNon-sterileNon-sterileSame
25Electrical safetyLead wire meets Clause 8.5.2.3 ofAAMI/ANSI ES60601-1Lead wire meets IEC 60601-1Same
26OtherPerformanceGood electrical conductivity, goodadhesive propertyGood electrical conductivity,good adhesive propertySame

7. Conclusions

Based on the conclusions drawn from the non-clinical tests demonstrate that the device is as safe, as effective, and performs as well as the legally marketed devices identified in the submission. Thus the subject device is substantially equivalent to the predicate device K090198.

§ 882.1320 Cutaneous electrode.

(a)
Identification. A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.(b)
Classification. Class II (performance standards).