K Number
K220560

Validate with FDA (Live)

Device Name
Dermatac Drape
Manufacturer
Date Cleared
2022-07-14

(136 days)

Product Code
Regulation Number
878.4780
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Dermatac™ Drape is an accessory to the following V.A.C.® Therapy Negative Pressure Wound Therapy Systems:

3M™ ActiV.A.C.™, 3M™ V.A.C.® Simplicity, and 3M™ V.A.C.Via™ Negative Pressure Wound Therapy . Systems, which are integrated wound management systems for use in acute, extended and home care settings. . The V.A.C.ULTA™ and V.A.C.RX4™ Negative Pressure Wound Therapy Systems are integrated wound management systems for use in acute care settings and other professional healthcare environments where product use is conducted by or under the supervision of a qualified healthcare professional.

When used on open wounds, these systems are intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention, by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation, and by removing exudate and infectious material. Open wound types include: chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure or venous insufficiency) flaps and grafts.

When used on closed surgical incisions, they are intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudates via the application of negative pressure wound therapy.

The Dermatac™ Drape is also an accessory to the V.A.C. VERAFLO™ Therapy (Instillation) provided by the V.A.C.ULTATM Therapy Unit.

V.A.C. VERAFLO™ Therapy is indicated for patients who would benefit from vacuum assisted drainage and controlled delivery of topical wound treatment solutions and suspensions over the wound bed.

V.A.C. VERAFLO™ Therapy is indicated for patients with chronic, sub-acute and dehisced wounds, partial-thickness burns. ulcers (such as diabetic, pressure or venous insufficiency) flaps and grafts.

Device Description

The Dermatac™ Drape is a semi-occlusive wound drape that is used as an accessory to the V.A.C.® Therapy and V.A.C. VERAFLO™ Therapy Systems. The Dermatac™ Drape is a single-use, sterile covering that provides a sealed environment for delivery of these therapies. It also allows for a moist wound environment.

The drape consists of a polyurethane film with acrylic adhesive with a perforated silicone layer. The perforations in the silicone laver expose the acrylic adhesive coated on the polyurethane film. The acrylic adhesive secures the drape to the periwound skin, thus creating a sealed wound environment.

The therapy systems with which it is used are comprised of the following:

  • Software controlled therapy unit that provides negative pressure and, in the case of ● V.A.C. VERAFLO™ Therapy, a pump for controlled delivery of topical wound solutions.
  • Disposable canister which collects wound exudate and, in the case of V.A.C. ● VERAFLO™ Therapy, instilled solutions
  • Foam dressing for placement into the wound ●
  • Semi-occlusive drape that covers the dressing ●
  • Tubing set that connects the dressing to the therapy unit. ●
AI/ML Overview

The provided text is a 510(k) premarket notification for a medical device called the Dermatac™ Drape. This document focuses on demonstrating substantial equivalence to a predicate device, rather than proving the device meets specific performance criteria through a study involving AI or human interpretation of medical images.

Therefore, many of the requested points related to AI acceptance criteria, training/test set sample sizes, expert ground truth, MRMC studies, and standalone algorithm performance cannot be answered based on the provided text. The document describes a physical medical device (a wound drape), not a diagnostic algorithm.

However, I can extract the information that is present:

1. A table of acceptance criteria and the reported device performance:

Since this is a 510(k) for a physical device, the "acceptance criteria" are not framed as typical performance metrics for an AI system. Instead, the testing focuses on validating the fundamental functions and safety of the drape.

Acceptance Criteria (Test Performed)Reported Device Performance
V.A.C.™ Negative Pressure Maintenance System TestMaintained negative pressure within specifications and managed fluid exudate without unexpected alarms.
Package Integrity testingSterile barrier integrity maintained throughout its labeled shelf life.
Product performance testing of dressing components after ETO sterilization (Dressing Extensibility, Moisture Vapor Transmission Rate, Peel Adhesion Force, Wet Peel Adhesion Testing, Release Liner Testing)Product functioned as intended throughout its labeled shelf life.
Biocompatibility Testing (Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Material-Mediated Pyrogenicity, Subacute Systemic Toxicity, Genotoxicity, Implantation)Favorable biocompatibility test data for all relevant endpoints.

2. Sample sizes used for the test set and the data provenance:

  • Sample sizes: Not explicitly stated for each test. The document refers to "testing" which implies a sample was used, but the exact number of units or iterations is not provided.
  • Data provenance: Not specified in terms of country of origin. The studies are described as "non-clinical tests." The document states "Clinical and Pre-clinical testing were not necessary to demonstrate equivalence," implying these were laboratory or benchtop tests rather than patient-based studies.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. This is not a study requiring expert interpretation or ground truth establishment in the context of medical image analysis. The tests are physical/chemical performance evaluations.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Not applicable. This is not a study requiring adjudication of expert interpretations.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. This is a physical medical device, not an AI system. No MRMC study was performed.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • No. This is a physical medical device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • The "ground truth" for the performance tests of the physical device would be based on engineering specifications, material science standards, and established biological/chemical safety protocols (e.g., ISO standards for biocompatibility tests). For example, "maintaining negative pressure within specifications" implies a predefined acceptable range, which serves as the "ground truth" for that test.

8. The sample size for the training set:

  • Not applicable. This is a physical medical device, not an AI system that requires a "training set."

9. How the ground truth for the training set was established:

  • Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

July 14, 2022

3M Melanie Avila Regulatory Affairs Manager 6203 Farinon Drive San Antonio, Texas 78249

Re: K220560

Trade/Device Name: Dermatac™ Drape Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OMP Dated: February 28, 2022 Received: February 28, 2022

Dear Melanie Avila:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Julie Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220560

Device Name Dermatac™ Drape

Indications for Use (Describe)

The Dermatac™ Drape is an accessory to the following V.A.C.® Therapy Negative Pressure Wound Therapy Systems:

3MTM ActiV.A.C.TM, 3M™ V.A.C.® Simplicity, and 3MTM V.A.C.Via™ Negative Pressure Wound Therapy .

Systems, which are integrated wound management systems for use in acute, extended and home care settings. . The V.A.C.ULTA™ and V.A.C.RX4™ Negative Pressure Wound Therapy Systems are integrated wound management systems for use in acute care settings and other professional healthcare environments where product use is conducted by or under the supervision of a qualified healthcare professional.

When used on open wounds, these systems are intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention, by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation, and by removing exudate and infectious material. Open wound types include: chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure or venous insufficiency) flaps and grafts.

When used on closed surgical incisions, they are intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudates via the application of negative pressure wound therapy.

The Dermatac™ Drape is also an accessory to the V.A.C. VERAFLO™ Therapy (Instillation) provided by the V.A.C.ULTATM Therapy Unit.

V.A.C. VERAFLO™ Therapy is indicated for patients who would benefit from vacuum assisted drainage and controlled delivery of topical wound treatment solutions and suspensions over the wound bed.

V.A.C. VERAFLO™ Therapy is indicated for patients with chronic, sub-acute and dehisced wounds, partial-thickness burns. ulcers (such as diabetic, pressure or venous insufficiency) flaps and grafts.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

3M Health Care Business Group 6203 Farinon San Antonio. TX 78249

Contact Person: Melanie Avila Email: mavila9@mmm.com Phone: 210-275-5038 Date Prepared: 14 February 2022

Name of Subiect Device: Dermatac™ Drape Predicate Device: V.A.C. DERMATAC™ Drape (K200390) Common or Usual Name: Dressing component of Negative Pressure Wound Therapy (NPWT) System Classification Name: Neqative Pressure Wound Therapy Powered Suction Pump (and components) Regulatory Number: 21 CFR 878.4780 Regulatory Class: Class II Product Code: OMP

Device Description

The Dermatac™ Drape is a semi-occlusive wound drape that is used as an accessory to the V.A.C.® Therapy and V.A.C. VERAFLO™ Therapy Systems. The Dermatac™ Drape is a single-use, sterile covering that provides a sealed environment for delivery of these therapies. It also allows for a moist wound environment.

The drape consists of a polyurethane film with acrylic adhesive with a perforated silicone layer. The perforations in the silicone laver expose the acrylic adhesive coated on the polyurethane film. The acrylic adhesive secures the drape to the periwound skin, thus creating a sealed wound environment.

The therapy systems with which it is used are comprised of the following:

  • Software controlled therapy unit that provides negative pressure and, in the case of ● V.A.C. VERAFLO™ Therapy, a pump for controlled delivery of topical wound solutions.
  • Disposable canister which collects wound exudate and, in the case of V.A.C. ● VERAFLO™ Therapy, instilled solutions
  • Foam dressing for placement into the wound ●
  • Semi-occlusive drape that covers the dressing ●
  • Tubing set that connects the dressing to the therapy unit. ●

Intended Use / Indications for Use

The Dermatac™ Drape is an accessory to the following V.A.C.® Therapy Negative Pressure Wound Therapy Systems:

{5}------------------------------------------------

3M™ ActiV.A.C.™, 3M™ V.A.C.® Simplicity, and 3M™ V.A.C.Via™ Negative Pressure Wound . Therapy Systems, which are integrated wound management systems for use in acute, extended and home care settings.

The V.A.C.ULTA™ and V.A.C.RX4™ Negative Pressure Wound Therapy Systems are . integrated wound management systems for use in acute care settings and other professional healthcare environments where product use is conducted by or under the supervision of a qualified healthcare professional.

When used on open wounds, these systems are intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention, by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and perfusion, and by removing exudate and infectious material. Open wound types include: chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure or venous insufficiency) flaps and grafts.

When used on closed surgical incisions, they are intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudates via the application of negative pressure wound therapy.

The Dermatac™ Drape is also an accessory to the V.A.C. VERAFLO™ Therapy (Instillation) provided by the V.A.C.ULTA™ Therapy Unit.

V.A.C. VERAFLO™ Therapy is indicated for patients who would benefit from vacuum assisted drainage and controlled delivery of topical wound treatment solutions and suspensions over the wound bed.

V.A.C. VERAFLO™ Therapy is indicated for patients with chronic, acute, traumatic, sub-acute and dehisced wounds, partial-thickness burns, ulcers and venous insufficiency), flaps and grafts.

Summary of Technological Characteristics

Comparison of subject and predicate device:

  • Minor changes to the material formulation .
  • . Minor changes to device packaging.
  • . Additional manufacturing location
  • . Minor changes to labeling

The indications for use, technological characteristics and principles of operation have not changed.

A table comparing the key features of the subject and predicate devices is provided below.

Summary of the technological characteristics of the device compared to the predicate device [21 CFR 807.92(a)(6)]

{6}------------------------------------------------

CharacteristicSubject Device:Predicate Device: Dermatac Drape, K200390
Indicated Wound TypesIdenticalChronic Acute Traumatic Subacute Dehisced wounds Partial-thickness burns Ulcers (such as diabetic, pressure or venous insufficiency) Flaps and grafts Surgically closed incisions (For V.A.C. Therapy only)
V.A.C. Negative Pressure Wound Therapy Units3M™ ActiV.A.C.™ 3M™ V.A.C.® Simplicity 3M™ V.A.C.Via™ V.A.C.ULTA™ V.A.C.RX4™ACTIV.A.C.™ V.A.C. SIMPLICITY™ V.A.C.VIA™ V.A.C. FREEDOM™ V.A.C.ULTA™, INFOV.A.C™ V.A.C.RX4™
Use environment/Care Setting of dressing kitIdenticalAcute, extended and home care settings

Performance Data

Summary of non-clinical tests conducted for determination of substantial equivalence:

  • . V.A.C.™ Negative Pressure Maintenance System Test demonstrates the Dermatac™ Drape maintains negative pressure within specifications and manages fluid exudate without unexpected alarms.
  • . Package Integrity testing to ensure the sterile barrier integrity is maintained throughout its labeled shelf life.
  • Product performance testing of dressing components after ETO sterilization to verify . the product functions as intended throughout its labeled shelf life.
    • o Dressing Extensibility
    • Moisture Vapor Transmission Rate o
    • Peel Adhesion Force o
    • Wet Peel Adhesion Testing o
    • o Release Liner Testing
  • Summary of biocompatibility Testing .

{7}------------------------------------------------

EndpointStudy Type(Test System)Guidelines
CytotoxicityMEM Elution(L-929 Mouse Fibroblast Cells)USFDA GLP 21CFR58ISO 10993-5 (2009)ISO 10993-12 (2007)
SensitizationGuinea Pig Maximization Sensitization(Hartley guinea pigs)USFDA GLP 21CFR58ISO 10993-10 (2010)ISO 10993-12 (2012)
IrritationIntracutaneous Reactivity(New Zealand White Rabbit)USFDA GLP 21CFR58ISO 10993-10 (2010)ISO 10993-12 (2012)
Acute SystemicToxicityAcute Systemic Injection(Swiss Mice)USFDA GLP 21CFR58ISO 10993-11 (2006)ISO 10993-12 (2012)
Material-MediatedPyrogenicityRabbit Pyrogenicity Test(New Zealand White Rabbits)USFDA GLP 21CFR58ISO 10993-11 (2006)ISO 10993-12 (2012)
Subacute SystemicToxicity32-day Repeated Dose Subacute Toxicity(Sprague Dawley rats)US FDA GLP 21CFR58ISO 10993-11 (2006)ISO 10993-12 (2012)
GenotoxicityBacterial Mutagenicity - Ames Assay(Salmonella typhimurium [TA97a, TA98, TA100and TA1535] and Escherichia coli [WP2-uvrA])USFDA GLP 21CFR58ISO 10993-3 (2014)ISO 10993-12 (2012)
ImplantationIn Vitro Mouse Lymphoma with ExtendedTreatment(L5178Y cells)USFDA GLP 21CFR58ISO 10993-6 (2016)

The product had favorable biocompatibility test data for all relevant endpoints.

In all instances, the Dermatac™ Drape functioned as intended and all test results observed were as expected.

Clinical and Pre-clinical testing were not necessary to demonstrate equivalence. In addition, human factors enqineering testing was not required since the subject device has the same user interface and use environment as the predicate.

Conclusions

The subject device is as safe and effective as the predicate device. The subject device's fundamental technology and principles of operation are unchanged compared to the predicate device. The subject device's Intended Use remains the same from the predicate device as cleared under K200390.

The minor differences between the subject device and its predicate do not significantly affect the safety or effectiveness of the device, nor did they represent a change in intended use. The performance data

{8}------------------------------------------------

demonstrates that the Dermatac™ Drape is as safe and effective as the predicate. Thus, the Dermatac™ Drape is substantially equivalent to the predicate.

§ 878.4780 Powered suction pump.

(a)
Identification. A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.(b)
Classification. Class II.