K Number
K220298

Validate with FDA (Live)

Device Name
CerroZone Mobile
Manufacturer
Date Cleared
2022-07-01

(149 days)

Product Code
Regulation Number
880.5045
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CerroZone Mobile is intended as a room recirculating air cleaner. The system is used for filtering out and inactivating airborne particles (i.e., bacteria and viruses) from the air for medical purposes.

Device Description

The CerroZone Mobile is a medical recirculating air cleaner that uses multiple inactivation processes consisting of reactive oxidizing species (ozone) and ultraviolet radiation to inactivate bacteria and viruses in the air. The CerroZone Mobile may be used in medical facilities. Once turned on, inlet air is passed through CerroZone Mobile's inline flow-through design which inactivates bacteria and viruses in a single pass flow-through. The device generates 220 cubic feet of air per minute (CFM), or 2.36 air changes per hour (ACH) in a standard 5,600 cu. Ft room. This process of air being inlet, purified, then outlet, lasts approximately 1.2 seconds. A 4 LOG reduction of airborne particles in a standard 579 cu. Ft room is achieved in 30 minutes or less. A 5.67 LOG, or 99.9998%, reduction is achieved in 45 minutes or less.

AI/ML Overview

The CerroZone Mobile device is a medical recirculating air cleaner designed to filter out and inactivate airborne bacteria and viruses. The following information summarizes its acceptance criteria and the studies conducted.

1. Table of Acceptance Criteria and Reported Device Performance:

Test MethodologyPurposeAcceptance CriteriaReported Device Performance
MS2 bacteriophage aerosolization in a sealed environmental bioaerosol chamberTo evaluate the efficacy of the CerroZone Mobile at reducing viability of aerosolized MS2 bacteriophage by a combination of entrainment and destructionGreater than 4 log reduction (99.99%)Average net log reduction for MS2 bacteriophage: 5.67 ± 0.19 / 60 minutes
Methicillin Resistant Staphylococcus epidermidis (MRSE) aerosolization in a single-pass flow-through chamber connected to the CerroZone MobileTo evaluate the efficacy of the CerroZone Mobile at reducing viability of aerosolized bacteria by a combination of entrainment and destructionGreater than 3 log reduction (99.9%) in single-pass testingDemonstrated effectiveness against Staphylococcus epidermidis
Ozone EmissionsTo assess ozone emissionsEmissions within the maximum acceptable level of ozone given in 21 CFR 801.415 (0.050 ppm)Emits between 0.000 and 0.010 ppm of Ozone with a measured absolute max of 0.036 ppm. All emissions fall within the acceptable level.

2. Sample Size Used for the Test Set and Data Provenance:

The document does not explicitly state the sample sizes for the biological aerosolization tests (MS2 bacteriophage and MRSE). The studies appear to be laboratory-based non-clinical tests rather than human subject or field trials, so terms like "country of origin" and "retrospective/prospective" might not be strictly applicable in the same way they would be for clinical studies. The context implies these were controlled experiments conducted to assess device performance under specific conditions.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

This information is not provided in the document. The "ground truth" for these non-clinical tests would have been established by the direct measurement of reduction in viable microorganisms using standard microbiological techniques, rather than expert interpretation.

4. Adjudication Method for the Test Set:

Not applicable. The tests involved direct measurement of microbial reduction, not subjective interpretation requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

No, an MRMC comparative effectiveness study was not done. The CerroZone Mobile is a medical device for air cleaning, not an interpretative diagnostic tool, so such a study would not be relevant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

These were standalone performance tests of the device itself (algorithm not applicable), evaluating its ability to reduce airborne pathogens directly.

7. The Type of Ground Truth Used:

The ground truth used for the efficacy studies was based on quantitative laboratory measurements of microbial viability (e.g., plaque assays for bacteriophages, colony counts for bacteria) before and after exposure to the device. For ozone emissions, the ground truth was direct quantitative measurement of ozone concentration.

8. The Sample Size for the Training Set:

Not applicable. This device is a physical air cleaner, not an AI/ML algorithm. Therefore, there is no "training set" in the context of machine learning.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 1, 2022

CerroZone % Matthieu Kirkland Regulatory Specialist Ram+ 2251 San Diego Avenue, Ste B-257 San Diego, California 92110

Re: K220298

Trade/Device Name: CerroZone Mobile Regulation Number: 21 CFR 880.5045 Regulation Name: Medical Recirculating Air Cleaner Regulatory Class: Class II Product Code: FRF Dated: May 23, 2022 Received: May 24, 2022

Dear Matthieu Kirkland:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220298

Device Name CerroZone Mobile

Indications for Use (Describe)

The CerroZone Mobile is intended as a room recirculating air cleaner. The system is used for filtering out and inactivating airborne particles (i.e., bacteria and viruses) from the air for medical purposes.

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for "CERR*ZONE". The word "CERR" is in black, and the word "ZONE" is in blue. There is a star symbol between the two words. The letters are all capitalized, and there is a trademark symbol in the upper right corner of the word "ZONE".

510(k) Summary K220298

DATE PREPARED

July 01, 2022

MANUFACTURER AND 510(k) OWNER

CerroZone P.O Box 66800, St. Louis, MO 63166 +1-618-337-6000 Telephone: Official Contact: Marlon E. Robinson, Director of Operations

REPRESENTATIVE/CONSULTANT

Matthieu Kirkland, M.S. Allison C. Komiyama, Ph.D., RAC RQM+ Telephone: +1 (412) 899-7424 mkirkland@rqmplus.com, akomiyama@rqmplus.com Email:

DEVICE INFORMATION

Proprietary Name/Trade Name:CerroZone Mobile
Common Name:Medical Recirculating Air Cleaner
Regulation Number:21 CFR 880.5045
Class:II
Product Code:FRF
Premarket Review:Infection Control and Plastic Surgery Devices (DHT4B)
Review Panel:General Hospital

PREDICATE DEVICE IDENTIFICATION

The CerroZone Mobile is substantially equivalent to the following predicate:

510(k)NumberPredicate Device Name / ManufacturerPrimaryPredicateReferenceDevice
K200321Novaerus NV1050 / Novaerus US Inc
K203189AEROCURE-MD / Aerobiotix, Inc.

The predicate and reference devices have not been subject to a design related recall.

DEVICE DESCRIPTION

The CerroZone Mobile is a medical recirculating air cleaner that uses multiple inactivation processes consisting of reactive oxidizing species (ozone) and ultraviolet radiation to inactivate bacteria and viruses in the air. The CerroZone Mobile may be used in medical facilities. Once turned on, inlet air is passed through CerroZone Mobile's inline flow-through design which inactivates bacteria and viruses in a single pass flow-through. The device generates 220 cubic feet of air per minute (CFM), or 2.36 air changes per hour (ACH) in a standard 5,600 cu. Ft room. This process of air being inlet, purified, then outlet, lasts approximately 1.2 seconds. A 4 LOG reduction of airborne particles in a standard 579 cu. Ft room is achieved in 30 minutes or

{4}------------------------------------------------

less. A 5.67 LOG, or 99.9998%, reduction is achieved in 45 minutes or less.

INDICATIONS FOR USE

The CerroZone Mobile is intended as a room recirculating air cleaner. The system is used for filtering out and inactivating airborne particles (i.e., bacteria and viruses) from the air for medical purposes.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

The following table summarizes the similarities and differences between the subject and predicate device.

{5}------------------------------------------------

Subject DevicePrimary Predicate DeviceComparison
CerroZoneNovaerus US Inc.
CerroZone MobileCurrentNovaerus NV1050K200321
ImageImage: CerroZone Mobile CurrentImage: Novaerus NV1050
Indications for UseThe CerroZone Mobile is intended as a room recirculating air cleaner. The system is used for filtering out and inactivating airborne particles (i.e., viruses and bacteria) from the air for medical purposesThe Novaerus NV1050 is intended as a room recirculating air cleaner. The system is used for filtering out and inactivating airborne particles from the air for medical purposesSimilar
Product Codes /Regulation NumberFRF/ 21 CFR 880.5045FRF/ 21 CFR 880.5045Identical
Regulation DescriptionMedical recirculating air cleanerMedical recirculating air cleanerIdentical
Type of UseOver-the-Counter UseOver-the-Counter UseIdentical
Mechanism of ActionMicroorganisms including viruses and bacteria, are inactivated by the device via damage by multiple inactivation processes:Reactive Oxidizing Species (ozone)Ultraviolet RadiationAlso, a filter to trap the resulting virus/bacteria particulates:Particulate FilterActivated Carbon FilterMicroorganisms including viruses and bacteria, are inactivated by the device via damage by multiple inactivation processes:Reactive Oxidizing Species (ozone)Ultraviolet RadiationIonsElectronsHigh Electric FieldsAlso, a filter to trap the resulting virus/bacteria particulates:Particulate FilterActivated Carbon FilterSimilar
Reduction of BiologicalAgentsMS2 phage reduced by 4 log reduction in 30 minutes or less and 5.67 log reduction in 45 minutes or less when operating at full fan speed in a standard room of 579 ft³ (16 m³)MS2 phage reduced by 3 log reduction in 10 minutes and 4 log reduction in 15 minutes when operating at full fan speed in a room of 580ft³ (16.4 m³)Similar
Ozone EmittedWithin the maximum acceptable level of ozone given in 21 CFR 801.415Within the maximum acceptable level of ozone given in 21 CFR 801.415Identical
InstallationFree standingFree standingIdentical
Reactive Oxidizing Species(Ozone) SourceHigh-output UVC generating lamp elementsHigh-output Plasma Field generating coil elementsSimilar
Reactive Oxidizing Species(Ozone) RemovalActivated Carbon FilterProprietary Catalyst SubstrateActivated Carbon FilterSimilar
Air SourceCentrifugal FanCentrifugal FanIdentical
Device Air Changes PerHour (ACH)2.36 device air changes per hour in a 5,600 ft³ room2 device air changes per hour in a 3,200 – 16,000 ft³ roomSimilar
UV Light or ReactiveOxidative Species (Ozone)Exposure Safety FeaturesA catalyst substrate converts any ozone generated back into oxygen. An activated carbon filter absorbs any remaining ozone.If > 0.03 ppm of ozone is measured the external ozone sensors, the CerroZone Mobile will automatically shut down. The CerroZone Mobile also utilizes a second and redundant externalAn activated carbon filter absorbs any ozone generated.Similar
ozone sensor that runs on a different circuit. Assuch, if one external ozone senor were to stopfunctioning, another sensor will be active.Safety feature confirmed by UL 867 & CSAC22.2#187
Input Voltage120 Volt110 VoltDifferent
DimensionsUnit Dimensions:Height: 59 inWidth: 29 inDepth: 17 inUnit Dimensions:Height: 59 inWidth: 27.5 inDiameter: 8.27 inDifferent

{6}------------------------------------------------

{7}------------------------------------------------

SUMMARY OF NON-CLINICAL TESTING

The following tests were performed to demonstrate safety based on current industry standards:

Electromagnetic Compatibility and Electrical Safety: The subject device was tested in compliance to EN/IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests and UL 867 & CSA C22.2#187 Standard for Electrostatic Air Cleaners.

Determination of Turbulent Air Flow and/or Site Contamination: Risk of turbulent air flow and/or potential site contamination in areas that have a sterile field or controlled air flow has been assessed via computational fluid dynamics per ASHRAE/ASHE Standard 170-2017 Ventilation of Health Care Facilities.

Ozone Emissions: Ozone Emissions of the CerroZone Mobile have been assessed per Section 40 of UL 867 & CSA C22.2#187 Electrostatic Air Cleaners. Results of the test show that the CerroZone Mobile emits between 0.000 and 0.010 ppm of Ozone with a measured absolute max of 0.036 ppm. All Ozone emissions measured fall within the maximum acceptable level of ozone, 0.050 ppm, given in 21 CFR 801.415.

Test MethodologyPurposeAcceptance CriteriaResults
MS2 bacteriophage wereaerosolized into a sealedenvironmental bioaerosolchamber containing theCerroZone Mobile.To evaluate the efficacy ofthe CerroZone Mobile atreducing viability ofaerosolized MS2bacteriophage by acombination ofentrainment anddestructionGreater than 4 logreduction (99.99%)Average net log reduction /timeMS2 bacteriophage,$5.67 \pm 0.19$ / 60 mins
Methicillin ResistantStaphylococcus epidermidis(MRSE) were aerosolizedinto a single-pass flowthrough chamberconnected to theCerroZone Mobile.To evaluate the efficacy ofthe CerroZone Mobile atreducing viability ofaerosolized bacteria by acombination ofentrainment anddestructionGreater than 3 logreduction (99.9%) in single-pass testingDemonstratedeffectiveness againstStaphylococcus epidermidis

SUMMARY OF CLINICAL TESTING

No clinical data were provided in order to demonstrate substantial equivalence.

CONCLUSION

CerroZone Inc. concludes that the nonclinical tests demonstrate that the CerroZone Mobile is as safe, as effective, and performs as well as or better than the legally marketed predicate device.

§ 880.5045 Medical recirculating air cleaner.

(a)
Identification. A medical recirculating air cleaner is a device used to remove particles from the air for medical purposes. The device may function by electrostatic precipitation or filtration.(b)
Classification. Class II (performance standards).