Search Results
Found 21 results
510(k) Data Aggregation
(124 days)
FRF
The CerroZone Mini is intended as a room recirculating air cleaner. The system is used for filtering out and inactivating airborne particles (i.e., bacteria and viruses) from the air for medical purposes.
The CerroZone Mini is a medical recirculating air cleaner that uses multiple inactivation processes, including use of reactive oxidizing species (ozone) and ultraviolet radiation, to inactivate bacteria and viruses in the CerroZone Mini may be used in medical facilities. Once turned on, inlet air is passed through CerroZone Mini's inline flow-through design which inactivates bacteria and viruses in a single pass flow-through. The device generates 110 cubic feet of air per minute (CFM), or 2.00 air changes per hour (ACH) in a standard 5,600 cubic foot room. The process of air being inlet, purified, then outlet, lasts approximately 1.2 seconds. A 4 LOG reduction, or 99.99% reduction, of airborne particles in a standard 579 cubic foot room is achieved in 45 minutes or less.
The prompt requests information about a study proving that a medical device meets acceptance criteria. The provided text is an FDA 510(k) clearance letter for the CerroZone Mini, a medical recirculating air cleaner. The document details the device, its indications for use, comparison to a predicate device, and summarizes non-clinical testing.
However, the provided text does not contain the specifics of a study proving device acceptance criteria in the way one might expect for an AI/algorithm-driven medical device, which is what the prompt implies by asking about sample sizes for test and training sets, expert ground truth, adjudication methods, and MRMC studies. The device in question is an air cleaner, and the "studies" mentioned are laboratory tests assessing its ability to reduce airborne biological agents and ozone emissions, not clinical trials or AI performance evaluations involving human-in-the-loop scenarios.
Therefore, many of the requested points are not applicable to the data provided for this specific device. I will address only the points for which information is available in the text.
Here's the analysis based on the provided document:
1. A table of acceptance criteria and the reported device performance
Test Methodology | Purpose | Acceptance Criteria | Reported Device Performance |
---|---|---|---|
MS2 bacteriophage aerosolized into a sealed environmental bioaerosol chamber containing the CerroZone Mini. | To evaluate the efficacy of the CerroZone Mini at reducing viability of aerosolized MS2 bacteriophage by a combination of entrainment and destruction. | Greater than 4 log reduction (99.99%). | Average net log reduction / time, MS2 bacteriophage, 5.76 ± 0.20 / 45 mins. |
Methicillin Resistant Staphylococcus epidermidis (MRSE) aerosolized into a single-pass flow through chamber that modeled the CerroZone Mini. | To evaluate the efficacy of the CerroZone Mini at reducing viability of aerosolized bacteria by a combination of entrainment and destruction. | Greater than 3 log reduction (99.99%) in single-pass testing. | Demonstrated effectiveness against Staphylococcus epidermidis. |
Ozone Emissions assessed per Section 40 of UL 867 & CSA C22.2 No.187. | To ensure that ozone emissions are within acceptable levels. | Within the maximum acceptable level of ozone, 0.050 ppm, given in 21 CFR 801.415. | Emits between 0.000 and 0.002 ppm of Ozone with a measured absolute max of 0.002 ppm. |
Electromagnetic Compatibility and Electrical Safety (EN/IEC 60601-1-2, UL 867 & CSA C22.2 No.187). | To demonstrate safety regarding electromagnetic disturbances and electrical safety. | Compliance to EN/IEC 60601-1-2 and UL 867 & CSA C22.2 No.187. | The subject device is equivalent to a predicate device that tested in compliance. |
2. Sample sizes used for the test set and the data provenance
The document does not specify exact "sample sizes" in terms of number of runs or specific data points for the biological agent reduction tests, nor does it explicitly state the precise "data provenance" (e.g., country of origin, retrospective/prospective). These are laboratory efficacy and safety tests, not clinical studies with human subjects or large datasets typical for AI model validation. The tests were performed in a controlled laboratory environment.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. "Ground truth" in the context of this device's testing refers to measured physical/biological parameters (e.g., log reduction of bacteria/viruses, ozone concentration), not expert interpretation of medical images or other data typically associated with AI/diagnostic devices. The results are based on quantitative laboratory measurements.
4. Adjudication method for the test set
Not applicable. There was no need for adjudication as the "ground truth" was established through direct laboratory measurements, not subjective evaluations or consensus processes.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is an air cleaner, not an AI-assisted diagnostic or decision support tool for human readers. No MRMC study was performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable beyond the device's inherent standalone function as an air cleaner. There is no AI algorithm in the traditional sense that performs a diagnostic or interpretive function. The device's performance is its inherent physical and chemical action on airborne particles.
7. The type of ground truth used
The ground truth for the biological agent reduction tests was established through direct quantitative laboratory measurements of the reduction in viable MS2 bacteriophage and Staphylococcus epidermidis concentrations. For ozone emissions, the ground truth was direct measurement of ozone concentration in ppm. For electrical compatibility and safety, the ground truth was compliance with specified international and national standards (EN/IEC 60601-1-2, UL 867, CSA C22.2 No.187), verified through standard engineering tests.
8. The sample size for the training set
Not applicable. This device is an air cleaner with a physical mechanism of action, not a machine learning model that requires a "training set."
9. How the ground truth for the training set was established
Not applicable as there is no training set for this device.
Ask a specific question about this device
(92 days)
FRF
The Airdog X8 Air Purifier is a mobile air cleaner intended to remove particles from the air for medical purposes. The device is intended for indoor use only. The Airdog X8 has been demonstrated to effectively inactivate H3N2 and reduce Staphylococcus albus by 4 log with L5 speed.
The Airdog X8 Air Purifier (X8) is a mobile air cleaner. The air from the room enters the X8 and passes a pre-filter, ionization frame, ionization field, collecting plates and carbon filter that captures particular matter and biological agents from the air. The X8 contains multiple PCB modules, device status indicators, power switch and a display panel, it allows user to select between different operation parameters, including auto-mode, sleep mode and wind speed. The Ionization wireframe and the collecting plate can be cleaned routinely and are reusable. X8 is powered from an AC wall outlet. It is intended to be used indoors only. The device is not intended to be connected to other instruments or medical devices.
The provided text is a 510(k) Summary for the Airdog X8 Air Purifier. It lacks the details typically found in a study proving a device meets specific acceptance criteria, especially for an AI/ML-driven device with complex performance metrics like those involving human readers or expert consensus. This document pertains to an air purifier, and its "performance data" is focused on physical and biological filtration capabilities, not on an AI's diagnostic or assistive accuracy.
Therefore, I cannot extract information related to:
- Sample size used for the test set and data provenance (e.g., country of origin, retrospective/prospective)
- Number of experts used to establish ground truth and their qualifications
- Adjudication method for the test set
- Multi-Reader Multi-Case (MRMC) comparative effectiveness study or effect size
- Standalone (algorithm only) performance
- Type of ground truth used (expert consensus, pathology, outcomes data)
- Sample size for the training set
- How ground truth for the training set was established
However, based on the provided text, I can infer the "acceptance criteria" and "reported device performance" related to the air purifier's function.
Here's a breakdown of what can be extracted:
1. A table of acceptance criteria and the reported device performance
While not explicitly called "acceptance criteria" with pass/fail thresholds in the typical AI/ML sense, the document lists specifications and test results that serve as performance benchmarks.
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Biological Agent Reduction (H3N2 Influenza virus) | 99.99% reduction with L5 speed operation (Airdog X8) |
Biological Agent Reduction (Staphylococcus albus) | 4 log reduction with L5 speed (Airdog X8) |
Particle Filtration (PM2.5) | 4-log reduction in PM2.5 particles in 120 minutes in a 30m³ chamber (Airdog X8) |
Ozone Emission | Meets the requirements of UL867 ( |
Ask a specific question about this device
(270 days)
FRF
The MA-40 and MA-112 are intended for use as room recirculating air cleaner. The systems are used for filtering out and inactivating airborne particles from the air for medical purposes.
The MA-40 has been demonstrated to reduce the following airborne microorganisms by >99.99% in a test chamber of 83ft3 (580 ft3)
BIOAEROSOL TYPE | SPECIES | DEVICE FAN SPEED | TOTAL TRIAL TIME (MINUTES) |
---|---|---|---|
Virus | MS2 bacteriophage | Speed 3 | 30 |
Speed 2 | 45 | ||
Speed 1 | 90 | ||
Virus | Phi-X 174 bacteriophage | Speed 3 | 30 |
Bacterial | Staphylococcus Epidermidis | Speed 3 | 30 |
Bacterial | Escherichia coli | Speed 3 | 30 |
Bacterial | Bacillus Subtilis Endospore | Speed 3 | 30 |
Speed 2 | 45 | ||
Speed 1 | 120 | ||
Mold | Aspergillus Brasiliensis | Speed 3 | 30 |
The MA-112 has been demonstrated to reduce the following airborne microorganisms by >99.99% in a test chamber of 83ft3 (580 ft3)
BIOAEROSOL TYPE | SPECIES | DEVICE FAN SPEED | TOTAL TRIAL TIME (MINUTES) |
---|---|---|---|
Virus | MS2 bacteriophage | Speed 4 | 20 |
Speed 3 | |||
Speed 2 | |||
Speed 1 | |||
Virus | Phi-X 174 bacteriophage | Speed 4 | 20 |
Bacterial | Staphylococcus Epidermidis | Speed 4 | 20 |
Bacterial | Escherichia coli | Speed 4 | 20 |
Bacterial | Bacillus Subtilis Endospore | Speed 4 | 20 |
Speed 3 | |||
Speed 2 | |||
Speed 1 | |||
Mold | Aspergillus Brasiliensis | Speed 4 | 20 |
Not Found
This document is a 510(k) premarket notification from the FDA, indicating that the device being reviewed is a Medical Recirculating Air Cleaner. Therefore, the device in question is not an AI/ML powered device. The information requested in your prompt (e.g., acceptance criteria for AI algorithms, sample sizes for AI model training/testing, expert adjudication, MRMC studies) is typically associated with the evaluation of AI/ML-driven medical devices.
Since the document provided is for a physical air cleaner and not an AI/ML device, most of the requested information does not apply. However, I can extract what is provided in the document related to acceptance criteria and performance of this physical medical device.
Here's an analysis based on the provided document:
The acceptance criteria for the MA-40 and MA-112 air cleaners are based on their ability to reduce airborne microorganisms. The study proves that the devices meet these criteria by demonstrating a reduction of specific bioaerosol types by >99.99% within certain timeframes in a test chamber.
1. A table of acceptance criteria and the reported device performance
The acceptance criterion for both devices is to reduce airborne microorganisms by >99.99%. The tables provided show the time taken to achieve this reduction for different bioaerosol types and device fan speeds. The device performance (achieving >99.99% reduction) is reported for each scenario.
BIOAEROSOL TYPE | SPECIES | DEVICE FAN SPEED | TOTAL TRIAL TIME (MINUTES) to achieve >99.99% reduction (MA-40) | TOTAL TRIAL TIME (MINUTES) to achieve >99.99% reduction (MA-112) | Acceptance Criteria: Reduction by >99.99% |
---|---|---|---|---|---|
For MA-40 | |||||
Virus | MS2 bacteriophage | Speed 3 | 30 | Not applicable (MA-40 specific) | Met |
Speed 2 | 45 | Not applicable (MA-40 specific) | Met | ||
Speed 1 | 90 | Not applicable (MA-40 specific) | Met | ||
Virus | Phi-X 174 bacteriophage | Speed 3 | 30 | Not applicable (MA-40 specific) | Met |
Bacterial | Staphylococcus Epidermidis | Speed 3 | 30 | Not applicable (MA-40 specific) | Met |
Bacterial | Escherichia coli | Speed 3 | 30 | Not applicable (MA-40 specific) | Met |
Bacterial | Bacillus Subtilis Endospore | Speed 3 | 30 | Not applicable (MA-40 specific) | Met |
Speed 2 | 45 | Not applicable (MA-40 specific) | Met | ||
Speed 1 | 120 | Not applicable (MA-40 specific) | Met | ||
Mold | Aspergillus Brasiliensis | Speed 3 | 30 | Not applicable (MA-40 specific) | Met |
For MA-112 | |||||
Virus | MS2 bacteriophage | Speed 4 | Not applicable (MA-112 specific) | 20 | Met |
Speed 3 | Not applicable (MA-112 specific) | "20" (implied through position, no specific time given for lower speed without breaking the row) | Met | ||
Speed 2 | Not applicable (MA-112 specific) | "20" (implied through position) | Met | ||
Speed 1 | Not applicable (MA-112 specific) | "20" (implied through position) | Met | ||
Virus | Phi-X 174 bacteriophage | Speed 4 | Not applicable (MA-112 specific) | 20 | Met |
Bacterial | Staphylococcus Epidermidis | Speed 4 | Not applicable (MA-112 specific) | 20 | Met |
Bacterial | Escherichia coli | Speed 4 | Not applicable (MA-112 specific) | 20 | Met |
Bacterial | Bacillus Subtilis Endospore | Speed 4 | Not applicable (MA-112 specific) | 20 | Met |
Speed 3 | Not applicable (MA-112 specific) | "20" (implied through position) | Met | ||
Speed 2 | Not applicable (MA-112 specific) | "20" (implied through position) | Met | ||
Speed 1 | Not applicable (MA-112 specific) | "20" (implied through position) | Met | ||
Mold | Aspergillus Brasiliensis | Speed 4 | Not applicable (MA-112 specific) | 20 | Met |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size: The test "sample" here refers to the conditions under which the device was tested. The performance was demonstrated "in a test chamber of 83ft³ (580 ft³)". It's not a biological sample size in the sense of patient data. The document does not specify the number of individual tests or replicates run for each bioaerosol type and fan speed.
- Data Provenance: The document does not specify the country of origin of the data or whether the study was retrospective or prospective. It is a report of performance in a controlled test chamber.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This question is not applicable as this is a physical device being tested for its ability to filter and inactivate microorganisms, not an AI/ML algorithm requiring expert interpretation of medical images or data. The "ground truth" is measured by the reduction of bioaerosols in a controlled environment.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This is not an adjudication process for human interpretation or AI output. The performance is based on quantitative measurements in a laboratory setting.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI-powered device, and therefore no MRMC study involving human readers and AI assistance would be conducted.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable, as this is not an AI/ML algorithm. The "standalone" performance is the device's ability to purify air without human intervention beyond turning it on and setting the fan speed.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" is the measured reduction of specific bioaerosol concentrations within the test chamber. This is determined through laboratory methods suitable for quantifying microorganisms, which is an objective measurement rather than a subjective expert assessment.
8. The sample size for the training set
Not applicable. This is a physical device, not an AI model that requires a training set.
9. How the ground truth for the training set was established
Not applicable. There is no training set for this type of device.
Ask a specific question about this device
(264 days)
FRF
The RIA SafeGuard is intended as a room recirculating air cleaner. The system is used for filtering out airborne particles PM2.5 from the air for medical purposes, and is intended for over-the-counter use.
The RIA SafeGuard has wind speed of 1.3 m/s and airflow of 135 CMH (m3/h). Using the HEPA H13 filter, it provides a 4 log (99.99%) reduction in suspended particulate PM2.5 in 120 minutes within a 29.75 m³ room.
RIA SafeGuard (SG020MS120) is an air purifier that contains the highefficiency HEPA technology and Bi-Polar Ionization (BPI) unit to remove 99.99% of particles as small as 2.5 microns from the air stream. SafeGuard addresses suspended particles with a single wind speed of 1.3 m/s, providing a 4 log reduction in PM2.5 particulates within 120 minutes in a 29.75 m³ space. The airflow is 135 CMH (m3/h). It is intended to be placed in a room such as medical care institutions and hospitals etc. RIA SafeGuard is a product that is designed as one-touch, easy to use, and no contact with patients. RIA SafeGuard has built in a blue LED as the switch-on indicator. All the maintenance should be done by a trained technician. Changing the HEPA is the only recommended maintenance to do by a non-trained technician, recommended renewing every 12 months.
The provided text describes the mechanical and electrical performance of an air purification device, not an AI/ML-enabled medical device. Therefore, the information typically required for an AI/ML device's acceptance criteria, such as details about ground truth, expert adjudication, MRMC studies, or training set data, is not present in this document.
However, I can extract the acceptance criteria and study details relevant to this specific device based on the provided text.
Here's a breakdown of the acceptance criteria and the study that proves the device meets them, according to the document:
1. A table of acceptance criteria and the reported device performance:
Test Name | Purpose | Acceptance Criteria | Reported Performance |
---|---|---|---|
Filtration of Particles | Demonstration of Device Air Cleaner Efficacy for 4 Log Reduction of PM2.5 Air Particulates in 120 minutes in a 29.75 m³ room. | Reduction rate > 99.99% (4 log reduction) | Pass |
Ozone Emissions UL867 | Determine the ozone emissions of the device. | The maximum ozone concentration emitted is below the 0.050 ppm standard. | Pass |
Electrical Safety Testing IEC 60601-1:2005, IEC 60601-1:2005+AMD1:2012, IEC 60601-1:2005+AMD1:2012+AMD2:2020 | Determine the general requirements for basic safety and essential performance of the device. | Meet the test specification according to the standards. | Pass |
Electromagnetic Compatibility (EMC) Testing IEC 60601-1-2:2014 | Determine the electromagnetic compatibility, emissions and immunity of the device. | Meet the test specification according to the standards. | Pass |
2. Sample sized used for the test set and the data provenance:
- Sample Size: The document does not specify a "sample size" in the context of a test set for an AI/ML model. Instead, it refers to a single device (RIA SafeGuard, Model Number: SG020MS120) that underwent non-clinical testing.
- Data Provenance: The tests were conducted on the subject device, RIA SafeGuard. The exact location or origin of the testing data is not explicitly stated, but the manufacturer is RIA Tech Co., Ltd. located in Taipei City, Taiwan (R.O.C.). The tests were "non-clinical," implying they were conducted in a laboratory or controlled environment, rather than involving patient data (retrospective or prospective).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not applicable to this device's testing. "Ground truth" in the context of an AI/ML model usually refers to labeled data verified by human experts. For a physical device like an air cleaner, the "ground truth" is established through direct physical measurements against predefined engineering and safety standards (e.g., measuring particle reduction with sensors, ozone concentration with detectors). No human experts were involved in establishing the "ground truth" for these physical performance metrics.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
This information is not applicable. Adjudication methods are relevant for subjective interpretations of data, typically in AI/ML model validation where multiple experts might disagree on labels. For the physical performance tests conducted, the results are objective measurements that do not require adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This information is not applicable. An MRMC study is relevant for evaluating the impact of an AI/ML system on human performance (e.g., how radiologists improve diagnostic accuracy with AI assistance). The RIA SafeGuard is an air cleaning device, not an AI/ML diagnostic tool, and therefore, an MRMC study was not performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
This information is not applicable. This question pertains to AI/ML algorithm performance. The "Filtration of Particles" test can be considered an analogue to a standalone performance test for the device in that it measures the device's inherent ability to clean air without human interaction determining its efficacy. However, it's not an "algorithm-only" test.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
The "ground truth" for this device's performance was established through objective physical measurements against established standards and design specifications.
- For "Filtration of Particles": The ground truth was the measured reduction of PM2.5 particulates using scientific instruments.
- For "Ozone Emissions": The ground truth was the measured ozone concentration.
- For "Electrical Safety" and "Electromagnetic Compatibility": The ground truth was compliance with the specified IEC standards determined through testing.
8. The sample size for the training set:
This information is not applicable. This device is a physical product, not an AI/ML model that undergoes training with a dataset.
9. How the ground truth for the training set was established:
This information is not applicable, as there is no training set for this type of device.
Ask a specific question about this device
(557 days)
FRF
The Airdog is a mobile air cleaner intended to be used to remove particles from the air for medical purposes. The device is intended for indoor use only. The Airdog has been demonstrated to effectively inactivate H3N2 and reduce Staphylococcus albus by 4 log with L4 speed.
The Airdog X5 Recirculating Air Cleaner is a floor-standing air cleaner with an air quality sensor. The air from the room enters the X5 and passes a pre-filter, ionization frame, ionization field, collecting plates and carbon layer that captures particular matter and virus from the air. The X5 contains multiple PCB modules, air quality sensor, device status indicators, power switch and a display panel it allows user to select between different operation parameter, including auto-mode, sleep mode and wind speed. The Ionization Wireframe and the Collecting Plate can be cleaned routinely and are reusable. X5 is powered from an AC wall outlet. It is intended to be used indoors only.
This document is a 510(k) Pre-market Notification from the FDA regarding the Airdog X5 Recirculating Air Cleaner (model KJ300F-X5). It's a regulatory approval document and thus primarily focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed scientific study report with all the requested specifics.
Based on the provided text, here's the information regarding acceptance criteria and the study that proves the device meets them:
1. A table of acceptance criteria and the reported device performance
Performance Metric | Acceptance Criteria / Predicate Performance (K200321 Novaerus NV1050) | Reported Device Performance (Airdog X5) |
---|---|---|
Reduction of Biological Agents | Bacillus Globigii endospores and MS2 phage reduced by 3 log in 10 minutes and 4 log in 15 minutes. | H3N2 Influenza virus inactivating by 99.99% (4 log) with L4 speed. Staphylococcus albus reduced by 99.99% (4 log) in 1 hour. |
Filtration of Particles | 4 log reduction in 0.5 to 2.0 μm sized particles in 10 minutes in a 580ft³ (16.4m³) sealed room. | Produces a 4 log reduction in PM2.5 particles in 120 minutes in a 30 m³ chamber. |
Ozone Emitted | Below 10 ppb (Implies meeting regulatory requirements, which are typically |
Ask a specific question about this device
(163 days)
FRF
The Qorda QD is intended as a room recirculating air cleaner. The system is used for filtering out and inactivating airborne particles from the air for medical purposes.
The Qorda QD1 has been demonstrated to remove the following organisms under the following exposure conditions:
| Organism | Fan Speed | Average Maximum log
reduction | Exposure Time |
|----------------------------------------|------------|----------------------------------|---------------|
| MS2 bacteriophage | Turbo MODE | 5.6 log reduction | 20 minutes |
| MS2 bacteriophage | Sleep MODE | 4.9 log reduction | 45 minutes |
| Bacillus subtilis
endospores | Turbo MODE | 4.4 log reduction | 20 minutes |
| Bacillus subtilis
endospores | Sleep MODE | 4.3 log reduction | 60 minutes |
Oorda OD1 is a freestanding medical recirculation air cleaner that facilitates movement with wheels. The device includes two pre-filters, two deodorizing filters, two HEPA filters - 3 filters in the front panel and 3 filters in the rear panel), and a manual for users.
The device can adjust the air volume in four stages through manual operation (Level 1, Level3, Turbo) and Sleep mode. When operating in sleep mode, the equipment LED and PlasmaWave are turned off and operated with low noise.
The device contains the following additional functions.
(1)Child lock: Button lock function that prevents children or pets from changing their current motion mode even if they accidentally operate device.
(2)Filter Replacement Guide: The remaining filter life is displayed on the front display of the device and tells you when to replace the filter.
(3)Adjustment of illuminance: You can adjust the illuminance in four stages (100%, 50%, 20%, OFF).
(4)You can turn on or off the PlasmaWave.
(5)The timer allows the equipment to turn off automatically after a certain period of time.
Qorda QD1 inhales air in both directions on the front and rear of the device and purifies it through the four cleaning steps below:
[Pre-Filter]
The air is drawn in through the cover front, passes through the Pre-Filter then moves to two filters stages(CD Carbon Filter, True HEPA Filter). The Pre-Filter protects CD Carbon Filter, True HEPA Filter inside the Pre-Filter.
[CD Carbon Filter]
Reduces any ozone generated as a byproduct of the plasma field. It also prolongs the True HEPA Filter lifespan.
[True HEPA Filter]
True HEPA Filter captures 99.99%* of airborne ultrafine particles as small as 0.003 microns. *Based on independent laboratory test conducted on particles as small as 0.003 microns in size.
[PlasmaWave]
PlasmaWave Technology has been tested by an independent third-party to remove particles from the air for medical purposes.
Here's an analysis of the acceptance criteria and study information for the Qorda QD1 device, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Test Name | Acceptance Criteria | Reported Device Performance |
---|---|---|
Ozone Emissions | Ozone emitted to be |
Ask a specific question about this device
(149 days)
FRF
The CerroZone Mobile is intended as a room recirculating air cleaner. The system is used for filtering out and inactivating airborne particles (i.e., bacteria and viruses) from the air for medical purposes.
The CerroZone Mobile is a medical recirculating air cleaner that uses multiple inactivation processes consisting of reactive oxidizing species (ozone) and ultraviolet radiation to inactivate bacteria and viruses in the air. The CerroZone Mobile may be used in medical facilities. Once turned on, inlet air is passed through CerroZone Mobile's inline flow-through design which inactivates bacteria and viruses in a single pass flow-through. The device generates 220 cubic feet of air per minute (CFM), or 2.36 air changes per hour (ACH) in a standard 5,600 cu. Ft room. This process of air being inlet, purified, then outlet, lasts approximately 1.2 seconds. A 4 LOG reduction of airborne particles in a standard 579 cu. Ft room is achieved in 30 minutes or less. A 5.67 LOG, or 99.9998%, reduction is achieved in 45 minutes or less.
The CerroZone Mobile device is a medical recirculating air cleaner designed to filter out and inactivate airborne bacteria and viruses. The following information summarizes its acceptance criteria and the studies conducted.
1. Table of Acceptance Criteria and Reported Device Performance:
Test Methodology | Purpose | Acceptance Criteria | Reported Device Performance |
---|---|---|---|
MS2 bacteriophage aerosolization in a sealed environmental bioaerosol chamber | To evaluate the efficacy of the CerroZone Mobile at reducing viability of aerosolized MS2 bacteriophage by a combination of entrainment and destruction | Greater than 4 log reduction (99.99%) | Average net log reduction for MS2 bacteriophage: 5.67 ± 0.19 / 60 minutes |
Methicillin Resistant Staphylococcus epidermidis (MRSE) aerosolization in a single-pass flow-through chamber connected to the CerroZone Mobile | To evaluate the efficacy of the CerroZone Mobile at reducing viability of aerosolized bacteria by a combination of entrainment and destruction | Greater than 3 log reduction (99.9%) in single-pass testing | Demonstrated effectiveness against Staphylococcus epidermidis |
Ozone Emissions | To assess ozone emissions | Emissions within the maximum acceptable level of ozone given in 21 CFR 801.415 (0.050 ppm) | Emits between 0.000 and 0.010 ppm of Ozone with a measured absolute max of 0.036 ppm. All emissions fall within the acceptable level. |
2. Sample Size Used for the Test Set and Data Provenance:
The document does not explicitly state the sample sizes for the biological aerosolization tests (MS2 bacteriophage and MRSE). The studies appear to be laboratory-based non-clinical tests rather than human subject or field trials, so terms like "country of origin" and "retrospective/prospective" might not be strictly applicable in the same way they would be for clinical studies. The context implies these were controlled experiments conducted to assess device performance under specific conditions.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
This information is not provided in the document. The "ground truth" for these non-clinical tests would have been established by the direct measurement of reduction in viable microorganisms using standard microbiological techniques, rather than expert interpretation.
4. Adjudication Method for the Test Set:
Not applicable. The tests involved direct measurement of microbial reduction, not subjective interpretation requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
No, an MRMC comparative effectiveness study was not done. The CerroZone Mobile is a medical device for air cleaning, not an interpretative diagnostic tool, so such a study would not be relevant.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
These were standalone performance tests of the device itself (algorithm not applicable), evaluating its ability to reduce airborne pathogens directly.
7. The Type of Ground Truth Used:
The ground truth used for the efficacy studies was based on quantitative laboratory measurements of microbial viability (e.g., plaque assays for bacteriophages, colony counts for bacteria) before and after exposure to the device. For ozone emissions, the ground truth was direct quantitative measurement of ozone concentration.
8. The Sample Size for the Training Set:
Not applicable. This device is a physical air cleaner, not an AI/ML algorithm. Therefore, there is no "training set" in the context of machine learning.
9. How the Ground Truth for the Training Set Was Established:
Not applicable, as there is no training set for this type of device.
Ask a specific question about this device
(322 days)
FRF
The Novaerus NV 1050 is intended as a room recirculating air cleaner. The system is used for filtering out and inactivating airborne particles from the air for medical purposes.
The Novaerus NV1050 (NV1050) is a free-standing cabinet of a size that can be maneuvered into position by a single person. The NV1050 is powered from an AC wall outlet. The NV1050 circulates the room air through its cabinet. The air is drawn in through the front of the cabinet, passes through a pre-filter then a Dielectric Barrier Discharge Plasma generator ("plasma generator") stage, and a further two filters (HEPA, Carbon) The cleaned air is exhausted out of the top panel of the cabinet. The device has simple controls: On/Off and fan speed setting from 1 to 5. There is a button to initiate a filter blockage test to check and inform the user if the filters are getting close to requiring changing. The only routine maintenance is a calendar-based filter change schedule indicated in the User Manual; the indicator light on the front panel acts as a reminder that the filters will soon need attention. The airflow path through the NV1050 is: - A general pre- filter to remove particles from the input air flow. . - A bank of three (3) plasma generators, each Plasma Generator consists of ● two (2) plasma Coils. Micro-Organisms, including virus and bacteria, are inactivated by the plasma generator. This occurs through damage by plasma constituents (ions, electrons, reactive oxidizing species, ultraviolet radiation, high electric fields) - A HEPA (High-efficiency Particulate Air) filter to trap the resulting virus/bacteria particulates. - . An activated carbon filter to trap any ozone in the airstream before it is output to the environment.
The provided text describes the Novaerus NV1050 medical recirculating air cleaner and its substantial equivalence to a predicate device. It includes information on non-clinical testing to demonstrate performance against acceptance criteria.
Here's the breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Test name/Methodology/Standard name | Purpose | Acceptance Criteria | Result |
---|---|---|---|
IEC60601-1:2005/A1;2012 | Device electrical safety | Pass the requirements of the consensus standard | Pass |
IEC60601-1-2:2014 | Device Electromagnetic Compatibility | Pass the requirements of the consensus standard | Pass |
Ozone emissions | Confirm ozone emissions are below the maximum permitted levels |
Ask a specific question about this device
(130 days)
FRF
The Airsonett AIR-4 is a Mobile Air Cleaner intended to be used to remove particles from the air for medical purposes. The device is intended for home use only.
Airsonett is based on the Temperature controlled Laminar Airflow (TLA) technology. The air from the room enters the Airsonett and passes a filter that captures allergens and other particles. The filtered air is cooled to slightly below the ambient room temperature and is supplied with a low velocity from the air supply nozzle. Since the filtered air is slightly cooler, and therefore heavier than the surrounding air, the filtered air will descend slowly from the air supply nozzle by means of gravity in a laminar manner (non-turbulent). This descending colder air counteracts the body convection, displaces the allergen load in the breathing zone and thus dramatically reduces the level of inhalant allergens for the patient all through the night.
Air is drawn in through the air intake at the floor level and through the HEPA filter. A silent Blower (fan) brings airflow through the filter. The air is directed through the Cooler/Heater and divided into a cool respectively warm air flow. The cool air flow is directed through the Air guidance arm (neck) and out through the Airshower (Air Supply Nozzle). The Airshower can be altered in height, by adding/removing and combining the neck parts, to adapt to different types of environments. The warm air flow is directed to the Warm air outlet. On its way to the outlet the air flow passes the electronics and transports away the extra heat produced by the electronics.
Airsonett is based on a microprocessor controlled supervisory system. This system controls the device behaviour by measuring temperatures, controlling thermoelectric modules and fan unit, monitoring user interaction as well as management of internal timers to keep track of the total ontime for the system and time since last filter change. The control functions (Front Panel Board) and power distribution function (Power Electronics Board) are allocated on two separate circuit boards.
Here's a breakdown of the acceptance criteria and study information for the Airsonett device, based on the provided text:
Acceptance Criteria and Device Performance
Study Endpoint | Acceptance Criteria | Reported Device Performance |
---|---|---|
Air quality in treated air envelope (clean zone) | Filter Filtration efficiency ≥99.5% of particles with size Ø≥0.5µm | Conforms with filtration efficiency ≥99.5%. |
Air quality in treated air envelope (clean zone) | Clean zone Filtration efficiency ≥99.5% of particles with size Ø≥0.5µm, which is equivalent to -Class 1000 according to FED STD 209E and -Class 6 according to ISO 14644-1 in environments of |
Ask a specific question about this device
(491 days)
FRF
Filtering out airborne particulates from air for medical purposes. HEPAirX 100 is intended for over-the-counter use.
Not Found
This is a letter from the FDA regarding a 510(k) premarket notification for a medical device called HEPAirX 100, a Medical Recirculating Air Cleaner. The document is a clearance letter, meaning the device has been found substantially equivalent to a predicate device.
Unfortunately, this document does not contain the information requested regarding acceptance criteria, study details, sample sizes, ground truth establishment, or expert qualifications. This type of information would typically be found in the 510(k) summary or the full submission, not in the clearance letter itself.
The letter confirms the device's classification, regulation number (21 CFR 880.5045), and product code (FRF). It also includes the "Indications for Use" statement, which specifies that the HEPAirX 100 is for "Filtering out airborne particulates from air for medical purposes" and is intended for over-the-counter use.
To answer your questions, I would need access to the full 510(k) submission document, specifically the sections related to performance testing and clinical data (if any was required and submitted).
Ask a specific question about this device
Page 1 of 3