K Number
K220284
Date Cleared
2022-07-08

(157 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands or fingers to prevent contamination between patient and examiner.

Device Description

Powder Free Blue Nitrile Examination Glove meets all the current specifications listed under the ASTM Specification D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application. The principle operation and mechanism of this device is to prevent contamination between patient and examiner and this principle is achieved through testing of barrier, physical properties and other testing stated in the performance data. This device is for over-the counter single use.

AI/ML Overview

The document provided describes the acceptance criteria and performance data for a Powder Free Nitrile Examination Glove (Blue), K220284, in comparison to a predicate device. This is a medical device submission, and the studies described are non-clinical performance data to demonstrate substantial equivalence, not studies related to AI or diagnostic algorithms. Therefore, many of the requested fields related to AI/algorithm performance (e.g., MRMC studies, standalone performance, training set details) are not applicable.

Here's the information that can be extracted:

1. Table of Acceptance Criteria and Reported Device Performance

Test MethodologyPurposeAcceptance CriteriaReported Device PerformanceResult
ASTM D6319-19: DimensionsTo measure the physical length and width of the glovesLength: XS-Min 220mm, S-Min 220mm, M-Min 230mm, L-Min 230mm, XL-Min 230mmMinimal Value for Length: XS: 240mm, S: 246mm, M: 240mm, L: 250mm, XL: 251mmPASS
Width: XS-70 ± 10mm, S-80 ± 10mm, M-95 ± 10mm, L-110 ± 10mm, XL-120 ± 10mmMinimal Value for Width: XS: 73mm, S: 84mm, M: 93mm, L: 104mm, XL: 111mmPASS
ASTM D6319-19: Physical PropertiesTo measure the tensile strength and elongation before and after agingTensile Strength: Before Aging: 14.0 MPa, After Aging: 14.0 MPaTensile Strength: Before Aging: Minimal Value: 19.10 MPa, After Accelerated Aging: Minimal Value: 19.0 MPaPASS
Ultimate Elongation: Before Aging: min. 500%, After Aging: min. 400%Ultimate Elongation: Before Aging Minimal Value: 520%, After Aging Minimal Value: 480%PASS
ASTM D6319-19: ThicknessTo measure the physical palm and finger thickness of the glovesPalm-Min. 0.05mmPalm Minimal Value: XS: 0.07mm, S: 0.07mm, M: 0.07mm, L: 0.07mm, XL: 0.07mmPASS
Finger-Min. 0.05mmFinger Minimal Value: XS: 0.11mm, S: 0.11mm, M: 0.11mm, L: 0.11mm, XL: 0.12mmPASS
ASTM D6319-19: Freedom from Holes TestingTo detect the presence of holes in gloveMeet AQL 2.5 at G1Meet AQL 2.5 at G1PASS
ISO 10993-10: Primary Skin IrritationTo determine the irritation potential of glove when exposed to skin surfaceNot an irritantUnder the condition of study, not an irritantPASS
ISO 10993-10: Dermal SensitizationTo determine the sensitization potential of glove when exposed to skin surfaceNot a SensitizerUnder the condition of study, not a sensitizerPASS
ISO 10993-11: Acute Systemic ToxicityTo evaluate the adverse systemic reaction potential from gloveNot induce systemic toxicityDid not induce systemic ToxicityPASS
ASTM D6124-06: Residual Powder AnalysisTo check the amount of powder residue from glove surfaceResidual powder test (< 2 mg/glove)Average 0.10mg/glovePASS

2. Sample size used for the test set and the data provenance

The document specifies "Minimal Value" and "Average" for various measurements, indicating that multiple samples were tested, but it does not provide the exact sample size for each test (e.g., number of gloves tested for dimensions, physical properties, etc.).

Data provenance: The testing was performed by KGM Gloves Sdn. Bhd. for their own device. The location of the manufacturer is Malaysia. The testing data is non-clinical performance data, likely gathered prospectively as part of product development and quality control.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. The "ground truth" here refers to objective physical and chemical properties measured by standardized tests (e.g., ASTM and ISO standards) and laboratory procedures, not expert interpretation of medical images or other subjective assessments.

4. Adjudication method for the test set

Not applicable. The tests involve objective measurements and standardized protocols, not human adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This document is for a physical medical device (examination gloves), not an AI-powered diagnostic or assistive tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an algorithmic device.

7. The type of ground truth used

The ground truth used consists of objective measurements and laboratory test results against established international and national standards (ASTM D6319-19, ISO 10993-10, ISO 10993-11, ASTM D6124-06). For example, a glove's length is its measured length, and its tensile strength is the result obtained from the specified test method.

8. The sample size for the training set

Not applicable. There is no "training set" as this is not a machine learning/AI device.

9. How the ground truth for the training set was established

Not applicable. There is no "training set" as this is not a machine learning/AI device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 8, 2022

KGM Gloves Sdn. Bhd. Ooi Seng RA Manager Lot 18893 & 18894, Jalan Perusahaan 9, Kawasan Perusahaan Kamunting. Perak 34600 Malaysia

Re: K220284

Trade/Device Name: Powder Free Nitrile Examination Gloves (Blue) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: May 10, 2022 Received: May 16, 2022

Dear Ooi Seng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220284

Device Name Powder Free Nitrile Examination Gloves (Blue)

Indications for Use (Describe)

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands or fingers to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

__ Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

1.0 510 (K) SUMMARY 510(k) Number: K220284

2.0 Submitter:

KGM Gloves Sdn. Bhd. Lot 18893 & 18894, Jalan Perusahaan 9, Kawasan Perusahaan Kamunting, 34600 Kamunting, Perak, Malaysia. Tel: +605-891 3612 Fax: +605-891 6612

Establishment Registration Number: 3020763216

Contact Person:

Ooi Loon Seng (Mdm) i) (ii) Contact No: +6012-4612918 Email: loonsenq54@gmail.com

Melissa Oon Contact No: +6012-4730656 Email: melissaoon@gmail.com

Date of Preparation : January 15, 2022

3.0 Name of the Device

Trade Name: Powder Free Nitrile Examination Glove (Blue) Classification Name: Patient Examination Glove Device Classification: Class I Regulation Number: 21 CFR 880.6250 Product Code: LZA

4.0 Identification of The Legally Marketed Predicate Device

Predicate Device Name: Powder Free Blue Nitrile Examination Glove Predicate 510(K) Number: K210375 Manufacturer's Name: Duramitt Sdn. Bhd. Device Classification: Class I Regulation Number: 21 CFR 880.6250 Product Code: LZA

5.0 Description of Device

Powder Free Blue Nitrile Examination Glove meets all the current specifications listed under the ASTM Specification D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application. The principle operation and mechanism of this device is to prevent contamination between patient and examiner and this principle is achieved through testing of barrier, physical properties and other testing stated in the performance data. This device is for over-the counter single use.

6.0 The Intended Use of Device

The Powder Free Nitrile Examination Glove (Blue) is a disposable device intended for medical purposes that is worn on the examiner's hands or fingers to prevent contamination between patient and examiner.

{4}------------------------------------------------

7.0 Summary of the Technological Characteristic Comparison of the Device

Provided below is a comparison of the subject device and the predicate device

Characteristics andParametersProposed Device -Powder Free NitrileExamination GlovesPredicate Device -Powder Free BlueNitrile ExaminationGloves (K210375)Comparison
Product CodeLZALZAIdentical
Intended UseA patient examinationglove is a disposabledevice intended formedical purposes that isworn on the examiner'shands or fingers toprevent contaminationbetween patient andexaminer.A patient examinationglove is a disposabledevice intended formedical purposes that isworn on the hand orfinger to preventcontamination betweenpatient and examiner.Identical
Device ClassificationClass IClass IIdentical
Device materialNitrile(Acrylonitrile butadiene)Nitrile(Acrylonitrile butadiene)Similar
ColourBlueBlueSimilar
Instruction forUse on LabelingSingle Use only.Single Use only.Identical
ConstructionAmbidextrousAmbidextrousIdentical
Freedom from holesMeet AQL 2.5 at G1Meet AQL 1.5 at G1Same
Length:XS-Minimum 220mmS-Minimum 220mmM-Minimum 230mmL-Minimum 230mmXL-Minimum 230mmMinimal Value for Length:XS: 240mmS: 246mmM: 240mmL : 250mmXL: 251mmAverageXS:-S: 242mmM: 242mmL : 243mmXL:-Similar
Width:-XS-70 ± 10mmS-80 ± 10mmM-95 ± 10mmL-110 ± 10mmXL-120 ± 10mmMinimal Value for Width:XS: 73mmS: 84 mmM: 93 mmL: 104 mmXL: 111mmAverageXS:-S: 84mmM: 96mmL : 107mmXL:-Similar
Palm ThicknessMinimum 0.05mmPalm Minimal Value:XS: 0.07mmS: 0.07mmM: 0.07mmAverageXS:-S: 0.06mmM: 0.06mmSimilar
L : 0.07mmL : 0.06mm
XL: 0.07mmXL:-
Finger ThicknessMinimum 0.05mmFinger Minimal Value:XS: 0.11mmS: 0.11mmM: 0.11mmL : 0.11mmXL: 0.12mmAverageXS:-S: 0.08mmM: 0.08mmL : 0.08mmXL:-Similar
Tensile Strength(Before age)Minimum 14.0MPaMinimum 19.1MPaAverage 19.04MPaSimilar
Tensile Strength(After Age)Minimum 14.0MPaMinimum 19.0MPaMinimum 14.0MPaDifferent
Ultimate Elongation(Before age)Minimum 500%Minimum 520%Average 560%Similar
Ultimate Elongation(After age)Minimum 400%Minimum 480%Average 510%Similar
Residual powder test(<2mg/glove)Average 0.10mg/gloveAverage 0.59mg/gloveSimilar
Primary Skin IrritationUnder the condition ofstudy, not an irritantUnder the condition ofstudy, not an irritantSame
Dermal SensitizationUnder the condition ofstudy, not a sensitizerUnder the condition ofstudy, not a sensitizerSame
Acute SystemicToxicityNot induce systemictoxicityNot induce systemictoxicitySame

{5}------------------------------------------------

Summary of Non-Clinical Performance Data 8.0

Provided in the table below is the summary of the non-clinical testing performed with the subject device. The results demonstrate that the subject device met the acceptance criteria.

Test MethodologyPurposeAcceptance CriteriaResultsWidth:XS-70 ± 10mmS-80 ± 10mmM-95 ± 10mmL-110 ± 10mmXL-120 ± 10mmMinimal Value for Width:XS: 73mmS: 84 mmM: 93 mmL: 104 mmXL: 111mmPASS
ASTM D6319-19:DimensionsTo measure thephysical length andwidth of the glovesLength:XS-Minimum 220mmS-Minimum 220mmM-Minimum 230mmL-Minimum 230mmXL-Minimum 230mmMinimal Value forLength:XS: 240mmS: 246mmM: 240mmL : 250mmXL: 251mmASTM D6319-19:PhysicalPropertiesTo measure the tensile strength and elongation before and after agingTensile Strength: Before Aging: 14.0 MPa After Aging: 14.0 MPaTensile Strength: Before Aging: Minimal Value: 19.10 MPa After Accelerated Aging: Minimal Value: 19.0 MPa
Ultimate Elongation:Before Aging: min. 500%After Aging: min. 400%Ultimate Elongation: Before Aging Minimal Value: 520% After Aging Minimal Value: 480%PASS
ASTM D6319-19:ThicknessTo measure the physical palm and finger thickness of the glovesPalm-Min. 0.05mmPalm Minimal Value:XS: 0.07mmS: 0.07mmM: 0.07mmL : 0.07mmXL: 0.07mm
Finger-Min. 0.05mmFinger Minimal Value:XS: 0.11mmS: 0.11mmM: 0.11mmL : 0.11mmXL: 0.12mmPASS

{6}------------------------------------------------

{7}------------------------------------------------

ASTM D6319-19:Freedom fromHolesTestingTo detect thepresence of holein gloveMeet AQL 2.5 at G1Meet AQL 2.5 at G1PASS
ISO 10993-10:Primary SkinIrritationTodeterminetheirritation potential ofglove when expose toskin surface.Not an irritantUnder the condition ofstudy, not an irritantPASS
ISO 10993-10:DermalSensitizationTodeterminethesensitizationpotential of glovewhen expose to skinsurfaceNot a SensitizerUnder the condition ofstudy, not a sensitizerPASS
ISO 10993-11:Acute SystemicToxicityToevaluatetheadversesystemicreactionpotentialfrom gloveNot induce systemic toxicityDid not induce systemicToxicityPASS
ASTM D6124-06:Residual PowderAnalysisTo check the amountof powder residuefrom glove surfaceResidual powder test( $<2$ mg/glove)Average 0.10mg/glovePASS

Summary of Clinical Performance Data 9.0

Not applicable - Clinical data is not needed for the subject device, gloves.

10.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K210375.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.