K Number
K220265

Validate with FDA (Live)

Date Cleared
2023-09-24

(601 days)

Product Code
Regulation Number
862.1117
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ADVIA Centaur® NT-proBNPII (PBNPII) assay is for in vitro diagnostic use in the quantitative determination of N-terminal pro-brain natriuretic peptide (NT-proBNP) in human serum and plasma (EDTA and lithium heparin) using the ADVIA Centaur® XP system.

In the Emergency Department (ED) and Outpatient (OP) populations, measurements of NT-proBNP are used as an aid in the diagnosis of heart failure (HF) in patients with clinical suspicion of new onset or worsening HF.

Device Description

The ADVIA Centaur® NT-proBNPII (PBNPII) assay kit includes the Primary Reagent ReadyPack and the Calibrator. The Primary Reagent ReadyPack contains Lite Reagent, Solid Phase Reagent, and Ancillary Well Reagent. The Calibrator includes Low and High Calibrators which are lyophilized.

AI/ML Overview

The provided document discusses the ADVIA Centaur® NT-proBNPII (PBNPII) assay, an in vitro diagnostic device for measuring N-terminal pro-brain natriuretic peptide (NT-proBNP) to aid in the diagnosis of heart failure in Emergency Department (ED) and Outpatient (OP) populations. The submission aims to demonstrate substantial equivalence to the predicate device, the Roche Elecsys proBNP II assay (K072437).

Here's an analysis of the acceptance criteria and the studies that support them:


1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state "acceptance criteria" as a single, consolidated table with pass/fail values for all performance characteristics. Instead, it presents various performance studies with their results. Based on the "Comparison of Technological Characteristics with the Predicate Device" and the "Performance Characteristics" sections, we can infer some criteria and compare the device's performance.

Performance CharacteristicAcceptance Criteria (Implied/Predicate)Reported Device Performance (ADVIA Centaur PBNPII)
Intended UseAid in diagnosis of suspected congestive heart failure.Aid in diagnosis of HF in ED and OP populations with clinical suspicion of new onset or worsening HF.
MeasurementQuantitativeQuantitative
TechnologyChemiluminescence immunoassayChemiluminescence immunoassay (1-step sandwich)
Sample TypePlasma and SerumHuman serum, plasma (EDTA and lithium heparin)
Assay Range5-35,000 pg/mL (Predicate)35-35,000 pg/mL
Hook EffectNo hook effect up to 300,000 pg/mL (Predicate)No hook effect up to 300,000 pg/mL (will report >35,000 pg/mL)
Precision (Total CV)For Serum: e.g., <4.6% @ 44 pg/mL, <1.8% @ 2,410 pg/mL (Predicate)For Serum: 6.7% @ 116 pg/mL, 4.0% @ 271 pg/mL, 2.3% @ 380 pg/mL, 2.1% @ 806 pg/mL, 2.0% @ 1,597 pg/mL, 2.8% @ 25,073 pg/mL. QCs also within acceptable limits (presumably).
ReproducibilityNot explicitly stated, typically within acceptable CV limits across sites.Total CVs ranging from 3.0% to 5.0% for serum samples (141-10428 pg/mL) and 3.3% to 4.5% for controls.
LinearityN/A (implied to be demonstrated across assay range)Linear for 35-35,000 pg/mL
Limit of Blank (LoB)LoB < LoD (Predicate)13 pg/mL
Limit of Detection (LoD)5.00 pg/mL (Predicate)20 pg/mL
Limit of Quantitation (LoQ)50.0 pg/mL (Predicate)35 pg/mL
Interference (HIL & Other)Bias not to exceed 10% (Implied)Bias due to hemoglobin, bilirubin, lipemia <10%; various other substances also do not interfere (bias <10%).
Cross-ReactivityNot Detectable / <1.0% for most (Implied, similar to predicate)Most cross-reactants <1.0% or not detectable. ProBNP (glycosylated) 1.0%-19.0%, ProBNP (non-glycosylated) 13.0%-30.0%.
Specimen EquivalencyRegression equation close to y=x+0, r close to 1 (Implied)Plasma (EDTA, Lithium heparin), SST, RST vs. Serum: Regression equations y=1.00x +/- small constant, r ~0.999-1.000.
Clinical Performance (ED Pop.)Likelihood Ratios (LR+, LR-) and Post-test Risk (Implied to be clinically useful for diagnosis)Age-dependent rule-in/out cutoffs provided with LR+ and LR- (0.06-0.09 for Negative LR-, suggesting good rule-out).
Clinical Performance (OP Pop.)Sensitivity, Specificity, PPV, NPV (Implied to be clinically useful for diagnosis)Overall: Sensitivity 86.0%, Specificity 64.3%, PPV 34.4%, NPV 95.5% (at 125 pg/mL cutoff). Detailed breakdown by age/sex also provided.

Note: The "Acceptable Criteria" values for precision were primarily taken directly from comments in the tables (indicated by "<" followed by a number), which appear to be the applicant's internal criteria rather than directly from the predicate. The predicate's precision data is shown for comparison purposes in the "Comparison of Technological Characteristics."

2. Sample Size Used for the Test Set and Data Provenance

  • Precision and Reproducibility:
    • Precision: Assayed in duplicate in 2 runs per day for 20 days. Number of samples not explicitly stated but implied to be multiple levels of serum and QCs (6 serum, 3 QC samples listed).
    • Reproducibility: N=90 samples assayed in duplicate in 2 runs per day for 5 days at 3 sites.
  • Linearity: Not specified.
  • Detection Limit (LoB, LoD, LoQ): Not specified.
  • Interference (HIL & Other Substances): Not specified.
  • Cross-Reactivity: Not specified.
  • Specimen Equivalency: N=50 samples for each tube type comparison to serum.
  • Expected Values (Reference Study Group): 723 apparently healthy subjects (362 females, 361 males).
  • Clinical Performance (ED Population): 3128 subjects with signs and symptoms of acute HF who presented to the ED. 1148 adjudicated as acute HF, 1980 as without HF.
  • Clinical Performance (OP Population): 1033 OP subjects with signs and symptoms of new onset HF. 185 adjudicated as new onset HF, 848 as without HF.

Data Provenance: The document does not explicitly state the country of origin for the clinical study data or if it was retrospective or prospective. However, the term "prospectively enrolled" is used for both the ED and OP clinical performance studies, indicating these were prospective studies.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • Clinical Performance (ED Population): An "independent central adjudication panel of expert clinicians (Cardiologists)" determined the diagnosis and severity of HF. The exact number of experts is not stated, nor are their specific qualifications (e.g., number of years of experience).
  • Clinical Performance (OP Population): An "independent central adjudication panel of expert clinicians (Cardiologists)" determined the diagnosis of new onset HF. Similar to the ED population, the exact number and specific qualifications are not detailed.

4. Adjudication Method (for the test set)

The adjudication method used by the "independent central adjudication panel of expert clinicians (Cardiologists)" is not explicitly described (e.g., 2+1, 3+1, none). It only states they "determined" the diagnosis.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was done. This device is an in vitro diagnostic assay (laboratory test), not an imaging or interpretive device that typically involves human readers in the same way an MRMC study would apply. Therefore, there is no effect size reported for human readers improving with or without AI assistance.

6. Standalone Performance Study

Yes, a standalone (algorithm only) performance study was performed for the assay. The entire "Performance Characteristics" section (9.1 through 9.12) describes the analytical and clinical performance of the ADVIA Centaur PBNPII assay in isolation, without human interpretation of the assay results in the context of the study design. The clinical utility is assessed by how accurately the assay's quantitative results correlate with the adjudicated clinical diagnosis.

7. Type of Ground Truth Used

  • Clinical Performance (ED and OP Populations): Clinical diagnosis adjudicated by an "independent central adjudication panel of expert clinicians (Cardiologists)." This falls under expert consensus clinical diagnosis.
  • Analytical Performance (Precision, Linearity, Detection Limits, Interference, Cross-Reactivity, Specimen Equivalency): These are quantitative measurements against established reference methods, standards, or known concentrations, serving as the ground truth for analytical validity.

8. Sample Size for the Training Set

The document describes an in vitro diagnostic assay for measuring biomarkers. These types of assays typically do not have a "training set" in the same sense as machine learning algorithms, which require labeled data for model development. Instead, they involve method development and validation experiments. The document describes studies for:

  • Establishing expected values (reference study group, N=723).
  • Establishing disease study groups for statistical analysis of correlation (ED population N=1148, OP population N=185).
  • Various analytical validation studies such as precision, linearity, and detection limits.

Therefore, the concept of a "training set" for the assay, in the context of algorithm development, is not directly applicable or discussed. The studies mentioned above contribute to the overall validation and characterization of the assay's performance.

9. How the Ground Truth for the Training Set Was Established

As mentioned above, the concept of a "training set" for an assay like this is not directly applicable. If we consider the data used to characterize the assay's performance as analogous to "training data" (i.e., the data used to define the assay's characteristics and clinical cut-offs), then:

  • Clinical Cut-offs: The clinical cut-offs for the ED and OP populations were established based on the measured NT-proBNP values in patient populations (ED and OP disease study groups) where the diagnosis of heart failure was determined by an "independent central adjudication panel of expert clinicians (Cardiologists)." This expert consensus clinical diagnosis serves as the ground truth for establishing the clinical utility of the various cut-off values.
  • Reference Intervals: The reference interval was established using 723 apparently healthy subjects. The "healthy" status would have been determined through standard medical screening and criteria, serving as the ground truth for defining the normal range.
  • Analytical Studies: Ground truth for analytical studies (e.g., linearity, detection limit) is established through reference materials, spiked samples with known concentrations, and comparison to established methods or gold standard instrumentation.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

September 24, 2023

Siemens Healthcare Diagnostics Inc. Asha Gartland Regulatory Affairs Specialist 511 Benedict Ave Tarrytown, NY 10591

Re: K220265

Trade/Device Name: ADVIA Centaur® NT-proBNPII (PBNPII) Regulation Number: 21 CFR 862.1117 Regulation Name: B-Type Natriuretic Peptide Test System Regulatory Class: Class II Product Code: NBC Dated: November 18, 2022 Received: November 18, 2022

Dear Asha Gartland:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Marianela Perez-torres -S

Marianela Perez-Torres, Ph.D. Acting Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220265

Device Name ADVIA Centaur® NT-proBNPII (PBNPII)

Indications for Use (Describe)

The ADVIA Centaur® NT-proBNPII (PBNPII) assay is for in vitro diagnostic use in the quantitative determination of N-terminal pro-brain natriuretic peptide (NT-proBNP) in human serum and plasma (EDTA and lithium heparin) using the ADVIA Centaur® XP system.

In the Emergency Department (ED) and Outpatient (OP) populations, measurements of NT-proBNP are used as an aid in the diagnosis of heart failure (HF) in patients with clinical suspicion of new onset or worsening HF.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

This 510(k) Summary is being submitted in accordance with the requirements of 21 CFR 807.92 and the Safe Medical Device Act of 1990.

The assigned 510(k) Number is: K220265 1. Date Prepared September 22nd , 2023 Applicant Information 2. Asha Gartland Contact: Regulatory Affairs Specialist Address: Siemens Healthcare Diagnostics Inc. 511 Benedict Avenue Tarrytown, NY 10591-5097 845-327-8296 Phone: Email asha.gartland@siemens-healthineers.com

Regulatory Information 3.

Table 1. Regulatory Information for ADVIA Centaur® NT-proBNPII (PBNPII) Assay

Trade NameADVIA Centaur® NT-proBNPII (PBNPII)
Common NameTest, Natriuretic Peptide
Classification NameB-type natriuretic peptide test system
FDA ClassificationClass II
Review PanelClinical Chemistry (75)
Product CodeNBC
Regulation Number21 CFR 862.1117

Predicate Device Information 4.

Predicate Device Name: Roche Elecsys proBNP II assay

510(k) Number: K072437

{4}------------------------------------------------

Intended Use / Indications for Use 5.

The ADVIA Centaur® NT-proBNPII (PBNPII) assay is for in vitro diagnostic use in the quantitative determination of N-terminal pro-brain natriuretic peptide (NT-proBNP) in human serum and plasma (EDTA and lithium heparin) using the ADVIA Centaur® XP system.

In the Emergency Department (ED) and Outpatient (OP) populations, measurements of NT-proBNP are used as an aid in the diagnosis of heart failure (HF) in patients with clinical suspicion of new onset or worsening HF.

Special Conditions for Use Statement 6.

For Prescription Use

7. Device Description

ComponentVolumeIngredients
ADVIA Centaur PBNPII Primary Reagent ReadyPack (included in assay kit)
Lite Reagent7.5 mL/packMonoclonal sheep anti-human NT-proBNP F(ab')2fragment antibody (~0.36 µg/mL) labeled with acridiniumester in buffer; bovine serum albumin (BSA); bovinegamma globulin; preservatives
Solid Phase Reagent20.0 mL/packMonoclonal sheep anti-human NT-proBNP antibody (~2µg/mL) labeled with biotin bound to streptavidin magneticparticles (~220 mg/L) in buffer; BSA; bovine gammaglobulin; sheep gamma globulin; preservatives
Ancillary Well Reagent7.5 mL/packBuffer; BSA; bovine gamma globulin; sheep gammaglobulin; preservatives
ADVIA Centaur PBNPII Calibrator (included in assay kit)
ADVIA Centaur PBNPII Lowand High Calibrators2.0 mL/vialLyophilizedAfter reconstitution, low or high levels of NT-proBNPantigen; buffer; BSA; preservatives

{5}------------------------------------------------

Candidate DevicePredicate
ItemADVIA CentaurNT-proBNP (PBNPII) assayRoche Elecsys proBNP II assay(K072437)
Intended UseThe ADVIA Centaur® NT-proBNPII (PBNPII)assay is for in vitro diagnostic use in thequantitative determination of N-terminalpro-brain natriuretic peptide (NT-proBNP) inhuman serum and plasma (EDTA and lithiumheparin) using the ADVIA Centaur® XPsystem.In the Emergency Department (ED) andOutpatient (OP) populations, measurementsof NT-proBNP are used as an aid in thediagnosis of heart failure (HF) in patientswith clinical suspicion of new onset orworsening HF.Immunoassay for the in vitroquantitative determination ofN-terminal pro-Brain natriureticpeptide in human serum and plasma.This assay is used as an aid in thediagnosis of individuals suspected ofhaving congestive heart failure.The test is further indicated for therisk stratification of patients withacute coronary syndrome andcongestive heart failure. The test mayalso serve as an aid in the assessmentof increased risk of cardiovascularevents and mortality in patients at riskfor heart failure who have stablecoronary artery disease.
Indications for UseSame as Intended Use (Candidate)Same (for Predicate)
Similarities
MeasurementQuantitativeSame
TechnologyChemiluminescenceSame
Sample typePlasma and SerumSame
StandardizationTraceable to synthetic human NT-proBNP(1-76) through internal standardSame
Calibration2 levelsSame
CalibratorsLyophilizedSame
Hook EffectNo hook effect up to 300,000 pg/mLSame
Differences
Operating Principle1-Step Sandwich immunoassay2-Step Sandwich immunoassay
Assay Range35-35,000 pg/mL5-35,000 pg/mL
Candidate DevicePredicate
ItemADVIA CentaurNT-proBNP (PBNPII) assayRoche Elecsys proBNP II assay(K072437)
Sample Volume20 μL15 μL
LoB13 pg/mLLoB<LoD
LoD20 pg/mL5.00 pg/mL
LoQ35 pg/mL50.0 pg/mL
Detection AntibodyMonoclonal sheep anti-human NT-proBNPF(ab')2 fragmentantibody (~0.36 µg/mL) labeled withacridinium ester in bufferMonoclonal anti-NT-proBNP antibody(sheep) labeled with rutheniumcomplex 1.1 µg/mL in buffer
Capture AntibodyMonoclonal sheep anti-human NT-proBNPantibody (~2 µg/mL)labeled with biotin bound to streptavidinmagnetic particles(~220 mg/L) in bufferBiotinylated monoclonalanti-NT-proBNP antibody (mouse) 1.1µg/mL in buffer
Precision (Serum)(Total CV)6.7% @ 116 pg/mL4.0% @ 271 pg/mL2.3% @ 380 pg/mL2.1% @ 806 pg/mL2.0% @ 1,597 pg/mL2.8% @ 25,073 pg/mL4.2% @ 144 pg/mL2.9% @ 418 pg/mL2.8% @ 4,778 pg/mL4.6% @ 44.0 pg/mL2.6% @ 126 pg/mL1.8% @ 2,410 pg/mL3.8% @ 33,606 pg/mL2.8% @ 82.0 pg/mL1.6% @ 2,318 pg/mL
Intended UsePopulation(s)Emergency Department (ED) andOutpatient (OP)Single Population
Clinical Cut-OffED PopulationRule out HF:All ages <300 pg/mLIndeterminate<50 years ≥300 to ≤ 450 pg/mL50-75 years ≥ 300 to ≤ 900 pg/mL>75 years ≥ 300 to ≤ 1800 pg/mLRule-in HF< 50 years > 450 pg/mL50–75 years > 900 pg/mL>75 years > 1800 pg/mLOP PopulationRule out HFAll ages <125 pg/mL125 pg/mL for <75 years450 pg/mL for ≥ 75 years
Candidate DevicePredicate
ItemADVIA CentaurNT-proBNP (PBNPII) assayRoche Elecsys proBNP II assay(K072437)
Clinical DataPresentationED PopulationLikelihood Ratio – age dependent rule-out(negative), indeterminate and rule in(positive) cut offsOP PopulationSensitivity and Specificity and PositivePredictive Value (PPV) and NegativePredictive Value (NPV) for ages ≤75 and >75years and sexSensitivity and Specificity vs age (<45,45-54, 55-64, 65-74, ≥ 75 and <75years) and sex

Comparison of Technological Characteristics with the Predicate 8. Device

{6}------------------------------------------------

510(k) Summary

{7}------------------------------------------------

510(k) Summary

Performance Characteristics: ADVIA Centaur PBNPII assay 9.

Precision 9.1

Precision was determined in accordance with CLSI Document EP05-A3. Samples were assayed in duplicate in 2 runs per day for 20 days. The following results were obtained:

RepeatabilityWithin-Lab
SpecimenTypeMean of 80repspg/mL(pmol/L)SDpg/mL(pmol/L)% CVAcceptableCriteria(%CV)SDpg/mL(pmol/L)% CVAcceptableCriteria(%CV)
Serum 1116 (13.7)6.1 (0.720)5.3<2.57.8 (0.920)6.7ಸಿ ಸರ್ಕಾರಿ ಸಾಧಿಸಿದ ಸಾಧಿಸಿದ ಮಾರ್ಗ್ ಸಾಧಿಸಿದ ಸಂಸ್ಕೃತಿಗಳು ಮಾರ್ಗ್ ಸಾಧಿಸಿದ ಸಂಸ್ಕೃತಿಗಳು ಮಾರ್ಗ್ ಸಾಮಾನ ಮಾರ್ಗ್ ಸಾಮಾನ್ಯ ಸಂಸ್ಕೃತಿಗಳಿಗೆ ಮಾರ್ಗ್ ಸಾಮಾನ ಮಾರ್ಗ್ ಸಾಮಾನ್ಯ ಸಂಸ್ಕೃತಿಗಳಿಗೆ ಮಾರ್ಗ್ ಮ
Serum 2271 (32.0)9.6 (1.13)3.5ಸ್ಸ್ ನ10.8 (1.27)4.0ಸಿಕೊಡುವುದು ಸಹಾಯಿತು. ಸಾಮಾನ್ಯ ಸಂಪರ್ಕಗಳು ಸಾಮಾನ ಮಾರ್ಥಿಸಿದ ಮಾರ್ಗ್ ಮಾರ್ಗ್ ಸಾಮಾನ್ಯ ಸಂಸ್ಕೃತಿಗಳು ಮಾರ್ಗ್ ಮಾರ್ಗ್ ಮಾರ್ಗ್ ಮಾರ್ಗ್ ಸಾಮಾನ್ಯ ಸಾಮಾನ್ಯ ಸಂಸ್ಕೃತಿಗಳು ಮಾರ್ಗ್ ಮಾರ್ಗ್ ಮಾರ್ಗ್ ಮಾರ್ಗ್ ಸ
Serum 3380 (44.8)5.4 (0.637)1.4ಸ್ಸ್ ನ8.9 (1.05)2.3ಸಿಕೊಡುವುದು ಸಹಾಯಿತು. ಸಾಮಾನ್ಯ ಸಂಪರ್ಕಗಳು ಸಾಮಾನ ಮಾರ್ಥಿಸಿದ ಮಾರ್ಗ್ ಮಾರ್ಗ್ ಸಾಮಾನ್ಯ ಸಂಸ್ಕೃತಿಗಳು ಮಾರ್ಗ್ ಮಾರ್ಗ್ ಮಾರ್ಗ್ ಮಾರ್ಗ್ ಸಾಮಾನ್ಯ ಸಾಮಾನ್ಯ ಸಂಸ್ಕೃತಿಗಳು ಮಾರ್ಗ್ ಮಾರ್ಗ್ ಮಾರ್ಗ್ ಮಾರ್ಗ್ ಸ
Serum 4806 (95.1)15.3 (1.81)1.9<716.7 (1.97)2.1<10
Serum 51597 (188)25.7 (3.03)1.6<731.6 (3.73)2.0<10
Serum 625,073(2959)416.4 (49.1)1.7<7690.1(81.4)2.8<10
QC 1144 (17.0)5.6 (0.661)3.9ಸ್ಸ್ ನ6.1 (0.720)4.2ಸಿಕೊಡುವುದು ಸಾಧಿಸಿದ ಸಮಾರಿಕೆಯಿಂದ ಸಿರ್ವಹಿಸಿದ ಸಂಸ್ಕೃತಿಗಳು ಸಾಮಾನ ಮಾರ್ಥಿಕ ಸಾಮಾನ ಮಾರ್ಥಿಕ ಸಾಮಾನ್ಯ ಸಂಕರ್ಥಿಸಿದ ಮಾರ್ಥಿಕ ಸಾಮಾನ ಮಾರ್ಥಿಕ ಸಾಮಾನ ಮತ್ತು ಮಾರ್ಗ್ ಮತ್ತು ಮಾರ್ಗ್ ಮಾರ್ಗ್ ಮಾರ್ಗ್ ಮಾರ್
QC 2418 (49.3)8.7 (1.03)2.1ಸ್ಸ್ ನ12.1 (1.43)2.9ಸಿಕೊಡುವುದು ಸಹಾಯಿತು. ಸಾಮಾನ್ಯ ಸಂಪರ್ಕಗಳು ಸಾಮಾನ ಮಾರ್ಥಿಸಿದ ಮಾರ್ಗ್ ಮಾರ್ಗ್ ಸಾಮಾನ್ಯ ಸಂಸ್ಕೃತಿಗಳು ಮಾರ್ಗ್ ಮಾರ್ಗ್ ಮಾರ್ಗ್ ಮಾರ್ಗ್ ಸಾಮಾನ್ಯ ಸಾಮಾನ್ಯ ಸಂಸ್ಕೃತಿಗಳು ಮಾರ್ಗ್ ಮಾರ್ಗ್ ಮಾರ್ಗ್ ಮಾರ್ಗ್ ಸ
QC 34778 (564)93.0 (11.0)1.9<7134.7(15.9)2.8<10

{8}------------------------------------------------

9.2 Reproducibility

Reproducibility was determined in accordance with CLSI Document EP05-A3. Samples (N=90) were assayed in duplicate in 2 runs per day for 5 days at 3 sites.

MeanRepeatabilityBetween RunBetween DayBetween SiteReproducibility
SampleIDNSD%CVSD%CVSD%CVSD%CVSD%CV
pg/mL (pmol/L)pg/mL(pmol/L)pg/mL (pmol/L)pg/mL (pmol/L)pg/mL(pmol/L)pg/mL (pmol/L)
SerumA90141 (16.6)4.1 (0.484)2.9%0 (0.000)0.0%0.8 (0.094)0.6%5.7 (0.673)4.0%7 (0.826)5.0%
SerumB90324 (38.2)7.6 (0.897)2.4%0 (0.000)0.0%2.6 (0.307)0.8%6.7 (0.791)2.1%10.5 (1.24)3.2%
SerumC90489 (57.7)8.5 (1.00)1.7%0 (0.000)0.0%4.9 (0.578)1.0%15 (1.77)3.1%18 (2.12)3.7%
SerumD90824 (97.2)14.7 (1.73)1.8%0 (0.000)0.0%9.8 (1.16)1.2%18.5 (2.18)2.2%25.5 (3.01)3.1%
SerumE901641 (194)29.4 (3.47)1.8%13.9 (1.64)0.8%9.6 (1.13)0.6%35 (4.13)2.1%48.7 (5.75)3.0%
SerumF9010428 (1231)184.6 (21.8)1.8%49.7 (5.86)0.5%95.7 (11.3)0.9%374.3 (44.2)3.6%431.1 (50.9)4.1%
SerumG9019469 (2297)349.1 (41.2)1.8%147.4 (17.4)0.8%0 (0.000)0.0%643.1 (75.9)3.3%746.5 (88.1)3.8%
SerumH9028860 (3405)696.8 (82.2)2.4%0 (0.000)0.0%127.4 (15.0)0.4%736.1 (86.9)2.6%1021.6 (121)3.5%
Control190146 (17.2)4 (0.472)2.8%1.2 (0.142)0.8%1.7 (0.201)1.2%4.8 (0.566)3.3%6.6 (0.779)4.5%
Control290430 (50.7)9.3 (1.10)2.2%3.8 (0.448)0.9%1.9 (0.224)0.4%12.9 (1.52)3.0%16.5 (1.95)3.8%
Control3904965 (586)89 (10.5)1.8%42.9 (5.06)0.9%32.4 (3.82)0.7%129 (15.2)2.6%165.7 (19.6)3.3%

The following results are representative of the performance of the assay:

9.3 Linearity

Linearity testing was performed in accordance with CLSI Document EP06-Ed2. The ADVIA Centaur PBNPII assay is linear for the measuring interval of 35-35,000 pg/mL (4.13-4130 pmol/L).

9.4 Detection Limit

The limit of blank (LoB), limit of detection (LoD), and the limit of quantitation (LoQ) were determined as described in CLSI protocol EP17-A2. The ADVIA Centaur PBNPII assay has an LoB of 13 pg/mL (1.53 pmol/L), an LoD of 20 pg/mL (2.36 (pmol/L), and an LoQ of 35 pg/mL (4.13 pmol/L).

ਰੇ' ਦ Interference

Hemolysis, Icterus, Lipemia (HIL)

Interference testing was performed in accordance with CLSI Document EP07-ed3. The following substances do not interfere with the assay when present in serum at the concentrations indicated.

{9}------------------------------------------------

Bias due to these substances does not exceed 10% at NT-proBNP concentrations of 123–137 pg/mL (14.5-16.2 pmol/L) and 1612-1813 pg/mL (190-214 pmol/L).

SubstanceSubstance Test Concentration
Hemoglobin1000 mg/dL
Bilirubin, conjugated60.0 mg/dL
Bilirubin, unconjugated60.0 mg/dL
Lipemia (Intralipid)3000 mg/dL

Other Substances

The following substances do not interfere with the assay when present in serum at the concentrations indicated. Bias due to these substances does not exceed 10% at NT-proBNP concentrations of 116–163 pg/mL (13.7–19.2 pmol/L) and 1478–2225 pg/mL (174–263 pmol/L).

{10}------------------------------------------------

Siemens Healthcare Diagnostics 510(k) Summary

SubstanceSubstance TestConcentrationSubstanceSubstance TestConcentration
Abciximab21.0 µg/mLL-dopa (Levodopa)0.750 mg/dL
Acetaminophen20.0 mg/dLLidocaine8.00 mg/dL
Acetylcysteine15.0 mg/dLLisinopryl16.0 µg/mL
Allopurinol2.50 mg/dLLovastatin16.0 µg/mL
Amiodarone20.0 42.0 µg/mLL-Thyroxine60.0 µg/dL
Amlodipine Besylate4.00 µg/mLMethyldopa2.50 mg/dL
Ampicillin5.30 7.50 mg/dLMethylprednisolone0.783 mg/dL
Ascorbic Acid5.00 mg/dLMetoprolol tartrate15.0 mg/dL
Atenolol1.00 mg/dLMetronidazole12.3 mg/dL
Atorvastatin32.0 mg/dLMilrinone lactate2.40 µg/mL
Biotin3510 ng/mLMolsidomin (Placeholder)0.018 mg/dL
Bisoprolol0.258 µg/mLNicotine0.100 mg/dL
Caffeine6.00 10.8 mg/dLNifedipine6.00 mg/dL
Calcium dobesilate6.00 mg/dLNitrofurantoin40.0 µg/mL
Captopril5.00 mg/dLNitroglycerine0.160 µg/mL
Carvedilol4.32 mg/dLOxazepam12.0 µg/mL
Cefoxitin660 mg/dLOxytetracycline100 µg/mL
Chloramphenicol7.80 mg/dLPhenobarbital69.0 mg/dL
Chlordiazepoxide1.00 mg/dLPhenprocoumon (Placeholder)1.50 mg/dL
Cholesterol500 mg/dLPhenylbutazone32.1 mg/dL
Cinnarizine3.00 mg/dLPhenytoin6.00 mg/dL
Clopidogrel bisulfate30.0 µg/mLPravastatin0.021 mg/dL
Creatinine30.0 mg/dLProbenecid200 µg/mL
Cyclosporine4000 ng/mLPropafenone HCL30.0 mg/dL
Digitoxin75.0 ng/mLPropanolol0.150 mg/dL
Digoxin390 ng/mLProtein: Albumin6.00 g/dL
Diltiazem120 µg/mLProtein: Total11.1 g/dL
Dipotassium EDTA9.00 mg/mLQuinidine20.0 µg/mL
Dipyridamole30.0 µg/mLRetavase (reteplase)3.33 mg/dL
Diclofenac60.0 µg/mLRheumatoid Factor1500 IU/mL
Disopyramide40.0 µg/mLRifampicin (Rifampin)4.80 mg/dL
Dopamine16.0 mg/dLSalicylic Acid60.0 mg/dL
Doxycycline hyclate1.80 mg/dLSimvastatin32.0 µg/mL
Enalapril maleate16.0 µg/mLSotalol hydrochloride0.510 mg/dL
Epinephrine0.050 mg/dLSpironolactone7.50 mg/dL
Erythromycin13.8 mg/dLStreptokinase150,000 U/L
Furosemide6.00 mg/dLSulfamethoxazole320 µg/mL
Gentamycin sulfate3.51 mg/dLTheophylline4.00 mg/dL
HAMA (Human Anti-Mouse Antibody)800 ug/LTolbutamide150 mg/dL
Heparin330 U/dLTorasemide1.50 mg/dL
Hydralazine20.0 µg/mLTriglycerides1000 mg/dL
Hydrochlorothiazide20.0 µg/mLTrimethoprim64.0 µg/mL
Immunoglobulin G (IgG)5.00 g/dLVerapamil96.0 µg/mL
Indomethacin16.0 µg/mLUrokinase150,000 U/L
Insulin0.160 mg/dLWarfarin8.00 mg/dL
Isosorbide dinitrate6.00 mg/dL

{11}------------------------------------------------

Siemens Healthcare Diagnostics 510(k) Summary

{12}------------------------------------------------

9.6 Cross-Reactivity

Cross-reactivity was determined in accordance with CLSI Document EP07-ed3. Cross-reactants were tested at NT-proBNP concentrations of 0 pg/mL (0 pmol/L) and 126–158 pg/mL (14.9–18.6 pmol/L).

Cross-reactantCross-reactant ConcentrationCross-reactivity (%)
Adrenomedullin1.00 ng/mLNot Detectable< 1.0%
Aldosterone0.600 ng/mLNot Detectable< 1.0%
Angiotensin I0.600 ng/mLNot Detectable< 1.0%
Angiotensin II0.600 ng/mLNot Detectable< 1.0%
Angiotensin III1.00 ng/mLNot Detectable< 1.0%
ANP283.10 µg/mLNot Detectable< 1.0%
Arg-Vasopressin1.00 ng/mLNot Detectable< 1.0%
BNP323.50 µg/mLNot Detectable< 1.0%
CNP322.20 µg/mLNot Detectable< 1.0%
DNP1.00 ng/mLNot Detectable< 1.0%
Endothelin20.0 pg/mLNot Detectable-5.0%
preproANP26-553.50 µg/mLNot Detectable< 1.0%
preproANP56-921.00 ng/mLNot Detectable< 1.0%
preproANP104-1231.00 ng/mLNot Detectable< 1.0%
proBNP (glycosylated)3000 pg/mL1.0%19.0%
proBNP (non-glycosylated)3000 pg/mL13.0%30.0%
Renin; human50.0 ng/mLNot Detectable< 1.0%
Urodilatin3.50 µg/mLNot Detectable< 1.0%
VNP1.00 ng/mLNot Detectable< 1.0%

{13}------------------------------------------------

9.7 Specimen Equivalency

Specimen equivalency was determined with the Passing-Bablok regression model in accordance with CLSI Document EP09c-ed3.

Tube (y) vs. Serum (x)Regression EquationSample Intervalpg/mL (pmol/L)rN
Plasma (Dipotassium EDTA) vsSerum (no gel barrier)y = 1.00x + 1 pg/mL(y = 1.00x + 0.118 pmol/L)50-30,134(5.90-3556)1.00050
Plasma (Lithium heparin) vs Serum (nogel barrier)y = 1.00x + 4 pg/mL(y = 1.00x + 0.472 pmol/L)57-30,390(6.73-3586)0.99950
BD Vacutainer® SST™ Tube vs Serum(no gel barrier)y = 1.00x - 2 pg/mL(y = 1.00x - 0.236 pmol/L)54-30,122(6.37-3554)1.00050
BD Vacutainer® RST™ Tube vs Serum(no gel barrier)y = 1.00x + 4 pg/mL(y = 1.00x + 0.472 pmol/L)53-30,237(6.25-3568)1.00050

9.8 Expected Values

9.8.1 Reference Study Group

Expected values were established non- parametrically on the ADVIA Centaur XP system in accordance with CLSI Document EP28-A3c on a population of 723 apparently healthy subjects (362 females and 361 males) without HF.

SexAge(years)NMean pg/mL(pmol/L)SD pg/mL(pmol/L)Median pg/mL(pmol/L)95th Percentilepg/mL (pmol/L)% < 125pg/mL% < 300pg/mL
Male<5012056 (6.61)95.0 (21.2)< 35 (4.13)124 (14.6)N/Aa97.5
50-7512178 (9.20)136.4 (16.1)< 35 (4.13)322 (38.0)N/Aa94.2
>7512056 (6.61)50.6 (5.97)< 35 (4.13)154 (18.2)N/Aa98.3
Female<5012257 (6.73)33.1 (3.91)37 (4.37)133 (15.7)N/Aa100
50-7512077 (9.09)124.0 (14.6)40 (4.72)192 (22.7)N/Aa97.5
>7512089 (10.5)216.4 (25.5)43 (5.07)178 (21.0)N/Aa97.5
OverallOverall72369 (8.14)124.9 (14.7)< 35 (4.13)163 (19.2)92.097.5

a The cut-off value is not applicable for the age group indicated.

As with all in vitro diagnostic assays, each laboratory should determine its own reference interval for the diagnostic evaluation of patient results. Consider these values as guidance only.

{14}------------------------------------------------

9.8.2 Disease Study Group

ED Population 9.8.2.1

Samples were obtained from 1148 patients (476 females and 672 males) that were adjudicated with a diagnosis of acute HF in the ED. ADVIA Centaur PBNPII values are presented for subjects with acute HF (ED Population- Disease Study Group) by subgroups based on age group and NYHA Functional Class. Each group of females and males was divided into three age groups as follows: < 50 years, 50-75 years and>75 years.

The data for the ED population with acute HF for combined male and female (All) is presented in this summary.

ED Population – All Subjects with Acute HF
Age Group< 50 years50–75 years> 75 years
Mean pg/mL (pmol/L)6004 (708)6820 (805)8591 (1014)
SD pg/mL (pmol/L)7601 (897)8182 (965)8984 (1060)
Median pg/mL (pmol/L)3552 (419)3677 (434)5295 (625)
95th Percentile pg/mL (pmol/L)25,401 (2997)26,473 (3124)34,451 (4065)
% > 300 pg/mL94.395.098.4
% > 450 pg/mL90.9N/A*N/A*
% > 900 pg/mLN/A*83.8N/A*
% > 1800 pg/mLN/A*N/A*85.8
N230481437
  • The cut-off value is not applicable for the age group indicated.

9.8.2.2 OP Population

Samples were obtained from 185 patients (102 females and 83 males) adjudicated with new onset of HF.

The data for the OP population with new onset of HF for combined male and female (All) is presented in this summary.

OP Population – All Subjects with Adjudicated New Onset HF
Age Group≤ 75 years> 75 years
Mean pg/mL (pmol/L)1313 (155)2418 (285)
SD pg/mL (pmol/L)3756 (443)5536 (653)
Median pg/mL (pmol/L)449 (53.0)661 (78.0)
95th Percentile pg/mL (pmol/L)4547 (537)13,518 (1595)
% > 125 pg/mL82.196.1
N13451

{15}------------------------------------------------

NYHA Functional ClassaNYHA INYHA IINYHA IIINYHA IV
Mean pg/mL (pmol/L)b880 (104)3340 (394)7139 (842)7466 (881)
SD pg/mL (pmol/L)1161 (137)6262 (739)8480 (1001)8277 (977)
Median pg/mL (pmol/L)387 (46)996 (118)4018 (474)4409 (520)
5th Percentile pg/mL(pmol/L)135 (15.9)73 (8.6)261 (30.8)449 (52.9)
95th Percentile pg/mL(pmol/L)2612 (308)17357 (2048)29643 (3498)30,824 (3637)
N4c265573474

The NYHA classifications for the ED and OP population with acute HF for combined male and female (All) is presented in this summary.

a Unable to determine NYHA classification for 16 subjects.

b p = <0.0001 for increasing mean NT-ProBNP concentration between NYHA Class II and NYHA Classes III and IV per Jonckheere-Terpstra. The mean NT-ProBNP concentration for NYHA Class IV are not statistically different.

ৎ One male subject, identified as an outlier with an abnormally high NT-ProBNP concentration, was excluded from the NYHA I classification summary.

OP Population – Receiver Operating Characteristic Curve

The Receiver Operating Characteristic (ROC) curve presents the clinical sensitivity and specificity for the enrolled 185 subjects diagnosed with new onset HF and 848 subjects without HF. The area under ROC curve for the ADVIA Centaur PBNPII assay is 0.839 with a 95% confidence interval of 0.804 to 0.868.

Image /page/15/Figure/8 description: This image shows a receiver operating characteristic (ROC) curve. The ROC curve plots the true positive rate (sensitivity) against the false positive rate (1-specificity). The area under the curve (AUC) is 0.839 with a 95% confidence interval of (0.804, 0.868). The ROC curve is a graphical representation of the performance of a binary classifier system as its discrimination threshold is varied.

{16}------------------------------------------------

510(k) Summary

9.9 Dilution Recovery

Samples that exceed the high end of the measuring interval may be diluted 1:5 and 1:10 with ADVIA Centaur Multi-Diluent 1 using automated dilution. The percent recovery of diluted samples up to 63,128 pg/mL (7449 pmol/L) ranged from 80.5% to 88.2% with a mean recovery of 83.9%.

9.10 High Dose Hook

High NT-proBNP concentrations can cause a paradoxical decrease in the RLUs (high-dose hook effect). In this assay, patient samples with NT-proBNP concentrations above the measuring interval and as high as 300,000 pg/mL (35,400 pmol/L) will report >35,000pg/mL (4130 pmol/L).

9.11 Stability

The ADVIA Centaur NT-proBNP (PBNPII) reagents and calibrators are stable until the date printed on the box label when stored at 2-8°C.

The onboard stability and calibration interval of the ADVIA Centaur NT-proBNP (PBNPII) reagents is 36 days. The ADVIA Centaur PBNPII Calibrators are stable for 48 hours at 2-8°C and 31 days at < -20°C after reconstitution. The Calibrators are stable onboard at Room Temperature for 6 hours.

9.12 Clinical Performance

ED Population 9.12.1

A total of 3128 subjects with signs and symptoms of acute HF who presented to the ED, were prospectively enrolled in a multi-site clinical evaluation of the ADVIA Centaur PBNPII assay. Diagnosis and severity of HF were determined by an independent central adjudication panel of expert clinicians (Cardiologists).1148 subjects were adjudicated as acute HF and 1980 subjects were adjudicated as without HF.

Clinical results were evaluated and Post-test risk (%), Positive Likelihood ratio (LR+), and respective two-sided confidence intervals (CI) were computed. Subgroup analyses were performed by using age group and sex.

Clinical Agreement between ADVIA Centaur PBNPII Results and Adjudicated Diagnosis Entire ED Population by Age Group with combined male and female (All) is presented in this summary:

Adjudicated Diagnosis
Age Group(years)NT-proBNPRule-in Cutoff(pg/ml)PositiveNegativeTotalPost-testRisk (%)95% CI Post-testRisk (%)LR+95% CI LR+*
<50Positive> 45020912333263.0% (209/332)(57.5%,68.2%)5.01(4.25, 5.91)
Indeterminate300 - ≤4508192729.6% (8/27)(13.8%,50.2%)1.24(0.55, 2.80)
Negative<300135365492.4% (13/549)(1.3%,4.0%)0.07(0.04, 0.12)
Total230678908Pre-Test Risk = 25.3% (230/908)

{17}------------------------------------------------

Adjudicated Diagnosis
Age Group(years)NT-proBNPRule-in Cutoff(pg/ml)PositiveNegativeTotalPost-testRisk (%)95% CI Post-testRisk (%)LR+95% CI LR+*
50-75Positive> 90040318558868.5% (403/588)(64.6%,72.3%)3.71(3.25, 4.24)
Indeterminate300 - ≤9005416021425.2% (54/214)(19.6%,31,6%)0.58(0.43, 0.77)
Negative<300244754994.8% (24/499)(3.1%,7.1%)0.09(0.06, 0.13)
Total4818201301Pre-Test Risk = 37.0% (481/1301)
>75Positive> 180037517454968.3% (375/549)(64.2%,72.2%)2.38(2.10, 2.69)
Indeterminate300-≤18005518624122.8% (55/241)(17.7%,27.7%)0.33(0.25, 0.43)
Negative<30071221295.4% (7/129)(2.2%,10.9%)0.06(0.03, 0.13)
Total437482919Pre-Test Risk = 47.6% (437/919)

*Two-sided 95% score intervals will be present risk and positive likelihood ratio estimates.

9.12.2 OP Population

A total of 1033 OP subjects with signs and symptoms of new onset HF were prospectively enrolled in a multi-site clinical evaluation of the ADVIA Centaur PBNPII assay. The clinical performance was assessed using a single cut-off value of 125 pg/mL.

Diagnosis and severity of HF were determined by an independent central adjudication panel of expert clinicians (Cardiologists).185 subjects were adjudicated as new onset HF and 848 subjects were adjudicated as without HF.

Diagnostic accuracy was computed as the Clinical Sensitivity, Clinical Specificity, Negative Predictive (NPV), and Positive Predictive (PPV).

The OP population clinical performance for All subjects is presented below:

Cut off ValueSensitivitySpecificityPPVNPV
125 pg/mL86.0% (159/185)64.3% (545/848)34.4% (159/462)95.5% (545/571)
(80.1%, 90.6%) *(60.9%, 67.5%) *(30.1%, 39.0%) *(93.4%, 97.0%) *

*Confidence Interval

The OP population clinical performance by sex and age group is presented below:

{18}------------------------------------------------

Siemens Healthcare Diagnostics

510(k) Summary

GroupSensitivitySpecificityPPVNPV
Males≤ 75 yearsEstimate85.5% (53/62)68.9% (179/260)39.6% (53/134)95.2% (179/188)
95% CI*74.2%, 93.1%62.8%, 74.4%31.2%, 48.4%91.1%, 97.8%
Females≤ 75 yearsEstimate79.2% (57/72)70.8% (335/473)29.2% (57/195)95.7% (335/350)
95% CI*67.5%, 87.8%67.0%, 74.9%23.0%, 36.2%93.0%, 97.6%
Males> 75 yearsEstimate90.5% (19/21)37.2% (16/43)41.3% (19/46)88.9% (16/18)
95% CI*69.6%, 98.8%22.8%, 51.7%27.0%, 56.8%65.3%, 98.6%
Females> 75 yearsEstimate100% (30/30)20.8% (15/72)34.5% (30/87)100% (15/15)
95% CI*88.4%,100%(2.2%, 32.0%24.6%, 45.4%78.2%, 100%

*CI=Confidence Interval

10. Conclusions

ADVIA Centaur NT-proBNP II (PBNPII) assay is substantially equivalent in analytical performance to the currently marketed predicate device, the Roche Elecsys proBNP II assay cleared under 510(k) Number: K072437.

§ 862.1117 B-type natriuretic peptide test system.

(a)
Identification. The B-type natriuretic peptide (BNP) test system is an in vitro diagnostic device intended to measure BNP in whole blood and plasma. Measurements of BNP are used as an aid in the diagnosis of patients with congestive heart failure.(b)
Classification. Class II (special controls). The special control is “Class II Special Control Guidance Document for B-Type Natriuretic Peptide Premarket Notifications; Final Guidance for Industry and FDA Reviewers.”