K Number
K220252

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-04-14

(73 days)

Product Code
Regulation Number
872.3630
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The High Retention Attachment System is designed to support fixed, partial or full arch restorations on endosseous dental implants in the mandible or maxilla for the purpose of restory function. It is used in fixed hybrid restorations that can be attached with a snap-in system. The High Retention Attachment System is compatible with the following implants: (list of compatible implants and diameters provided in the document).

Device Description

The High Retention Attachment System is a system that provides rigid connection of fixed, partial, and full arch restorations (fixed/detachable hybrid dentures) to endosseous dental implants. It consists of abutments, attachment housings, inserts and seating and removal tools. The abutments are provided in various cuff heights with the implant and abutment connection specific to the Zest Anchors, LLC or OEM implant. The subject device abutments are made of titanium alloy conforming to ASTM F136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401) and are TiN (titanium nitride) coated. The High Retention Attachment System is as provided with an abutment (LOCATOR Attachment) and the surgical instrumentation necessary to place the implant. The High Retention Attachment System is designed to accommodate a path of insertion diverqence of up to 20° per implant and no more than 40° of divergence between implants.

AI/ML Overview

This document describes a 510(k) premarket notification for a dental device, the "High Retention Attachment System." It does not contain information about an AI/ML-driven medical device, hence no study details are provided to address the acceptance criteria for such a device. The content is primarily focused on demonstrating substantial equivalence to a predicate device through material, design, and intended use comparisons, rather than performance testing against specific acceptance criteria for an AI algorithm.

Therefore, I cannot provide the requested information from the provided text as it does not pertain to the performance and validation of an AI/ML medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Zest Anchors, LLC Richard Hines Regulatory Affairs Manager 2875 Loker East Carlsbad, California 92026

Re: K220252

Trade/Device Name: High Retention Attachment System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: March 24, 2022 Received: March 28, 2022

Dear Richard Hines:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K220252

Device Name

High Retention Attachment System

Indications for Use (Describe)

The High Retention Attachment System is designed to support fixed, partial or full arch restorations on endosseous dental implants in the mandible or maxilla for the purpose of restory function. It is used in fixed hybrid restorations that can be attached with a snap-in system. The High Retention Attachment System is compatible with the following implants:

Implant ManufacturerImplant SystemImplant Diameter
Ace SurgicalInfinity OCTAGON3.3, 4.1, 4.8
Infinity TRI-CAM3.5, 4.3, 5.0
Infinity Internal Hex3.7, 4.1, 4.7, 5.1
BioHorizonsTapered Internal3.0
Tapered Plus3.8
Laser-Lok3.0
Biomet 3i3iT3, NanoTite, OSSEOTITE3.25, 3.75, 4.0
Blue Sky BioQuattro3.3, 4.1, 4.8
One Stage3.3, 4.1, 4.8
CamlogSCREW-LINE ROOT-LINE 23.3, 3.8, 4.3, 5.0
SCREW-LINE3.3, 3.8, 4.3, 5.0
DentsplyAstra Tech OsseoSpeed EV3.6, 4.2, 4.8
Xive Frialit-23.4, 3.8, 4.5, 5.5
Ankylos C/X3.5
Astra Tech OsseoSpeed TX3.5, 4.0, 4.5, 5.0
Hiossen, Inc.SS3.5, 4.0, 4.5
ET, TS3.5, 4.0, 4.5, 5.0, 6.0, 7.0
US3.5, 3.3, 4.5, 4.0, 5.0, 6.0, 7.0
Implant DirectSwish Plus4.1, 4.8
SwishTapered4.1, 4.8
InterActive3.2, 3.7, 4.3, 5.0
ReActive3.7, 4.2, 4.7, 5.7
RePlant3.5, 4.3, 5.0
Legacy 1, 2, 3, 43.7, 4.2
Legacy 2, 3, 43.2, 4.7, 5.2
Legacy 1, 35.7
Legacy 2, 45.7, 7.0
Legacy 14.7
IDSMegaGen Any Ridge3.5, 4.0, 4.5, 5.0, 5.5, 6.0, 6.5, 7.0, 7.5, 8.0
KeystonePrimaConnex3.5, 4.1, 5.0
Genesis3.5
TILOBEMAXX7.0, 8.0, 9.0
MISC13.3, 3.75, 4.2
V33.3, 3.9, 4.3, 5.0
SEVEN3.3
M43.3
NobelReplace, Replace Select, Nobel Speedy3.5, 4.3, 5.0
NobelActive, NobelParallel CC, NobelReplace CC3.5, 4.3, 5.5
Branemark, Nobel Speedy, Groovy3.3, 3.75, 4.0, 5.0
OCO BiomedicalEngage3.25, 4.0, 5.0
Southern ImplantsTri-Nex3.5, 4.3, 5.0
Tri-MAX77.0
StraumannRoxolid SLActive, Roxolid SLA3.3, 4.1, 4.8
Zest Dental SolutionsLocator Overdenture Implant System2.4, 2.9, 3.5, 3.9, 4.4, 4.9
Zimmer DentalTrabecular Metal, Tapered Screw-Vent, Screw-Vent,Advent3.7, 4.1, 4.7, 6.0
Spline Reliance Cylinder3.25, 4.0, 5.0
Spline Twist3.75

{3}------------------------------------------------

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Zest Dental Solutions. The logo has the letters "ZD" in a bold, teal font on the left side. To the right of the letters, the words "ZEST DENTAL" are written in a bold, dark teal font, with the word "SOLUTIONS" written in a lighter teal font below it. The logo is simple and modern, and the colors are clean and professional.

510(K) Summary – K220252 High Retention Attachment System

i. General Information on Submitter

Applicant:Zest Anchors, LLC
Address:2875 Loker Avenue, EastCarlsbad, CA 92010 USA
Telephone:442-244-4835, ext. 480
Cellphone:949-395-8795
Contact Person:Richard Hines
Contact Title:Regulatory Affairs Manager
Email:richard.hines@zestdent.com
Date Prepared:April 13, 2022
Establishment Registration:3006796211

ii. General Information on Device

Proprietary Name:High Retention Attachment System
Common Name:Dental Implant Abutment
Classification Name:Endosseous dental implant abutment(21 CFR 872.3630)
Regulatory Class:II
Product Code:NHA (Abutment, Implant, Dental, Endosseous)
  • iii. Predicate Device
Predicate Device510(k) Number
High Retention Attachment System: Zest Anchors, LLCK213391
Reference Devices510(k) Number
Locator Overdenture Implant System, Zest Anchors, LLCK120198
Locator Overdenture Implant System, Zest Anchors, LLCK203701

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Zest Dental Solutions. The logo has the letters "ZD" on the left side in a blue gradient color. To the right of the letters is the text "ZEST DENTAL" in a bold, sans-serif font, with the word "SOLUTIONS" underneath in a smaller font.

iv. Description of Device

The High Retention Attachment System is a system that provides rigid connection of fixed, partial, and full arch restorations (fixed/detachable hybrid dentures) to endosseous dental implants. It consists of abutments, attachment housings, inserts and seating and removal tools. The abutments are provided in various cuff heights with the implant and abutment connection specific to the Zest Anchors, LLC or OEM implant. The subject device abutments are made of titanium alloy conforming to ASTM F136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401) and are TiN (titanium nitride) coated.

The High Retention Attachment System is as provided with an abutment (LOCATOR Attachment) and the surgical instrumentation necessary to place the implant.

The High Retention Attachment System is designed to accommodate a path of insertion diverqence of up to 20° per implant and no more than 40° of divergence between implants.

The abutments are identical to the High Retention Attachment System devices cleared in the predicate. K213391, with the main difference being an additional implant system compatibility. The additional compatible implant system, Zest Anchors, LLC's Locator Overdenture Implant (LODI) implants, were previously cleared in K120198 and K203701 and are included as reference devices, in the K220252 submission.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image is a logo for Zest Dental Solutions. The logo features the letters "ZD" in a bold, blue font on the left side. To the right of the letters, the words "ZEST DENTAL" are written in a dark blue, sans-serif font. Below "ZEST DENTAL", the word "SOLUTIONS" is written in a smaller, light blue font.

Indication for Use V.

The High Retention System is designed to support fixed, partial or full arch restorations on endosseous dental implants in the mandible or maxilla for the purpose of restoring masticatory function. It is used in fixed hybrid restorations that can be attached with a snap-in system. The High Retention Attachment System is compatible with the following implants:

Implant ManufacturerImplant SystemImplant Diameter
Ace SurgicalInfinity OCTAGON3.3, 4.1, 4.8
Infinity TRI-CAM3.5, 4.3, 5.0
Infinity Internal Hex3.7, 4.1, 4.7, 5.1
BioHorizonsTapered Internal3.0
Tapered Plus3.8
Laser-Lok3.0
Biomet 3i3iT3, NanoTite, OSSEOTITE3.25, 3.75, 4.0
Blue Sky BioQuattro3.3, 4.1, 4.8
One Stage3.3, 4.1, 4.8
CamlogSCREW-LINE ROOT-LINE 23.3, 3.8, 4.3, 5.0
SCREW-LINE3.3, 3.8, 4.3, 5.0
DentsplyAstra Tech OsseoSpeed EV3.6, 4.2, 4.8
Xive Frialit-23.4, 3.8, 4.5, 5.5
Ankylos C/X3.5
Astra Tech OsseoSpeed TX3.5, 4.0, 4.5, 5.0
Hiossen, Inc.SS3.5, 4.0, 4.5
ET, TS3.5, 4.0, 4.5, 5.0, 6.0, 7.0
US3.5, 3.3, 4.5, 4.0, 5.0, 6.0, 7.0
Implant DirectSwish Plus4.1, 4.8
SwishTapered4.1, 4.8
InterActive3.2, 3.7, 4.3, 5.0
ReActive3.7, 4.2, 4.7, 5.7
RePlant3.5, 4.3, 5.0
Legacy 1, 2, 3, 43.7, 4.2
Legacy 2, 3, 43.2, 4.7, 5.2
Legacy 1, 35.7
Legacy 2, 45.7, 7.0
Legacy 14.7
IDSMegaGen Any Ridge3.5, 4.0, 4.5, 5.0, 5.5, 6.0, 6.5, 7.0, 7.5, 8.0
KeystonePrimaConnex3.5, 4.1, 5.0
Genesis3.5
TILOBEMAXX7.0, 8.0, 9.0
MISC13.3, 3.75, 4.2
V33.3, 3.9, 4.3, 5.0
SEVEN3.3
M43.3
NobelReplace, Replace Select, NobelSpeedy3.5, 4.3, 5.0
NobelActive, NobelParallel CC, NobelReplace CC3.5, 4.3, 5.5
Branemark, Nobel Speedy, Groovy3.3, 3.75, 4.0, 5.0
OCO BiomedicalEngage3.25, 4.0, 5.0
Southern ImplantsTri-Nex3.5, 4.3, 5.0
Tri-MAX77.0
StraumannRoxolid SLActive, Roxolid SLA3.3, 4.1, 4.8
Zest Dental SolutionsLocator Overdenture Implant System2.4, 2.9, 3.5, 3.9, 4.4, 4.9
Zimmer DentalTrabecular Metal, Tapered Screw-Vent, Screw-Vent, Advent3.7, 4.1, 4.7, 6.0
Spline Reliance Cylinder3.25, 4.0, 5.0
Spline Twist3.75

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the logo for Zest Dental Solutions. The logo features the letters "ZD" in a bold, blue font on the left side. To the right of the letters, the words "ZEST DENTAL" are stacked on top of the word "SOLUTIONS" in a smaller, blue font. The logo is simple and professional, and it is likely used to represent the company's brand.

vi. Predicate Device Comparison

The following table compares the Indications for Use and key technological

characteristics of the subject and predicate device:

Characteristic/ FeatureHigh RetentionAttachment System(Subject Device)High RetentionAttachmentSystem(Predicate Device- K213391)LocatorOverdentureImplant System(ReferenceDevice #1 -K120198)The LocatorOverdentureImplant System(ReferenceDevice #2 -K203701)Comparison
Reason forpredicate orreferencen/aIndications, function,material of insert,mechanical (retention)performanceDesign of implant added to compatibility tablePredicate is unchanged andreference device is added tocleared compatibility table
Indicationfor useLocator HighRetentionAttachment Systemis designed tosupport fixed, partialor full archrestorations onendosseous dentalimplants in themandible or maxillafor the purpose ofrestoring masticatoryfunction. It is used infixed hybridrestorations that canbe attached with asnap-in system.High RetentionAttachment System isdesigned to supportfixed, partial or full archrestorations onendosseous dentalimplants in the mandibleor maxilla for the purposeof restoring masticatoryfunction. It is used infixed hybrid restorationsthat can be attached witha snap-in system.LocatorOverdentureImplant System isdesigned to retainoverdentures orpartial dentures inthe mandible ormaxilla.The LocatorOverdentureImplant System isdesigned to retainoverdentures orpartial dentures inthe mandible ormaxilla.Immediate loading isindicated when goodprimary stability hasbeen achieved and withappropriate occlusalloading.Similar:The subject and the primarypredicate K223391 have identicalindications for use, with theexception of an additional implantsystem compatibility for the subjectdevices.The additional compatible implantsystem, Zest Anchors, LLC'sLocator Overdenture Implant(LODI) implants, were previouslycleared in K120198 and K203701,and are included as referencedevices.
FDAProductCodeNHA (Abutment, Implant,Dental, Endosseous) 21 CFR872.3630NHA (Abutment,Implant, Dental,Endosseous) 21CFR 872.3630DZE (Implant,Endossesous,Root-Form) 21CFR 872.3640DZE (Implant,Endossesous, Root-Form) 21 CFR872.3640Substantially Equivalent:The subject device havedemonstrated substantialequivalence to the primarypredicate device, K213391, in thatthey utilize identical materials,fundamental designs, intendeduse and principles of operation.The reference devices, K120198and K203701, are only includedas reference devices for additionalimplant system compatibilities.
DESIGN
Abutmentplatformdiameter2.3 mm to 7.0 mm,multiple systems2.3 mm to 7.0mm, multiplesystemsN/AN/ASubstantially Equivalent.Predicate and subjectabutment platformdiameters are identical.
AbutmentangleStraightStraightN/AN/ASubstantially Equivalent.Predicate and subject abutmentangles are identical
Abutment /implantinterfaceConical, External Hex,Internal Hex, InternalMulti LobeConical,External Hex,Internal Hex,Internal MultiLobeExternal HexExternal HexSubstantially Equivalent.Predicate and subject deviceabutment/implant interfaces areidentical.
DivergenceAllowance20°/40° (except whennot allowed by theimplant manufacturer)20°/40° (exceptwhen not allowedby the implantmanufacturer)N/AN/ASubstantially Equivalent.Divergence allowance isunchanged from the predicatedevice.
ProsthesisattachmenttypePEEK insert retainedin denture attachmenthousingPEEK insertretained indentureattachmenthousingN/AN/ASubstantially Equivalent.Predicate and subject deviceprosthesis attachment type areidentical.
Materials
Abutmentor ImplantMaterialAbutment: Ti-6AI-4VELIAbutment: Ti-6AI-4V ELIImplant: Ti6AI-4V ELIImplant: Ti6AI-4V ELISubstantially Equivalent.both the abutment and theimplants in the subject,predicate, and referencedevices are made from thesame titanium alloy (Ti-6Al-4V ELI).
DeviceMaterialSurfaceTreatmentAbutment: TiNAbutment: TiNImplant:ResorbableBlast Media(RBM) usingMCD ApataticAbrasiveImplant:ResorbableBlast Media(RBM) usingMCD ApataticAbrasiveSubstantially Equivalent.Predicate and subject device useidentical surface treatmentprocess.
ProstheticretentioncomponentPEEKPEEKN/AN/ASubstantially Equivalent.Predicate and subject deviceretention components areidentical.
ReferenceImplantSizeN/AN/A2.4mm, 2.9mm3.5mm, 3.9mm,4.4mm, 4.9mmReference Devices are clearedfor implants of 2.4mm to 4.9mmin diameter
Sterilization
SterileNoNoYesYesSubstantially Equivalent.Predicate and subject device areboth supplied non-sterile.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Zest Dental Solutions. The logo has the letters "ZD" in a bold, blue font on the left side. To the right of the letters are the words "ZEST DENTAL" in a smaller, blue font, with the word "SOLUTIONS" underneath in a lighter blue. The logo is simple and modern, and the use of blue gives it a professional and trustworthy feel.

As noted in the table above, the subject and predicate devices are identical. Fatigue testing according to ISO 14801: 2016 was performed for Zest Anchors, LLC's Locator Overdenture Implant (LODI) implants (K120198/K203701) and leveraged to support modifications to the K213391 compatibility table.

Summary of Non-Clinical Performance Testing vii.

No new testing was performed as a part of this submission for the determination of substantial equivalence.

Fatigue testing, sterilization, and biocompatibility testing are all leveraged from previous submissions

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Zest Dental Solutions. The logo features the letters "ZD" in a bold, blue font on the left side. To the right of the letters, the words "ZEST DENTAL" are written in a similar bold, blue font. Below "ZEST DENTAL", the word "SOLUTIONS" is written in a lighter blue color.

MR safety for the subject device system was based on the "worst-case" of the entire system including all variations (all compatible implant bodies, dental abutments, and fixation screws) and material composition. The "worst-case" assessment includes an evaluation of the High Retention Attachment System devices in the MRI environment using scientific rationale and published literature (i.e., Woods, Terry O., Jana G. Delfino, and Sunder Rajan. "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795).

In addition, the "worst-case" MR Safety rationale addressed parameters per the FDA guidance, "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque.

viii. Substantial Equivalence

The risk management activities and results of the other activities described above provide reasonable assurance that the subject devices have demonstrated substantial equivalence to the predicate devices in the that they utilize the same materials and fundamental designs and also have the same intended use and principles of operation.

§ 872.3630 Endosseous dental implant abutment.

(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)