K Number
K220249

Validate with FDA (Live)

Date Cleared
2022-05-26

(115 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition these gloves were tested for use with Chemotherapy drugs in accordance with ASTM D6978 standards. Gloves is direct contact and not to be worn for more than 24 hours.

Device Description

The subject device is Sterile Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs And Fentanyl Citrate (Blue). The subject device is a patient examination glove made from nitrile latex compound, blue color, powder free and sterile (Per 21 CFR 880.6250, class I). The device is direct contact and not to be worn for more than 24 hours. The device meets all the specifications in ASTM D6319-19, Standard specification for Nitrile Examination Gloves. Additionally, the gloves have been tested for biocompatibility per 10993-11 and permeability to chemotherapy drugs and Fentanyl Citrate per ASTM D6978-05(2019).

AI/ML Overview

This document is a 510(k) premarket notification for "Sterile Nitrile Powder Free Examination Gloves Tested For Use With Chemotherapy Drugs And Fentanyl Citrate (Blue)". It details the device's characteristics, performance, and comparison to a predicate device.

Here's the breakdown of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

The device's performance is tested against various physical properties and chemical permeation criteria.

Physical Properties (based on ASTM D6319-19):

ItemsAcceptance CriteriaReported Device Performance (Results)
LengthMinimum 230mmAll size ≥230
Palm WidthXS: 70±10mm76-78mm
S: 80±10mm86-88 mm
M: 95±10mm96-98mm
L: 110±10mm106-108 mm
XL: 120±10mm116-118 mm
XXL: 130±10mm126-128 mm
ThicknessPalm: 0.05mm (min)0.07-0.08mm
Finger: 0.05mm (min)0.09-0.11mm
Tensile Strength, Before Aging14MPa, min15.7-20.6 MPa
Tensile Strength, After Accelerated Aging14MPa, min15.5-19.5 MPa
Ultimate Elongation, Before Aging500%, min500-550%
Ultimate Elongation, After Accelerated Aging400%, min410-510%
Freedom from holesG-I, AQL 2.5Meet and above AQL2.5 requirements
Powder-Content≤ 2 mg per glove≤ 2 mg, meet requirements

Chemotherapy Drug and Fentanyl Citrate Permeation (based on ASTM D6978 standards):

Chemotherapy Drug / Fentanyl Citrate ConcentrationReported Minimum Breakthrough Detection Time (Minutes)
Bleomycin Sulfate 15mg/ml (15000 ppm)>240
Busulfan 6mg/ml (6,000 ppm)>240
Carboplatin 10mg/ml (10,000 ppm)>240
Carmustine 3.3 mg/ml (3,300 ppm)22.4
Chloroquine 50mg/ml (50,000ppm)>240
Cisplatin 1mg/ml (1,000 ppm)>240
Cyclophosphamide 20mg/ml (20,000 ppm)>240
Cyclosporin 100 mg/ml (100,000 ppm)>240
Cytarabine HCL, 100 mg/ml (100,000 ppm)>240
Dacarbazine 10 mg/ml (10,000 ppm)>240
Daunorubicin HCL, 5 mg/ml (5,000 ppm)>240
Docetaxel HCL, 10 mg/ml (10,000 ppm)>240
Doxorubicin HCL, 2 mg/ml (2,000 ppm)>240
Epirubicin HCL, 2 mg/ml (2,000 ppm)>240
Etoposide, 20 mg/ml (20,000 ppm)>240
Fludarabine, 25 mg/ml (25,000 ppm)>240
Fluorouracil, 50mg/ml (50,000ppm)>240
Gemcitabine, 38mg/ml (38,000ppm)>240
Idarubicin HCL, 1mg/ml (1,000ppm)>240
Ifosfamide, 50mg/ml (50,000ppm)>240
Irinotecan, 20mg/ml (20,000ppm)>240
Mechlorethamine HCI, 1mg/ml (1,000ppm)>240
Melphalan, 5mg/ml (5,000ppm)>240
Methotrexate, 25mg/ml (25,000ppm)>240
Mitomycin C, 0.5mg/ml (500ppm)>240
Mitoxantrone HCL, 2mg/ml (2,000ppm)>240
Oxaliplatin, 5mg/ml (5,000ppm)>240
Paclitaxel, 6mg/ml (6,000ppm)>240
Paraplatin, 10mg/ml (10,000ppm)>240
Retrovir, 10mg/ml (10,000ppm)>240
Rituximab, 10mg/ml (10,000ppm)>240
Thiotepa, 10mg/ml (10,000ppm)46.3
Topotecan, 1mg/ml (1,000ppm)>240
Trisenox, 1mg/ml (1,000ppm)>240
Velcade, 1mg/ml (1,000ppm)>240
Vincristine Sulfate, 1mg/ml (1,000ppm)>240
Fentanyl Citrate Injection (100 mcg/2ml)>240

Note: For Carmustine and Thiotepa, the document explicitly states warnings due to low permeation times, even though they have a reported BDT.

Biocompatibility Studies (ISO 10993-10, ISO 10993-11, ISO 10993-5):

Technological CharacteristicsStandard / Test / FDA GuidanceResult Summary (Acceptance Criteria & Performance)
Primary Skin Irritation TestISO 10993-10:2010Meets (Under the conditions of the study, not an irritant)
Dermal Sensitization AssayISO 10993-10:2010Meets (Under the conditions of the study, not a sensitizer)
In Vitro Cytotoxicity TestISO 10993-5:2009The test article extract showed potential toxicity to L929 cells. Cytotoxicity concern was addressed by acute systematic toxicity testing.
Acute Systemic Toxicity TestISO 10993-11:2017Meets (Under the conditions of this study, there was no evidence of systemic toxicity.)

2. Sample size used for the test set and the data provenance

The document specifies the standards used for testing, which implicitly defines the sample sizes and testing methodologies. However, it does not explicitly state the exact sample sizes used for each specific test (e.g., number of gloves tested for each chemotherapy drug).

  • Standards Used: ASTM D6319-19 (for dimensions, physical properties, freedom from holes, powder content) and ASTM D6978-05(2019) (for permeability to chemotherapy drugs and Fentanyl Citrate). Biocompatibility testing was done per ISO 10993-10, ISO 10993-11, and ISO 10993-5.
  • Data Provenance: Not explicitly stated, but based on the owner's identification (Grand Work Plastic Products Co., Ltd. in Hebei, China) and the contact information (Hongray USA Medical Products Inc. in Chino, California), the testing was likely conducted by or for the manufacturer, potentially in China or a contracted lab. The data is retrospective for the purpose of this 510(k) submission, as the tests have already been performed.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This question is not applicable as the document describes the testing of physical and chemical properties of gloves against established industry standards. The "ground truth" here is the performance requirement defined by these standards, not expert consensus on medical images or clinical outcomes. The tests are objective measurements, not subjective expert interpretations.

4. Adjudication method for the test set

Not applicable. The tests performed are objective, quantitative measurements against predefined specifications (e.g., minimum length, minimum tensile strength, breakthrough detection time). There is no "adjudication" in the sense of reconciling differing expert opinions.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a submission for sterile nitrile examination gloves, not an AI-powered diagnostic device. No human readers or AI assistance are involved in the performance evaluation of the gloves.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical medical device (gloves), not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for the device's performance is objective, laboratory-measured data based on established international standards:

  • Physical properties (dimensions, strength, elongation, hole detection, powder content) are measured quantitatively according to ASTM D6319-19, ASTM D5151-19, and ASTM D6124-06(2017).
  • Chemical permeation (breakthrough detection time for chemotherapy drugs and Fentanyl Citrate) is measured according to ASTM D6978-05(2019).
  • Biocompatibility (skin irritation, sensitization, systemic toxicity, cytotoxicity) is assessed using in vitro and in vivo tests guided by ISO 10993 standards.

8. The sample size for the training set

Not applicable. This is a physical medical device, not a machine learning model that requires a training set.

9. How the ground truth for the training set was established

Not applicable. This is a physical medical device, not a machine learning model.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

May 26, 2022

Grand Work Plastic Products Co., Ltd % Kathy Liu Project Manager Hongray USA Medical Products Inc. 3973 Schaefer Avenue Chino, California 91710

Re: K220249

Trade/Device Name: Sterile Nitrile Powder Free Examination Gloves Tested For Use With Chemotherapy Drugs And Fentanyl Citrate (Blue) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC, ODO Dated: April 18, 2022 Received: April 26, 2022

Dear Kathy Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bifeng Qian, M.D., Ph.D Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220249

Device Name

Sterile Nitrile Powder Free Examination Gloves Tested For Use With Chemotherapy Drugs And Fentanyl Citrate (Blue)

Indications for Use (Describe)

Glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition these gloves were tested for use with Chemotherapy drugs in accordance with ASTM D6978 standards.

Gloves is direct contact and not to be worn for more than 24 hours

Chemotherapy DrugMinimum BDT (Minutes)
Bleomycin Sulfate 15mg/ml (15000 ppm)>240
Busulfan 6mg/ml (6,000 ppm)>240
Carboplatin 10mg/ml (10,000 ppm)>240
Carmustine 3.3 mg/ml (3,300 ppm)22.4
Chloroquine 50mg/ml (50,000ppm)>240
Cisplatin 1mg/ml (1,000 ppm)>240
Cyclophosphamide 20mg/ml (20,000 ppm)>240
Cyclosporin 100 mg/ml (100,000 ppm)>240
Cytarabine HCL, 100 mg/ml (100,000 ppm)>240
Dacarbazine 10 mg/ml (10,000 ppm)>240
Daunorubicin HCL, 5 mg/ml (5,000 ppm)>240
Docetaxel HCL, 10 mg/ml (10,000 ppm)>240
Doxorubicin HCL, 2 mg/ml (2,000 ppm)>240
Epirubicin HCL, 2 mg/ml (2,000 ppm)>240
Etoposide, 20 mg/ml (20,000 ppm)>240
Fludarabine, 25 mg/ml (25,000 ppm)>240
Fluorouracil, 50mg/ml (50,000ppm)>240
Gemcitabine, 38mg/ml (38,000ppm)>240
Idarubicin HCL, 1mg/ml (1,000ppm)>240
Ifosfamide, 50mg/ml (50,000ppm)>240
Irinotecan, 20mg/ml (20,000ppm)>240
Mechlorethamine HCI, 1mg/ml (1,000ppm)>240
Melphalan, 5mg/ml (5,000ppm)>240
Methotrexate, 25mg/ml (25,000ppm)>240
Mitomycin C, 0.5mg/ml (500ppm)>240
Mitoxantrone HCL, 2mg/ml (2,000ppm)>240
Oxaliplatin, 5mg/ml (5,000ppm)>240
Paclitaxel, 6mg/ml (6,000ppm)>240
Paraplatin, 10mg/ml (10,000ppm)>240
Retrovir, 10mg/ml (10,000ppm)>240
Rituximab, 10mg/ml (10,000ppm)>240
Thiotepa, 10mg/ml (10,000ppm)46.3
Topotecan, 1mg/ml (1,000ppm)>240
Trisenox, 1mg/ml (1,000ppm)>240
Velcade, 1mg/ml (1,000ppm)>240
Vincristine Sulfate, 1mg/ml (1,000ppm)>240

{3}------------------------------------------------

  • Please note that the following drugs have extremely low permeation times: Carmustine: 22.4 minutes, Thiotepa: 46.3 minutes WARNING: DO NOT USE WITH CARMUSTINE WARNING: DO NOT USE WITH THIOTEPA
Fentanyl Citrate and Concentration
Fentanyl Citrate Injection (100 mcg/2ml)

Minimum Breakthrough Detection Time in Minutes >240

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Donggao Industrial Zone, Zanhuang, Hebei, 050000, China

Product: Sterile Nitrile Powder Free Examination Gloves Tested For Use With Chemotherapy Drugs And Fentanyl Citrate (Blue)

510(K) SUMMARY

This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR §807.92.

The assigned 510(K) number is: K220249

Date Prepared: April 18, 2022

1. Owner's Identification:

Mrs. Wu Yuli Grand Work Plastic Products Co., Ltd. Donggao Industrial Zone, Zanhuang, Hebei, 050000, China Tel: 86-311-66179668

Contact: Ms. Kathy Liu, Project Manager Address: 3973 Schaefer Ave., Chino, CA 91710 Tel: 909-590-1611

2. Name of the Device:

Trade Name: Sterile Nitrile Powder Free Examination Gloves Tested For Use With Chemotherapy Drugs And Fentanyl Citrate (Blue) Common Name: Exam Gloves Classification Name: Patient Examination Glove Classification Regulation: 880.6250 Classification Panel: 880 General Hospital and Personal Use Product Code: LZA, LZC and QDO Device Class: Class I

3. Predicate Device Information:

Hartalega SDN.BHD. Sterile Nitrile Powder Free Examination Glove Tested for use with Chemotherapy Drugs and Fentanyl Citrate (Blue) (K201531)

4. Device Description:

The subject device is Sterile Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs And Fentanyl Citrate (Blue). The subject device is a patient examination glove made from nitrile latex compound, blue color, powder free and sterile (Per 21 CFR 880.6250, class I). The device is direct contact and not to be worn for more than 24 hours. The device meets all the specifications in ASTM D6319-19, Standard specification for Nitrile Examination Gloves.

Additionally, the gloves have been tested for biocompatibility per 10993-11 and permeability to chemotherapy drugs and Fentanyl Citrate per ASTM D6978-05(2019).

5. Indications for Use:

A patient examination glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs & Fentany] Citrate and provides protection against following chemotherapy drugs and the minimum breakthrough time are listed as follows.

Chemotherapy DrugMinimum BDT (Minutes)
Bleomycin Sulfate 15mg/ml (15000 ppm)>240
Busulfan 6mg/ml (6,000 ppm)>240
Carboplatin 10mg/ml (10,000 ppm)>240

{5}------------------------------------------------

Donggao Industrial Zone, Zanhuang, Hebei, 050000, China

Product: Sterile Nitrile Powder Free Examination Gloves Tested For Use With Chemotherapy Drugs And Fentanyl Citrate (Blue)

Carmustine 3.3 mg/ml (3,300 ppm)22.4
Chloroquine 50mg/ml (50,000ppm)>240
Cisplatin 1mg/ml (1,000 ppm)>240
Cyclophosphamide 20mg/ml (20,000 ppm)>240
Cyclosporin 100 mg/ml (100,000 ppm)>240
Cytarabine HCL, 100 mg/ml (100,000 ppm)>240
Dacarbazine 10 mg/ml (10,000 ppm)>240
Daunorubicin HCL, 5 mg/ml (5,000 ppm)>240
Docetaxel HCL, 10 mg/ml (10,000 ppm)>240
Doxorubicin HCL, 2 mg/ml (2,000 ppm)>240
Epirubicin HCL, 2 mg/ml (2,000 ppm)>240
Etoposide, 20 mg/ml (20,000 ppm)>240
Fludarabine, 25 mg/ml (25,000 ppm)>240
Fluorouracil, 50mg/ml (50,000ppm)>240
Gemcitabine, 38mg/ml (38,000ppm)>240
Idarubicin HCL, 1mg/ml (1,000ppm)>240
Ifosfamide, 50mg/ml (50,000ppm)>240
Irinotecan, 20mg/ml (20,000ppm)>240
Mechlorethamine HCI, 1mg/ml (1,000ppm)>240
Melphalan, 5mg/ml (5,000ppm)>240
Methotrexate, 25mg/ml (25,000ppm)>240
Mitomycin C, 0.5mg/ml (500ppm)>240
Mitoxantrone HCL, 2mg/ml (2,000ppm)>240
Oxaliplatin, 5mg/ml (5,000ppm)>240
Paclitaxel, 6mg/ml (6,000ppm)>240
Paraplatin, 10mg/ml (10,000ppm)>240
Retrovir, 10mg/ml (10,000ppm)>240
Rituximab, 10mg/ml (10,000ppm)>240
Thiotepa, 10mg/ml (10,000ppm)46.3
Topotecan, 1mg/ml (1,000ppm)>240
Trisenox, 1mg/ml (1,000ppm)>240
Velcade, 1mg/ml (1,000ppm)>240
Vincristine Sulfate, 1mg/ml (1,000ppm)>240
Fentanyl Citrate Injection (100 mcg/2ml)>240
  • Please note that the following drugs have extremely low permeation times: Carmustine: 22.4 minutes, Thiotepa: 46.3 minutes WARNING: DO NOT USE WITH CARMUSTINE WARNING: DO NOT USE WITH THIOTEPA

6. Specification for Nitrile Gloves:

ItemsAcceptance CriteriaResults
LengthMinimum 230mmAll size ≥230
Palm WidthXS: 70±10mm76-78mm
S: 80±10mm86-88 mm
M:95±10mm96 -98mm
L:110±10mm106-108 mm
XL: 120±10mm116-118 mm

{6}------------------------------------------------

Donggao Industrial Zone, Zanhuang, Hebei, 050000, China

Product: Sterile Nitrile Powder Free Examination Gloves Tested For Use With Chemotherapy Drugs And Fentanyl Citrate (Blue)

XXL: $130\pm10$ mm126-128 mm
ThicknessPalm: 0.05mm (min)0.07-0.08mm
Finger: 0.05mm (min)0.09-0.11mm
Tensile Strength, Before Aging14MPa, min15.7-20.6 MPa
Tensile Strength, After Accelerated Aging14MPa, min15.5-19.5 MPa
Ultimate Elongation, Before Aging500%, min500-550%
Ultimate Elongation, After Accelerated Aging400%, min410-510%
Freedom from holesG-I, AQL 2.5Meet and above AQL2.5 requirements
Powder-Content$\leq$ 2 mg per glove$\leq$ 2 mg, meet requirements

Gloves meet all the specification listed in ASTM D 6319-19.

7. Comparison of Subject Device and Predicate Device:

CharacteristicsSubject DevicePredicate DeviceRemark
K220249K201531
Trade NameSterile Powder Free NitrileExamination Gloves (Blue), Testedfor Use with Chemotherapy DrugsNitrile Examination GlovesSterile Tested for Use withChemotherapy Drugs claim, WhiteSimilar
Product CodeLZA, LZC and QDOLZA, LZC and QDOSame
Regulation Number21 CFR 880.625021 CFR 880.6250Same
ClassIISame
Indications for UseSubject device is a disposable deviceintended for medical purpose that isworn on the examiner's hand orfinger to prevent contaminationbetween patient and examiner. Thedevice is for over-the-counter use.Predicate device is a disposable deviceintended for medical purpose that is wornon the examiner's hand or finger toprevent contamination between patient andexaminer. The device is for over-the-counter use.Same
DescriptionSterile Examination glove made ofNitrile and color Blue and testedwith chemotherapy DrugsSterile Examination glove made of Nitrileand color White and tested withchemotherapy DrugsSame
Design FeatureAmbidextrousAmbidextrousSame
Material useNitrile Synthetic LatexNitrile Synthetic LatexSame
ColorBlueBlueSame
Sterility ISO 11737-2EN ISO 11137-2SterilitySterilitySame
Single usedSingle usedSingle usedSame
Non Sterile or SterileSterileSterileSame
10993-10:2010 Skin IrritationStudyUnder the conditions of thestudy, not an irritantUnder the conditions of the study, notan irritantSame
10993-10:2010 MaximizationSensitization StudyUnder the conditions of thestudy, not a sensitizerUnder the conditions of the study, not asensitizerSame
10993-5:2009 In VitroUnder the conditions of this study,Under the conditions of this study, theSame

{7}------------------------------------------------

Donggao Industrial Zone, Zanhuang, Hebei, 050000, China

Product: Sterile Nitrile Powder Free Examination Gloves Tested For Use With Chemotherapy Drugs And Fentanyl Citrate (Blue)

Cytotoxicity Testthe test article extract showed potential toxicity to L929 cells.Cytotoxicity concern was addressed by acute systematic toxicity testing.test article extract showed potential toxicity to L929 cells.Cytotoxicity concern was addressed by acute systematic toxicity testing.
ISO 10993 Part 11Acute Systemic Toxicity TestUnder the conditions of this study, there was no evidence of systemic toxicity.Under the condition of this study, Nitrile Examination glove, sterile shows no adverse biological reactionSame
TechnologicalCharacteristicsSubject DeviceK220249Predicate DeviceK201531Remark
LengthMinimum 230mmMinimum 230mmSame
Palm Width (size) (mm)
XS70±1070±10Same
S80±1080±10Same
M95±1095±10Same
L110±10110±10Same
XL120±10120±10Same
XXL130±10/Different
Thickness(mm)
FingerMinimum 0.05Minimum 0.05Same
PalmMinimum 0.05Minimum 0.05Same
Tensile Strength, Before Aging14MPa, min14MPa, minSame
Ultimate Elongation,Before Aging500%, min500%, minSame
Tensile Strength, AfterAccelerated Aging14MPa, min14MPa, minSame
Ultimate Elongation, AfterAccelerated Aging400%, min400%, minSame
Freedom from holesIn accordance with ASTM D5151-19, following ASTMD6319- 19, G-I, AQL 2.5In accordance with ASTM D5151-19, following ASTMD6319- 19, G-I, AQL 2.5Same
Residual Powder<2 mg per glove<2 mg per gloveSame

Dimensions and Performance Comparison Table:

Chemotherapy Permeation Comparison Claim:

Tested Chemotherapy Drug andConcentrationMinimum BDT (Minutes)Remark
Predicate DeviceSubject Device
Bleomycin Sulfate 15mg/ml (15000 ppm)/>240Different
Busulfan 6mg/ml (6,000 ppm)/>240Different
Carboplatin 10mg/ml (10,000 ppm)/>240Different
Carmustine 3.3 mg/ml (3,300 ppm)26.322.4Similar

{8}------------------------------------------------

Donggao Industrial Zone, Zanhuang, Hebei, 050000, China

Product: Sterile Nitrile Powder Free Examination Gloves Tested For Use With Chemotherapy Drugs And Fentanyl Citrate (Blue)

Chloroquine 50mg/ml (50,000ppm)/>240Different
Cisplatin 1mg/ml (1,000 ppm)>240>240Same
Cyclophosphamide 20mg/ml (20,000 ppm)>240>240Same
Cyclosporin 100 mg/ml (100,000 ppm)/>240Different
Cytarabine HCL, 100 mg/ml (100,000 ppm)/>240Different
Dacarbazine 10 mg/ml (10,000 ppm)>240>240Same
Daunorubicin HCL, 5 mg/ml (5,000 ppm)/>240Different
Docetaxel HCL, 10 mg/ml (10,000 ppm)/>240Different
Doxorubicin HCL, 2 mg/ml (2,000 ppm)>240>240Same
Epirubicin HCL, 2 mg/ml (2,000 ppm)/>240Different
Etoposide, 20 mg/ml (20,000 ppm)>240>240Same
Fludarabine, 25 mg/ml (25,000 ppm)/>240Different
Fluorouracil, 50mg/ml (50,000ppm)>240>240Same
Gemcitabine, 38mg/ml (38,000ppm)/>240Different
Idarubicin HCL, 1mg/ml (1,000ppm)/>240Different
Ifosfamide, 50mg/ml (50,000ppm)/>240Different
Irinotecan, 20mg/ml (20,000ppm)/>240Different
Mechlorethamine HCI, 1mg/ml (1,000ppm)/>240Different
Melphalan, 5mg/ml (5,000ppm)/>240Different
Methotrexate, 25mg/ml (25,000ppm)>240>240Same
Mitomycin C, 0.5mg/ml (500ppm)>240>240Same
Mitoxantrone HCL, 2mg/ml (2,000ppm)/>240Different
Oxaliplatin, 5mg/ml (5,000ppm)/>240Different
Paclitaxel, 6mg/ml (6,000ppm)>240>240Same
Paraplatin, 10mg/ml (10,000ppm)/>240Different
Retrovir, 10mg/ml (10,000ppm)/>240Different
Rituximab, 10mg/ml (10,000ppm)/>240Different
Thiotepa, 10mg/ml (10,000ppm)97.146.3Similar
Topotecan, 1mg/ml (1,000ppm)/>240Different
Trisenox, 1mg/ml (1,000ppm)/>240Different
Velcade, 1mg/ml (1,000ppm)/>240Different
Vincristine Sulfate, 1mg/ml (1,000ppm)>240>240Same
Fentanyl Citrate Injection (100 mcg/2ml)>240>240Same

8 Summary of Non-Clinical Testing

Non-clinical tests were performed to verify that the subject device will meet the acceptance criteria of the performance test shown below:

Technological CharacteristicsStandard / Test /FDA GuidanceResult Summary
DimensionsASTM D6319-19Meets
Physical PropertiesASTM D6319-19Meets
Freedom from holesASTM D6319- 19ASTM D5151-19Meets
Residual PowderASTM D 6124-06(2017)ASTM D6319- 19Meets
Primary Skin Irritation TestISO 10993-10:2010Meets
Dermal Sensitization AssayISO 10993-10:2010Meets
Acute Systemic ToxicityISO 10993-11:2017Meets

{9}------------------------------------------------

Donggao Industrial Zone, Zanhuang, Hebei, 050000, China

Product: Sterile Nitrile Powder Free Examination Gloves Tested For Use With Chemotherapy Drugs And Fentanyl Citrate (Blue)

The subject device meets the applicable requirements for patient examination gloves with regards to dimension and sizes, physical properties, freedom from holes and powder Residuals as found in the following standards: ASTM D6139, ASTM D5151, ASTM D6124.The subject device passes biological reactivity testing for dermal sensitization, irritation and acute systemic toxicity with accordance ISO10993-10 and ISO 10993-11.

9. Conclusion:

The conclusions drawn from the nonclinical tests demonstrate that the Subject device is as safe, as effective, and performs as well as or better than the legally marketed predicated device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.