K Number
K220231

Validate with FDA (Live)

Device Name
GyroStim
Date Cleared
2022-04-27

(90 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GyroStim is intended to assist in the treatment of balance disorders and vestibular dysfunction in adult patients, age 18 and older, up to 400lb. It is intended to be used by a trained medical professional as prescribed by physician in a clinical environment.

Device Description

GyroStim is an AC powered computer-controlled multi-axial positional and rotational chair. The device consists of a patient chair mounted within a rotational frame assembly that is capable of rotating in the pitch and yaw axes. The rotational frames are automated with computer-controlled drive systems capable of executing preprogrammed positional and rotational motion profiles. The system provides clinicians with a safe and efficient means for administering a wide range of specific positional and rotational motions for inducing stimulation of the vestibular system.

GyroStim operational software runs on a pc-based desktop computer. The software can capture, monitor, and display the patient's subjective response to each run. The response data may be used by the clinician for clinical decision support (CDS) to assist with progress evaluation, for profile selection and rate of therapy advancement, for individualizing therapy sessions, and for monitoring and preventing overstimulation.

Patient data is collected and stored and may be accessed for graphing, analysis, and reports.

The software includes libraries of prewritten profiles. A library of pre-written positional profiles located on the RUN BPPV page contains a collection of profiles for executing commonly used canalith repositioning maneuvers. A library of pre-written motion profiles located on the RUN GyroStim page contains a collection of profiles for executing motion induced vestibular stimulation.

Specifically, the library of motion profiles contains 30 levels of motion induced vestibular stimulation intensity, ranging from Level 1 to Level 30. The lowest level of motion induced vestibular stimulation intensity is Level 1, which rotates at a rate of 1 revolution per minute (RPM) or 0.016667 Hz. The highest level of motion induced vestibular stimulation intensity is Level 30. which rotates at 30 RPM or 0.5Hz. The intensity of the 30 levels differs and increments linearly at a rate of 1 RPM or 0.01667Hz per level; i.e.. Level 1 rotates at 1 RPM or 0.01667 Hz, Level 2 rotates at 2 RPM or 0.03334 Hz, Level 3 rotates at 3 RPM or 0.05 Hz, and so on up to Level 30 which rotates at 30 RPM or 0.5 Hz. Additionally, each of the 30 Levels offer a selection of run durations at 15 seconds (s), 30s, 60s, and 180s.

When configured with the integrated patient transfer system, GyroStim provides accessibility to the large population of patients with moderate to severe mobility challenges, many of whom are limited or unable to engage in the existing standard of care for balance disorders and vestibular dysfunction.

The GyroStim assembly, operational software, and libraries of profiles provide clinicians with a safe, effective, and efficient medical device for assisting in the treatment of balance disorders and vestibular dvsfunction.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the GyroStim device, based on the provided text:

Acceptance Criteria and Device Performance

Item TestedTest MethodPurposeAcceptance CriteriaReported Device Performance
V4-GYROSTIM III-04A (Electrical Safety)ANSI/AAMI/ES 60601-1, Ed 3.1Electrical safetyComplies with ANSI/AAMI/ES 60601-1, Ed 3.1Pass
V4-GYROSTIM III-04A (EMC)ANSI/AAMI/IEC 60601-1-2, Ed 4EMCComplies with Class A, Professional Healthcare Environment of ANSI/AAMI/IEC 60601-1-2, Ed 4Pass
V4-GYROSTIM III-04, V4-GYROSTIM III-04A (Static Loading Safety Factors)Engineering Analysis per Clause 9.8.2 and Table 21 of IEC 60601-1, Ed 3.1Determination of static loading safety factorsSafety factor determined to be greater than the 2.5x requirements of Table 21 of IEC 60601-1Pass, Min. safety factor, 27.5x
V4-GYROSTIM III-04, V4-GYROSTIM III-04A (Floor Loading Distribution)Engineering AnalysisDetermination of floor loading distribution for installation into non-residential environmentsComplies with uniform and concentrated live load requirements per IBC, 2018*Pass
V4-GYROSTIM III-04, V4-GYROSTIM III-04A (Reliability and Fatigue)Engineering Analysis (including fatigue and dynamic loading, and retrospective data analysis)Reliability analysis, including fatigue and dynamic loading, and retrospective data analysisDevice shall remain reliable and not cause patient or end user injury for 10-year shelf lifePass
V4-GYROSTIM III-04, V4-GYROSTIM III-04A, GyroStim software (Software Verification)IEC 62304:2006/A1:2016, clause 5.7 and FDA's Guidance for Industry and FDA Staff, "Guidance for the Content for the Premarket Submissions for Software Contained in Medical Devices".Software verificationSoftware complies with pre-defined requirements of the software requirement specification for a Class B / "moderate" level of concern due to software failurePass
V4-GYROSTIM III-04, V4-GYROSTIM III-04A, GyroStim software (Accuracy of Controls)Measurement of rotational accuracy in pitch and yaw directionsAccuracy of ControlsSystem meets the defined accuracy requirements for RPM, acceleration, and degrees of rotationPass
V4-GYROSTIM III-04, V4-GYROSTIM III-04A (Summative Usability Study)IEC 62366-1, 2015 +A2020 and FDA's Guidance for Industry and FDA Staff, "Applying Human Factors and Usability Engineering to Medical Devices"Summative Usability StudyNo instances of use errors or close calls with potential for harmPass, none observed

Study Details

The provided document describes a preliminary clinical study and various non-clinical bench (performance) tests.

1. Clinical Study:

  • Sample size used for the test set and the data provenance:
    • The document states that the study "utilized retrospective data from the two treatment groups: GyroStim group, and SOC Group." However, it does not specify the exact sample size (number of patients) used in either group.
    • The data provenance is retrospective. The country of origin of the data is not specified.
  • Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    • The assessments and treatments were performed by a "vestibular and balance medical specialist (DPT)."
    • The document does not mention a number of experts for establishing ground truth or their specific qualifications beyond "DPT." It implies one such specialist conducted the assessments and treatments.
  • Adjudication method: Not described or applicable, as it was a retrospective comparison of treatment outcomes, not a diagnostic study requiring expert consensus on individual cases.
  • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    • No, this was not an MRMC study. The GyroStim is a physical device for therapy, not an AI for image interpretation or diagnosis by human readers.
  • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    • The document describes the performance of the device itself in both non-clinical and clinical contexts. The "algorithm" (software) of GyroStim is part of the device's control system, and its verification was done as a standalone test (software verification). However, the overall clinical study evaluated the device in use by a trained medical professional, so it's not a purely "algorithm-only" performance in the clinical sense.
  • The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
    • Outcomes data obtained from "pre and post intervention assessments" using "quantitative equipment and methods, including SOT, DHI, ABC, DVA, and mDGI." The "ground truth" for effectiveness was based on the improvement in these objective and subjective balance scores.
  • The sample size for the training set: Not applicable. The clinical study was a comparison of treatment outcomes using an existing device (GyroStim) against standard of care. There is no mention of a machine learning model being trained on a separate dataset for clinical decision-making within the scope of this FDA summary. The "RUN GyroStim profiles" are pre-written profiles within the software, not learned models from a training set in the typical AI sense.
  • How the ground truth for the training set was established: Not applicable, as there was no explicit training set described for an AI/ML model for clinical decision-making.

2. Non-Clinical Bench (Performance) Testing:

  • Sample size used for the test set and the data provenance: This refers to physical units of the device (V4-GYROSTIM III-04A, V4-GYROSTIM III-04, GyroStim software). The sample size implicitly refers to the device units tested to ensure compliance with the specified standards. The data provenance is from internal engineering analysis and testing.
  • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: These tests rely on established engineering standards and test methods (e.g., IEC, ANSI/AAMI). The "ground truth" is compliance with these standards, verified by qualified engineers or test facilities. The specific number or qualifications are not detailed beyond adherence to the standards.
  • Adjudication method: Not applicable for bench testing. Pass/fail criteria are defined by the standards.
  • MRMC study: Not applicable.
  • Standalone performance: Yes, these are standalone performance tests of the device and its software components against engineering and safety standards.
  • Type of ground truth: Defined engineering and safety standards (e.g., ANSI/AAMI/ES 60601-1, IEC 60601-1-2, IEC 62304, IBC 2018).
  • Sample size for training set: Not applicable.
  • How ground truth for training set was established: Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 27, 2022

UltraThera Technologies, Inc. Kevin Maher President 2 North Cascade Avenue, Suite 640 Colorado Springs, Colorado 80903

Re: K220231

Trade/Device Name: GyroStim Regulatory Class: Unclassified Product Code: LXV Dated: January 26, 2022 Received: January 27, 2022

Dear Kevin Maher:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220231

Device Name GyroStim

Indications for Use (Describe)

The GyroStim is intended to assist in the treatment of balance disorders and vestibular dysfunction in adult patients, age 18 and older, up to 4001b. It is intended to be used by a trained medical professional as prescribed by physician in a clinical environment.

Restricted Device (per 21 CFR 801.420 and CFR 801.421).

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Ultrathera Technologies, Inc. The logo features a red diamond shape with a star in the center and a curved line extending from the top left to the bottom right. To the right of the diamond is the word "ULTRATHERA" in gold letters, and below that is the phrase "TECHNOLOGIES, INC." in red letters.

510(k) Summary

1. Sponsor/ Applicant

UltraThera Technologies, Inc. 2 North Cascade Ave., Suite 640 Colorado Springs, Colorado 80903

Kevin Maher President Phone: (719) 685-7883 Fax: (719) 418-6916 Email: kmaher@ultrathera.com

Summary Preparation Date: January 26, 2022

2. Device

Trade NameGyroStim
Model NumbersV4-GYROSTIM III-04 - GyroStim G 1 - Stationary canopy, no laser optionV4-GYROSTIM III-04 - GyroStim G 3 - Stationary canopy with laser targetoption
Model NumbersV4-GYROSTIM III-04A - GyroStim G 3X - Sliding canopy with laser targetoption
ClassificationUnclassified (pre-amendment)
Product CodeLXV
510(k) NumberK220231
Review PanelEar, Nose, and Throat (ENT) Devices

3. Predicate Device

Vesticon Epley Omniax™, 510(k) # K071973

Device Description 4.

GyroStim is an AC powered computer-controlled multi-axial positional and rotational chair. The device consists of a patient chair mounted within a rotational frame assembly that is capable of rotating in the pitch and yaw axes. The rotational frames are automated with computer-controlled drive systems capable of executing preprogrammed positional and rotational motion profiles. The system provides clinicians with a safe and efficient means for administering a wide range of specific positional and rotational motions for inducing stimulation of the vestibular system.

GyroStim operational software runs on a pc-based desktop computer. The software can capture, monitor, and display the patient's subjective response to each run. The response data may be used by the clinician for clinical decision support (CDS) to assist with progress evaluation, for profile selection and rate of therapy advancement, for individualizing therapy sessions, and for monitoring and preventing overstimulation.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Ultrathera Technologies, Inc. The logo features a red diamond shape with a gold line running through it on the left side. To the right of the diamond is the word "ULTRATHERA" in gold, and below that is the phrase "TECHNOLOGIES, INC." in red.

Patient data is collected and stored and may be accessed for graphing, analysis, and reports.

The software includes libraries of prewritten profiles. A library of pre-written positional profiles located on the RUN BPPV page contains a collection of profiles for executing commonly used canalith repositioning maneuvers. A library of pre-written motion profiles located on the RUN GyroStim page contains a collection of profiles for executing motion induced vestibular stimulation.

Specifically, the library of motion profiles contains 30 levels of motion induced vestibular stimulation intensity, ranging from Level 1 to Level 30. The lowest level of motion induced vestibular stimulation intensity is Level 1, which rotates at a rate of 1 revolution per minute (RPM) or 0.016667 Hz. The highest level of motion induced vestibular stimulation intensity is Level 30. which rotates at 30 RPM or 0.5Hz. The intensity of the 30 levels differs and increments linearly at a rate of 1 RPM or 0.01667Hz per level; i.e.. Level 1 rotates at 1 RPM or 0.01667 Hz, Level 2 rotates at 2 RPM or 0.03334 Hz, Level 3 rotates at 3 RPM or 0.05 Hz, and so on up to Level 30 which rotates at 30 RPM or 0.5 Hz. Additionally, each of the 30 Levels offer a selection of run durations at 15 seconds (s), 30s, 60s, and 180s.

When configured with the integrated patient transfer system, GyroStim provides accessibility to the large population of patients with moderate to severe mobility challenges, many of whom are limited or unable to engage in the existing standard of care for balance disorders and vestibular dysfunction.

The GyroStim assembly, operational software, and libraries of profiles provide clinicians with a safe, effective, and efficient medical device for assisting in the treatment of balance disorders and vestibular dvsfunction.

5. Indications for Use

The GvroStim is intended to assist in the treatment of balance disorders and vestibular dysfunction in adult patients, age 18 and older, up to 400lb. It is intended to be used by a trained medical professional as prescribed by physician in a clinical environment.

6. Technological Characteristics and Substantial Equivalence

The GyroStim is substantially equivalent to the Epley Omniax regarding the intended use and technological characteristics, as explained in following table and discussion:

FeatureSubject DeviceUltraThera GyroStimPredicate DeviceVesticon EpleyOmniaxComparisonof Devices
510(k) NumberK220231K071973N/A
Intended UseThe GyroStim is intended toassist in the treatment ofbalance disorders andvestibular dysfunction.The Omniax is intended toassist in the diagnosis andtreatment of balancedisorders and vertigo,including benignparoxysmal positionalvertigoComparable
Target patientpopulationAdults, age 18 and older,maximum 400lbAdults, age 18 and older,maximum 350lbComparable
Use environmentProfessional healthcare /Professional healthcare /Same
FeatureSubject DeviceUltraThera GyroStimPredicate DeviceVesticon EpleyOmniaxComparisonof Devices
clinical environmentclinical environment
Delivery of TherapyWhole body positioning androtationWhole body positioningand rotationComparable
Library of ProfilesLibraries of positionalmaneuvers and rotationalprofilesLibrary of positionalmaneuversComparable
Intended UserTrained medical personnelTrained medical personnelSame
Frequency of useAs prescribed by physicianAs prescribed by physicianSame
Prescription and/orover-the-counter useBy prescription onlyBy prescription onlySame
Operator InterfaceComputer with operationalsoftware to control therotational and positionalmotion of the chairComputer with operationalsoftware to control therotational and positionalmotion of the chairSame
Positional andRotational MeansElectro-mechanical motionchairElectro-mechanical motionchairSame
Axis of Motion &Rotational DegreesPitch: 360°Yaw: 360°Pitch, 360°Yaw, 360°Roll, 360°Comparable
MaximumAcceleration180 deg/s2(Information is redacted)Unknown
Maximum Velocity180 deg/s(30 RPM)(Information is redacted)Unknown
Collects and reportspatient response tovestibular stimulationYesYesSame
Patient RestraintSystemFive-point harness andinterlocked lap and anklerestraintsTwo parallel over-shoulderstraps, lap beltComparable
Electrical SafetyPerformanceComplies withANSI/AAMI/ES 60601-1,Ed 3.1 and applicablecollateral standardsComplies with IEC 60601-1Same
EMCComplies withANSI/AAMI/IEC 60601-1-2,Ed 4, Class A, professionalhealthcare environmentComplies with IEC 60601-1-2Same

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Ultrathera Technologies, Inc. The logo consists of a red diamond shape with a stylized star inside, followed by the company name in gold lettering. Below the company name, the words "TECHNOLOGIES, INC." are written in red.

Non-clinical Bench (Performance) testing 7.

The performance of the GyroStim is verified based on the successful completion of following tests:

Item TestedTest MethodPurposeAcceptance CriteriaResults
V4-GYROSTIM III-04AANSI/AAMI/ES60601-1, Ed 3.1Electrical safetyComplies withANSI/AAMI/ES 60601-1,Ed 3.1Pass
V4-GYROSTIM III-04AANSI/AAMI/IEC60601-1-2, Ed 4EMCComplies with Class A,Professional HealthcareEnvironment ofANSI/AAMI/IEC 60601-Pass
Item TestedTest MethodPurposeAcceptance CriteriaResults
V4-GYROSTIM III-04,V4-GYROSTIM III-04AEngineeringAnalysis per Clause9.8.2 and Table 21of IEC 60601-1, Ed3.1Determination of staticloading safety factors1-2, Ed 4Safety factor determinedto be greater than the2.5x requirements ofTable 21 of IEC 60601-1Pass,Min.safetyfactor,27.5x
V4-GYROSTIM III-04,V4-GYROSTIM III-04AEngineeringAnalysisDetermination of floorloading distribution forinstallation into non-residentialenvironmentsComplies with uniformand concentrated liveload requirements perIBC, 2018*Pass
V4-GYROSTIM III-04,V4-GYROSTIM III-04AEngineeringAnalysisReliability analysis,including fatigue anddynamic loading, andretrospective dataanalysisDevice shall remainreliable and not causepatient or end user injuryfor 10-year shelf lifePass
V4-GYROSTIM III-04,V4-GYROSTIM III-04A,GyroStim softwareIEC 62304:2006/A1:2016,clause 5.7 andFDA's Guidance forIndustry and FDAStaff, "Guidance forthe Content for thePremarketSubmissions forSoftware Containedin Medical Devices".Software verificationSoftware complies withpre-definedrequirements of thesoftware requirementspecification for a ClassB / "moderate" level ofconcern due to softwarefailurePass
V4-GYROSTIM III-04,V4-GYROSTIM III-04A,GyroStim softwareMeasurement ofrotational accuracyin pitch and yawdirectionsAccuracy of ControlsSystem meets thedefined accuracyrequirements for RPM,acceleration, anddegrees of rotationPass
V4-GYROSTIM III-04,V4-GYROSTIM III-04A,IEC 62366-1, 2015+A2020 and FDA'sGuidance forIndustry and FDAStaff, "ApplyingHuman Factors andUsabilityEngineering toMedical Devices"Summative UsabilityStudyNo instances of useerrors or close calls withpotential for harmPass,noneobserved

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for Ultrathera Technologies, Inc. The logo features a red diamond shape with a stylized star inside on the left. To the right of the diamond is the word "ULTRATHERA" in large, gold, sans-serif font. Below "ULTRATHERA" is the text "TECHNOLOGIES, INC." in a smaller, red, sans-serif font.

8. Biocompatibility Testing

The biocompatibility evaluation for the GyroStim device was conducted in accordance with the FDA's Guidance "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" September 2020, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA.

Overall, it can be concluded that the GyroStim materials in limited, direct, or transient patient contact with intact skin pose a very low biocompatibility risk based upon materials characterization, common use in healthcare and commercial applications, and history of safe use.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Ultrathera Technologies, Inc. The logo features a red diamond shape with a white star in the center and two yellow lines curving around the star. To the right of the diamond is the word "ULTRATHERA" in yellow, and below that is the phrase "TECHNOLOGIES, INC." in red.

Clinical Testing 9.

A preliminary clinical study was conducted to compare the outcomes of the RUN GyroStim profiles to outcomes of the current standard of care (SOC) for treatment of balance disorders and vestibular dysfunction. The study utilized retrospective data from the two treatment groups: GyroStim group, and SOC Group.

The clinical comparison included data from pre and post intervention assessments from both patient groups. The assessments and treatments were performed by a vestibular and balance medical specialist (DPT) and with guantitative equipment and methods, including SOT, DHI, ABC, DVA, and mDGI. Both groups received the same intake, balance education, and training for at-home daily exercise routines. The only variable between the two groups was the treatment: treatment with GyroStim, or treatment with SOC. Assessment data from multiple vestibular/balance disorders were collected retrospectively and presented for both groups.

Data comparison was conducted for both treatment groups to compare the following:

  • difference in pre- and post- objective or subjective balance scores .
  • duration of treatment time between pre- and post- therapy balance scores .

The analysis of retrospective clinical test data from patients with vestibular and balance disorder indicated that the RUN GyroStim profiles provided treatment that was as effective as the current SOC.

The retrospective data also indicated that treatment with RUN GyroStim profiles provided a statistically relevant (p<0.0001) average balance score improvement per week of treatment when compared to SOC. Overall, the preliminary clinical study indicated that treatment with GyroStim provided substantially equivalent treatment of vestibular/balance disorders as the current standard of care.

10. Conclusion:

The subject device, GyroStim, has a substantially equivalent intended use and technological characteristics as the cited predicate device. The minor changes or differences presented do not raise new questions of safety or effectiveness. The nonclinical test results have demonstrated the subject device is as safe and can perform with substantial equivalence as the predicate device. The preliminary clinical study provides evidence that the subject device is as effective as the current standard of care for balance disorders and vestibular dysfunction.

Therefore, it is concluded that no new questions of safety and effectiveness were raised. We conclude that the subject device is substantially equivalent to the cited predicate device.

N/A