K Number
K220155

Validate with FDA (Live)

Date Cleared
2022-05-23

(124 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Gloves have been tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

Device Description

Powder-free Nitrile Examination Gloves Tested for use with Chemotherapy Drugs

AI/ML Overview

The provided document describes the FDA clearance for "Powder-free Nitrile Examination Gloves Tested for use with Chemotherapy Drugs" (K220155). This is a medical device, specifically examination gloves, and not an AI/ML powered device. Therefore, the questions related to AI/ML device performance, such as sample sizes for test and training sets, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, and ground truth establishment for training, are not applicable to this submission.

However, the document does contain information about the acceptance criteria and the study that proves the device meets those criteria for its intended use with chemotherapy drugs.

Here's an analysis based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Chemotherapy Drug Permeation)Reported Device Performance (Breakthrough Time)
>240 minutes for Bleomycin Sulfate, Bortezomib, Busulfan, Carboplatin, Chloroquine, Cisplatin, Cyclophosphamide, Cytarabine, Dacarbazine, Daunorubicin HCl, Docetaxel, Etoposide, Fluorouracil, Gemcitabine, Idarubicin HCl, Ifosfamide, Irinotecan, Mechlorethamine HCl, Melphalan, Methotrexate, Mitomycin C, Mitoxantrone, Paclitaxel, Triclosan, Trisenox, Vincristine Sulfate.>240 minutes for the above-listed drugs.
Lower permeation times for specifically identified drugs.45.4 minutes for Carmustine (BCNU)
15.6 minutes for ThioTepa

Note: The document explicitly states "WARNING: Not recommended for use with Carmustine and ThioTepa" due to their low permeation times. This indicates that while the testing was performed, the acceptance criteria for these specific drugs effectively resulted in a caveat for their use, highlighting that the "accepted" performance for other drugs is >240 minutes.

2. Sample size used for the test set and the data provenance

The document does not specify the sample size used for the test set (number of gloves or repetitions of testing). It also does not provide information on data provenance (e.g., country of origin, retrospective or prospective). This type of detail is typically found in the full study report, which is not included in this FDA clearance letter.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This is a physical device (gloves) and its performance against chemical permeation, not an AI/ML diagnostic or prognostic tool requiring expert ground truth for interpretation. The "ground truth" is the chemical permeation itself, measured objectively.

4. Adjudication method for the test set

Not applicable. As this is not an AI/ML device, there's no "adjudication" in the sense of reconciling expert opinions on diagnoses or interpretations. The permeation testing follows a standardized protocol (ASTM D6978-05).

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI/ML device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an AI/ML device.

7. The type of ground truth used

The "ground truth" for this device's performance regarding chemotherapy drug permeation is established through objective laboratory testing following a recognized standard. Specifically, it used ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs. This standard defines the methodology to measure the breakthrough time of various chemotherapy drugs through the glove material.

8. The sample size for the training set

Not applicable. This is not an AI/ML device, so there is no training set in the context of an algorithm.

9. How the ground truth for the training set was established

Not applicable. As this is not an AI/ML device, there is no training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 23, 2022

Motex Healthcare (Anhui) Co., Ltd. % Jen Ke-Min Official Correspondent Roc Chinese-European Industrial Research Society No. 58, Fu Chiun Street Hsin Chu City, Taiwan 300113 Taiwan

Re: K220155

Trade/Device Name: Powder-free Nitrile Examination Gloves Tested for use with Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Non-powdered patient examination glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: April 6, 2022 Received: April 13, 2022

Dear Jen Ke-Min:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bifeng Qian, M.D., Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K220155

Device Name

Powder-free Nitrile Examination Gloves Tested for use with Chemotherapy Drugs

Indications for Use (Describe)

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Gloves have been tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

Chemotherapy Drug Permeation:

Chemotherapy Drugs Tested

Bleomycin Sulfate, 15 mg/ml (15,000 ppm) >240 minutes Bortezomib (Velcade), 1 mg/ml (1,000ppm) >240 minutes Busulfan, 6 mg/ml (6,000 ppm) >240 minutes Carboplatin, 10 mg/ml (10,000 ppm) Carmustine (BCNU), 3.3 mg/ml (3,300 ppm) Chloroquine, 50 mg/ml (50,000 ppm) Cisplatin, 1.0 mg/ml (1,000 ppm) Cyclophosphamide (Cytoxan), 20.0 mg/ml (20,000 ppm) Cytarabine, 100 mg/ml (100,000 ppm) Dacarbazine, 10.0 mg/ml (10,000 ppm) Daunorubicin HCI, 5 mg/ml (5,000 ppm) Docetaxel, 10 mg/ml (10,000 ppm) Etoposide, 20.0 mg/ml (20,000 ppm) Fluorouracil, 50.0 mg/ml (50,000 ppm) Gemcitabine, 38 mg/ml (38,000 ppm) Idarubicin HCI, 1 mg/ml (1,000 ppm) Ifosfamide, 50 mg/ml (50,000 ppm) Irinotecan, 20 mg/ml (20.000 ppm) Mechlorethamine HCI, 1 mg/ml (1,000 ppm) Melphalan, 5 mg/ml (5,000 ppm) Methotrexate, 25 mg/ml (25,000ppm) Mitomycin C, 0.5 mg/ml (500 ppm) Mitoxantrone, 2 mg/ml (2,000 ppm) Paclitaxel, 6.0 mg/ml (6.000 ppm) ThioTepa, 10.0 mg/ml (10,000 ppm) Triclosan, 2 mg/ml (2,000 ppm) Trisenox, 1 mg/ml (1,000 ppm) Vincristine Sulfate, 1.0 mg/ml (1,000 ppm)

*WARNING: Not recommended for use with Carmustine and ThioTepa.

The maximum testing time is 240 minutes. Please note that the following drugs have low permeation times: 45.4 minutes

Carmustine (BCNU), 3.3 mg/ml (3,300 ppm) ThioTepa, 10.0 mg/ml (10,000 ppm)

FORM FDA 3881 (6/20)

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

240 minutes 45.4 minutes* >240 minutes >240 minutes >240 minutes >240 minutes >240 minutes >240 minutes >240 minutes >240 minutes >240 minutes >240 minutes >240 minutes >240 minutes >240 minutes >240 minutes >240 minutes >240 minutes >240 minutes >240 minutes >240 minutes 15.6 minutes* >240 minutes >240 minutes >240 minutes

15.6 minutes

Breakthrough Time

PSC Publishing Services (301) 443-674

{3}------------------------------------------------

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

/ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20)

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.